key: cord-1044501-u1djm3zh authors: Melissa, Liang Meishi; Ting, Chua Sze; Manful, Chan Man Fung; Li, Chionh Hui; Clara, Chooi Chi Yuen; Jones, Clare Louise; Tan, Sarah Li Hui; Veronica, Soong Si Min; Xuan, Tan Yi; Zhen, Low Xi title: Effectiveness of Fever Tent Setup in Covid-19 Pandemic – from Radiology's Perspective date: 2021-12-30 journal: J Med Imaging Radiat Sci DOI: 10.1016/j.jmir.2021.12.007 sha: d1ebeeba4b6ea3768a2996f703c3b875c3db5aa0 doc_id: 1044501 cord_uid: u1djm3zh Introduction: This paper describes our experience in setting up a dedicated imaging facility within a temporary fever tentage in an acute tertiary hospital in Singapore during the coronavirus disease 2019 (COVID-19) pandemic. We review in detail the effectiveness of the setup and its role from the radiological perspective. Methods: The dedicated imaging facility within the temporary fever tentage was equipped with a computer-on-wheels (COWs) to access patients’ medical records and a portable x-ray machine to allow for a smooth workflow. Radiation dose measurements were acquired around the imaging facility using phantoms and dosimeters to ensure radiation safety. Results: Due to its rapid nature and availability as a screening tool, chest x-ray (CXR) is the most widely used imaging modality during the COVID-19 pandemic. Our dedicated fever tent setup minimizes possible in-hospital transmission between both patients and staff, and provides a more streamlined workflow to tackle the high workload. It allowed us to reduce the time required for each radiograph, providing timely imaging services and radiological reports for expedient clinical screening. Discussion: The close collaboration between Radiology and Emergency Department in setting up the fever tentage is a crucial tool in allowing us to manage the COVID-19 pandemic. The fever tentage imaging facility is a highly effective tool, providing the means to handle the increased patient load in a streamlined and safe manner during a pandemic. Conclusion: This paper provides insights and guidelines in setting up a dedicated imaging service within the fever tent for future infectious disease outbreak contingency plans. The rapid onset of the COVID-19 pandemic has brought about much apprehension and has threatened to overwhelm healthcare systems globally. Since the first reported case in Wuhan, China on December 2019 1 , the virus"s rapid spread prompted the WHO to declare it a pandemic in March 2020 2 . As of August 2020, the cumulative cases stand at 18.4 million with 700 000 deaths globally 3 . Prevention and containment of virus transmission is deemed as the surest way to reduce morbidity and mortality 4 even with vaccination 5 . Highly infectious, the COVID-19 virus manifests itself in patients with symptoms ranging from a mild to severe respiratory illness 6 . Medical imaging techniques, such as the Chest X-Ray (CXR), Chest computed tomography (CT) and even ultrasound (of lungs), in conjunction with RT-PCR (reverse transcription polymerase chain reaction) testing 7, 8 was used in its diagnosis. In Singapore, there are 9 public and 8 private hospitals 9 island wide to serve a population of 5.7 million 10 . Singapore confirmed its first imported case on the 23 rd January 2020 and to date (31 st August 2020) has seen 56812 confirmed COVID-19 cases 11 . At its peak during the month of April 2020, the highest recorded daily number of cases was 1425 12 and a total of 5390 suspected cases were reviewed in our institution. In view of the rapid spread in both community and in foreign worker dormitories, the Singapore government has implemented two months of heightened safe distancing measures (Circuit Breaker period) on the 7 th April 2020 13 . Tighter measures were implemented to curb the transmission of COVID-19. Figure 1 shows the situational update in Singapore from the months of March to July 2020. In this paper, we review on the effectiveness of the fever tent setup during the pandemic in "the institution" from the radiological perspective. Being an acute tertiary institution in close proximity to a number of foreign worker dormitories, we received a total of 2450 confirmed COVID-19 patients as of August 2020. Due to the longer turnaround time of the serological RT-PCR (Reverse transcription polymerase chain reaction) testing 2 , as well as the highly variable and non-specific nature of COVID-19 symptomology, CXRs are still widely used for screening suspect cases internationally 14, 15 . During the months of February to August, there was a sharp increase in screening chest radiographs. For example, there was a threefold increase in CXR cases from February (1904 CXRs) to March (6100 CXRs). The COVID-19 virus is highly contagious. It is transmissible by airborne droplets and close contact with secretions. To handle both the increased demand and highly contagious nature of the cases, the Radiology Department collaborated with the Emergency Department to set up a fever tent. We will proceed to discuss the benefits and some challenges Radiology faced with this setup. Due to the retrospective nature of the study with no patients or patients" data involvement, ethics approval was not required for this study. In our hospital, all COVID-19 related X-rays were imaged within the dedicated fever imaging areas as detailed below. Screening and triaging was conducted at the entrance of the Emergency Department, where patients with respiratory symptoms and/or close contacts with COVID-19 patients were triaged into the fever zone. In the fever zone, there are various fever imaging facilities (Table 1 ). pandemic. An X-ray facility was constructed within the fever tent itself utilising a digital radiography (DR) portable X-ray machine. A retrospective review of the average time taken for a CXR in fever tent X-ray and fixed X-ray Room for the month of May 2020 was conducted. The structure of the fever tent erected with wooden planks was segregated into areas for doctor consultations, patient waiting areas, Covid-19 swab areas and an imaging facility for CXRs (Fig. 2) . The portable X-ray machine was parked at a fixed location while maintaining the standard of source to image distance (SID) of 180cm. The direction of the primary beam was distanced away from the patient waiting area. In order to cater to patients of various heights, a motorised, remote-operated adjustable stand was used to hold the imaging detector. With all the main components of the X-ray unit fixed, the radiographers, who are already limited by their comprehensive personal protective equipment (PPE), had fewer variables to deal with and this helped to reduce physical fatigue and examination time. This was especially important because, although sheltered, the tent was located outdoors. The high ambient temperature and humidity of our climate made it physically taxing on the radiographers and staff. The PPE required to be worn by the radiographers comprise of: hair net, disposable isolation gown, gloves and eye goggles. These conditions were also unfavourable to the portable X-ray machine. This required the implementation of larger industrial sized air coolers to achieve feasible working conditions for our staff and equipment. Within the x-ray imaging area (Fig. 3) , privacy curtains were used to partition the changing and imaging areas. Multiple electrical points essential for supplying the portable X-ray machine, cooler, and COWs were mounted at accessible locations in the fever tent. Electronic signature pads were connected to the computers to document pregnancy declaration for female patients of child bearing age. Waste bins, contaminated linen bins and consumables storage were also made available. In terms of infection control, it was ensured that proper infection control guidelines and standards were adhered to in the fever tent. Machines and equipment that came into contact with patients were disinfected with disinfectant wipes after every examination. X-ray examinations were also performed with a 5-minute interval in between to allow the disinfectant solution to dry fully before the next examination. Additionally, single-use disposable plastic bags were used to cover the detector plate to prevent cross-contamination between patients. All patients were required to wear facemasks and advised to sanitise their hands upon entering the ED facility. Located in a relatively small makeshift area, radiation safety was our immediate concern. Multiple portable X-ray shields (of 0.2 mm Pb thickness) were placed in an overlay fashion to allay concerns with scatter radiation To further confirm this, radiation readings (Fig 4) were measured at various locations. Multiple chest x-ray exposures (66kVp, 6mAs) were made on phantoms and dosimeters were used to document exposure readings. The readings were taken at least three times at four sides of the fever tent x-ray facility to ensure accuracy. The results showed that the areas behind lead shields and not in line with the direct beam were comparable to background radiation, not exceeding 10μSv/hr 16 . This was in accordance with guidelines from the National Environment Agency (NEA) -Radiation Protection and Nuclear Science Division (RPNSD) who was consulted, and approved the licence to perform radiographs at the premise. Radiation signs and barricades were put up to warn and restrict human traffic from entering these unsafe zones (primary beam area). Prior to making any x-ray exposures, the radiographer will ensure that only the patient undergoing the x-ray examination is present in the examination room. An oral warning would be given before making any x-ray exposures. Sufficient time would be given to the other healthcare personnel nearby so that they can move away from the examination room and be at a safe distance away from the radiation source and standing behind the lead shield. The portable x-ray machine must sound an audible warning when a radiation exposure is made. Radiographers who are working in the fever tent are to wear a Thermo-Luminescence Dosimeter(TLD) personal radiation monitoring device for monitoring of individual radiation dose received while performing the chest X-ray at fever tent. The CXRs done in the fever tent were performed using digital radiography ( Therefore, CXR standardised techniques performed for COVID-19 screening at the fever tent is preferred. DR technology has an inherent wide dynamic range and post-processing capability which reduce the number of retakes due to exposure errors 20 . Safe radiologic practices ensure that patient dose is kept as low as reasonably achievable (ALARA) while ensuring diagnostic image quality 21 . Our institution is fully equipped with electronic medical record (EMR) and electronic devices such as COWs and laptops which were used to access imaging orders. Electronic signature pad was connected to the computers to document pregnancy declaration for female patients within child bearing age. By eliminating the hardcopy forms, it reduced cross infection for healthcare workers. On the other hand, electronic devices can be easily cleaned with alcohol based wipes. All imaging requests were filtered according to patient location and examination type. They were performed either within the fever tent itself, the fixed X-ray room or performed portably at the fever facility. Upon performing the examination, all COVID-19 suspect cases had "COVID" remarked in the EMR system by the radiographers to alert the radiologist for STAT The labelling of COVID suspects in the system allows easy and consistent shunting into a new worklist folder. This independent list was to be regularly checked by the radiologist and allowed for a quick turn-around-time from completion of the study to finalization of the report within approximately one hour. This service was provided around the clock, extending up to two hours overnight. In May 2020, a total of 2035 CXR were done at the fixed x-ray room and fever tent. The average time taken to complete the CXR from the request initiation to completion were 19 .92mins and 12.78mins at the fixed x-ray room and fever tent respectively. The implementation of the temporary X-ray facility within the fever tent has allowed us to improve our efficiency of COVID screening radiographs. During the period 11 th February 2020 to 31 st July 2020, a total of 14234 chest radiographs were performed in our temporary fever tent, with a total of 1693 hours saved with the implementation of the fever tent x-ray facility. The CXR workflow and process in the fever tent area was greatly accelerated by designated The infrastructure designated for use in the fever tent X-ray area also helped increase the efficiency in performing X-ray procedures. The wireless compatible portable DR machine was used in the fever tent X-ray area. With the wireless capabilities, it allowed the radiographer to retrieve the worklist and send acquired images seamlessly. The wireless nature of the DR set-up also obviated the need for processing of cassettes by the reader, as compared to the Computed Radiography (CR) in the fixed X-ray room. This minimised movement and improved efficiency in this otherwise cramped space. Using DR would also omit the need to process cassettes. Having fewer touch points could lead to better infection control. The placement of a portable workstation in the fever tent X-ray facility enabled radiographers wireless access to the EMR easily to view X-ray orders as well as to check patent"s identity during the examination. The workstation also allowed female patients of child bearing age (12-55 years old) to sign on the electronic pregnancy declaration form prior to X-ray. This saved time as radiographers did not have to print the hardcopy forms to scan into EMR after the procedure. Initial radiation safety concern of the fever tent was alleviated. The X-ray area within the tent was enclosed by several overlapping lead shields that were specifically arranged to minimize leakage of scatter radiation. This helped to eradicate any unnecessary radiation exposure to staff who are nearby. Radiation dose measurements taken across multiple tent locations showed no significant increase in radiation dose. This is in keeping with National Environmental Agency (NEA)"s licensing for radiation facilities, not exceeding 10μSv/hr 16 26 . In other sites with tent, there were no dedicated x-ray service within the temporary facility, patients who required imaging were escorted through a separate exit and a dedicated pathway directly into the main ED 27, 28 . However, this approach would result in the lack of seamless experience for the patients. Imaging findings of COVID-19 on the chest radiograph are often subtle, particularly in earlymild disease, with only 69% of admitted patients presenting with an abnormal chest radiograph. When reporting radiographs from this COVID-19 specific folder, the radiologist is made aware of the clinical context and can thus tailor his sensitivity for COVID-19 specific findings. For example, the most frequently encountered findings of "dry" air space opacities 14, 29 , which can be subtle and non-specific in a different clinical context. The label of the radiograph into the "COVID folder", much like any good clinical history, would cue the radiologist into increasing his sensitivity and clinical suspicion for subtle pulmonary consolidations. This would be in addition to the usually reported abnormalities such as pulmonary masses, rib fractures, pneumothoraxes etc. On the ground, we did not find an increase in other overlooked findings during this period. With a system and mindset to answer specific clinical question, the fever tent helped to facilitate clearer communication between the emergency physicians and the radiologists and minimizes secondary calls or questions which might delay the patient"s isolation or deisolation status. Made aware of the clinical suspicion and appropriate context for COVID-19 pneumonia, the radiologist would tend to make a specific mention to include COVID-19 related findings, tailoring his or her report to address the explicit purpose of the study -to assess for possible COVID pneumonia. It is important to note that this occured in a period before rapid test kits were available, thus chest radiographs were an essential screening tool, playing a large role in triaging potential COVID patients. Lastly, the flexible nature of the fever tent setup meant that hospital staff were always able to adapt the setup to accommodate the current situation. This allows the effective channelling of available resources to ensure maximisation of manpower, space and machines. For example, during the surge of COVID-19 patients from the foreign worker dormitories, more chairs were added and the fever tent was expanded to a bigger area to put up with the increased number of patients. This also extended to the imaging facility within the tent where more machines could be brought in or equipment could be moved around to suit the situation. The unprecedented COVID-19 pandemic has drastically changed the lives of people all around the world. Within the radiology department, we were challenged to reconfigure our setup and work processes in response to the needs of frontline imaging services to aid in timely diagnosis for the patients who presented to our hospital. Lessons learnt from the COVID-19 pandemic should be incorporated into the design and planning of future radiology departments and facilities. This will enable our future healthcare facilities to be equipped with imaging services that are integrated with the hospital"s outbreak containment and infection control strategies. With the change to the environment and workflow, working in the fever tent raised inevitable challenges. The most pertinent issues such as the portable machine experiencing intermittent Wireless-Fidelity (Wi-Fi) connection issues due to the instability of Wi-Fi connection, detector artefacts, space constraints and warm environment as discussed below. The portable X-ray machine is connected to the patient"s worklist wirelessly while the imaging detector is connected to the portable machine via WI-FI. This allows acquired images to be sent to the facility"s picture archiving and communication system (PACS) in a few seconds. However, instability of the hospital"s Wi-Fi network connection at the fever tent would disrupt this workflow, as the fever tent is not within range of the Wi-Fi coverage. Patients would then have to be brought to the fixed X-ray room for their procedure, potentially delaying diagnosis. Also, it was noted by the engineer that signal emitting devices would disrupt the connection between the detector and the machine. To resolve this, the position of the detector on the erect stand was adjusted to receive a better signal and to increase the distance of the patients" mobile device from the detector. In addition, a transmission cable was also available as a backup to mitigate the intermittent connection issues. However, it was used only as a contingency plan with infection control considerations and would be cleaned with disinfectant wipes after usage. Singapore has a typically tropical coastal climate; it is warm and humid throughout the year. There is little variation in temperature and its daily temperature ranges from between 23°C to 33°C 30 . The outdoor fever tent has no air-conditioning and the temperature can fluctuate between 28-36°Celsius (C) in the day and 24-28°C at night. When the detector absorbs high energy from the X-ray photons, heat is also naturally produced. Coupled with that, the detector"s temperature can increase by a magnitude of 7°C above the calibrated optimal working temperature (26°C) causing intermittent overheating of the detector. This overheating may result in intermittent degradation of image quality with artefacts (Fig. 5 ). Cooling industrial fans and a laptop cooler were installed in the facility to reduce the ambient temperature and cool the detector respectively. Radiographers performed daily quality check on hardware and warning signs. Daily test exposures were also made to ensure optimal image quality. Due to the pandemic, the portable machine in the tent has been heavily utilised to screen the surge of patients. Wear and tear artefacts inevitably ensued as seen by the radiopaque speckled appearance on the image ( Fig 6) . To resolve the issue of recurrent artefacts, regular detector calibration by the engineer is required. However, due to Covid-19 safety measures, staff contact within the fever tent had to be minimised, and radiographers had to learn to troubleshoot and resolve the issue. There were also times where calibration-trained radiographers were not rostered on duty when an incident occurred. This problem was overcome by training more radiographers to be proficient in performing detector calibration and to have more regular detector calibration sessions conducted. In addition, regular updates and communication between supervisors and the pandemic team ensured that everyone knew how to handle the problem when encountered. Regular online continual professional development (CPD) sessions with examples of digital radiographic artifacts were also carried out. This helped to educate the radiographers on how to identify these artefacts and thereafter know how to resolve the artefact or prevent the artefacts on subsequent CXR images. Multiple lead shields were used to demarcate the X-ray area. This limited the space and thus could only accommodate ambulant patients. Hence, all X-rays for non-ambulant patients, bedbound or requiring wheelchair assistance would have to be done in the fixed X-ray room. Heat exhaustion was another challenge faced by the radiographers working in the fever tent. In line with hospital"s infection control requirements, it was mandated that they had to don full PPE despite the warm and humid weather. However, this was overcome with the increase in manpower to allow for more regular breaks to rest and rehydrate during the pandemic shift. Radiographers were segregated into 2 main teams, clean and pandemic with rotations of 2-3- week cycles. Radiographers on shift worked minimum 40hours per week, averaged over one or 2-week shift roster. At the start, the pandemic team consisted of 8 radiographers per week, with 2 radiographers per shift. To cope with the extra demand, an addition of 4 radiographers per week were allocated to the pandemic team, totalling up to 3 radiographers per shift. Despite the challenges faced, the many advantages of the fever tent have clearly shown that it has been an effective strategy to manage the pandemic in terms of patient segregation in imaging. This collaboration between Radiology and ED to set up the tent was crucial and contributed much towards the hospital"s needs to cope with the surge in the number of patients in this emergency. Figure 2 shows the overall layout of the fever tent with dedicated x-ray area, consultation rooms, nurse station, swab area, ECG station and patient holding area. The perimeter of the fever tent area is demarcated by wooden planks. Figure 3 shows the layout of the fully functional fever tent x-ray area, with portable machine (radiation sign), computer-on-wheels (C.O.W). changing area, changing area and industrial cooler. The x-ray area was lined with multiple lead shields in an overlay fashion for radiation protection. The x-ray tube was positioned at 180cm away from the imaging plate for consistency of chest x-rays. Figure 4 shows the average radiation survey meter (RSM) readings, 0.31μSv/hr, 0.41μSv/hr, 0.38μSv/hr and 0.54μSv/hr at four regions. The readings were taken with the primary x-ray beam (blue arrow) directed at a phantom with chest x-ray exposures (66kVp, 6mAs). Figure 5 shows an chest x-ray with multiple horizontal linear artefacts which were caused by overheating of x-ray detector. Figure 6 is a chest x-ray image with radiopaque speckled appearance as shown by the blue arrows. A Novel Coronavirus from Patients with Pneumonia in China WHO Director-General's opening remarks at the media briefing on COVID-19 -11 Accessed 17 WHO Coronavirus Disease (COVID-19) Dashboard Accessed 26 Organisation WH. COVID-19 advice for the public: Getting vaccinated The role of chest radiography in confirming covid-19 pneumonia Thoracic imaging tests for the diagnosis of COVID-19 statistics/singapore-health-facts/health-facilities Singapore Population CIRCUIT BREAKER TO MINIMISE FURTHER SPREAD OF COVID-19 minimise-further-spread-of-covid-19 Frequency and Distribution of Chest Radiographic Findings in Patients Positive for COVID-19 Clinical utility of chest radiography for severe COVID-19 Summary of the Radiation Protection (Ionising Radiation) Regulations Bontrager's textbook of radiographic positioning and related anatomy Chest radiography for radiologic technologists.(DIRECTED READING) Continuing education in anaesthesia, critical care & pain A major advantage of digital imaging for general radiography is the potential for reduced patient dose so film/screen systems should be phased out as unnecessarily hazardous Analyses of patient dose and image quality for chest digital radiography Experience summary of a COVID-19 designated community hospital and its operation model Temporary hospitals in times of the COVID pandemic. An example and a practical view Outcomes of COVID-19-Related ARDS Patients Hospitalized in a Military Field Intensive Care Unit. Mil Med. 2021. 25. Operational considerations for case management of COVID-19 in health facility and community Emergency department preparation for COVID-19: accelerated care units Emergency Department Management of the Covid-19 Pandemic How to organize an emergency department during the COVID-19 pandemic An update on COVID-19 for the radiologist -A British society of Thoracic Imaging statement Assessment of temperature extremes and climate change impacts in Singapore The authors would like to thank Dr Yeong Kuan Yuen, Dr Bernard Wee, Dr Gary Choa and all colleagues from Ng Teng Fong General Hospital Radiology Department and Emergency Department for their invaluable assistance and support in the preparation of this paper. We thank all frontline workers for their dedication and effort during this COVID-19 epidemic.