key: cord-1047459-fdvwfy29 authors: Al-Qatatsheh, Ahmed; Morsi, Yosry; Zavabeti, Ali; Zolfagharian, Ali; Salim, Nisa; Z. Kouzani, Abbas; Mosadegh, Bobak; Gharaie, Saleh title: Blood Pressure Sensors: Materials, Fabrication Methods, Performance Evaluations and Future Perspectives date: 2020-08-11 journal: Sensors (Basel) DOI: 10.3390/s20164484 sha: 2cf6a96f9b5336b13d23e2ec7cd74aa7b567b367 doc_id: 1047459 cord_uid: fdvwfy29 Advancements in materials science and fabrication techniques have contributed to the significant growing attention to a wide variety of sensors for digital healthcare. While the progress in this area is tremendously impressive, few wearable sensors with the capability of real-time blood pressure monitoring are approved for clinical use. One of the key obstacles in the further development of wearable sensors for medical applications is the lack of comprehensive technical evaluation of sensor materials against the expected clinical performance. Here, we present an extensive review and critical analysis of various materials applied in the design and fabrication of wearable sensors. In our unique transdisciplinary approach, we studied the fundamentals of blood pressure and examined its measuring modalities while focusing on their clinical use and sensing principles to identify material functionalities. Then, we carefully reviewed various categories of functional materials utilized in sensor building blocks allowing for comparative analysis of the performance of a wide range of materials throughout the sensor operational-life cycle. Not only this provides essential data to enhance the materials’ properties and optimize their performance, but also, it highlights new perspectives and provides suggestions to develop the next generation pressure sensors for clinical use. Cardiovascular diseases caused 31% of deaths worldwide [1] , and recently, they had the highest confirmed death cases in Italy and China during the novel pandemic known as the coronavirus disease 2019 [2, 3] . In return, the demand for an accurate home-diagnostic tool for blood pressure measurements, along with other vital signs (e.g., temperature, respiratory rate) has increased massively. These tools, especially if enabled with telemedicine, will not only help assess a patient's health status, triage the patient to appropriate care, determine potential diagnoses, and predict recovery, but also, it will help provide real-time medical monitoring, for instance, people in home-quarantine [4, 5] . Hence, improving the precision and accuracy in blood pressure measurements can help significantly with early diagnosis and cardiovascular risk stratification [6] [7] [8] [9] , because inadequate performance in blood pressure measurement will increase current levels of fatal stroke and fatal myocardial infections [10] , as well as impose an avoidable financial burden [11] . At the beginning of the twenty-first century, the use of sensors and mobile internet begins to provide a platform to continuously monitor all vital signs [12] [13] [14] [15] [16] [17] [18] [19] [20] , including blood pressure. Not only does this help reduce the risk of cardiovascular complications, but also it supports making accurate and real-time healthcare data available for healthcare professionals at the office to assist select the best treatment strategies and consider the impact on patient outcomes [21] [22] [23] [24] [25] [26] . Furthermore, this type of monitoring can save millions of lives around the globe annually [27] [28] [29] . Advancements in engineering and material science have been the main driver in the development of sensor technologies during the past decade [29] [30] [31] . Indeed, tactile sensors, and more precisely, skin-like soft electronics begin to transform healthcare [32] [33] [34] . In return, several studies highlight the crucial implications of this field and indicate that a timely review is necessary [35] [36] [37] . Since most studies focus on device functionality [38] [39] [40] , there is a need to investigate device clinical performance and capabilities beyond proof-of-concept measurements outside of the laboratory [41] , following standardized evaluation approaches [42] . By precisely studying the unique nature of medical needs and evaluating the functionality of sensing principles and materials, we will comprehensively identify materials' properties and their associated performance in line with structure strategies needed for accurate and continuous blood pressure measurement. Also, we will identify challenges along with future research opportunities. We aim to create a crosslink between healthcare practice and material science following a transdisciplinary approach illustrated in (Figure 1 ) to emphasize the importance of design and fabrication elements that have been either overlooked or compromised. Sensors 2020, 20, x FOR PEER REVIEW 2 of 85 home-quarantine [4, 5] . Hence, improving the precision and accuracy in blood pressure measurements can help significantly with early diagnosis and cardiovascular risk stratification [6] [7] [8] [9] , because inadequate performance in blood pressure measurement will increase current levels of fatal stroke and fatal myocardial infections [10] , as well as impose an avoidable financial burden [11] . At the beginning of the twenty-first century, the use of sensors and mobile internet begins to provide a platform to continuously monitor all vital signs [12] [13] [14] [15] [16] [17] [18] [19] [20] , including blood pressure. Not only does this help reduce the risk of cardiovascular complications, but also it supports making accurate and real-time healthcare data available for healthcare professionals at the office to assist select the best treatment strategies and consider the impact on patient outcomes [21] [22] [23] [24] [25] [26] . Furthermore, this type of monitoring can save millions of lives around the globe annually [27] [28] [29] . Advancements in engineering and material science have been the main driver in the development of sensor technologies during the past decade [29] [30] [31] . Indeed, tactile sensors, and more precisely, skin-like soft electronics begin to transform healthcare [32] [33] [34] . In return, several studies highlight the crucial implications of this field and indicate that a timely review is necessary [35] [36] [37] . Since most studies focus on device functionality [38] [39] [40] ., there is a need to investigate device clinical performance and capabilities beyond proof-of-concept measurements outside of the laboratory [41] , following standardized evaluation approaches [42] . By precisely studying the unique nature of medical needs and evaluating the functionality of sensing principles and materials, we will comprehensively identify materials' properties and their associated performance in line with structure strategies needed for accurate and continuous blood pressure measurement. Also, we will identify challenges along with future research opportunities. We aim to create a crosslink between healthcare practice and material science following a transdisciplinary approach illustrated in (Figure 1 ) to emphasize the importance of design and fabrication elements that have been either overlooked or compromised. The theoretical and practical framework behind accurate blood pressure measurement is complex and, sometimes, overlooked entirely [43] , therefore, understanding the effect of different approaches for blood pressure measurement is essential for developing accurate sensing materials suitable for medical use. The volume of blood ejected by the heart into the arteries, the elastance or The theoretical and practical framework behind accurate blood pressure measurement is complex and, sometimes, overlooked entirely [43] , therefore, understanding the effect of different approaches for blood pressure measurement is essential for developing accurate sensing materials suitable for medical use. The volume of blood ejected by the heart into the arteries, the elastance or stiffness of the walls of the arteries, and the rate at which the blood flows out of the arteries altogether affect blood In a healthy individual, systolic blood pressure and diastolic blood pressure are 110-115 mmHg and 70-75 mmHg, respectively [44] [45] [46] . A blood pressure measure out of this range may be associated with the incidence of several cardiovascular events (e.g., stroke, heart failure, and end-stage renal disease) [6, 47] . Since arterial pressure varies continuously during the cardiac cycle, the morphological shape of different signals associated with blood pressure varies as well, as seen in Figure 2c . This shape is acquired as an electrocardiograph (ECG), ballistocardiograph (BCG), and phonocardiograph (PCG) signals [48] [49] [50] . Also, the morphological shapes and values of blood pressure wave vary when it travels from the highly elastic central arteries to the stiffer ones [51] , and when it travels away from the heart either upward (i.e., towards the head) or downward (i.e., towards the foot) in an upright position ( Figure 3 ) [43, 47] . In the next sections, we will discuss blood pressure measurement techniques and carefully evaluating their transducing modalities and materials for accurate real-time monitoring. Invasive blood pressure is directly measured by an intravascular catheter unit, which comprises of three main components: an intra-arterial cannula, an infusion tube, and a transducer [52, 53] . The intra-arterial cannula is a short and parallel-sided cannula made of different materials such as Teflon ® In a healthy individual, systolic blood pressure and diastolic blood pressure are 110-115 mmHg and 70-75 mmHg, respectively [44] [45] [46] . A blood pressure measure out of this range may be associated with the incidence of several cardiovascular events (e.g., stroke, heart failure, and end-stage renal disease) [6, 47] . Since arterial pressure varies continuously during the cardiac cycle, the morphological shape of different signals associated with blood pressure varies as well, as seen in Figure 2c . This shape is acquired as an electrocardiograph (ECG), ballistocardiograph (BCG), and phonocardiograph (PCG) signals [48] [49] [50] . Also, the morphological shapes and values of blood pressure wave vary when it travels from the highly elastic central arteries to the stiffer ones [51] , and when it travels away from the heart either upward (i.e., towards the head) or downward (i.e., towards the foot) in an upright position ( Figure 3 ) [43, 47] . In the next sections, we will discuss blood pressure measurement techniques and carefully evaluating their transducing modalities and materials for accurate real-time monitoring. Minimally-invasive blood pressure measurement is based on nonvascular implantable miniaturized sensors that are compatible with body tissues, and these devices can provide real-time monitoring of the cardiac cycle [69] , including intravascular [70] , intraocular [71] and intracranial [72] using different MEMS-based implantable blood pressure sensors including Au-PI diaphragms [73] and Si nanomembranes [74] . The accuracy of a minimally invasive approach, in contrast to the invasive, is still controversial, and it may be due to the drift in sensitivity over a long time that affects long-term accuracy [75] . The full-occlusion technique includes auscultatory [76] , oscillometry [77] , and palpatory [78] . auscultatory and oscillometry are comparable to a gold standard [6, 79] , unlike palpatory, which is not used because obtaining a diastolic blood pressure measurement is difficult and may lead to considerable error [78] . The accuracy of the oscillometry method can be highly affected by muscle contraction, noise artifacts, artery stiffness, age, and physical health [76] , hence, validation and recalibration are crucial [6, 76] . Auscultatory and oscillometry methods are intermittent [80] and different cuff types [81] [82] [83] [84] [85] [86] and fabrics [87] [88] [89] [90] [91] may lead to different blood pressure measurements [92, 93] . 2 Invasive blood pressure is directly measured by an intravascular catheter unit, which comprises of three main components: an intra-arterial cannula, an infusion tube, and a transducer [52, 53] . The intra-arterial cannula is a short and parallel-sided cannula made of different materials such as Teflon ® [54] , PU [55] , PVC [56] , Vialon [57] or silicone rubber [58] to reduce thrombosis and bacterial infections [59] . The cannula is connected to an infusion tube, and the catheter-tip is the pressure sensing component in the transducer assembly. The transducer assembly conventionally utilizes MEMS technology to convert pressure waves into electrical signals [60, 61] using silicon-based [62] and non-silicon based MEMS, such as Ti/Pt metallic wire coated with PI/SU [63] , PEDOT: PSS with a Ag protective layer on a flexible PDMS substrate [64] and PVF 2 [65] . Other non-catheter-based pressure sensors include a capacitive-based bioresorbable POMaC/PGS/Mg on a (PHB/PHV) substrate sensor [66] . The invasive approach is accurate and free of operator bias. Indeed, it is considered the gold standard for all other measures [67, 68] . Minimally-invasive blood pressure measurement is based on nonvascular implantable miniaturized sensors that are compatible with body tissues, and these devices can provide real-time monitoring of the cardiac cycle [69] , including intravascular [70] , intraocular [71] and intracranial [72] using different MEMS-based implantable blood pressure sensors including Au-PI diaphragms [73] and Si nanomembranes [74] . The accuracy of a minimally invasive approach, in contrast to the invasive, is still controversial, and it may be due to the drift in sensitivity over a long time that affects long-term accuracy [75] . The full-occlusion technique includes auscultatory [76] , oscillometry [77] , and palpatory [78] . auscultatory and oscillometry are comparable to a gold standard [6, 79] , unlike palpatory, which is not used because obtaining a diastolic blood pressure measurement is difficult and may lead to considerable error [78] . The accuracy of the oscillometry method can be highly affected by muscle contraction, noise artifacts, artery stiffness, age, and physical health [76] , hence, validation and recalibration The selection of the most suitable modality can efficiently enhance the accuracy of the non-invasive blood pressure measurement based on wearable devices. For instance, ballistocardiography (BCG) and seismocardiography have recently attracted attention [146] and several BCG and SCG transducing modality sensors utilizing different conductive materials have become available including PVDF [147] and PVDF/Au electrodes in a skin soft electronic tattoo [148] , Au/PET and graphene/PMMA at thermal release tape (TRT) or tattoo paper [149] , biaxial PP film based on EmFi [150] and a polysilicon surface-micromachined and monolith silicon based on MEMS [150, 151] . Wearable and fixable design and fabrication of electrocardiography (ECG) and impedance cardiography (ICG) electrodes include the use of PEDOT: PSS and Ag plated electrodes [149] , PVDF-based electrodes [152] and high-grade Ag coated fibers on a Textrode [153] . Wet Ag/AgCl electrodes are highly inconvenient for long-term applications because they lead to skin rashes and allergies with prolonged use [154] and when replacing wet electrodes with dry ones, dry electrodes have been reported to compromise user's safety due to direct electrical contact between the skin and the electrode [155] .Using non-contact capacitive coupled electrodes (CCEs) to overcome this safety issue may limit the use of ECG and ICG in continuous blood pressure monitoring [156] . Also, ECG and ICG provide spot measurement and are not suitable for long-term cardiac tracking for wearable devices worn on the wrist [157] . Skin patch sensors based on electromagnetic (EM) detection were built from a conductive trace of copper to measure intravascular stroke volume [124] and intracranial blood pressure [158] . Also, a magnetoelastic skin curvature sensor along with ECG electrodes was used to measure blood pressure in the carotid [159] . The use of the EM modality needs high levels of calculation complexity and relies on many mathematical assumptions that may affect the measurement accuracy. Optical transduction based on PPG has disadvantages that can limit its usability for accurate, wearable, and continuous blood pressure measurement. The PPG working principle assumes that blood has a constant light absorptivity. In contrast, blood light absorptivity is highly affected by blood composition, particularly substances that have high absorptivity at NIR, such as hemoglobin [160, 161] . Also, PPG requires direct contact with the skin that may cause discomfort to the user [162] . Furthermore, the use of the PPG-based transducing modality requires a stable contact force between the sensor and the measurement site [163] . The LED in PPG has a relatively limited light penetration depth (i.e., up to 8 mm). Hence, PPG use is limited to superficial arteries such as radial arteries and peripheral arterioles in fingers and ear lobes [164] . Furthermore, blood pressure waveforms cannot be extracted accurately from arteries adjacent to veins because any volumetric circulatory change in the artery will simply interfere with that in the vein [115] . Other transducing modalities including phonocardiography (PCG) using a PVDF-based sensor [165] , tonoarteriography (TAG) using a flexible piezoresistive pressure sensor [166] and ultrasound using a piezo-pillar with filling epoxy on Cu and Cu/Sn electrode at Pi substrate [125] are promising. However, they rely on several mathematical assumptions, as well as physiological artifacts and measurements, including blood viscosity, vessel radius, and beam inclination that may affect their measurement accuracy. ( Figure 5 ) illustrates transducing modalities used in blood pressure measurement. On the other hand, mechanical and acoustic transducing modalities have remarkably paved the way for more efficient signal-feature extraction that can reflect the insightful information on blood pressure dynamics [163, [167] [168] [169] [170] . In Table 2 we summarize transducing modalities along with some of the sensing principles associated with blood pressure measurement. Sensors 2020, 20, x; doi: FOR PEER REVIEW www.mdpi.com/journal/sensors phonocardiography (PCG) using a PVDF-based sensor [165] , tonoarteriography (TAG) using a flexible piezoresistive pressure sensor [166] and ultrasound using a piezo-pillar with filling epoxy on Cu and Cu/Sn electrode at Pi substrate [125] are promising. However, they rely on several mathematical assumptions, as well as physiological artifacts and measurements, including blood viscosity, vessel radius, and beam inclination that may affect their measurement accuracy. ( Figure 5 ) illustrates transducing modalities used in blood pressure measurement. On the other hand, mechanical and acoustic transducing modalities have remarkably paved the way for more efficient signal-feature extraction that can reflect the insightful information on blood pressure dynamics [163, [167] [168] [169] [170] . In Table 2 we summarize transducing modalities along with some of the sensing principles associated with blood pressure measurement. In this section, we review the fundamental sensing principles and evaluate the feasibility of each to identify those with the required performance. Creating a conductive network within an insulating matrix can be explained using the percolation theory [176] . The percolation theory explains the behavior of the composite matrix while transforming from an insulator to a conductor by increasing the content of a conducting filler or fiber gradually till the content reaches "the percolation threshold," where the relationship between the measured electrical resistivity of the composite matrix and filler volume can be expressed as in Equation (1) [177] : where ρ resistivity of composites, ρ 0 resistivity of conductive filler, v filler volume, v c percolation threshold and t = critical exponent. The piezoresistivity of the composite matrix dramatically increases several orders of magnitude as a result of the formation of continuous electron paths or conducting networks immediately after the filler volume fraction exceeds a percolation threshold. When the nanocomposite matrix reached an acceptable level of resistivity, the resistance-based electrical signal can be transduced into an applied pressure [178] . Therefore, the change in composite resistivity is mainly derived from two factors: The deformation in the composite geometry that may lead to changes in its length and cross-section area. The change in resistivity of the composite by changing the resistivity and/or volume of the conductive filler. The piezoresistive sensing principle is widely used in wearable sensors, including the use of graphene and its derivatives [179, 180] , CNT and CB [181] [182] [183] , metal NPs and NWs [184, 185] , conductive polymers [186, 187] and MXenes [188, 189] . In a two-parallel-plate capacitor, capacitance is directly proportional to the relative static permittivity of the material between the two plates, the area of overlap of the two plates, and the electric conductivity of the material that comprise the two plates. Capacitance is inversely proportional to the separation distance between the two plates. A small amount of exerted force may cause the plates to deflect and capacitance to change [190] . Like the piezoresistive one, the piezocapacitive sensing principle is widely used in wearable sensors, including the use of graphene and its derivatives [191, 192] , CNT [193, 194] , metal NPs and NWs [195] [196] [197] , conductive polymers [198, 199] and MXenes [200] . Optical pressure sensors are devices through which light is often guided into a fiber or an optical waveguide. The applied pressure subsequently modifies the light intensity or wavelength through a Fiber Bragg Grating (FBG) configuration [201] . FBG wearable sensors have gained interest as they have electromagnetic immunity and high sensitivity that make them ideal for the use in an MRI environment with no threat to the patient and no influence over the quality of imaging [202] [203] [204] while using electrically non-conductive, bio-compatible and low optical permeability materials such as PDMS [205] and PA [206] . However, FBG-based sensing is constrained with the necessity of using the sophisticated signal analysis to detect breathing and cardiac activity [203] , the little resistance to mechanical stress [206] and the limited range of Bragg wavelengths [204] . The field effect transistor (FET) pressure sensor utilizes transducing materials in the gate or channel region while controlling the flow of electrical current. The FET modalities include the graphene FET pressure sensor and its derivative [207, 208] , PVDF multimodal FET sensor [209] and pentacene/ P(VDF-TrFe) multifunctional OFET sensor. For most FET-based pressure sensors, the sensing mechanisms generally relies on the change in the capacitance of the dielectric layer, hence a small change in capacitance will generate a relatively high current signal output due to transistor signal amplification function [210, 211] . The triboelectric pressure sensor utilizes the effect of contact-induced electrification [212] [213] [214] . Generally, a material would become electrically charged after it comes into physical contact with another dissimilar material, and the strength of charges are different for different materials [212] [213] [214] [215] . Triboelectric sensors include a textile-based sensor made of Ag-coated fabric [215] , an eardrum-inspired active sensor made of an ITO-coated nylon thin film laminated onto a PET substrate [216] , a membrane-based triboelectric sensor [217] , a Downy-structured triboelectric nanogenerator (D-TENG)-based sensor [218] , a 3D cellular sensor array [219] , a flexible weaving constructed self-powered pressure sensor (WCSPS) [220] and a shape-memory PU (SMPU)-based sensor [221] . D-TENG-based and 3D cellular sensor array sensors showed encouraging results for blood pressure measurement [218, 219] , and flexible weaving constructed self-powered pressure sensors showed a discrepancy of about 0.87-3.65% when compared to a commercial cuff-based device [220] . However, triboelectric-based pressure sensors usually show a relatively low limit of detection [216, 217] and its mechanism does not provide a stable and exact signal output that is sensitive to the magnitude, direction, and location of the applied stress in real-time [222] . Furthermore, the mechanism is profoundly affected by the amount of kinetic energy and momentum generated [223, 224] that do not suit this type of application for accurate, yet continuous, blood pressure measurement and monitoring throughout all-day activities. Because of the findings we presented above, and because the triboelectric mechanism produces signals sensitive to humidity variations regardless of the use of a hydrophobic substrate structure [222] we will limit our coverage, in this review, to a few examples. The design and fabrication of hydrophobic surfaces, suitable for this mechanism, remain challenging for different combinations of materials to ensure stable operation under various service conditions. ( Figure 6 ) illustrates the sensing principles we studied in this review. triboelectric-based pressure sensors usually show a relatively low limit of detection [216, 217] and its mechanism does not provide a stable and exact signal output that is sensitive to the magnitude, direction, and location of the applied stress in real-time [222] . Furthermore, the mechanism is profoundly affected by the amount of kinetic energy and momentum generated [223, 224] that do not suit this type of application for accurate, yet continuous, blood pressure measurement and monitoring throughout all-day activities. Because of the findings we presented above, and because the triboelectric mechanism produces signals sensitive to humidity variations regardless of the use of a hydrophobic substrate structure [222] we will limit our coverage, in this review, to a few examples. The design and fabrication of hydrophobic surfaces, suitable for this mechanism, remain challenging for different combinations of materials to ensure stable operation under various service conditions. ( Figure 6 )illustrates the sensing principles we studied in this review. We have carefully reviewed and considered several factors to retain the highest performance of different design and fabrication strategies. Setting a performance review criterion is crucial to select functional materials that can maintain essential qualities for interacting with biological systems in wearable devices [225] . For instance, mechanical pliability is important for devices that are in direct contact with some regions of the skin to minimize discomfort and respond to various strains associated with the body's motion. Equally, suitability for medical use is important, too. The dimensions of our review criteria cover mainly the essential performance elements necessary for accurate and continuous blood pressure measurement when compared to a gold standard (Figure 7 ), including sensitivity, the limit of detection, and response time. Depending on information availability, additional dimensions were included, such as hysteresis, and simplicity in design. For example, hysteresis is a measure of the difference between the received signal and the applied force. High hysteresis is a significant disadvantage, and its effect, therefore, should be reduced. We have carefully reviewed and considered several factors to retain the highest performance of different design and fabrication strategies. Setting a performance review criterion is crucial to select functional materials that can maintain essential qualities for interacting with biological systems in wearable devices [225] . For instance, mechanical pliability is important for devices that are in direct contact with some regions of the skin to minimize discomfort and respond to various strains associated with the body's motion. Equally, suitability for medical use is important, too. The dimensions of our review criteria cover mainly the essential performance elements necessary for accurate and continuous blood pressure measurement when compared to a gold standard (Figure 7 ), including sensitivity, the limit of detection, and response time. Depending on information availability, additional dimensions were included, such as hysteresis, and simplicity in design. For example, hysteresis is a measure of the difference between the received signal and the applied force. High hysteresis is a significant disadvantage, and its effect, therefore, should be reduced. Active materials, electrode, and substrate are the essential components of a typical wearable sensor [226] [227] [228] . Besides, a dielectric material, which is an electrical insulator that can be polarized by an applied electric field, is used in some other sensors such as flexible FET-type sensors [211] . Flexibility, stretchability, and conductivity are amongst the most critical properties, hence, developing suitable functional materials with proper electrical and mechanical performance is of great importance to ensure maintaining high performance under various operating conditions. In this section, we review the development of active, substrate, and electrode materials, and study the effect of the use of different design and fabrication strategies of different materials on sensor performance. The substrate is considered the primary source for the stretchability of wearable sensors, and directly determines the level of comfort and long-term durability. The substrate is the outermost layer, which may directly contact body tissues, hence affects the safety and performance of the sensor. Materials used as substrates are primarily made of either polymers or natural materials, and their properties will depend on their chemical structures and processing approaches. PDMS is one of the most widely used material as a substrate for its high stretchability. It is non-toxic, non-flammable, and hydrophobic with acceptable processability. PET has good transparency, high creep resistance, and acceptable printability. Cellulose paper, on the other side, is biodegradable, biocompatible, recyclable, and inherently flexible, but its durability and stability are still amongst the most significant challenges for enhancing its properties. Table 3 illustrates some of the materials used as a substrate in wearable sensors along with their features and Figure 8 depicts different strategies for substrate design and fabrication. Active materials, electrode, and substrate are the essential components of a typical wearable sensor [226] [227] [228] . Besides, a dielectric material, which is an electrical insulator that can be polarized by an applied electric field, is used in some other sensors such as flexible FET-type sensors [211] . Flexibility, stretchability, and conductivity are amongst the most critical properties, hence, developing suitable functional materials with proper electrical and mechanical performance is of great importance to ensure maintaining high performance under various operating conditions. In this section, we review the development of active, substrate, and electrode materials, and study the effect of the use of different design and fabrication strategies of different materials on sensor performance. The substrate is considered the primary source for the stretchability of wearable sensors, and directly determines the level of comfort and long-term durability. The substrate is the outermost layer, which may directly contact body tissues, hence affects the safety and performance of the sensor. Materials used as substrates are primarily made of either polymers or natural materials, and their properties will depend on their chemical structures and processing approaches. PDMS is one of the most widely used material as a substrate for its high stretchability. It is non-toxic, non-flammable, and hydrophobic with acceptable processability. PET has good transparency, high creep resistance, and acceptable printability. Cellulose paper, on the other side, is biodegradable, biocompatible, recyclable, and inherently flexible, but its durability and stability are still amongst the most significant challenges for enhancing its properties. Table 3 illustrates some of the materials used as a substrate in wearable sensors along with their features and Figure 8 depicts different strategies for substrate design and fabrication. PDMS materials can be processed quickly to form approximately any shape such as "bumpy" [237] that can be used to squeeze out an air gap to increase the capacitance. This design is simple and provides flexibility, but it cannot provide the desired limit of detection while maintaining capacitance. In addition to their stretchability and simplicity, several design and fabrication strategies for microstructure elastomer include a naturally molded substrate [242] , and patterned substrate [238, 243] where the combination of microstructure elastomers and active materials (i.e., sensing elements) are sandwiched by the patterned electrodes to create a resistive response that changes whenever a stimulus (e.g., pressure) is applied. In respect to their low limit of detection and relatively high sensitivity, the sensors with naturally molded and patterned substrates are, sometimes, associated with low durability. The use of the elastomer composite substrate may overcome this drawback and others and provide a practical approach where active materials (e.g., CNTs) are incorporated into the elastomer matrix (e.g., PDMS) [239] . Active materials in the elastomer composite highly improve electrical conductivity and are useful for the reinforcement of the elastomer due to their low density and high aspect ratio. When such fillers are added to an elastomer, its mechanical properties, including tensile strength and hardness, increase, and this will depend on loading, dispersion, and alignment in the composite matrix; yet, this may affect the sensor sensitivity and reduce its limit of detection. As a result, the fabrication of a pressure sensor towards practical applications is becoming a challenging task. There are different approaches to achieve high sensitivity, such as spray deposition [244] and multilayer spray coating [245] , but they require sophisticated fabrication approaches. Other approaches, including porous composites with active material coatings [240] , textile, and thin films [236] and printed electronics [241] have been recently utilized in tactile sensors, and they are used in different flexible electronic applications, such as artificial skin [246] . These approaches may help provide better mechanical properties in comparison with that of composite elastomers. However, the need to achieve high sensitivity without affecting other pressure performance elements, such as the limit of detection and durability, is still not adequately addressed. For example, Ag nanowire (NW) coated on cotton as a substrate [236] provides relatively high sensitivity (i.e., 3.4 kPa −1 ) and rapid response and relaxation time (i.e., <50 ms), but with relatively less flexibility (i.e., > 5000 cycles). In contrast, CB at PU as a substrate [240] provides relatively higher sensitivity (i.e., 16.4 kPa −1 ), a faster response time (i.e., 20 ms), and higher durability (i.e., 50,000 cycles) but with a relatively low limit of detection PDMS materials can be processed quickly to form approximately any shape such as "bumpy" [237] that can be used to squeeze out an air gap to increase the capacitance. This design is simple and provides flexibility, but it cannot provide the desired limit of detection while maintaining capacitance. In addition to their stretchability and simplicity, several design and fabrication strategies for microstructure elastomer include a naturally molded substrate [242] , and patterned substrate [238, 243] where the combination of microstructure elastomers and active materials (i.e., sensing elements) are sandwiched by the patterned electrodes to create a resistive response that changes whenever a stimulus (e.g., pressure) is applied. In respect to their low limit of detection and relatively high sensitivity, the sensors with naturally molded and patterned substrates are, sometimes, associated with low durability. The use of the elastomer composite substrate may overcome this drawback and others and provide a practical approach where active materials (e.g., CNTs) are incorporated into the elastomer matrix (e.g., PDMS) [239] . Active materials in the elastomer composite highly improve electrical conductivity and are useful for the reinforcement of the elastomer due to their low density and high aspect ratio. When such fillers are added to an elastomer, its mechanical properties, including tensile strength and hardness, increase, and this will depend on loading, dispersion, and alignment in the composite matrix; yet, this may affect the sensor sensitivity and reduce its limit of detection. As a result, the fabrication of a pressure sensor towards practical applications is becoming a challenging task. There are different approaches to achieve high sensitivity, such as spray deposition [244] and multilayer spray coating [245] , but they require sophisticated fabrication approaches. Other approaches, including porous composites with active material coatings [240] , textile, and thin films [236] and printed electronics [241] have been recently utilized in tactile sensors, and they are used in different flexible electronic applications, such as artificial skin [246] . These approaches may help provide better mechanical properties in comparison with that of composite elastomers. However, the need to achieve high sensitivity without affecting other pressure performance elements, such as the limit of detection and durability, is still not adequately addressed. For example, Ag nanowire (NW) coated on cotton as a substrate [236] provides relatively high sensitivity (i.e., 3.4 kPa −1 ) and rapid response and relaxation time (i.e., <50 ms), but with relatively less flexibility (i.e., >5000 cycles). In contrast, CB at PU as a substrate [240] provides relatively higher sensitivity (i.e., 16.4 kPa −1 ), a faster response time (i.e., 20 ms), and higher durability (i.e., 50,000 cycles) but with a relatively low limit of detection (i.e., 91 Pa) that does not fulfill blood pressure measurement requirements. In the next section, we will address this challenge in a more systematic approach. The active materials represent a crucial component to respond to external stimuli. Generally, high performing pressure sensors need to pose high conductivity, exceptional chemical stability, and durability, as well as high flexibility as such, adding to excellent performance that includes fast response and high sensitivity. Nano conducting materials including carbon materials, such as CNT [239] and graphene [247] , NWs and nanoparticles (NPs) [236, 248] , metal-organic-framework (MOF), Mxens [249, 250] and conducting polymers [251] are among the most commonly used conductive components (i.e., active materials) that can be either embedded into or placed on the elastomeric polymer substrates. Figure 9 illustrates the different types of active materials used in sensors. Sensors 2020, 20, x FOR PEER REVIEW 17 of 85 (i.e., 91Pa) that does not fulfill blood pressure measurement requirements. In the next section, we will address this challenge in a more systematic approach. The active materials represent a crucial component to respond to external stimuli. Generally, high performing pressure sensors need to pose high conductivity, exceptional chemical stability, and durability, as well as high flexibility as such, adding to excellent performance that includes fast response and high sensitivity. Nano conducting materials including carbon materials, such as CNT [239] and graphene [247] , NWs and nanoparticles (NPs) [236, 248] , metal-organic-framework (MOF), Mxens [249, 250] and conducting polymers [251] are among the most commonly used conductive components (i.e., active materials) that can be either embedded into or placed on the elastomeric polymer substrates. Figure 9 illustrates the different types of active materials used in sensors CB is a form of paracrystalline carbon with different sizes and shapes; it has a relatively high surface-area-to-volume ratio and electrical conductivity [252] . CNTs are widely used as active materials due to their high electrical and thermal conductivity [253] , great mechanical flexibility, and excellent carrier mobility [254] . There are generally two types of CNTs: SWCNTs and MWCNTs. SWCNTs are cylindrical with a diameter on the order of nanometres that may vary due to strain and different modes of wrapping. An SWCNT results when one layer of carbon atoms in two-dimensional hexagonal lattices is wrapped into the form of a long cylinder (i.e., aspect ratio 300-1000) [255] . Utilizing different types of sensing principles, carbon materials as active materials have been used with PU as CB at PU sponge [240] and MWCNT with rGO at PU foam [256] . Also, it has been used with PDMS as CNT [257, 258] , aligned carbon nanotube (ACNT) with graphene [259] and vertically aligned carbon nanotube (VACNT) [260] , as well as EcoFlex rubber as CNT [194, 244] . In addition to the role of volume fraction and aspect ratio of the carbon materials in nanocomposite matrix in improving electrical and mechanical properties, the orientation of active materials within the nanocomposite matrix and substrate morphology (e.g., cell/pore size) can play a pivotal role in tremendously enhancing the matrix sensitivity, the limit of detection, and the durability. The effect of orientation and morphology is illustrated in Figures 10-12 . CB is a form of paracrystalline carbon with different sizes and shapes; it has a relatively high surface-area-to-volume ratio and electrical conductivity [252] . CNTs are widely used as active materials due to their high electrical and thermal conductivity [253] , great mechanical flexibility, and excellent carrier mobility [254] . There are generally two types of CNTs: SWCNTs and MWCNTs. SWCNTs are cylindrical with a diameter on the order of nanometres that may vary due to strain and different modes of wrapping. An SWCNT results when one layer of carbon atoms in two-dimensional hexagonal lattices is wrapped into the form of a long cylinder (i.e., aspect ratio 300-1000) [255] . Utilizing different types of sensing principles, carbon materials as active materials have been used with PU as CB at PU sponge [240] and MWCNT with rGO at PU foam [256] . Also, it has been used with PDMS as CNT [257, 258] , aligned carbon nanotube (ACNT) with graphene [259] and vertically aligned carbon nanotube (VACNT) [260] , as well as EcoFlex rubber as CNT [194, 244] . In addition to the role of volume fraction and aspect ratio of the carbon materials in nanocomposite matrix in improving electrical and mechanical properties, the orientation of active materials within the nanocomposite matrix and substrate morphology (e.g., cell/pore size) can play a pivotal role in tremendously enhancing the matrix sensitivity, the limit of detection, and the durability. The effect of orientation and morphology is illustrated in sensitivity and limit of detection without affecting flexibility. ACNTs architecture is essential for achieving enhanced fundamental characterization. For instance, it was shown that electron mobility of highly ordered CNTs films is 43 times higher than that of a random CNTs network [261] hence the CNTs volume faction in the nanocomposite matrix can be reduced to create a conducting percolation network when CNTs are aligned. Likewise, lowering CNTs volume fraction in nanocomposite matrix improves mechanical properties; storage modulus of ACNTs nanocomposite improved by approximately 40.0% as compared to the random CNTs [262] and tensile strength, as well as elongation of ACNTs fibrous composite enhanced by 150.0% and 62.5%, respectively, when compared to randomized CNTs fibrous composite with similar volume fraction [263] . sensitivity and limit of detection without affecting flexibility. ACNTs architecture is essential for achieving enhanced fundamental characterization. For instance, it was shown that electron mobility of highly ordered CNTs films is 43 times higher than that of a random CNTs network [261] hence the CNTs volume faction in the nanocomposite matrix can be reduced to create a conducting percolation network when CNTs are aligned. Likewise, lowering CNTs volume fraction in nanocomposite matrix improves mechanical properties; storage modulus of ACNTs nanocomposite improved by approximately 40.0% as compared to the random CNTs [262] and tensile strength, as well as elongation of ACNTs fibrous composite enhanced by 150.0% and 62.5%, respectively, when compared to randomized CNTs fibrous composite with similar volume fraction [263] . Figure 11 depicts the performance of a CB based sensor inspired by spider crack mechanisms [240] that naturally increases sensitivity without negatively affecting mechanical properties. In this design, rigid and spherical CB NPs were selected as a conductive filler to form conductive layers on PU Sponge rather than fillers with a higher aspect ratio that could result in entangling or staking with each other, hence reducing flexibility and durability. Another carbon material used as an active material is graphene, which is a 2D allotrope of a single sp 2 hybridized carbon atom graphite layer comprised of a monoatomic hexagonal lattice structure [264] . Graphene has unique properties, such as large surface area (i.e., up to 263 m 2 g −1 ), high stiffness (i.e., up to 1.0 TPa), high thermal conductivity (i.e., up to 5000 W m −1 k −1 ), high electron mobility (i.e., up to 250,000 cm 2 V −1 S −1 ), high electrical conductivity (i.e., 200,000 cm 2 V −1 s −1 ), and tuneable bandgap [265] . GO is a chemically modified graphene with O2 functional groups such as epoxides, alcohols, and carboxylic acids. GO has received a great deal of attention because it readily exfoliates as single sheets in water [266] . rGO, on the other hand, is prepared from rGO by thermal, chemical, or electrical treatments. Hence there are always some defects resulting from unreduced O2 functional groups in or on the rGO surface, and subsequently, rGO cannot have the perfect graphene structure, which was described above [267] . Graphene has been used with PET substrates, such as a graphene electrode [190, 247, 268] , a graphene FET [269] and a stencil mask [192] . Also, it has been used with PDMS, such as a graphene film [242] or as an rGO at PDMS [238] or as an rGO at PDMS and ITO at PET [243] . Also, it has been used as a graphene FET [270] and as a laser-induced graphene (LIG) [271] (Table 4) . Carbon compounds such as CNT and CB have been around for around thirty years, and they will keep attracting great interest within the scientific community owing to their superior mechanical, electric, and optoelectronic properties. Still, bio-inspired designs, fabrication strategies, other than the use of aligned primary conductive constituent in laminar composites, and alignment techniques need to be further investigated to meet the performance requirements of medical applications. Table 4 . Summary of active materials, sensing principles used in wearable sensors along with their sensing properties. Figure 12 shows two different orientation techniques of CNTs on PDMS substrates to increase sensitivity and limit of detection without affecting flexibility. ACNTs architecture is essential for achieving enhanced fundamental characterization. For instance, it was shown that electron mobility of highly ordered CNTs films is 43 times higher than that of a random CNTs network [261] hence the CNTs volume faction in the nanocomposite matrix can be reduced to create a conducting percolation network when CNTs are aligned. Likewise, lowering CNTs volume fraction in nanocomposite matrix improves mechanical properties; storage modulus of ACNTs nanocomposite improved by approximately 40.0% as compared to the random CNTs [262] and tensile strength, as well as elongation of ACNTs fibrous composite enhanced by 150.0% and 62.5%, respectively, when compared to randomized CNTs fibrous composite with similar volume fraction [263] . Another carbon material used as an active material is graphene, which is a 2D allotrope of a single sp 2 hybridized carbon atom graphite layer comprised of a monoatomic hexagonal lattice structure [264] . Graphene has unique properties, such as large surface area (i.e., up to 263 m 2 g −1 ), high stiffness (i.e., up to 1.0 TPa), high thermal conductivity (i.e., up to 5000 W m −1 k −1 ), high electron mobility (i.e., up to 250,000 cm 2 V −1 S −1 ), high electrical conductivity (i.e., 200,000 cm 2 V −1 s −1 ), and tuneable bandgap [265] . GO is a chemically modified graphene with O 2 functional groups such as epoxides, alcohols, and carboxylic acids. GO has received a great deal of attention because it readily exfoliates as single sheets in water [266] . rGO, on the other hand, is prepared from rGO by thermal, chemical, or electrical treatments. Hence there are always some defects resulting from unreduced O 2 functional groups in or on the rGO surface, and subsequently, rGO cannot have the perfect graphene structure, which was described above [267] . Graphene has been used with PET substrates, such as a graphene electrode [190, 247, 268] , a graphene FET [269] and a stencil mask [192] . Also, it has been used with PDMS, such as a graphene film [242] or as an rGO at PDMS [238] or as an rGO at PDMS and ITO at PET [243] . Also, it has been used as a graphene FET [270] and as a laser-induced graphene (LIG) [271] (Table 4 ). [194] • No blood pressure measurement application but heart pulse. [190] • No blood pressure measurement application but heart pulse. • Stability for around 5000 loading/unloading cycles. 6 The applied voltage is 1.0 mV. 7 • Response time is 30ms. 5 [195] • No blood pressure measurement application. Stability for around 5000 loading/unloading cycles. 6 • Not all properties were reported. Carbon compounds such as CNT and CB have been around for around thirty years, and they will keep attracting great interest within the scientific community owing to their superior mechanical, electric, and optoelectronic properties. Still, bio-inspired designs, fabrication strategies, other than the use of aligned primary conductive constituent in laminar composites, and alignment techniques need to be further investigated to meet the performance requirements of medical applications. Graphene has been used with PI, such as a porous graphene (PG) sponge [276] , interdigital electrode (IDE) [285] and with PU as rGO [281] or rGO with GO [273] . Furthermore, graphene has been used with materials such as silicon nitride in MEMS [274] , PEN [278] , PANI wrapped sponge [279] , PMMA [277] , DN hydrogels [280] and EcoFlex rubber [284] . Surprisingly, graphene has been also used with other unconventional materials such as tissue paper [275] polystyrene balls [283] and even a 3M VHB Tape [282] . Getting a flexible 3D graphene structure with excellent 2D electrical properties was common across strategies used to design and fabricate the top-performing sensors with high sensitivity and flexibility, as illustrated in Figure 13 . In the work of Xia and co-workers [242] a 3D graphene structure (i.e., fingerprint-like 3D graphene) was produced in a mechanism different from that of the seed-induced 2D graphene growth of the chemical vapor deposition (CVD) method. This mechanism considers the edge seed-induced growth in a hydrogen-rich environment. Furthermore, the use of hierarchal structure PDMS films (i.e., molded from natural leaves) improves the sensitivity as it provides support and abundant contact sites. Jia and co-workers [282] present a novel technique, inspired by the microstructure of the human skin surface, where gradient rGO wrinkles are produced, through the reduction of coated GO on a pre-strained 3M VHB Tape. The high sensitivity is achieved through a three-stage contribution mechanism: the dome-like microstructure, caused by the expansion of the underlying rGO layer, increases the active contact area between the sensing layer and the counter electrode leading to an increase in the electron flow path. The gradual disappearance of air gaps, between the stack of GO sheets resulting from chemical treatment when undergoing applied pressure, will make the contact sites among the sheets to increase rapidly, and therefore the electron flow path will further increase significantly. The compression in the wrinkle ridge formed by the flexible matrix will increase the rate of change in the contact area that is positively correlated with sensitivity. Figure 14 illustrates the design and fabrication techniques of the work of Zhu and co-workers [271] and Jia and co-workers [282] , as well as graphene and rGO properties. Zhu and co-workers [271] offer a scalable fabrication technology for producing 3D patterned graphene films (i.e., 50 µm) from PDMS with high conductivity and excellent mechanical properties. Direct laser scribing, in ambient air, provides high temperatures that can break Si-C, Si-O, and C-H bonds in PDMS. In return, Si and H atoms can combine with either the O 2 in the air or the O 2 atoms in PDMS to form SiO 2 and gaseous water; the remaining carbon atoms are arranged into graphite structures.This process may be further elaborated through mechanisms of photothermal/photochemical process where the sp 3 (i.e., carbon atoms are converted to sp 2 ) carbon atoms by pulsed laser irradiation [298] . Jia and co-workers [282] present a novel technique, inspired by the microstructure of the human skin surface, where gradient rGO wrinkles are produced, through the reduction of coated GO on a pre-strained 3M VHB Tape. The high sensitivity is achieved through a three-stage contribution mechanism: the dome-like microstructure, caused by the expansion of the underlying rGO layer, increases the active contact area between the sensing layer and the counter electrode leading to an increase in the electron flow path. The gradual disappearance of air gaps, between the stack of GO sheets resulting from chemical treatment when undergoing applied pressure, will make the contact sites among the sheets to increase rapidly, and therefore the electron flow path will further increase significantly. The compression in the wrinkle ridge formed by the flexible matrix will increase the rate of change in the contact area that is positively correlated with sensitivity. Figure 14 illustrates the design and fabrication techniques of the work of Zhu and co-workers [271] and Jia and co-workers [282] , as well as graphene and rGO properties. Graphene has high electrical conductivity, it is more durable than diamond, yet it can be stretched by a quarter of its length, like soft rubber. With such an array of unique properties, Graphene allows for ground-breaking biomedical applications. Still, fabrication strategies, other than embedding aligned conductive material between layers of elastomeric material by different deposition or laser-induced techniques, and alignment techniques need to be further investigated to meet the performance requirements of medical applications. As more focus has been devoted recently to develop highly efficient piezoresistive pressure sensors, active materials other than carbon-based materials were studied. Ferroelectric polymers such as PVDF is considered as an essential category of frequently used active materials in terms of its flexibility and ease of processing [299] . PVDF is a semicrystalline homopolymer with the molecular formula [-CH2-CF2-]n, the crystallite polymorphs have five phases (α, β, Υ, δ, and ε) [300] The range of electrical dipole moments of the PVDF monomer is 5.0 × 10 −3 • C.m for α-PVDF and 8.0 × 10 −3 • C.m for β-PVDF [301, 302] hence, conducting polymers are advantageous over other active materials due to their sensitivity and flexibility [303] PVDF has been used in many sensors [287] as PVDF-HFP and PEDOT [286] as well as, along with rGO [251] Furthermore, PEDOT: PSS and an aqueous PUD elastomer blend were used as a composite polymer with high conductivity [176] Another application is the utilization of PPy@PVA-co-PE Nanofibers at elastic POE Nanofibers [288] The composite film of rGO/PVDF showed an excellent performance that can measure the static and dynamic tactile and thermal signals, and this performance has been achieved through the use of rGO that boosted the sensitivity of Electrospun PVDF-TrFe Nanofibers, which in return, maintained flexibility in the nanocomposite. (Figure 15 ) illustrates the high performance of work [251] . Sensors 2020, 20, x FOR PEER REVIEW 37 of 85 electrodes [304] However, there is a need for investigation for new strategies to enhance mechanical properties and reusability further to meet the performance requirements of medical applications. Unlike nanoparticles, which are on the order of nanometres with different shapes (e.g., semiround, nano-cubes, spheres) [290] , nanowires (NWs) are structures with a length typically on the order of microns and diameter on the order of tens of several hundred nanometres (i.e., high aspect ratio >1000) [305, 306] . NWs have excellent mechanical properties when they are scaled below 100 nanometres. However, their resistivity change in response to applied stresses [307] and this makes them favourable for the use in a wide range of applications of piezoresistivity. Ultralong 1D metal NWs can be used to increase sensitivity without affecting flexibility because of their high aspect ratio lowers the percolation threshold of filler materials, resulting in high conductivity [308] . This novel and successful strategy have been used with ultrathin Au NWs (e.g., 2 nm) and high aspect ratio (e.g., >10,000) to provide outstanding performance for wearable sensors [185] with excellent biomedical capabilities [289] . Another strategy is the utilization of the imprinting technique to control the patterns and orientation of functional materials by template restriction [309] to assemble Au NPs in densely packed micro/NWs [290] . The effect of the adoption of these two strategies on sensor performance is illustrated in Figure 16 . Owing to their high electrical conductivity, biocompatibility, and exceptional mechanical properties, metal nanoparticles have rarely been applied in flexible pressure sensors [290] . The use of metal nanoparticles may be an opportunity for simple fabrication strategies that strike a balance between meeting the performance of medical applications without an increase in fabrication complexity and cost. Conducting polymers have tuneable conductivity and high processability through dispersion. They can play a pivotal role in tactile sensing by providing conductivity exceeding that of bulk metal electrodes [304] However, there is a need for investigation for new strategies to enhance mechanical properties and reusability further to meet the performance requirements of medical applications. Unlike nanoparticles, which are on the order of nanometres with different shapes (e.g., semi-round, nano-cubes, spheres) [290] , nanowires (NWs) are structures with a length typically on the order of microns and diameter on the order of tens of several hundred nanometres (i.e., high aspect ratio >1000) [305, 306] . NWs have excellent mechanical properties when they are scaled below 100 nm. However, their resistivity change in response to applied stresses [307] and this makes them favourable for the use in a wide range of applications of piezoresistivity. Ultralong 1D metal NWs can be used to increase sensitivity without affecting flexibility because of their high aspect ratio lowers the percolation threshold of filler materials, resulting in high conductivity [308] . This novel and successful strategy have been used with ultrathin Au NWs (e.g., 2 nm) and high aspect ratio (e.g., >10,000) to provide outstanding performance for wearable sensors [185] with excellent biomedical capabilities [289] . Another strategy is the utilization of the imprinting technique to control the patterns and orientation of functional materials by template restriction [309] to assemble Au NPs in densely packed micro/NWs [290] . The effect of the adoption of these two strategies on sensor performance is illustrated in Figure 16 . Owing to their high electrical conductivity, biocompatibility, and exceptional mechanical properties, metal nanoparticles have rarely been applied in flexible pressure sensors [290] . The use of metal nanoparticles may be an opportunity for simple fabrication strategies that strike a balance between meeting the performance of medical applications without an increase in fabrication complexity and cost. Sensors 2020, 20, x FOR PEER REVIEW 38 of 85 Figure 16 . Effect of adoption ultrathin NWs and imprinting to control patterns and orientation of functional materials on sensor performance. MOF materials are a class of chemical compounds consisting of metal ions or clusters coordinated to organic ligands to form one-dimensional structure, or two or three-dimensional structures [310] with the capability to form ultrahigh porous materials [311] , with up to 90% free volume, and a huge internal surface area extending 6000 m 2 /g [312] Although most of the MOF materials are generally associated with a lack of electrical conductivity [313] , a few have been reported as electronic conductors [313, 314] . MOF materials have been thoroughly studied for major interests such as heterogeneous catalysis and biomedical applications [315, 316] . MXene materials are two-dimensional transition metal carbides, carbonitrides and nitrides with the general formula Mn+1XnTx, where M represents an early transition metal, X is carbon and/or nitrogen, and Tx stands for surface terminations (e.g., OH, O, and F) [317] . In addition to the high flexibility of their interconnected networks in elastic matrices [318] , Mxene materials show a unique combination of metallic conductivity and hydrophilicity resulting from the free electrons of transition metal carbide or nitride backbone and Mxene materials surface terminations, respectively [317] that allow them to be used in different applications such as energy storage [319] and pressure sensors [250] . Increasing sensitivity without affecting flexibility is a common objective in tactile sensor design and fabrication. As illustrated in Figure 17 , several attempts have been performed utilizing different MOF materials such as the NW arrays of CuTCNQ at PI substrate [249] and C-MOF-PANIF at PU sponge [291] , MOF-5 derived porous carbon and PDMS composite [292] as well as two different types of MXene nanosheet materials by sandwiching porous MXene-impregnated tissue paper between a biodegradable PLA and IDE-coated PLA thin sheets [200] and by fabricating a flexible sensor based on Ti3C2-MXene with IDEs [250] . MOF materials are a class of chemical compounds consisting of metal ions or clusters coordinated to organic ligands to form one-dimensional structure, or two or three-dimensional structures [310] with the capability to form ultrahigh porous materials [311] , with up to 90% free volume, and a huge internal surface area extending 6000 m 2 /g [312] Although most of the MOF materials are generally associated with a lack of electrical conductivity [313] , a few have been reported as electronic conductors [313, 314] . MOF materials have been thoroughly studied for major interests such as heterogeneous catalysis and biomedical applications [315, 316] . MXene materials are two-dimensional transition metal carbides, carbonitrides and nitrides with the general formula M n+1 X n T x , where M represents an early transition metal, X is carbon and/or nitrogen, and Tx stands for surface terminations (e.g., OH, O, and F) [317] . In addition to the high flexibility of their interconnected networks in elastic matrices [318] , Mxene materials show a unique combination of metallic conductivity and hydrophilicity resulting from the free electrons of transition metal carbide or nitride backbone and Mxene materials surface terminations, respectively [317] that allow them to be used in different applications such as energy storage [319] and pressure sensors [250] . Increasing sensitivity without affecting flexibility is a common objective in tactile sensor design and fabrication. As illustrated in Figure 17 , several attempts have been performed utilizing different MOF materials such as the NW arrays of CuTCNQ at PI substrate [249] and C-MOF-PANIF at PU sponge [291] , MOF-5 derived porous carbon and PDMS composite [292] as well as two different types of MXene nanosheet materials by sandwiching porous MXene-impregnated tissue paper between a biodegradable PLA and IDE-coated PLA thin sheets [200] and by fabricating a flexible sensor based on Ti 3 C 2 -MXene with IDEs [250] . The NW arrays of CuTCNQ at PI substrate [249] shows a low limit of detectability with higher flexibility due to the high sensitivity and flexibility provided by the combination of MOF nanocrystal arrays and PI substrate. Other design and fabrication strategies, such as the C-MOF-PANIF at PU sponge [291] show higher sensitivity but slower response time. On the other side, the use of Ti 3 C 2 -MXene with IDEs [250] shows high sensitivity but relatively a higher limit of detection when compared with the work of Fu and co-workers, and Wang and co-workers. This high sensitivity might be due to the high strength of MXene materials as the effective Young's modulus of a single layer of Ti 3 C 2 Tx is 0.33 ± 0.03 TPa, which is the highest among the mean values reported in nanoindentation experiments for other solution-processed 2D materials [320] . The NW arrays of CuTCNQ at PI substrate [249] shows a low limit of detectability with higher flexibility due to the high sensitivity and flexibility provided by the combination of MOF nanocrystal arrays and PI substrate. Other design and fabrication strategies, such as the C-MOF-PANIF at PU sponge [291] show higher sensitivity but slower response time. On the other side, the use of Ti3C2-MXene with IDEs [250] shows high sensitivity but relatively a higher limit of detection when compared with the work of Fu and co-workers, and Wang and coworkers. This high sensitivity might be due to the high strength of MXene materials as the effective Young's modulus of a single layer of Ti3C2Tx is 0.33 ± 0.03 TPa, which is the highest among the mean values reported in nanoindentation experiments for other solution-processed 2D materials [320] . Ideally, materials used for the design and fabrication of an electrode could maintain excellent conductivity under large stretchability without affecting its sensing stability, as this plays a pivotal role in maintaining the high accuracy of the signals generated in response to changes in the applied pressure. The adoption of different design and fabrication strategies, as illustrated in figure 18 , may lead to a trade-off between the electrode performance and the sensor performance. Ideally, materials used for the design and fabrication of an electrode could maintain excellent conductivity under large stretchability without affecting its sensing stability, as this plays a pivotal role in maintaining the high accuracy of the signals generated in response to changes in the applied pressure. The adoption of different design and fabrication strategies, as illustrated in Figure 18 , may lead to a trade-off between the electrode performance and the sensor performance. Carbon materials such as CNTs and Graphene composites are preferred for use as electrode over generally used metals, including Ag, Au, and Cu, for their high conductivity and flexibility. [325] Nanocomposite-based electrodes can provide mechanical flexibility and high conductivity for flexible sensing devices, reduce the material cost of Nobel metal deposition, and complex fabrication (e.g. Lithography) and boost sustainability using environmentally friendly materials. [321] The conductive texture-based electrode provides an expandable tactile and strain sensor array. [324] Furthermore, their woven structured power-generating arrays can be integrated into a single multifactional device for piezoelectric energy harvesting and tactile sensing. The gauge factor of such sensors is much higher than that of the conventional metallic strain gauges. However, this gauge factor is less than that of a nanocomposite, for instance, made of MoS2/GF/Ecoflex. [284] Although this woven structure, as a strain sensor, shows acceptable linearity and sensitivity, it shows a Carbon materials such as CNTs and graphene composites are preferred for use as electrode over generally used metals, including Ag, Au, and Cu, for their high conductivity and flexibility [325] . Nanocomposite-based electrodes can provide mechanical flexibility and high conductivity for flexible sensing devices, reduce the material cost of Nobel metal deposition, and complex fabrication (e.g., lithography) and boost sustainability using environmentally friendly materials [321] . The onductive texture-based electrode provides an expandable tactile and strain sensor array [324] . Furthermore, their woven structured power-generating arrays can be integrated into a single multifactional device for piezoelectric energy harvesting and tactile sensing. The gauge factor of such sensors is much higher than that of the conventional metallic strain gauges. However, this gauge factor is less than that of a nanocomposite, for instance, made of MoS2/GF/Ecoflex [284] . Although this woven structure, as a strain sensor, shows acceptable linearity and sensitivity, it shows a hysteresis of a maximally 4.6% difference during stretching and releasing, whereas, a pressure sensor based percolative metal NPs arrays with Ag IDEs shows an insignificant level of hysteresis induced by a 1KPa applied pressure while maintaining a limit of detection and a sensitivity of 0.5 Pa and 0.13 kPa −1 , respectively [248] . Printed electrodes can provide tremendous physical, mechanical, and electrical properties, such as low resistivity and high flexibility, while exhibiting a low limit of detection under low voltage without compromising on reliability and durability [322, 326] . However, there are several challenges associated with this technology that include cost, synthesis, and choice of ink. For example, although both inkjet and screen printing technologies provide good flexibility at relatively low cost, both can offer a conductivity depending on the adhesion between the active materials and substrate that, in return, can be affected by several elements including the viscosity and temperature (i.e., curing) of the ink [327, 328] . The use of conducting polymers such as PEDOT: PSS seeks to bridge the gaps in the performance associated with other electrode materials, including poor stretchability, low conductivity, and little stability [323] . Though, the acidic nature of PEDOT: PSS (pH is between 1 and 2) can cause corrosion to any metallic interface with the PEDOT: PSS-based electrodes that can degrade the overall performance of the device [329] . Sensor-based metallic electrodes, as illustrated in (/unload cycles. These variations can be explained while taking into consideration the difference in the two dielectric and supporting materials applied, and the fabrication strategies followed, as well as the morphology and physical, mechanical, and electrical properties of the Ag used. Figure 19 shows different design and fabrication strategies, including the use of Pt, Ag, and some metal oxides. Using Pt coated polymeric nanofibers (i.e., nanohair) to fabricate a highly sensitive strain-gauge sensor, the work of Pang and co-workers [293] shows ultra-sensitivity with high flexibility when compared with that of the use of ITO electrode [294] , which shows a better response time and a lower limit of detection. In the work of Shuai and co-workers [297] and the work of Wu and co-workers [295] , tactile sensors are based on the use of Ag in both electrodes. However, they showed significant variations in flexibility in terms of the number of load/ unload cycles. These variations can be explained while taking into consideration the difference in the two dielectric and supporting materials applied, and the fabrication strategies followed, as well as the morphology and physical, mechanical, and electrical properties of the Ag used. Sensors 2020, 20, x FOR PEER REVIEW 37 of 81 mechanical, and electrical properties of the Ag used. Figure 19) , show different design and fabrication strategies, including the use of Pt, Ag, and some metal oxides. Using Pt coated Polymeric Nanofibers (i.e. Nanohair) to fabricate a highly sensitive strain-gauge sensor, the work of Pang and co-workers [293] shows ultra-sensitivity with high flexibility when compared with that of the use of ITO electrode, [294] which shows a better response time and a lower limit of detection. In the work of Shuai and co-workers, [297] and the work of Wu and co-workers, [295] the tactile sensors are based on the use of Ag in both electrodes. However, they showed significant variations in flexibility in terms Figure 19 . Performance of some sensors based on metallic electrodes. Wearable medical sensors are usually subject to scratches and mild damages that may limit their robustness and reduce their operational lifetime while affecting their mechanical and electrical Wearable medical sensors are usually subject to scratches and mild damages that may limit their robustness and reduce their operational lifetime while affecting their mechanical and electrical capabilities. Equally, they need to maintain clean and safe during all-day activities while support a green environment. Being challenged to take new steps in the design of novel material concepts, researchers have become attracted to electronics that can mimic human skin, enable health monitoring such that they can self-clean [330, 331] or self-heal after the wear and tear of every day, [332] [333] [334] [335] or under environmental stresses [336] [337] [338] and unpredictable damages [339] . Several approaches and strategies have been articulated to enable built-in capabilities for autonomous cleaning [340] [341] [342] , damage repair [343] [344] [345] [346] [347] [348] , diagnosis and reporting [349] , as well as healing [350] and degradation [351] . We believe that understanding external and internal factors affecting device performance throughout its operational life cycle can help reduce material wastes and repair costs, as well as increase efficiency, safety, and reliability, which are highly desirable in clinical applications. Capitalizing on the work of Patrick and coworkers [348] , we developed an operational lifecycle of a wearable sensor that expands to cover four essential functions: self-protection, self-diagnosis and reporting, self-healing, and self-degradation while providing self-cleaning perpetually ( Figure 20 illustrates strategies for autonomous functions throughout the operational lifecycle. Generally, self-cleaning strategies aim to resist microbiota adhesion to a surface, such that they can aggravate biofilm formation. [352] A biofilm is created when planktonic organisms adhere to a surface excreting a slimy, glue-like substance that enables adhesion, and forming shear-resistant, matrix-embedded multicellular communities. [353, 354] The use of antibiotics can largely aggravate biofilm formation, though, multi/pan-drug resistant infectious organisms have become a major public health issue. [355, 356] Besides, antibiotic is limited to bacterial infections only. [357] Inspired by the Lotus Effect, [358] material design that resists microbiota adhesion to a surface has become a powerful alternative strategy. [359] [360] [361] [362] Adhesion of microbiota to solid surfaces is a complex process; [363] hence, several material properties need to be considered to ensure an effective material design strategy. The following is a summary of such features: Surface Energy: Bacterial adhesion is affected by the physicochemical properties of the bacterial cell; hence, surface free energy is one of the most influencing factors of bacterial adhesion at the early stages of biofilm formation. [364] [365] Materials with relatively low surface free energy reduce the Generally, self-cleaning strategies aim to resist microbiota adhesion to a surface, such that they can aggravate biofilm formation [352] . A biofilm is created when planktonic organisms adhere to a surface excreting a slimy, glue-like substance that enables adhesion, and forming shear-resistant, matrix-embedded multicellular communities [353, 354] . The use of antibiotics can largely aggravate biofilm formation, though, multi/pan-drug resistant infectious organisms have become a major public health issue [355, 356] . Besides, antibiotics are limited to bacterial infections only [357] . Inspired by the Lotus Effect [358] , material design that resists microbiota adhesion to a surface has become a powerful alternative strategy [359] [360] [361] [362] . Adhesion of microbiota to solid surfaces is a complex process [363] ; hence, several material properties need to be considered to ensure an effective material design strategy. The following is a summary of such features: Surface Energy: Bacterial adhesion is affected by the physicochemical properties of the bacterial cell; hence, surface free energy is one of the most influencing factors of bacterial adhesion at the early stages of biofilm formation [364, 365] . Materials with relatively low surface free energy reduce the tendency of bacterial adhesion to surfaces when compared to that of higher surface energies [320, 366] . Indeed, this is almost always applicable to viruses since the wet and dry depositions of viruses are usually associated with either bacteria or particulate matter (PM) [367, 368] . However, the earlier the first interfacial adhesion is initiated, the sooner the effect of other forces take place [369] . Surface Chargeability: The cell surface charge is created from the dissociation or protonation of functional groups that include carboxylate, phosphate and amine moieties [370, 371] . With few exceptions, most bacterial and viral cell surfaces carry negative charges under certain physicochemical conditions [372, 373] . Hence, positively charged surfaces attract bacterial cells, and electrostatic repulsion disrupts cell contact with negatively charged surfaces. However, electrostatic repulsion disruption can be compromised by extracellular structures that promote adhesion, including fimbriae, flagella, curli, and pili [374] . Surface Roughness and Microtopography: The substrate topography at the micro/nanoscale does influence bacterial detachment behavior, particularly, the number of cells attached to the surface, and their orientation relative to the surface details [375, 376] . The irregularities on a surface promote bacterial adhesion and biofilm formation, while a smooth surface does not support bacterial adhesion and biofilm formation [377] . This bacterial adhesion can be explained through the large surface area produced by a rough surface in comparison with a smooth one. Surface Wettability: The wettability of either bacterial cells or the surface in contact does affect bacterial adhesion. Bacteria with a hydrophobic cell surface prefer hydrophobic material surfaces, whereas those with a hydrophilic cell surface prefer hydrophilic surfaces [331, 378] . Furthermore, bacteria adhesion can be mediated by the properties of the suspending medium, including surface tension, pH, ionic strength, and temperature [379] . Solid materials under several environmental conditions attract various organic and inorganic matters to adsorb to their surfaces before microbiota adhesion, forming a layer called the conditioning film [380] [381] [382] . Bacterial strains may influence the effect of conditioning films on bacterial adhesion, and this can significantly modify surfaces physicochemical and topographical properties, and subsequently lead to unpredictable deviations from the expected outcomes [380] . Developing mechanisms that consider the effect of surface properties on microbiota adhesion, many studies recommend different strategies to resist such an adhesion. However, more than one mechanism is probably involved at the same time, as these mechanisms have different ranges of action [380] . Hence, we summarize below some of these strategies based on experimental outcomes: The selection of substrates with surface free energy in the range 20-27 mN/m [330, 331] , including the use of coating materials such as silicone elastomer, perfluoroalkoxyalkane (PFA) and polytetrafluoroethylene (PTFE) [330] . More information is available in the work of Van der Mei and co-workers [383] . Also, surface free energy difference can provide an accurate and straightforward thermodynamic measure for quantitatively predicting bacterial adhesion [384] . The selection of hydrophobic substrates with increasing surface stiffness, including substrate surfaces made of PDMS in the range 2.4-2.8 MPa [340] [341] [342] . On more rigid polymer-brush coatings, Gram-positive bacteria (i.e., Staphylococcus epidermidis) desorb more readily. In contrast, softer polymer-brush coatings deform upon adhesion that strongly increases the normally oriented adhesion force, which impedes desorption, decreases the elastic modulus (i.e., stiffness) drastically, and increases the loss factor (i.e., viscous portion) [385, 386] . Shrink-induced superhydrophobic substrate surfaces, such as PE, PS, and PC, highly prevent E. coli bacteria growth [387] . However, the exposure of a substrate surface to a certain medium (e.g., PU to human urine) will render the surface more hydrophilic [388, 389] . Also, the adhesion behaviors of Gram-positive bacteria on wettable surfaces are crucially mediated by peptidoglycan (PGN) through direct interaction with the surfaces, but with the highest adhesion to surface when using superhydrophobic bioinspired hierarchal surfaces (e.g., a rose petal-like structure) [390] . Contrariwise, the adhesion behavior of Gram-negative bacteria on wettable surfaces does not consistently agree with that of the Gram-positive bacteria, but generally, they show low adhesion behavior when superhydrophobic surfaces were used [390] . Another potential factor is the sufficiency of the amount of the extracellular polysaccharides (EPS) that the bacteria form to anchor to a surface adequately. It has been observed that Gram-negative bacteria strains (i.e., Pseudomonas aeruginosa) were unable to colonize at a superhydrophobic titanium surface fabricated by femtosecond laser ablation. In contrast, Gram-positive bacteria strains (i.e., Staphylococcus aureus) were able to colonize the same surface structure [360] . The selection of substrates with smooth surfaces that have a minimum height difference between the highest peak and the lowest valley (i.e., Z range), including substrate surfaces made of PDMS with a Z-range of 15.5 and polystyrene 96-microwell plates with a z-range of 34.8 [379] . Changing the roughness and microtopography using an engineered surface of a sharkskin microstructure patterned on a PDMS substrate reduced S. aureus bacteria adhesion by 93.0% [391] and similarly, using an engineered surface of a sharkskin microstructure imprinted with TIO 2 NPs on a PET substrate reduced E. coli bacteria adhesion by 70.0-85.0% [392] . Other materials such as the superhydrophobic surface of a PS fibrous web hydrophobized by PFDTS vapor deposition with reduced solid area fraction displayed self-cleaning ability [393] Also, bacterial adhesion (i.e., Pseudomonas aeruginosa) to the surface of PS colloidal crystals (i.e., spheres of 1500 nanometers in diameter) was found to be less than that of PS colloidal crystals but with a smaller diameter (i.e., spheres of 450 nanometers) suggesting that the greater spacing between favorable sites on the 1500-nanometer colloidal crystal hindered the early-stage biofilm formation by separation of cell bodies [359] . The use of electrical current for electro-eradication/inactivation of bacteria including the use of pulsed electric field (PEF) as an inactivation approach due to its effect on the cell wall and coat architecture of Bacillus pumilus [394] and the use of a 5.0 mA direct current (DC) for more than 40 min in deionized water shows an over 99.999 % inactivation of E. coli and Staphylococcus aureus on a uniform conductive membrane of blended carboxyl-functionalized MWCNT with a polysulfone polymer [395] . Increasing device expected lifetime by implementing preventive measures protecting it against different causes of damage is the main objective for self-protection strategies. Coatings generally provide mechanical [396] , chemical [397] and weathering [398, 399] protection that increases resistance to wear, and this can be achieved through polymeric and smart coatings. Polymeric coatings (e.g., PDMS [400, 401] , PMMA [179, 402] and PVDF [403] [404] [405] ) can form organic thin films onto a substrate to significantly modify surface reactivity towards various elements including corrosion, adhesion, and wear [406] . Likewise, inorganic coatings (e.g., graphene [407] , Ag [408] and Au [409] ) can modify surface properties providing corrosion and damage protection. Organic coatings can leverage substrate corrosion and damage resistance. For example, the use of PMMA coating increased adhesion between graphene and PDMS substrate [410] and PVA coatings increased adhesion between GNS and PU yarn [411] . Also, coating, as a self-protection strategy, can extend device longevity by providing a self-degradation property. For instance, TiO 2 coatings are widely applied due to their safe use [412] and photocatalytic efficiency utilized in degrading organic pollutants into nontoxic inorganic molecules [413] , since their efficiency can be further enhanced in the presence of a piezoelectric field [414] . Besides, TiO 2 photocatalytic efficiency can be further utilized in photoinduced self-decomposition [415] . Silk and silk coatings can be applied to protect against mechanical and chemical degradation threats due to their nontoxicity [416] and exceptional, unique properties [417] [418] [419] . Since not long ago smart coatings have been increasingly applied to tactile sensors [420] where smart materials are designed to provide protection and remain passive unless promoted to perform a stimuli-based function when activated through certain changes within their environment such as magnetic fields, pH, light or temperature [421, 422] . While encapsulated-based and memory-shape assisted self-healing coatings are also self-protection strategies that we will discuss in more detail under self-healing [423, 424] , the microvascular network is a self-healing strategy, but it can also be applied to protect against thermal-induced stresses. The use of microvascular networks to support a thermal management capability is principally applied to preserve the stiffness-destroying effect of the high temperatures [425] . Tactile sensors are not usually subject to extreme temperatures. In contrast, their temperature is significantly limited to either the user's skin in contact with the wearable sensor or the external surroundings [426] , nevertheless, fluctuation in outdoor and indoor temperature, for instance, can lead to a significant change in the cardiac cycle, which affects blood pressure [427, 428] and equally, can induce hysteresis, which affects sensor sensitivity [429] . Therefore, microvascular networks may play a pivotal role in providing thermal stability to reduce/eliminate undesirable thermal hysteresis. Super-wettability is another effective strategy for self-protection. Super-wettability can provide self-cleaning [430, 431] , chemical-shielding [432, 433] and anti-corrosion [434, 435] effects, to name a few. For instance, partially embedded perfluorosilane-coated graphene into TPU by a dissolution and resolidification method can make the resulting composite withstand the abrasion by sandpaper for 20.0 m and a strain up to 400% without losing superhydrophobicity [436] . The principal objective for self-diagnosis and report strategy is to indicate that a device has experienced mechanical stress or damage [437] . Microcracks in polymeric materials are often difficult to detect, yet they can jeopardize the device performance leading to failure. Strategies that enhance detection of damage are necessary to improve functionality, ensure safety, and boost reliability [438] . Since their inception, chromogenic-based polymeric materials have been creating a fundamental interest in the development of materials that change their color in response to an external stimulus [439] . Chromogenic-based self-diagnosis and reporting strategies include mechanochromic [440] , thermochromic [441] and electrochromic [442] . Depending on the intensity of mechanical stress that a polymer can be subject to, the affected molecules within the polymeric chain can induce a chemical transformation in the responsive material (i.e., mechanophores) accompanied by a change in optical properties. Mechanophores include NITEC [443] and SP [444, 445] which changes color in bulk polymers from yellow to red [446] , as it has preferential mechanochemical activation in the stress direction [447] . Aside from mechanochromic, we found that less attention has paid to other chromogenic self-diagnosis and report strategies, particularly, the thermo-and electrochromic. Mechanophoreand mechanochemical-based sensors and probe molecules in the mechanochemistry research area have progressed considerably over the past years [448] [449] [450] [451] [452] [453] . In contrast, thermo-and electrochromic have advanced differently with less attention on thermochromic fluorescence [454] , thermochromic sensors [455] and electrochromic sensors [456] . Self-diagnosis and self-reporting, as well as, self-damage sensing that correlates the electrical conductivity of the CNTs or graphene flakes composite with damage severity [457] [458] [459] , or utilizes electrical resistance tomography mapping [460] can be further improved using the unique properties of thermo-and electrochromic polymeric materials. Finally, chromogenic based encapsulation can be applied by micro-and nanocapsules [461, 462] or microvascular networks (i.e., hollow glass fibers full of reactive liquids) to monitor the crack propagation. The main purpose of mechanisms for self-healing is to restore original mechanical and electrical capabilities without affecting functional performance [463] . Self-healing strategies can be grouped into different categories when considering strategies by which the healing mechanism is integrated into the targeted material. One categorization is based on the capability of the material to self-heal [464] , in which the automatic (i.e., intrinsic) self-healing materials do not need an external healing agent, and non-automatic (i.e., extrinsic) materials do require human intervention or may require external energy or pressure to trigger the self-healing mechanism [465, 466] . Another categorization is based on the damage length scale [348] . In this review, we will discuss self-healing strategies while considering self-healing mechanisms and their associated agents, if required, as well as enablers. Chemical-based self-healing is a strategy, which utilizes the built-in capabilities in the targeted material to initiate a chemical reaction whenever physical damage occurs. This strategy takes place at the smallest scale of the targeted material, where autonomous self-healing mechanisms based on chemical reactions are triggered through several classes, including a covalent bond [467] , supramolecular interaction [468] , hydrogen-bonding [469] , ionic-interactions [470] and π-π stacking [471] . Chemo-mechanical based self-healing is another strategy, which utilizes a rehealing agent within the host matrix (i.e., targeted material) to initiate a chemical reaction in response to mechanical damage [472] . Mainly, there are two chemo-mechanical self-healing classes; this includes encapsulation [423, 473] and microvascular networking [345, 474] . Other strategies utilize a re-healing agent to initiate a chemical reaction with external stimuli, such as; magnetic field [475, 476] , UV [477] and light and heat [478] as well as, memory shape-assisted self-healing induced by heat [479] and magnetic fields, and UV [480] . Table 5 summarizes a list of self-healing mechanisms along with some of their structure design properties. Regaining a high conductivity after the damage is crucial for sensor functionality, hence, different strategies were developed to ensure full recovery utilizing several self-healing mechanisms. For instance, graphene derivatives, particularly GO, offer various modifiable oxygen-based groups (e.g., OH-groups), and its conductivity can be further improved by reducing it into rGO [490, 491] . These properties can be considered before the design and fabrication of a self-healable graphene-based sensor. The OH-groups on GO can help form polymeric chains leading to a self-healable structure based on hydrogen-bonds [486, 492] , ionic interactions [493] and interfacial metal-ligand coordination [494] . Equally important is the GO content in the composite matrix since it affects thermal, electrical, and mechanical properties inversely. It was observed that thermal conductivity healing could be completed even at the lowest content, while electrical conductivity was low at low continents and became higher till the percolation limit. Then, mechanical healing gradually increased with content until it reached a maximal level of 70% [495] . CNTs are favorable in sensors due to their high electric conductivity and ease of functionalization [496, 497] . Hence, it is perfect for self-healable sensing structures based on metal-ligand coordination [498] , electrostatic interactions [499] , hydrogen bond [500, 501] In microcapsule-based self-healing mechanism [502] , healing temperature [503] and CNT content [504] affect the thermal, electrical, and mechanical properties differently. When using G-CNT altogether, the G-CNT heterostructures may induce a synergistic effect that can improve its tensile strength and self-healing property [505] . In addition to GO and CNT, different conductive materials, such as conductive polymer-based sensing structures, NWs and NPs, and MXenes are favorable to self-healing based on hydrogen bonds, ionic coordination interactions and physical entanglements (e.g., CPH(TOCNF/PAA-PPy) [506] ) and hydrogen-bond (e.g., Ag NWs [507] and MXene [508] ). The progress that has been achieved in self-healing is very promising, and though novel materials and innovative strategies in a more comprehensive approach are needed. For instance, wearable devices need to be healed and fully restored to their functional performance under all-day activities and standard working conditions (e.g., at skin temperature or during daylight) with minimal external interactions. Likewise, wearable devices need to be healed utilizing biocompatible and biodegradable materials that are safe, yet reliable, for both humans and the environment. The world produces as much as 50 million tons of electronic and electrical waste (e-waste) a year, and only 20% of this is formally recycled. For instance, there is 100 times more gold in a ton of e-waste than in a ton of gold ore [509] . In medical applications, the large motion-induced artefacts that occur from the loose coupling between wearable devices and a body part undermine device clinical relevance and accuracy [510] [511] [512] and increase the need for stable components of the wearable devices that safely interface with the skin without causing complications similar to contact dermatitis [143, 513] . It is thus urgent to develop environmentally friendly materials, of low cost, and high reusability meeting the required performance of electrical devices as the diverse and hazardous nature of this waste makes recycling difficult and expensive [514] . Degradability is important since self-protected and self-healable materials will continue performing until they ultimately reach the end of their operational life. Hence controlled degradation is not only required when the device performance has become unrestorable to its original state, such that self-healing is no more feasible but also is required when the device is needed for only a certain time. Therefore, the design and fabrication of robust devices that will decompose only when exposed to a certain stimulus are highly desirable. While degradable materials can be broken down into smaller constituent pieces, biodegradable materials can be broken into low molecular weight products as biologically benign or physiological conditions [515, 516] . The concept of biodegradability can be further expanded to cover biocompatibility, where biocompatible materials can fulfill their desired functions in contact with a living system without producing an adverse effect or deleterious changes [517] . Biodegradable and biocompatible materials, knowns as bioresorbable materials [518] , have attracted much attention in the past decade due to the potential environmental problems caused by traditional nonbiodegradable polymers, and the wide variety of application scenarios include their use in a wide variety of sensors to collect essential physiological information, such as blood pressure, strain, and temperature to name a few [519, 520] . Inorganic nanomaterials generally have low biodegradability that can be a challenge in biomedical applications while avoiding nanotoxicity as a result of inhalation or skin exposure [521] . For instance, indium tin oxide or tin-doped indium oxide (ITO) are widely used as excellent inorganic transparent electrical conductors. However, studies suggested that ITO nanoparticles (NPs) possess genotoxic potential on human lungs [522] , while others suggested ITO possesses toxicity on cells and organs [523, 524] and a material safety data sheet on ITO stated that ITO causes skin irritation and severe eye irritation [525] . Although a range of toxicological studies has been conducted assessing various physicochemical characteristics of nanomaterials, the results have been inconsistent and definitive rules cannot yet be established [521] and reviews on the potential for nanoparticles to penetrate the skin have been mixed [526] . For instance, silver nanoparticles (ranging from 9.8-48.8 nm) coated with PVP has been reported, in vitro, to penetrate through both intact and damaged full-thickness human skin [527] . Besides, shape dependency and active metallic facets of silver nanoparticles have been reported, both in vitro and in vivo, to exhibit diverse skin penetration. It has also been reported that triangular nanoparticles, which appear as 2 nm thick and equilateral triangle plates with an average side length of 50 nm exhibit the lowest skin penetration measured in terms of density and concertation of silver nanoparticles in blood, and rod-shaped nanoparticles, which appears with an average and diameter of 50 and 20 nm, respectively, exhibits the highest [528] . Unlike Ag NPs and ITO, MXenes are 2D inorganic compounds having low cytotoxicity and can be used in antibacterial applications [529] . Natural and synthetic polymers are perfectly suited for biodegradable and biocompatible electronics as they provide a wide variety of material properties and functions that can be tuned using chemical structural modifications allowing conformal contact with complete tissue surface at cost-effective fabrication strategies [530, 531] . Tables 6 and 7 summarise some properties of natural and synthetic polymers utilized in biodegradable and/or biocompatible sensors as substrates, dielectrics, active materials (i.e., conductors), or semiconductors. Interestingly, natural fabrics such as silk can be tuned to increase its susceptibility to specific sterilization techniques utilized to control its degradation rate [417] . Moreover, silk and silk coatings can provide corrosion resistance while performing as a smart coating [612] and likewise properties of some polymeric materials can be modified to offer longer shelf-life for end-use and allow the controlled degradation in response to UV light, for instance, in some special applications [613] . [185] or an electrode [587, 588] 2.0 CNTs Organic -103 − 6.7 × 106 [553] Yes [589] Yes [590] • CNT conductivity will highly depend on many parameters, such as structure and purity. • Likewise, CNTs biodegradability will depend on many parameters, including concentration, size, and functionalization. However, the remaining part of CNTs that stay in the body and accumulate may lead to unknown long-term effects [590] . • Electrical conductivity is for pristine Graphene. Conductivity will highly depend on many parameters, including the structure and quality of as-made materials. They are used as an excellent conductor in electromechanically based sensors [593] , and biosensors, [594] , as well as gas separation [595] . They are used as an active material [610] or an electrode [611] 6. Outlook Biomaterials ( Figure 21 ) have attracted considerable attention due to their exceptional performance and are, therefore, well-received as one of the future building blocks of digital healthcare. For blood pressure measurements and real-time monitoring, wearable sensors have not comprehensively reached a medically accepted level of functionality to replace intermittent cuff-based devices. This gap has been seemingly associated with a disconnect between the two knowledge areas, particularly, blood pressure measurement approaches and their related modalities at one side, and functional materials design and fabrication strategies and their relevant sensing mechanisms at the other one. Invasive and minimally invasive blood pressure measurement does not support continuous measurement and real-time monitoring. Hence, ECG and PPG signals are mostly used to monitor health status without disturbing the users during daily activities. It would be beneficial to incorporate more comprehensive information collected from multi-modality signals such as mechanical (i.e., piezoresistive), thermal (i.e., thermoresistive), or even humidity to further improve the performance of current wearable sensors rather than considering a single sensing principle as illustrated in this review. Apart from the rapid development demonstrated in this review, there is a bright future for tactile sensors utilizing smart materials (e.g., piezoelectric, self-healing, self-powering, self-cleaning), additive manufacturing, big data analytics (e.g., artificial intelligence) and cloud computing to fulfill healthcare demand for personalized medicine and remote monitoring. We found that sensitivity was the focus for most of the developed devices [247] [248] [249] 251, [257] [258] [259] [260] [274] [275] [276] [277] [278] [279] [280] [281] [282] [283] [284] [285] and those with piezoresistive mechanisms generally showed high performance when compared with others [242, 259, 271, 283, 293] . Although piezocapacitive-based sensors still show excellent detectability and sensitivity, they are more susceptible to noise resulting from field interaction and fringing capacitance, as well as, other factors such as temperature [192, 247, 294, 614] . FET-based sensors show high sensitivity with excellent response time due to their perfect functionalities of signal transduction and amplification, but their flexibility measured under continuous cycles of loading/unloading, is still a challenge [224, 269, 270, 278] while the best performing device of this category shows a tiny bending radius lower than 0.02 cm with no significant variation in its electrical characteristics after more than 200 cycles [209] a piezoresistive-based sensor with a similar sensitivity can show consistent performance under continuous cycles of loading/unloading on the order of thousands. Also, FET based sensors require relatively higher operating voltage when compared to that of piezoresistive. This variation in performance should be considered as a motive, not only for further development of existing sensing principles but also for further investigation of novel mechanisms to go beyond meeting high sensitivity requirements to include linear performance without hysteresis. The high sensitivity requirements have been achieved through the effective utilization of newly developed functional materials, such as PVDF/rGO [251] or ultrathin NWs [289] , the optimization of device geometry, such as the design inspired by insects' sensing capabilities [240] or plant leaves' morphology [242] or the microstructure of human skin [282] . Other strategies include the reduction of active materials concentration in nanocomposite matrix through alignment [260] , the utilization of novel additives manufacturing techniques such as LIG [271] , or the imprinting technique to control Invasive and minimally invasive blood pressure measurement does not support continuous measurement and real-time monitoring. Hence, ECG and PPG signals are mostly used to monitor health status without disturbing the users during daily activities. It would be beneficial to incorporate more comprehensive information collected from multi-modality signals such as mechanical (i.e., piezoresistive), thermal (i.e., thermoresistive), or even humidity to further improve the performance of current wearable sensors rather than considering a single sensing principle as illustrated in this review. Apart from the rapid development demonstrated in this review, there is a bright future for tactile sensors utilizing smart materials (e.g., piezoelectric, self-healing, self-powering, self-cleaning), additive manufacturing, big data analytics (e.g., artificial intelligence) and cloud computing to fulfill healthcare demand for personalized medicine and remote monitoring. We found that sensitivity was the focus for most of the developed devices [247] [248] [249] 251, [257] [258] [259] [260] [274] [275] [276] [277] [278] [279] [280] [281] [282] [283] [284] [285] and those with piezoresistive mechanisms generally showed high performance when compared with others [242, 259, 271, 283, 293] . Although piezocapacitive-based sensors still show excellent detectability and sensitivity, they are more susceptible to noise resulting from field interaction and fringing capacitance, as well as, other factors such as temperature [192, 247, 294, 614] . FET-based sensors show high sensitivity with excellent response time due to their perfect functionalities of signal transduction and amplification, but their flexibility measured under continuous cycles of loading/unloading, is still a challenge [224, 269, 270, 278] while the best performing device of this category shows a tiny bending radius lower than 0.02 cm with no significant variation in its electrical characteristics after more than 200 cycles [209] a piezoresistive-based sensor with a similar sensitivity can show consistent performance under continuous cycles of loading/unloading on the order of thousands. Also, FET based sensors require relatively higher operating voltage when compared to that of piezoresistive. This variation in performance should be considered as a motive, not only for further development of existing sensing principles but also for further investigation of novel mechanisms to go beyond meeting high sensitivity requirements to include linear performance without hysteresis. The high sensitivity requirements have been achieved through the effective utilization of newly developed functional materials, such as PVDF/rGO [251] or ultrathin NWs [289] , the optimization of device geometry, such as the design inspired by insects' sensing capabilities [240] or plant leaves' morphology [242] or the microstructure of human skin [282] . Other strategies include the reduction of active materials concentration in nanocomposite matrix through alignment [260] , the utilization of novel additives manufacturing techniques such as LIG [271] , or the imprinting technique to control the patterns and orientation of functional materials by template restriction [290] . The design and fabrication of wearable sensors with autonomous capabilities include wireless transmission [615, 616] self-powered [220, 617] , self-healed [485] and self-degraded [519] will certainly enable continuous diagnosis of cardiovascular disease. However, these features and similar ones are no longer considered desirable, but they are becoming as crucial as other performance measures since they will allow automatic reparation of device malfunctions and disposal. Furthermore, the integrated sensing modality of tactile, temperature, and humidity [420, 456, 618, 619] to extract additional features from other signals (e.g., respiratory) [620] with accuracy under the interface of strong body movement in real-time [621], will boost the performance of pressure measurements. The performance of wearable sensors, capitalizing on the recent advancement in machine learning and cloud computation, can be further boosted by selecting optimal features that can contribute to dynamic blood pressure changes. In return, this will provide an accurate blood pressure measurement noninvasively and continuously, help enable early prevention and personalized treatment of hypertension, and reduce its burden on society. Finally, there is an essential need for a multidisciplinary approach encompasses of different knowledge areas, mainly, data, material, medical, and engineering sciences to ensure seamless integration between various sensor components and architecture, and address other challenges associated with sensor overall performance, as well as, evaluate the impact of each on device reliability and efficiency. Figure 22 summarises future development in blood pressure measurement and real-time monitoring. Sensors 2020, 20, x FOR PEER REVIEW 56 of 85 fabrication of wearable sensors with autonomous capabilities include wireless transmission [615, 616] self-powered [220, 617] , self-healed [485] and self-degraded [519] will certainly enable continuous diagnosis of cardiovascular disease. However, these features and similar ones are no longer considered desirable, but they are becoming as crucial as other performance measures since they will allow automatic reparation of device malfunctions and disposal. Furthermore, the integrated sensing modality of tactile, temperature, and humidity [420, 456, 618, 619] to extract additional features from other signals (e.g., respiratory) [620] with accuracy under the interface of strong body movement in real-time [621], will boost the performance of pressure measurements. The performance of wearable sensors, capitalizing on the recent advancement in machine learning and cloud computation, can be further boosted by selecting optimal features that can contribute to dynamic blood pressure changes. In return, this will provide an accurate blood pressure measurement noninvasively and continuously, help enable early prevention and personalized treatment of hypertension, and reduce its burden on society. Finally, there is an essential need for a multidisciplinary approach encompasses of different knowledge areas, mainly, data, material, medical, and engineering sciences to ensure seamless integration between various sensor components and architecture, and address other challenges associated with sensor overall performance, as well as, evaluate the impact of each on device reliability and efficiency. Figure 22 summarises future development in blood pressure measurement and real-time monitoring. Funding: This research received no external funding. Funding: This research received no external funding. The authors declare no conflict of interest. World Health Organization. World Health Organization Cardiovascular Diseases (CVDs) Report; World Health Organization (WHO) COVID-19 Surveillance Group Characteristics of COVID-19 Patients dying in Italy Report Based on Available Data on World Health Organization (WHO)-China Joint Mission Statement on the Need for Increased Access to Telehealth to Combat Community Spread of COVID-19 Managing emerging infectious diseases: Should travel be the fifth vital sign? ESH/ESC Guidelines for the management of arterial hypertension: The Task Force for the management of arterial hypertension of the European Society of Hypertension (ESH) and of the European Society of Cardiology (ESC) Mechanisms of plaque formation and rupture Early Detection of Cardiac Impairment and Prediction of RV Hypertrophy in Patients With CTD Early detection and management of the high-risk patient with elevated blood pressure. Vasc. Health Risk Manag The importance of accurate blood pressure measurement Clinical Requirements of Future Patient Monitoring in the Intensive Care Unit: Qualitative Study Point-of-care testing based on smartphone: The current state-of-the-art Single-layered ultra-soft washable smart textiles for all-around ballistocardiograph, respiration, and posture monitoring during sleep The application of personal glucose meters as universal point-of-care diagnostic tools Point-of-care approaches for cancer diagnostics Automatic smartphone-based microfluidic biosensor system at the point of care General-purpose passive wireless point-of-care platform based on smartphone A smartphone-integrated ratiometric fluorescence sensing platform for visual and quantitative point-of-care testing of tetracycline Nanomaterials-based enzyme electrochemical biosensors operating through inhibition for biosensing applications Moving toward Precision Therapy Multiclass classifier based cardiovascular condition detection using smartphone mechanocardiography The Vision of Industrie 4.0 in the making-a case of future told, tamed, and traded High-performance medicine: The convergence of human and artificial intelligence A new paradigm in the treatment of the cardiovascular disease continuum: Focus on prevention Clinical decision-making of cardiologists regarding admission and treatment of patients with suspected unstable angina or non-ST-elevation myocardial infarction: Protocol of a clinical vignette study Secondary prevention of cardiovascular diseases and application of technology for early diagnosis Preventing heart disease in the 21st century: Implications of the pathobiological determinants of Atherosclerosis in youth (PDAY) study Bio-compatible cyber-physical system for cloud-based customizable sensor monitoring of pressure conditions Electronic Skin: Achievements, Issues and Trends tactile image sensors employing camera: A review Skin sensors are the future of health care Accessing analytes in biofluids for peripheral biochemical monitoring Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care Nanomaterial-enabled wearable sensors for healthcare Bio-Integrated wearable systems: A comprehensive review Wearables in medicine Estimating Accuracy at exercise intensities: A comparative study of self-monitoring heart rate and physical activity wearable devices Wearable devices in clinical trials: Hype and hypothesis Investigating sources of inaccuracy in wearable optical heart rate sensors Clinical assessment of a non-invasive wearable MEMS pressure sensor array for monitoring of arterial pulse waveform, heart rate and detection of atrial fibrillation Recommendations for assessment of the reliability, sensitivity, and validity of data provided by wearable sensors designed for monitoring physical activity The meaning of blood pressure McDonald's Blood Flow in Arteries: Theoretical, Experimental and Clinical Principles Elastic properties and Windkessel function of the human aorta. Cardiovasc. Drugs Ther Principles and techniques of blood pressure measurement Ballistocardiogram: Mechanism and potential for unobtrusive cardiovascular health monitoring A novel heart sound activity detection framework for automated heart sound analysis Reconfigurable architecture for multi-lead ecg signal compression with high-frequency noise reduction Central blood pressure: Current evidence and clinical importance Accuracy of invasive arterial pressure monitoring in cardiovascular patients: An observational study Infusion thrombophlebitis: A prospective comparison of 645 Vialon and Teflon cannulae in anaesthetic and postoperative use In vitro biocompatibility of a polyurethane catheter after deposition of fluorinated film Infection risks associated with peripheral vascular catheters Medical grade tubing: Criteria for catheter applications Silicone and polyurethane tunneled infusion catheters: A comparison of durability and breakage rates Catheter-related thrombosis: A practical approach Electrostatic MEMS microphones Design and optimization of a three-terminal piezoresistive pressure sensor for catheter based in vivo biomedical applications A novel ultra-miniature catheter tip pressure sensor fabricated using silicon and glass thinning techniques Polyimide/SU-8 catheter-tip MEMS gauge pressure sensor MEMS-based flexible force sensor for tri-axial catheter contact force measurement PVF2catheter-tip transducers for pressure, sound and flow measurements Biodegradable and flexible arterial-pulse sensor for the wireless monitoring of blood flow Machine learning methods for real-time blood pressure measurement based on photoplethysmography Gold standard is an appropriate term Accuracy of non-invasive and minimally invasive hemodynamic monitoring: Where do we stand? Continuous in vivo blood pressure measurements using a fully implantable wireless SAW sensor An implantable intraocular pressure transducer: Initial safety outcomes Advances in intracranial pressure monitoring and its significance in managing traumatic brain injury A stainless-steel-based implantable pressure sensor chip and its integration by microwelding Bioresorbable optical sensor systems for monitoring of intracranial pressure and temperature Long term, implantable blood pressure monitoring systems From Korotkoff and Marey to automatic non-invasive oscillometric blood pressure measurement: Does easiness come with reliability? A model-based retrospective analysis of the fixed-ratio oscillometric blood pressure measurement Jamshed, D. Palpatory method of measuring diastolic blood pressure Convenient measurement of systolic pressure: The reliability and validity of manual radial pulse pressure measurement Clinical Investigation of Intermittent Automated Measurement Type Effect of cuff design on auscultatory and oscillometric blood pressure measurements Impact of cuff positioning on blood pressure measurement accuracy: May a specially designed cuff make a difference? Review: A century of confusion; which bladder for accurate blood pressure measurement? A new double cuff sphygmotonometer for accurate blood pressure measurement Measurement of blood pressure during pregnancy: Evaluation of the 'TriCUFF' Accuracy of a single rigid conical cuff with standard-size bladder coupled to an automatic oscillometric device over a wide range of arm circumferences Blood Pressure Cuff and to a Method of Making the Same Disposable Medical Pressure Cuffs and Method of Production Recyclable or Biodegradable Blood Pressure Cuff High durability and low toxicity antimicrobial coatings fabricated by quaternary ammonium silane copolymers Noninvasive intraocular pressure monitoring: Current insights Method of deflection corrected tonometry with phantom vessel experiments Noninvasive validation of central and peripheral augmentation index estimated by a novel wrist-worn tonometer Development of a palpable carotid pulse pressure sensor using electromagnetic induction Measuring intraocular pressure Comparison of Goldmann applanation tonometry, rebound tonometry and dynamic contour tonometry in normal and glaucomatous eyes Digest of technical papers (Cat. No.03TH8664). In A Soft Contact Lens with a MEMS Strain Gage Embedded For Intraocular Pressure Monitoring Soft wearable contact lens sensor for continuous intraocular pressure monitoring High resolution non-invasive intraocular pressure monitoring by use of graphene woven fabrics on contact lens Cantilever arrayed blood pressure sensor for arterial applanation tonometry Development of a multi-array pressure sensor module for radial artery pulse wave measurement Pulse wave analysis with two tonometric devices: A comparison study Oscillometric blood pressure measurement used for calibration of the arterial tonometry method contributes significantly to error A novel minimally occlusive cuff method utilizing ultrasound vascular imaging for stress-free blood pressure measurement: A-proof-of-concept study Noninvasive cardiac output monitoring in cardiothoracic surgery patients: Available methods and future directions Measurement of blood pressure Hydrostatic Finger Cuff for Blood Wave Formanalysis And Dagnostic Support Removable Polyurethane Adhesives with Improved Temperature Resistance Properties Disposable and Detachable Sensor For Continuous Non-Invasive Arterial Blood Pressure Monitoring Innovation for Noninvasive Hemodynamic Management; Edwards Lifesciences The blood supply of various skin areas as estimated by the photoelectric plethysmograph Continuous blood pressure measurement from invasive to unobtrusive: Celebration of 200th birth anniversary of Carl Ludwig Passive wearable skin patch sensor measures limb hemodynamics based on electromagnetic resonance Do the Finapres and Colin radial artery tonometer measure the same blood pressure changes following deflation of thigh cuffs? Non-invasive continuous arterial pressure monitoring with Nexfin does not sufficiently replace invasive measurements in critically ill patients Comparing volume-clamp method and intra-arterial blood pressure measurements in patients with atrial fibrillation admitted to the intensive or medium care unit Simulation of blood flow in deformable vessels using subject-specific geometry and spatially varying wall properties Beyond blood pressure: Pulse wave analysis-a better way of assessing cardiovascular risk? Future Cardiol Cardiac output, oxygen consumption and arteriovenous oxygen difference following a sudden rise in exercise level in humans Passive self resonant skin patch sensor to monitor cardiac intraventricular stroke volume using electromagnetic properties of blood Monitoring of the central blood pressure waveform via a conformal ultrasonic device ESC guidelines on the management of stable coronary artery disease: The Task Force on the management of stable coronary artery disease of the European Society of Cardiology Relation between blood pressure and pulse wave velocity for human arteries Comprehensive assessment of coronary artery stenoses: Computed tomography coronary angiography versus conventional coronary angiography and correlation with fractional flow reserve in patients with stable angina Clinical outcomes of fractional flow reserve by computed tomographic angiography-guided diagnostic strategies vs. usual care in patients with suspected coronary artery disease: The prospective longitudinal trial of FFR(CT): Outcome and resource impacts study Diagnostic performance of noninvasive fractional flow reserve derived from coronary computed tomography angiography in suspected coronary artery disease: The NXT trial (Analysis of coronary blood flow using ct angiography: Next steps) Diagnostic Accuracy of computed tomography-derived fractional flow reserve: A systematic review Optimal electrocardiographic lead systems: Practical scenarios in smart clothing and wearable health systems Flexible polymeric dry electrodes for the long-term monitoring of ECG Comfortable textile-based electrode for wearable electrocardiogram Development of printed and flexible dry ECG electrodes Bonfiglio, A. fully textile, pedot:pss based electrodes for wearable ecg monitoring systems Skin conformal polymer electrodes for clinical ECG and EEG recordings Van Hoof, C. Soft, comfortable polymer dry electrodes for high quality ECG and EEG recording Accuracy and precision of non-invasive cardiac output monitoring devices in perioperative medicine: A systematic review and meta-analysisdagger Toward Ubiquitous Blood Pressure Monitoring via Pulse Transit Time: Theory and Practice Pulse transit time technique for cuffless unobtrusive blood pressure measurement: From theory to algorithm Diagnostic approach in allergic and irritant contact dermatitis Guideline contact dermatitis: S1-Guidelines of the German Contact Allergy Group (DKG) of the German Dermatology Society (DDG), the Information Network of Dermatological Clinics (IVDK), the German Society for Allergology and Clinical Immunology (DGAKI), the Working Group for Occupational and Environmental Dermatology (ABD) of the DDG, the Medical Association of German Allergologists (AeDA), the Professional Association of German Dermatologists (BVDD) and the DDG Advanced materials for health monitoring with skin-based wearable devices Unobtrusive nocturnal heartbeat monitoring by a ballistocardiographic sensor in patients with sleep disordered breathing A Chair-based unconstrained/nonintrusive cuffless blood pressure monitoring system using a two-channel ballistocardiogram A chest-laminated ultrathin and stretchable e-tattoo for the measurement of electrocardiogram, seismocardiogram, and cardiac time intervals Epidermal Electronic Systems for Sensing and Therapy Study on Ballistocardiogram Acquisition in a Moving Wheelchair with Embedded Sensors Refinement and evaluation of a hydraulic bed sensor Development of a piezoelectric polyvinylidene fluoride (PVDF) polymer-based sensor patch for simultaneous heartbeat and respiration monitoring Textrode-enabled transthoracic electrical bioimpedance measurements-towards wearable applications of impedance cardiography A direct comparison of wet, dry and insulating bioelectric recording electrodes Wearable sensors for ECG measurement: A review Textile-based capacitive sensors for respiration monitoring Acoustic sensing as a novel wearable approach for cardiac monitoring at the wrist Non-invasive electromagnetic skin patch sensor to measure intracranial fluid-volume shifts Method for continuous nondisturbing monitoring of blood pressure by magnetoelastic skin curvature sensor and ECG Quantitative hemoglobin tomography with diffuse near-infrared spectroscopy: Pilot results in the breast Optical blood pressure estimation with photoplethysmography and FFT-based neural networks Skin photoplethysmography-A review Wearable Photoplethysmographic Sensors-Past and Present Noninvasive pulse oximetry utilizing skin reflectance photoplethysmography High Signal-to-noise ratio phonocardiogram using a shielded pvdf film sensor A flexible tonoarteriography-based body sensor network for cuffless measurement of arterial blood pressure Calibration and validation of homeostasis parameters estimates produced by a DSP embedded in a wheelchair Pulse transit time measurement using seismocardiogram, photoplethysmogram, and acoustic recordings: Evaluation and comparison Highly skin-conformal microhairy sensor for pulse signal amplification Flexible piezoresistive sensor patch enabling ultralow power cuffless blood pressure measurement Mapping the human body for vibrations using an accelerometer Conformable amplified lead zirconate titanate sensors with enhanced piezoelectric response for cutaneous pressure monitoring Non-invasive blood pressure estimation from ecg using machine learning techniques Design and implementation of a portable impedance cardiography system for noninvasive stroke volume monitoring Flexible organic/inorganic hybrid near-infrared photoplethysmogram sensor for cardiovascular monitoring Highly stretchable resistive pressure sensors using a conductive elastomeric composite on a micropyramid array Introduction to Percolation Theory Design and application of a high sensitivity piezoresistive pressure sensor for low pressure conditions Laser-Printed, Flexible Graphene Pressure Sensors Highly sensitive wearable 3D piezoresistive pressure sensors based on graphene coated isotropic non-woven substrate Carbon nanotube-based piezoresistive sensors fabricated by microwave irradiation Nano carbon black-based high performance wearable pressure sensors ZnO nanorods/carbon black-based flexible strain sensor for detecting human motions High Resolution E-Jet Printed Temperature Sensor on Artificial Skin Highly conductive, stretchable and biocompatible Ag-Au core-sheath nanowire composite for wearable and implantable bioelectronics High sensitivity, wearable, piezoresistive pressure sensors based on irregular microhump structures and its applications in body motion sensing Highly sensitive, self-powered and wearable electronic skin based on pressure-sensitive nanofiber woven fabric sensor Microchannel-confined mxene based flexible piezoresistive multifunctional micro-force sensor Hydrophobic and stable mxene-polymer pressure sensors for wearable electronics Tunable-Sensitivity flexible pressure sensor based on graphene transparent electrode Graphene-based three-dimensional capacitive touch sensor for wearable electronics Graphene oxide as high-performance dielectric materials for capacitive pressure sensors Eraser-based eco-friendly fabrication of a skin-like large-area matrix of flexible carbon nanotube strain and pressure sensors Highly sensitive, flexible, and wearable pressure sensor based on a giant piezocapacitive effect of three-dimensional microporous elastomeric dielectric layer Natural plant materials as dielectric layer for highly sensitive flexible electronic skin A flexible and highly sensitive capacitive pressure sensor based on conductive fibers with a microporous dielectric for wearable electronics High-Performance paper-based capacitive flexible pressure sensor and its application in human-related measurement Self-Powered human-health monitoring through aligned pvdf nanofibers interfaced skin-interactive piezoelectric sensor A Wearable transient pressure sensor made with mxene nanosheets for sensitive broad-range human-machine interfacing Transparent, optical, pressure-sensitive artificial skin for large-area stretchable electronics Vital sign measurement using fbg sensor for new wearable sensor development, fiber optic sensing-principle, measurement and applications. Shien-Kuei Liaw Fiber Bragg grating-based sensor for monitoring respiration and heart activity during magnetic resonance imaging examinations Wireless, portable fiber bragg grating interrogation system employing optical edge filter Magnetic resonance imaging compatible non-invasive fibre-optic sensors based on the bragg gratings and interferometers in the application of monitoring heart and respiration rate of the human body: A comparative study Cardio-Respiratory monitoring in archery using a smart textile based on flexible fiber bragg grating sensors Piezoelectric graphene field effect transistor pressure sensors for tactile sensing Electronic biosensing with flexible organic transistor devices Ultrathin, flexible and multimodal tactile sensors based on organic field-effect transistors User-interactive electronic skin for instantaneous pressure visualization Flexible field-effect transistor-type sensors based on conjugated molecules Reviving vibration energy harvesting and self-powered sensing by a triboelectric nanogenerator Large-Scale and washable smart textiles based on triboelectric nanogenerator arrays for self-powered sleeping monitoring Progress in triboelectric nanogenerators as self-powered smart sensors Wireless textile-based sensor system for self-powered personalized health care Eardrum-inspired active sensors for self-powered cardiovascular system characterization and throat-attached anti-interference voice recognition Membrane-Based self-powered triboelectric sensors for pressure change detection and its uses in security surveillance and healthcare monitoring Triboelectric nanogenerator enabled body sensor network for self-powered human heart-rate monitoring Epidermis-Inspired ultrathin 3d cellular sensor array for self-powered biomedical monitoring Flexible weaving constructed self-powered pressure sensor enabling continuous diagnosis of cardiovascular disease and measurement of cuffless blood pressure Shape memory polymer-based self-healing triboelectric nanogenerator Triboelectric generators and sensors for self-powered wearable electronics Transparent triboelectric nanogenerators and self-powered pressure sensors based on micropatterned plastic films Sliding-triboelectric nanogenerators based on in-plane charge-separation mechanism The rise of plastic bioelectronics Flexible, stretchable sensors for wearable health monitoring: Sensing mechanisms, materials, fabrication strategies and features Recent advances in tactile sensing technology. Micromachines Integration of heterogeneous materials for wearable sensors High strain biocompatible polydimethylsiloxane-based conductive graphene and multiwalled carbon nanotube nanocomposite strain sensors In tunable and wearable high performance strain sensors based on laser patterned graphene flakes SWEATCH': A wearable platform for harvesting and analysing sweat sodium content Lancet-free and label-free diagnostics of glucose in sweat using Zinc Oxide based flexible bioelectronics Pencil-on-paper: Electronic devices Textile building blocks: Toward simple, modularized, and standardized smart textile Wearable contactless respiration sensor based on multi-material fibers integrated into textile Silver nanowires coated on cotton for flexible pressure sensors Development of a flexible PDMS capacitive pressure sensor for plantar pressure measurement Highly sensitive flexible three-axis tactile sensors based on the interface contact resistance of microstructured graphene Carbon nanotube based elastomer composites -an approach towards multifunctional materials Large-Area compliant, low-cost, and versatile pressure-sensing platform based on microcrack-designed carbon black @ polyurethane sponge for human-machine interfacing 3D Printed stretchable tactile sensors CVD growth of fingerprint-like patterned 3D graphene film for an ultrasensitive pressure sensor Microstructured graphene arrays for highly sensitive flexible tactile sensors Skin-like pressure and strain sensors based on transparent elastic films of carbon nanotubes Flexible polymer transistors with high pressure sensitivity for application in electronic skin and health monitoring Spreading properties of cosmetic emollients: Use of synthetic skin surface to elucidate structural effect Flexible, tunable, and ultrasensitive capacitive pressure sensor with microconformal graphene electrodes An ultrahigh resolution pressure sensor based on percolative metal nanoparticle arrays Solution-processed large-area nanocrystal arrays of metal-organic frameworks as wearable, ultrasensitive, electronic skin for health monitoring A highly flexible and sensitive piezoresistive sensor based on MXene with greatly changed interlayer distances An ultra-sensitive and rapid response speed graphene pressure sensors for electronic skin and health monitoring Carbon Black as a Polymer Filler Electrical properties of carbon nanotube based fibers and their future use in electrical wiring Flexible CMOS integrated circuits based on carbon nanotubes with sub-10 ns stage delays Effect of carbon nanotube aspect ratio on the thermal and electrical properties of epoxy nanocomposites. Fuller. Nanotubes Carbon Nanostruct Highly exfoliated MWNT-RGO ink-wrapped polyurethane foam for piezoresistive pressure sensor applications Reverse-micelle-induced porous pressure-sensitive rubber for wearable human-machine interfaces Giant tunneling piezoresistance of composite elastomers with interlocked microdome arrays for ultrasensitive and multimodal electronic skins Flexible and highly sensitive pressure sensors based on bionic hierarchical structures Wearable resistive pressure sensor based on highly flexible carbon composite conductors with irregular surface morphology Transparent electronics based on transfer printed aligned carbon nanotubes on rigid and flexible substrates Aligned multi-walled carbon nanotube-reinforced composites: Processing and mechanical characterization Aligned carbon nanotube-epoxy composites: The effect of nanotube organization on strength, stiffness, and toughness Graphene: Status and prospects Graphene based materials: Past, present and future Ultra-thin membranes for sensor applications Stepwise Reduction of Graphene Oxide (GO) and its effects on chemical and colloidal properties Transparent, Sensitive, and Crosstalk-Free Capacitive Tactile Sensor Array Based on Graphene Electrodes and Air Dielectric Mechanosensation-active matrix based on direct-contact tribotronic planar graphene transistor array Integrated arrays of air-dielectric graphene transistors as transparent active-matrix pressure sensors for wide pressure ranges Fabrication of low-cost and highly sensitive graphene-based pressure sensors by direct laser scribing polydimethylsiloxane Silk-molded flexible, ultrasensitive, and highly stable electronic skin for monitoring human physiological signals A flexible and highly pressure-sensitive graphene-polyurethane sponge based on fractured microstructure design Graphene based piezoresistive pressure sensor Graphene-Paper pressure sensor for detecting human motions Highly sensitive and flexible tactile sensor based on porous graphene sponges for distributed tactile sensing in monitoring human motions Capacitive pressure sensing with suspended graphene-polymer heterostructure membranes Graphene tribotronics for electronic skin and touch screen applications A flexible pressure sensor based on rGO/polyaniline wrapped sponge with tunable sensitivity for human motion detection Reduced graphene oxide-based double network polymeric hydrogels for pressure and temperature sensing Multi-dimensional flexible reduced graphene oxide/polymer sponges for multiple forms of strain sensors Skin-inspired flexible and high-sensitivity pressure sensors based on rGO films with continuous-gradient wrinkles An ultrasensitive flexible pressure sensor for multimodal wearable electronic skins based on large-scale polystyrene ball@reduced graphene-oxide core-shell nanoparticles Highly sensitive and flexible strain-pressure sensors with cracked paddy-shaped mos2/graphene foam/ecoflex hybrid nanostructures Piezoresistive pressure sensor based on synergistical innerconnect polyvinyl alcohol nanowires/wrinkled graphene film Wearable high-performance pressure sensors based on three-dimensional electrospun conductive nanofibers High performance piezoelectric devices based on aligned arrays of nanofibers of poly(vinylidenefluoride-co-trifluoroethylene) A nanofiber based artificial electronic skin with high pressure sensitivity and 3D conformability A wearable and highly sensitive pressure sensor with ultrathin gold nanowires Gold nanoparticle densely packed micro/nanowire-based pressure sensors for human motion monitoring and physiological signal detection A wearable breathable pressure sensor from metal-organic framework derived nanocomposites for highly sensitive broad-range healthcare monitoring Multifunctional sensor based on porous carbon derived from metal-organic frameworks for real time health monitoring A flexible and highly sensitive strain-gauge sensor using reversible interlocking of nanofibres High sensitivity flexible capacitive pressure sensor using polydimethylsiloxane elastomer dielectric layer micro-structured by 3-D printed mold Theoretical study and structural optimization of a flexible piezoelectret-based pressure sensor Highly sensitive flexible pressure sensors with microstructured rubber dielectric layers Highly sensitive flexible pressure sensor based on silver nanowires-embedded polydimethylsiloxane electrode with microarray structure Laser-induced porous graphene films from commercial polymers Highly sensitive textile strain sensors and wireless user-interface devices using all-polymeric conducting fibers The influence of chain structure on the equilibrium melting temperature of poly(vinylidene fluoride) Analysis Method: FTIR studies of β-phase crystal formation in stretched PVDF films Semi-crystalline fluorinated polymers A Percolation Model for Piezoresistivity in Conductor-Polymer Composites Conducting polymer electrodes for electroencephalography Synthesis of silver nanoparticles with different shapes III-V nanowires and nanowire optoelectronic devices Riel, H. manipulating surface states of iii-v nanowires with uniaxial stress Highly stretchable and highly conductive metal electrode by very long metal nanowire percolation network Plasmonic nano-printing: Large-area nanoscale energy deposition for efficient surface texturing Mechano-regulated metal-organic framework nanofilm for ultrasensitive and anti-jamming strain sensing Design and synthesis of an exceptionally stable and highly porous metal-organic framework Introduction to metal-organic frameworks Integration of a (-Cu-S-)n plane in a metal-organic framework affords high electrical conductivity Nanostructuration of PEDOT in porous coordination polymers for tunable porosity and conductivity Engineering Metal-organic framework catalysts for C-C and C-X coupling reactions: Advances in reticular approaches from Metal-organic framework-based nanomaterials for biomedical applications Electronic and optical properties of 2D transition metal carbides and nitrides (MXenes) Flexible, stretchable and electrically conductive MXene/natural rubber nanocomposite films for efficient electromagnetic interference shielding 2D metal carbides and nitrides (MXenes) for energy storage Elastic properties of 2D Ti3C2T x MXene monolayers and bilayers Ionogel infiltrated paper as flexible electrode for wearable all-paper based sensors in active and passive modes Low-voltage, high-sensitivity and high-reliability bimodal sensor array with fully inkjet-printed flexible conducting electrode for low power consumption electronic skin Highly conductive stretchable all-plastic electrodes using a novel dipping-embedded transfer method for high-performance wearable sensors and semitransparent organic solar cells Multifunctional woven structure operating as triboelectric energy harvester, capacitive tactile sensor array, and piezoresistive strain sensor array One-Step Fabrication of Stretchable Copper Nanowire Conductors by a Fast Photonic Sintering Technique and Its Application in Wearable Devices Flexible, transparent patterned electrodes based on graphene oxide/silver nanowire nanocomposites fabricated utilizing an accelerated ultraviolet/ozone process to control silver nanowire degradation High-resolution patterning of graphene by screen printing with a silicon stencil for highly flexible printed electronics Flexible electronics based on micro/nanostructured paper Functional solid additive modified PEDOT:PSS as an anode buffer layer for enhanced photovoltaic performance and stability in polymer solar cells Surface free energy effect on bacterial retention Physico-chemistry of bacterial transmission versus adhesion Intrinsically stretchable and healable semiconducting polymer for organic transistors Mimicking human and biological skins for multifunctional skin electronics Preparation of artificial skin that mimics human skin surface and mechanical properties Wettability manipulation of overflow behavior via vesicle surfactant for water-proof surface cleaning Route towards sustainable smart sensors: Ferroelectric polyvinylidene fluoride-based materials and their integration in flexible electronics Organic Semiconductors-Based Devices Electrical Reliability to Environmental Stress Multifunctional skin-inspired flexible sensor systems for wearable electronics Emerging flexible and wearable physical sensing platforms for healthcare and biomedical applications Stiffness of cross-linked poly(dimethylsiloxane) affects bacterial adhesion and antibiotic susceptibility of attached cells How bacteria respond to material stiffness during attachment: A role of escherichia coli flagellar motility Cyclic-di-GMP and oprF are involved in the response of pseudomonas aeruginosa to substrate material stiffness during attachment on polydimethylsiloxane (PDMS) Self-healing polymer composites: Prospects, challenges, and applications Self-healing materials for electronic applications: An overview Self-healing materials with microvascular networks Self-healing materials for soft-matter machines and electronics Recent development and biomedical applications of self-healing hydrogels Polymers with autonomous life-cycle control Self-diagnosis of smart structures based on dynamical properties The grand challenges in smart materials research Nanogenerators for wearable bioelectronics and biodevices Bacterial biofilms: A common cause of persistent infections Biofilms: Formation, Development and Properties Biofilm formation as microbial development antibiotic restriction might facilitate the emergence of multi-drug resistance Liquid-infused silicone as a biofouling-free medical material The International Bank for Reconstruction and Development/The World Bank Purity of the sacred lotus, or escape from contamination in biological surfaces A. colloidal crystals delay formation of early stage bacterial biofilms Bacterial retention on superhydrophobic titanium surfaces fabricated by femtosecond laser ablation Biofilm bridges forming structural networks on patterned lubricant-infused surfaces Investigating the BSA protein adsorption and bacterial adhesion of Al-alloy surfaces after creating a hierarchical (micro/nano) superhydrophobic structure A brief recap of microbial adhesion and biofilms Physico-chemistry from initial bacterial adhesion to surface-programmed biofilm growth Detachment studies on microfouling in natural biofilms on substrata with different surface tensions Preventing bacterial adhesion onto surfaces: The low-surface-energy approach Deposition rates of viruses and bacteria above the atmospheric boundary layer Concentrations and size distributions of airborne influenza A viruses measured indoors at a health centre, a day-care centre and on aeroplanes Detachment and successive re-attachment of multiple, reversibly-binding tethers result in irreversible bacterial adhesion to surfaces Electric double layer interactions in bacterial adhesion to surfaces Electrostatic behavior of the charge-regulated bacterial cell surface Zeta potential of selected bacteria in drinking water when dead, starved, or exposed to minimal and rich culture media Zeta Potential and aggregation of virus-like particle of human norovirus and feline calicivirus under different physicochemical conditions Role of fimbriae, flagella and cellulose on the attachment of salmonella typhimurium ATCC 14028 to plant cell wall models Bacteria pattern spontaneously on periodic nanostructure arrays Effect of micro-and nanoscale topography on the adhesion of bacterial cells to solid surfaces Strength of bacterial adhesion on nanostructured surfaces quantified by substrate morphometry Bacterial adhesion: From mechanism to control Printed paper-based arrays as substrates for biofilm formation Micro-and Nanotopography sensitive bacterial attachment mechanisms: A review Biomaterial-centered infection: Microbial adhesion versus tissue integration Physico-chemistry of initial microbial adhesive interactions-its mechanisms and methods for study A reference guide to microbial cell surface hydrophobicity based on contact angles Quantitatively predicting bacterial adhesion using surface free energy determined with a spectrophotometric method Normally oriented adhesion versus friction forces in bacterial adhesion to polymer-brush functionalized surfaces under fluid flow Substrate viscosity plays an important role in bacterial adhesion under fluid flow Shrink-induced superhydrophobic and antibacterial surfaces in consumer plastics Factors affecting in vitro adherence of ureteral stent biofilm isolates to polyurethane Effect of bacterial, urine and substratum surface tension properties on bacterial adhesion to biomaterials Bioinspired hierarchical surface structures with tunable wettability for regulating bacteria adhesion Impact of engineered surface microtopography on biofilm formation of Staphylococcus aureus Bioinspired Photocatalytic shark-skin surfaces with antibacterial and antifouling activity via nanoimprint lithography Surface characteristics influencing bacterial adhesion to polymeric substrates Cell wall as a target for bacteria inactivation by pulsed electric fields Inactivation of bacteria by electric current in the presence of carbon nanotubes embedded within a polymeric membrane Graphene-coated spandex sensors embedded into silicone sheath for composites health monitoring and wearable applications Substrate-Independent, Transparent oil-repellent coatings with self-healing and persistent easy-sliding oil repellency High-performance uv protective waterborne polymer coatings based on hybrid graphene/carbon nanotube radicals scavenging filler The Addition of graphene to polymer coatings for improved weathering Polydimethylsiloxane-titania nanocomposite coating: Fabrication and corrosion resistance A piezo-resistive graphene strain sensor with a hollow cylindrical geometry Electrochemical study of a hybrid polymethyl methacrylate coating using SIO2 nanoparticles toward the mitigation of the corrosion in marine environments Corrosion protection of electrospun PVDF-ZnO superhydrophobic coating Sandwich-structured rGO/PVDF/PU multilayer coatings for anti-corrosion application High-performance flexible yarn for wearable piezoelectric nanogenerators Polymer Tribological Coatings A comprehensive review on graphene-based anti-corrosive coatings Highly Conductive PVA/Ag Coating by Aqueous in Situ Reduction and Its Stretchable Structure for Strain Sensor Enculescu, I. Direct and contactless electrical control of temperature of paper and textile foldable substrates using electrospun metallic-web transparent electrodes All-graphene strain sensor on soft substrate An ultraflexible polyurethane yarn-based wearable strain sensor with a polydimethylsiloxane infiltrated multilayer sheath for smart textiles Therapeutic Goods Administration. Literature Review on the Safety Of Titanium Dioxide and Zinc Oxide Nanoparticles in Sunscreens Self-powering/self-cleaning electronic-skin basing on PVDF/TiO2 nanofibers for actively detecting body motion and degrading organic pollutants Enhanced photocatalytic degradation performance by fluid-induced piezoelectric field TiO2 nanoparticles immobilized on polyacrylonitrile nanofibers mats: A flexible and recyclable photocatalyst for phenol degradation Silk fibroin as edible coating for perishable food preservation Strategies for tuning the biodegradation of silk fibroin-based materials for tissue engineering applications Polymorphic regenerated silk fibers assembled through bioinspired spinning Carbonized silk fabric for ultrastretchable, highly sensitive, and wearable strain sensors A Superhydrophobic smart coating for flexible and wearable sensing electronics Control of molecular transport through stimuli-responsive ordered mesoporous materials Stimuli-responsive photonic polymer coatings Epoxy self-healing coating by encapsulated epoxy ester resin in poly (urea-formaldehyde-melamine) microcapsules Shape-memory assisted scratch-healing of transparent thiol-ene coatings Retention of mechanical performance of polymer matrix composites above the glass transition temperature by vascular cooling Highly precise multifunctional thermal management-based flexible sensing sheets Acute effects of temperature exposure on blood pressure: An hourly level panel study Relationship between outdoor temperature and blood pressure Flexible pressure sensor with high sensitivity and low hysteresis based on a hierarchically microstructured electrode Repellent materials. Robust self-cleaning surfaces that function when exposed to either air or oil Superhydrophilic and superhydrophobic nanostructured surfaces via plasma treatment Superomniphobic surfaces for effective chemical shielding Designing bioinspired anti-biofouling surfaces based on a superwettability strategy Ultrafast electrosynthesis of high hydrophobic polypyrrole coatings on a zinc electrode: Applications to the protection against corrosion Robust super-hydrophobic coating prepared by electrochemical surface engineering for corrosion protection A stretchable and super-robust graphene superhydrophobic composite for electromechanical sensor application Composites for Spacecraft Failure detection and prevention for cyber-physical systems using ontology-based knowledge base Chromic Phenomena: Technological Applications of Color Chemistry Mechanochromic Fluorescent polymers with aggregation-induced emission features Temperature sensitive colour-changed composites Tin oxide-based electrochromics Self-reporting fluorescent step-growth raft polymers based on nitrile imine-mediated tetrazole-ene cycloaddition chemistry Mechanical reactivity of two different spiropyran mechanophores in polydimethylsiloxane Mechanochemically Active Soft Robots Flexible multifunctional sensors for wearable and robotic applications Role of mechanophore orientation in mechanochemical reactions Empowering mechanochemistry with multi-mechanophore polymer architectures Polymer mechanochemistry: From destructive to productive Stress-responsive polymers containing cyclobutane core mechanophores: Reactivity and mechanistic insights Dynamic phosphorescent probe for facile and reversible stress sensing Self-reporting dynamic covalent polycarbonate networks 3D-printed mechanochromic materials Mechanochromic and thermochromic fluorescent properties of cyanostilbene derivatives Thermochromic polymer film sensors for detection of incipient thermal damage in carbon fiber-epoxy composites A chameleon-inspired stretchable electronic skin with interactive colour changing controlled by tactile sensing Damage detection using self-sensing concepts Coupled carbon nanotube network and acoustic emission monitoring for sensing of damage development in composites Autonomic restoration of electrical conductivity using polymer-stabilized carbon nanotube and graphene microcapsules Damage identification in carbon fiber reinforced polymer plates using electrical resistance tomography mapping Monitoring crack appearance and healing in coatings with damage self-reporting nanocapsules A Robust damage-reporting strategy for polymeric materials enabled by aggregation-induced emission Tough and water-insensitive self-healing elastomer for robust electronic skin Self-healing materials Self-healing polymers and composites Self-healing polymeric materials Dynamic covalent polymer networks: From old chemistry to modern day innovations Extremely Fast Self-Healable Bio-Based Supramolecular Polymer for Wearable Real-Time Sweat-Monitoring Sensor A super-stretchable and tough functionalized boron nitride/PEDOT:PSS/poly(N-isopropylacrylamide) hydrogel with self-healing, adhesion, conductive and photothermal activity Three-dimensional self-healable touch sensing artificial skin device A healable supramolecular polymer blend based on aromatic π−π stacking and hydrogen-bonding interactions Encapsulation-Based Self-Healing Polymers and Composites A chemo-mechanically coupled model for capsule-based self-healing polymer materials Microvascular network optimization of self-healing materials using non-dominated sorting genetic algorithm II and experimental validation Repairing polymers using oscillating magnetic field All-printed magnetically self-healing electrochemical devices UV-triggered self-healing polyurethane with enhanced stretchability and elasticity Light-and heat-triggered polyurethane based on dihydroxyl anthracene derivatives for self-healing applications Shape memory-enhanced electrical self-healing of stretchable electrodes Metallo-, thermo-, and photoresponsive shape memory and actuating liquid crystalline elastomers A high performance self-healing strain sensor with synergetic networks of poly( Ɛ -caprolactone) microspheres, graphene and silver nanowires Ultrafast self-healing and injectable conductive hydrogel for strain and pressure sensors Self-healing, fully functional, and multiparametric flexible sensing platform Highly stretchable and self-healable mxene/polyvinyl alcohol hydrogel electrode for wearable capacitive electronic skin Self-healing and stretchable 3D-printed organic thermoelectrics Rehealable, fully recyclable, and malleable electronic skin enabled by dynamic covalent thermoset nanocomposite Continuous self-healing life cycle in vascularized structural composites Self-healing graphene-based composites with sensing capabilities Magnetic field triggered multicycle damage sensing and self healing Electrochemistry-assisted microstructuring of reduced graphene oxide-based microarrays with adjustable electrical behavior The route to functional graphene oxide Self-healable multifunctional electronic tattoos based on silk and graphene Self-healable, super tough graphene oxide-poly(acrylic acid) nanocomposite hydrogels facilitated by dual cross-linking effects through dynamic ionic interactions Scalable Manufactured Self-Healing Strain Sensors Based on Ion-Intercalated Graphene Nanosheets and Interfacial Coordination Effect of graphene content on the restoration of mechanical, electrical and thermal functionalities of a self-healing natural rubber High-field electrical transport in single-wall carbon nanotubes Carbon nanotubes: Present and future commercial applications Self-healing, highly sensitive electronic sensors enabled by metal-ligand coordination and hierarchical structure design 3D printing of a self-healing nanocomposite for stretchable sensors Extremely stretchable strain sensors based on conductive self-healing dynamic cross-links hydrogels for human-motion detection Self-healing strain sensors based on nanostructured supramolecular conductive elastomers Electrically conductive self-healing polymer composite coatings Effect of capsule addition and healing temperature on the self-healing potential of asphalt mixtures Recovery of electro-mechanical properties inside self-healing composites through microencapsulation of carbon nanotubes Synergistic effect of different graphene-CNT heterostructures on mechanical and self-healing properties of thermoplastic polyurethane composites A Skin-inspired stretchable, self-healing and electro-conductive hydrogel with a synergistic triple network for wearable strain sensors applied in human-motion detection Flexible sandwich structural strain sensor based on silver nanowires decorated with self-healing substrate MXenes stretch hydrogel sensor performance to new limits The Platform for Accelerating the Circular Economy (PACE). The E-Waste Coalition A New Circular Vision for Electronics Wrist-worn optical and chest strap heart rate comparison in a heterogeneous sample of healthy individuals and in coronary artery disease patients The validity and inter-device variability of the Apple Watch for measuring maximal heart rate Wearable multichannel photoplethysmography framework for heart rate monitoring during intensive exercise Wearable health device dermatitis: A case of acrylate-related contact allergy E-waste in the international context -A review of trade flows, regulations, hazards, waste management strategies and technologies for value recovery Biomaterials-The Intersection of Biology and Materials Science biodegradable polymeric materials in degradable electronic devices On the mechanisms of biocompatibility Materials, processes, and facile manufacturing for bioresorbable electronics: A review A stretchable and biodegradable strain and pressure sensor for orthopaedic application Biodegradable and highly deformable temperature sensors for the internet of things Allergic Responses Induced by the Immunomodulatory Effects of Nanomaterials upon Skin Exposure In vitro evaluation of the cellular effect of indium tin oxide nanoparticles using the human lung adenocarcinoma A549 cells Comparison of the toxicity of sintered and unsintered indium-tin oxide particles in murine macrophage and epidermal cells Metal oxide nanoparticles with low toxicity Material safety data sheet (MSDS): Indium tin oxide Nanoparticle dermal absorption and toxicity: A review of the literature Human skin penetration of silver nanoparticles through intact and damaged skin Shape-Dependent Skin Penetration of Silver Nanoparticles: Does It Really Matter? Emerging two-dimensional monoelemental materials (Xenes) for biomedical applications A Highly skin-conformal and biodegradable graphene-based strain sensor Flexible, highly sensitive pressure sensor with a wide range based on graphene-silk network structure Porosity, density and mechanical properties of the paper of steam exploded bamboo microfibers controlled by nanofibrillated cellulose Bio-Multifunctional Smart Wearable Sensors for Medical Devices From paper to nanopaper: Evolution of mechanical and physical properties Cellulose-derivative-based gate dielectric for high-performance organic complementary inverters Paper-based disposable molecular sensor constructed from oxide nanowires, cellulose nanofibers, and pencil-drawn electrodes Physical and mechanical properties of PLA, and their functions in widespread applications -A comprehensive review Mapping the mechanical and electrical properties of commercial silicone elastomer formulations for stretchable transducers Ultra-highly sensitive, low hysteretic and flexible pressure sensor based on porous MWCNTs/Ecoflex elastomer composites Flexible piezocapacitive sensors based on wrinkled microstructures: Toward low-cost fabrication of pressure sensors over large areas Highly stretchable capacitive sensor with printed carbon black electrodes on barium titanate elastomer composite Flexible Polycaprolactone and Polycaprolactone/Graphene Scaffolds for Tissue Engineering A highly sensitive biodegradable pressure sensor based on nanofibrous dielectric Synthetic bioresorbable polymers. In Durability and Reliability of Medical Polymers Dielectric properties of biocompatible and biodegradable polycaprolone and polylactide and their nanocomposites in the millimeter wave band. AIP Conf. Proc. 2012, 1459 Measurement of nonlinear mechanical properties of PDMS elastomer Poly(dimethylsiloxane) thin films as biocompatible coatings for microfluidic devices: Cell culture and flow studies with glial cells A highly sensitive pressure sensor using a Au-patterned polydimethylsiloxane membrane for biosensing applications Flexible pressure sensor with ag wrinkled electrodes based on PDMS substrate Flexible piezoelectric nanofibers/polydimethylsiloxane-based pressure sensor for self-powered human motion monitoring Enhancing performance of triboelectric nanogenerator by filling high dielectric nanoparticles into sponge PDMS film tensile and surface mechanical properties of polyethersulphone (PES) and polyvinylidene fluoride (PVDF) membranes Biocompatibility of polyethersulfone membranes for cell culture systems Biodegradable poly(ethylene succinate) (PES). 1. Crystal growth kinetics and morphology Study on properties of barium titanate/polyethersulfone dielectric composites prepared by physical dispersion method Study on effect of the polyethylene terephthalate/nanoclay nanocomposite film on the migration of terephthalic acid into the yoghurt drinks simulant Synthetic Biomaterials for Regenerative Medicine Applications Tailoring piezoresistive sensitivity of multilayer carbon nanotube composite strain sensors A novel means of fabricating microporous structures for the dielectric layers of capacitive pressure sensor Highly flexible room temperature NO2 sensor based on MWCNTs-WO3 nanoparticles hybrid on a PET substrate Inkjet-printed sensors on flexible substrates Enhanced thermal and mechanical properties of polyimide/graphene composites Biocompatibility of polyimides: A mini-review Preparation and characterization of biodegradable condensation polyimide Gold/polyimide-based resistive strain sensors Poly Lactic-co-Glycolic Acid (PLGA) as Biodegradable Controlled Drug Delivery Carrier Mechanical properties of PVA material for tissue engineering applications Poly (vinyl alcohol) as versatile biomaterial for potential biomedical applications Biodegradation of poly (vinyl alcohol) based materials Poly (vinyl alcohol) gate dielectric in organic field-effect transistors Study on the mechanical properties of different silkworm silk fibers High-strength, durable all-silk fibroin hydrogels with versatile processability toward multifunctional applications Methods and Applications of Multilayer Silk Fibroin Laminates Based on Spatially Controlled Welding in Protein Films silk composite electronic textile sensor for high space precision 2d combo temperature-pressure sensing Stretchable, biocompatible, and multifunctional silk fibroin-based hydrogels toward wearable strain/pressure sensors and triboelectric nanogenerators Flexible organic thin-film transistors with silk fibroin as the gate dielectric Graphene-based wireless bacteria detection on tooth enamel Unexpected Rheological Behavior of a Hydrophobic Associative Shellac-Based Oligomeric Food Thickener Mechanical properties of liquid-filled shellac composite capsules Natural resin shellac as a substrate and a dielectric layer for organic field-effect transistors Shellac films as a natural dielectric layer for enhanced electron transport in polymer field-effect transistors Biocompatible and Biodegradable Materials for Organic Field-Effect Transistors Evaluation of the biocompatibility of a coating material for an implantable bladder volume sensor Silicones for Stretchable and Durable Soft Devices: Beyond Sylgard-184 Transparent conducting films based on graphene oxide/silver nanowire hybrids with high flexibility Biocompatibility of silver nanoparticles and silver ions in primary human mesenchymal stem cells and osteoblasts Stretchable/flexible silver nanowire Electrodes for energy device applications Stretchable electronic skin based on silver nanowire composite fiber electrodes for sensing pressure, proximity, and multidirectional strain Biocompatibility and applications of carbon nanotubes in medical nanorobots Overview of carbon nanotubes for biomedical applications The pristine graphene produced by liquid exfoliation of graphite in mixed solvent and its application to determination of dopamine A brief review on graphene/inorganic nanostructure composites: Materials for the future Graphene-based sensors for human health monitoring Recent advances in graphene-based biosensor technology with applications in life sciences Graphene-based membranes for CO 2 separation Electrical properties of molybdenum disulfide MoS2. Experimental study and density functional calculation results Probing the biocompatibility of MoS2 nanosheets by cytotoxicity assay and electrical impedance spectroscopy Highly sensitive 1T-MOS pressure sensor with wide linearity based on hierarchical microstructures of leaf vein as spacer Conducting polymers and their biomedical applications. In Biosynthetic Polymers for Medical Applications The biocompatibility of polyaniline and polypyrrole: A comparative study of their cytotoxicity, embryotoxicity and impurity profile Study of the electrical characteristics of polyaniline prepeared by electrochemical polymerization facile fabrication of multi-hierarchical porous polyaniline composite as pressure sensor and gas sensor with adjustable sensitivity Polyaniline nanofiber wrapped fabric for high performance flexible pressure sensors High electrical conductivity and carrier mobility in oCVD PEDOT thin films by engineered crystallization and acid treatment Toxicity evaluation of PEDOT/biomolecular composites intended for neural communication electrodes Highly conductive PEDOT: PSS electrode by simple film treatment with methanol for ITO-free polymer solar cells Why does the electrical conductivity in PEDOT: PSS decrease with PSS content? A study combining thermoelectric measurements with impedance spectroscopy Adaptable pressure textile sensors based on a conductive polymer Biodegradable Transparent Substrate Based on Edible Starch-Chitosan Embedded with Nature-Inspired Three-Dimensionally Interconnected Conductive Nanocomposites for Wearable Green Electronics A Highly sensitive piezoresistive pressure sensor based on graphene oxide/polypyrrole @ polyurethane sponge Extending the pressure sensing range of porous polypyrrole with multiscale microstructures A silk-based coating containing GREDVY peptide and heparin on Mg-Zn-Y-Nd alloy: Improved corrosion resistance, hemocompatibility and endothelialization Accelerating the degradation of polyolefins through additives and blending A polymer-based capacitive sensing array for normal and shear force measurement Textile-Based wireless pressure sensor array for human-interactive sensing Wireless body sensor networks based on metamaterial textiles Hierarchical elastomer tuned self-powered pressure sensor for wearable multifunctional cardiovascular electronics Highly sensitive and multifunctional tactile sensor using free-standing ZnO/PVDF thin film with graphene electrodes for pressure and temperature monitoring A Sensor array using multi-functional field-effect transistors with ultrahigh sensitivity and precision for bio-monitoring