key: cord-1047574-xt3uv53d authors: Genoud, V.; Stortz, M.; Waisman, A.; Berardino, B. G.; Verneri, P.; Dansey, V.; Salvatori, M.; Remes Lenicov, F.; Levi, V. title: Extraction-free protocol combining Proteinase K and heat inactivation for detection of SARS-CoV-2 by RT-qPCR date: 2020-12-17 journal: nan DOI: 10.1101/2020.12.16.20248350 sha: c435c1710782cbd30ef17d9de71d6476e76fc2a3 doc_id: 1047574 cord_uid: xt3uv53d Real-time reverse transcription PCR (RT-qPCR) is the gold-standard technique for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) detection in nasopharyngeal swabs specimens. The analysis by RT-qPCR usually requires a previous extraction step to obtain the purified viral RNA. Unfortunately, RNA extraction constitutes a bottleneck for early detection in many countries since it is expensive, time-consuming and depends on the availability of commercial kits. Here, we describe an extraction-free protocol for SARS-CoV-2 detection by RT-qPCR from nasopharyngeal swab clinical samples in saline solution. The method includes a treatment with proteinase K followed by heat inactivation (PK+HID method). We demonstrate that PK+HID improves the RT-qPCR performance in comparison to the heat-inactivation procedure. Moreover, we show that this extraction-free protocol can be combined with a variety of multiplexing RT-qPCR kits. The method combined with a multiplexing detection kit targeting N and ORF1ab viral genes showed a sensitivity of 0.99 and a specificity of 0.99 from the analysis of 106 positive and 106 negative clinical samples. In conclusion, PK+HID is a robust, fast and inexpensive procedure for extraction-free RT-qPCR determinations of SARS-CoV-2. The outbreak of the coronavirus disease 2019 , caused by SARS-CoV-2, was declared a pandemic by World Health Organization on March 11, 2020 . Confronted by the lack of a vaccine or specific treatment, early detection, isolation and contact tracing are fundamental strategies for reducing the spread of this disease [1] . RT-qPCR is the gold standard method for SARS-CoV-2 determinations [2] . These assays typically involve collecting the clinical specimen usually with nasopharyngeal swabs, extracting the RNA from the sample and analyzing the presence of the viral RNA through RT-qPCR. Nowadays, the RNA extraction step is considered a bottleneck of SARS-CoV-2 RT-qPCR determinations [3] . This procedure can be automatized but commercial extraction robots are very expensive and may not be affordable for most clinical labs in low-income countries. In contrast, manual RNA-extraction is a quite cumbersome procedure that includes several washing and centrifuging steps at a Biosafety level 2 laboratory and relatively expensive extraction kits. In Argentina, a trained lab technician processes approximately 24 samples every 1.5 hours. Additionally, the skyrocketing increment on the clinical use of RT-qPCR due to the COVID-19 pandemic led to a worldwide shortage of RNA purification reagents [4, 5] thus, many research groups around the world have concentrated their efforts on either simplifying or eliminating the extraction step [6] [7] [8] [9] [10] [11] [12] [13] [14] [15] [16] . Björn Reinius at the Karolinska Institute in Sweden [17] optimized an extraction-free sample preparation protocol that only includes a very simple heat-inactivation step at 95°C during 5 min (HID method). Through HID RT-qPCR, they were able to detect with high sensitivity the relatively short N1 and N2 regions of the viral RNA which are preserved after the heating step. This group also performed an exhaustive analysis of media formulation compatible with the HID method [17] . All rights reserved. No reuse allowed without permission. preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this this version posted https://doi.org/10.1101/2020.12. 16.20248350 doi: medRxiv preprint More recently, the Centers for Disease Control and Prevention (CDC) has approved a similar thermal-inactivation protocol (1 min at 95°C) that can be used with Quantabio UltraPlex 1-Step ToughMix (4X) and is sufficient for viral inactivation [18] . This protocol also detects N1 and N2 sequences of the viral RNA. Many protocols traditionally used for extraction-free preparation of clinical samples included treatments with proteinase-K (PK), a commonly used protease that degrade RNases and hence, it helps to preserve RNA integrity [19, 20] . A recent work suggested that PK may facilitate extractionfree determinations of SARS-CoV-2 although a major loss in sensitivity with a 6-units shift in CT values on the determination of the envelope (E) gene [21] may prevent its uses in clinical analysis. Other recent preprint [22] tested a protocol based on PK and heat inactivation using a reduced number of samples (N=17) in UTM medium with relative low CT values, and claim a good sensitivity on the determination of the E gene. Although these results might seem contradictory with recent findings showing suboptimal detection of the E amplicon on HID samples in UTM medium [17] , the efficiency of the PCR is very dependent on the detection kit, transport medium brand, thermocycler, etc. Here, we combined these strategies and optimized the PK+HID method, a very simple protocol for extraction-free RT-qPCR determinations of SARS-CoV-2. Our results position PK+HID as an inexpensive, versatile and reliable alternative to bypass the traditional RNA extraction procedure and suggest that it could substantially contribute to improve the testing capacity in countries with limited resources and facilities. Sample collection. Nasopharyngeal swabs were collected during screening of non-hospitalized symptomatic cases and close contacts to these cases at diverse hospital and clinical centers at the Buenos Aires area (AMBA, Área Metropolitana de Buenos Aires), Argentina. Swabs were deposited in 2-5 ml of saline solution. These clinical samples were processed at the Instituto de Investigaciones Biomédicas en Retrovirus y SIDA (INBIRS, Buenos Aires, Argentina), a specialized center currently dedicated to SARS-CoV-2 diagnosis by RT-qPCR. For routine diagnosis, RNA extraction was performed from 300 µl of swab samples using a Chemagic 360-D (Perkin-Elmer), an automated extraction equipment, according to the manufacturer's instructions. This procedure does not include an RNA-concentration step. All rights reserved. No reuse allowed without permission. preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. For experiments with variable PK concentrations, 10 µl of PK stock solutions with the adequate concentration were added. Samples were incubated at 55°C for 15 min and then at 98°C for 5 min in a thermal cycler with heated lid. Finally, inactivated samples were cooled to 4°C and kept at this temperature until RT-qPCR analysis. All rights reserved. No reuse allowed without permission. preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. GeneFinder detection kits were analyzed with the CFX Manager software (BioRad) to obtain CT values. Data obtained with 2019-nCoV CDC were analyzed with the free software LinRegPCR [23] to obtain CT values. This software was also used to calculate PCR efficiencies and the initial amount of template copies (in arbitrary units). All rights reserved. No reuse allowed without permission. preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this this version posted December 17, 2020. ; https://doi.org/10.1101/2020.12.16.20248350 doi: medRxiv preprint We used a reduced number of SARS-CoV-2 positive (N=6) and negative (N=3) nasopharyngeal swab samples collected in saline solution to test if their preincubation with PK before the thermal inactivation step improves the performance of the HID extraction-free method. Although previous results [17] suggested that saline solution is suboptimal for direct RT-qPCR detection, this is one of the transport medium recommended by CDC and it is also commonly used to collect clinical samples in Argentina. In these experiments, we mixed up 90 μl of the samples with 10 μl of PK 10 mg/ml (final concentration 1 mg/ml) in 0.2 ml PCR tubes and incubated them in a thermocycler during 15 min at 55°C followed by 5 min at 98°C (PK+HID samples). As controls, we used heat-inactivated aliquots omitting the preincubation with PK at 55°C (HID samples) and RNA samples obtained through a standard extraction protocol (purified RNA samples) from the same nasopharyngeal swabs. These samples were analyzed by RT-qPCR using the 2019-nCoV CDC Diagnostic Panel which detects the N1 and N2 viral regions and the human RNase P (RP) gene as internal control. Table) . All rights reserved. No reuse allowed without permission. preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this this version posted December 17, 2020. ; https://doi.org/10.1101/2020.12.16.20248350 doi: medRxiv preprint All rights reserved. No reuse allowed without permission. preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this this version posted December 17, 2020. ; https://doi.org/10.1101/2020.12.16.20248350 doi: medRxiv preprint 8 The analysis of the amplification curves shows that the PCR efficiency (EPCR) was the lowest for the HID treatment followed by PK+HID and then by the RNA extraction treatment (Fig 1c) . This result suggests that the PK treatment may contribute to degrade unknown components of the sample matrix that interfere with RT-qPCR. Additionally, we observed a higher PCR amplification efficiency for the N1 viral gene compared to N2, in line with previous results [17, 24] . We also analyzed the amount of amplicon copies in PK+HID relative to that determined in HID samples and observed that this ratio was higher than 1 for N2 and RP suggesting that PK treatment also protected these RNA regions from RNase´s action (Fig 1d) . PK+HID involves the addition of the PK buffer and thus a 10 % dilution of the saline medium used to collect nasopharyngeal swabs. Therefore, we next evaluate if the improvement of PK+HID treatment compared to heat-inactivation is only due to the dilution of the saline medium. We analyzed five positive samples processed by the standard PK+HID protocol or by HID but adding to the samples the same amount of PK buffer without PK (HID´ samples). S2 Fig shows that CT values for PK+HID samples were lower compared to HID´ samples; the analyses of amplification curves indicate that nPK+HD was higher than nHID´ supporting that PK contributes with the preservation of the integrity of the viral RNA. Therefore, the dilution of saline solution could not completely explain the better performance of PK+HID. Nevertheless, the efficiency of PCR amplification for HID´ samples was slightly lower only for N2 amplicon suggesting that the dilution of saline solution in PK+HID samples also contributes to increase the efficiency of PCR. We next tested if using a higher volume of PK+HID samples could improve SARS-CoV-2 detection in the RT-qPCR determinations. Although the sensitivity of the assay is expected to be higher with more copies of the viral RNA template, a previous report also suggested that increasing too much this volume may have the opposite effect due to RT-qPCR inhibitors in the sample matrix [17] . We observed that the amplification curves shifted to lower cycles when using instead of -qPCR determination resulting in lower CT values for the N1 region (0.3 to 1 units, S2 Table) . This result suggests that a higher sample volume could contribute to the detection of SARS-CoV-2 in samples with low viral titers. Finally, we tested if the performance of the PK+HID method depends on PK concentration in the range of 0.1 to 1 mg/ml. Fig 1d shows that CT values obtained for the viral and human amplicons did not change within this PK concentration range. These values were also lower than those observed in the absence of PK, confirming the improving role of PK in this extraction-free method. All rights reserved. No reuse allowed without permission. preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this this version posted December 17, 2020. ; https://doi.org/10.1101/2020.12.16.20248350 doi: medRxiv preprint We next ran RT-qPCR determinations of SARS-CoV-2 in PK+HID and purified RNA samples prepared from positive and negative nasopharyngeal swab samples. Based on the results showed in the previous section, we used 1 mg/ml PK for the PK-(i.e. between the volumes assayed in the previous section) in the following RT-qPCR assays. For this particular experiment, we only analyzed the N1 viral gene and the RP human internal control to maximize the number of samples loaded in the PCR multiwell plate. In these experiments, we correctly identified all positive (27/27) and negative (6/6) samples. Fig 2 also shows that the CT values determined for the viral sequence N1 in PK+HID positive samples strongly correlated with those values measured in purified RNA samples supporting that the PK+HID treatment robustly preserves the information on the viral loads in the swab specimens. All rights reserved. No reuse allowed without permission. preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this this version posted December 17, 2020. ; https://doi.org/10.1101/2020.12.16.20248350 doi: medRxiv preprint In the previous sections, we showed that the PK+HID method resulted in an efficient alternative to bypass RNA extraction when detecting the N1 and N2 regions of the viral RNA. However, most laboratories use PCR kits that target other viral genes often combined in multiplex setups. Therefore, we next evaluated if the improved PK+HID method could be used with multiplexing detections kits. We first used DisCoVery (Safecare Biotech Hangzhou), a SARS-CoV-2 multiplex RT-qPCR detection kit that targets the SARS-CoV-2 genes N and ORF1ab and human RNase P gene. Relevantly, many worldwide providers offer detection kits targeting one or both of these genes. All rights reserved. No reuse allowed without permission. preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. All rights reserved. No reuse allowed without permission. preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. We correctly identified 44/46 positive and 3/3 negative samples. S2 Table compiles the CT values of the false-negative samples (see Discussion) . Additionally, the internal control (IC) of two other PK+HID samples was not detected making mandatory repeating the assay; these invalid samples were not included in S3 Table. We determined the amplification efficiency of the viral regions N and ORF1ab and the human gene RP and found that these values were similar for PK+HID and purified RNA samples (Fig 3b) . We also compared the amplification curves obtained using this detection kit and swab samples that were either treated by HID or PK+HID or subjected to the RNA extraction procedure (S3 Fig). PK treatment lowered the CT values in comparison to those obtained with HID only, with a mean + SEM difference in CT values of 2.7 + 0.7 (N), 3.1 + 1.0 (ORF1ab) and 1.4 + 0.2 (RP). These results, obtained with a different detection kit, agree with those showed in Fig 1 also supporting that the PK treatment improves the performance of the heat-inactivation method. The work of Reinius and colleagues [17] confirmed the virus inactivation after HID treatment. As a control, we also run infectivity assays as described in A common practice in many clinical labs involves freezing/thawing reagents several times. Therefore, we tested if repeated cycles of freezing/thawing of the PK stock solution may affect the performance of PK+HID. To address this issue, we processed 10 positive samples by PK+HID using a PK solution that was previously subjected to 10 cycles of freezing/thawing or a PK solution thawed only once. The mean differences in CT values between these samples were 0.17 + 0.15 (N), -0.05 + 0.20 (ORF1ab) and 0.01 + 0.14 (RP) (Ndata = 10) showing that the PK stock solution is stable for up to 10 cycles of freezing/thawing. To analyze if the PK+HID method can be combined with other multiplexing kits, we prepared PK+HD and purified RNA samples from 94 randomly-selected clinical specimens and ran RT-qPCR All rights reserved. No reuse allowed without permission. preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this this version posted December 17, 2020. ; https://doi.org/10.1101/2020.12.16.20248350 doi: medRxiv preprint 13 determinations using the GeneFinder detection kit which amplifies N, E and RdRp regions of the viral RNA and the human RP gene as internal control. The analysis of CT values measured in PK+HID and RNA samples showed stronger correlations for the N amplicon in comparison to E and RdRp amplicons (Fig 4a) . Additionally, some of the Npositive RNA samples (also considered SARS-CoV-2 positive in clinical diagnosis [24] ) showed negative results for the E and RdRp amplicons (3/38 and 4/38, respectively); this difference increased in PK+HID samples with 8 more samples showing negative results for both amplicons while being N-positive. As mentioned above, other authors also reported a poor detection and large right-shifts in CT values for E amplicon in extraction-free samples [17] . All rights reserved. No reuse allowed without permission. preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this this version posted December 17, 2020. ; https://doi.org/10.1101/2020.12.16.20248350 doi: medRxiv preprint genes in randomly-selected clinical specimens. Nasopharyngeal swab samples were processed for RNA extraction (purified RNA) or subjected to treatment with PK 1 mg/ml (55 o C for 15 min) followed by heat- We also observed that the amplification efficiency of viral amplicons N, E and RdRp was similar for PK+HID and purified RNA samples whereas EPCR of the internal control RP was slightly lower for PK+HID samples (Fig 4b) . These results support that the PK+HID treatment contributes to inactivate inhibitors of the PCR reaction. To quantitatively analyze the performance of the PK+HID treatment combined with the use of GeneFinder kit, we estimated the accuracy, sensitivity and specificity of detection as previously described [25] obtaining values of 0.95, 0.92 ± 0.05 and 0.96 ± 0.04, respectively. In the validation assay described above, swab samples were first analyzed by the usual RT-qPCR procedure including the RNA extraction step; 10 to 24 h later, we processed the same nasopharyngeal swabs by PK+HID. Therefore, we asked if the extra-time that swab samples remain at 4 C could lead to a partial degradation of viral RNA and consequently, to an apparently lower sensitivity of the PK+HID method. To test this hypothesis, we performed the validation experiment running simultaneously the RNA extraction and PK-HID procedures using 106 positive and 106 negative randomly-selected samples and analyzed the RNA and PK+HID samples by RT-qPCR with the detection kit DisCoVery (S5 Fig); unfortunately, we could not run these experiments with the GeneFinder detection kit used above because the provider run out of stock during the pandemic outbreak. The analyses of the data showed that the accuracy was 0,990 and the sensitivity and All rights reserved. No reuse allowed without permission. preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this this version posted December 17, 2020. ; https://doi.org/10.1101/2020.12.16.20248350 doi: medRxiv preprint specificity were both 0.99 ± 0.01. These results strongly support that PK+HID is a reliable protocol for extraction-free RT-qPCR SARS-CoV-2 determinations. In this last validation assay, we also analyzed the amplification curves of positive samples with with this kit. We could detect 29/29 and 5/6 positive and negative samples, respectively suggesting that PK+HID could be combined with these type of detection kits. The gold standard method for detection of SARS-CoV-2 in nasopharyngeal swab samples is RT-qPCR, a technique with exquisite sensitivity and high specificity commonly applied to analyze purified RNA samples. The RNA extraction step has nowadays become the bottleneck in COVID-19 testing, especially during pandemic outbreaks when personnel, resources and biosafety compatible facilities to perform the testing are scarce. The fast evolution of COVID-19 pandemic and the necessity of performing massive and rapid RT-qPCR determinations to test, trace and isolate SARS-CoV-2 positive patients encouraged the scientific community to evaluate alternative methods to bypass the laborious RNA extraction step. We mentioned above heat inactivation methods that make use of a fast thermal-treatment of the samples and allow the identification of SARS-CoV-2 positive samples targeting N1 and N2 viral genes [17] . These results are consistent with other works that also tested the performance of heatinactivated respiratory samples in RT-qPCR assays [8, 11, [26] [27] [28] [29] . All rights reserved. No reuse allowed without permission. preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this this version posted December 17, 2020. ; https://doi.org /10.1101 /10. /2020 Here, we combined heat inactivation with a pretreatment with PK, protease routinely used in nucleic acid preparations and thus widely available in most clinical laboratories. We mentioned previous works using PK to treat nasopharyngeal swab samples; the protease is also starting to be used in the analysis of saliva samples for the detection of SARS-CoV-2 in extraction-free RT-qPCR determinations [21, 22, 30, 31] . In line with these reports, our data shows that preincubation with PK improves the performance of the heat-inactivation protocol in RT-qPCR assays targeting N1 and N2 viral genes with the 2019-nCoV CDC Diagnostic Panel. However, a validation with a larger number of samples should be done to compare the sensitivities of both methods. We verified that the performance of the PK+HID does not depend on PK concentration in the range of 0.1 -1 mg/ml; this robustness of the PK+HID method is appealing since small variations in the PK concentration will not significantly affect the results of SARS-CoV-2 determinations and also allows reducing even more the costs of sample preparation for RT-qPCR assays. Using the 2019-nCoV CDC Diagnostic Panel, we could correctly identify the SARS-CoV-2 positive and negative analyzed samples and, for the positive samples, we obtained CT values in PK+HID samples very close to those of purified RNAs (2.7 units for N1) whereas heat inactivation without PK treatment produces higher shifts in CTs (5 units for N1). Remarkably, we demonstrated that the PK+HID method could be used in combination with other RT-qPCR kits that target different viral genes in multiplex RT-qPCR reactions. Specifically, we used DisCoVery (targeting N and ORF1ab genes) and Genefinder (targeting N, E and RdRp genes). The detection of these genes was unexpected since the HID method did not provide satisfactory results in the detection of larger amplicons [17] . We found that CT values of both N and ORF1ab viral amplicons detected with DisCoVery were similar to those measured for the purified RNA samples, in line with our observations with the CDC kit. Although the detection of the E and RdRp genes with the Genefinder kit was weaker than the detection of the N gene, the relatively high detection sensitivity of the PK+HID method estimated from the analysis of 94 randomly selected samples suggests that the PK+HID method can be used as an alternative, extraction-free RT-qPCR analysis of nasopharyngeal swab samples in clinical diagnostics. All rights reserved. No reuse allowed without permission. preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. We should emphasize that the success of an extraction-free method strongly depends on the correct combination of transport medium and detection kit. Merindol et al [32] found that saline solution swab medium impaired the performance of the AllplexTM 2019-nCoV Assay. Although, in our hands, saline solution swab medium could be used with 2019-nCoV CDC, GeneFinder TM COVID-19 Plus RealAmp and DisCoVery SARS-CoV-2RT-PCR detection kits. Since the extraction-free protocol may reduce the efficiency of the PCR, it is critical to select a PCR kit that preserves sensitivity. While most bibliography, as well as our own findings point to N as the most sensitive viral region, ORF1ab gene also showed a good performance with our extraction-free protocol. Our results indicate that the combination of proteinase K preincubation with heat inactivation is a very robust and reliable procedure for extraction-free RT-qPCR determinations of SARS-CoV-2. We should mention that preparation of the 94 PK+HID samples required to complete the 96-well PCR plate (2 wells are needed for the positive and negative controls) takes approximately 60 min by one operator whereas manual extraction of the same number of samples using commercial kits would take 5-6 hours of work in a Biosafety Level 2 area with continuous use of a benchtop centrifuge. Additionally, PK treatment involves an estimated cost of 0.37 USD per sample whereas the cost of standard RNA extraction procedures is clearly superior (6-10 USD per sample in Argentina). We should also highlight that the HID method combined with the 2019-nCoV CDC Diagnostic Panel is an inexpensive option since it does not require extra reagents as PK+HID. The simplicity and relatively short period of time required for the protocol and the full-availability and low price of the required reagents place PK+HID as a cheap, simple and fast alternative to traditional RNA extraction protocols. All rights reserved. No reuse allowed without permission. preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. All rights reserved. No reuse allowed without permission. preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this this version posted December 17, 2020. All rights reserved. No reuse allowed without permission. preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this this version posted December 17, 2020. ; https://doi.org/10.1101/2020.12.16.20248350 doi: medRxiv preprint Effectiveness of isolation, testing, contact tracing, and physical distancing on reducing transmission of SARS-CoV-2 in different settings: a mathematical modelling study The explosion of new coronavirus tests that could help to end the pandemic Overcoming the bottleneck to widespread testing: a rapid review of nucleic acid testing approaches for COVID-19 detection preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this this version posted PubMed Central RNA Extraction Kits for COVID-19 Tests Are in Short Supply in US. The Scientist Magazine FDA chief warns of supply 'pressure' on reagents for coronavirus tests Validation of an extraction-free RT-PCR protocol for detection of SARS-CoV2 RNA. medRxiv (Posted as a preprint) Clinical validation of innovative, low cost, kit-free, RNA processing protocol for RT-PCR Posted as a preprint). 2020. Epub Extraction-free SARS-CoV-2 detection by rapid RT-qPCR universal for all primary respiratory materials Use of a simplified sample processing step without RNA extraction for direct SARS-CoV-2 RT-PCR detection Comparison of media and standards for SARS-CoV-2 RT-qPCR without prior RNA preparation Direct-RT-qPCR Detection of SARS-CoV-2 without RNA Extraction as Part of a COVID-19 Testing Strategy: From Sample to Result in One Hour PubMed Central PMCID: PMC7459950. All rights reserved. No reuse allowed without permission preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this this version posted Evaluating the efficacy of RT-qPCR SARS-CoV-2 direct approaches in comparison to RNA extraction A simple method for SARS CoV-2 detection by rRT-PCR without the use of a commercial RNA extraction kit COVID-19 diagnostics for resource-limited settings: Evaluation of "unextracted" qRT-PCR Laboratory management for SARS-CoV-2 detection: a user-friendly combination of the heat treatment approach and rt-Real-time PCR testing RNA by direct RT-qPCR on nasopharyngeal specimens without extraction of viral RNA Massive and rapid COVID-19 testing is feasible by extraction-free SARS-CoV-2 RT-PCR Comparative Evaluation of Three Homogenization Methods for Isolating Middle East Respiratory Syndrome Coronavirus Nucleic Acids From Sputum Samples for Real-Time Reverse Transcription PCR PubMed Central PMCID: PMC4940489. All rights reserved. No reuse allowed without permission preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this this version posted Comparison of three methods for extraction of viral nucleic acids from blood cultures Pretreatment of the clinical sample with Proteinase K allows detection of SARS-CoV-2 in the absence of RNA extraction. bioRxiv (Posted as a preprint) A streamlined approach to rapidly detect SARS-CoV-2 infection, avoiding RNA extraction. medRxiv Amplification efficiency: linking baseline and bias in the analysis of quantitative PCR data Analytical sensitivity and efficiency comparisons of SARS-CoV-2 RT-qPCR primer-probe sets Understanding diagnostic tests 1: sensitivity, specificity and predictive values Fast SARS-CoV-2 detection by RT-qPCR in preheated nasopharyngeal swab samples Direct RT-qPCR detection of SARS-CoV-2 RNA from patient nasopharyngeal swabs without an RNA extraction step While DJS was employed at IXIS LLC at the time of this study, his employment there did not create a competing interest. Further, IXIS LLC had no involvement in this study An alternative workflow for molecular detection of SARS Euro All rights reserved. No reuse allowed without permission the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this this version posted Evaluation of simple nucleic acid extraction methods for the detection of SARS-CoV-2 in nasopharyngeal and saliva specimens during global shortage of extraction kits Rapid, convenient and efficient kit-independent detection of SARS-CoV-2 RNA SARS-CoV-2 detection by direct rRT-PCR without RNA extraction preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this this version posted All rights reserved. No reuse allowed without permission doi: medRxiv preprint CT RNA (gene, kit) CTPK+HID (gene, kit) RP, DisCoVery) nd (N, DisCoVery) nd (ORF1ab, DisCoVery) 23.2 (RP, DisCoVery) (RP, DisCoVery) nd (N, DisCoVery) nd (ORF1ab, DisCoVery) 26.0 (RP, DisCoVery) RP, Genefinder) 40.1 (N, GeneFinder) nd (RdRp, GeneFinder) nd (E, GeneFinder) 30.3 (RP, Genefinder) (RP, Genefinder) 41.3 (N, GeneFinder) nd (RdRp, GeneFinder) nd (E, GeneFinder) 28.2 (RP, Genefinder) RP, Genefinder) nd (N, GeneFinder) nd CT values of false-negative samples detected with the PK+HID method, using the results obtained with purified RNA samples as reference V. L. thanks J. C. Reboreda and A. Quaglino for their support. All rights reserved. No reuse allowed without permission. preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.The copyright holder for this this version posted December 17, 2020. All rights reserved. No reuse allowed without permission. preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.The copyright holder for this this version posted December 17, 2020. (b) Ratio between the relative amplicon copy amounts detected using PK+HID or RNA purification for those nasopharyngeal swab samples that presented CT values >30 for any of the viral genes in purified RNA samples (Ndata =10 and 12 for N and ORF1ab genes, respectively). All rights reserved. No reuse allowed without permission. preprint (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity.The copyright holder for this this version posted December 17, 2020. ; https://doi.org/10.1101/2020.12.16.20248350 doi: medRxiv preprint