key: cord-1023961-p2a87gq8 authors: Parker, C. W.; Singh, N.; Tighe, S.; Blachowicz, A.; Wood, J. M.; Seuylemezian, A.; Vaishampayan, P.; Urbaniak, C.; Hendrickson, R.; Laaguiby, P.; Clark, K.; Clement, B. G.; O'Hara, N. B.; Couto-Rodriguez, M.; Bezdan, D.; Mason, C.; Venkateswaran, K. title: End-to-End Protocol for the Detection of SARS-CoV-2 from Built Environments date: 2020-08-18 journal: nan DOI: 10.1101/2020.08.16.20172668 sha: 6502121c751fb5d2a347317ecac61b4ae6acd2ac doc_id: 1023961 cord_uid: p2a87gq8 Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), the virus that causes coronavirus disease 2019, is a respiratory virus primarily transmitted from person to person through inhalation of droplets or aerosols, laden with viral particles. However, as some studies have shown, virions can remain infectious for up to 72 hours on surfaces, which can lead to transmission through contact. For this reason, a comprehensive study was conducted to determine the efficiency of protocols to recover SARS-CoV-2 from surfaces in built environments. This end-to-end (E2E) study showed that the effective combination of monitoring SARS-CoV-2 on surfaces include using an Isohelix swab as a collection tool, DNA/RNA Shield as a preservative, an automated system for RNA extraction, and reverse transcriptase quantitative polymerase chain reaction (RT-qPCR) as the detection assay. Using this E2E approach, this study showed that, in some cases, SARS-CoV-2 viral standards were still recovered from surfaces as detected by RT-qPCR for as long as eight days even after bleach treatment. Additionally, debris associated with specific built environment surfaces appeared to negatively impact the recovery of RNA, with Amerstat inhibition as high as 90% when challenged with an inactivated viral control. Overall, it was determined that this E2E protocol required a minimum of 1,000 viral particles per 25 cm2 to successfully detect virus from test surfaces. When this method was employed to evaluate 368 samples collected from various built environmental surfaces, all samples tested negative, indicating that the surfaces were either void of virus or below the detection limit of the assay. 1 Abstract 26 Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), the virus that causes 27 coronavirus disease 2019, is a respiratory virus primarily transmitted from person to person 28 through inhalation of droplets or aerosols, laden with viral particles. However, as some studies 29 have shown, virions can remain infectious for up to 72 hours on surfaces, which can lead to 30 transmission through contact. For this reason, a comprehensive study was conducted to 31 determine the efficiency of protocols to recover SARS-CoV-2 from surfaces in built 32 environments. This end-to-end (E2E) study showed that the effective combination of monitoring 33 SARS-CoV-2 on surfaces include using an Isohelix swab as a collection tool, DNA/RNA Shield 34 as a preservative, an automated system for RNA extraction, and reverse transcriptase quantitative 35 polymerase chain reaction (RT-qPCR) as the detection assay. Using this E2E approach, this 36 study showed that, in some cases, SARS-CoV-2 viral standards were still recovered from 37 surfaces as detected by RT-qPCR for as long as eight days even after bleach treatment. 38 Additionally, debris associated with specific built environment surfaces appeared to negatively 39 impact the recovery of RNA, with Amerstat inhibition as high as 90% when challenged with an 40 inactivated viral control. Overall, it was determined that this E2E protocol required a minimum 41 of 1,000 viral particles per 25 cm 2 to successfully detect virus from test surfaces. When this 42 method was employed to evaluate 368 samples collected from various built environmental 43 surfaces, all samples tested negative, indicating that the surfaces were either void of virus or 44 below the detection limit of the assay. 45 The ongoing severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) (the virus 47 responsible for coronavirus disease 2019; pandemic has led to a global slow down 48 Introduction 4 (RNA)-based detection be applied to clinical specimens (4, 5) but no such policies have been set 83 for environmental monitoring (6). Since the risk of infection from contaminated surfaces is of 84 serious concern, the need for environmental surface monitoring, along with understanding the 85 effectiveness of cleaning and disinfection is critical. In this study we outline a comprehensive 86 approach to characterize and develop an effective environmental monitoring plan that can be 87 used to understand viral persistence and elimination. 88 The ability to collect and analyze samples is fundamental to any microbial monitoring analysis. 89 During this study, a noninfectious and replication-deficient virus was used as a surrogate for the 90 SARS-Cov-2 virus to inoculate representative test surfaces and analyzed for recovery efficiency. 91 Several sampling strategies were evaluated for collecting samples from various materials. 92 Experimental parameters such as the method of viral inoculation of each surface type, collection 93 and transport, and analysis techniques were used to determine viral recovery efficiency, total 94 biomass, species-specific recovery, background contaminant levels, inhibitory factors, as well as 95 sampling and detection anomalies. 96 The overall objective of the study was to develop a standardized end-to-end (E2E) protocol for 97 the detection of SARS-CoV-2 from built environmental surfaces and to determine the minimum 98 number of RNA copies needed on fomites to positively detect virus within the limit of detection 99 of our assay. This study included collecting ~400 samples from seven surface types common to 100 materials found in the built environment and measuring the recovery efficiency of the surrogate 101 SARS-CoV-2 virus. After establishing the E2E protocol, further reproducibility studies were 102 conducted by a second laboratory for verification. 103 5 To determine the impact of the swabs and DRS transfer medium on the percent recovery 106 of viral particles, the SeraCare AccuPlex SARS-CoV-2 reference material was added to tubes 107 containing water and DRS solution, either directly into the tubes or inoculated onto swabs first, 108 before RNA extraction followed by RT-qPCR. The resulting viral copy numbers were then 109 compared and computed to understand the effects of swabs, DRS solution, and various other 110 combinations in the recovery of viral particles (Figure 1) . The RNA copies detected from 111 Accuplex placed into the water suspension (no swab) was used as the 100% positive copy 112 number reference to calculate other combinations. For practical applications, swabs should either 113 be placed in water or in a transport medium like DRS so that samples could be transported and 114 processed in the laboratory. Relative to Accuplex in water (no swab), there was a 12 % loss of 115 viral load when Accuplex solution soaked on the swab before being placed in water (swab effect). 116 Similarly, when Accuplex was placed directly into DRS instead of water (no swab), the recovery 117 was 71% (DRS effect). The double effect of the swab and DRS on viral recovery was 118 significantly less (p=0.0008), with only 21% recovery ( Figure 1) . 119 An automated RNA extraction system was compared to a manual extraction where AccuPlex 121 was inoculated on swabs containing DRS ( Figure 2 ). There were no significant differences 122 between the automated system and the manual extraction method. Subsequently, several viral 123 transport media were also compared with water. No significant differences were observed 124 between the three different viral transport media ( Figure 2 ). All tested combinations yielded 125 between 183 and 204 Nucleocapsid N1 fragment copies per 5µL RNA extract. To characterize 126 the extraction efficiency of the automated process, the Accuplex viral particles were directly 127 added to 96-well PCR plates, and subjected to thermal/enzymatic treatments before performing 128 7 Initially, when viral particles were desiccated in an Eppendorf tube and RNA extracted 151 directly, a 11% loss of RNA due to desiccation was documented when compared to the solution 152 that was not dried. The average percent recovery of the viral particles directly inoculated onto the 153 swabs in DRS was ~23% ( Figure 3B) . However, when desiccated on materials, the highest 154 percent of RNA recovery after Day 1 was observed for PETG material (1.68%), followed by PSS 155 (0.57%) and BSS (0.21%) ( Figure 3C ). The lowest observed recovery was from FRP at 0.03%. 156 ( Figure 3C ). On Day 2, viral recovery decreased on PETG (0.6%) and PSS coupons (0.16%); 157 however, no decrease was noted on Day 2 for the BSS and FRP materials ( Figure 3C ). After 158 treatment with 0.6% bleach, the recovery from BSS and PETG materials decreased to below 159 detection limit (BDL), while the bleach was not effective in the removal of RNA from PSS as 160 traces of RNA could still be detected (0.45% recovery), while only 0.03% recovery was 161 observed on FRP. This might be because the RT-qPCR assay could detect very short fragments 162 of RNA (~70 bp) and hence likely amplified degraded nucleic acids. This test revealed that viral 163 persistence on surfaces varies, and in some cases (such as on PSS) viral RNA can be recovered 164 after cleaning with bleach. 165 Reverse-transcription loop-mediated isothermal amplification (RT-LAMP) is a single step 167 colorimetric presence-absence assay that can be used as a narrow range semi-quantitative assay 168 to determine the presence of SARS-CoV-2. When used as described by the manufacturer, the 169 colorimetric RT-LAMP assay generates a yellow color for a positive result or remains 170 unchanged (pink) for a negative result. Samples with borderline results have a gradient color 171 change between yellow and pink. All samples can be further analyzed to obtain narrow range 172 semi-quantitative results by measuring the resulting DNA product with the Qubit DNA broad 173 8 range quantification kit. Since the dynamic range between positive and negative is narrow, 174 negative reactions (pink) have final DNA concentrations of <50 ng/µl post amplification, and 175 full-color positives have a post amplification of ~550 ng/µl in a 25ul reaction. Samples that are 176 borderline will have DNA yields between 50 and 550. For these studies, both positive viral 177 controls were lysed directly (Supplemental Table 1 ), and the lowest limit of detection was 178 determined at 5 and 12.5 copies per reaction for AccuPlex, and NATrol, respectively ( Figure 4A) . 179 Since the RT-LAMP assay is not truly quantitative, values < 150 ng µL -1 DNA concentration 180 were considered negative, and values ≥ 150 ng µL -1 were considered positive. 181 The samples collected from the inoculated coupon ( Figure 3A ) and analyzed by were further analyzed using the RT-LAMP assay ( Figure 4B ). When the RT-qPCR results were 183 compared with the RT-LAMP assay results, >300 copies were definitively positive with RT-184 LAMP assay (yellow coloration). However, for samples with concentrations near the limit of 185 detection (LOD) for RT-LAMP assay (~12.5 copies/µL), the results between RT-qPCR and RT-186 LAMP were less correlative. ( Figure 4B ). For samples that exhibited discrepancies between the 187 two assays (BSS2 and PETG3), Sanger sequencing was performed and confirmed SARS-CoV-2 188 sequences ( Figure 4C ) indicating the reliability of the RT-LAMP assay. 189 Since many of the chemicals associated with cleaning, disinfection, and indigenous chemical 191 constituents of the materials could have PCR inhibitors, we conducted several experiments to 192 determine this potential. The precision cleaned uninoculated surface materials (25 cm 2 ) described 193 above (BSS, PSS, PETG, and FRP), were swabbed and placed in DRS media along with 5000 194 copies of the NATrol viral control. These samples were processed along with a positive control 195 that included a swab in DRS media with NATtrol viral control but not exposed to any test 196 9 surfaces. Results indicated that all swabs used for sample collection of surface materials 197 demonstrated similar recovery rate as the controls not used in surface sampling. Analysis of 198 variance (ANOVA) indicated that swabs used to sample both BSS and FRP had similar recovery 199 rates of 25.2% and 24.3%, respectively, while PSS had 30.8% and PETG had 36.0% 200 (Supplemental Figure 1) . Swabs sampling from PETG had a significantly higher recovery rate 201 than FRP (p=0.0001) and BSS (p=0.0006), while PSS was significantly higher than FRP 202 (p=0.0152) and BSS (p=0.0439) surface types. The recovery percentages exhibited for all swabs 203 were within a standard deviation (Average 29.5% +/-5.5% standard deviation). These recovery 204 rates from various tested materials (24% to 36%) were similar to the positive control that were 205 not exposed to any test surfaces (20% to 25%; Figure 3B ), which demonstrated that the precision 206 cleaning did not leave residual organics or debris that could inhibit the RT-qPCR assay. The 207 difference in the recovery was attributed solely to the DRS-swab combination, since viral 208 particles spiked in water with swab without DRS solution demonstrated an ~88% recovery. 209 To determine whether the detection of SARS-CoV-2 virus would be affected by the debris 211 associated with built environment materials (stainless-steel metal, Amerstat, plastic, copper, 212 painted surfaces, and wood), samples were collected and efficiency of the E2E procedure was 213 tested. The collected materials in DRS solution were inoculated with and without AccuPlex (500 214 copies) viral standards prior to RNA extraction and RT-qPCR assay. As expected, there was a 215 significant inhibition in the recovery of the AccuPlex viral RNA from all surface materials 216 swabbed, ranging from 50% recovery for stainless-steel to only 4% and 8% for Amerstat and the 217 painted surfaces, respectively ( Figure 5) . Wood, copper, and plastic surfaces exhibited 218 intermediate recovery of viral particles at a level of ~20%. On average, the recovery of AccuPlex 219 10 viral RNA from the stainless-steel was significantly higher than all other tested surfaces (p=0.05 220 to 0.004), but the AccuPlex RNA recovery was more variable with a range from 15% to 100%. 221 Barring one or two outliers, the recovery from plastics was consistent ( Figure 5) . 222 To ascertain whether the decreased AccuPlex viral RNA recovery was due to the interaction 223 of the environmental debris with the organics in extraction reagents, the post-extract of the six 224 materials were spiked with 500 copies of synthetic fragments obtained from Integrated DNA 225 Technologies (IDT) prior to being subjected to RT-qPCR. In contrast to the results above, which 226 showed large inhibition, amplification of the spiked synthetic IDT fragments after the RNA 227 extraction was largely unaffected. The copper resulted in the lowest recovery at 77%, followed 228 by plastic, wood, and painted surfaces (~84%), while the stainless-steel and the Amerstat 229 exhibited 90% recovery when compared to the control ( Figure 5, Supplemental Figure 2 ). These 230 results underscore how the type of environmental surface can influence the recovery of viral 231 molecules, while the RNA purification kit chemistry might account for a small percentage of 232 such inhibition. 233 The E2E assay was repeated using the same materials by an independent laboratory for 235 reproducibility and verification purposes. The independent evaluation included LOD 236 determination of RT-qPCR assay, RNA extraction efficiency of automated system/manual kits, 237 and recovery of NATrol viral particles from various built environment material surfaces. The 238 results of the second laboratory evaluation were comparable and or equivalent to the results 239 presented here. A standalone report is included in Data Set-1. 240 241 11 The E2E protocol developed during this study could confirm viral presence from built 242 environment surfaces only when ≥1,000 viral particles per 25 cm 2 were present due to the losses 243 associated with swab collection, transportation solution, RNA extraction, and material surface 244 retention. Despite these limitations, the combination of using Isohelix swab, DRS as 245 transportation medium, automated RNA extraction, and RT-qPCR assay was determined to be 246 the best available E2E protocol during March 2020 to reproducibly detect and measure SARS-247 CoV-2 from built environment surfaces. The E2E process implemented during this study are 248 shown in Figure 6 . The samples collected were from seven different materials found in 10 249 buildings, including stainless steel, Amerstat, plastic, copper, and painted surfaces. None of the 250 368 samples collected tested positive for SARS-CoV-2 (i.e., RT-qPCR amplification for N1 gene 251 was BDL) using the E2E process developed during this study. Since the detection sensitivity of 252 the E2E process implemented was 1,000 viral particles per 25 cm 2 , the samples collected from 253 built environmental surfaces were either devoid of the targeted virus or BDL of the E2E assay. 254 The current clinical method for screening potential SARS-CoV-2 virus patients require an 256 initial throat and or nasopharyngeal swab sample collection (9). Unlike clinical samples, fomites 257 and high-touch surfaces that become contaminated with the virus display lower concentrations of 258 the virus (10), which are often difficult to detect due to method limitations and, in some cases, 259 inhibitory materials. For this reason, robust methods are imperative for the recovery and 260 detection of SARS-CoV-2 from environmental surfaces. Previous studies have analyzed a variety 261 of methods for viral recovery from surfaces (11); however, there are substantial number of 262 variables that can impact collection, processing, and quantification of viral particles. Despite the 263 World Health Organizations "How To" guide for SARS-CoV-2 surface sampling in hospital 264 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 18, 2020. . settings, there has not been a comprehensive study that adequately addresses all the issues 265 associated with an E2E assay for SARS- During this study, Isohelix swabs were selected over Copan swabs due to easier handling and 267 higher sensitivity for sample collection, and this approach has been successfully used by other 268 studies (12). Furthermore, our results demonstrated that automated RNA extraction was as 269 efficient (13, 14) as manual kits for extracting synthetic SARS-CoV-2, which has been 270 previously noted using phenol-chloroform (15). However, despite these shortcomings, two distinct molecular technologies (RT-qPCR and RT-280 LAMP assay) evolved to become part of the mainstream research toolkit for both clinical and 281 environmental testing (add mason wired paper here). The benefits of these assays, when run in 282 tandem, help resolve data associated with the more challenging and complex environmental 283 sample type to accurately detect and quantify SARS-CoV-2 from various surface materials. 284 While comparing RT-qPCR, RT-LAMP, and ddPCR to detect SARS-CoV-2 from surface 285 samples, ddPCR was found to be the most accurate and repeatable diagnostic tool. However, 286 ddPCR is more expensive and requires a specialized ddPCR instrument. In contrast, the RT-287 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 18, 2020. . https://doi.org/10.1101/2020.08.16.20172668 doi: medRxiv preprint 13 LAMP assay is faster and less labor intensive than ddPCR, is relatively inexpensive, and requires 288 minimal instrumentation to operate (e.g., a heat block, water bath or thermocycler). The results 289 of the RT-LAMP assay are colorimetric and the low infrastructure requirements make the assay 290 ideal for field testing as demonstrated in past studies (22, 23). Since RT-LAMP is a narrow range 291 semi-qualitative assay, accurate quantification is best performed by RT-qPCR; the gold standard 292 for viral RNA detection (24-26). For our studies, we selected RT-qPCR for its wide dynamic 293 range, throughput, and sensitivity (2 copies/µL) as the primary analysis tool, with RT-LAMP as 294 the confirmatory assay. 295 Once an infected person begins shedding SARS-CoV-2 viral particles, the primary route of 296 infection is via respiration; either through droplets or aerosols that are expelled during normal 297 speech, respiration, and especially sneezing, and unintentionally inhaled by even healthy 298 individuals (27). Although the virus appears to be primarily transmitted through air, SARS-CoV-299 2 can remain viable on surfaces for up to 72 hours (6, 7). Thus, similar to other respiratory 300 viruses (28), it is likely that a major route of SARS-CoV-2 infection comes from contact with 301 infected surfaces followed by inadvertent touching of the face and mouth. This pattern was 302 observed in a study in a Wenzhou, China, where numerous individuals became infected, despite 303 not having any direct contact with known patients (29). These findings, in combination with 304 virus longevity on surfaces, strongly suggest that transmission is not just limited to aerosols. 305 Additionally, preliminary research suggests that the infective dose is lower for SARS-CoV-2 in 306 comparison to other respiratory infections (30, 31). These findings highlight the importance of 307 effective environmental surveillance, surface monitoring, and proper sanitization methods to 308 eliminate the virus. 309 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 18, 2020. . 14 In this study, we analyzed several environmental surface materials that were inoculated with 310 a known concentration of SARS-CoV-2 viral reference standard to determine the recovery 311 efficiencies for each material. While each material has characteristics that contribute to recovery, 312 surface roughness and hydrophobicity are important contributors. It has been reported that 313 surface roughness is a key mitigating factor for lower recovery of biological materials (32). For 314 the test surfaces evaluated in this study, FRP had a textured surface and resulted in a lower 315 NATrol viral recovery. Even though the PETG surface had similar surface roughness as BSS and 316 PSS, the surface texture of PETG might have enabled the recovery of more virions. This was 317 evident from visual observations revealing that the dried virus inoculum was easily dissociated 318 and resulted in higher recoveries-a likely result of both the smoothness and hydrophobicity. 319 The decreasing amount of recoverable NATrol viral particles over time from the surface 320 materials are likely attributed to the combination of desiccation time (Day 1 to 8) and the use of a 321 disinfectant. Unlike all other materials, PSS demonstrated nearly the same RNA copy numbers 322 persisting from Day 1 desiccation until Day 8 post-bleach. This might be due to the RT-qPCR 323 method, which was targeting only 67 to 71 bp amplicons that could still easily be detected from 324 virion fragments degraded by the disinfectant. Despite desiccation, the chemical nature of the 325 paint associated with PSS material might have allowed viral fragments to persist even after 326 cleaning with bleach. Furthermore, after applying bleach, the pigmentation of the paint was 327 altered indicating a chemical reaction had occurred, which may have enabled easier removal of 328 viral particles from the surfaces. Similarly, recovery of viral fragments was documented on the 329 cruise ship Diamond after hypochlorite disinfection of contaminated rooms (33). These 330 surprising positive results were likely due to the detection of degraded RNA fragments still being 331 detected in the short amplicon RT-qPCR method. Due to short fragment amplification, even 332 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 18, 2020. . https://doi.org/10.1101/2020.08.16.20172668 doi: medRxiv preprint 15 significantly degraded RNA can be detected. In order to avoid these false positive results after 333 the bleach treatments, samples should be tested using an alternative technique that targets longer 334 RNA fragments, such as To detect the high concentration of SARS-CoV-2 virus in clinical samples, it has been shown 336 that direct amplification was possible with a maximum sample input of 2 µL (34), whereas for 337 environmental samples, a RNA purification step was mandatory due to the PCR inhibitory 338 substances (13, 35) . In addition, the concentration of the target molecules during RNA extraction 339 would allow larger volume input (10-fold more) and increased the detection limit (2 copies per 340 µL of RNA extract). The collection of microorganisms from environmental surfaces have been 341 documented to yield ~1 to 10% of biological materials due to issues in their removal from the 342 surfaces as well as challenges associated with their dissociation from the swab (13, 35). During 343 our study, the DRS chemistry in combination with environmental debris and RNA extraction has 344 compounded losses an additional ~80%. 345 The E2E process implemented to survey SARS-CoV-2 virus presence for built environment 346 surfaces (n=368 samples) exhibited no viral incidence (or <1,000 viral particles per 25 cm 2 ), 347 which might be attributed to a highly controlled practices that were strictly adhered. These 348 practices included but were not limited to admitting limited number of employees at a given time 349 period, training "Safe at Workplace," enforcing social distancing, wearing masks, practicing 350 personal hygiene, and deep-cleaning of the environmental surfaces might have limited the viral 351 contamination in these built environment surfaces. However, high traffic areas like hospitals, 352 restaurants, cruise ships, and subways might show a different pattern(s) of viral adherence and 353 persistence on fomites and surfaces (33, 36). 354 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 18, 2020. . https://doi.org/10. 1101 When examining all elements, the optimized E2E protocol implemented during this study 356 indicated that only ~0.5 to 2% of the viral particles could be recovered from a variety of built 357 environment surfaces and a minimum of 1,000 target molecules (viruses) per 25 cm 2 were 358 needed to positively detect the virus. During this study, it was established that 1% of NATrol 359 viral particles were recovered due to sample collection (swabs) and transportation solution 360 (DRS), and that the RNA extraction step accounted for a further 90% loss of target molecules. 361 These data reflect an overall E2E process efficiency of 0.1%, meaning that at least 1,000 copies 362 need to be present for successful and reproducible detection of the SARS-CoV-2 virus from 363 environmental surfaces. 364 Two noninfectious, replication-deficient, encapsulated SARS-CoV-2 viral reference 367 standards were used during this study, including the SeraCare AccuPlex (Milford, MA; Cat#: 368 0505-0126), which contained the ORF1a, RdRp, E, and N sequences, and the ZeptoMetix 369 NATtrol (Buffalo, NY; Cat#: NATSARS(COV2)-ERC), which contained the entire RNA 370 sequence. The AccuPlex and NATtrol stocks were purchased at a concentration of 5 x 10 3 and 5 371 x 10 4 viral particles per mL, respectively. These concentrations were confirmed in-house using 372 digital droplet PCR (ddPCR) to be within 1.25% accurate (Supplemental Table 1 ). 373 Digital droplet PCR was performed using the BioRad QX200 instrument with the IDT 374 primer/probe set for N1 and N2 with a modified probe quencher of Iowa Black ZEN/IBFQ (Cat# 375 10006770) along with the BioRad One-Step RT ddPCR advanced supermix (Cat #1864021). 376 Four methods were used for extraction of RNA from these reference materials (i.e., AccuPlex 377 and NATrol) and consisted of the following: (a) direct lysis at 75˚C for 5 min, (b) direct lysis of 378 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 18, 2020. . a 1:1 mixture of sample to nuclease-free water (15ul:15ul) to which 3ul of Proteinase K (Qiagen) 379 and 0.8ul of RNase inhibitor (Ribolock, Thermo Scientific EO0381) was added and incubated at 380 50˚C for 10 minutes followed by freeze thaw -80˚C to + 95˚C for 4 minutes, (c) utilization of 381 viral RNA extraction kits such as the QIAamp Viral RNA Mini Kit (Qiagen, Germantown, MD; 382 Cat #52904) and (d) the RNeasy Micro kit (Qiagen; Cat #: 74004). Volumes of 1, 2, 3, 5, and 7 383 µL of the Accuplex and NATrol viral standards were analyzed using methods 1 and 2 for ddPCR 384 to determine exact copy number. 385 Two protocols, involving sample collection and transport medium, were tested during 387 this study. The first protocol in Supplemental Figure 3A preservative. Sampled Isohelix swabs were broken off into the sample tube and transferred to the 400 lab at room temperature. Once in the lab, samples were extracted for nucleic acids via the 401 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The standard methodology for viral RNA extraction in this study involved using the 409 surface samples collected in DRS (~200 μL), and processing them using the automated Maxwell This extraction tube was incubated at room temp for 10 minutes and 56°C for additional 10 415 minutes. Samples were transferred to Maxwell cartridges for extraction using the Viral Total 416 Nucleic Acid program of the instrument. Purified RNA was eluted into a 60 μL of UltraPure 417 molecular grade water and divided into two aliquots. Samples were stored at −80°C with one 418 aliquot used for downstream RT-qPCR analysis while the other aliquot was archived for later use. 419 In order to compare the efficiency of the extraction protocols and the effects of the 420 DNA/RNA Shield on RNA amplification, four sets of extraction fluids were prepared in 421 triplicate. Set one was prepared with 100 µL of AccuPlex in 100 µL of UltraPure water; set two 422 was prepared with 100 µL of AccuPlex in 100 µL 95% EtOH; and sets three and four were 423 prepared with 100 µL of AccuPlex in DRS. Sets one, two, and three were all processed on the 424 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 18, 2020. Hence, only the N1 primer set was used for developing the E2E protocol. Samples that resulted 440 in a N1 positive results were further confirmed with the N2 primer set. A significantly higher 441 viral copy number was detected using the IDT reference material (1.28-fold) in comparison to 442 Twist (P < 0.05) when assessed with RNA extracted from AccuPlex as a benchmark control 443 (Supplemental Figure 4E) . 444 To determine the limits of detection of the RT-qPCR assay, a two-fold dilution series from 0 445 to 200 viral RNA copies per reaction volume (5µL; 12 replicates), were conducted and indicated 446 a LOD of 10 viral RNA copies per 5 µL reaction volume (2 copies/µl; Supplemental Figure 5A ). 447 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 18, 2020. . https://doi.org/10.1101/2020.08.16.20172668 doi: medRxiv preprint 20 Among the 12 replicates that theoretically contained one viral RNA copy (5 copies/5µl), five did 448 not reach the cycle threshold (Ct) and were thus considered as BDL. All no template controls 449 (NTC) were negative. As expected, the standard deviation of Ct values increased as the molecule 450 concentration decreased (<10 copies) (Supplemental Figure 5B) . (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 18, 2020. . https://doi.org/10.1101/2020.08.16.20172668 doi: medRxiv preprint 21 There are some unresolved issues with the RT-qPCR, which repeatedly detected higher 470 amounts of AccuPlex and NATrol in standard controls than the quantity that was being measured 471 by dd-qPCR. Supplemental Figure 6A -B shows the number of copies per mL of AccuPlex and 472 NATrol that were calculated based on RT-qPCR runs. The red lines demarcate the reported 473 concentration of viral particles for AccuPlex and NATrol. Supplemental Figure 6C shows that 474 there was a 2.69-fold higher concentration of viral particles for AccuPlex and 3.5-fold higher for 475 NATrol. 476 Titrations were performed on both AccuPlex and NATtrol viral particles for estimating copy 484 numbers. Direct RNA extraction was performed by mixing viral reference particles at a ratio of 485 1:1 (15 µL to 15 µL to water) and adding 1 µL of an RNase inhibitor (RiboLock-ThermoFisher 486 EO0381) and 3 µL of Proteinase K (Qiagen Germantown Maryland), with incubation at 50˚C for 487 10 min followed by immediately freezing at -80˚C. After freezing, the controls were 488 immediately incubated at 95˚C for 4 min, followed by duplicate titrations into the RT-LAMP 489 reaction master mix and incubation at 65˚C for 42 min. 490 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 18, 2020 . . https://doi.org/10.1101 22 Four of the most common high-touch surface materials were used in this study including bare 492 302 stainless steel (BSS), painted 302 stainless steel (PSS; white acrylic paint 168130-Rust-493 Oleum, Vernon Hills, IL), polyethylene terephthalate modified with glycol (PETG), and 494 fiberglass-reinforced plastic (FRP). All materials were smooth on a macroscale, except for FRP, 495 which exhibited an irregular, textured surface. These materials were fabricated as test "coupons" 496 of 25 cm 2 square at Jet Propulsion Laboratory (JPL) and sand tumbled to deburr. Throughout, 497 coupons were handled carefully as to limit surface damage and scratching. 498 Unless otherwise indicated, all precision cleaning was performed in a Class 100 biohazard 500 hood or a Class 100 laminar flow bench. Care was taken in handling, and high-grade chemicals 501 were used to minimize contamination. Coupons were precision cleaned based on each individual 502 material's best practices, as outlined in JPL's standard protocols (42). In short, BSS was cleaned 503 per JPL D-51981 type IV (subsequent baths of solvent, detergent AquaVantage 815 GD; Brulin 504 Holding Company, Indianapolis, IN), alkaline, and final passivation). The passivation consisted 505 of a 30-min exposure to 5 M nitric acid at 24°C. Due to the paint's associated chemical attributes 506 and susceptibility to solvents, PSS was rinsed with deionized water. The FRP wood laminate and 507 PETG were both cleaned per JPL D-51981 type V method C (solvent bath followed by deionized 508 water rinse). After cleaning, the product cleanliness level was tested to the level 100 (which 509 means particles of <0.5 µm not exceed 100 particle counts). The cleaned test coupons were 510 individually sealed in an antistatic Amerstat bag, until use. 511 Precision-cleaned coupons were opened aseptically in a biosafety cabinet and placed into 513 individual, sterile Petri dishes. Aliquots of 10 µl of the NATrol control were spotted (n=10) onto 514 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 18, 2020. . each test coupon in evenly spaced rows of 3, 4, and 3 spots and covered with a lid. Triplicate 515 coupons of each of the following material types were prepared, including several controls: (i) a 516 BSS coupon remained uninoculated (NC BSS) and were processed alongside as a negative 517 control; (ii) a swab negative control in DRS; (iii) a swab with 5,000 copies of NATtrol in DRS; 518 and (iv) 5,000 copies of NATtrol control extracted directly from Maxwell. All test coupons were 519 loaded into a modified GasPak System, Anaerobic Jar 150 LG (Cat#: 260607; Becton Dickinson, 520 Franklin Lakes, NJ) with the palladium catalyst removed and a valve port drilled into the top of 521 the lid; no desiccation beads or reactants were added. The lid port was hooked up to a vacuum 522 line to provide negative pressure on the jar. Coupons were dried at room temperature for 18 523 hours, sampled, and immediately extracted for RNA (Day 1; initial collection). After initial 524 swabbing, coupons were stored at room temperature for 24 hours at standard pressure, and 525 swabbed again with a fresh swab, followed by viral RNA extraction (Day 2; secondary 526 collection). Coupons were subsequently stored at room temperature and standard pressure for 527 another 5 days, after which coupons were treated with 10% bleach (0.6% v/v sodium 528 hypochlorite) using Kimwipes (Kimberly-Clark, Irving, TX) using a wiping pattern vertically, 529 horizontally, and diagonally. Coupons were allowed to dry for 30 min, swabbed, and extracted 530 for RNA (Day 8; collection after bleach treatment). 531 Test coupon were sampled over a 25 cm 2 area using Isohelix MS-02 buccal swabs (Cell 533 Projects, Kent, UK). Prefilled 2 mL tubes containing 200 µL of DRS and labeled with a unique 534 barcode (Cat. No.: R1100-96-1) were used for each sample. The Isohelix swab was dipped into 535 DRS solution for 15 s prior to sampling to ensure the swab was sufficiently moistened. The 536 moistened swab was then held against the sample surface at a 45-degree angle and dragged in a 537 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 18, 2020. . https://doi.org/10.1101/2020.08.16.20172668 doi: medRxiv preprint 24 raster pattern across the 25 cm 2 area. To ensure good coverage over the sample area, the raster 538 pattern was repeated three times in different directions (horizontal, vertical, and diagonal) , 539 rotating the swab head 180 degrees between the horizontal and vertical passes. The swab was 540 held to the surface perpendicular to the direction of travel to ensure that the maximal surface area 541 was covered by the swab during each pass in the raster pattern. After sample collection, each 542 swab head was transferred into the same barcoded tube used to pre-moisten the swab by 543 aseptically breaking and twisting the head off into the tube. Sample specific metadata (e.g., 544 surface type and finish) were recorded for each barcoded tube. Environmental sampling of built 545 environment surfaces was conducted in an identical manner. Environmental samples and field 546 control samples were collected in a similar manner, but instead of dragging the moistened swab 547 across a surface, the moistened swab for the field control was waved in the air for 2 min prior to 548 breaking the head off into a barcoded tube. After collection of all samples, DRS collection tubes 549 were stored at room temperature for up to three hours before RNA extraction. 550 Environmental Debris in the Recovery of Viral Particle/RNA 551 Various materials, including metal, Amerstat, plastic, wood, copper plate, and painted 552 surface, were tested to assess whether the debris associated with the environmental surface 553 affected the recovery and detection of the viral RNA. Each surface was sampled with two swabs, 554 which were preserved in DRS. The DRS from both collection tubes corresponding to one type of 555 the environmental surface was pooled together, mixed, and divided into two 200 μL aliquots. 556 One aliquot was inoculated with 100 μL of AccuPlex control and the other with 100 μL of 557 UltraPure water and extracted using the Maxwell RSC using the protocol for the RT-qPCR 558 analysis. The percent recovery for each tested material was determined when compared to the 559 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 18, 2020. . https://doi.org/10.1101/2020.08.16.20172668 doi: medRxiv preprint 25 control containing 100 μL of AccuPlex SARS-CoV-2 in 200 μL of UltraPure water. Three 560 technical RT-qPCR replicates of biological samples were used for the analysis. 561 Test surfaces were evaluated to determine if any RNA extraction inhibition was observed for 563 the AccuPlex viral particles using the Maxwell RSC system. In order to determine the potential 564 effect of the Maxwell RSC kit, 500 copies of the IDT synthetic fragments (2019-CoV N Positive 565 Control, Integrated DNA Technologies, Inc., San Jose, CA) were added to each test reaction and 566 5 μL of each extracted sample. 20 µl of the Luna master mix (Cat#: E3006; NEB, Ispsich MA) 567 was added, the plate sealed and analyzed using the QuantStudio 6F instrument. The percent of 568 the inhibition for each tested material was determined by comparing to the control containing 569 500 copies. Three technical RT-qPCR replicates of biological samples were used for the analysis. 570 In order to determine whether debris or chemistry associated with the built environment 572 surface materials contain PCR inhibitors, uninoculated test coupons were swabbed following the 573 standard swabbing procedure outlined above. The swabs were transferred to a tube containing 574 DRS and 100 µL of the NATtrol viral particles. Appropriate negative and positive controls were 575 also included. Sample were then extracted for RNA using the Maxwell RSC and were analyzed 576 via RT-qPCR. Inhibition was determined by comparison of the extraction ratios between positive 577 control and the sample reaction mixtures. 578 For the field data collection and associated metadata characteristics of samples, JPL 580 Information and Technology Solutions Directorate created a custom mobile application (Safe 581 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 18, 2020. . Workspace Analysis Barcode Scanner, or SWABS) that uses an iPhone to capture tracking 582 metadata at each stage of the sample collection and analysis process. The vendor-provided 583 barcode of each sample container was chosen as the unique identifier of each sample tracked 584 through each phase of analysis. In the first phase (sampling), the container barcode was scanned 585 and relevant metadata such as the name of the collection personnel, location, surface material 586 properties, time of collection, material lot numbers, and optional description details were 587 recorded for each sample. To accelerate data entry and eliminate human errors, barcode scanning 588 with the iPhone camera was used to exactly identify each sample, and data in common were 589 retained and automatically reused. In the second phase (RNA extraction), the sample was 590 scanned once again, and details of the Maxwell machine identifier, extraction tip size, and lot 591 numbers were added. For the third phase (archival), analysts added details of the cryo box 592 location, extraction tip size, and lot numbers for each sample. At the final qPCR stage, the record 593 was completed with the qPCR machine identifier, extraction tip size, and lot numbers, as well as 594 the Ct score and copy number determined by the analysis. At each stage of the analysis 595 procedure, the operations procedure version number was also recorded to track which 596 documented procedure was followed at the time when each sample was processed. All of these 597 sample processing data were gathered and stored in a centralized database at JPL. Leveraging 598 this database, the SWABS web application makes these data accessible for viewing, searching, 599 or editing, as well as providing reporting capabilities to communicate and summarize any of the 600 data on demand. Although the SWABS application streamlines and improves the accuracy of the 601 processing metadata recording process as a whole, it is most advantageous during the initial 602 sample collection phase owing to its mobile platform that allows its users to move around freely 603 within the workspace environment while minimally encumbered by support equipment. 604 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 18, 2020 . . https://doi.org/10.1101 Statistical Analyses 605 All statistical analyses were performed using GraphPad Prism Version 8.2.0 (GraphPad 606 Software, San Diego, California USA). Specifically, Welch's t-test and a two-way ANOVA 607 followed by a post-hoc sample correction were computed. Outliers were screened using the rOut 608 method from the robustX R package (https://CRAN.R-project.org/package=robustX). Oscar Rendon Perez (coupon material preparation and precision cleaning). We would also like to 620 thank Brent Mcwatters, Mark Powell, and the members of the JPL Information and Technology 621 Solutions Directorate that rapidly created an iPhone app that was used for metadata collection 622 during this project. We are indebted to the personnel involved in JPL shipping / receiving for 623 their help during this pandemic and the facility managers associated with the surface collection 624 locations. Additionally, Mikael Kubista from the TATAA Biocenter for helping to develop our 625 RT-qPCR limit of detection protocols, thanks to MetaSUB consortium members (especially 626 Benjamin Young) for generating the sample collection SOP, and Julie Dragon for RT-LAMP 627 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 18, 2020 . . https://doi.org/10.1101 28 assay SOP and continued support. Christopher Fleming and Frank Tansley Laboratory, and their employees make no warranty, expressed or implied, or assume any liability 644 or responsibility for the accuracy, completeness, or usefulness of information, apparatus, product, 645 or process disclosed in this manuscript, or represents that its use would not infringe upon 646 privately held rights. The use of, and references to any commercial product, process, or service 647 does not necessarily constitute or imply endorsement, recommendation, or favoring by the US 648 Government, NASA, California Institute of Technology, or Jet Propulsion Laboratory. Views 649 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. Work for this study way also completed by Biotia, Inc. Biotia and its employees make no 653 warranty, expressed or implied, or assume any liability or responsibility for the accuracy, 654 completeness, or usefulness of information, apparatus, product, or process disclosed in this 655 manuscript, or represents that its use would not infringe upon privately held rights. 656 KV coordinated with all authors in designing the concept, executed the study, implemented 658 the project, involved in the data analyses, and wrote the manuscript. CWP was involved in 659 establishing the RT-qPCR assay with NKS, and was instrumental in executing the E2E process, 660 and wrote the manuscript along with KV. NKS carried out the QA/QC of RT-qPCR assay and 661 analyzed the data. ST helped design and setup the experimental methods, provided training, 662 conducted RT-LAMP, qPCR, ddPCR, interpretation of data, and assisted in writing and editing 663 the manuscript. PL carried out lab work pertaining to protocol optimization of RT-LAMP, qPCR, 664 and ddPCR assays as well as Sanger sequencing. NKS, JMW, CWP, RH, and KC carried out 665 environmental sampling from various built environment surfaces. JMW managed metadata 666 collection and curation, helped analyze data, and was crucial in surface sampling as well as 667 drafting portions of the manuscript. AB performed RNA extraction of the collected samples, 668 contributed to data analysis, interpretation, and drafted parts of the manuscript. CU helped 669 analyze the data, write the paper, and assisted in the study design. RH helped with sample 670 collection and contributed in writing about the swab and DNA extraction methodology process 671 selection for the paper. KC planned and coordinated built-environment sampling and supported 672 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 18, 2020. . https://doi.org/10.1101/2020.08.16.20172668 doi: medRxiv preprint Isohelix swab (◆). Each extraction condition was then divided by the average copy numbers 848 generated from the water with no swab (theoretical highest yield) to get percent recovery and 849 plotted with columns representing their mean percentage. Welch's t test was used to determine 850 significant differences between extraction conditions, significance (p<0.05) denoted by '*'. 851 Direct PCR was treated as 100% to calculate the extraction efficeny of the other extraction 858 methods (recorded with in the columns). Significant differences were determined by Welch's t 859 test, significance (p<0.05) denoted by '*'. 860 were then sacrificed to extract RNA directly without removing from the surface. In addition, 866 aliquots of viral particles that were not desiccated but inoculated in DRS and swab materials 867 were also processed (DRS Swab, ■). (C) Viral particles were collected from the seeded surfaces 868 with Isohelix swabs and DRS, extracted on the Maxwell RSC, and quantified using RT-qPCR 869 assay. Viral RNA copy number for each condition was divided by an extraction control to 870 calculate percent recovery for Day 1 (▼), Day 2 (•), and Day 8 post bleach (■). Statistical 871 significance was determined by Welch's t test with significance (p<0.05) denoted by '*'. 872 Values that were not tested were marked as not applicable (NA), and values that were 883 undetectable were recorded as BDL. 884 with inactivated viral AccuPlex particles prior to RNA extraction and quantification with RT-887 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 18, 2020. . Painted Surface (•), and Wood (•) was compared to a positive control (•) and reported as 889 percent recovery compared to that positive control mean. Each column represents average 890 percent recovery for respective surface type. Significance (p<0.05) denoted by '*', based on 891 Welch's t test. 892 Figure 6 : Environmental surface testing using E2E protocol. The optimized E2E protocol for 893 detecting SARS-CoV-2 virus on surfaces is a 5 part procedure: (1) Surface sample collection; (2) 894 Viral transport medium; (3) RNA extraction; (4) RT-qPCR assay; and (5) Test results. 895 Tables 896 NA 897 898 All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 18, 2020. . 71.2% 20.6% All rights reserved. No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted August 18, 2020. Loop mediated isothermal amplification (LAMP) assays 691 as a rapid diagnostic for COVID-19 What do we learn from SARS-CoV-1 to SARS-CoV-2: 693 Evidence from global stock markets Uncertainties about the transmission routes of 2019 698 novel coronavirus Applying principles of behaviour change to 700 reduce SARS-CoV-2 transmission Aerosol and surface stability of 703 SARS-CoV-2 as compared with SARS-CoV-1 705 Detection of Severe Acute Respiratory Syndrome Coronavirus 2 RNA on Surfaces in 706 Quarantine Rooms Persistence of coronaviruses on 708 inanimate surfaces and their inactivation with biocidal agents Exaggerated risk of transmission of COVID-19 by fomites Epidemiology, transmission dynamics and control of SARS: 715 the 2002-2003 epidemic Global Genetic Cartography of 724 Urban Metagenomes and Anti-Microbial Resistance Evaluation of procedures for the collection, 728 processing, and analysis of biomolecules from low-biomass surfaces Performance 731 evaluation of the Maxwell 16 System for extraction of influenza virus RNA from diverse 732 samples Phenol-734 chloroform-based RNA purification for detection of SARS-CoV-2 by RT-qPCR: 735 32 comparison with automated systems Combination of RT-qPCR testing and clinical 738 features for diagnosis of COVID-19 facilitates management of SARS-CoV-2 outbreak The Architecture Direct RNA nanopore sequencing of full-length coronavirus genomes provides 747 novel insights into structural variants and enables modification analysis Assessment of Capture and Amplicon-Based Approaches 752 for the Development of a Targeted Next-Generation Sequencing Pipeline to Personalize 753 Lymphoma Management SARS-CoV-2 detection using 755 digital PCR for COVID-19 diagnosis, treatment monitoring and criteria for discharge. 756 medRxiv Real-time reverse transcription 758 loop-mediated isothermal amplification for rapid detection of West Nile virus Rapid and real-time detection of Chikungunya virus by reverse 762 transcription loop-mediated isothermal amplification assay Comparison of two 764 extraction methods for the detection of hepatitis A virus in semi-dried tomatoes and 765 murine norovirus as a process control by duplex RT-qPCR Respiratory virus shedding in exhaled breath and efficacy of face masks Significance of fomites in the spread of respiratory and 777 enteric viral disease Indirect virus transmission in 779 cluster of COVID-19 cases Clinical features of COVID-19 in elderly patients: A 781 comparison with young and middle-aged patients Role of membrane surface 785 morphology in colloidal fouling of cellulose acetate and composite aromatic polyamide 786 reverse osmosis membranes Environmental sampling for severe acute respiratory syndrome 788 coronavirus 2 (SARS-CoV-2) during a coronavirus disease (COVID-19) outbreak aboard 789 a commercial cruise ship DIRECT RT-794 qPCR DETECTION OF SARS-CoV-2 RNA FROM PATIENT NASOPHARYNGEAL 795 SWABS WITHOUT AN RNA EXTRACTION STEP Air, 801 Surface Environmental, and Personal Protective Equipment Contamination by Severe 802 Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) From a Symptomatic Patient Evaluation of procedures for the collection, 806 processing, and analysis of biomolecules from low-biomass surfaces Comparative analysis of methods for the 809 purification of DNA from low-biomass samples based on total yield and conserved 810 microbial diversity Global Genetic Cartography of 820 Urban Metagenomes and Anti-Microbial Resistance Detection of 2019 novel coronavirus 826 (2019-nCoV) by real-time RT-PCR. Euro surveillance : bulletin Europeen sur les 827 maladies transmissibles Shotgun Transcriptome and Isothermal Profiling of SARS-CoV-2 Infection Reveals 836 Unique Host Responses, Viral Diversification, and Drug Interactions General Cleaning of Materials No reuse allowed without permission. (which was not certified by peer review) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity