key: cord-1023200-4rxjc4pp authors: Carter, Nikki; Clausen, Maryam; Halpin, Rebecca A.; Blackmore, Colin; Cai, Kang; Delpuech, Oona; Kohlmann, Alexander; Magnusson, Otto; March, Ruth; O’Neill, Daniel; Prakash, Kasthuri; Sherwood, James; Sundin, Tabetha; Swift, Jason; Tarakameh, Azar; Wijdicks, Marilou; Wise, Daniel; Fidock, Mark title: A novel automated SARS-CoV-2 saliva PCR test protects a global asymptomatic workforce date: 2021-06-16 journal: Sci Rep DOI: 10.1038/s41598-021-92070-w sha: 87e7d7c0c04b014f0f1bb7d191bbfd7bfb118f69 doc_id: 1023200 cord_uid: 4rxjc4pp Regular PCR testing of nasopharyngeal swabs from symptomatic individuals for SARS-CoV-2 virus has become the established method by which health services are managing the COVID-19 pandemic. Businesses such as AstraZeneca have also prioritised voluntary asymptomatic testing to keep workplaces safe and maintain supply of essential medicines to patients. We describe the development of an internal automated SARS-CoV-2 testing programme including the transformative introduction of saliva as an alternative sample type. 1. Employee requests a test via a bespoke IT application that includes the employee recording potential symptoms for COVID-19 and any other information required by local laws. 2. An HCP confirms that the individual is asymptomatic, has provided informed consent and orders the test. 3 . Collection of the sample as advised by manufacturers' and local health authorities' instructions. www.nature.com/scientificreports/ As expected, this testing cascade in asymptomatic individuals requires high analytical sensitivity. In our asymptomatic cohort, positive samples had a mean Ct value of 32 (Ct range of 40-15), where a Ct of 32 corresponds to a sample with a viral RNA load of approximately 1 × 10 5 copies per mL, consistent with reported values in similar populations 3 . Interestingly, although higher Ct values are associated with asymptomatic samples, compared to samples from symptomatic cohorts, these differences are neither consistent nor statistically significant 4 . Recent publications have proposed that saliva collection reflects peak infection, predicts COVID-19 outcome and is associated with lower overall cost 5-7 compared to nasopharyngeal collection. This project was facilitated by a bespoke IT application accessible via mobile phones, desktops or other computers. Test and sample data is linked to each AstraZeneca employee via their personal record identifier and a unique sample identity. The IT app ensures that the connection between this identity and sample is segregated and encrypted, while personally sensitive information is de-identified and high-fidelity sample tracking is facilitated. An average of 0.7% samples tested to date were positive in this voluntary internal programme, and individuals and their close contacts were advised to self-isolate according to national guidelines (https:// www. ecdc. europa. eu/ sites/ defau lt/ files/ docum ents/ Guida nce-for-disch arge-and-ending-of-isola tion-of-people-with-COVID-19. pdf; https:// www. cdc. gov/ coron avirus/ 2019-ncov/ if-you-are-sick/ quara ntine. html). Overall, this programme has achieved its objective of maintaining a safe workplace environment by keeping the infection rate approximately tenfold lower than that observed in the respective communities. Establishing asymptomatic testing. genesig COVID-19 Real-Time PCR (UK and Sweden). AstraZeneca laboratories in Europe used the genesig COVID-19 Real-Time PCR assay (PrimerDesign) (PD) (https:// www. fda. gov/ media/ 136309/ downl oad), approved in Europe for use with NPS and saliva samples. This assay detects the ORF1ab gene and can detect 0.33 copies of whole viral genome RNA/µL, with a sensitivity of ≥ 95% and 100% specificity. The bioinformatic analysis of SARS-CoV-2 genomic epidemiology published on the GISAID EpiCoV database is reviewed weekly by PD to verify that the assay is effective for new variants. The assay was verified for linearity, reproducibility, user, temporal and instrument variability at AstraZeneca laboratories. Assessments were performed using SARS-CoV-2 RNA contrived samples (Twist Synthetic SARS-CoV-2 RNA (Control 1 and Control 2), Twist Bioscience, https:// www. twist biosc ience. com/ resou rces/ produ ctsheet/ twist-synth etic-sars-cov-2-rna-contr ols), demonstrating that the assay was performing within acceptable limits as laid out in the instructions for use (IFU). Linearity was shown to be high, with 93% of 14 experiments meeting the slope acceptance criteria (1.0 ± 0.15) and 100% of 18 runs having an R 2 ≥ 0.99 and meeting the acceptance criteria of ≥ 0.95. Limit of detection (LoD) was confirmed as 1 copy/µL in 98% (41/42) of replicates, which is suitable for viral detection in asymptomatic individuals. Intra-inter assay precision experiments indicated that the assay detected the contrived samples consistently. When an average Ct from replicate plates (same RNA extraction) was used for classification, 100% detection was observed for dilutions from 20 to 200,000 copies of RNA per well across 4 days of testing. In the AstraZeneca USA CLIA laboratory, the PD kit was not approved for use and available at the time of laboratory set up. Hence, assessments were performed using the TaqPath RT-PCR COVID-19 Combo Kit (ThermoFisher Scientific Inc.) (TF) (https:// www. fda. gov/ media/ 137450/ downl oad). The TF Kit has specific target sequences for 3 genes: ORF1ab, N Protein, S protein. According to the manufacturer, the LoD of the TF Kit is 10 copies of whole viral genome RNA, which will detect ≥ 95% positive samples. Verification experiments used the same panel of Twist SARS-CoV-2 RNA contrived samples and successfully demonstrated assay linearity and dynamic range down to 10 copies per qRT-PCR reaction LoD. Reproducibility was demonstrated using spiked-in RNA level of 10 copies per reaction. Transition to saliva samples. Sample collection. Following a comparison of extraction and amplification efficiency for different saliva sample collection devices with appropriate regulatory approval, GeneFix (https:// isohe lix. com/ wp-conte nt/ uploa ds/ 2020/ 12/ GeneF ix-broch ure-2020_ v6. pdf) was selected for use in Europe, and DNA Genotek OMNIgene (https:// www. fda. gov/ media/ 143416/ downl oad) in the US. These devices incorporate a funnel to facilitate saliva collection into a tube and contain a proprietary stabilisation buffer, enabling remote collection. Aligned to the instructions for use, employees were advised to be well hydrated and to fast for 30 min before sample collection 8 . Employees were supervised but self-collected their saliva samples 9 . Sample handling. To simplify the protocol, the lysis buffer, proteinase K and the internal extraction control (IEC) could be premixed up to 2 h in advance of use and added to a saliva sample simultaneously without impairing stability or sensitivity of RNA extraction. Clinical sensitivity. 20 paired saliva samples and NPS specimens were obtained from patients determined to be positive for SARS-Cov-2 by a separate qRT-PCR assay (AllPlex 2019-nCOV, https:// www. fda. gov/ media/ 137178/ downl oad) performed no more than 3 days prior to collection. Analysis demonstrated that the saliva sample was positive in every matched pair in which the NPS was positive (Fig. 2a) . Analytical validity. The analytical validity of saliva testing was evaluated through a comparison with NPS testing for impact on extraction, amplification and sensitivity and SARS-CoV-2 detection using contrived sam- The sensitivity for detecting SARS-CoV-2 was similar in contrived saliva and nasal samples for both assays used (Fig. 2b) . In both cases assay failures were only seen at the lowest concentration and at comparable rates. Saliva did not impact extraction or amplification. Interlaboratory assay concordance was monitored and maintained by the regular evaluation of the Qnostics SARS-CoV-2 Analytical Q Panel 01 [Cat # SCV2AQP01-B [USA] ]. Optimisation of saliva testing. Saliva testing protocols were optimised as follows to ensure a low intrinsic test failure rate. This resulted in a failure rate of < 0.5% of samples per run. Input volumes between 50 and 600 µL were evaluated, 600 µL limited RNA binding to the Beckman Coulter beads during extraction whereas 50 µL resulted in increased assay failures (https:// www. beckm an. com/ search# q= C5852 9AA&t= coveo-tab-techd ocs). The input volume for both PCR tests was standardised at 200 µL. Extending the reverse transcription time to 30 minutes for saliva samples (N=8) reduced assay failure and improved sensitivity (Ct values), although the effect size was relatively small. The introduction of automated sample handling (Biomek i5 and i7 liquid handlers for transfer from collection tube to plate, lysis and heat inactivation and magnetic bead extraction of RNA, https:// www. beckm an. com/ landi ng/ ppc/ liquid-handl ers/ biomek/ i-series) reduced processing time by around 20% and reduced test failures 10-fold, from 5 to 0.5%, compared to manual testing. (iv) Addition of agents One of the major challenges in the transition to automated testing of saliva samples was sample viscosity. The addition of 50 µL of 1M dithiothreitol (DTT, Sigma Aldrich) to 2 ml saliva plus stabiliser reduced sample viscosity and improved automated tube to plate transfer. Addition of DTT did not affect test sensitivity of either synthetic contrived or clinical SARS-CoV-2 samples. Study participation and methods. Informed consent was obtained from all participants within this internal SARS-CoV-2 testing program, consent was reviewed by the AstraZeneca Bioethics Advisory Board. This study was approved by the AstraZeneca internal review body with all methods performed in accordance with local relevant guidelines and standard operating procedures from AstraZeneca. Sourced samples for methods development were obtained according to AstraZeneca Human Biosamples policy, which includes review of informed consent and ethics approvals. Saliva or nasopharyngeal swab specimens for detection of SARS-CoV-2 Patterns of SARS-CoV-2 testing preferences in a national cohort in the United States Viral dynamics in asymptomatic patients with COVID-19 The early phase of the COVID-19 outbreak in Lombardy Temporal profiles of viral load in posterior oropharyngeal saliva samples and serum antibody responses during infection by SARS-CoV-2 Saliva viral load is a dynamic unifying correlate of COVID-19severity and mortality The sensitivity and costs of testing for SARS-CoV-2 infection with saliva versus nasopharyngeal swabs Challenges in use of saliva for detection of SARS CoV-2 RNA in symptomatic outpatient Saliva offers a sensitive, specific and non-invasive alternative to upper respiratory swabs for SARS-CoV-2 diagnosis The authors would like to acknowledge the following for their contributions to the testing programme: Daisy Belfield, Graham Belfield, Martin Booth, Helen Brown, Benjamin Guscott, Linda Kitching, Ingela Maxvall, Patrick O'Shea, Marta Potapo, Katya Ribeiro Passos, Jonathan Sumner, Paul Turner, Abhishek Upadhyay, Liisa Vilén. Each author has made substantial contributions to the conception and design of the work, or the software used, including data analysis and interpretation. All authors have contributed to the draft and have reviewed the submitted manuscript. Correspondence and requests for materials should be addressed to M.F. Publisher's note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations. License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http:// creat iveco mmons. org/ licen ses/ by/4. 0/.