key: cord-1031279-7qd8bc6w authors: Mengelle, C.; Mansuy, J.‐M.; Sandres‐Sauné, K.; Barthe, C.; Boineau, J.; Izopet, J. title: Prospective evaluation of a new automated nucleic acid extraction system using routine clinical respiratory specimens date: 2012-04-12 journal: J Med Virol DOI: 10.1002/jmv.23281 sha: d3d707e85976bb088c3037683fc1a00fe9b0a32e doc_id: 1031279 cord_uid: 7qd8bc6w The aim of the study was to evaluate the MagNA Pure 96™ nucleic acid extraction system using clinical respiratory specimens for identifying viruses by qualitative real‐time PCR assays. Three extraction methods were tested, that is, the MagNA Pure LC™, the COBAS Ampliprep™, and the MagNA Pure 96™ with 10‐fold dilutions of an influenza A(H1N1)pdm09 sample. Two hundred thirty‐nine respiratory specimens, 35 throat swabs, 164 nasopharyngeal specimens, and 40 broncho‐alveolar fluids, were extracted with the MagNA Pure 96™ and the COBAS Ampliprep™ instruments. Forty COBAS Ampliprep™ positive samples were also tested. Real‐time PCRs were used to identify influenza A and influenza A(H1N1)pdm09, rhinovirus, enterovirus, adenovirus, varicella zoster virus, cytomegalovirus, and herpes simplex virus. Similar results were obtained on RNA extracted from dilutions of influenza A(H1N1)pdm09 with the three systems: the MagNA Pure LC™, the COBAS Ampliprep™, and the MagNA Pure 96™. Data from clinical respiratory specimens extracted with the MagNA Pure 96™ and COBAS Ampliprep™ instruments were in 98.5% in agreement (P < 0.0001) for influenza A and influenza A(H1N1)pdm09. Data for rhinovirus were in 97.3% agreement (P < 0.0001) and in 96.8% agreement for enterovirus. They were in 100% agreement for adenovirus. Data for cytomegalovirus and HSV1‐2 were in 95.2% agreement (P < 0.0001). The MagNA Pure 96™ instrument is easy‐to‐use, reliable, and has a high throughput for extracting total nucleic acid from respiratory specimens. These extracts are suitable for molecular diagnosis with any type of real‐time PCR assay. J. Med. Virol. 84:906–911, 2012. © 2012 Wiley Periodicals, Inc. The recent emergence of a new strain of influenza, influenza A(H1N1)pdm09, has emphasized the need for a rapid, high throughput system for identifying viruses in respiratory specimens. The diagnosis must also be able to differentiate between influenza A(H1N1)pdm09 itself and other respiratory viruses that may appear as complications. An influenza A(H1N1)pdm09 infection might lead to hospitalization in an intensive care unit where the number of available beds is limited [Fuhrman et al., 2011] . Molecular diagnosis of viral infections has become the ''gold standard'' in hospital laboratories and has replaced increasingly cellular virus culture and direct immunoassays. While real-time PCRs have improved biological management, the nucleic acid extraction step remains time-consuming and fully automated extraction instruments are needed. High quality realtime PCR assays to identify a broad range of viruses can only be performed on a high quality total nucleic acid extract. The first automated systems able to extract nucleic acids from various clinical specimens had a limited capacity, no more than 24 or 32 samples. Several high throughput automated platforms are now available. They include the COBAS Ampliprep TM (Roche Diagnostics, Meylan, France) [Sandres-Saune et al., 2007; Schumacher et al., 2007; Sizmann et al., 2007; Alp and Hascelik, 2009] , the NucliSENS easyMAG (bioMérieux, Boxtel, The Netherlands) [Loens et al., 2007; Perandin et al., 2009; Pillet et al., 2009] , the QIAGEN QIAsymphony SP (Qiagen, Valencia, CA) [Miller et al., 2010] , and the m2000 system (Abbott Laboratories, North Chicago, IL) [Scott et al., 2009 ]. All of these systems are capable of extracting high quality virus nucleic acid, but they are still timeconsuming and require 2-4 hr to extract 96 samples. The recently available MagNA Pure 96 TM (Roche Diagnostics) claims to be able to extract nucleic acids from 96 samples in less than 1 hr. The aim of this study was to examine the performance of the MagNA Pure 96 TM system and its convenience for routine use in comparison to the COBAS Ampliprep TM . The first test samples were a series of 10-fold dilutions of influenza A(H1N1)pdm09. They were all extracted using the MagNA Pure LC TM , the COBAS Ampliprep TM , and the MagNA Pure 96 TM instruments (Roche). A series of clinical respiratory specimens were then analyzed using the MagNA Pure 96 TM and the COBAS Ampliprep TM instruments by real-time PCRs. Dilutions of a control specimen. The influenza A(H1N1)pdm09 intralaboratory quality control (IQC) was diluted in minimum essential medium to obtain five 10-fold dilutions. This IQC was prepared from infected culture of Madin and Darby canine kidney cells with a clinical respiratory specimen. The nucleic acids were extracted from each dilution in duplicate using three extraction systems: the MagNA Pure LC TM , the COBAS Ampliprep TM , and the MagNA Pure 96 TM . Routine clinical specimens. A total of 239 clinical respiratory samples were collected prospectively during 1 week between January 27, 2010 and February 4, 2010 by the Department of Virology, CHU Toulouse, France. The nucleic acids were extracted from these samples using the COBAS Ampliprep TM and the MagNA Pure 96 TM instruments. They were 35 throat samples collected on virologically adapted Virocult 1 (Kitvia, Labarthe Inard, France), 164 nasopharyngeal swab samples collected on Virocult 1 , and 40 broncho-alveolar fluids collected in sterile tubes. Samples were tested for several viruses, including influenza A, influenza A(H1N1)pdm09, rhinovirus, enterovirus, adenovirus, varicella zoster virus, cytomegalovirus, and herpes simplex type 1 (Table I) . COBAS Ampliprep TM selected positive samples. Respiratory clinical samples tested positive for adenovirus (N ¼ 15), cytomegalovirus (N ¼ 9), enterovirus (N ¼ 9), and HSV (N ¼ 7) were also tested. They were nasopharyngeal specimens (adenovirus and enterovirus) and bronchoalveolar fluids (cytomegalovirus and HSV). Three systems were used: the MagNA Pure LC Total Nucleic Kit 1 (Roche) on the MagNA Pure LC TM instrument, the COBAS Ampliprep Total Nucleic Acid Isolation kit 1 (TNAI; Roche) on the COBAS Ampliprep TM instrument [Mengelle et al., 2008] , and the MagNA Pure 96 DNA and Viral NA Small Volume Kit 1 (Roche) on the MagNA Pure 96 TM . Input and output volumes were 200 and 100 ml on the MagNA Pure TM systems. The volumes were 500 and 75 ml on the COBAS Ampliprep TM . PCR tests used are described in Table I . Influenza A and influenza A(H1N1)pdm09 viruses were detected with the RealTime ready Inf A/H1N1 Detection Set 1 with the RNA virus master 1 according to manufacturer's instructions on the Light Cycler 480 TM system (Roche Diagnostics). The set contained specific primer/probe mixes for detecting influenza A matrix protein 2 (M2) and the Mexico variant specific hemagglutinin HA1 (H1) in a single pass. Rhinovirus was detected on the Light Cycler 2.0 TM real-time instrument with an in-house technique [Kares et al., 2004] . Samples underwent one step RT-PCR using the SYBR Green technique, and then melting curve analysis was used to identify rhinovirus. Enterovirus [Verstrepen et al., 2002] , adenovirus [Heim et al., 2003] , varicella zoster virus [Espy et al., 2000] , and cytomegalovirus [Mengelle et al., 2003] were detected on the Light Cycler 2.0 TM using inhouse real-time PCRs. Herpes simplex types 1 and 2 were detected on the Light Cycler 1.0 TM , as previously described [Mengelle et al., 2004] . FRET technology was used on the melting curve analysis to differentiate between HSV-1 and HSV-2. Data were analyzed using Stata TM software 9.2 (StataCorp, College Station, TX). The Kappa Cohen test was used to compare the performances of the assays. A P-value of <0.05 was considered significant. The five 10-fold dilutions of the IQC influenza A(H1N1)pdm09 were extracted each twice with the three extraction instruments, the MagNA Pure LC TM , the COBAS Ampliprep TM , and the MagNA Pure 96 TM . Each extract was tested for influenza A (target M2) and influenza A(H1N1)pdm09 (target H1). Influenza A and influenza A(H1N1)pdm09 viruses were detected in RNA extracted with all the three instruments from the undiluted IQC and in the 10 À1 , 10 À2 , and 10 À3 dilutions. Both targets were detected in both 10 À4 dilutions extracted with the MagNA Pure LC TM and the MagNA Pure 96 TM (Table II) , but only every other target was detected in 10 À4 dilutions extracted with the COBAS Ampliprep TM instrument. The H1 target was detected once, in one 10 À5 dilution extracted with the MagNA Pure LC TM instrument. Both targets were negative in 10 À5 dilutions extracted with the MagNA Pure 96 TM and the COBAS Ampliprep TM instruments. Two hundred thirty-nine clinical respiratory samples were extracted prospectively with the COBAS Ampliprep TM and the MagNA Pure 96 TM instruments. As the number of positive samples was too small, 40 additional COBAS Ampliprep TM positive samples were extracted with the MagNA Pure 96 TM instrument. Prospective samples and retrospective positive samples were pooled for analysis. Detection of RNA viruses. Sixty-four samples were tested for influenza A and influenza A(H1N1) pdm09. Three samples were influenza A (M2) positive, whereas 60 were negative. One nasopharyngeal sample was M2 positive with the MagNA Pure 96 TM and M2 negative with the COBAS Ampliprep TM . Two samples were influenza A(H1N1)pdm09 positive, and 61 were negative by both extraction methods. One nasopharyngeal sample was positive for H1 after MagNA Pure 96 TM extraction and negative after COBAS Ampliprep TM extraction. The detections of influenza A and influenza A(H1N1)pdm09 were in 98.5% concordance (P < 0.0001). Seventy-four samples were tested for rhinovirus: 16 samples were positive and 56 negative with both extraction systems. Two samples were MagNA Pure 96 TM positive and COBAS Ampliprep TM negative. The detections were in 97.3% concordance (P < 0.0001). The results for enterovirus agreed 96.8%: 10 samples were positive and 20 were negative. Both techniques disagreed on one sample which was MagNA Pure 96 TM negative (Table III) . Detection of DNA viruses. The two extraction systems agreed 100% for adenovirus: 15 samples were positive and 26 samples were negative, and for varicella zoster virus: six negative samples. Twenty-one broncho-alveolar samples were tested for cytomegalovirus: 10 were negative and 10 were positive by both methods. One sample was MagNA Pure 96 TM positive and COBAS Ampliprep TM negative. There was thus 95.2% concordance (P < 0.0001). Detection of HSV1-2 gave 95.2% concordant results: 40 samples gave identical results, 10 samples were positive, and 30 were negative (P < 0.0001). The 10 positive results gave identical herpes virus types (HSV-1). One sample was MagNA Pure 96 TM positive/ COBAS Ampliprep TM negative and one sample gave the opposite result. Both samples were bronchoalveolar specimens containing very little herpes simplex type 1 (Table IV) . Practicality. The MagNA pure 96 TM and the COBAS Ampliprep TM use the same technology: lysis to release nucleic acids that are then bound to the silica surface of added magnetic particles due to the chaotropic salt conditions and the high ionic strength of the lysing reagent. The reagents supplied with both instruments are ready-to-use and may be kept at room temperature for about 30 days. However, the instruments are used very differently: samples are dispensed in steps of 24 samples onto the Positive 15 0 0 0 0 0 10 1 11 10 1 11 Negative 0 26 26 0 6 6 0 10 10 1 30 31 Total 15 26 41 0 6 6 10 11 21 11 31 42 Automated Extraction for Respiratory Samples COBAS Ampliprep TM , so that extraction of 96 samples takes at least 4 hr. The MagNA Pure 96 TM instrument uses a 96-nozzle pipette head and 8-96 samples can be processed in a single run of less than 1 hr. The MagNA pure 96 TM also requires an input volume of 200 ml, and elution provides 100 ml. Most real-time PCR tests require an input volume of 5-10 ml, so at least seven viruses can be detected in a single extract. The performance of a new commercially available high throughput extraction system was evaluated and was compared with other, older systems on a standard sample and on routine clinical samples. Similar results were obtained when RNA was extracted from dilutions of influenza A(H1N1)pdm09 with three systems, the MagNA Pure LC TM , the COBAS Ampliprep TM , and the MagNA Pure 96 TM . Previous studies on the quantification of EBV [Mengelle et al., 2008] and CMV [Mengelle et al., 2011] in whole blood found a good correlation between the MagNA Pure LC TM system and the COBAS Ampliprep TM . Thus, these data show the importance of comparing the new extraction system, the MagNA Pure 96 TM , with another highly automatic instrument that uses identical technology, the COBAS Ampliprep TM . For this purpose, total virus nucleic acid (DNA and RNA) was extracted from respiratory specimens with both instruments and then tested for several viruses (influenza A and influenza A(H1N1)pdm09, rhinovirus, enterovirus, adenovirus, varicella zoster virus, cytomegalovirus, and herpes virus types 1 and 2) by real-time PCRs. Previous studies have shown that high throughput automated extraction systems can be used for nasopharyngeal aspirates [Chan et al., 2008] , plasma [Miller et al., 2010] , and EDTA whole blood samples [Pillet et al., 2009 ]. Chan et al. compared two automatic extraction systems (the Nuclisens easyMAG and the Qiagen BioRobot 9604) with the manual QIAamp (Qiagen) extraction method. They found no difference in the sensitivities of all three methods. Furthermore, until recently the MagNA Pure LC TM system has provided an easy-to-use platform for extracting various clinical specimens. But we believe this is the first report of using the MagnA Pure 96 TM instrument to extract nucleic acids from a large number of routine clinical respiratory specimens, including broncho-alveolar fluids, throat samples, and nasopharyngeal samples that were collected in 1 week during the influenza A(H1N1)pdm09 pandemic. The results show that the two extraction systems performed very similarly on all the clinical respiratory specimens regardless the real-time PCR technology used. Moreover, the methods performed similarly regardless of the structure of the viruses (DNA and RNA, with or without an envelope). The viruses tested can serve as models for all other viruses, including influenza B, respiratory syncytial virus, and coronaviruses. Some of theses viruses, such as varicella zoster virus, are rarely found in respiratory infections, but as this was a prospective study, all the items requested by the clinicians were tested. The use of this new automated system entails no change in sample preparation. Samples are diluted and mixed in minimum essential medium before extraction, and no extra medium, such as N-acetyl cysteine, is necessary. The MagNA Pure 96 TM can extract 96 samples in less than 1 hr and several extractions can be run in a single day if necessary. The advent of the MagNA Pure 96 TM has lead to huge changes in the organization of the laboratory. Two extractions a day are now run, so clearing all daily incoming clinical specimens. In addition, as elution volume is 100 ml, only a single extraction is needed even if several PCR tests are to be done. This greatly reduces the technician's handson time and hence the cost of the entire PCR test. As most PCR assays are now carried out on the same day, the waiting time for results has decreased dramatically, as have the number of phone calls from clinicians. The MagNA Pure 96 TM software is easy-to use, so that technicians do not require a long initial training. Maintenance is also easy; it requires only a few minutes a day. In conclusion, the MagNA Pure 96 TM high throughput automated extraction system is a reliable device for extracting total virus nucleic acids from a broad range of clinical respiratory specimens in order to detect viruses by various real-time PCR tests. A batch of 96 samples can be extracted in less than 1 hr, which allows for very rapid molecular diagnosis and thus identification of viral pathogens. Operator time is reduced thanks to ready-to use reagents that can be kept at room temperature. These features are all of great economic importance, in addition to providing rapid diagnosis and improving patient management. Comparison of 3 nucleic acid isolation methods for the quantification of HIV-1 RNA by Cobas Taqman real-time polymerase chain reaction system Comparison of the NucliSens easyMAG and Qiagen BioRobot 9604 nucleic acid extraction systems for detection of RNA and DNA respiratory viruses in nasopharyngeal aspirate samples Diagnosis of varicella-zoster virus infections in the clinical laboratory by LightCycler PCR Adult intensive-care patients with 2009 pandemic influenza A(H1N1) infection Rapid and quantitative detection of human adenovirus DNA by real-time PCR Real-time PCR for rapid diagnosis of entero-and rhinovirus infections using LightCycler Evaluation of NucliSens easyMAG for automated nucleic acid extraction from various clinical specimens Automated extraction and quantification of human cytomegalovirus DNA in whole blood by real-time PCR assay Use of two real-time polymerase chain reactions (PCRs) to detect herpes simplex type 1 and 2-DNA after automated extraction of nucleic acid Comparison of two highly automated DNA extraction systems for quantifying Epstein-Barr virus in whole blood Comparison of 2 highly automated nucleic acid extraction systems for quantitation of human cytomegalovirus in whole blood Comparison of QIAGEN automated nucleic acid extraction methods for CMV quantitative PCR testing Performance evaluation of the automated NucliSens easyMAG nucleic acid extraction platform in comparison with QIAamp Mini kit from clinical specimens Comparative evaluation of a commercially available automated system for extraction of viral DNA from whole blood: application to monitoring of Epstein-Barr virus and cytomegalovirus load Detection and quantitation of HCV RNA using real-time PCR after automated sample processing Fully automated quantification of human immunodeficiency virus (HIV) type 1 RNA in human plasma by the COBAS AmpliPrep/COBAS TaqMan system Evaluation of the Abbott m2000 RealTime human immunodeficiency virus type 1 (HIV-1) assay for HIV load monitoring in South Africa compared to the Roche Cobas AmpliPrep-Cobas Amplicor, Roche Cobas AmpliPrep-Cobas TaqMan HIV-1, and BioMerieux NucliSENS EasyQ HIV-1 assays Fully automated quantification of hepatitis C virus (HCV) RNA in human plasma and human serum by the COBAS AmpliPrep/COBAS TaqMan system Evaluation of a rapid real-time RT-PCR assay for detection of enterovirus RNA in cerebrospinal fluid specimens We thank Roche Molecular Diagnosis for providing the MagNA Pure96 TM reagents used in this study. The English text was checked by Dr. Owen Parkes.