key: cord-1041185-69afkw7x authors: Yelin, Idan; Aharony, Noga; Shaer Tamar, Einat; Argoetti, Amir; Messer, Esther; Berenbaum, Dina; Shafran, Einat; Kuzli, Areen; Gandali, Nagham; Shkedi, Omer; Hashimshony, Tamar; Mandel-Gutfreund, Yael; Halberthal, Michael; Geffen, Yuval; Szwarcwort-Cohen, Moran; Kishony, Roy title: Evaluation of COVID-19 RT-qPCR test in multi-sample pools date: 2020-05-02 journal: Clin Infect Dis DOI: 10.1093/cid/ciaa531 sha: b1dc6c975b290058faee40892e00981f606d2e57 doc_id: 1041185 cord_uid: 69afkw7x BACKGROUND: The recent emergence of SARS-CoV-2 lead to a current pandemic of unprecedented scale. Though diagnostic tests are fundamental to the ability to detect and respond, overwhelmed healthcare systems are already experiencing shortages of reagents associated with this test, calling for a lean immediately-applicable protocol. METHODS: RNA extracts of positive samples were tested for the presence of SARS-CoV-2 using RT-qPCR, alone or in pools of different sizes (2-, 4-, 8- ,16-, 32- and 64-sample pools) with negative samples. Transport media of additional 3 positive samples were also tested when mixed with transport media of negative samples in pools of 8. RESULTS: A single positive sample can be detected in pools of up to 32 samples, using the standard kits and protocols, with an estimated false negative rate of 10%. Detection of positive samples diluted in even up to 64 samples may also be attainable, though may require additional amplification cycles. Single positive samples can be detected when pooling either after or prior to RNA extraction. CONCLUSIONS: As it uses the standard protocols, reagents and equipment, this pooling method can be applied immediately in current clinical testing laboratories. We hope that such implementation of a pool test for COVID-19 would allow expanding current screening capacities thereby enabling the expansion of detection in the community, as well as in close organic groups, such as hospital departments, army units, or factory shifts. The ongoing pandemic of the recently-emerged SARS-CoV-2 is critically challenging health systems worldwide. The virus is characterized by fever and severe acute respiratory syndrome 1, 2 . As of March 17, the World Health Organization (WHO) has reported over 170,000 cases with over 10,000 new diagnoses added in 24 hours 3 . Detecting carriers of the virus is fundamental to response efforts. It ensures the quarantine of COVID-19 patients to prevent local spread 1 , and more broadly informs national response measures 4 . Nevertheless, as monitoring capacity is limited, testing in most countries is generally focused on acutely ill patients, while potentially infectious carriers at the community remain undiagnosed. As many countries are already experiencing shortages of diagnosis kits and factories struggling to keep with the demand [5] [6] [7] , it has become important to come up with new ways to conserve the reagents used for diagnostic tests. At the same time, as the disease is novel, it is of value to validate any modifications to the testing process before universal adoption 8,9 . Pooling diagnostic tests has been applied in other infectious diseases and is especially attractive as it requires no additional training, equipment, or materials. In this method, first suggested by Dorfman in 1943 10 and perfected over the years [11] [12] [13] , samples are mixed and tested at a single pool, and subsequent individual tests are made only if the pool tests positive. In addition to being used in the clinic for infectious disease diagnostics in previous epidemics 14, 15 , pooling has been proven to work for RT-qPCR 16, 17 , a time-consuming step for which the reagents are expected to be in short supply 18 . Nonetheless, as SARS-CoV-2 is a novel pathogen, it is unclear how diluting a sample containing its RNA would affect the sensitivity of this assay and the false-negative rate. A c c e p t e d M a n u s c r i p t 5 Here, we test the ability of the standard RT-qPCR test for detecting a single positive sample within a pool of negative samples. First, pooling RNA extracts of clinical samples, we tested previouslyconfirmed positive samples alone and combined with an increasing number of previously-confirmed negative samples and found that positive samples can still be well observed in pools of up to 32 samples, and possibly even 64 with additional PCR cycles. Second, pooling prior to RNA extraction was tested and successfully demonstrated for 8-sample pools. Swabs from both nostrils and the throat were previously collected by healthcare providers and sent to the Virology laboratory at the Rambam Health Care Campus, Haifa, Israel. A volume of 130 microliter of the transport swab buffer was mixed with 270 microliter lysis buffer and RNA was extracted using magLEAD (Precision System Science). We obtained samples tested between March 4-15, 2020. This study was granted exemption from IRB approval for use of deidentified discarded RNA samples of COVID-19 tests by The Rambam Health Care Campus IRB committee. The original diagnostic run at the Rambam Health Care Campus was robust. Positive samples had on average 135 ± 32 -fold stronger fluorescence relative to negatives, and the positive samples reached the threshold, which we set at a fluorescence of 300 according to CDC guidelines 8 , at the 25.5± 6.1 cycle ( Figure 1A) . The five positive samples selected similarly averaged at Ct of 24.5±3.1 and maximum fluorescence of 5164±912 ( Figure 1B) . Compared to unpooled tests, the Ct of pools increased, on average, by 3.6±2.1, 2.9±0.45 and 2.4±0.11 for genes E, RdRp and N, respectively (Fig. 4, Supplementary Fig. 2) . The average increase in Ct was 2.9, which is in accordance with an 8 fold dilution of the initial samples. We found that a single clinical sample with SARS-CoV-2 RNA can be consistently detected in a pool of up to 16 to 32 samples. Pooling this way leads to only a linear increase in the threshold cycle (Ct). Even for a relatively restrictive cutoff, pooling would reach sensitivity of 96% for a pool size of 16 samples. For high-fold pooling, our data shows an estimated false negative rate of 10% (1 out of 10), which is relatively small compared to the inherent clinical sensitivity of the standard assay 20 . We further demonstrate that pooling works well also when applied prior to RNA extraction. RT-qPCR could be further optimized for the detection of low-concentration RNA. For instance, additional amplification cycles could lower detection limit allowing better detection for pools of more than 32 samples, which based on extrapolation of the data we expect would allow the 64sample pools of positive sample #5 to cross the threshold. In addition, some abnormalities as with duplicate B in positive samples #5 and #2 could have been due to interference from contamination in one or more of the samples. The unusual peaks for positive sample #2 could be due to a changing salt concentration that disturbed the TaqMan probe. Since both these issues could be solved by A c c e p t e d M a n u s c r i p t 10 further diluting the RNA samples with water, it is worthwhile to explore whether diluting samples with different ratios of water could improve the integrity of the signal in pooled results. While we tested samples with a range of different signal strengths, the detection of samples with even lower signals may warrant the use of smaller pools. Alternatively, adding a few additional PCR cycles could be considered as a means to increase detection rate of such low viral load samples. In general, as RT-qPCR kits and protocols vary internationally, use of suggested pooling may require validation for each specific setting. These results can be used not only for pooling, but also in multiplexing and any other signal compression techniques where samples are mixed to reduce the number of tests. We hope that this proof-of-concept will encourage others to develop mathematical and computational tools tailored for the pooling of SARS-CoV-2 tests. Pooling is especially useful for routine community survey and for monitoring of cohesive groups. Local and global epidemic response critically depend on determining carriage frequency in the population, which is greatly enabled by pooling techniques. Furthermore, pooling techniques can be used for routine monitoring of essential work groups, such as hospital staff, military units, and factory workers. While the frequency of infection in these groups may be low, diagnosing even a single positive person typically requires quarantine of the entire group to prevent further spread in the community. In these surveillance applications, pooling may allow more routine monitoring and detection of low frequency of carriage thereby informing policy makers, reducing transmission, and alleviating the strain on healthcare services. Triaging patients in the outbreak of the 2019 novel coronavirus Discovery of a novel coronavirus associated with the recent pneumonia outbreak in humans and its potential bat origin Coronavirus disease 2019 (COVID-19) Situation Report COVID-19) Situation Summary China, faces such a shortage of coronavirus test-kits that people say getting one is like 'winning the lottery The US is struggling to test more people for the coronavirus. Now it's facing a shortage of the materials used to run those tests Global shortage of Covid-19 test kits hits Australia as other nations limit exports Use Only Real-Time RT-PCR Protocol for Identification of 2019-nCoV High sensitivity detection of SARS-CoV-2 using multiplex PCR and a multiplex-PCRbased metagenomic method The Detection of Defective Members of Large Populations Optimality of group testing in the presence of misclassification Sequential prevalence estimation with pooling and continuous test outcomes Pooled-testing procedures for screening high volume clinical specimens in heterogeneous populations A methodology for deriving the sensitivity of pooled testing, based on viral load progression and pooling dilution Screening for the Presence of a Disease by Pooling Sera Samples Evaluation of the pooling of swabs for real-time PCR detection of low titre shedding of low pathogenicity avian influenza in turkeys High-throughput pooling and real-time PCR-based strategy for malaria Thermo Fisher to produce millions of coronavirus diagnostic tests -STAT Diagnostic detection of 2019-nCoV by real-time RT-PCR Detection of SARS-CoV-2 in Different Types of Clinical Specimens Guidance and standard operating procedure: COVID-19 virus testing in NHS laboratories A c c e p t e d M a n u s c r i p t