key: cord-1042811-f1peohdf authors: Bertoli, Simona; Leone, Alessandro; De Amicis, Ramona; Foppiani, Andrea; Osio, Diana; Battezzati, Alberto title: Effects of wearing a FFP2 mask on indirect calorimetry measurements: A pilot study date: 2020-11-10 journal: Clin Nutr ESPEN DOI: 10.1016/j.clnesp.2020.10.015 sha: 66321c109bb891dedafab6a6cdd7842e1f7a3950 doc_id: 1042811 cord_uid: f1peohdf BACKGROUND & AIMS: During the coronavirus disease 2019 (COVID-19) pandemic the use of Indirect calorimetry (IC) during nutritional rehabilitation program requires special precautions due to possible contagions for patients and health professionals. We evaluated in a sample of healthy subjects the agreement between oxygen consumption (VO(2) mL/min), carbon dioxide production (VCO(2) mL/min), respiratory quotient (RQ) and resting energy expenditure (REE kcal/24 h/day) measured by IC with and without a filtering facepiece mask. MATERIALS: 10 subjects with a mean (SD) age of 43 (10) years and a body mass index of 25.2 (5.8) kg/m(2) underwent indirect calorimetry both with and without a class 2 filtering facepiece mask (FFP2), in random order. The limits of agreement (LOA) and the concordance correlation coefficient (CCC) were used to evaluate the interchangeability of the measurement conditions. RESULTS: The LOA between REE measured with and without FFP2 (−111 to 189 kcal/day) were comparable to those for repeated IC tests without wearing masks and CCC (0.95) showed substantial agreement. CONCLUSIONS: We observed high agreement between REE measured by IC with and without FFP2 mask. These procedures are interchangeable in clinical practice. Measuring resting energy expenditure (REE) by indirect calorimetry (IC) plays a central role in the management of the nutritional rehabilitation program. IC is required to establish a correct nutritional diagnosis and to prescribe the optimal energy intake for a specific patient energy requirements, both in adults and in children and specifically in presence of diseases that can impair REE, making common predictive formulas unreliable [1] . On 9 March 2020, the government of Italy declared a national quarantine. The lockdown measures have included the temporary closure of all non-essential activities, and increased measures aimed to increase personal hygiene, symptom monitoring, early diagnosis, patient isolation, and the suspension of all non-deferred medical visits, including nutritional assessment and indirect calorimetry in order to guarantee social distancing. In phase 2 of these epidemic, clinical activities are opening and new standardized procedures are needed to protect both patients and health professionals. Concerning REE measurement, an increasing number of queries has been raised related to the use and safety of IC, beside ordinary cleaning procedures [2] . Manufacturing companies have released comprehensive cleaning and disinfection manuals to ease and promote adequate infection prevention procedures. Specifically, the canopy needs validated cleaning and Abbreviations: BMI, body mass index; CCC, concordance correlation coefficient; COVID-19, coronavirus disease 2019; FFP2, class 2 facepiece filtering mask; IC, indirect calorimetry; LOA, limits of agreement; REE, resting energy expenditure; RQ, respiratory quotient; SD, standard deviation; VCO 2 , carbon dioxide production; VO 2 , oxygen consumption; nmVCO 2 , carbon dioxide production not wearing class 2 filtering facepiece mask; nmVO 2 , oxygen consumption not wearing class 2 filtering facepiece mask; wmVCO 2 , carbon dioxide production wearing class 2 filtering facepiece mask; wmVO 2 , oxygen consumption wearing class 2 filtering facepiece mask. disinfection procedures for any reusable parts and a high bacterial/ viral efficiency filter connected to the patient canopy expiratory port protecting the device from any contaminated flow and finally, many instruments are provided of electropneumatic devices to guarantee that no air flows from the device to the patient. However, all these best practice guidelines [3] do not foresee that the patient may wear a mask during the procedure, thus increasing the level of biological risk for both the patient and the healthcare personnel and requiring an higher level of individual protection devices. The aim of the present study was to evaluate the agreement between REE measured by IC with and without a filtering facepiece mask in a sample of healthy subjects. A convenience sample of 10 subjects (7 women and 3 men) with a mean (SD) age of 43 (10) years was consecutively recruited among the University personnel and IC exams were conducted during 3 days at the International Center for the Assessment of Nutritional Status (ICANS). All subjects underwent a clinical history and a physical examination. Acute and chronic organ diseases were reasons for exclusion. Participants were healthy volunteers who were enrolled in a nutritional study that included indirect calorimetry among other investigations. The whole study protocol had been approved by the local Ethical committee of the University of Milan (report #21/2020) and all volunteers provided their fully informed written consent prior to participate in the study. This manuscript reports only the results relative to indirect calorimetry, whereas the full study results will be reported separately. Weight and height were measured according to standard guidelines [4] . Body mass index (BMI) was calculated as weight (kg)/[stature(m)] 2 . An open-circuit ventilated-hood system IC (Q-NRG, Cosmed, Roma, Italy) [5] was used to measure oxygen consumption (VO 2 ), carbon dioxide production (VCO 2 ) from ventilation gases in spontaneous breathing, to calculate REE. The IC was calibrated according to the manufacturer's instructions: calibration of the flowmeter was performed using a certified 3 L calibration syringe, while calibration of zero, span and delay alignment of the gas analyzers was performed using a certified calibration gas. All subjects received in random order two consecutive VO 2 (mL/ min), VCO 2 (mL/min) measurements both with and without a class 2 filtering facepiece (FFP2) mask. Each measurement was done by one trained dietitian early in the morning (between 09.00 and 10.00) and after twelve hours' fast. Subjects lay awake completely at rest supine on a bed in a thermally neutral environment (24 C). Gas exchange was measured by transparent ventilated canopy equipped with antibacterial filter and disposable veil. To avoid gas leakages, the subject's head was carefully wrapped with veil. Canopy hood underwent high-level disinfection, immediately after each use, with a soft cloth strayed with a cleaning solution for 30 s followed by air drying. A 15-min resting period was observed before starting the measurements in order for the subjects to acclimatize themselves to the canopy and instrument noise. For each measurement period, respiratory gas samples were taken for at least 10 min and the data collected during the first 5 min were discarded, as recommended [6] A 5-min steady state was defined as the first five consecutive stable 1-min readings with a coefficient of variation < 10% for VCO 2 and VO 2 [7] . After steady state was achieved, the FFP2 mask status was switched, and the second measurement period was begun always discarding the first 5 min of measurement. Respiratory quotient (RQ) was calculated as the ratio VCO 2 /VO 2 during steady state. To calculate REE modified Weir equation was applied [8] using the mean VO 2 and VCO 2 during steady state: REE (kcal/day) ¼ [3.941VO 2 (mL/min) þ 1.106VCO 2 (mL/min)] Â 1.44 Values of continuous variables are reported as mean and standard deviation (SD). The proportion of subjects who reached the steady state was recorded as well as the time to complete each set of measures. The duration of the studies was then compared among mask status by paired T-test. Bland and Altman's method was used to calculate the limits of agreement (LOA) between measurements taken with (wmVO 2 , wmVCO 2 and wmREE) and without mask (nmVO 2 , nmVCO 2 and nmREE). For each measurement of interest, bias was defined as the difference between the values obtained with and without mask. Pitman's test was used to evaluate proportional bias [9] . Lin's concordance correlation coefficient (CCC) was calculated as further measure of agreement [10] . The strength of the agreement was graded as poor when <0.90, moderate between 0.90 and 0.95, substantial between 0.95 and 0.99 and almost perfect when >0.99. Table 1 gives the measurements of the study subjects. All subjects could equally reach a steady state of VO 2 and VCO 2 with and without mask. Provided that the measures in the first 5 min were always discarded, the consecutive 5 minutes steady state conditions were achieved in similar times, 11.9 (2.2) and 11.4 (2.7) minutes whether wearing and not wearing the mask (p ¼ 0.674). Table 2 reports the agreement between the measurements of interest with and without mask. Figure 1 gives the corresponding LOA and CCC plots. The fixed bias was low for all measurements. In detail, a fixed bias of 6.01 mL/ min (9.86), 3.19 mL/min (16.19), À0.01 (0.06) and 39 kcal/day (77) was found for wmVO 2 , wmVCO 2 , wmRQ and wmREE, respectively. No proportional bias was found for any measurement of interest. wmREE and wmVO 2 showed a substantial (CCC ¼ 0.95, 0.99) agreement with the respective reference measurements. Lower CCC values were found for wmVCO 2 and wmRQ. This study was performed to assess whether wearing a FFP2 mask during indirect calorimetry may affect measured REE. All patients reached steady state of measured gas exchanges and showed a high agreement between VO 2 , VCO 2 and REE in a similar time estimated with and without FFP2 mask. No previous studies investigated this issue that in COVID-19 era has become of great interest in the clinical practice of nutritional rehabilitation program. We used Lin's CCC because it is one of the less biased measures of agreement [10] , and the REE CCC value of 0.95 showed substantial agreement. The variability we observed among the tests with different mask status can be explained by the inherent variability of IC exams, because the LOA are comparable to those previously reported for repeated IC tests without wearing masks [5] . Mean intra-subject variability for VO 2 and VCO 2 of our new generation indirect calorimeter within the mean interval of 5 min is À3 ± 14 and À3 ± 12 mL/min, respectively [5] . In our study, the differences with and without filter mask were in line with this intra-subject variability: VO 2 and VCO 2 values tended to be higher with the filter mask but were not statistically significant. Our work provides useful information for the management of IC during the COVID-19 pandemic but has some limitations. Specifically, the main limitations of this study stay in the limited sample number and in the selection of subject in a wide range of BMI, but without significant comorbidities potentially affecting the IC performance. In conclusion, wearing a filtering facepiece did not affect REE measured by IC. Not wearing the mask may expose the health personnel and other patients to infective hazard, thus we recommend wearing a mask during indirect calorimetry because this practice does not affect REE estimates. Our data support the use of a mask as the standard practice during COVID-19 pandemic. This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors. Author contributions SB and AB were responsible of the conception and design of the study. DO and RDA acquired the data. AL and AB performed analysis. SB and AB provided interpretation of data. SB, AB, and AL drafted the article. AB, AL and AF revised it critically for important intellectual content. All authors gave final approval of the version to be submitted. Indirect calorimetry and nutritional problems in clinical practice Indirect calorimetry: technical aspects ESPEN expert statements and practical guidance for nutritional management of individuals with sars-cov-2 infection Anthropometric standardization reference manual A new indirect calorimeter is accurate and reliable for measuring basal energy expenditure, thermic effect of food and substrate oxidation in obese and healthy subjects. E Spen Eur E Measurement reliability and reactivity using repeated measurements of resting energy expenditure with a face mask, mouthpiece, and ventilated canopy Measured and predicted resting energy expenditure in clinically stable patients New methods for calculating metabolic rate with special reference to protein metabolism Statistical techniques for comparing measurers and methods of measurement: a critical review A concordance correlation coefficient to evaluate reproducibility The authors declare that there is no conflict of interest regarding the publication of this article.