key: cord-1046412-epgv6rp1 authors: Perez Lorenzo, Luis Javier; Fernandez de la Mora, Juan title: Exceeding a resolving power of 50 for virus size determination by differential mobility analysis() date: 2020-09-12 journal: J Aerosol Sci DOI: 10.1016/j.jaerosci.2020.105658 sha: 1ef6b343d8ab5a9436ad98bff86160735665352b doc_id: 1046412 cord_uid: epgv6rp1 A recently described DMA designed for high resolution viral particle analysis (Perez-DMA; Perez-Lorenzo et al, 2020) is modified to decrease the relative peak full width at half maximum (FWHM) below previously achieved ≈3.3%. The electrode radii at the outlet slit (R(1) = 1.01 cm; R(2) = 2 cm) and the working length are almost unchanged (L = 114.9 vs. 116 mm). The laminarization trumpet and the radius of the curve merging the trumpet to the working section are both considerably widened to improve gas flow laminarization. DMA evaluation with salt clusters is improved by reducing the flow resistance at the gas outlet, to reach substantially larger sheath gas flow rates Q ˜ 1700 L/min. Tests with tetraheptylammonium bromide clusters with a center rod diverging at 3(o) demonstrate FWHM<2.7%, without indications of performance loss due to turbulence even at 1700 L/min. Correcting these high flow rate data for diffusive broadening reveals a maximal DMA FWHM in the limit of non-diffusing particles and zero sample flow, FWHM(∞) = 1.8%. An uncorrected peak width approaching 2% is independently demonstrated at much lower flow rates of sheath gas with two recently described bee virus particle standards having singularly narrow size distributions at mean diameters of 38 and 17 nm. Correcting raw 38 nm particle peak widths for broadening due to diffusion and aerosol to sheath gas flow rate ratio q/Q shows an even more ideal response with FWHM(∞)<1%, where this value includes nonidealities in the DMA as well as possible lack of monodispersity in the viral particles. Following the development of electrospray ionization (ESI, Fenn et al., 1989) it became possible to bring to the gas phase large ions such as proteins and viruses for analysis by mobility and mass spectrometry, either separately or in tandem. The approach has been quite mature for many years in the case of proteins. However, it remains in a relatively primitive state for viral particles, in spite of the great interest offered by the possibility to diagnose unknown viral infections based on measuring the mobility or the mass of the virus involved. Mass spectrometry has been severely limited at viral sizes due to the large number of elementary charges (z>100) typically carried by electrosprayed viral particles. The combination of this large z with the slight mass heterogeneity inevitable in viral genomes precludes the identification of isolated mass peaks in full viruses, not just with high resolution mass spectrometers, but even when combining mobility separation and mass spectrometry in tandem. Conventional mass measurements have been possible only on empty viral capsids made up of just proteins, where there is no mass variation between different capsids (Uetrech et al. 2008) . The only successful mass measurements to date on full viral particles have all been based on the independent measurement of both charge and mass made possible by exquisitely sensitive single-particle charge measurement techniques developed by Jarrold and colleagues (Lutomski at al. 2017) . It is important to note that all viral mass spectrometry studies made to date rely on fully charged electrospray ions, because no mass spectrometer exists able to cover the mass range of viral particles carrying only one or a few elementary charges. Mobility analysis has typically considerably less resolving power than mass spectrometry, but its size range is adequate to analyze singly charged viral particles with sizes of suggested the need to develop better DMAs specially designed for the 20-60 nm range. Our first attempt in this direction resulted in a previously described instrument , nicknamed the Perez-DMA. Its inner and outer electrode radii R 1 =1.01cm; R 2 = 2cm and axial length L=11.6 cm were all fairly conventional. We abandoned Reischl's bullet geometry (supported only on its downstream end) in favor of a center rod (supported both upstream and downstream) in order to achieve good centering in such a long instrument. To avoid an excessively long and heavy device, the inlet laminarization trumpet was small relative to our previous bullet-based DMAs. A key problem in that study was the lack of narrow size standards in the viral size range, needed to evaluate the DMA performance. As a reference, in order to prove unambiguously a resolution of 30 in mobility, the FWHM inherent to the standard itself had to be well below 1/30 in mobility, say 1/50. This implies a FWHM of about 1/100 in diameter. Since such outstanding standards had not yet been developed, we were forced to test that instrument with the only truly monomobile particles then available to us: singly charged cluster ions of a few nm produced by a bipolar source electrospraying alcohol solutions of various relatively large salts. This required a DMA able to deal with both, ~1 nm clusters and viral particles, which had to operate correctly at flow rates from above 1000 L/min down to ≈10 L/min. These diverse constraints forced a limit of 30 to the maximum resolution we were able to establish for that instrument. This prevented us from knowing for sure if this maximum demonstrated resolving power was caused by limitations of the DMA, the mobility standard, or both. The present study represents our new attempt to overcome that previous impasse, by seeking substantial improvements in the DMA, the mobility/size standard, and the quality of the electrospray. The effort has been facilitated by two innovations. First a better DMA passing an increased flow rate of sheath gas while still maintaining the flow laminar for evaluation with small clusters. Second, we use two new viral size standards recently described by Fernandez de la . The overall structure of the experimental apparatus is shown in Figure 1 . Design of the Perez-LT DMA. The LT denomination of the new instrument refers to its Long laminarization Trumpet. Its geometry and main geometrical characteristics are given in Figure 2 and Table 1 J o u r n a l P r e -p r o o f (i) The flow rate of sheath gas Q was limited to 1200 L/min by the flow resistance at the outlet. An increased Q naturally reduces the diffusive contribution to peak width for a given monomobile cluster. This problem has been addressed by the low-resistance outlet [12] shown in a three-dimensional rendition in Figure 2 . This piece achieves the same goal as its conical ancestor, of centering the rod with respect to the outer electrode. In addition, the cone is now carved with six symmetric rectangular openings occupying ≈80% of the channel cross section to pass the sheath gas into the exhaust chamber [14] . The flow resistance is also substantially reduced at the four laminarization screens [5] due to their much wider cross section (150 cm 2 versus 32 cm 2 ). The former perforated inlet plate [3] centering the upstream end of the center rod offers now far less resistance by using 6 radial bars that connect its outer and inner rings (left inset to Figure 2 ) and block just 18% of the laminarization screen section. (ii) Suspected imperfect flow laminarization and flow quality in the working section, due to an insufficiently wide set of laminarization screens, and to an excessive curvature in the final region of the entry flow (originally forced by the desire of a compact instrument). The quality of the inlet flow has two important effects on resolution. First, directly through the broadening effects of free stream turbulence surviving from the complex flow prevailing upstream the laminarizing screens. Second, the high Reynolds number laminar flow initially established by a suitable inlet geometry still has a tendency to become turbulent, and this transition is accelerated by any surviving free stream turbulence. The maximal flow rate that can be achieved prior to this transition is accordingly increased by a better laminarizer. These circumstances have driven us to provisionally drop the criterion of having a light instrument, in favor of achieving a much better laminarization. The new design therefore incorporates a much wider and longer laminarization trumpet [6] , and a smoother transition with a larger curvature radius joining the end of the conical region of the trumpet (arrow [6] ) to the point where the sheath gas meets the stream of sample gas entering through the inlet port [9] . The higher curvature of the prior design in this region was seen in flow calculations to accelerate the near-wall fluid upstream the inlet slit beyond the mean speed at that section. This situation subsequently forced a local flow deceleration at the outer edge of the boundary layer, which tends to destabilize the boundary layer. (iii) Conical inner electrode. The original Perez DMA had a conical inner electrode diverging at an angle of 1 o . This resulted in a gradual flow acceleration through the working section, which contributes to boundary layer stability (Hoppel, 1968 (Hoppel, , 1970 Fernandez de la Mora, 2002) . Nevertheless, we observed signs of resolution loss in some prototypes at flow rates below the maximum achievable. This we attributed to some incipient form of turbulent transition, caused either by an insufficient convergence angle, or by inadequate laminarization, as discussed in point (ii). To cover for these two possibilities, in addition to the larger laminarization structure of point (ii), we have tested two center rod geometries, with divergence angles 0.5 o and 3 o . The latter center rod [10] is the one shown in Figure 2 . Other relevant changes are schematically indicated in Table 2 . The electric field surrounding the anti-static tube (AST) [16] bringing the monodisperse aerosol from the high voltage of the center rod to ground at the outlet tube provides now a larger effective length of approximately uniform axial field, and therefore a better transmission. The ideal situation to avoid internal radial fields and associated electrophoretic losses is when the AST is immersed in a purely axial external uniform field, which is favored by the two flat electrodes seen in Figure 2 (top) at the beginning and the end of the tube [16] . Figure 3 gives more detail comparing the prior (a) and the new (b) outlet geometries. In the past we did not pay adequate attention to the effect of the electric field surrounding laterally the AST, as may be surmised from examining Figure 3a for the original Perez-DMA. In this case, although the AST is separated by a PEEK insulator from the high voltage electrode around it, the surrounding null external field penetrates through the upper part of the insulator, almost as if the insulator did not exist, and the potential along the AST remains constant down to the point where the metallic piece surrounding the AST ends. Most of the length of the AST is therefore ineffective in bringing the aerosol closer to ground potential. The effective length of the AST naturally depends on its conductivity and its dielectric constant and that of the insulator tube. However, for typical values of these parameters the insulator tube is relatively ineffective in shielding the AST from the external fields. The effective length of the AST in the LT configuration ( Figure 3b ) is essentially its whole length. 6 rectangular vents (30° off axis) Sheath outlet area at throat (cm 2 ) 3.5 7.4 AST effective length (mm) ≈8 ≈47 a The outer edge of the LT meshes is bound by a cylindrical instead of a conical surface. The DMA functions such that the product VZ of the classification voltage V and the electrical mobility Z of a particle is strictly proportional to sheath gas flow rate Q. Accordingly, accurate values of the DMA constant VZ/Q and Q must be known in order to infer mobility from measured voltage. The constant VZ/Q is fixed by purely electrostatic considerations and, like a capacity, depends only on the geometry (Tammet 1970) . For a strictly cylindrical geometry with inner and outer electrode radii R 1 , R 2 and classification length L this geometric constant is given by Equation (1a) (Knutson and Whitby 1975) . For an axisymmetric shape varying along the axial direction Equation (1b) defines the dimensionless shape factor K based on the radii at the exit slit (x=L). For a slowly varying geometry K may be approximated by Equation (1c) and K is a function of α only. In our past work, the lack of a sensor able to measure Q with precision led us to infer Z with the help of a cluster standard of mobility Z o , by exploiting the inverse relation between mobility and classification voltage: Z o V o =ZV when operating under fixed flow conditions. Unfortunately this method is problematic when the mobilities of the cluster standard and the particle of interest differ by several orders of magnitude, as is the case with most viruses. The main problem is that, at a flow rate suitable for virus selection, the classification voltage of a ~1nm standard is rather small, resulting in a broad peak and an inaccurate determination of V o . Accordingly, a reliable mobility inference should rather be based directly on a good measurement of the flow rate. For our non-cylindrical DMA, a proper determination of Q is also necessary to obtain the geometrical constant ZV/Q, since the expression provided by (1c) is only an approximation. Q was measured with a Sensirion Inc. SFM3000 Mass Flow Meter covering up to 200 sL/min (notice that the flow rate standard of this instrument is at 20 o C at 1013 mbar, and is maintained here). The manufacturer claims a maximal error of ±2% of the measured flow rate, with reproducibility of ±0.75%. This suggested the interest of calibrating more accurately the instrument to reduce the ±2% flow measurement ambiguity to ±0.75%. This we have achieved with a Half-Mini DMA (Fernandez de la Mora, 2017) by measuring the classification voltages V of positive singly charged ion clusters of tetraheptylammonium bromide (Ude et al. 2005) produced by a bipolar ion source (Fernandez de la Mora and Barrios, 2017). We first noticed that the ratio VZ/Q r based on the uncorrected flowmeter reading Q r varied at most by ±2% down at least to 25 sL/min. The Half-Mini DMA has a cylindrical working section (inner and outer radii R 1 = 4 mm, R 2 = 7 mm) over 19.03 mm out of its 20 mm length, with a modest change in R 2 (from 7 to 7.43) mm over the first 0.97 mm of its length. Its geometric constant is accordingly expected to be quite close to that of a cylinder. Use of the approximation (1c) shows that this constant differs from that for a cylinder (with R 1 = 4 mm, R 2 = 7 mm; L=20 mm) by a factor of 0.9984. This miniscule non-ideality is due to the fact that the departure from a cylindrical shape is initially with zero slope, so the region where R 2 is effectively different from 7 mm is considerably shorter than 0.97 mm. Therefore the error involved in approximating the geometry of this calibrating DMA as strictly cylindrical with R 2 =7 mm is well below 0.5%. Accordingly, the ratio VZ/Q is precisely known without the need of a calibration, and was used to obtain the flow rate Q. By comparing this actual Q to the value given by the Sensirion flowmeter we have established the calibration relation Q = 1.025Q r -0.4509 L/min, from which we are confident Q is known within better than 1% when Q r <200 L/min. We have confirmed in subsequent experiments with other DMAs that the ratio Q/(ZR) based on the calibrated Q(Q R ) relation is considerably more constant than the group Q r /(ZV), supporting the manufacturer's claim on the relatively high reproducibility of the measure of their sensor. At flow rates above 200 L/min Q was inferred from the assumption that it is strictly proportional to the voltages at which a given peak appears, with the proportionality constant determined for a given viral particle from a spectrum taken at Q < 200 sL/min. DMA circuit. The DMA was operated in a closed circuit of sheath gas with a pump a heat exchanger and a HEPA filter identical to those of Perez-Lorenzo et al. (2020) (Figure 1 ). The flow rate Q of sheath gas was controlled with a feedback loop that fixed the rotation speed of the blower with variations well below 1%. At flow rates below about ~100 lit/min the blower speed was fixed at 3200 rpm, and Q was controlled with a throttle valve inserted in the flow circuit. The experiments with the highest flow rates reported (up to 1700 L/min) could be achieved only by removing the valve and the flowmeter from the circuit, and by using a HEPA filter of high cross section. Viral particle preparation. The sample was prepared at BVS Inc. by the same procedure previously described (Fernandez-Garcia et al, 2020; Wick, 2015) , with more detail on this singularly narrow viral standard given by Fernandez de la . Briefly, 6 g of infected bees collected from the vicinity of a beehive are added to 40 ml of water purified by reverse osmosis, mixing them in a blender for 2 minutes. The material is then drained through a three layer cheesecloth to remove large fragments such as insect wings and legs. It is subsequently centrifuged at 20,000 relative centrifugal gravities to separate to the bottom of the vial cell fragments and other comparably large objects. The supernatant recovered is mixed with 500 ml of deionizes (DI) water and run past a 500 kDa hollow fiber tangential filter that removes particles smaller than about 7 nm and reduces the volume to 2 ml. This sample is then diluted 10fold into DI water, cleaned with a 0.45 mm mixed cellulose ester filter, and shipped to Yale. These samples are harmless to humans and can be handled at a facility with biosafety level 1 (BSL1). We have no absolute certainty on the nature of the virus in a given sample. A first provisional determination was made at BVS based on the particle size measured with TSI's commercial system. When studied at BVS, the sample used in this work showed the presence of two main viral peaks attributed to the Sacbrood virus and the Chronic Bee Paralysis Virus (CBPV), both very common in bees (Chen and Siede, 2007) . When studied a few days later at Yale, only the CBPV was present. Unlike the Sacbrood virus, this particle was robustly preserved during the shipment from BVS to Yale, and, once refrigerated, maintained high stability in DI water for many weeks. The stability in 90 mM aqueous ammonium acetate is not so good, with low mobility tails developing in a few days, that can be clearly seen with high resolution instrumentation (Fernandez de la Mora et al. 2020, Figure 4 ). The aqueous BVS samples were first analyzed as received and gave fairly sharp viral peaks over a slowly varying background. The background could be considerably reduced and the peaks narrowed down by several stages of dialysis in a Thermo Scientific Slide-A-lyzer mini dialysis device (88405) with a 20 kDa mass cut-off. We typically exchanged the DI water three times, dialyzing for periods of 2, 2 and 7 hours. All the high resolution spectra shown here correspond to the further dialyzed samples, to which ammonium acetate was added to a final concentration of 90 mM. Further confirmation that the sample used in this work is indeed CBPV is given by the fact that the mobility spectra obtained at Yale showed a second sharp peak at about 17 nm, in addition to J o u r n a l P r e -p r o o f the CBPV expected at 38 nm. The 17 nm peak is frequently associated with the CBPV, and does not appear alone due to its inability to replicate independently (Overton et al. 1982) . It is consequently referred to as the CBPV satellite (CBPVS). In any case, the precise assignment of a viral species to our sample is of no great relevance to our DMA evaluation here, as long as the particles offer a narrow size distribution. We shall accordingly refer to these standards as CBPV and CBPVS, without concern for the correctness of the assignment. Aerosol generation. A sample flow of 0.6 L/min was drawn by the detector pump from the ambient (without dehumidifying) through a HEPA filter, and went through a bipolar ES source. Figure 4 shows raw mobility spectra for positively charged THABr clusters at the lowest and the highest settings of the blower used, showing a drastic increase in resolution as the flow rate goes up. The bipolar source used produces dominantly singly charged ions (Fernande de la Mora and Barrios, 2017), so the ordering of the peaks in the spectrum reveals directly their composition: monomer (bare cation), dimer, trimer, etc, from left to right. The upper spectrum taken at 3200 rpm resolves up to 18 clusters. The lower spectrum taken at 9000 rpm goes only up to the octamer. Note that the noise level is about 0.1 fA, as expected for this electrometer. As a result of the rather low signal, the zero drift of the electrometer is not negligible. It was corrected by assuming a linear variation of the baseline with the voltage, such that the horizontal regions between isolated peaks would be indeed flat. As an example, in the data shown at 9000 rpm (lower continuous black line) the last datum that reads approximately 0 fA when corrected would have read 0.4 fA, and the flat theoretically horizontal regions between peaks 1-6, if not corrected, would have had an unphysical finite upward slope. Uncorrected data are shown also for reference as the dotted line falling slightly above the corrected continuous line for 9000 rpm. Spectra were taken at the pump settings and flow rates indicated in Table 3 . The peaks for the various clusters in these spectra taken at several pump speeds were fitted to Gaussian curves to obtain the resolving powers shown in Figure 5 . The figure includes the continuous gray line C V 1/2 scaling with peak voltage as theoretically expected from diffusive broadening (Rosell et al. 1996) . The coefficient C (1.55 Volt -1/2 ) used is not obtained by any rigorous theoretical analysis, but empirically as a low-voltage fit (gray line in Figure 5 ) to the upper envelope of the data: 1/FWHM Diff = C V 1/2 ; C =1.55 Volt -1/2 . (2) The value of C analogously obtained for the prior Perez DMA model was 1.35 Volt -1/2 , indicating a superior performance of the LT version, perhaps as a result of the better laminarization or the larger convergence angle. One can see at all pump settings that the data for the lower voltages (smaller clusters) approach best the diffusive limit. Yet, past a certain voltage, the resolution shows a drastic reduction. This reduction cannot be due to turbulence, since each series is obtained at fixed flow rate with only the size of the cluster varying. We believe it is primarily an artificial consequence of the fact that the larger clusters are not completely resolved from each other (top spectrum in Figure 4 above 0.4 kV), so that either the tails from neighbor cluster peaks or ion impurities between main members of the cluster series create a background that is not accounted for by the Gaussian fitting. The resolutions plotted in Figure 5 for each Q cannot accordingly be trusted past the voltage at which they fall markedly below the diffusion line, which starts at the voltage at which subsequent peaks are no longer baseline separated from each other. This corruption naturally sets in at smaller voltages for smaller pump speeds, so the upper envelope of the various data sets provides a more reliable measure of the real resolution, uncorrupted by peak merging. For instance, the spectrum of Figure 4 corresponding to a pump speed of 9000 rpm shows isolated peaks for all the clusters included (monomer to octamer), so all the open circles of Figure 5 ought to give uncorrupted resolutions. The resolutions measured at 9000 rpm for three of the data exceeds 37, well above the best value of about 30 observed with the prior Perez DMA. If the zero drift correction had not been implemented, the maximal resolving power would have been 34, still substantially higher than in our earlier study with the shorter laminarization trumpet. The fact that the measured resolution at increasing voltages falls clearly below the diffusive limit indicates that there is some level of non-ideality with this instrument. In other words, while the ideal peak width is 0 for non-diffusing particles in the limit q/Q→0, the actual value is non zero: FWHM→ FWHM ∞ > 0 when q=0 and diffusion is negligible. (3a) Concrete mechanisms possibly contributing to FWHM ∞ are (i) imperfect axisymmetry of either the instrument or the flow field. (ii) Even in a macroscopically symmetric device, tiny scratches on the walls may create substantial asymmetries by promoting local boundary layer transition or separation in certain regions of the DMA and not in others. (iii) Imperfect stretching or localized dirt in the laminarization screens. (iv) A finite size distribution of the viral standard. (v) Any unsteadiness in the flow. These may include not just high frequency flow unsteadiness, but could also result from an excessive fast scan of the DMA voltage compared with the particle residence time in the DMA or the detector (this last problem does not arise in this work) More generally, we consider three separate sources of peak broadening: (i) diffusive, (ii) the contribution due to finite aerosol flow rate (Knutson and Whitby, 1975) , and (iii) an additional source collecting the nonidealities intrinsic to the instrument. Their contributions to peak width can be considered to be independent, and thus approximated by the quadratic additive rule FWHM 2 = FWHM Diff 2 +FWHM ∞ 2 +(q/Q) 2 . (3b) The contribution (q/Q) 2 due to the finite aerosol flow rate q is negligible for the present cluster measurements, but is important in the analysis of viral particles considered in the following subsection. The theoretical lines used as references for the cluster measurements are accordingly based on the following simplified form of (2-3) FWHM 2 = 1/(C V 1/2 )+FWHM ∞ , (3c) with C=1.55 Volt -1/2 and FWHM ∞ determined by a fit to the data. The excellent fit to the 9000 rpm data (excluding the last point) shown as a dashed black line in Figure 5 is obtained with FWHM ∞ =0.018, implying an intrinsic DMA resolving power 1/FWHM ∞ = 55.6. This extrapolated diffusion-corrected performance is specific to the conditions of this experiment, with an aerosol flow rate q = 1.3 L/min and a sheath gas flow rate Q = 1700 L/min. Nevertheless, the fact that the cluster data are well fit by a single FWHM ∞ over a wide range of flow rates J o u r n a l P r e -p r o o f supports the interesting conclusion that the quality of the flow does not appear to deteriorate from the smallest to the largest flow rates explored (565-1700 L/min). In addition to determining the resolution of the LT-3° DMA, cluster standards have been used to obtain the calibration constants collected in Table 4 for the LT-3° and the LT-0.5° Perez DMAs. The measured values are included in convenient units in the second column, while the third column re-expresses this measured value as the dimensionless shape factor K defined in Equation (1b). The fourth column is the approximate value computed for the shape factor K via Equation (1c), which is fairly close to the measurement for the two angles studied. Figure 6 compares these two measured data (including the purely theoretical unity datum at α=0) with the approximation (1c) for the dimensionless shape factor K(α). The resolving power of a different DMA prototype using the LP-0.5 o inner rod was also briefly investigated with ion clusters, but was not as high as that found in the Perez-LP-3 o prototype due to some flow or geometric imperfection that remains to be identified. We do not think the problem with this prototype is related to the 0.5 o angle. Note finally that the largest Q achieved corresponds to a maximal gas velocity of only 30 m/s, so compressibility effects are negligible under all conditions investigated. Mobility spectra obtained with the LT-3 o DMA for the 3 viral particles produced by our sample have been previously presented and analyzed with the goal of characterizing the particles themselves (Fernandez de la Mora et al. 2020). Our present analysis of the same data is aimed at evaluating the DMA, especially at the modest flow rates required for virus analysis. The prior study by Perez-Lorenzo et al (2020) did explore this key low flow rate region (down to Q=47 L/min) with protein particles having relatively broad size distributions (FWHM>6%). Their conclusion that the DMA response was close to that ideally expected up to q/Q ratios of 5-10% was based on spectra showing peak widths between 6% and 8%, and provided no direct assurance of the DMA's ability to reach resolving powers of even 20 with particles larger than 20 nm. The situation is now more favorable thanks to the new viral standards. Figures 7a-b show mobility spectra for the CBPV and CBPVS viruses, where each viral particle is measured several times (and appears at several voltages) during a single voltage scan involving different Q values. The smaller peak to the left of the CBPV peak is too weak to yield a precise width, and will accordingly not be used here as a mobility standard. In experiments carried out several months after those shown in Figure 7a we have observed that the dominant peak and its shoulder decrease and increase in height with time. This evolution has enabled a more accurate determination of the width of the slightly more mobile peak, which was seen to be comparable to that studied here for the slightly less mobile peak. Data on inverse peak width are reported in Figure 8 for the two main particles. Note first that both figures show resolving powers even larger than the cluster experiments of Figure 5 . The width reported has been inferred following several criteria. Gaussian all refers to a fit of the whole peak to a Gaussian curve. This fit reflects best the real FWHM of the measured peak, including its right tail (attributed to involatile impurities in the solvent incorporated as contaminant residues into some viral particles). Gaussian Left implies a fitting of the left and top of the peak to a Gaussian, ignoring the right side of the peak potentially contaminated with involatile residues from the solution. This value reflects better the intrinsic width of the particles by removing contamination from ES imperfections. Linear Left and Linear All similarly refer to a fit to either the whole or the left half of the peaks to a triangular shape. That the peaks are approximately triangular, as predicted theoretically in the absence of diffusion (Knutson and Whitby, 1975) , can be seen in Figure 7b for the satellite (shown in more detail in Figure 6 of Fernandez de la Mora 2020). The real peaks are missing the three sharp edges of the triangle at its top and its two bottom regions, as a result of either diffusion, or of a finite width of the real particle size distribution (see Figure 2b of Fernandez de la Mora 2017). Accordingly, the height of the theoretical triangle exceeds slightly that of the experimental peak. This leads to a smaller FWHM for the triangular shape than for the actual peak, explaining why the Gaussian fit data in Figure 7 lie below the linear (triangular) fit data. At the modest values of Q used for the data shown in Figure 7 , the (q/Q) 2 contribution from the sample flow rate variable ignored in our prior analysis of the cluster data is no longer irrelevant. J o u r n a l P r e -p r o o f Prediction (3b) is shown in Figure 8a for the same value C=1.55 previously inferred from cluster studies, and for two values of FWHM ∞ . This quantity was associated in the prior cluster data with any possible non-ideal broadening mechanism from geometrical or flow imperfections in the DMA. This was then justified because the cluster standards are believed to be strictly monomobile. The same has never been established for any viral standard, so that now FWHM ∞ combines nonidealities in the response of the instrument with the perhaps finite width of the mobility distribution of the viral particles. It is noteworthy that, once the diffusive and q/Q corrections have been included in the theory, it fits the data with a value of FWHM ∞ approximately half-way between 0% and 1%. In other words, even if the DMA had a perfectly ideal response, we can state that the intrinsic width of the size distribution of this virus is about 0.5%. Alternatively, even if the virus was perfectly monodisperse, the resolving power of the DMA for non-diffusing particles at q/Q→0 would be about 200! These inferences are based on the linear-left criterion to determine the resolution. The choice is justified because our goal is to gauge the imperfections on either the viral size distribution or the DMA response, rather than the evident imperfections in the electrospraying process responsible for the asymmetry of the peaks. We conclude from the analysis of the data for the CBPV particle that both the DMA and the particle standard are closer to the ideal than could have been anticipated. The fact that the values of FWHM ∞ inferred are independent of the flow regime (Q) both for the CBPV (FWHM ∞ ≈0.5%) and the cluster data (FWHM ∞ ≈2%), yet differ so much among themselves is puzzling. This paradox is illuminated by the resolution data for the CBPV Satellite (Figure 8b) , which span flow rates intermediate between those used for the CBPV and the cluster studies. In this case no Gaussian analysis is used as the peaks look manifestly triangular. The theoretical lines shown in Figure 8b represent Equation (3b) with C=1.45 and FWHM ∞ of 0% and 2%, as indicated in the legend. The choice of C=1.45 used here differs slightly from the C =1.55 obtained in the prior cluster measurements and the C =1.35 reported by Perez-Lorenzo et al. 2020, perhaps because the effective diffusivity is enhanced by free stream turbulence, and therefore depends on the quality of the laminarizer and the Reynolds number. In spite of the scatter of the data, one can clearly recognize a region of high resolution to the right, and another of lesser resolution to the left, with a jump (marked with a vertical line) from one to the other taking place somewhere in the interval 159 < Q(L/min) < 177 (q=0.6 L/min). The corresponding critical DMA Reynolds numbers defined in Table 4 is in the range 1874 < Re * < 2087. These are values typical for turbulent transition in a tube in the absence of laminarization, and it is most unusual for an instability to set in in a DMA so close to this critical condition. The general situation is also unusual for a variety of reasons. First, the discontinuity seen is rather mild, as FWHM ∞ changes from about 1.8-2% at Q>170 L/min, to about 0.5% below it. This small effect was undetectable in our previous Perez DMA design. All other prior work with supercritical DMAs running laminarly at Reynolds numbers larger than 2000 has involved considerably shorter working lengths L, which decreases the time available for the growth of perturbations. These prior studies have also tended to work always at Re>> 2000, whence, if a regime change took place at Re=2000 it would have been missed. In perspective of the unusual situation, we have looked for other possible explanations for the jump in Figure 8b . One possibility is that the quality of the electrospray deteriorated slightly at one point as the series scanning over Q was carried out. This hypothesis is not unlikely, since achieving the rather high resolving powers recorded in Figures 8 requires that all components of the system (virus, ES, charge reduction, DMA) be close to their best. The data of Figure 8a do not allow resolution of the "turbulence versus ES quality" ambiguity because the maximal flow rate achieved in work with CBPV is substantially below the ~160 L/min at which the transition was observed in the CBPVS study. DMAs designed to run at high Reynolds numbers often exhibit nonideal behavior as a result of the natural tendency of some level of upstream turbulence to survive through the laminarization elements, or of some form of new turbulence to develop in the working section. Nevertheless, we have shown here that it is possible to achieve flow conditions where the DMA operates almost ideally. This regime is not easy to attain, but has been obtained here by using a wide laminarization trumpet with a particularly smooth transition into the working section. This ideal regime was not easy either to identify with prior evaluation methods relying on cluster standards useful mainly at fairly high flow rates. Fortunately, the availability of two recently identified viral standards with diameters of 17 and 38 nm now allows high resolution probing of DMAs at modest flow rates compatible with this ideal behavior. Exceptionally low uncorrected FWHM values have as a result been measured, approaching 2%. Once corrections are made for peak broadening by diffusion and the finite aerosol flow rate, we estimate a FWHM ∞ of 0.5%, including both DMA nonidealities and the finite width of the viral particles. This implies that the two viral particles tested are monomobile for all practical purposes, and also indicates that the DMA operates close to ideally. By future improvements in signal/noise we hope to obtain better values of the Q dependence of the uncorrected FWHM, and better bounds for the corrected DMA and virus FWHM ∞ . A confirmation of the difficulties involved in achieving near-ideal DMA performance is given by the fact that only one of the two Perez-LT DMA prototypes tested performed as exceptionally as reported here. We are confident that the performance of the inferior prototype was unrelated to the different inner rod angle, being rather due to one of the many usual causes for non-ideal DMA response. Planar differential mobility analyzer with a resolving power of 110 Siede R. Honey bee viruses Electrospray Ionization for Mass Spectrometry of Large Biomolecules High-resolution mass spectrometry of viral assemblies: Molecular composition and stability of dimorphic hepatitis B virus capsids Diffusion broadening in converging differential mobility analyzers High-Resolution Mobility Analysis of Charge-Reduced Electrosprayed Protein Ions A Bipolar electrospray source of singly charged salt clusters of precisely controlled composition Expanded flow rate range of high-resolution nanoDMAs via improved sample flow injection at the aerosol inlet slit Mobility analysis of Proteins by charge-reduction in a bipolar Electrospray source Singularly narrow viral size and mobility standards from the 38.3 nm chronic bee paralysis virus and its 17.5 nm satellite Virus size analysis by gas phase mobility measurements: Resolution limits The ions in the troposphere: Their interactions with aerosols and the geoelectric 9eld Measurement of the mobility distribution of tropospheric ions Analysis and handling of bio-nanoparticles and environmental nanoparticles using electrostatic aerosol mobility Macromolecule Analysis Based on Electrophoretic Mobility in Air: Globular Proteins Aerosol Classification by Electric Mobility: Apparatus Theory and Applications How to avoid interfering electrochemical reactions in ESI MS analysis Hepatitis B Virus Capsid Completion Occurs through Error Correction Relationships between the RNA components of chronic bee-paralysis virus and those of chronic bee-paralysis virus associate A high resolution DMA covering the 1-67 nm size range Improved DMA for high-resolution viral particle studies. To be submitted to Virion structure and genome delivery mechanism of sacbrood honeybee virus Chronic bee paralysis: A disease and a virus like no other? Sizing nanoparticles and ions with a short differential mobility analyzer The Aspiration Method for the Determination of Atmospheric-ion Spectra. Israel Program for Scientific Translations. Jerusalem. The original work in Russian Resolving Oligomers from Fully Grown Polymers with IMS-MS Molecular monodisperse mobility and mass standards from Electrosprays of tetra-alkyl ammonium halides Ion mobility mass spectrometry of proteins and protein assemblies Allmaier (2019) Virus-like particle size and molecular weight/mass determination applying gas-phase electrophoresis (native nES GEMMA Bionanoparticles as Candidate Reference Materials for Mobility Analysis of Nanoparticles We thank Kaartikey Misra, César Peláez, Jerome J. Schmitt III and the Nanoengineering Corporation (NEC) for their many contributions to this study. We gratefully acknowledge support from NEC and the NIH SBIR Phase I Grant Number: 1 R43 GM131542-01A1 (National Institute of General Medical Sciences, under Dr. Dmitriy Krepkiy). This study would not have been possible without the viral samples contributed by Mr. David Wick from BVS Inc (Stevensville, Montana). Following Yale rules JFM declares an interest in the company NanoEngineering commercializing the DMA here described