key: cord-1026320-ninete26 authors: Chen, Grant H.; Hellman, Samuel; Irie, Takeshi; Downey, Robert J.; Fischer, Gregory W. title: Regulating inspiratory pressure to individualize tidal volumes in a simulated two-patient, one-ventilator system date: 2020-06-30 journal: Br J Anaesth DOI: 10.1016/j.bja.2020.06.043 sha: 1a2948714ebe3d6ba397d813766bd735233fd468 doc_id: 1026320 cord_uid: ninete26 nan Subsequently, this strategy was validated on four sheep in 2008 for 12 h, and on two awake humans for 10 min. 6, 7 However, prolonged support of multiple patients per ventilator is challenging due to the inability to compensate for variability in patient size and pulmonary compliance, which can also vary over the course of the disease. 8 COVID-19 patients typically require 10 days of mechanical ventilatory support, and the inability to individualize tidal volume during this time could lead to hyperventilation or hypoventilation. Therefore, numerous national societies have warned against splitting ventilators to support multiple patients. 9 In order to safely ventilate multiple patients, systems must allow individualized control of patients' tidal volumes and ensure changes in one patient do not affect the other. We present a solution affording patient-specific peak inspiratory pressure (PIP) adjustment for multiple patients using a single ventilator. During the preparation of this manuscript, we became aware of the Pressure-Regulated Ventilator Splitting (PReVentS) group solution 10 , which tackles some of these issues; similarities and differences are discussed. In our design, an adjustable fixed-pressure regulator was added at the inspiratory limb of each simulated patient's breathing circuit. Critically, the regulators have adjustable diaphragms set relative to atmospheric pressure such that the pressure for each patient is fixed; adjustments to inspiratory pressure on the ventilator do not affect delivered PIP. Thus, airway pressures (and consequently tidal volumes) are independently modulated for each patient independently of one another. Both Drager Apollo (manufacturer, location) and Puritan-Bennett 840 (manufacturer, location) ventilators were used for testing; the data presented are from a Drager Apollo. The ventilators were separately attached to two lung simulators using Y-pieces to split standard 60-inch ventilator circuits into parallel inspiratory configurations ( Figure 1A Figure 1B shows the observed tidal volumes as a function of compliance. For (b), lung 2 PIP was observed as we varied the PIP of lung 1. Figure 1C demonstrates that changing the inspiratory pressure conditions for lung 1 did not significantly change the tidal volume for lung 2. This was true even when lung 1 was completely and suddenly occluded, simulating cough or bronchospasm. Using two lung simulators placed in parallel as described, we were able to adjust two inspiratory pressures independently and achieve physiologic tidal volumes in multiple model lungs of different compliances and sizes. Allowing the inspiratory pressure to be independently modified may permit two or more patients to be connected to a single ventilator without depending on matching of patients' lung compliances and sizes. Through controlling the pressure in each circuit, appropriate tidal volumes can be achieved while lung compliance varies between patients and/or changes over time (e.g., with developing acute respiratory distress syndrome). We have not used this circuit on patients. We recommend patients be sedated and paralysed for utilization of this method. In addition, we only tested the adjustment and regulation of inspiratory pressure in relation to tidal volume. While individualized PEEP was not the focus of this test, expiratory limb 4116ANNKE regulators (manufacturer, location) could achieve individual PEEP control. The PReVentS preprint discloses pressure-regulating valves on both the inspiratory and expiratory limbs to adjust PIP and PEEP. In their system, a pressure difference within the circuit will rise and fall with varying input pressures. However, it may be advantageous to use regulators that allow a fixed absolute pressure setting relative to atmosphere, which does not change as other parameters are modified. Post-market modification of valves from AMBU bags (manufacturer, location) may be more susceptible to failure during use compared to our use of commercial pressure regulators (produced under ISO standards). Our data do not cover the full range of clinical parameters. For our studies, inspiratory times were kept fixed, although in actual patients, inspiratory times may be intermittently adjusted. Furthermore, this scheme is not intended as a permanent solution for ventilating multiple patients, and should be used only with hospital administration approval and acknowledgement of the unique ethical considerations during a crisis (such as the COVID-19 pandemic). 11, 12 Although the COVID-19 pandemic inspired our designs, it may have utility in other mass casualty scenarios such as natural disasters, terrorist attacks, and battlefield medicine. Future versions should aim to extend to more than two patients per ventilator. GWF is a consultant for and on the speaker's bureau of Edwards LifeSciences. All other authors have no conflicts to declare. Supported, in part, by US National Cancer Institute Cancer Center Support Grant P30CA008748. The Italian coronavirus disease 2019 outbreak: recommendations from clinical practice Covid-19 in critically ill patients in the Seattle region -case series Outbreak of pneumonia of unknown etiology in Wuhan China: the mystery and the miracle Clinical characteristics of 138 hospitalized patients with 2019 novel coronavirus-infected pneumonia in Wuhan, China A single ventilator for multiple simulated patients to meet disaster surge Increasing ventilator surge capacity in disasters: ventilation of four adult-human-sized sheep on a single ventilator with a modified circuit Simultaneous ventilation of two healthy subjects with a single ventilator Use of a single ventilator to support 4 patients: laboratory evaluation of a limited concept Anesthesia Patient Safety Foundation, American Association of Critical-Care Nurses Ventilato/Joint-Statement-Patients-Single-Ventilator.pdf?lang=en-US Pressure-Regulated Ventilator Splitting (PReVentS) -A COVID-19 Supporting more than one patient with a single mechanical ventilator: useful last resort or unjustifiable risk? Implementing shared ventilation must be scientific and ethical, or it risks harm