key: cord-1046696-dkukwnov authors: de Jongh, Frans H.C.; de Vries, Heder J.; Warnaar, Rob S.P.; Oppersma, Eline; Verdaasdonk, Rudolf; Heunks, Leo M.A.; Doorduin, Jonne title: Ventilating two patients with one ventilator: technical setup and laboratory testing date: 2020-07-06 journal: ERJ Open Res DOI: 10.1183/23120541.00256-2020 sha: ea4cfca32b3cf2fe18c56181a76e27bfc609a174 doc_id: 1046696 cord_uid: dkukwnov With a modified circuit, it is feasible to ventilate two patients with one ventilator over a relevant range of compliances. Adding inspiratory resistance allows individual titration of tidal volume, and incorporating one-way valves prevents pendelluft. https://bit.ly/3ex8SYP Technically, two ventilator circuits can be connected in parallel to one ventilator by adding T-connectors to the inspiratory and expiratory limbs. This was first reported using a setup to ventilate four test lungs, but very limited measurements were presented [5] . Another study demonstrated that ventilating four sheep with one ventilator was feasible for 12 h with frequent blood gases and repositioning [6] . In a more detailed laboratory evaluation, it was concluded that a single ventilator to support four patients is possible, but that tidal volume (V T ) cannot be controlled for each subject and depends on respiratory compliance [7] . To what extent different lung compliances affect V T distribution was not described, nor the strategies to control V T individually. The aim of the current study was to design an improved setup to ventilate two patients with one ventilator and evaluate V T distribution at different lung compliances. Furthermore, we aimed to individualise V T by adding inspiratory resistances (R insp ) in the circuit. Two breathing circuits (RT380; Fisher & Paykel, Auckland, New Zealand) were connected in parallel to a mechanical ventilator (Evita Infinity V500; Dräger, Lubeck, Germany) by T-connectors with one-way valves (#1954000; Intersurgical, Wokingham, UK) to the inspiratory and expiratory ventilator limbs (figure 1a). The Y-piece of both breathing circuits were connected to a single port of the Dual Adult Lung simulator (Michigan Instruments, Grand Rapids, MI, USA) with calibrated pneumotachographs (Vyntus; Vyaire Medical, Mettawa, IL, USA) and HEPA filters positioned between Y-piece and the test lung. Multiple experimental conditions were tested in pressure control (PC) mode (inspiratory time:expiratory time ratio 1:2, positive end-expiratory pressure (PEEP) 12 cmH 2 O, inspiratory rise time 0.1 s). First, compliance of both test lungs (C lung ) was gradually increased (10-70 mL·cmH 2 O −1 ) at an inspiratory peak pressure (P peak ) of 22 cmH 2 O at 25 breaths per min and thereafter 15 breaths per min. Second, C lung,right was gradually increased while C lung,left was fixed at 50 mL·cmH 2 O −1 . Third, three levels of R insp were added to the right test lung by compressing the tube with an adjustable clamp, while no R insp was added to the left test lung. C lung was equal in both test lungs (50 mL·cmH 2 O −1 ) and P peak was set at 30 cmH 2 O at 15 breaths per min. This experiment was repeated with C lung,right at 70 mL·cmH 2 and without one-way valves. Measurements for each experimental condition were performed for 1 min. V T to each test lung was calculated as an average over all breaths. V T is expressed in mL·kg −1 considering a patient with a predicted bodyweight of 80 kg. Figure 1b shows that V T in both test lungs remain comparable when C lung is gradually increased at 25 breaths per min. Due to the longer respiratory time constant at higher compliance and the short inspiratory time there is a deflection in the V T curve, which becomes linear at 15 breaths per min. Increasing C lung,right while keeping C lung,left constant results in a constant V T in the left lung, except for very low compliances. Figure 1c shows that adding resistance to the inspiratory limb of the right test lung decreases V T , while V T in the left test lung remains constant. In a setup without one-way valves, peak cough pressure, simulated by manual compression of the left test lung, was transmitted to the right test lung (both 10 cmH 2 O) and 68% of cough volume (190 mL) was inspired by the right test lung. Adding one-way valves to the setup completely prevented this pendelluft effect. Note that the one-way valves in the T-pieces we used were adapted to provide airflow in the shown directions. In our setup, we used a pneumotachograph (PNT) to record separate tidal volumes (V T ) and airway opening pressures. In clinical practice, a capnograph or pressure port could also be used to facilitate individual patient monitoring. b) V T distribution between the right and left test lung (blue and green solid lines) at increasing test lung compliance (C lung , lower x-axis) at 25 breaths per min and for the left test lung at 15 breaths per min (grey line). The dashed green line represents V T of the left test lung when C lung,left is kept constant (upper x-axis), while C lung,right is gradually increased (blue solid line, lower x-axis). Note that at very low C lung,right (10-20 mL·cmH 2 O −1 ), V T in the left lung becomes lower (dashed green line) despite constant C lung,left . This can be explained by the fast inflation of the right test lung at very low compliance (low time constant), which causes the ventilator to drastically decrease inspiratory flow and thereby averts complete inflation of the left test lung. c) Effect of adding inspiratory resistance (R insp ) to the right test lung (blue lines On average, we found that V T measured by the ventilator was 5% higher than the sum of the V T delivered to both test lungs (measured using the pneumotachographs). Experiments were repeated with other intensive care unit ventilators (Getinge Servo-U/Servo-I, Getinge, Mansfield, UK; Philips Respironics V680, Philips, Guildford, UK; Hamilton Medical G5 and G6, Hamilton Medical, Bonaduz, Switzerland) as well as another breathing circuit (#2000000S, Intersurgical, Wokingham, UK), and provided similar results. The major finding of the current bench study is that with a modified ventilator circuit it is technically feasible to ventilate two patients with one ventilator over a clinically relevant range of compliances in PC mode. Second, adding R insp allows individual titration of V T . Third, incorporating one-way valves in the circuit improves safety of the setup by preventing pendelluft. In agreement with a previous study, V T was distributed evenly between test lungs with a similar compliance [6] . In addition, we tested this over a large range of clinically relevant compliances and found that if compliance changes in one test lung, V T is preserved in the other test lung (except at very low compliance). In other words, changes in respiratory mechanics in one patient will not affect the other patient. This is of clinical relevance because respiratory mechanics in critically ill patients are subject to constant change or differ due to anthropometric differences. Importantly, our findings only apply to PC mode. In volume control mode, changes in respiratory mechanics in one patient, or worse a complete airway obstruction, will distribute V T to the other patient. By adding R insp (i.e. increasing the respiratory time constant) to one test lung, we were able to control V T . As shown in figure 1c, this may facilitate ventilating two patients with different respiratory mechanics with the same protective V T . Note that this only applies when flow does not reach zero at end-inspiration and requires careful monitoring of V T . In contrast to other studies [5] [6] [7] , we added one-way valves to the circuit to prevent backflow between patients for two reasons. First, to avoid high pressures, for instance with coughing or airway suctioning, being transmitted to the other patient. Expiratory pressure with coughing may exceed 100 cmH2O. Second, to limit infectious cross-contamination, although this requires further studies. Recently, multiple professional societies in the USA issued a consensus statement with multiple reasons for avoiding ventilating multiple patients with one ventilator [8] . Many of these reasons apply to volume control modes, which, as discussed here, should be avoided. Furthermore, adding additional R insp and one-way valves to the circuit as in our setup allows partial individualisation and prevents pendelluft. Nevertheless, several important practical considerations remain. First, patient breathing effort in PC mode should be absent or ventilator triggering disabled, as triggering by one patient will result in severe patient-ventilator asynchrony for the other patient. This approach is therefore not applicable during weaning. Second, respiratory rate, I:E ratio, inspiratory oxygen fraction and PEEP cannot be individualised. Third, individual monitoring is required and ideally includes capnography, flow and inspiratory pressure near the endotracheal tube, and pulse oximetry. Fourth, alarm settings of V T should be adapted to multiple patients. Stringent alarm limits on V T may be used to monitor whether one of the patient's respiratory mechanics are changing. Finally, we would like to emphasise that our approach is off-label ventilator use and should only be considered in emergency situations, after ethical board approval, consultation of all stakeholders and adequate training. Sharing ventilators between two patients may be a life-saving rescue intervention in health crisis or in low-resource countries. Future studies should be performed to improve this strategy and further develop clinical safety testing. Positive-pressure ventilation equipment for mass casualty respiratory failure Augmentation of hospital critical care capacity after bioterrorist attacks or epidemics: recommendations of the Working Group on Emergency Mass Critical Care Critical care during epidemics The Other Option Is Death': New York Starts Sharing of Ventilators A single ventilator for multiple simulated patients to meet disaster surge Increasing ventilator surge capacity in disasters: ventilation of four adult-human-sized sheep on a single ventilator with a modified circuit Use of a single ventilator to support 4 patients: laboratory evaluation of a limited concept Joint statement on multiple patients per ventilator Conflict of interest: F.H.C. de Jongh has nothing to disclose. H.J. de Vries has nothing to disclose. R.S.P. Warnaar has nothing to disclose. E. Oppersma has nothing to disclose. R. Verdaasdonk has nothing to disclose. L.M.A. Heunks reports a grant from ZonMw (CVID technology grant from Dutch goverment) during the conduct of the study; and a grant from Liberate Medical for expiratory muscle research and a grant from Orion Pharma for a levosimendan study, outside the submitted work. J. Doorduin has nothing to disclose. Support statement: This study was supported by a grant from "ZonMw program Creative solutions CoVID-19" (number 5000.9953). Funding information for this article has been deposited with the Crossref Funder Registry.