key: cord-1024745-ec498wqp authors: Aoyama, Takahiro; Uto, Koichiro; Shimizu, Hidetoshi; Ebara, Mitsuhiro; Kitagawa, Tomoki; Tachibana, Hiroyuki; Suzuki, Kojiro; Kodaira, Takeshi title: Physical and dosimetric characterization of thermoset shape memory bolus developed for radiotherapy date: 2020-10-22 journal: Med Phys DOI: 10.1002/mp.14516 sha: 4cf467a47a5ea0e1e2e4b63187d49e952ccb9ab4 doc_id: 1024745 cord_uid: ec498wqp PURPOSE: We developed a thermoset shape memory bolus (shape memory bolus) made from poly‐ε‐caprolactone (PCL) polymer. This study aimed to investigate whether the shape memory bolus can be applied to radiotherapy as a bolus that conformally adheres to the body surface, can be created in a short time, and can be reused. METHODS: The shape memory bolus was developed by cross‐linking tetrabranch PCL with reactive acrylate end groups. Dice similarity coefficient (DSC) was used to evaluate shape memory characterization before deformation and after restoration. In addition, the degree of adhesion to the body surface and crystallization time were calculated. Moreover, dosimetric characterization was evaluated using the water equivalent phantom and an Alderson RANDO phantom. RESULTS: The DSC value between before deformation and after restoration was close to 1. The degree of adhesion of the shape memory bolus (1.9%) was improved compared with the conventional bolus (45.6%) and was equivalent to three‐dimensional (3D) printer boluses (1.3%–3.5%). The crystallization time was approximately 1.5 min, which was clinically acceptable. The dose calculation accuracy, dose distribution, and dose index were the equivalent compared with 3D boluses. CONCLUSION: The shape memory bolus has excellent adhesion to the body surface, can be created in a short time, and can be reused. In addition, the shape memory bolus needs can be made from low‐cost materials and no quality control systems are required for individual clinical departments, and it is useful as a bolus for radiotherapy. Sufficient dose to the skin surface is difficult to prescribe in high-energy x-ray and electron beam used in radiotherapy because of the build-up effect. 1 Therefore, if the tumor is on the skin surface, a bolus is usually used to increase the surface dose. 1, 2 Conventional flat-type bolus (conventional bolus) is useful for areas with flat body surface, such as the abdomen and back. However, in body areas with a non-flat surface, such as the face and breast, dose distribution on the surface may not lead to significant improvements because of poor adhesion. 1 In recent years, a patient-specific bolus [three-dimensional (3D) bolus] has been made using 3D printer technology to improve the degree of adhesion on non-flat areas. [3] [4] [5] [6] Improvement of dose distribution and dose index using 3D boluses compared with conventional boluses has been reported. [7] [8] [9] [10] [11] Three-dimensional bolus is made for several hours to several days using body surface data obtained from computed tomography (CT) images or surface scanner images of patients. There are two methods of applying the 3D bolus to radiotherapy planning. In the first method, we use the 3D bolus created in advance for CT simulation (CTs). 8, 9, 11 This method requires the 3D bolus before CTs; hence, additional examination is needed to obtain body surface data. In addition, making 3D bolus sometimes requires several days; therefore, the local control rate of the tumor may reduce owing to delayed treatment initiation. 12 Another method is applying a virtual bolus for radiotherapy planning. 13 This method does not affect the treatment schedule. However, the treatment accuracy will be reduced if the shape or density of the virtual bolus is different from actual 3D bolus. Furthermore, 3D boluses cannot be reused unlike conventional boluses because they are designed specifically for patients. Therefore, it is necessary to dispose them every time treatment ends, which consumes a lot of medical resources. In particular, environmental pollution during disposal can be a problem with some materials, including acrylonitrile butadiene styrene resin, when used in 3D printing. 14 Therefore, the 3D bolus cannot be made in a short time and cannot be reused, although it has excellent adhesion to the patient's body surface. A thermoplastic mask made of poly-ϵ-caprolactone (PCL) using the immobilization system adheres to the body surface within a few minutes. 15-17 Thus, the thermoplastic boluses using the materials of immobilization systems are now available in the market 18 ; however, the shape cannot be restored and reused because the thermoplastic bolus is not chemically cross-linked. Characteristics such as excellent adhesion, short time, and reuse are important in terms of clinical and environmental aspects; however, there is no bolus that has all these characteristics. We developed a thermoset shape memory bolus using PCL polymer with shape memory characteristics and investigated whether the shape memory bolus can be applied to radiotherapy that conformally adheres to the body surface, can be created in a short time, and can be reused. Our group has been conducting research on thermoset PCL polymers with shape memory characteristics. In this study, the shape memory bolus was developed by changing the molecular weights of the thermoset PCL used in a previous study. 19 The procedure for developing the shape memory bolus from thermoset PCL is presented in Section 2.A and Fig. 1 . The shape memory bolus was developed by cross-linking tetrabranch PCL with acrylate end groups FIG. 1. Schematic illustrations of (a) macromonomer synthesis and (b) cross-linking reaction. Medical Physics, 47 (12), December 2020 (macromonomer) according to a previously reported protocol. 20, 21 Briefly, tetrabranch PCL was synthesized by ringopening polymerization of ϵ-caprolactone [Tokyo Chemical Industry (TCI) Co., Ltd, Tokyo, Japan] with pentaerythritol (TCI Co., Ltd, Tokyo, Japan) as initiators. Then, acryloyl chloride (TCI Co., Ltd, Tokyo, Japan) was reacted with the hydroxyl end group of the branched chains. The structures and molecular weights were estimated using 1 H nuclear magnetic resonance spectroscopy (JEOL, Tokyo, Japan). The average degree of polymerization of tetrabranch PCL was 100 [ Fig. 1(a) ]. The tetrabranch PCL with average degrees of polymerization of 100 was abbreviated as 4b100. The tetrabranch PCL macromonomers were dissolved at 50 wt% in xylene containing 1.5 wt% (against polymer) benzoyl peroxide (BPO; Sigma-Aldrich, St. Louis, MO, USA). The solution was injected between glass slides with a 0.2-cm thick Teflon spacer. Then, thermal polymerization was performed at 80°C overnight to obtain the crosslinked PCL [ Fig. 1(b) ]. From these processes, the shape memory bolus with a thickness of 0.15 cm was developed [ Fig. 2(a) ]. The shape of the developed bolus before deformation and after restoration was evaluated to investigate the shape memory characteristics. The Aquilion LB CT system (Canon Medical Systems, Tochigi, Japan) was used under the following conditions: tube voltage = 120 kV, tube current = 300 mA, matrix size = 512 × 512 pixels, field of view = 550 mm, and slice thickness = 1 mm. The evaluation procedure is as follows: (a) contouring from CT images of the shape memory bolus using RayStation version 6. where A and B are the two structures that were evaluated. DSC and OI can range from 0 (no overlap) to 1 (complete overlap). A is the structure of the bolus before deformation, and B is the structure after restoration. Procedures from 1 to 5 were performed three times using different samples, and the average values AE standard deviation was calculated. In addition, the DSC value was calculated by equation 1 when only the position was changed without deformation. The RANDO phantom with the shape memory bolus was scanned [ Fig. 3 (a)] to evaluate the degree of adhesion of the shape memory bolus to the body surface. Using CT images with scanning range of AE2 cm from the reference line, the air gap between the RANDO phantom and the bolus was manually contoured (Fig. 4) . The degree of adhesion was calculated by the following formula. Degree of adhesion % ½ ¼ Volume of the air gap Volume of the bolus  100 (3) In addition, a 3D bolus [ Fig.3 (b)] was generated using the NJB-300W personal 3D printer (Ninjabot, Shizuoka, Japan) with a polylactic acid (PLA) filament (3D bolus PLA ). The marching cube algorithm was applied while converting from DICOM to STL, and the Laplacian smoothing filter was utilized to smoothen the polygonal mesh with the RAD modeler (tetraface Inc., Tokyo, Japan), which is a 3D-modeling software. The printing parameters are presented in Table I Furthermore, the crystallization time after softening and stopping the heating was also measured. The water equivalent phantom was used to evaluate the dose calculation accuracy of the shape memory bolus in the treatment planning system. Next, treatment plans were compared using the RANDO phantom with virtual nasal cavity cancer to evaluate the dose distributions between the shape memory bolus and the 3D boluses. Four treatment plans using water equivalent phantoms and four types of boluses (shape memory bolus, 3D bolus PLA, 3D bolus PU , and conventional bolus) were created to evaluate the dose calculation accuracy after passing through the bolus. The shape memory bolus had a physical density of 1.15 g/cm 3 and thickness of 0.45 cm with three layers. The physical densities of the 3D bolus PLA , 3D bolus PU , and conventional bolus were 1.25, 1.03, and 1.03 g/cm 3 , respectively, with a thickness of 0.5 cm in all the boluses. These densities were examined using the objects tested. The bolus was placed on a 30 cm stack of solid water HE (nominal density = 1.032 AE 0.005 g/cm 3 ) (Sun Nuclear Co., Melbourne, Australia) on a couch top, and the source surface distance was set at 100 cm (Fig. 5) . The x-ray energy was 6 MV using TrueBeam (Varian Medical Systems, California, USA), the field size was 10 cm × 10 cm, and MU was 100. The percentage depth dose (PDD) on the central axis was obtained using a plane-parallel chamber (PTW34001) (PTW, Freiburg, German). In addition, absolute doses of 1, 5, 10, and 20 cm in depth were measured using a farmer chamber (PTW30013) (PTW, Freiburg, Germany). Dose measurements were performed five times, and the average of the five measurements was taken as the measurement value. Dose calculation was performed with the same geometric arrangement as dose measurement with RayStation. Four types of boluses were contoured and overwritten with the physical density of each bolus. 22 Collapsed cone convolution version 3.4 was used as the dose calculation algorithm with RayStation, and the dose calculation grid size was 0.1 cm. The PDD on the central axis obtained using the treatment planning system was compared with the measured value of the plane-parallel chamber. The PDD curves were normalized at a depth of 10 cm, and the wall thickness of the plane-parallel chamber was considered by shifting the measured PDD. In addition, the dose difference between the calculated doses and the measured doses with the plane-parallel and farmer chamber was evaluated as follows: The depth-scaling factor was applied by overwriting with 1.03, which is the density of the water equivalent phantom, and the applied fluence-scaling factor was 1. To investigate whether the shape memory bolus can produce the equivalent dose distribution and dose index as the 3D boluses, three treatment plans were created for virtual nasal cavity cancer. Using CT images of the shape memory bolus, the 3D bolus PLA , and the 3D bolus PU , treatment plans of 200 cGy/fraction were created for a virtual planning target volume (PTV: 26.2 mL) in the nasal cavity region of the RANDO phantom. Rings with 1 or 2 cm away from the target edge were contoured as organ at risk (OAR_1cm, 59.3 mL; OAR_2cm, 104.9 mL) (Fig. 6) . Two types of plans were created: single static field with field size of 4 cm × 4 cm and gantry angle of 0°and single arc volumetric-modulated arc therapy (VMAT) with a gantry angle of 181°to 179°. The MU value was scaled to include 95% volume of the PTV at 200 cGy (D 95% = 200 cGy). Under these conditions, D 98% , D 50% , and D 2% of PTV and D mean and D 1% of OARs were calculated from dose-volume histogram (DVH). The shape memory characteristics were evaluated using the DSC value before deformation and after restoration by the method presented in Section 2.B.1. The calculated DSC value was 0.979 AE 0.006. The DSC value was 0.975 AE 0.008 when only the position was changed without deformation. When the DSC values before deformation and after restoration were very close to 1 and almost equal to that without deformation, the shape memory bolus had an almost perfect shape memory characteristic. Additionally, because the OI value was 0.977 AE 0.011, there was no volume change. The degree of adhesion of the boluses calculated by the method presented in Section 2.B.2 is 1.9% (shape memory bolus), 3.5% (3D bolus PLA ), 1.3% (3D bolus PU ), and 45.6% (conventional bolus) ( Table II) . The adhesion of the shape memory bolus was improved compared with the conventional bolus and was equivalent to the 3D bolus PLA and the 3D bolus PU . In addition, the crystallization time of the shape memory bolus after softening was 1.5 min at room temperature. The calculated PDD curve and the measured values using the plane-parallel ion chamber and their differences are presented in Fig. 7 . For four boluses, the dose differences of calculated and measured values were within AE5.0% (range, −4.8% to 0.7%) in the build-up region and within AE1.5% (range, −1.2% to 0.5%) after the build-up region [Figs. 7(a)-7(d)]. The dose differences between the calculated dose and the measured dose using a farmer-type ion chamber at depths of 1, 5, 10, and 20 cm are presented in Table III . The dose difference was within AE1.5% (range, −1.4% to 0.5%) in all boluses, which agreed well. The dose distribution and DVH of the single static field plan and VMAT plan with the shape memory bolus, 3D bolus PLA , and 3D bolus PU are presented in Figs. 8 and 9 . The dose index is presented in Table IV . The quality of the dose distribution and dose index were equivalent in all plans. In this study, the shape memory bolus was developed using the thermoset PCL polymers and was confirmed to have almost perfect shape memory characteristics. Because the shape memory bolus has a covalent network structure after cross-linking (i.e., chemically cross-linked PCL), it is more chemically stable than the PCL used in thermoplastic masks. Therefore, the shape memory characteristics are not lost unless physically destroyed even if a high dose is irradiated and deformation and restoration are repeated. It can probably be used for at least 6 months to 1 yr, similar to commercially available thermoplastic masks. In addition, PCL has been approved for cleaning with 70% alcohol, as per the World Health Organization's (WHO) recommended cleaning procedures for COVID-19. 23,24 Therefore, the shape memory bolus made of PCL can be safely reused. Even when the shape memory bolus is destroyed and discarded, there are few adverse effects on the environment owing to the biodegradability of PCL. 25 When the average degree of polymerization of tetrabranch PCL was changed to 35 (4b35) from 100, the degree of adhesion of the 4b35 (3.4%, data not indicated) was slightly decreased compared with 4b100 (1.9%). It is considered that the deformation ability in 4b35 was reduced because the average degree of polymerization was less than that of 4b100. 21 Meanwhile, as the average degree of polymerization is larger, it is considered that the deformation ability of the polymer was large. However, excessive polymerization might not be suitable for a bolus because chemical cross-linking becomes insufficient and stability during heating decreases. In addition, the crystallization time also changes because of the change in the average degree of polymerization. 19, 20 It is considered that the shape memory bolus made from 4b100 is optimal for the bolus because the crystallization time of 1.5 min at room temperature is not too short to mimic the shape of the body surface and be clinically acceptable. The dose difference of the shape memory bolus in the PDD and absolute dose was within AE1.5%, excluding the build-up region, and was the same as other boluses ( Fig. 6 and Table III ). It may be noted that an ionization chamber was used for dose measurement. Thus, the evaluation of surface dose might not have been performed accurately. Detailed examinations, such as the assessment of varying bolus thickness, film analysis, and Monte Carlo simulation, should be performed prior to clinical use. The treatment plan for virtual tumor using the shape memory bolus was equivalent to the 3D boluses in both dose distribution and dose index (Figs. 7 and 8 and Table IV) . These results were consistent with other studies reporting that the dose calculation of 3D bolus was accurately calculated using treatment planning system and the dose distribution was improved. 7, 18 The characteristics of 3D boluses depend on the creation process, the 3D printer model, type of slicer software, and setting parameters. Therefore, even if the same material is used, the same bolus cannot be created. 26 Moreover, 3D boluses have many unclear points regarding changes during treatment period and durability and safety against radiation. 27 Therefore, to use the 3D bolus clinically, it is necessary to build a quality control flow for each facility, which requires a lot of time and labor costs. 13, 27 The shape memory bolus is advantageous as it is easy to operate because its creation procedure is similar to that of immobilization systems. In addition, PCL has long been used in thermoplastic masks for radiotherapy, and its durability and safety against radiation during the treatment period are guaranteed. Moreover, PCL polymers are inexpensive ($10/kg) compared with commercially available patient-specific 3D boluses ($2381/piece 8 ). Furthermore, PCL used in shape memory boluses has been approved by the Food and Drug Administration (FDA); hence, it is highly reliable in clinical use. Our findings confirmed that the shape memory bolus developed in this study has excellent adhesion to the body surface, can be created in a short time with low costs, can be reused, and was indicated to be useful as a bolus for radiotherapy. By contrast, once the shape memory bolus is crystallized, the shape cannot be deformed or restored unless it is reheated. Therefore, to facilitate adhesion to the patient's body surface in the event of surface changes, reshaping after reheating is required. In addition, the shape memory bolus must be shaped by pressing it on the patient's skin, which can be painful if there are wounds. There are two major limitations in this study. First, because the thickness of the shape memory bolus was 0.15 cm, there was a need to stack for compensation of buildup distance. One of the methods to increase the thickness of the bolus was the use of a thicker Teflon spacer (e.g., 3.0-5.0 mm). Accordingly, it would be possible to have large distance between glass slides to inject the solutions and to create thicker bolus. However, this may cause difficulty in achieving cross-linked uniformity due to the thermal diffusion during cross-linking reaction and affect the physical characterization of the bolus. Second, we did not evaluate the degree of adhesion to the soft areas (e.g., breast). Because it was difficult to adhere to soft areas owing to the amount of PCL polymer deformation, the 3D bolus may have better adhesion to the soft areas. To address these limitations, additional evaluation is required in the future. We developed a thermoset shape memory bolus made from PCL polymer and evaluated the physical and dosimetric characterization. The developed bolus had almost perfect shape memory characteristics and conformally adhered to the body surface. The crystallization time was approximately 1.5 min, which was clinically acceptable. In addition, dosimetric characterization was equivalent to 3D boluses. Therefore, the shape memory bolus has excellent adhesion to the body surface, can be created in a short time, can be reused, and is useful as a bolus for radiotherapy. Khan's the Physics of Radiation Therapy Utilization of custom electron bolus in head and neck radiotherapy New technique for developing a proton range compensator with use of a 3-dimensional printer Three-dimensional printer-generated patient-specific phantom for artificial in vivo dosimetry in radiotherapy quality assurance Preparation and fabrication of a full-scale, sagittal-sliced, 3D-printed, patient-specific radiotherapy phantom Towards 3D printed multifunctional immobilization for proton therapy: initial materials characterization Efficacy of patient-specific bolus created using three-dimensional printing technique in photon radiotherapy Development and validation of a 3D-printed bolus cap for total scalp irradiation Improving 3D-printing of megavoltage X-rays radiotherapy bolus with surface-scanner Design and production of 3D printed bolus for electron radiation therapy Clinical applications of 3-dimensional printing in radiation therapy Does delay in starting treatment affect the outcomes of radiotherapy? A systematic review Clinical implementation of 3D printing in the construction of patient specific bolus for electron beam radiotherapy for non-melanoma skin cancer Physical and chemical effects of ingested plastic debris on short-tailed shearwaters, Puffinus tenuirostris, in the North Pacific Ocean Inter-and intrafraction patient positioning uncertainties for intracranial radiotherapy: a study of four frameless, thermoplastic mask-based immobilization strategies using daily cone-beam CT Six dimensional analysis with daily stereoscopic x-ray imaging of intrafraction patient motion in head and neck treatments using five points fixation masks Dosimetric characterization of a rigid, surface-contourspecific thermoplastic bolus material Temperature-responsive poly (ϵ-caprolactone) cell culture platform with dynamically tunable nano-roughness and elasticity for control of myoblast morphology Temperature-responsive crosslinked poly (ϵ-caprolactone) membrane that functions near body temperature Shape-memory surface with dynamically tunable nano-geometry activated by body heat Assessing the effect of electron density in photon dose calculations CIVCO. Keeping Patients Safe and Comfortable with CIVCO Thermoplastics Alcohol and COVID-19: what you need to know Biodegradation of poly-ϵ-caprolactone under controlled composting conditions Dosimetric characterization of 3D printed bolus at different infill percentage for external photon beam radiotherapy Material matters: analysis of density uncertainty in 3D printing and its consequences for radiation oncology We are grateful to Mr. Fumiya Tanaka and Hiroki Adachi of Hitachi, Ltd., for the useful discussions. We thank Mr. Wataru Takahashi of TOYO MEDIC Co., Ltd. and Mr. Kazunari Shiohara of Engineering System Co., Ltd. for the helpful advice. This work was supported by JSPS KAKENHI Grant Number JP19H00439 and JP20H01105. The authors have no relevant conflict of interest to disclose. Author to whom correspondence should be addressed. Electronic mail: aoyamat@aichi-cc.jp; Telephone: +81-52-762-6111; Fax: +81-52-752-8390.