key: cord-1034918-nzuano6l authors: İnci, Kamil; Boyacı, Nazlıhan; Kara, İskender; Gürsel, Gül title: Assessment of different computing methods of inspiratory transpulmonary pressure in patients with multiple mechanical problems date: 2021-09-03 journal: J Clin Monit Comput DOI: 10.1007/s10877-021-00751-8 sha: f79baa08a7e1dccfa6fb44f5cfc2cecb56c0ee43 doc_id: 1034918 cord_uid: nzuano6l While plateau airway pressure alone is an unreliable estimate of lung overdistension inspiratory transpulmonary pressure (PL) is an important parameter to reflect it in patients with ARDS and there is no concensus about which computation method should be used to calculate it. Recent studies suggest that different formulas may lead to different tidal volume and PEEP settings. The aim of this study is to compare 3 different inspiratory PL measurement method; direct measurement (PL(D)), elastance derived (PL(E)) and release derived (PL(R)) methods in patients with multiple mechanical abnormalities. 34 patients were included in this prospective observational study. Measurements were obtained during volume controlled mechanical ventilation in sedated and paralyzed patients. During the study day airway and eosephageal pressures, flow, tidal volume were measured and elastance, inspiratory PL(E), PL(D) and PL(R) were calculated. Mean age of the patients was 67 ± 15 years and APACHE II score was 27 ± 7. Most frequent diagnosis of the patients were pneumonia (71%), COPD exacerbation(56%), pleural effusion (55%) and heart failure(50%). Mean plateau pressure of the patients was 22 ± 5 cmH(2)O and mean respiratory system elastance was 36.7 ± 13 cmH(2)O/L. E(L)/E(RS)% was 0.75 ± 0.35%. Mean expiratory transpulmonary pressure was 0.54 ± 7.7 cmH(2)O (min: − 21, max: 12). Mean PL(E) (18 ± 9 H(2)O) was significantly higher than PL(D) (13 ± 9 cmH(2)O) and PL(R) methods (11 ± 9 cmH(2)O). There was a good aggreement and there was no bias between the measurements in Bland–Altman analysis. The estimated bias was similar between the PL(D) and PL(E) (− 3.12 ± 11 cmH(2)O) and PL(E) and PL(R) (3.9 ± 10.9 cmH(2)O) measurements. Our results suggest that standardization of calculation method of inspiratory PL is necessary before using it routinely to estimate alveolar overdistension. Mechanical ventilation (MV) is a life saving treatment modality in patients with respiratory failure. But unless perform appropriate settings according to patient requirements and thoraco-pulmonary mechanics it might cause ventilator induced lung injury (VILI) [1] . Current standard of care for ARDS includes limiting tidal volume (V T ) and plateau pressures (P plat ). However, P plat alone is an unreliable estimate of lung overdistension during MV especially in patients with increased chest wall elastance like chest wall edema, kyphoscoliosis, and intra-abdominal hypertension. In these situations pleural pressure increases and more force is necessary to distend the chest wall [2] . Because of this reason estimation of transpulmonary pressure (PL) is recommended [3] . Recent clinical studies suggest that adjusting ventilator settings based on PL measurements may improve oxygenation, and reduce VILI [3, 4] . The measurement of PL needs the estimation of pleural pressure(Ppl) from Pes [5] [6] [7] [8] [9] . PL is the real distending force of the lung parenchyma and in general it is calculated as the difference between the Pplat and the Ppl [5] [6] [7] [8] [9] [10] [11] . On the other hand there are different usage trends of PL in the treatment of ARDS patients. First, these measurements can be used to adjust PEEP. Increased chest wall elastance (E CW ) elevates P pl and may cause negative PL. As negative PL may lead to lung or airway collapse, adjusting PEEP to achieve positive end-expiratory PL prevents collapse and optimizes lung mechanics [3, 4, 11, 12] . In general for this aim expiratory PL is used. Second, PL measurements may be used monitor for lung overdistension and to account for the increased E CW during tidal breathing. For this aim measurement of inspiratory PL is recommended [3, 7] . On the other hand there are 3 different calculation method to find PL inspiratory and there is no consencus about which one should be used. Among the three approaches directly-measured method (PL D ), is the most frequently used method in literature. With this method PL is simply calculated as the absolute difference between airway and Pes and inspiratory Pes is used for this aim [3] . Another frequently used method is the elastance-derived method (PL E ) [2] . It estimates the end-inspiratory PL as the product of P plat times the ratio between lung (E L ) and respiratory system elastance (E RS ). In this method delta P es is used as P es value. Third method is release-derived method (PLR) and is calculated as the difference between P aw and P es from end-inspiration to atmospheric pressure and requires deconnection from the ventilator of the patient [13, 14 ] (Fig. 1) . Targeting an inaccurate PL could be potentially dangerous as it may lead to over-or underinflation of the lung and there is no adequate information about the yield of PL in general ICU patients with multiple mechanical abnormalities and which measurement method should be used [15] . In general studies related with transpulmonary pressure are performed in patients with ARDS or experimental ARDS models aiming to protect patients from VILI. On the other hand, ARDS patients consists of 7-10% of ICU patients and majority of medical ICU patients are patients with chronic cardiopulmonary diseases [16] [17] [18] . They may have atelectasis, edema, risk of overdistension also during mechanical ventilation. There are very few studies investigating other patients with respiratory failure and other etiologies than ARDS. The aim of this study is to compare 3 inspiratory PL measurement methods computed in patients with multiple mechanical abnormalities. The study was approved by the ethics committee of the Gazi University Faculty of Medicine (28.12.2015/166) and performed at ICU of Department of Pulmonary Critical Care Medicine of Gazi University School of Medicine. Before enrollement written informed consent was obtained by the patient or their family. The study population comprised 34 patients > 18 years old, with different pathologies. If the patient has a contraindication to esophageal catheterization (recent upper GI surgery, bleeding varices) then was excluded from the study. All patients were intubated with an endotracheal tube no:8, sedated, paralyzed and kept in supine position. On the study day before the measurements propofol (10 mg/mL at a rate of 25-90 mg/h depending on need) was delivered intravenously and neuromuscular blocker was given till measurements were completed. The patients were ventilated with an AVEA ventilator (Viasys Health Care, Yorba Linda, CA). Baseline ventilator settings were; tidal volume 6 mL/ kg, PEEP 5 cmH 2 O, and respiratory rate 16 breaths/min. Respiratory rate was adjusted to maintain arterial carbon dioxide (PaCO 2 ) between 35 and 45 cm H 2 O and FiO 2 was set to maintain arterial oxygen saturation (%SO 2 ) between 90 and 95%. Eosephageal pressure (Pes) was measured with the AVEA ventilator and eosephageal balloon catheter (AVEA Smart-Cath Esophageal Pressure Monitoring Tube Set; 8 Fr adult. Viasys Health Care (Avea™; Carefusion)). First eosephageal ballon was positioned in the stomach to check the presence of positive deflection. Then, it was retracted until it reached the lower third of the esophagus(35-40 cm); in this position, an inspiratory occlusion was made to check for concordant changes in airway and esophageal pressure and inflated with 1.0-1.5 mL of air [8, 9, 19 ]. During an inspiratory and expiratory pauses, the airway and eosephageal pressures were measured. Following instrumentation, lung mechanics, peak (P peak ), and end-inspiratory plateau (P platei ) pressures; end-inspiratory esophageal pressure (P esei ) at P platei (P esPlat ) at PEEP (P esPEEP ) and at athmospheric pressure (P esZEEP ) the difference between these two pressures (dPes); inspiratory PL pressures (cmH 2 O); arterial blood gas values (pH, PO 2 , PCO 2 ) were recorded. Elastance of the respiratory system (E RS ), lung (E L ) and chest wall (E CW ) during tidal inflation was calculated from the change in P plat minus PEEP divided by the tidal volume E RS : The following formulas were used for assessment of transpulmonary pressures and compared with each other. PL D is based on measured 'absolute' value of P es . This method has an assumption that absolute P es by itself can be used as a surrogate of P pl and then PL is directly calculated as follows usng absolute values; Directly measured (PL D ) = P plat-Pes,ei where P plat, Pes,ei are airway and esophageal pressure during an end-inspiratory occlusion [3, 20] . Elastance-derived method (PL E ) does not require the measurement of P es atmospheric pressure but simply the change in P es : estimates the end-inspiratory PL as the product of plateau pressure times the ratio between lung (E L ) and respiratory system elastance (E RS ). [2, 4] In release derived method (PL R ) transpulmonary pressure is computed as the difference between P plat and P es from endinspiratory to atmospheric pressure (Fig. 1 ). [13, 14] Lastly we calculated accuracy values of PL E and PL R accepting PL D method as gold standard method. We accepted thereshold values for PL E as 25 cm H 2 O and for the others as 20 cmH 2 O for this analysis. Data are reported as mean ± SD or as median(IQ), unless otherwise specified, as appropriate. Statistical significance was defined as p < 0.05. Baseline and physiologic variables were compared by Student's t test for variables that were normally distributed and by Wilcoxon sign rank test for variables that were not normally distributed. To calculate correlations between the methods we used Pearson correlation coefficient. Two different formula were compared with Bland-Altman analysis(Linear regression and one sample t tests done for the Bland-Altman analyses). Accuracy of the different measurement methods were also calculated. SPSS version 20.0 (IBM, Armonk, NY, USA) was used for all statistical analyses and figures. Fourty-one patients were included in the study at the beginning but in 7 patients quality of measurements were not good and they were excluded from the study (In 2 patients we noticed later that eosephageal ballon was in the trachea. In 2 patients swinging of the eosephageal pressure tracing was not enough or there was no ossilation during the recordings and in 3 patients we noticed presence of spontaneous breathing during recordings). Demographics, diagnosis and comorbidities of the patients were given in Table 1 . Admission blood gas values: pH: 7.30 ± 0.17, PaO 2 : 73 ± 28 mmHg:, PaCO 2 :52 ± 21 mmHg, HCO 3 :24 ± 6 mEq, Sat O 2 %: 87 ± 11%. According to statistical analysis results there were good correlation and agreement and there was no bias between the measurements. Elastance derived method measured PL significantly higher than other formulas(p < 0.05). Respiratory mechanics were presented in Table 2 . There was a significant difference between the mean values of measurements according to Wilcoxon signed rank test results[13 ± 9 cmH 2 O for PL D and 18 ± 9 cmH 2 O for PL E , (p = 0.014)], and there was a good aggreement and there Correlation analysis showed good and significant correlation between the measurements (Spearmann correlation analysis: R: 0.619, p: 0.005). Since release derived PL (PL R ) calculated only in 26 patients all statistical analysis including Bland-Altman was performed in them. There was a significant difference between the mean values of measurements (PLR 11 ± 9, PL E 15.2 ± 10, p: 0.005). In Bland-Altman analysis there was a good aggreement and there was no bias between the measurements (Bias (regression): p:0.776). Agreement (One sample t test): Mean difference: 3.9 ± 10,9, p: 0.080. upper confidence limit: 25.26((10.9 × 1.96) + (3.9)) lower confidence limit: − 17.46) (− (10.9 × 1.96)- [3, 9] ). Correlation analysis showed good and significant correlation between the measurements (R:0.615, p: 0.002). Bland-Altman plots of the comparisons were given in Fig. 2 . Accuracy values of the PL E and PL R were 79% and 77% respectively when we consider PL D as gold standard method. Our study results showed that there was a good correlation and agreement between the measurements and there was no bias between them, but elastance derived method calculated PL significantly higher than the other two methods. These findings are in line with the results of other studies performed in patients with ARDS [21, 22] . Inspiratory PL is an important parameter to reflect lung overdistension and PL D is the most frequently used method fort his aim. But there is progressively increasing clinical data and debate about the differences between the PL D and PL E . Since there is no concensus about which computation method should be used to calculate PL recent literature have measured and presented the results of both measurement method in their studies. Nearly in all of them inspiratory PL calculated with PL E method is about 5 to 8 cmH 2 O higher than the PL D . In this clinical study we confirmed these results in our ICU patient population with mixed pulmonary mechanical problems. Computation method is important while using PL to set PEEP and tidal volume. Recent studies suggest that different formulas may lead different ventilator settings [23] . In a recent study comparing obese and nonobese COVID19 patients' respiratory mechanics Mezidi Berges' et al. aimed to compare estimated alveolar distension with measurement of PL E , during individual PEEP titration using two different targets: P plat 28-30 cmH 2 O or positive PLexpiratory 0-5 cmH 2 O [22] . They also monitored overdistension using PL D method. When they titrated PEEP levels according to EXPRESS protocol, PEEP was titrated on the basis of P plat . Mean PL E was 20 cmH 2 O, and PL D was significantly lower than PL E with a mean difference of 8.5 cmH 2 O (p < 0.0001). With the PLexpi protocol, mean Fig. 2 Comparison of different formulas with Bland-Altman graphic PL E was 24 cmH 2 O, and PL D was significantly lower than PL E with a mean of difference of 8.7 cmH 2 O (p < 0.0001). One of the possible explanations of why these three different method give different estimates of PL is that they uses different Pes values to calculate PL. While absolute inspiratory P es value is used to calculate PL D , delta P es (difference P es end-inspiratory and P es end expiratory) is used to calculate PL E . On the other hand PL R uses P es measured at P ZEEP at atmospheric pressure during release maneuver. This method requires deconnection of patient from the ventilator [13, 14] . As previously shown by Pelosi et al. and recently confirmed by experimental and clinical studies there is not a single P pl value among the thorax [24] [25] [26] [27] [28] . In these studies authors placed pleural catheters to the dependent and nondependent lung regions and compared P pl with the P es and showed that the static P pl increases from nondependent to dependent regions in supine position, creating a vertical pressure gradient. In Tilmonts' study PL values calculated using P es were close to those obtained from the dependent pleural catheter but higher than those obtained from the nondependent pleural catheter both during expiration and inspiration in ventilated lung transplant recipients. During controlled ventilation, the absolute value of P es was higher than the P pl of nondependent lung regions and could therefore underestimate the highest level of lung stress in nondependent lung regions. In addition, the elastance-derived method seemed useful to prevent this pitfall. Results of these studies suggest that the direct approach gives a better estimation of expiratory and inspiratory PL of the dependent lung, therefore possibly useful to set PEEP, while the elastances ratio approach better estimates the inspiratory PL of the nondependent lung so that it should be used to set the appropriate tidal volume (or pressure support) and as a target to reduce VILI. Therefore recent studies suggest that both approaches may give results for clinical application but with different meanings. Talmor and colleagues, assuming that P es equals the P pl , proposed to set PEEP as to reach a positive P L at end expiration. The rationale of this approach is that being the P es equal to P pl , a negative PL corresponds to a collapsed lung. Setting PEEP values according to inspiratory PL D they avoided overdistention of lung. This approach was proposed and tested in humans and resulted in better compliance and oxygenation, but not in a decreased mortality when compared to the traditional FiO 2 /PEEP table approach [3] . A recent randomized controlled trial was unable to show any survival benefit using this approach compared to an empirical high PEEP-FiO 2 titration table in patients with moderate to severe ARDS [20] . To eliminate the influence of the mediastinal weight and assuming that its influence is constant throughout the respiratory cycle, Gattinoni et al. proposed PL E method [2, 29] . This method assumes that the lung and respiratory system pressure-volume curves are linear in the range of PEEP and the tidal volume used in the clinical setting. The reason why this method does not use absolute P es value to calculate P L is that while delta P pl equals to delta P es , absolute P pl is not equal to absolute P es in all lung regions. This method estimates the portion of P aw that is spent to inflate the lungs, and the portion that is required to move the chest wall, based on the relative contribution of E L and E CW to E RS . In general 30 cmH 2 O of P pl in a general ARDS population is associated with a PL of approximately 21 cmH 2 O, being the average E L /E RS equal to 0.7. On the other hand, the E CW / E CW is not a fixed value, but may vary in this population from 0.2 to 0. 8 [13] . In contrast to this study we found PL E significantly higher than PL R in our study. Approximately 30% of our study population had obesity and intraabdominal hypertension (IAH) in our study. Obesity increases baseline IAP of approximately 9-15 mmHg. Obese patients have reduced lung volume with increased atelectasis, lung and chest wall compliance. Based on previous studies' data, Pelosi et al. found a relationship between IAP and chest wall elastance. They recommended that PL D measurement should be required if IAP is higher than 12 mmHg or if chest wall elastance is likely to be increased for other reasons [30, 31] . Assessment of PL seems important to understand the effect of different ventilator settings and to optimize interventions. Therefore before use it routinely for assessment of lung overdistension it is important to know how to measure it and which upper limit should be used needs to be determined and standardized. Despite all three methods reflect lung stress there are different reports giving acceptable upper limit of PL. In general limiting PLE lower than 25 cmH 2 O and PL D lower than 20 mmHg have been proposed in ARDS patientsy [5, 6] . Major limitation of our study is limited number of patients. Having more patients and to compare different calculation methods in patients with different pathologies such as obesity chest wall problems would have been more informative. Since targeting an inaccurate PL could be potentially dangerous as it may lead to over-or underinflation of the lung and thus could cause VILI and every method for the estimation of PL is based on assumptions, the choice of which method and threshold value to use should be standardize. These issues need to be supported by further trials. Conflict of interest All authors declare that they have no conflict of interest. Ethical approval The study was approved by the ethics committee of the Gazi University Faculty of Medicine (28.12.2015/166). Informed consent Written informed consent was obtained from all patients. Ventilator-induced lung injury Bench-tobedside review: chest wall elastance in acute lung injury/acute respiratory distress syndrome patients Mechanical ventilation guided by esophageal pressure in acute lung injury ECMO criteria for influenza A (H1N1)-associated ARDS: role of transpulmonary pressure Guiding ventilation with transpulmonary pressure Transpulmonary pressure: importance and limits Esophageal and transpulmonary pressures in acute respiratory failure The application of esophageal pressure measurement in patients with respiratory failure Esophageal pressure monitoring: why, when and how? Lung stress and strain during mechanical ventilation for acute respiratory distress syndrome Should we titrate positive end-expiratory pressure based on an end-expiratory transpulmonary pressure? Prospective observational study to evaluate the effect of different levels of positive end-expiratory pressure on lung mechanics in patients with and without acute respiratory distress syndrome The assessment of transpulmonary pressure in mechanically ventilated ARDS patients Interpretation of the transpulmonary pressure in the critically ill patient Value and limitations of transpulmonary pressure calculations during intra-abdominal hypertension for the LUNG SAFE Investigators and the ESICM Trials Group. Epidemiology, patterns of care, and mortality for patients with acute respiratory distress syndrome in intensive care units in countries Fifty years of research in ARDS: the epidemiology of acute respiratory distress syndrome: a th birthday review Outcome of acute hypoxaemic respiratory failure: insights from the lung safe study Transpulmonary pressure monitoring during mechanical ventilation: a bench-to-bedside review Effect of titrating positive end-expiratory pressure (PEEP) with an esophageal pressure-guided strategy vs an empirical high PEEP-FIO2 strategy on death and days free from mechanical ventilation among patients with acute respiratory distress syndrome a randomized clinical trial Transpulmonary pressures in obese and nonobese COVID-19 ARDS Ann PEEP titration in moderate to severe ARDS: plateau versus transpulmonary pressure Pleural pressure and optimal positive end-expiratory pressure based on esophageal pressure versus chest wall elastance: incompatible results Recruitment and derecruitment during acute respiratory failure: an experimental study Esophageal manometry and regional transpulmonary pressure in lung injury Regional distribution of transpulmonary pressure Oesophageal pressure as a surrogate of pleural pressure in mechanically ventilated patients Determinants of the esophageal-pleural pressure relationship in humans Targeting transpulmonary pressure to prevent ventilatorinduced lung injury Prevalence of intraabdominal hypertension in critically ill patients: a multicentre epidemiological study Chest wall mechanics and abdominal pressure during general anaesthesia in normal and obese individuals and in acute lung injury Publisher's Note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations