key: cord-1019167-elyk7dgc authors: Mao, Rui; Qi, Lifei; Wang, Zhuo; Liu, Hongtao; Du, Yuguang title: Helix loop-mediated isothermal amplification of nucleic acids date: 2018-05-23 journal: RSC advances DOI: 10.1039/c8ra01201f sha: 539705717ffef790970d65095722319c67f55faf doc_id: 1019167 cord_uid: elyk7dgc Isothermal nucleic acid amplification has played a key role in the point of care test (POCT). In this study, a helix loop-mediated isothermal amplification (HAMP) method with high specificity, efficiency and rapidity was developed. The MERS-Cov orf1b gene was chosen for the validation and optimization of HAMP. The HAMP analysis was performed at a constant temperature of 61–65 °C and yielded a self-primed spiral structure with no introduction of exogenous gene sequence by two pairs of specially designed primers. The primers for helix loop formation were composed of two complementary primers including the helix forward primer and the helix reverse primer, the 3′ ends of which were complementary to their respective target nucleic acids. HAMP assay can be monitored by fluorescence signals with the addition of Eva Green in the reaction mixture. In addition, an accelerated HAMP was developed after the addition of acceleration probe, which could be finished within 75 min with a sensitivity of 10 copies per reaction. Further, a reverse transcription-HAMP (RT-HAMP) was proven to be feasible for RNA detection by combining the reverse transcriptase with DNA polymerase. Finally, both the HAMP and RT-HAMP assay were visually conducted by using Hydroxynaphthol blue (HNB) as a chromogenic indicator. All in all, it is suggested that the HAMP assay would have great potential in POCT applications. Nucleic acid amplication is one of the most valuable tools in life science research because of its applications in molecular diagnosis, forensic investigations, and biomedical developments. Polymerase chain reaction (PCR) was the rst and is still the most popular amplication technology for nucleic acid analysis. However, PCR is usually complex and timeconsuming, requiring sophisticated thermal cycling instruments and well-trained operators. 1 These disadvantages have limited its application to some special elds such as point-ofcare use in resource-limited areas. 2 In an effort to reduce the resource requirements of PCR, a variety of isothermal amplication techniques have been developed and applied widely, including self-sustained sequence replication reaction (3SR), 3 rolling circle amplication (RCA), 4 strand displacement amplication (SDA), 5 helicasedependent amplication (HDA), 6 loop-mediated isothermal amplication (LAMP) 7 and recombinase polymerase amplication (RPA), 8 Cross-priming amplication (CPA), 9 and Polymerase Spiral Reaction (PSR), 10 etc. 3SR, NASBA, SDA, HDA and RPA, require multiple enzymes and rigorous optimization, which may increase the assay cost. As to the single-enzyme based methods, LAMP and CPA require four or more primers, which may pose challenges for primer design, 11 although specic primer-design soware is available. Among the above-mentioned methods, PSR, based on auto nucleic acid synthesis with spiral structure (the scheme is presented in Fig. 1a) , has been successfully applied to the rapid detection of toxA gene. 12 However, an exogenous botanic gene was introduced to the primer design, which may inuence the selection of appropriate primers. Furthermore, the overlap part in the starting procedure of helix self-priming elongation is relatively short, and this may affect the self-annealing and elongation efficiency, which may easily lead to the amplication bias and false positive/negative results. In this study, a helix loop-mediated isothermal amplication (HAMP) method inspired by spiral structure in PSR was developed. Similar to LAMP, HAMP was performed with two pairs of primers at a constant temperature. Unlike PSR, there is no exogenous sequences introduced in primer design to start a self-priming helix structure formation. The structure can be formed by slightly modied PCR primer and the overlap part of helix loop in HAMP is larger compared with that in PSR which would enhance the stability of the self-priming structure. In addition, the acceleration probe for HAMP reaction will dramatically improve the amplication efficiency and sensitivity. Moreover, an Mg 2+ indicator, Hydroxynaphthol blue (HNB) was introduced into the reaction system to offer a visual judgement method for end-point result based on the colour change. Overall, the HAMP method may offer an alternative for the point of care test. A 550 bp DNA fragment of the MERS-orf1b genome (GenBank: KU308549.1) with a T7 promoter sequence cloned into a pUC57simple vector was provided by BGI Biological Engineering Technology and Services Co. Ltd (Shenzhen, China). Other DNA fragments of MERS-CoV, such as orf1a (GenBank: KX108946.1), N2 (GenBank: KU242424.1), N3 (GenBank: KJ156873.1), upE (GenBank: KX108944.1) were also provided by BGI. The copy number of the plasmid was determined using DNA copy number calculator available at http://cels.uri.edu/gsc/ cndna.html. The articial RNA of MERS-orf1b was prepared by RiboMax™ Large Scale RNA Production System following the recommended instruction. Primers and probes (listed in Table 1 ) were provided by BGI Biological Engineering Technology and Services Co. Ltd (Shenzhen, China). The Bst 2.0 WarmStart DNA polymerase, AMV Reverse Transcriptase and 10 Â ThermoPol reaction buffer (including 200 mM Tris-HCl, 100 mM KCl, 100 mM (NH 4 ) 2 SO 4 , 20 mM MgSO 4 , and 1% Triton X-100) were purchased from New England BioLabs (Ipswich, MA, USA). Deoxynucleotide triphosphates (dNTPs) and RiboMax™ Large Scale RNA Production Systems were from Promega (Madison, WI, USA). Eva Green and GelRed were obtained from Biotium (Hayward, CA, USA). Restricted enzymes were purchased from Thermo Fisher Scientic (Shanghai, China). DNA purication kit and Nuclease-free water were purchased from Tiangen Biotech Co., Ltd. (Beijing, China). Other reagents, unless specied, were obtained from Sigma-Aldrich (St. Louis, MO, USA). The real-time uorescence detection of nucleic acid was performed using StepOne™ Real-Time PCR detection system (The Applied Biosystems) at 1 min intervals under a constant temperature and followed by the melting curve analysis procedure. Gels were imaged by Tanon 4100 Gel Image Analysis System (Tanon Science & Technology Co., Ltd.). Nucleic acid concentration was determined by using the Quawell UV-Vis Spectrophotometer Q5000 (Quawell Technology, Inc.). HAMP reactions were carried out in a 25 ml reaction mixture system as followed: 8 U Bst 2.0 WarmStart DNA polymerase, 2.5 ml 10 Â ThermoPol reaction buffer, 1 M betaine, 6 mM MgSO 4 , 1.4 mM each dNTP and appropriate nucleic acid template. For one reaction, 1.6 mM HF, 1.6 mM HR, 0.2 mM F2 and 0.2 mM R2 primers were used. In addition, 0.8 mM Fin and Rin primers were added to accelerate the HAMP reaction. At last, nucleasefree water was added to obtain a nal volume of 25 ml. For application of HAMP on detecting of RNA targets, 5 U AMV Reverse Transcriptase instead of water was further added into the reaction mixture. The reactions were performed at 63 C on StepOne™ real-time PCR system or water bath. HAMP amplication products were determined by gel electrophoresis, uorescence-based detection using the real time PCR or direct visual detection using Mg 2+ indicator. For gel electrophoresis analysis, the amplication products were rstly digested using the restriction enzyme HindIII found in MERS-orf1b gene, and then analysed by the electrophoresis Fig. 1 The schematic presentation of PSR and HAMP system and primer design of HAMP system. (a) Representative scheme about the PSR system; (b) representative scheme about the HAMP system; (c) localization of primers and restriction enzyme cutting sites in the nucleotide sequence of MERS-orf1b (part) for HAMP method. Paper in 1% agarose gel in 1 Â TAE buffer (40 mM Tris-acetate, 1 mM EDTA, pH 8.0). DNA was stained with GelRed (Biotium, USA) and further visualised by Gel Image Analysis System. For uorescence-based detection, Eva Green, a sequence-unspecic nucleic acid labelling dye used for real time of monitoring PCR and LAMP reactions, 13 was used for quantitatively analysing amplication results of HAMP based on the uorescence intensity detected by the StepOne™ real-time PCR detection system. For visual detection, 1 ml Hydroxynaphthol blue (HNB) solution (3 mM) was added to the reaction mixture. A positive reaction was indicated by a colour change from violet to sky blue, while a negative reaction remained on violet colour. The Plasmid DNA of pUC57-orf1b was used as the template with 10-fold serial dilution and resulted in 10 8 , 10 7 , 10 6 , 10 5 , 10 4 , and 10 3 copies of the template per reaction respectively. And the articial RNA of MERS-orf1b gene was synthetized by RiboMax™ Large Scale RNA Production System. For the amplication of RNA template, the same serial dilution was performed using the RNA of MERS-orf1b. And 10 8 , 10 7 , 10 6 , 10 5 , 10 4 , 10 3 and 10 2 copies of RNA per reaction were used to evaluate the sensitivity of reverse transcription-HAMP (RT-HAMP). The Plasmid DNA of orf1a, N2, N3 and UpE were used for specicity evaluation of HAMP. All experiments described above were replicated to ensure the reproducibility. Similar to LAMP and PSR, HAMP is dependent on auto-cycling strand displacement by a DNA polymerase with high strand displacement activity and a set of two specially designed primers and two outer primers. The reaction mechanisms of PSR and HAMP were shown in Fig. 1a and b, respectively. As to the primer design of HAMP, the sequences (typically 20-24 nt) at both ends of the target region for amplication in a DNA were designated as F1c and R1. Two sequences (typically 20-24 nt) inside both ends of F1c and R1 were designated as Finc and Rin (not shown in Fig. 1b) and two sequences (typically 17-21 nt) outside the ends of F1c and R1 were designated as F2c and R2. Given this structure above, the sequences of Helix forward primer (HF) and Helix reverse primer (HR) were designed as the following: the primer HF contained two parts, i.e., the reverse sequence of R1 (R1r) and the complementary sequence of F1c (F1). The primer HR contained two parts: the reverse complementary sequence of F1c (F1r) and the sequence of R1. Evidently, the sequences of HF and HR were reverse to each other. The two inner primers (accelerate probe) consisted of Rin and the sequence (Fin) complementary to Finc. The two outer primers consisted of the sequence of R2 and the complementary sequence of F2c (F2). The mechanism of HAMP is shown in Fig. 1b . Firstly, the two strands of DNA double helix are separated in the presence of betaine at 61-65 C. Then, the F1 segment of HF primer hybridizes to the F1c for the target sequence amplication (structure 1 to structure 2). The outer primer F2 hybridizes with F2c to initiate the DNA synthesis and then displace the newly-produced DNA strand containing HF primer sequence at 5 0 end (structure 4). Next, the R1 segment of HR primer hybridizes to the R1c sequence in the newlyproduced DNA (structure 5). The outer primer R2 hybridizes with R2c sequence in the newly-produced DNA to initiate the DNA synthesis (structure 6), and then displace the DNA strand containing HR primer sequence at 5 0 end (structure 8). The R1rc and F1c of 5 0 regions are reverse complementary to R1 and F1r of 3 0 regions on the DNA strand of structure 8. Thus, the strand could curl to form a helix-loop to start self-primed DNA synthesis (structure 9). Finally, a single helix structure is formed (structure 10) and the non-complementary pairing region can be annealed by accelerated probes and a partially double helix structure may be formed (structure 11). Basically, the primer design for HAMP reaction is relatively simple. The two target sites (F1 and R1) and two outer primer sites (F2 and R2) could all be selected by the common PCR primer design soware (e.g., DNASTAR, DNAMAN). Similar to PSR, the reactions of polymerization and displacement in HAMP system are activated by a single target to produce multiple target copies and dependent on auto-cycling strand displacement by a self-primed single helix structure. However, unlike PSR, no exogenous sequences are introduced in the HAMP reaction. Furthermore, a longer overlap part in the helix loop is formed in the HAMP reaction, which may promote the efficiency of self-annealing and elongation. The amplication can be further accelerated by the addition of acceleration probes, which may improve the reaction efficiency and sensitivity by accelerating isothermal amplication and reducing detection time. MERS-CoV orf1b gene (Fig. 1c) was chosen for systemic verication of HAMP using the selected primers ( Table 1 ). As shown in Fig. 2b -c, the crossing threshold (C t ) values of 3 independent reactions for detecting 10 6 copies of MERS-CoV orf1b gene were all about 51 min (Fig. 2b ) and the melting temperature were 84.32 C (Fig. 2c) . Both the real-time amplication curve and melting curve analysis indicated a good repeatability of HAMP. Furthermore, the result of negative control indicated that non-specic amplication failed to occur. The HAMP amplication product was further determined by enzyme digestion of HindIII, which cuts the correct amplication product into 146 bp fragments. As shown in Fig. 2a , the digested product showed a strong band approximately 100-200 bp. Two bands from undigested products were excised from the gel and sequenced. The sequencing result conrmed that the DNA bands were corresponded to the predicted HAMP ampli-cation product of MERS-CoV orf1b sequence. The MERS-CoV orf1b plasmid DNA was used as target nucleic acids to determine the sensitivity of HAMP. A serial dilution of MERS-CoV orf1b plasmid was used as templates. As shown in Fig. 3a , the amplication by HAMP was completed within 80 min. As low as 10 3 copies of template was detected by HAMP, indicating a good sensitivity, although the amplication efficiency is relatively low at this concentration of template. Meanwhile, the standard amplication curve of HAMP showed a ne regression coefficient (R 2 ¼ 0.9969) (Fig. 3b) , suggesting the potential application of HAMP for quantitative analysis. In addition, primary evaluation of HAMP specicity was conducted by applying the fragments of MERS-CoV (orf1a, N2, N3 and UpE) and no amplication was observed (data not shown). The acceleration probes were designed to further improve amplication efficiency of HAMP. The probe targeted the MERS-CoV orf1b sequence between F1 and R1, and annealed to the single strand part of helix loop during HAMP amplication (Fig. 1a, structure 11 ). The amplication product of HAMP using acceleration probes was also analysed by enzyme digestion and gel electrophoresis. The digested product is similar to that of HAMP, and the negative control showed no amplication product (Fig. 2a) . The amplication products were also excised form the gel and further conrmed by sequencing. These results indicated that addition of acceleration probe did not inuence the specicity of HAMP reaction. But different bands were formed by comparing their products (lane 1 and lane 4), implying that the annealing of acceleration probe may produce different structures of nucleic acids with similar nucleic acid sequence. The amplication efficiency of HAMP was further analysed. Results showed that the addition of acceleration probes shortened the threshold time in the amplication curve by approximately 20 min compared to that in HAMP with the same template concentration, indicating a dramatic improvement in the HAMP efficiency (Fig. 3a and c) . However, the regression coefficient on the standard amplication curve (R 2 ¼ 0.9834) was worse than that in HAMP (Fig. 3d) . Theoretically, the developed HAMP produced no non-specic products when the appropriate accelerated primers were introduced. Noticeably, accelerated primers can recognize the specic sequences included in the amplication products, hence the DNA synthesis will exhibit higher efficiency in HAMP analysis. The accelerated primers were also applied to PSR. 12 Since the isothermal nucleic acid amplication was induced by the same helix loop structure, the HAMP method showed similar amplication efficiency and sensitivity in comparison with PSR. The feasibility of HAMP for RNA detection was demonstrated by combining the reverse transcriptase with DNA polymerase. In Standard amplification curve obtained by plotting the C t value versus the plasmid copy number. The DNA concentration was as followed: 10 8 copies, 10 7 copies, 10 6 copies, 10 5 copies, 10 4 copies, 10 3 copies, 10 2 copies and negative control (nuclease-free water). (c) Representative isothermal amplification plot by real time PCR analysis on serially diluted dsDNA with addition of acceleration probe. Amplification was performed at 63 C for 65 min. (d) Standard amplification curve obtained by plotting the C t value versus the plasmid copy number. The DNA concentration was as followed: 10 6 copies, 10 5 copies, 10 4 copies, 10 3 copies, 10 2 copies, 10 1 copies, 10 0 copies and negative control (nuclease-free water). amplification curve obtained by plotting the C t value versus the plasmid copy number. The RNA concentration was as followed: 10 8 copies, 10 7 copies, 10 6 copies, 10 5 copies, 10 4 copies, 10 3 copies, 10 2 copies and negative control (nuclease-free water). this assay, a serial dilution of the articial RNA of MERS-orf1b gene was uses as templates. As shown in Fig. 4 , the detection limit of RT-HAMP was 10 4 copies within 70 min with a regression coefficient (R 2 ¼ 0.9723). The result suggested that the RT-HAMP was suitable for RNA target amplication. The amplication efficiency of RT-HAMP is likely determined by the activity of AMV reverse transcriptase and DNA polymerase, more suitable enzymes may be screened to optimize our amplication system in the future work. HNB was applied to the visual detection of HAMP and RT-HAMP. As shown in Fig. 5 , visual detection of HAMP and RT-HAMP for nucleic acids worked well in this study aer HNB was chosen as colorimetric agent, which would make it applicable in resource limit areas. Nucleic acid amplication usually produced a large amount of pyrophosphate ion by-products, which may react with Mg 2+ to form insoluble magnesium pyrophosphate, accompanied by the decrease in Mg 2+ concentration in reaction mixture. 14 On the other hand, some chromogenic agents such as HNB can chelate Mg 2+ to form a stable chemical complex with the purple colour produced. With the consumption of Mg 2+ , the solution colour will change into sky blue. Based on the above, the colorimetric monitoring method was developed for LAMP reaction, 15, 16 and proved to be also suitable in HAMP. In recent years, the point of care test (POCT) has been paid more attention for its huge potential in global healthcare. 17 At the same time, the POCT should meet the guidelines proposed by the World Health Organization (WHO) for developing diagnostic techniques, namely, ASSURED (Affordable, Sensitive, Specic, User friendly, Robust and Rapid, Equipment free and Deliverable). 18 In accordance with the guidelines by WHO, the HAMP system developed in this study allowed a good specicity and a sensitivity of 10 DNA copies per reaction within 75 min, and the method is suitable for HNB-based visual detection by naked eye. It is suggested that HAMP assay would be of great potential in the application to POCT. There are no conicts to declare. Visual detection of HAMP and RT-HAMP for nucleic acids. The concentration of both plasmid DNA and mimic RNA was 10 6 copies in the amplification system. Amplification was performed at 63 C for 60 min. Reaction 1 and 2, MERS-orf1b plasmid DNA; reaction 3 and 4, negative control of 1 and 2 The authors thank Chinese Academy of Sciences for the funding of this research (Grant No. CXJJ-15Z012).