key: cord-1021599-i8rbqens authors: Zhou, Yuyan; Ni, Zhong; Ni, Yuenan; Liang, Binmiao; Liang, Zongan title: Comparison of Actual Performance in the Flow and Fraction of Inspired O(2) among Different High-Flow Nasal Cannula Devices: A Bench Study date: 2021-05-04 journal: Can Respir J DOI: 10.1155/2021/6638048 sha: 07341fb00e935e7308e6b3135c0cb9e554b5ac79 doc_id: 1021599 cord_uid: i8rbqens BACKGROUND: High-flow nasal cannula (HFNC) oxygen therapy has been recommended for use in coronavirus disease 2019 (COVID-19) patients with acute respiratory failure and many other clinical conditions. HFNC devices produced by different manufacturers may have varied performance. Whether there is a difference in these devices and the extent of the differences in performance remain unknown. METHODS: Four HFNC devices (AIRVO 2, TNI softFlow 50, HUMID-BH, and OH-70C) and a ventilator with an HFNC module (bellavista 1000) were evaluated. The flow was set at 20, 25, 30, 35, 40, 45, 50, 60, 70, and 80 L/min, and the FiO(2) was set at 21%, 26%, 30%, 35%, 40%, 45%, 50%, 60%, 70%, 80%, and 90%. Then, one side of the cannulas was clipped to simulate the compression, bending, or blocking of the nasal cannulas. The flow and FiO(2) of the delivered gas were recorded and compared among settings and devices. RESULTS: The actual-flow and actual-FiO(2) delivered by different settings and devices varied. AIRVO 2 had superior performance in flow and FiO(2) accuracy. bellavista 1000 and OH-70C had good performance in the accuracy of actual-flows and actual-FiO(2,) respectively. bellavista 1000 and HUMID-BH had a larger flow range from 10 to 80 L/min, but only bellavista 1000 could provide a stable flow with an excessive resistance up to 60 L/min. TNI softFlow 50 had the best flow compensation and could provide sufficient flow with excessive resistance at 20–50 L/min. CONCLUSIONS: The variation in flow, FiO(2) settings, and devices could influence the actual-flow and actual-FiO(2) delivered. AIRVO 2 and OH-70C showed better FiO(2) accuracy. TNI softFlow 50, bellavista 1000, and HUMID-BH could lower the risk of insufficient flow support due to accidental compression or blocking of the cannulas. In addition, ventilators with HFNC modules provided comparable flow and FiO(2) and could be an alternative to standalone HFNC devices. High-flow nasal cannula (HFNC) oxygen therapy can deliver heated and humidified high-flow oxygenated gas via nasal cannulas with flow ranges from 10 to 80 L/min in adults depending on the manufacturer. Such flow can cover the physiological inspiratory flow needed by patients to achieve a stable fraction of inspired O 2 (FiO 2 ) of 21-100%. Previous studies have shown that HFNC therapy is easy to apply and prevents nasal epistaxis [1, 2] . In addition, HFNC therapy has multiple physiological advantages, including generating positive end-expiratory pressure (PEEP) [3] , washing out dead space [4] , decreasing inspiratory effort [3, 5] , and improving lung volume and compliance [5] . Evidence suggests that HFNC therapy has some benefits in providing better oxygenation than conventional oxygen therapy (COT) [6, 7] . HFNC therapy could reduce intubation-related adverse events when used as a preoxygenation device [8] . In addition, HFNC therapy decreased the intubation rate but did not impact mortality in patients with acute hypoxemic respiratory failure [9] . Another study showed that HFNC therapy can reduce 90-day mortality compared with noninvasive ventilation (NIV) [10] . Among cardiothoracic surgery patients, compared with NIV, HFNC therapy did not result in a worse rate of reintubation [11] . HFNC therapy also reduced the reintubation rate compared with COT after extubation in lower-risk respiratory failure patients [12] . A recent observational study showed that, in acute hypercapnic exacerbation of chronic obstructive pulmonary disease (AECOPD) patients, HFNC therapy was effective in improving the dyspnea score, gas exchange, and mucus production [13] . HFNC therapy is noninferior to NIV as initial ventilatory support for CO 2 clearance in mildto-moderate AECOPD patients [14] . Recently, HFNC therapy has been recommended for use in coronavirus disease 2019 (COVID-19) patients with acute respiratory failure (ARF) [15] . Most of the previous studies in this field focused on exploring how parameter settings affect physiological effects, such as actual inhaled FiO 2 [16, 17] , humidification [18] [19] [20] [21] , the PEEP effect [17, 22, 23] , and dead space flushing [23] [24] [25] [26] [27] , on cylinder or 3D-printed models and even human volunteers [28] . According to previous studies, flow plays a determinant role in the physiological effects of HFNC therapy. HFNC devices produced by different manufacturers may have varied performance, especially in terms of flow delivery. Few studies have compared different HFNC devices by the actual-flow and actual-FiO 2 at the cannulas. e degree of the difference among the manufacturers remains unknown, and therefore, it is meaningful to compare different devices. Furthermore, the results might provide clinicians with a deeper understanding of the difference among settings and devices to help select the most suitable equipment for the specific clinical condition. Four HFNC devices (AIRVO 2, Fisher & Paykel Healthcare, Auckland, New Zealand; TNI softFlow 50, TNI Medical AG, Würzburg, Germany; HUMID-BH, RESPIRACARE, Shenyang, China; OH-70C, Microme, Hunan, China) and a ventilator with HFNC module (bellavista 1000, Imtmedical, Buchs, Switzerland) were included and tested using their own breathing circuits, humidification chambers, nasal cannulas, and other accessories. Details are shown in Table 1 . e Medical Intensive Care Unit (MICU) of the West China Hospital provided the high-flow equipment. A typical HFNC system comprises a flow generator, active heated humidifier, single-limb heated circuit, and nasal cannula [29] . e international standard for HFNC equipment, particularly the requirements for basic safety and essential performance of ventilatory high-flow therapy equipment (ISO/DIS 80601-2-90), is still under development by the International Organization for Standardization. e flow-generating mechanism and structure of these devices in this study are summarized in Table 2 . A VT PLUS HF gas flow analyzer (Fluke Biomedical) was used to measure the actual-flow rate of the gas delivered by HFNC devices. A MaxO 2 oxygen analyzer (Maxtec) was used to measure FiO 2 . Beijing Aerospace Changfeng Co., Ltd. provided the instruments mentioned above. Figure 1 helps to measure the parameters directly from the interface. Two 6 × 4 PU tubes provide ports for nasal cannulas and standardize the different diameters of different interfaces. e 22 mm cap allows attachment to different measuring instruments. is structure does not change the direction of the flow. us, the measuring instruments can record the most real performance of these HFNC devices. For FiO 2 measurement, cannulas were attached to a conventional ventilation limb, and a sampling tube allowed a small flow of delivered gas to pass through the oxygen analyzer. As Figure 1 shows, tests were performed separately because sampling for FiO 2 measurement can produce a side flow that may affect the flow measurement. Tests were performed in a ward of the Medical Intensive Care Unit, West China Hospital. e environmental temperature was controlled during the test procedure. Any equipment that might influence the testing was removed from the room. All HFNC devices and measuring instruments were adequately preheated and calibrated before testing. All the testing processes were repeated three times for FiO 2 testing and two times for flow testing at different times to reduce disturbance from possible environmental changes and avoid contingency. HFNC devices were set at 31°C, and the MR850 heated humidifier was set in noninvasive mode. As excess water vapor would affect the accuracy of the VT PLUS gas flow analyzer and MaxO 2 oxygen analyzer, no water was added to the humidification chamber during FiO 2 and flow measurements. e flow was set at 20, 25, 30, 35, 40, 45, 50, 60, 70, and 80 L/min depending on their maximum flow, which was named set-flow. At each level of set-flow, the FiO 2 was set at 21%, 26%, 30%, 35%, 40%, 45%, 50%, 60%, 70%, 80%, and 90%, which was named set-FiO 2 . e FiO 2 and flow rate of the delivered gas from the cannulas were named actual-FiO 2 and actual-flow, respectively. For each setting combination, after stabilization for 1 min, the actual-FiO 2 and actual-flow were recorded 3 times at an interval of 10 seconds. en, one side of the cannulas was clipped to increase the resistance of the nasal cannulas. After another 1 min of stabilization, the clipped-flow was recorded in the same way to estimate the ability to provide sufficient flow under extreme situations. Normally distributed variables are expressed as the mean ± SD, and nonnormally distributed variables are expressed as the median (interquartile range). e Kruskal-Wallis H test was used to compare the effect of different set-FiO 2 on actual-flows and set-flows on actual-FiO 2 in a single device. We also used the Kruskal-Wallis H test to compare the actual parameters in different devices under the same settings. e Wilcoxon signed rank test was Note: a Air supply may be responsible for flow accuracy. b Oxygen supply would affect the FiO 2 accuracy. Flowmeters cannot provide enough oxygen flow when higher FiO 2 at high-flow levels is needed. c Changes in settings would alter the input air and oxygen flow or pressure, which will challenge the air-oxygen mixer and cause an inaccurate output. d e FiO 2 monitor influences the FiO 2 accuracy. used to compare differences between settings and actual parameters. e analysis mentioned above can help us to learn the output accuracy of each tested device. e Wilcoxon rank test was used to compare actual-flows with clipped-flows at the same settings to show the effect of increased resistance (clipping one side of the cannulas) on actual-flows in each device. All statistical tests were 2-sided, and P < 0.05 was considered statistically significant. All statistical analyses were performed using IBM SPSS statistical software version 23 for Windows. Single Device and Different Devices. As Figure 2 (a) shows, there were significant differences between the actual-flows and the corresponding set-flows in all five devices at most set-flow levels (P < 0.001). AIRVO 2 had the minimum difference between actual-flows and set-flows among the five devices ( Figure 2) (Table 3) . Device. e influence of different settings on actual-flows in different devices is demonstrated in Supplemental Table 1 and Figure 2 . An increase in set-FiO 2 caused a significant increase in actual-flow at all set-flow levels in TNI softFlow 50, OH-70C, HUMID-BH, and bellavista 1000 (P < 0.001) but not AIRVO 2. For AIRVO 2, when set-flows were fixed at 20 and 25 L/min (P � 0.99 and 0.5), there was no statistically significant change in actual-flow, while set-FiO 2 increased (Figure 2(b) ; Table 3 ). Another interesting finding was that changes in set-FiO 2 influenced the actual-flow more significantly at higher set-flows in AIRVO 2 and TNI softFlow 50 (Figures 2(b) and 2(c) ). ere were significant differences in actual-flow before and after one side of the cannulas was clipped at any set-flow levels in all devices (P < 0.001) but not TNI softFlow 50. e flow change rate was defined as the change in flow after one side of the cannula was clipped and calculated as the actual-flow minus the clipped-flow and then divided by the actual-flow. Notably, TNI softFlow 50 had the best ability to provide the desired flow (flow change rate 0 ± 0.28%) and could almost fully compensate for the flow at various levels of set-flow (20-50 L/min) ( Table 4) . When the resistance increased by clipping one side of the cannulas, all five devices could provide a stable flow (flow change rate smaller than 10%) at low set-flow levels but presented different compensatory abilities at higher set-flow levels ( Figure 3 ). When the set-flow was higher than 40 L/ min, the flow change rate increased significantly. e influence of different settings on actual-FiO 2 in different devices is demonstrated in Supplemental Table 2 . ere was a significant difference between the actual-FiO 2 and set-FiO 2 in all five devices at all set-flow levels (P < 0.05 for all), except for AIRVO 2 at 30% (P � 0.717) (Figure 4 ). In addition, the actual-FiO 2 values of different devices presented significant differences at all set-FiO 2 levels (P < 0.001). Figure 4(a) and Table 5 show that the actual-FiO 2 of AIRVO 2 were the closest to the set-FiO 2 , while the actual-FiO 2 of the other four devices were lower or higher. Actual-FiO 2 of AIRVO 2, TNI softFlow 50, and HUMID-BH showed a trend that, with the increase in set-FiO 2 and setflow, the difference between the actual-FiO 2 and set-FiO 2 became greater (Figures 4(b) , 4(c), and 4(e)). Another interesting point was that the actual-FiO 2 of these devices were less affected by set-flows at several specific set-FiO 2 levels ( Table 5) . is study provides new insights into the changes and quantitative analysis of the actual output of each device under different settings. Among the five devices evaluated, the variation in flow settings, FiO 2 settings, and devices can influence the actual-flow and actual-FiO 2 delivered. e following points were obtained: (1) ere were significant differences between the actual and set values of flow and FiO 2 . Chikata et al. tested Optiflow (Fisher & Paykel Healthcare) and found a similar trend: as the set-FiO 2 increased, the differences between the actual-FiO 2 and the set-FiO 2 increased [16] . When set-FiO 2 was fixed, changes in set-flow would also affect the actual-FiO 2 . e airoxygen mixer might be the reason behind the influence of different FiO 2 values on the actual-flow and vice versa and the deviation of the measured flow and FiO 2 with the set-flow and FiO 2 . Changes in settings would alter the input flow or the air and oxygen pressure, which will challenge the air-oxygen mixer and cause an inaccurate output. e variation in oxygen flow below a certain threshold (decided by the accuracy of the built-in flow sensor and algorithm) might not trigger the adjustment of the air flow, resulting in the deviation of the final flow output. (2) TNI softFlow 50, bellavista 1000, and HUMID-BH showed a better ability to provide the desired flow when the resistance abnormally increased. Although the adjustment process was a complex systematic work, the performance of the turbine was the main determining factor. (3) AIRVO 2 had the best performance in the accuracy of actual-flow and actual-FiO 2 . In addition to the airoxygen mixer factor discussed above, the difference in the FiO 2 titration target can also partly explain the different performances of FiO 2 accuracy in manual adjustment devices (AIRVO 2, TNI softFlow 50, and HUMID-BH). In AIRVO 2, the real-time FiO 2 and flow were measured by an ultrasonic oxygen analyzer, which formed a feedback regulation and achieved the best accuracy. If the real-time FiO 2 is calculated by the flow of oxygen and air, the concentration of oxygen supply, which is considered to be 100% but usually is not, may cause a deviation in FiO 2 output. e sensitivity and vulnerability of the oxygen analyzers might also account for the difference between the actual-FiO 2 and the set-FiO 2 . Dai et al. [30] found that there was no significant difference in actual-FiO 2 and actual-flow between AIRVO 2 and HUMID-BH under various test conditions. However, in Dai's study, FiO 2 in both devices was titrated according to the oxygen analyzer (OxiQuant B, Envitec Corporation) at the end of the heated circuit. (4) Bellavista 1000, a ventilator with a built-in high-flow module, performed well in flow accuracy and was secondary to AIRVO 2. From the structure point, ventilators have the same or even a more advanced high-flow generator and air/oxygen mixing equipment as standalone HFNC devices. Ventilators with HFNC modules could provide comparable flow and FiO 2 and be an alternative to standalone HFNC devices. Bench evaluation in clinical conditions provides an informative way to assess the actual performance of HFNC devices. e difference in measured flow and FiO 2 with set values might mislead clinicians to overestimate or underestimate the patient's condition. e clinical consequences of these differences cannot be ruled out and need to be considered. More bench and clinical studies are needed to determine and quantify the degree and consequences of these differences. Increasing the number of devices tested, improving the airway models in bench studies, or conducting clinical tests in healthy volunteers or representative patients are important. However, though clinical studies can make the testing of a wide range of flow and FiO 2 levels on Varied effect with no obvious pattern e data are presented as the median (IQR). e FiO 2 deviation equals actual-FiO 2 minus the corresponding set-FiO 2 . a When the set-flow rate was 40 L/min and set-FiO 2 was 26%, 35%, 40%, 60%, and 70% (P < 0.488, 0.857, 0.322, 0.37, and 0.372, respectively), actual-FiO 2 was not significantly different from the corresponding set-FiO 2 . b When set-FiO 2 was fixed at 26%, 30%, and 50%, actual-FiO 2 under different set-flows had no significant difference in TNI softFlow 50 (P � 0.700, 0.155, 0.166). c e control system of TNI softFlow 50 stopped the rising process, triggering the "oxygen flow is too high" alarm. e alarm cannot be eliminated until the oxygen flow is lowered. d When set-FiO 2 was 35%, most of actual-FiO 2 at varied set-flows were not significantly different compared with set-FiO 2 (P � 0.439, 0.833, 0.103, and 0.37 for set-flow � 20, 25, 30, 45, and 60 L/min, respectively). e When set-FiO 2 was 26%, most of actual-FiO 2 at varied set-flows were not significantly different compared with set-FiO 2 (P � 0.37, 0.37, 0.857, and 0.713 for set-flow � 20, 25, 50, 60, and 70 L/min, respectively). the same patient more applicable in the clinical setting, it would be ethically questionable to do so. Finally, FiO 2 monitoring and feedback regulation are recommended for future HFNC devices. is technique ensures the output accuracy and improves the reliability and safety of the treatment based on our results in AIRVO 2. ere are several limitations in this study. First, there were unavoidable individual differences in the devices we evaluated. ere were inevitable possible sample differences in the same model and manufacturer as well. Second, we did not use water for all devices, which was not representative of the clinical use of HFNC devices. e gas was heated but not humidified, and how the absence of humidification and water vapor affected the testing results remains unknown. Future studies are needed to investigate this influence. All the tests were performed at 31°C, and no water was added to achieve accurate flow measurements and minimal equipment aging. According to the ideal gas equation, the measured flow may be higher in actual scenes at high setting temperatures. ird, the environmental temperature and humidity in the ward might affect the results. e variation in flow, FiO 2 settings, and devices can influence the actual-flow and actual-FiO 2 delivered. e clinical consequences of the deviation cannot be ruled out and need to be considered. AIRVO 2 and OH-70C showed better FiO 2 accuracy. TNI softFlow 50, bellavista 1000, and HUMID-BH could lower the risk of insufficient flow support due to accidental compression, bending, or blocking of the nasal cannulas. Ventilators with HFNC modules could provide comparable flow and FiO 2 and be an alternative to standalone HFNC devices. e data used to support the findings of this study are included within the supplementary information files. e authors declare that they have no conflicts of interest. High-flow nasal cannula oxygen therapy in adults: physiological benefits, indication, clinical benefits, and adverse effects Heated humidified high-flow nasal oxygen in adults Physiologic effects of high-flow nasal cannula oxygen in critical Care subjects Nasal high flow reduces dead space Physiologic effects of high-flow nasal cannula in acute hypoxemic respiratory failure Current evidence for the effectiveness of heated and humidified high flow nasal cannula supportive therapy in adult patients with respiratory failure High-flow oxygen therapy in acute respiratory failure Nasal high-flow preoxygenation for endotracheal intubation in the critically ill patient: a randomized clinical trial High flow nasal cannula compared with conventional oxygen therapy for acute hypoxemic respiratory failure: a systematic review and meta-analysis High-flow oxygen through nasal cannula in acute hypoxemic respiratory failure High-flow nasal oxygen vs noninvasive positive airway pressure in hypoxemic patients after cardiothoracic surgery Effect of postextubation high-flow nasal cannula vs conventional oxygen therapy on reintubation in low-risk patients High flow nasal therapy use in patients with acute exacerbation of copd and bronchiectasis: a feasibility study High flow nasal therapy versus noninvasive ventilation as initial ventilatory strategy in COPD exacerbation: a multicenter non-inferiority randomized trial Management of critically ill adults with COVID-19 F IO 2 in an adult model simulating high-flow nasal cannula therapy Factors affecting FiO 2 and PEEP during high-flow nasal cannula oxygen therapy: a bench study Humidity and inspired oxygen concentration during high-flow nasal cannula therapy in neonatal and infant lung models Humidification in very-high-flow nasal-cannula therapy in an adult lung model Humidification performance of two high-flow nasal cannula devices: a bench study Inspiratory tube condensation during highflow nasal cannula therapy: a bench study Mechanistic understanding of high flow nasal cannula therapy and pressure support with an in vitro infant model Positive end-expiratory pressure effect of 3 high-flow nasal cannula devices Correlation of high flow nasal cannula outlet area with gas clearance and pressure in adult upper airway replicas Nasal high flow clears anatomical dead space in upper airway models A high-flow nasal cannula system with relatively low flow effectively washes out CO2 from the anatomical dead space in a sophisticated respiratory model made by a 3D printer PEEP generated by high-flow nasal cannula in a pediatric model Impact of flow and temperature on patient comfort during respiratory support by high-flow nasal cannula High-flow nasal cannula oxygen therapy devices Influencing factors of actual-FiO2 difference during highflow nasal cannula therapy based on the active sevo respiratory simulation lung Acknowledgments e authors would like to thank Dr. Li Ling (Chinese Evidence-Based Medicine Center, West China Hospital, Sichuan University) for valuable advice regarding methodology. e instruments used for the measurement of all the devices, including a VT PLUS HF gas flow analyzer (Fluke Biomedical) and MaxO 2 oxygen analyzer (Maxtec), were provided by Beijing Aerospace Changfeng Co., Ltd. Supplemental Table 1 : effects of set-flow, set-FiO 2 , and devices on actual-flow. Supplemental Table 2 : effects of setflow, set-FiO 2 , and devices on actual-FiO 2 . (Supplementary Materials) Yuyan Zhou and Zhong Ni contributed equally to this study.