key: cord-1018407-k11asjaw authors: Wang, Xiang; Zhang, Qian; Zhang, Fang; Ma, Fenlian; Zheng, Wenzhi; Zhao, Zhihui; Bai, Yinglong; Zheng, Lishu title: Visual detection of the human metapneumovirus using reverse transcription loop-mediated isothermal amplification with hydroxynaphthol blue dye date: 2012-07-27 journal: Virol J DOI: 10.1186/1743-422x-9-138 sha: 7c19aadb1294d6b98fc8b947a98eefac0da6b870 doc_id: 1018407 cord_uid: k11asjaw BACKGROUND: Human metapneumovirus (hMPV) is a major cause of acute respiratory infections ranging from wheezing to bronchiolitis and pneumonia in children worldwide. The objective of this study is to develop a visual reverse transcription loop-mediated isothermal amplification (RT-LAMP) assay for the detection of hMPV and applied to the clinical samples. RESULTS: In this study, visual RT-LAMP assay for hMPV was performed in one step with the addition of hydroxynaphthol blue (HNB), and were used to detect respiratory samples. Six primers, including two outer primers (F3 and B3), two inner primers (FIP, BIP) and two loop primers (LF and LB), were designed for hMPV N gene by the online software. Moreover, the RT-LAMP assay showed good specificity and no cross-reactivity was observed with human rhinovirus (HRV), human respiratory syncytial Virus (RSV), or influenza virus A/PR/8/34 (H1N1). The detection limit of the RT-LAMP assay was approximately ten viral RNA copies, lower than that of traditional reverse transcriptase polymerase chain reaction (RT-PCR) 100 RNA copies. In the 176 nasopharyngeal samples, 23 (13.1%) were conformed as hMPV positive by RT-LAMP, but 18 (10.2%) positive by RT-PCR. CONCLUSION: Compared with conventional RT-PCR, the visual hMPV RT-LAMP assay performed well in the aspect of detect time, sensitivity, specificity and visibility. It is anticipated that the RT-LAMP will be used for clinical tests in hospital or field testing during outbreaks and in emergency. Osterhaus first reported and named the human metapneumovirus (hMPV) [1] ; it was also detected in Europe, America, Australia, and other countries [2] [3] [4] . In 2003, hMPV was reported in China [5] . hMPV is a negativesense, single-stranded RNA virus belonging to the genus Metapneumovirus in the subfamily Pneumovirinae of the family Paramyxoviridae. It often co-infects patients and causes acute respiratory diseases in different age groups together with other respiratory virus, and severe cases are suffered mostly by infants, the elderly, and patients with heart and lung diseases [6] [7] [8] [9] . Currently, the most common hMPV detection methods are by virus isolation, enzyme-linked immunoassays, McAb immuno-fluorescent assays and molecular biology methods based on the polymerase chain reaction (PCR, RT-PCR, nested PCR, and real-time PCR). However, sensitive cell lines for hMPV are limited. The cell line LLC-MK2 can reach a 100% infection within 5 days of isolation [10] , but this is far from fulfilling the requirements in terms of time efficiency for positive identification during outbreaks. In comparison, hMPV antigen detection requires the preparation of monoclonal antibodies; this entails significant costs and can be time-consuming. Although RT-PCR, qRT-PCR, and microarrays are suitable for high-throughput analyses and can produce rapid, specific, and sensitive results, the expensive instruments required for such tests restrict their application. LAMP is a novel nucleic acid isothermal amplification technique developed by Notomi [11] . LAMP relies on autocycling strand-displacement DNA synthesis performed by Bst DNA polymerase. Additionally, RNA can be amplified by reverse transcription coupled with LAMP (RT-LAMP) in one step, making it suitable to detect RNA viruses such as hMPV. RT-LAMP employs a set of six specially designed primers that recognize eight distinct sequences of the target, and only when three pairs of specific primers match the target sequence exactly can the gene amplification start. The whole reaction can be completed in 1.5 h and the amplified product shows a ladderlike pattern by agarose gel electrophoresis. Moreover, this method produces a large amount of amplified product, resulting in easier detection by visual judgment according to the turbidity of the reaction mixture. Therefore, LAMP and RT-LAMP method have been used as a powerful gene amplification tool due to its high specificity and sensitivity under isothermal condition. In this study, we established a RT-LAMP assay to detect hMPV, and the amplification result can be observed by hydroxynaphthol blue (HNB) dye-mediated visualization using the naked eye. Compared with conventional RT-PCR, the newly established visual RT-LAMP assay is simple, efficient, cost-effective and convenient that could be used as a diagnostic tool in clinical respiratory samples. As shown in Figure 1A , RT-LAMP products of hMPV RNA showed a ladder-like pattern upon agarose gel electrophoresis. When the sample tube did not contain target RNA, no amplification was seen. In addition, The amplification products can be observed by naked eyes under the natural light. The negative control reaction showed violet, but the color became sky blue for the positive amplification. Sequencing results of the recombinant plasmid pGEM-T-N217 indicated that the 217 bp target sequence was 100% homologous to the sequence of hMPV N gene used for the primers design (GenBank EF081360) (not shown). By sequence analysis, there exists a unique Msp I restriction site in the F1 region. After digestion of RT-LAMP products with the Msp I overnight, the 103 and 116 bp fragments were observed, and were in good accordance with those predicted theoretically from the expected structures. The results of sequence and digestion confirmed that the products were amplified specificly from the hMPV target region, ( Figure 1B ). Specificity and sensitivity of the hMPV RT-LAMP HRV, influenza virus A/PR/8/34 (H1N1), and RSV viral nucleic acid were amplified by RT-LAMP method using 3 pairs of primers of hMPV. No cross-reactivity was detected with the other three respiratory viruses, indicating a high level of specificity of the RT-LAMP assay for hMPV ( Figure 2A ). Positive color (sky blue) was only observed in the preparation of the hMPV, whereas none of the other viruses showed a color change. In Figure 2B , lanes 1-5 showed that decreases in the amount of nucleic acids (10 5 -10 copies/reaction) generated fading amplified band signals. Ten or more copies of the nucleic acid resulted in RT-LAMP-specific ladderlike bands, indicating a sensitivity level of at least 10 copies needed for amplification. RT-PCR products (3ul) were subjected to electrophoresis on 1% agarose gel stained with ethidium bromide, and a 457 bp specific band could be seen clearly. As shown in Figure 3 , RT-PCR assay could detect as little as 100 copies of RNA molecules, indicating less sensitive than the RT-LAMP method. To evaluate the clinical sensitivity of the RT-PCR assay and RT-LAMP assay under routine conditions, 176 nasopharyngeal specimens obtained from children with respiratory tract infections was chosen for investigation using both assays. 23 (13.1%) of these specimens were positive by the RT-LAMP assay, but 18 (10.2%) samples were identified hMPV positive by RT-PCR screening assay. Additionally, all RT-PCR positive specimens could be detected with RT-LAMP. This result suggested that RT-LAMP method was more feasible than RT-PCR when detecting hMPV from clinical samples, because of the higher sensitivity of RT-LAMP. Loop-mediated isothermal amplification assay (LAMP) was originally designed for genetic testing and applied to viruses and other pathogens [11] [12] [13] . Using six primers that target eight distinct gene sequences, the reaction is initiated only if all the primers match with the targets sequence correctly. Thus, these assays offer high levels of specificity, selectivity, sensitivity, and efficiency. In particular, reverse transcription and amplification reactions of RNA viruses can be completed in the meantime under isothermal environments, saving time and reducing experimental error. The most conserved segment within the N gene (GenBank EF081360) was selected as the target. The entire testing process under isothermal conditions of 65°C takes only 1.5 h (the DNA template must be predenatured at 95°C prior to the isothermal reaction). Additionally, results from the LAMP can be readily observed by the white precipitate of magnesium pyrophosphate generated during the reaction. Just because no laboratory equipment is needed for the determination of results, RT-LAMP can easily be used in field-testing. Moreover, A real-time monitoring turbidimeter was developed to monitor LAMP reactions [14] . Through continuous exploration, LAMP has been combined with a fluorescent dye for quantification in real-time reactions [15, 16] . For example, a metal ion indicator, hydroxynaphthol blue (HNB), was added to LAMP assay [17] [18] [19] . The HNB dye-based assay has a remarkable advantage compared with other color-based assays in that HNB is mixed prior to amplification. Thereby, the reaction tubes need not to be opened, which reduce the risk of cross-contamination. LAMP technology has been applied to a variety of nucleic acid detection assays for pathogens, such as H5N1-HPAIV, HBV, HIV-1, MTB, O157, trypanosomes, and plasmodium [15, 16, [20] [21] [22] [23] . This study established an RT-LAMP method with HNB dye to detect the hMPV N gene, which has significantly higher sensitivity (10 copies) levels than the conventional RT-PCR method and did not cross-react with RSV, HRV, or A/PR/8/34 (H1N1) viral nucleic acids. Restriction enzyme and sequence analyses also validated its specificity. In the testing of clinical respiratory samples, 18 hMPV positive samples were detected by the traditional RT-PCR method, but 23 hMPV positive samples by RT-LAMP method due to the higher sensitivity. The epidemiological analysis indicated that children with HMPV varied from 20 days to 146 months of age, and the majority of patients (82.6%, 18/23) were aged at under ≤5 years. The main clinical diagnoses included bronchopneumonia (52.2%, 12/ 23), acute respiratory tract infection (26.1%, 6/23), acute asthmatic bronchopneumonia (17.4%, 4/23) and pneumonia (4.3%, 1/23). In short, RT-LAMP with HNB dye was shown to be a sensitive and easy assay for detection of hMPV. Given these benefits, RT-LAMP is considered a useful and promising technique for the less equipped primary sector. It can also play an important role in the early diagnosis and control of diseases during outbreaks in remote areas or under field conditions. In brief, the RT-LAMP assay is a simple, rapid, and accurate nucleoside amplification method. In combination with HNB dye, the visual RT-LAMP is much easier for the result determination, and this method can be used for the diagnosis of hMPV not only in the laboratory but also in the field. Several hMPV genomic sequences isolated in China were downloaded from GenBank and aligned with each other using the AlignX function in Vector NTI Advance 10 (Invitrogen). The most conserved segment within the N gene of hMPV/CS038 N gene (GenBank EF081360) was selected as the target. All LAMP primers were designed by the online software http://primerexplorer.jp/e/ (Eiken Chemical Co., Ltd., Tokyo, Japan). Six primers were used for the RT-LAMP assay, including two outer primers (F3 and B3), a forward inner primers FIP (F1c-F2) , a backward inner primer BIP (B1c-B2), and two loop primers (LF and LB). The location and sequences of each primer were shown in Table 1 and Figure 4 . The predicted length of RT-LAMP is 217 bp. The RT-LAMP reaction was performed in a total volume of 25 μL mixture, containing 5 μM F3 and B3 (1 μL each primer), 40 μM FIP and BIP (1 μL each primer), 20 μM LF and LB (1 μL each primer), 10 mM dNTPs (2.5 μL), 25 mM betain (1 μL), 150 mM MgSO 4 (1 μL), 10× Bst buffer (2.5 μL), 8 U/μL Bst DNA polymerase (1 μL) (New England Biolabs), 10 U/μL AMV reverse transcriptase (1 μL) (Invitrogen), 3 mM HNB (1 μL), template RNA (1 μL) and double-distilled water (ddH 2 O; 8 μL). The mixture was incubated at 65°C for 1.5 h and then heated at 80°C for 5 min to terminate the reaction. Template RNA was replaced by ddH 2 O as negative control. A portion of each product was analyzed by 1% agarose gel electrophoresis stained with ethidium bromide and photographed under a UV transilluminator. In addition, the amplification product also can be visually inspected for the color change from violet to sky blue, and the negative control reaction should have remained violet. The tubes were observed by naked eyes and photographed under the natural light. After electrophoresis, DNA bands about 217-bp obtained from the RT-LAMP positive gel was extracted using a Gel/PCR DNA Fragments Extraction Kit (Geneaid), and was amplified specially by PCR method with the sense and antisense primers F3 and B3. Amplification was performed as follows: 94°C for 8 min, and then 40 cycles of 94°C for 30 s, at 50°C for 30 s, 72°C for 40 s, with a final 72°C extension for 10 min. The PCR was carried out in a total volume of 30 μL containing 10 μM F3 and B3 (1 μL each primer), 10 mM dNTPs (1 μL); 10× PCR buffer (3 μL), 5 U of rTaq (1 μL), extracted products (5 μL), and ddH 2 O (18 μL). The 217-bp product was extracted after 1% agarose gel electrophoresis, cloned into pGEM-T Easy Vector (Promega), and then transformed into competent DH5α cells (BioMed). The recombinant plasmid pGEM-T-N217 was extracted from positive clones was sequenced by Invitrogen (China). RT-LAMP products were digested with Msp I in a 20 μL reaction containing Msp I (1 μL) (Takara), 10 × Msp I Basal Buffer (2 μL), 0.1% bovine serum albumin (2 μL), LAMP products (5 μL), and ddH 2 O (10 μL), incubated at 37°C overnight, after which the DNA band pattern were analyzed on 12% polyacrylamide gel electrophoresis stained with ethidium bromide and photographed as above. To evaluate the cross-reactivity of the hMPV N gene primer set with other common respiratory viruses, viral nucleic acid extracted from HRV, hRSV, and influenza virus A/PR/8/34 (H1N1) strains were tested and a DEPC-treated water as negative control under the same conditions. To obtain the detection limit of the RT-LAMP assay, the recombinant plasmid pGEM-T-N217 linearized by the vector-specific restriction enzyme Nde I was transcribed in vitro using the T7 RiboMAX Express Large Scale RNA Production System (Promega) and digested with RNase-Free DNase Set (Qiagen). Then the RNA was purified using an RNeasy Mini Kit (Qiagen) according to the manufacturer's protocol and the concentration (μg/mL) was determined by Biophotometer (Eppendorf). The RNA copy number was calculated using the following formula: number of copies (copies/μL) = amount (ng) × 6.022 × 10 23 / length (base pairs) × 10 9 × 340. 10-fold serial dilutions of RNA ranging comprising 10 5 to a single copy were applied to determine RT-LAMP sensitivity limit. hMPV cDNA was obtained from viral nucleic acid samples using QIAGEN OneStep RT-PCR kit (Invitrogen) according to the manufacturer's instructions. We designed RT-PCR primers for hMPV N gene (forward: 5'-CAACATTCCGCAGAACCAGC -3' and reverse: 5'-GCCCATTTCACGCACCAGA -3'). Amplification was performed as follows: 50°C for 30 min, 95°C for 15 min, and then 40 cycles of 94°C for 30 s, at 57°C for 30 s, 72°C for 45 s, with a final 72°C extension for 10 min. Moreover, in order to evaluate the sensitivity of the RT-PCR assay, 457-nucleotide PCR product was extracted, cloned to pMD18-T vector (Takara), transcribed to RNA, and finally analyzed the sensitivity of RT-PCR. All the process and condition was similar to the sensitivity assessment for RT-LAMP described as above. To compare the availability of RT-PCR method and RT-LAMP method, 176 nasopharyngeal samples of the 176 children with acute respiratory tract infections were extracted for RNA using the Viral Nucleic Acid Extraction Kit (Geneaid) and were tested using visual RT-LAMP and RT-PCR methods developed as above. ACAGGAGTCTATTCATTGAGTATGGGAAAGCTTTAGGCTCATCTTCAACAGGAAGC TGTCCTCAGATAAGTAACTCATACCCTTTCGAAATCCGAGTAGAAGTTGTCCTTCG AAAGCAGAAAGTTTGTTTGTAAATATATTTATGCAAGCTTATGGAGCCGGTCAAAC TTTCGTCTTTCAAACAAACATTTATATAAATACGTTCGAATACCTCGGCCAGTTTG LF LB AATGCTAAGGTGGGGTGTCATTGCCAGATCATCTAACAACATAATGCTAGGGCATG TTACGATTCCACCCCACAGTAACGGTCTAGTAGATTGTTGTATTACGATCCCGTAC B1 Figure 4 hMPV N gene RT-LAMP primer design and target sequence location. Genomic DNA sequence of hMPV N gene (GeneBank Accession number: EF081360) from 659 to 875 nt in orientation 5' to 3'. The primers sequences are highlighted with Italic and bold, and the arrows show the primers direction from 5' to 3'. Osterhaus AD: A newly discovered human pneumovirus isolated from young children with respiratory tract disease Australian find suggests worldwide reach for metapneumovirus Evidence of human metapneumovirus in Australian children Characterization of human metapneumoviruses isolated from patients in North America Acute respiratory metapneumovirus infection in children under six years old in Beijing Human metapneumovirus infection in the Canadian population Human metapneumovirus detection in patients with severe acute respiratory syndrome Human metapneumovirus infection in the United States: clinical manifestations associated with a newly emerging respiratory infection in children Children with respiratory disease associated with metapneumovirus in Hong Kong Comparison of different cell lines and incubation times in the isolation by the shell vial culture of human metapneumovirus from pediatric respiratory samples Loop-mediated isothermal amplification of DNA A onestep reverse transcription loop-mediated isothermal amplification for detection and discrimination of infectious bursal disease virus Rapid and real-time detection of hepatitis A virus by reverse transcription loop-mediated isothermal amplification assay Real-time turbidimetry of LAMP reaction for quantifying template DNA Development and evaluation of realtime loop-mediated isothermal amplification for hepatitis B virus DNA quantification: a new tool for HBV management Rapid detection of HIV-1 by reversetranscription, loop-mediated isothermal amplification (RT-LAMP) Visual detection of turkey coronavirus RNA in tissues and feces by reverse-transcription loop-mediated isothermal amplification (RT-LAMP) with hydroxynaphthol blue dye Colorimetric detection of loop-mediated isothermal amplification reaction by using hydroxy naphthol blue Visual detection of pandemic influenza A H1N1 virus 2009 by reverse-transcription loopmediated isothermal amplification with hydroxynaphthol blue dye Efficacy of loop mediated isothermal amplification (LAMP) assay for the laboratory identification of mycobacterium tuberculosis isolates in a resource limited setting Udhayakumar V: Real-time fluorescence loop mediated isothermal amplification for the diagnosis of malaria LAMP for human African trypanosomiasis: a comparative study of detection formats Development and application of a loop-mediated isothermal amplification method on rapid detection Escherichia coli O157 strains from food samples Submit your next manuscript to BioMed Central and take full advantage of: • Convenient online submission • Thorough peer review • No space constraints or color figure charges • Immediate publication on acceptance • Inclusion in PubMed, CAS, Scopus and Google Scholar • Research which is freely available for redistribution The authors declare that they have no competing interests.Authors' contributions XW and LSZ designed this study; XW, QZ, FZ, WZZ, FLM performed this study; YLB and ZHZ collected the clinical specimens; WX and QZ wrote the paper; QZ, YLB and LSZ revised the manuscript critically. All authors read and approved the final manuscript.