key: cord-1020385-k7tgtcg8 authors: Radwan, Ahmed Bahgat; Paramparambath, Sreedevi; Cabibihan, John-John; Al-Ali, Abdulaziz Khalid; Kasak, Peter; Shakoor, Rana A.; Malik, Rayaz A.; Mansour, Said A.; Sadasivuni, Kishor Kumar title: Superior Non-Invasive Glucose Sensor Using Bimetallic CuNi Nanospecies Coated Mesoporous Carbon date: 2021-11-18 journal: Biosensors (Basel) DOI: 10.3390/bios11110463 sha: 0afca771c0d06b6b47be07bb1592f19a154767b7 doc_id: 1020385 cord_uid: k7tgtcg8 The assessment of blood glucose levels is necessary for the diagnosis and management of diabetes. The accurate quantification of serum or plasma glucose relies on enzymatic and nonenzymatic methods utilizing electrochemical biosensors. Current research efforts are focused on enhancing the non-invasive detection of glucose in sweat with accuracy, high sensitivity, and stability. In this work, nanostructured mesoporous carbon coupled with glucose oxidase (GO(x)) increased the direct electron transfer to the electrode surface. A mixed alloy of CuNi nanoparticle-coated mesoporous carbon (CuNi-MC) was synthesized using a hydrothermal process followed by annealing at 700 °C under the flow of argon gas. The prepared catalyst’s crystal structure and morphology were explored using X-ray diffraction and high-resolution transmission electron microscopy. The electrocatalytic activity of the as-prepared catalyst was investigated using cyclic voltammetry (CV) and amperometry. The findings show an excellent response time of 4 s and linear range detection from 0.005 to 0.45 mM with a high electrode sensitivity of 11.7 ± 0.061 mA mM cm(−2) in a selective medium. The accurate quantification of glucose is key to the diagnosis and management of patients with diabetes. In 2019, the International Diabetes Federation estimated that diabetes affects 463 million people, constituting 9.3% of the adult population aged 20-79, but with 1 in 2 remaining undiagnosed. Chronic hyperglycemia is a major risk parameter for long-term microvascular and macrovascular complications leading to increased morbidity and mortality, with 11.3% of global deaths being attributed to diabetes [1, 2] . Additionally, in patients with COVID-19, whilst long-term glycemic control has not been associated with adverse outcomes, the severity of hyperglycemia at admission has been allied to the prerequisite for mechanical ventilation, admission to the intensive care unit, and mortality, independent of age, diabetes, and hypertension [3, 4] . There is increasing use of bioelectrode interfaces with wearable bioelectronics to enable the continuous non-invasive transfer of physiological information [5] [6] [7] [8] [9] [10] . In addition, the detection of biochemical molecules to indicate physiology and metabolism can be achieved using wearable electrochemical biosensors [11] [12] [13] . These electrochemical bio-sensors are generally established on enzyme-electrochemical reactions with high specificity [14] . The key step for wearable electrochemical biosensors is the enzyme-electrode interface where The bimetallic Cu-Ni nanoparticles coated mesoporous carbon was obtained using the same method [11] . Briefly, Cu(CH 3 COO) 2 ·3H 2 O and Ni(CH 3 COO) 2 ·6H 2 O were placed in 20 mL N, N-dimethylformamide (DMF). The mesoporous carbon was added to the solution and left for 24 h until a homogenous mixture was obtained. Oleic acid and oleylamine were added to the solution and heated to 180 • C for 8 h using a 75 mL sealed Teflon-lined stainless autoclave. The acquired product was centrifuged at 10,000 rpm for 30 min and washed several times using ethanol and deionized water. Subsequently, the dried powder is heated up to 700 • C for 2 h at a heating rate of 5 • C per minute under an argon gas flow to obtain the final product of the mixed Cu-Ni nanoalloy decorated the mesoporous carbon (CuNi-MC). The catalyst ink was synthesized by adding 2 mL of ethanol to a mixture of 10 mg Ni/Cu-mesoporous composite, 2 mg GOx, and 40 µL nafion. The mixture was kept for sonication in an ice bath for 1 h. After that, 10 µL of the suspension was dropped on the GC electrode's active surface, which was left to dry at room temperature for 1 h to obtain a uniform thin coating on the surface. The artificial sweat solution was synthesized using 5.5 mM lactic acid, 22 mM urea, 10 mM KCl, and 25 mM uric acid [13] . The TEM analysis of the as-prepared catalyst was obtained by dispersing catalyst in IPA and a drop was left to dry on molybdenum grid to avoid interference with sample elements. A FEI-Thermo scientific-TALOS A FEG transmission electron microscope with a 200 kV beam voltage was used. Elemental analysis and mapping was performed using a FEI-Thermo scientific-SuperX EDS detector. The XRD analysis was carried using Rigaku, Miniflex2 Desktop, Tokyo, Japan, equipped with Cu K α radiations. The and phase study structural of the as-synthesized catalyst was carried out at a scanning step of 0.02 • in the 2θ range from 10 • to 90 • . Gamry-potentiostat/galvanostat (Ref. 600) was employed to perform the cyclic voltammetry (CV) and chronoamperometry experiments using a three-electrode cell configuration consisting of Ag/AgCl electrode as a reference electrode, the GC as the working electrode, and platinum wire as a counter electrode. The cyclic voltammetry was conducted at a potential range from −0.1 V to 0.8 V vs. Ag/AgCl at 25 mV/s, and the chronoamperometry was done at an applied potential of 0.6 V. Figure 1a shows the TEM image of the used mesoporous carbon synthesizing the CuNi-MC catalyst with less than 100 nm particle size. The XRD pattern of CuNi-MC is presented in Figure 1b . The average crystal size was estimated from the XRD patterns using the Debye-Scherer equation. where D is the crystallite size, λ is the wavelength of Cu-k α1 = 1.54060 Å, β can be estimated from the experimental peak width (FWHM) of the most intense peak (002). The average crystal size of the bimetallic Ni-Cu-nanoparticles was calculated in the range of 10 ± 2 nm. The amorphous peak positioned at 24.8 • was attributed to the mesoporous carbon. However, the peaks located at 43.9 • , 51. Figure 1a shows the TEM image of the used mesoporous carbon synthesizing the CuNi-MC catalyst with less than 100 nm particle size. The XRD pattern of CuNi-MC is presented in Figure 1b . The average crystal size was estimated from the XRD patterns using the Debye-Scherer equation. where D is the crystallite size, λ is the wavelength of Cu-kα1 = 1.54060 Å , β can be estimated from the experimental peak width (FWHM) of the most intense peak (002). The average crystal size of the bimetallic Ni-Cu-nanoparticles was calculated in the range of 10 ± 2 nm. The amorphous peak positioned at 24.8° was attributed to the mesoporous carbon. However, the peaks located at 43 The glucose sensitivity of the as-prepared catalyst was evaluated at variable pH of 7 and 12. Figure 3a shows the CV of the glucose sensor at pH 7, exhibiting no oxidationreduction peaks. Similar studies have been carried out exploring the effect of the pH on the electrode sensitivity, revealing that the catalyst sensitivity to glucose is boosted by increasing the pH [45] [46] [47] . The reduced species transferred the excess electrons to the Ni 2+ /Ni 3+ redox couple leading to an enhancement in the redox current. It is worth mentioning that ion contamination limitation can be eliminated employing alkaline conditions on the electrode surface, as the hydroxyl groups (OH ) suppress the chloride (Cl ) adsorption to the electrode surface [45, 47] . Accordingly, changing the real pH of the sweat solution in this work significantly enhanced the sensitivity of the as-prepared electrode. Figure 3b ,c exhibited the CV of the as-synthesized catalyst in the absence and presence of glucose oxidase, showing a distinct variation in their response to glucose. This was ascribed to the glucose oxidation by GOx enzyme-generating H 2 O 2 , which was reduced to H 2 O + O 2 at CuNi-MC surface and consequently increased the electrocatalytic currents. The glucose sensitivity of the as-prepared catalyst was evaluated at variable pH of 7 and 12. Figure 3a shows the CV of the glucose sensor at pH 7, exhibiting no oxidationreduction peaks. Similar studies have been carried out exploring the effect of the pH on tion to the electrode surface [45, 47] . Accordingly, changing the real pH of the sweat solution in this work significantly enhanced the sensitivity of the as-prepared electrode. Figure 3b ,c exhibited the CV of the as-synthesized catalyst in the absence and presence of glucose oxidase, showing a distinct variation in their response to glucose. This was ascribed to the glucose oxidation by GOx enzyme-generating H2O2, which was reduced to H2O + O2 at CuNi-MC surface and consequently increased the electrocatalytic currents. Figure 4 shows the CV of the electrode modified with CuNi-MC in sweat solution at different concentrations of glucose. The redox peak potential of CuNi-MC could be linked to the redox couple of Ni 2+ to Ni 3+ and Cu 2+ to Cu 3+ and then back to Ni 2+ and Cu 2+ . This finding indicates that the direct electrochemical reaction of GOx occurs on the CuNi-MC/GOx electrode surface, with GOx maintaining its electroactivity on the CuNi-MC/GOx surface. The current density increased by adding a higher amount of glucose, which could be accredited to the existence of many active sites due to the porous structure of the mesoporous carbon resulting in a high electron transfer rate. A Ni(OH)2 (Equation (5)) film is first constructed on the bimetallic CuNi surface, leading to Cu enrichment on the metal surface followed by oxidation to CuO and Cu(OH)2 (Equations (2) and (3)). Consequently, the film surface is composed of Cu(OH)2 and NiOOH, providing high electrical conductivity for the reaction [48] . Moreover, some CuO and Cu(OH)2 can be further oxidized to Cu 3+ oxide [49] . Mathew et al. [50] demonstrated that an electroactive species of Ni(OH)2 could be reconstructed with a predominant α-structure during the reaction of γ-NiO(OH) Figure 4 shows the CV of the electrode modified with CuNi-MC in sweat solution at different concentrations of glucose. The redox peak potential of CuNi-MC could be linked to the redox couple of Ni 2+ to Ni 3+ and Cu 2+ to Cu 3+ and then back to Ni 2+ and Cu 2+ . This finding indicates that the direct electrochemical reaction of GO x occurs on the CuNi-MC/GO x electrode surface, with GO x maintaining its electroactivity on the CuNi-MC/GO x surface. The current density increased by adding a higher amount of glucose, which could be accredited to the existence of many active sites due to the porous structure of the mesoporous carbon resulting in a high electron transfer rate. A Ni(OH) 2 (Equation (5)) film is first constructed on the bimetallic CuNi surface, leading to Cu enrichment on the metal surface followed by oxidation to CuO and Cu(OH) 2 (Equations (2) and (3)). Consequently, the film surface is composed of Cu(OH) 2 and NiOOH, providing high electrical conductivity for the reaction [48] . Moreover, some CuO and Cu(OH) 2 can be further oxidized to Cu 3+ oxide [49] . Mathew et al. [50] demonstrated that an electroactive species of Ni(OH) 2 could be reconstructed with a predominant α-structure during the reaction of γ-NiO(OH) with glucose while testing their enzymatic nickel hydroxide sensor in an alkaline medium. The generated α-Ni(OH) 2 will be oxidized to γ-NiO(OH) during the redox reaction, boosting the anodic current. Consequently, concentrated Ni 3+ will be formed on the surface and reduced to Ni 2+ during the glucose oxidation reaction, increasing the cathodic current. It is worth mentioning that few of γ-NiO(OH) species could be converted to β-Ni(OH) 2 , generating NiH or NiO as an irreversible reaction since the test was carried out in an alkaline medium where α-Ni(OH) 2 was not stable [50] . Nonetheless, the presence of higher content of Ni 2+ and Ni 3+ ions on the surface, smaller conversions could slightly influence the redox reaction and current [50] . Nevertheless, the bimetallic CuNi nanoparticles can efficiently restrain the unfavorable γ-NiOOH species construction, stabilizing β-NiOOH forms at pH 12 [51] . The presence of β-NiOOH species is considered to be an improved electroactive phase for superior electrochemical performance in an alkaline medium [52] . The presence of the CuNi bimetallic nanospecies facilitates the e transfer between the GC surface and the GO x , improving glucose detection. is worth mentioning that few of γ-NiO(OH) species could be converted to β-Ni(OH)2, generating NiH or NiO as an irreversible reaction since the test was carried out in an alkaline medium where α-Ni(OH)2 was not stable [50] . Nonetheless, the presence of higher content of Ni 2+ and Ni 3+ ions on the surface, smaller conversions could slightly influence the redox reaction and current [50] . Nevertheless, the bimetallic CuNi nanoparticles can efficiently restrain the unfavorable γ-NiOOH species construction, stabilizing β-NiOOH forms at pH 12 [51] . The presence of β-NiOOH species is considered to be an improved electroactive phase for superior electrochemical performance in an alkaline medium [52] . The presence of the CuNi bimetallic nanospecies facilitates the e¯ transfer between the GC surface and the GOx, improving glucose detection. Scheme 1 describes the glucose-sensing mechanism of the enzymatic biosensor. GOx catalyzes the glucose oxidation to gluconolactone in the existence of oxygen while generating (H2O2) hydrogen peroxide and water as by-products. In addition, further carboxylic acid, gluconic acid are formed as a result of the reaction of gluconolactone and water. On the other hand, (FAD + ) flavin adenine dinucleotide is employed as a redox cofactor for GOx to perform this oxidation reaction. FAD + acts as an electron acceptor, which will be reduced to FADH2 during the redox process [53] . Subsequent reaction with (O2) to yield hydrogen peroxide revive the FAD + cofactor. This process takes place at the anode surface, where the number of transferred electrons can be interrelated to the quantity of generated H2O2 and, therefore, the glucose content [53] . Scheme 1 describes the glucose-sensing mechanism of the enzymatic biosensor. GOx catalyzes the glucose oxidation to gluconolactone in the existence of oxygen while generating (H 2 O 2 ) hydrogen peroxide and water as by-products. In addition, further carboxylic acid, gluconic acid are formed as a result of the reaction of gluconolactone and water. On the other hand, (FAD + ) flavin adenine dinucleotide is employed as a redox cofactor for GOx to perform this oxidation reaction. FAD + acts as an electron acceptor, which will be reduced to FADH 2 during the redox process [53] . Subsequent reaction with (O 2 ) to yield hydrogen peroxide revive the FAD + cofactor. This process takes place at the anode surface, where the number of transferred electrons can be interrelated to the quantity of generated H 2 O 2 and, therefore, the glucose content [53] . NiOOH + Glucose → Ni(OH) 2 + Gluconolactone (9) CuOOH + Glucose → Cu(OH) + Gluconolactone (10) GOx(FAD) + Glucose → GOx(FADH2) + Gluconolactone (7) GOx(FADH2) → GOx(FAD) + 2e -+ 2H + (8) NiOOH + Glucose → Ni(OH)2 + Gluconolactone (9) CuOOH + Glucose → Cu(OH) + Gluconolactone (10) Scheme 1. Schematic illustration for glucose sensing mechanism based on CuNi-MC/GOx catalyst. The electrochemical behavior of CuNi-MC was explored at a scan rate that varied from 5 to 150 mV/s in sweat solution (Figure 5a ). It is noteworthy that by increasing the scan rate, the hydroxide ions (OH − ) penetrate the generated Ni(OH)2 layer giving rise to thickening the NiOOH wrought layer [54] . Both cathodic and anodic peak currents increased in proportion to the scan rates with a linear regression formula of ipa = 0.186-0.02 v 1/2 (n = 10, R = 0.9967) and ipc = -0.034 + 0.019 v 1/2 (n = 10, R = 0.9943), see Figure 5b . These outcomes specify that the electrochemical process of Ni(OH)2/NiOOH transformation is diffusion-controlled [54, 55] . The ipc shifted to higher negative values, and the ipa shifted positively by increasing the scan rates, leading to a larger peak-to-peak potential separation. The response of CuNi nanoalloys to glucose electro-oxidation is considerably better than that of pure Cu or Ni due to the synergistic electrochemical influence of the bimetallic nanoparticles since the unalloyed Ni is not robust under highly oxidizing conditions, and Cu has a stabilizing influence on the Ni surface [56] . The electrochemical behavior of CuNi-MC was explored at a scan rate that varied from 5 to 150 mV/s in sweat solution (Figure 5a ). It is noteworthy that by increasing the scan rate, the hydroxide ions (OH − ) penetrate the generated Ni(OH) 2 layer giving rise to thickening the NiOOH wrought layer [54] . Both cathodic and anodic peak currents increased in proportion to the scan rates with a linear regression formula of i pa = 0.186 − 0.02 v 1/2 (n = 10, R = 0.9967) and i pc = −0.034 + 0.019 v 1/2 (n = 10, R = 0.9943), see Figure 5b . These outcomes specify that the electrochemical process of Ni(OH) 2 /NiOOH transformation is diffusion-controlled [54, 55] . The i pc shifted to higher negative values, and the i pa shifted positively by increasing the scan rates, leading to a larger peak-to-peak potential separation. The response of CuNi nanoalloys to glucose electro-oxidation is considerably better than that of pure Cu or Ni due to the synergistic electrochemical influence of the bimetallic nanoparticles since the unalloyed Ni is not robust under highly oxidizing conditions, and Cu has a stabilizing influence on the Ni surface [56] . Figure 6a shows the corresponding amperometric response for the CuNi-MC/GOx based glucose sensor. With increasing the glucose content from 5 μm to 0.45 mM, the response of the modified electrode was enhanced due to the enhanced conductivity of the catalyst. Figure 6b exhibits a calibration plot between glucose and the current response of CuNi-MC/GOx modified electrode. The sensor showed a sensitivity of 11.7 ± 0.061 mA mM cm −2 with a linear range of 0.005 to 0.45 mM and a response time of 4 s. This can be Figure 6a shows the corresponding amperometric response for the CuNi-MC/GOx based glucose sensor. With increasing the glucose content from 5 µm to 0.45 mM, the response of the modified electrode was enhanced due to the enhanced conductivity of the catalyst. Figure 6b exhibits a calibration plot between glucose and the current response of CuNi-MC/GOx modified electrode. The sensor showed a sensitivity of 11.7 ± 0.061 mA mM cm −2 with a linear range of 0.005 to 0.45 mM and a response time of 4 s. This can be explained by the high electrocatalytic activity of the as-prepared catalyst, which acquires the uniformity of the glucose oxidase (GOx) enzyme to perform the electrochemical reaction between the electrode and glucose analyte. Some easily oxidizable compounds such as fructose, sucrose, dopamine, and uric acid coexist with glucose in human sweat. The selectivity of these compounds was analyzed with amperometric measurements concerning glucose concentration. There is no substantial interference with uric acid, dopamine, fructose, and sucrose (Figure 6c ). The molecular size of mesoporous carbon is 50 nm, whereas the molecular size of glucose is 0.9 nm. This might be the reason for enhancement in glucose sensitivity, as the glucose can be entrapped in the sites of the mesoporous carbon and interact with GOx enzyme with an increased surface area. Additionally, the molecular size of the other interacting species such as uric acid, dopamine, fructose, and sucrose is more or less similar to glucose. Nevertheless, the presence of GOx enzyme enhances the selectivity of glucose by interacting with the glucose molecules. Therefore, the mesoporous carbon coated on Cu-Ni is selective towards glucose rather than interfering species. The stability of the as-prepared catalyst has been assessed by recording the current response of the redox couple in 1 mM glucose solution after 20 consecutive voltammetric cycles, see Figure 7 . It is worth mentioning that the redox current response has insignifi- The limit of detection (LOD = 5.2 µm L −1 ) is calculated employing the following formula. LOD = 3σ m (11) where σ and m correspond to the standard deviation of the linear range produced after the addition of glucose and the slope of the fitted calibration curve, respectively. Figure 6d displayed that a fast current response was attained upon adding 40 µM glucose, reaching the steady-state current in less than 4 s. The rapid response time exhibits competent activity of the as-prepared catalyst toward glucose sensing. The stability of the as-prepared catalyst has been assessed by recording the current response of the redox couple in 1 mM glucose solution after 20 consecutive voltammetric cycles, see Figure 7 . It is worth mentioning that the redox current response has insignificantly reduced after a continuous scan of 20 cycles, demonstrating that the as-prepared sensor is robust and stable. The stability of the as-prepared catalyst has been assessed by recording the current response of the redox couple in 1 mM glucose solution after 20 consecutive voltammetric cycles, see Figure 7 . It is worth mentioning that the redox current response has insignificantly reduced after a continuous scan of 20 cycles, demonstrating that the as-prepared sensor is robust and stable. Table 1 shows a comparative study of the mesoporous carbon's sensitivity with Ni-Cu on glucose with different modified electrodes. The electrocatalytic activity with glucose for the electrode is depicted in terms of the electrode's high sensitivity. We introduced the mesoporous carbon along with CuNi, which enhanced the glucose sensitivity as the surface area for glucose interaction increased. The highly conductive mesoporous carbon increases the electron transport, which results in high sensitivity of glucose. The electrode modified with mesoporous carbon coated with Ni-Cu shows high sensitivity compared to other electrodes considered here. The composite mesoporous carbon coated with Ni-Cu nanoparticles can be used to fabricate wearable sensors to detect the amount of glucose in sweat. The as-synthesized catalyst's structural and morphological properties were assessed using XRD and TEM techniques. The electrochemical analysis showed a high electroactivity of the as-synthesized catalyst to glucose electro-oxidation. The catalyst exhibited glucose sensitivity at a wide linear range from 0.005 to 0.45 mM, a low detection limit of 5.2 µM (S/N = 3), high sensitivity of 11.7 mA mM cm −2 , and a fast response time of 4 s. The fabricated electrode's sensitivity range was compared with other previously established electrodes suggesting that the composite is a promising material for the fabrication of wearable sensors for glucose detection. The high sensitivity and selectivity of the glucose sensor can be explained on the basis of the molecular structure and its integrations. Global, Regional, and National Burden and Trend of Diabetes in 195 Countries and Territories: An Analysis from 1990 to 2025 Global and Regional Diabetes Prevalence Estimates for 2019 and Projections for 2030 and 2045: Results from the International Diabetes Federation Diabetes Atlas Single Iron Site Nanozyme for Ultrasensitive Glucose Detection Admission Hyperglycaemia as a Predictor of Mortality in Patients Hospitalized with COVID-19 Regardless of Diabetes Status: Data from the Spanish SEMI-COVID-19 Registry A Wearable Chemical-Electrophysiological Hybrid Biosensing System for Real-Time Health and Fitness Monitoring Non-Invasive Wearable Electrochemical Sensors: A Review Nanowires Pin Neurons: A Nano Metal Oxide Based Non-Enzymatic Electrochemical Sensors for Glucose Detection Conjugated Polymers for Biosensor Devices Wearable Salivary Uric Acid Mouthguard Biosensor with Integrated Wireless Electronics Fully Integrated Wearable Sensor Arrays for Multiplexed In Situ Perspiration Analysis Enzyme-Based Body-Worn Electronic Devices Wearable Electrochemical Sensors and Biosensors: A Review Direct Enzymatic Bioelectrocatalysis: Differentiating between Myth and Reality Principles of Direct (Mediator Free) Bioelectrocatalysis Oxygen Reduction on Redox Mediators May Affect Glucose Biosensors based on Unusual Response in Mediated Biosensors with an Oxidase/Peroxidase Bienzyme System A Graphene-based Electrochemical Device with Thermoresponsive Microneedles for Diabetes Monitoring and Therapy Wearable/Disposable Sweat-Based Glucose Monitoring Device with Multistage Transdermal Drug Delivery Module Electrochemical Glucose Sensing: Is There Still Room for Improvement? Flexible Enzymatic Glucose Electrochemical Sensor Based on Polystyrene-Gold Electrodes Correlation Between Sweat Glucose and Blood Glucose in Subjects with Diabetes Wearable Non-Invasive Epidermal Glucose Sensors: A Review The (In)Dependency of Blood and Sweat Sodium, Chloride, Potassium, Ammonia, Lactate and Glucose Concentrations during Submaximal Exercise. Graefe's Arch Miniature Biofuel Cell as a Potential Power Source for Glucose-Sensing Contact Lenses Progress of Advanced Nanomaterials in the Non-Enzymatic Electrochemical Sensing of Glucose and H 2 O 2 Highly Ordered Platinum-Nanotubule Arrays for Amperometric Glucose Sensing High-Performance Non-Enzymatic Glucose Sensors Based on CoNiCu Alloy Nanotubes Arrays Prepared by Electrodeposition Amperometric Biosensor Based on Multilayer Containing Carbon Nanotube, Plasma-Polymerized Film, Electron Transfer Mediator Phenothiazine, and Glucose Dehydrogenase Mediator-Free Electron-Transfer on Patternable Hierarchical Meso/Macro Porous Bienzyme Interface for Highly-Sensitive Sweat Glucose and Surface Electromyography Monitoring Highly Sensitive and Wearable In 2 O 3 Nanoribbon Transistor Biosensors with Integrated On-Chip Gate for Glucose Monitoring in Body Fluids Peroxidase-Like Activity of Smart Nanomaterials and Their Advanced Application in Colorimetric Glucose Biosensors Porous Cu-NiO Modified Glass Carbon Electrode Enhanced Nonenzymatic Glucose Electrochemical Sensors Diabetes Emergencies around Ramadan Study (DEARS)-A Multi-Center Study of Diabetes Emergencies Admitted before, during and after Ramadan in Malaysia Mitigation of Hypoglycemia During Ramadan Using the Flash Glucose Monitoring System Following Dose Adjustment of Insulin and Sulphonylurea in Patients Taking Multiple Glucose-Lowering Therapies (The PROFAST-IT Study) Biosensors and their Applications-A Review Recent Advances in Noninvasive Glucose Monitoring Ni Bimetallic Nanoparticles: Facile Synthesis, Characterization and its Application in Photodegradation of Organic Dyes Recent Advances in Enzymatic and Non-Enzymatic Electrochemical Glucose Sensing A Novel High Performance Non-Enzymatic Electrochemical Glucose Biosensor based on Activated Carbon-Supported Pt-Ni Nanocomposite Core-Shell Gold-Nickel Nanostructures as Highly Selective and Stable Nonenzymatic Glucose Sensor for Fermentation Process A Surface Analytical Examination of Passive Layers on CuNi Alloys: Part I. Alkaline Solution Constant Potential Amperometric Detection at a Copper-Based Electrode: Electrode Formation and Operation A Highly Sensitive Electrochemical Glucose Sensor Structuring with Nickel Hydroxide and Enzyme Glucose Oxidase Tailored Ni-Cu Alloy Hierarchical Porous Nanowire as a Potential Efficient Catalyst for DMFCs Electrocatalytic Oxidation of Glucose on Ni and NiCu Alloy Modified Glassy Carbon Electrode Glucose Oxidase-An Overview Cyclic Voltammetry Deposition of Nickel Nanoparticles on Tio2 Nanotubes and their Enhanced Properties for Electro-Oxidation of Methanol Facile synthesis of 2D/3D hierarchical NiCu bimetallic MOF for non-enzymatic glucose sensor Copper/Nickel Nanoparticle Decorated Carbon Nanotubes for Nonenzymatic Glucose Biosensor Bioelectrocatalysis and Surface Analysis of Gold Coated with Nickel Oxide/Hydroxide and Glucose Oxidase towards Detection of Glucose Smart Nanocomposites of Cu-Hemin Metal-Organic Frameworks for Direct Electrochemistry of Glucose Oxidase Immobilized on Au Nanoparticles-Functionalized 3D Hierarchically ZnO Nanostructures and its Application to Bioelectrochemical Glucose Sensor Vertically aligned Pt Nanowire Array/Au Nanoparticle Hybrid Structure as Highly Sensitive Amperometric Biosensors Multiwalled Carbon Nanotubes Coated with Cobalt(II) Sulfide Nanoparticles for Electrochemical Sensing of Glucose via Direct Electron Transfer to Glucose Oxidase The authors declare no conflict of interest.