key: cord-1020387-1xyjrjpf authors: Kim, Yong Kwan; Lim, Seong-In; Choi, Sarah; Cho, In-Soo; Park, Eun-Hye; An, Dong-Jun title: A novel assay for detecting canine parvovirus using a quartz crystal microbalance biosensor date: 2015-03-23 journal: J Virol Methods DOI: 10.1016/j.jviromet.2015.03.015 sha: b4028cf2aed4166e54dd30db86f37eddb14caa3a doc_id: 1020387 cord_uid: 1xyjrjpf Rapid and accurate diagnosis is crucial to reduce both the shedding and clinical signs of canine parvovirus (CPV). The quartz crystal microbalance (QCM) is a new tool for measuring frequency changes associated with antigen–antibody interactions. In this study, the QCM biosensor and ProLinker™ B were used to rapidly diagnosis CPV infection. ProLinker™ B enables antibodies to be attached to a gold-coated quartz surface in a regular pattern and in the correct orientation for antigen binding. Receiver operating characteristics (ROC) curves were used to set a cut-off value using reference CPVs (two groups: one CPV-positive and one CPV-negative). The ROC curves overlapped and the point of intersection was used as the cut-off value. A QCM biosensor with a cut-off value of −205 Hz showed 95.4% (104/109) sensitivity and 98.0% (149/152) specificity when used to test 261 field fecal samples compared to PCR. In conclusion, the QCM biosensor described herein is eminently suitable for the rapid diagnosis of CPV infection with high sensitivity and specificity. Therefore, it is a promising analytical tool that will be useful for clinical diagnosis, which requires rapid and reliable analyses. Canine parvovirus (CPV), a member of the genus Parvovirus within the family Parvoviridae, first emerged as a canine pathogen in the late 1970s (Appel et al., 1978; Kelly, 1978) . A few years after the first outbreak, CPV-2 was completely replaced by two antigenic variants, termed CPV-2a and CPV-2b (Parrish et al., 1988) . In 2000, a novel mutant of CPV, harboring an amino acid substitution at position 426 (CPV-2c), emerged in Italy and rapidly spread to several countries (Buonavoglia et al., 2001; Decaro and Buonavoglia, 2012) . Infection by canine CPV results in a highly contagious enteric disease with a high rate of morbidity and mortality (Larson and Schultz, 2007; Parrish, 1990 Parrish, , 1999 . CPV is a small, non-enveloped single-stranded DNA (ssDNA) virus with a genome of approximately 5.2 kb that encodes two structural (VP1 and VP2) and two non-structural (NS1 and NS2) proteins (Cotmore and Tattersall, 1987) . VP2 is the major capsid protein, and plays an important role as an antigenic determinant (Chang et al., 1992; Horiuchi et al., 1994) . * Corresponding author. Tel.: +82 31 467 1782; fax: +82 31 467 1800. Currently, a diagnosis of CPV is based on clinical signs, which include vomiting and diarrhea; however, a definitive diagnosis is difficult because these symptoms are common to other enteric diseases (Elia et al., 2007; Hirasawa et al., 1994) . Conventional detection methods such as electron microscopy (Teramoto et al., 1984) , virus isolation (Mochizuki et al., 1993) , latex agglutination (Veijalainen et al., 1986) , hemagglutination (Mochizuki et al., 1993; Teramoto et al., 1984) , and enzyme-linked immunosorbent assay (Drane et al., 1994; Teramoto et al., 1984) have been developed for the detection of CPV, and all are effective and accurate. Molecular detection techniques, including PCR (Mochizuki et al., 1993) , real-time PCR (Decaro et al., 2005) , nested PCR (Hirasawa et al., 1994) , and loop-mediated isothermal amplification (Cho et al., 2006; Mukhopadhyay et al., 2012) have also been used to diagnose CPV, although the sensitivity and specificity of these techniques are variable; however, a major drawback is that these methods are both laborious and time-consuming. The key to preventing the spread of CPV is early and rapid diagnosis. Infected dogs can then be isolated and given supportive treatment to reduce both morbidity and mortality. Therefore, a rapid, on-site diagnostic method is required for the early detection of CPV. The quartz crystal microbalance (QCM) is a nanogram-sensitive technique that utilizes acoustic waves generated by oscillating a piezoelectric, single crystal quartz plate to measure mass (Dixon, 2008). The piezoelectric effect is due to the electrical charge that accumulates in certain solid materials, such as crystals, in response to an applied mechanical stress. The QCM biosensor measures mass per unit area by measuring the change in the frequency of a quartz crystal resonator, and is a convenient method for detecting antigen-antibody interactions. The QCM biosensor is a "label-free technology" because binding is detected directly without the need for labeled reagents. Many studies have used the QCM biosensor to detect pathogenic microorganisms such as Vibrio cholerae, Bacillus anthracis, malaria, Salmonella Typhimurium, Leishmania infantum, and canine influenza virus (Cabral-Miranda et al., 2014; Carter et al., 1995; Cooper et al., 2001; Hao et al., 2009; Ittarat et al., 2013; Kim et al., 2014; Salam et al., 2013) . The aim of the present study was to use the QCM biosensor in conjunction with ProLinker TM B (used to immobilize CPV-specific antibodies onto a gold-coated quartz surface) to develop a suitable method for the rapid and accurate diagnosis of CPV. The change in mass at the quartz crystal surface induced by antigenantibody interactions generates a measurable change in frequency. The QCM biosensor described herein constitutes a rapid and sensitive method for diagnosing CPV infection in dogs. The 27 CPV-positive references (virus isolation) were obtained from the Korea veterinary culture collection (KVCC) (http://kvcc.kahis.go.kr) (Supplemental Table 1 ). There were also 40 CPV-negative references, which comprised of 36 CPV-negative fecal samples, two bacterial (Escherichia coli and salmonella) and two viral (canine coronavirus and canine rotavirus) isolates. All of the CPV-positive and CPV-negative reference samples were confirmed as positive and negative by PCR, respectively. The field samples comprised 261 fecal samples taken from dogs with suspected CPV which were being treated at animal hospitals in Seoul and Gyeonggi province in South Korea from 2012 to 2013. DNA was extracted using a Qiagen DNA extraction kit (Hilden, Germany), according to manufacturer's instructions. All samples were identified as CPV-positive or -negative by PCR using the following CPV-specific primers: forward, 5 -CAAATAGAGCATTGGGCTTACC-3 and reverse, 5 -CAATCTCCTTCTGGATATCTTC-3 . The PCR reaction mixture comprised 11 L of DNase-free water, 5 L of 5× PCR buffer, 1 L of each primer, 1 L of dNTP mix, 1 L of PCR enzyme mix, and 5 L of extracted DNA as a template (total volume 25 L). The PCR conditions were as follows: 94 • C for 5 min, followed by 35 cycles of 1 min at 94 • C, 1 min at 55 • C, and 1 min at 72 • C, with a final extension at 72 • C for 10 min. Supplementary Table 1 related to this article can be found, in the online version, at http://dx.doi.org/10.1016/j.jviromet. 2015.03.015. A commercial immunochromatography Ag kit was purchased from the Bionote Inc (Hwaseong, South Korea). The test device that comes in the kit shows the letters "T" (test line) and "C" (control line) on its surface. The control line should always be visible if the test has been performed properly. A purple "test line" is visible if sufficient CPV antigen is detected. Briefly, fecal samples were placed in sample collection tubes containing assay diluent and mixed vigorously. The tube was then left to stand for 1 min at room temperature to allow any particles to settle. Finally, four drops of sample solution were loaded into the sample hole using a disposable dropper and the results read after 10 min. An endpoint dilution assay was used to measure virus titer. Serial dilutions of KVCC-VR0900161 (Supplemental Table 1 ) were prepared and inoculated onto replicate Crandall Ress feline kidney (CRFK) cell cultures in 96-well plastic plates. The number of infected cell cultures at each virus dilution was then determined by looking for a cytopathic effect (CPE). After an incubation period for 4 days, wells in plates that displayed a CPE were scored as positive. The TCID 50 was calculated using the Reed & Muench endpoint method. The quartz was soaked in a solution containing MeOH and 35% HCl (1:1) for 30 min. A freshly prepared piranha solution (a 3:1 mixture of concentrated H 2 SO 4 and 30% H 2 O 2 ) was then added for 10 min. The quartz was then thoroughly washed with distilled water and dried under a stream of nitrogen (N 2 ) gas. The precleaned quartz was then gold-coated by thermal evaporation in a Magnetron Sputtering System (Cliotech, Seoul, South Korea) before immersion in a 3 mM ProLinker TM B solution (Proteogen, Chuncheon, South Korea), which contains exposed SH groups, for 1 h. The quartz was then rinsed sequentially with CHCl 3 , EtOH, and distilled water before drying under a stream of N 2 gas. A monoclonal anti-CPV IgG antibody was purchased from Median Diagnostics Inc (Chuncheon, South Korea). The monoclonal anti-CPV IgG antibody (50 g/mL) was mixed with PBS (1 mL containing 30% glycine) and 50 L of the mixture added to the gold-coated quartz and incubated overnight at 4 • C. The quartz was then rinsed twice with PBS and deionized water, and dried under a stream of N 2 gas. The antibody-coated quartz was stored at 4 • C until use. A X-delic XQ10 instrument (UBTgen, Seoul, South Korea) was used to measure frequency changes in the gold-plated quartz. Briefly, the monoclonal CPV IgG antibody-coated quartz was placed horizontally and loaded with 100 L of PBS. When the frequency reached equilibrium, 100 L of fecal sample was loaded onto the quartz surface. Real-time changes in resonant frequency were recorded until the signal again reached equilibrium. Each experiment was performed at room temperature and repeated three times. All of the CPV-positive (n = 27) and -negative reference samples (n = 40) were classified according to their measured frequency changes (Supplemental Table 2 ) and receiver operating characteristics (ROC) curves constructed (Fig. 1) . Reference samples were divided into two groups, namely, CPV-positive and CPV-negative. Although both groups showed clear separation, there was a point of overlap and this point of intersection was used as the cut-off value. The cut-off value was the threshold value that determined whether a sample was identified correctly as positive or negative. According to the ROC, the cut-off value was set at a frequency change of −205 Hz. The diagnostic sensitivity and specificity were calculated as follows: sensitivity = true positive/(true positive + false negative); specificity = true negative/(true negative + false positive). As A CPV-positive reference sample registered with the KVCC (KVCC-VR0900161) was used to set the detection limit. The TCID 50 /mL for this sample, as measured by a commercial immunochromatography Ag kit (CIAK) and PCR, was 10 3 and 10 2 , respectively (Table 1) . Using the test sample in the QCM biosensor revealed that frequency changes increased gradually as the concentration of the test sample increased (Fig. 2) . Frequency changes exceeding the cut-off value (−205 Hz) were generated at a minimum TCID 50 /mL value of 10 2 . If the QCM biosensor is to be of practical use in the field, the antibodies immobilized on the surface must retain long-term activity. Therefore, we stored the CPV IgG antibody coated quartz at 4 • C for 1, 6, 12, 15 and 18 months. The diagnostic accuracy at each time point was then tested using KVCC-VR0900161 (Supplemental Table 1 ) at 10 3 TCID 50 /mL. As shown in Supplemental Table 3 , the coefficient of variation (CV) at 1, 6, 12, 15 and 18 months increased gradually (1.8%, 3.0%, 4.7%, 7.5% and 12.4%) in a time-dependent manner. The frequency change and CV at each time point are shown in Fig. 3 . A CV < 5% was maintained for 12 months and thus quartzimmobilized CPV IgG antibody can be stored at 4 • C for up to 1 year, making them eminently suitable for field use. Supplementary Table 3 related to this article can be found, in the online version, at http://dx.doi.org/10.1016/j.jviromet. 2015.03.015. All suspected CPV-positive fecal samples were confirmed as CPV-positive or -negative by PCR before application to the QCM biosensor. Of the 261 field samples tested, 41.8% (109/261) were CPV-positive and 58.2% (152/261) were CPV-negative; of the 109 CPV-positive samples confirmed by PCR, CPV-2a, 2b and 2c comprised to 95.4% (104/109), 4.6% (5/109) and 0% (0/109), respectively. The QCM biosensor identified 104/109 as positive (95.4% sensitivity) and 149/152 as negative (98.0% specificity). However, the sensitivity and specificity of the CIAK were 80.7% (88/109) and 89.5% (136/152), respectively. The gold-coated quartz surface is a relatively new class of device to which biological recognition elements can be attached. When measuring the frequency change that caused the antigen-antibody interaction, it is important that the frequency is highly stabile and that background noise is kept to a minimum; this will ensure maximum sensitivity. Several factors may cause frequency drift, including the electrode surface and the viscosity of the liquid (Buttry and Ward, 1992; Tsionsky et al., 1995) . Previous studies used silver and gold electrodes as QCM biosensors (Kim et al., 2014; Su et al., 2000) ; however, gold electrodes produce more consistent results because the surface is less rough than that of silver (Su et al., 2000) . The roughness of the electrode surface is critical if frequency drift is to be minimized. Frequency drift is also associated with temperature (Rocklein and George, 2003; Martin et al., 1991; Wang et al., 2005) . If the temperature fluctuates, both real mass changes over time and apparent mass changes must be measured concurrently; thus temperature-dependent apparent mass changes must be characterized and minimized to allow the QCM biosensor to provide an accurate readout of the frequency change (Rocklein and George, 2003) . Over the last decade, the use of the QCM has been extended to the liquid phase (Bruckenstein et al., 1994) . The resonance frequency of the crystal responds to the properties of a contacting liquid. Inconsistences in liquid viscosity are easily overcome by using PBS or water; thus the electrode surface, the temperature, and the viscosity of the liquid must all be taken into account if frequency drift is to be minimized. If a QCM biosensor is to have both high sensitivity and specificity, important issues such as antibody orientation on the sensor surface must be resolved. There are two main approaches to preparing the surface on an antibody-based sensor: random antibody immobilization and site-directed antibody immobilization. The simplest technique is to allow random adsorption of the antibody onto the quartz surface. Although simple, this method has major drawbacks, including low stability and lower activity due to random orientation (Buijs et al., 1996; Makaraviciute and Ramanaviciene, 2013) . Direct antibody immobilization via a self-assembled monolayer has multiple advantages, namely, that the technique is simple and the covalent bonding means that the surface is both stable and reusable; however, a major drawback is the relatively low sensitivity due to random antibody orientation and the low availability of antibody binding sites (Kausaite-Minkstimiene et al., 2010; Tsai and Pai, 2009 ). Therefore, site-directed antibody immobilization methods have been developed to address these short-comings (Della Ventura et al., 2011; Hussack et al., 2009; Kwon et al., 2004; Um et al., 2011; Yang et al., 2013; Yoon et al., 2011) . A previous study showed that ProLinker TM B, a calixcrown-5 derivative, is able to orient antibodies in a site-directed manner, leading to a more consistent position on the surface and a significant increase in sensitivity for the target antigen (Lee et al., 2003) . As mentioned above, gold electrodes are the preferred carrier for use in QCM biosensors because they generate a stable frequency (Steegborn and Skládal, 1997) ; however, the major drawback is the high unit price of gold and thus gold-coated electrodes must be reusable to minimize the cost of each experiment. The antigen-antibody interaction is usually irreversible; however, the interaction can be disrupted by chaotropic reagents, organic solvents, or even ultrasonic radiation (Marazuela and Moreno-Bondi, 2002) . ProLinker TM B can easily be detached from quartz surfaces using piranha solution (Lee et al., 2003) . The gold-coated quartz crystal described herein is reusable, which significantly reduces the cost of the QCM biosensors used to detect CPV. In conclusion, the QCM biosensor described herein represents a promising analytical tool for use in clinical diagnosis, where rapid and reliable analyses are needed. Therefore, the QCM biosensor is a potential sensitive and specific assay for the rapid and early diagnosis of CPV infection. The authors have no conflicts of interest to declare. Status report: canine viral enteritis Dual quartz crystal microbalance oscillator circuit. Minimizing effects due to liquid viscosity, density, and temperature Changes in the secondary structure of adsorbed IgG and F(ab )2 studied by FTIR spectroscopy Evidence for evolution of canine parvovirus type 2 in Italy Measurement of interfacial processes at electrode surfaces with the electrochemical quartz crystal microbalance Detection of parasite antigens in Leishmania infantum-infected spleen tissue by monoclonal antibody-, piezoelectric-based immunosensors Quartz crystal microbalance detection of Vibrio cholerae O139 serotype Direct and sensitive detection of a human virus by rupture event scanning The autonomously replicating parvoviruses of vertebrates Multiple amino acids in the capsid structure of canine parvovirus coordinately determine the canine host range and specific antigenic and hemagglutination properties Detection of canine parvovirus in fecal samples using loop-mediated isothermal amplification Real-time PCR assay for rapid detection and quantitation of canine parvovirus type 2 in the feces of dogs Canine parvovirus-a review of epidemiological and diagnostic aspects, with emphasis on type 2c Light assisted antibody immobilization for bio-sensing Quartz crystal microbalance with dissipation monitoring: enabling real-time characterization of biological materials and their interactions Evaluation of a novel diagnostic test for canine parvovirus Detection of infectious canine parvovirus type 2 by mRNA real-time RT-PCR Rapid detection of Bacillus anthracis using monoclonal antibody functionalized QCM sensor Sensitive detection of canine parvovirus DNA by the nested polymerase chain reaction Mapping of determinants of the host range for canine cells in the genome of canine parvovirus using canine parvovirus/mink enteritis virus chimeric viruses Multivalent anchoring and oriented display of single-domain antibodies on cellulose Biosensor as a molecular malaria differential diagnosis Comparative study of random and oriented antibody immobilization techniques on the binding capacity of immunosensor An enteric disease of dogs resembling feline panleukopenia Detection of H3N2 canine influenza virus using a quartz crystal microbalance Antibody arrays prepared by cutinase-mediated immobilization on self-assembled monolayers Three-year serologic immunity against canine parvovirus type 2 and canine adenovirus type 2 in dogs vaccinated with a canine combination vaccine Pro-teoChip: a highly sensitive protein microarray prepared by a novel method of protein immobilization for application of protein-protein interaction studies Characterization of a quartz crystal microbalance with simultaneous mass and liquid loading Site-directed antibody immobilization techniques for immunosensors Fiber-optic biosensors-an overview Comparison of polymerase chain reaction with virus isolation and haemagglutination assays for the detection of canine parvoviruses in faecal specimens Development and evaluation of loop-mediated isothermal amplification assay for rapid and sensitive detection of canine parvovirus DNA directly in faecal specimens Canine host range and a specific epitope map along with variant sequences in the capsid protein gene of canine parvovirus and related feline, mink, and raccoon parvoviruses Emergence, natural history, and variation of canine, mink, and parvoviruses Host range relationships and the evolution of canine parvovirus Temperature-induced apparent mass changes observed during quartz crystal microbalance measurements of atomic layer deposition Real-time and sensitive detection of Salmonella Typhimurium using an automated quartz crystal microbalance (QCM) instrument with nanoparticles amplification Construction and characterization of the direct piezoelectric immunosensor for atrazine operating in solution Disposable, low cost, silver-coated, piezoelectric quartz crystal biosensor and electrode protection Comparison of enzyme-linked immunosorbent assay, DNA hybridization, hemagglutination, and electron microscopy for detection of canine parvovirus infections Surface plasmon resonance-based immunosensor with oriented immobilized antibody fragments on a mixed self-assembled monolayer for the determination of staphylococcal enterotoxin B Behavior of quartz crystal microbalance in nonadsorbed gases at high pressures Electrochemically oriented immobilization of antibody on poly-(2-cyanoethylpyrrole)-coated gold electrode using a cyclic voltammetry Latex agglutination test for detecting feline panleukopenia virus, canine parvovirus, and parvoviruses of fur animals Quartz crystal microbalance in elevated temperature viscous liquids: temperature effect compensation and lubricant degradation monitoring Immobilization of unraveled immunoglobulin G using well-oriented ZZ-His protein on functionalized microtiter plate for sensitive immunoassay Immobilization of antibodies on the selfassembled monolayer by antigen-binding site protection and immobilization kinetic control We are grateful for the technical assistance provided by Proteogen (Chuncheon, South Korea). This study was supported by a grant (project code number: B-AD20-2011-13-01) from the Animal and Plant Quarantine Agency (QIA), Ministry of Agriculture, Food and Rural Affairs, Republic of Korea.