key: cord-1016739-k12s0c8h authors: Cuong, Hoang Quoc; Hai, Nguyen Duc; Linh, Hoang Thuy; Hieu, Nguyen Trung; Anh, Nguyen Hoang; Ton, Tran; Dong, Tran Cat; Thao, Vu Thanh; Tuoi, Do Thi Hong; Tuan, Nguyen Duc; Loan, Huynh Thi Kim; Long, Nguyen Thanh; Thang, Cao Minh; Thao, Nguyen Thi Thanh; Lan, Phan Trong title: The Production of Standardized Samples with Known Concentrations for Severe Acute Respiratory Syndrome Coronavirus 2 RT-qPCR Testing Validation for Developing Countries in the Period of the Pandemic Era date: 2021-08-03 journal: Biomed Res Int DOI: 10.1155/2021/5516344 sha: 0e814eae182ce399b26324eec2d0dc24b6d1a59a doc_id: 1016739 cord_uid: k12s0c8h BACKGROUND: Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) has caused a pandemic of pneumonia spreading around the world, leading to serious threats to public health and attracting enormous attention. There is an urgent need for sensitive diagnostic testing implementation to control and manage SARS-CoV-2 in public health laboratories. The quantitative reverse transcription PCR (RT-qPCR) assay is the gold standard method, but the sensitivity and specificity of SARS-CoV-2 testing are dependent on a number of factors. METHODS: We synthesized RNA based on the genes published to estimate the concentration of inactivated virus samples in a biosafety level 3 laboratory. The limit of detection (LOD), linearity, accuracy, and precision were evaluated according to the bioanalytical method validation guidelines. RESULTS: We found that the LOD reached around 3 copies/reaction. Furthermore, intra-assay precision, accuracy, and linearity met the accepted criterion with an RSD for copies of less than 25%, and linear regression met the accepted R(2) of 0.98. CONCLUSIONS: We suggest that synthesized RNA based on the database of the NCBI gene bank for estimating the concentration of inactivated virus samples provides a potential opportunity for reliable testing to diagnose coronavirus disease 2019 (COVID-19) as well as limit the spread of the disease. This method may be relatively quick and inexpensive, and it may be useful for developing countries during the pandemic era. In the long term, it is also applicable for evaluation, verification, validation, and external quality assessment. A novel coronavirus that belongs to the Coronaviridae family has caused an outbreak of pneumonia spreading around the world, leading to serious threats to public health and attracting enormous attention [1, 2] . While waiting for the distribution of a vaccine across countries, in particular developing countries, or for the approval of new medicines, quantitative reverse transcription (RT-qPCR) is a key solution to combating this pandemic. Early detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) may be of importance for halting the spread of this disease in the community. Most diagnostic assays being applied for the diagnosis of coronavirus disease 2019 (COVID- 19) infections involve the RT-qPCR assay [3, 4] , which is obligatory, especially in the treatment and isolation of early-infected patients [5] . Furthermore, there is an urgent need for sensitive diagnostic testing implementation to control and manage SARS-CoV-2 in public health laboratories [6] . Up to now, the World Health Organization (WHO) has listed only 17 countries that have developed laboratories and protocols for the detection of SARS-CoV-2 [7] . Besides, WHO endorsed nations who have no testing capacity, and national COVID-19 laboratories with inadequate experience of SARS-CoV-2 testing are stimulated to send specimens to the WHO reference laboratory for confirmation [7] . Although a number of approaches are available to detect the SARS-CoV-2 viral genome, the sensitivity and specificity of SARS-CoV-2 testing depend on a variety of variables, such as the position of clinical specimens, the low viral load of the patient, the intermittent shedding, and the difference between multiple detection kits by producers [8, 9] . The majority of COVID-19 symptoms resemble those of the common flu or the cold. It is therefore critical that infected individuals be diagnosed early and accurately if this fatal disease is to be prevented from spreading widely. The identification in the early stages of the COVID-19 patients allows doctors to assist before serious complications have developed [10] . Furthermore, with a scarcity of chemical reagents for SARS-CoV-2 diagnosis and an increase in SARS-CoV-2 cases, the development of standardized specimens with known concentrations for validation is critical [11, 12] . Here, we synthesized RNA based on the genes published to estimate the concentration of inactivated virus samples for screening programs for COVID-19. 2.1. Sequence of the SARS-CoV-2 Virus and Synthesis of Artificial DNA. In this study, the sequence of the SARS-CoV-2 virus was used based on the database of the NCBI gene bank (GenBank NC_045512.2). We used the primer and probe sequences that were published in the previous study [13] . SARS-CoV-2's E gene from nucleotide no. 26132 to 26529 was as follows: ACACAATCGACGGT TCATCCGGAGTTGTTAATCCAGTAATGGAACCAATT TATGATGAACCGACGACGACTACTAGCGTGCCTTTG TAAGCACAAGCTGATGAGTACGAACTTATGTACTCA TTCGTTTCGGAAGAGACAGGTACGTTAATAGTTAAT AGCGTACTTCTTTTTCTTGCTTTCGTGGTATTCTTGC TAGTTACACTAGCCATCCTTACTGCGCTTCGATTGT GTGCGTACTGCTGCAATATTGTTAACGTGAGTCTTG TAAAACCTTCTTTTTACGTTTACTCTCGTGTTAAAA ATCTGAATTCTTCTAGAGTTCCTGATCTTCTGGTCT AAACGAACTAAATATTATATTAGTTTTTCTGTTTGG AACTTTAATTTTAGCCATGGCAG. DNA is synthesized in vitro by assembling singlestranded oligonucleotide fragments of 60-70 nucleotide length by the catalysis of Phusion™ High-Fidelity DNA Polymerase (2 U/μL) (Thermo Fisher Scientific) using the PCR technique. After assembly, product inspection was performed by 2% gel agarose electrophoresis. After assembly, the size of the DNA was determined by electrophoresis on 2% agarose gel and then transformed into the pJET1.2 plasmid (Phu Sa Corp), and the transformation was screened. The transformed plasmid was extracted and sequenced by using the E gene synthesized in vitro. The sequencing results showed that the E gene was synthesized to have the same sequence as the originally designed E gene. We used the MAFFT online server with the default parameters to align the SARS-CoV-2 genome sequences [14] , and the complete genome sequence from the NCBI gene bank (GenBank NC_045512.2) was used as a reference genome. The result of aligning the E gene sequence with the E gene sequence on GenBank showed that the two sequences are similar to each other, in which query covers 100% and has an E value of 0.0 ( Figure 1 ). E gene sequencing data is illustrated in Figure 2 . 2.2. Preparation of DNA-Carrying Plasmids (Recombinant DNA). The synthesized length of the gene fragments was connected to the pJET 1.2 plasmids using the cloning system (Phu Sa Corp). The recombinant plasmid was inserted into Escherichia coli (MAX Efficiency DH5α (Life Technologies, Carlsbad, CA; chemically competent,~10 9 colony-forming unit or CFU/μg pUC19)) on Luria-Bertani agar plates (Thermo Fisher Scientific, consisting of 10 g L −1 tryptone (peptone from casein), 5 g L −1 yeast extract, and 5 g L −1 sodium chloride) containing 100 μg/mL ampicillin. Five colonies on a pJET 1.2 plasmid were chosen for checking the transformation process by PCR with specific primers (pJET1.2, Phu Sa Corp; pJET1.2.Fw (5 ′ -d(CGACTCACT ATAGGGAGAGCGGC)-3 ′ ) and pJET1.2.Rv primers (5 ′ -d(AAGAACATCGATTTTCCATGGCAG)-3 ′ ). The modified vector-carrying strains were grown on Luria-Bertani broth overnight and recovered with the FavorPrep Plasmid Extraction Mini Kit (Favorgen, Biotech Corp) according to the manufacturer's recommended procedures. The transformation results were confirmed by genetic sequencing with specific primers and plasmid pJET1.2 primers, using the Sanger sequencing technique on the Applied Biosystems 3130 device. 2.3. RNA Synthesis In Vitro. The recombinant plasmid was extracted and straight-lined with 2 restriction enzymes NotI (NotI (10 U/μL), Thermo Fisher Scientific) and Kpn2I (BspEI (10 U/μL), Thermo Fisher Scientific) following the manufacturer's instructions. Straight-line plasmid fragments after treatment with a restriction enzyme were transcribed into RNA by the catalysis of the enzyme T7 RNA polymerase (Ambion™ T7 RNA Polymerase, cloned, 200 U/μL, Thermo Fisher Scientific) following the instructions of the manufacturers. The RNA was purified and preserved in diethyl-pyrocarbonate water. The RNA was then tested on a 1.2% agarose gel electrophoresis system using a 0.5X TBE buffer. The quantity of pure RNA product (ng/μL) was determined in triplicate on a NanoDrop by measuring absorbance at 260 nm with a spectrophotometer (NanoDrop™ 2000/2000c Spectrophotometer, Thermo Fisher Scientific). Based on the size and nature of the target RNA structure, the yield for each control was calculated using the Avogadro conversion factor (6, 022 × 10 23 ): number of copies = ðweight ðngÞ × 6, 022 × 10 23 Þ/ðlength × 330 × 10 9 Þ. [15] . In brief, these specimens were placed in a viral transport medium and refrozen; then, the virus was subsequently grown in Vero E6 cells (ATCC#C1008) in the Pasteur Institute in Ho Chi Minh City, Vietnam. Vero E6 cells were cultured in Dulbecco's minimal essential medium (DMEM, D1145, Sigma-Aldrich, US) supplemented with 10% heatinactivated fetal bovine serum (FBS, Gibco, Thermo Fisher, US) and antibiotics [16] . On the isolation day, 50 μL serum-free DMEM was added into columns 2-12 of a 96well culture plate. 100 μL clinical specimens were pipetted into column 1 and then serially diluted 2-fold across the plate. After trypsinization (Gibco Trypsin-EDTA, Thermo Fisher, US), Vero E6 cells were suspended in DMEM supplemented with 10% FBS, 200 IU/mL penicillin-streptomycin (Sigma-Aldrich, US), and 5 μg/mL amphotericin B (Sigma-Aldrich, US) with 2:5 × 10 5 cells/mL. 100 μL of cell suspension was directly added to the clinical swab dilutions and gently mixed by pipetting. The inoculated cultures were grown in a humidified atmosphere of 5% CO 2 at 37°C. The cytopathic effects (CPE) were observed in cells after incubation for three days. Standardized plaque assays were applied to SARS-CoV-2 according to the SARS-CoV and MERS-CoV protocols [17, 18] . The SARS-CoV-2 strain, SARS-CoV-2/human/VNM/n-CoV-19-02S/2020, was used in this study, and the genome sequence of the strain was deposited in GenBank (MT192773.1). In brief, the genome sequence length was 29,890 bp with no gaps and a high coverage of 1,897x. This strain was of the type betacoronavirus B and has 99.98 percent nucleotide-level similarity isolated in Wuhan and >90.56 percent pangolin-isolated SARS-CoV similarity. Four mutations have been identified as nonsynonymous mutations, such as G8388A (asparagine serine), A8987 T (phenylalanine isoleucine), and A8987 T (phenylalanine isoleucine) [9, 19] . 2.5. RT-qPCR for SARS-CoV-2. RNA synthesis was placed into a collection tube with 150 μL of virus preservation solution. In brief, 40 μL of cell lysate was transferred into a collection tube, followed by a vortex for 10 seconds. After standing at room temperature for 10 minutes, the collection tube was centrifuged at 1000 rpm/min for 5 minutes. The suspension was used for a quantitative reverse transcription PCR (RT-qPCR) assay of 2019-nCoV RNA. The E target gene was amplified and tested during the RT-qPCR assay with the forward primer 5 ′ -ACAGGTACGTTAA TAGTTAATA GCGT-3 ′ and reverse primer 5′-ATATTGCAGCAGTACGCACACA-3′ and the probe 5-′FAM ACACT AGCCATCCTTACTGCGCTT CGBBQ-3′. All oligonucleotides were synthesized and provided by TIB-Molbiol (Berlin, Germany). RT-qPCR was performed in triplicate on the LightCycler® 480 System using a 2019-nCoV nucleic acid detection kit according to the manufacturer's protocol (TIB-Molbiol reagents). The reaction mixture contained 5 μL of RNA, 12.5 μL of 2× reaction buffer provided with the Superscript III one-step RT-qPCR system with Platinum Taq Polymerase (Invitrogen, Darmstadt, Germany; 0.4 mM of each deoxynucleotide triphosphate (dNTP) and 3.2 mM magnesium sulfate), 1 μL of reverse transcriptase/Taq mixture from the kit, and 0.4 μL of a 50 mM magnesium sulfate solution (Invitrogen) [13] . RT-qPCR was performed under the following conditions: incubation at 55°C for 03 minutes and 95°C for 30 seconds, 45 cycles of denaturation at 95°C for 03 seconds, and extending and collecting the fluorescence signal at 60°C for 12 seconds. A cycle threshold value (Ct value) of 40 or more was defined as a negative test [9, 20] . (Table 1) . The artificial RNA, after determining the number of copies, is diluted to the appropriate concentration to investigate the detection limit, accuracy, and precision. Concentration of Synthesized RNA. In the present study, we tested a dilution series of nine replicates for synthesized RNA per concentration. The Ct value ranged from 22.4 to 23.69, and the number of copies was 48,300 copies/μL at the level of 10 4 dilutions. When dilution tends to be small, the Ct values are more likely to increase. There was quite consistent fluorescence at 22, 24, and 28 cycles per concentration ( Figure 5 ). Concentration of Synthesized RNA 3.3.1. Repeatability. Intra-assay precision was determined from assay results in each concentration run. The RSD for log copies of each concentration was 0.77-8.33%, which reached the accepted criterion of RSD < 25% (Table 2 ) [23, 24] . Similarly, intra-assay precision was performed from assay results on four different runs and on three different days. The RSD for log copies of each concentration was 2.80-13.30%, which meets the accepted criterion of an RSD of less than 25% (Table 3 ) [23, 24] . Synthesized RNA. Because the concentration of synthesized RNA was 48,300 copies/μL, we diluted it to 80% and 120% of the original concentration. RT-qPCR was then repeated seven times on different days. The percentage of RSD for log copies was 1.01-2.70% with an RSD of <25% [23] . The results also exhibited that the synthesized RNA specimens had a good extraction efficiency (Tables 4 and 5 ). In this study, the linearity was created based on the Ct value versus log copy. The linear regression was obtained from five independent assays performed on different days. We estimated the linear regression equation with Y = −3:68 x log copy/μL + 40:02. The R 2 of the standard linear equation reached 0.998, which meets the accepted criterion of R 2 > 0:98 ( Figure 6 ) [23, 24] . RT-qPCR was used in a dilution series of ten replicates run on three different days to estimate the Ct values based on the standard curves above to estimate the concentration of RNA extracted from the SARS-CoV-2 virus isolated from Vero E6 cells. The percentage of RSD for log copies was 0.54-14.61% with an RSD for log copies of <25% (Table 6 ) [23, 24] . 3.6. Limit of Detection (LOD). By establishing a standard curve, linearity was also confirmed and the LOD was calculated (~3 copies/reaction). In this study, intra-assay precision, accuracy, and linearity met the acceptance criteria with an RSD for log copies of less than 25% ( The current study reported the process of developing standardized specimens of known concentration, which were RNA extracted from the SARS-CoV-2 virus isolated, for evaluation, verification, validation, and external quality assessment. Standardized specimens are produced according to the international conference on harmonization guidelines for bioanalytical method validation [21] [22] [23] 25] . After creating the standard curve, RNA extracted from the inactivated virus was processed in a RT-qPCR system to estimate the standardized specimens at a known concentration. We found that the LOD reached 3 copies/reaction, and intra-assay precision, accuracy, and linearity met the accepted criterion with RSD for log copies of less than 25% [23, 24] . Besides, linear regression meets the accepted R 2 of 0.98 [21] [22] [23] [24] . It could help to assess the reliability of SARS-CoV-2 testing in order to improve the testing capacity of COVID-19 screening strategies [9] . This method may be relatively quick and inexpensive, making it useful for developing countries during the pandemic. The SARS-CoV-2 (E, N, and RdRP gene detection) test is a high-throughput RT-qPCR technique for the qualitative identification of SARS-CoV-2 nucleic acid in nasopharyngeal, nasal, and oropharyngeal swab samples from subjects suspected of COVID-19 [10] . In the current tests, the primer and probe sets are designed to detect three regions of the SARS-CoV-2 single-stranded RNA genome [10] . More specifically, the RdRP (RNA-dependent RNA polymerase) and N genes (a nucleocapsid protein) are SARS-CoV-2 specific, whereas the E gene (an envelope protein) is Sarbecovirus specific [10, 26] . Chu et al. suggested the ORF1b gene for [13] . Although our study only developed standardized specimens of known concentration based on the E gene, developing quick and accurate COVID-19 screening methods will also assist in identifying negative people and avoiding unnecessary COVID-19 quarantines, which have had a severe influence on social life and resulted in a significant economic crisis [10] . To detect the SARS-CoV-2 viral genome, a variety of approaches are available, including PCR-based SARS-CoV-2 detection (reverse transcription-quantitative PCR, reverse tran-scription digital PCR, and current isothermal amplification methods) and nonconventional methods (genome sequencing, clustered regularly interspaced short palindromic repeatbased COVID-19 detection, and nanoparticles). Of note, the sensitivity of the droplet digital PCR was found to be equal to or greater than that of the RT-qPCR [28, 29] . These approaches, however, are not equivalent to RT-qPCR tests in terms of cost, sensitivity, or specificity [30] . Until now, the gold standard technique is the molecular diagnosis of SARS-CoV-2 using the RT-qPCR assay [13, 31] . However, RT-qPCR assays depend on the similarity of SARS-CoV-2 to SARS-CoV, collecting time and location of the specimens [32] . Consequently, false-positive RT-qPCR results have been reported in the SARS-CoV-2 diagnosis output in recovery patients and asymptomatic infected patients [33] . RT-qPCR assays necessitate the use of relatively expensive instruments as well as highly trained personnel. Thus, these provisions limit diagnostic capacity expansion in several countries [30] . Taken together, the validation of rapid diagnostic tests for COVID-19 should be a priority for diagnosis and follow-up of patients both in the hospital and in the community, allowing us to detect cases early and isolate patients and close contacts rapidly. Access to reliable rapid diagnostic tests could improve the pressure on laboratories and enlarge the testing capacity to meet the most urgent medical and public health needs [13] . The SARS-CoV-2 pandemic posed a serious threat to human health. However, shortages of chemical reagents and healthcare workers are restraining testing capacity from the growing demand for COVID-19 diagnostics around the world, especially in resource-limited settings [9, 13] . However, strategies from WHO emphasized the vital role of early testing of suspected cases to halt virus spread and emphasized that the need for reliable assays to detect and laboratory confirm cases early [34] is therefore essential to evaluate the reliability of the RT-qPCR testing. While waiting for a vaccine to be distributed to every country, especially developing countries, or medicine to be approved, RT-qPCR is an essential solution in the fight against this pandemic. Standardized specimens of known concentration will contribute to improving the testing capacity and support screening, earlier diagnosis of infection, and isolation. In the long-term, standardized specimens of known concentrations are capable of evaluation, verification, validation, and external quality assessment. In the current study, we developed standardized specimens and followed the guidelines on bioanalytical method validation [21] [22] [23] 25] . Our study has several limitations. First, we only created standardized specimens of known concentration using the E gene, and our findings were unable to achieve a broad dynamic range [27] . Second, we only identify the purity of RNA based on UV absorbance, which is not strong enough for the quantification of the results, because results may be influenced by the impurity of the samples. However, RNA is pure (the ratio A260/A280 was 2.1). In summary, we synthesized RNA from the database of the NCBI gene bank to estimate the concentration of inactivated virus samples and provided a potential opportunity for reliable testing to diagnose COVID-19. From there, standardized specimens of known concentrations contribute to improving the testing capacity and supporting early diagnosis of infection as well as limiting the spread of the disease. This method may be relatively quick and inexpensive, and it may be useful for developing countries during the pandemic era. In the long term, standardized specimens of known concentrations are applicable for evaluation, verification, validation, and external quality assessment. Limit of detection RT-qPCR: Quantitative reverse transcription PCR SARS-CoV-2: Severe acute respiratory syndrome coronavirus 2 WHO: World Health Organization. The data used to support the findings of this study are available from the corresponding author upon request. The authors have no conflicts of interest. Severe respiratory illness caused by a novel coronavirus The epidemic of 2019-novelcoronavirus (2019-nCoV) pneumonia and insights for emerging infectious diseases in the future Development of a molecular-beacon-based multi-allelic real-time RT-PCR assay for the detection of human coronavirus causing severe acute respiratory syndrome (SARS-CoV): a general methodology for detecting rapidly mutating viruses A meta-analysis to evaluate the effectiveness of real-time PCR for diagnosing novel coronavirus infections Hospitalassociated outbreak of Middle East respiratory syndrome coronavirus: a serologic, epidemiologic, and clinical description Rapid establishment of laboratory diagnostics for the novel coronavirus SARS-CoV-2 in Bavaria World Health Organization, Laboratory testing for coronavirus disease 2019 (COVID-19) in suspected human cases: interim guidance Coronavirus disease 2019 (COVID-19): situation report Comparison of primer-probe sets among different master mixes for laboratory screening of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) Diagnostic techniques for COVID-19 and new developments Evaluation of WHO listed COVID-19 qPCR primers and probe in silico with 375 SERS-CoV-2 full genome sequences Comparative performance of SARS-CoV-2 detection assays using seven different primer/probe sets and one assay kit Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR Genome-wide analysis of SARS-CoV-2 virus strains circulating worldwide implicates heterogeneity Importation and human-to-human transmission of a novel coronavirus in Vietnam Isolation and characterization of SARS-CoV-2 from the first US COVID-19 patient Release of severe acute respiratory syndrome coronavirus nuclear import block enhances host transcription in human lung cells Cell host response to infection with novel human coronavirus EMC predicts potential antivirals and important differences with SARS coronavirus Clinical features, isolation, and complete genome sequence of severe acute respiratory syndrome coronavirus 2 from the first two patients in Vietnam Relationship of the cycle threshold values of SARS-CoV-2 polymerase chain reaction and total severity score of computerized tomography in patients with COVID 19 Committee for Medicinal Products for Human Use, Guideline on bioanalytical method validation Department of Health and Human Services, Food and Drug Administration Guidelines for validation of qualitative real-time PCR methods Verification of analytical methods for GMO testing when implementing interlaboratory validated methods Verification and validation of diagnostic laboratory tests in clinical virology Improved sensitivity using a dual target, E and RdRp assay for the diagnosis of SARS-CoV-2 infection: experience at a large NHS Foundation Trust in the UK Molecular diagnosis of a novel coronavirus (2019-nCoV) causing an outbreak of pneumonia Digital droplet PCR accurately quantifies SARS-CoV-2 viral load from crude lysate without nucleic acid purification Comparison of digital PCR and quantitative PCR with various SARS-CoV-2 primerprobe sets Nucleic acid testing of SARS-CoV-2 Identification of a novel coronavirus in patients with severe acute respiratory syndrome Comparative analysis of primer-probe sets for the laboratory confirmation of SARS-CoV-2 Risk-adapted treatment strategy for COVID-19 patients Laboratory readiness and response for novel coronavirus (2019-nCoV) in expert laboratories in 30 EU/EEA countries We would like to thank Dr. Ho Huu Tho from a military medical institution and colleagues at the Pasteur Institute in Ho Chi Minh City for supporting the study.