key: cord-1013406-7pwu2k5n authors: Freed, Nikki E; Vlková, Markéta; Faisal, Muhammad B; Silander, Olin K title: Rapid and Inexpensive Whole-Genome Sequencing of SARS-CoV-2 using 1200 bp Tiled Amplicons and Oxford Nanopore Rapid Barcoding date: 2020-07-18 journal: Biol Methods Protoc DOI: 10.1093/biomethods/bpaa014 sha: ea41abf1e8f3b436d814b3d673a3d785b1757231 doc_id: 1013406 cord_uid: 7pwu2k5n Rapid and cost-efficient whole-genome sequencing of SARS-CoV-2, the virus that causes COVID-19, is critical for understanding viral transmission dynamics. Here we show that using a new multiplexed set of primers in conjunction with the Oxford Nanopore Rapid Barcode library kit allows for faster, simpler, and less expensive SARS-CoV-2 genome sequencing. This primer set results in amplicons that exhibit lower levels of variation in coverage compared to other commonly used primer sets. Using five SARS-CoV-2 patient samples with C(q) values between 20 and 31, we show that high-quality genomes can be generated with as few as 10,000 reads (approximately 5 Mbp of sequence data). We also show that mis-classification of barcodes, which may be more likely when using the Oxford Nanopore Rapid Barcode library prep, is unlikely to cause problems in variant calling. This method reduces the time from RNA to genome sequence by more than half compared to the more standard ligation-based Oxford Nanopore library preparation method at considerably lower costs. Several approaches have been used to sequence the SARS-CoV-2 genome, including metagenomics [1] , sequence capture [2] , SISPA [3] , and multiplex PCR [2] [3] [4] [5] [6] , followed by next generation sequencing using either the Illumina or Oxford Nanopore platforms. Due to its simplicity and economy, using multiplexed PCR amplicons is perhaps the most common approach. This technique has been used to successfully sequence thousands of genomes over the first few months of the COVID-19 outbreak [7] . However, multiplexed amplicon sets often exhibit uneven amplification across the genome, with up to 100-fold differences in the concentration of different amplicons [4, 8] . In addition, the most common methods of library preparation for next-generation sequencing remain relatively expensive, even when samples are multiplexed. To alleviate these problems, here we describe an approach using multiplexed 1200 base pair (bp) tiled amplicons with the Oxford Nanopore Rapid Barcoding kit (SQK-RBK004). Briefly, two PCR 3 reactions are performed for each SARS-CoV-2 positive patient sample to be sequenced. One PCR reaction contains thirty primers that generate the odd numbered amplicons ('Pool 1'), while the second PCR reaction contains twenty eight primers that generate the even numbered amplicons ('Pool 2' ; Fig. 1 ). After PCR, the two amplicon pools are combined and can be used for a range of downstream sequencing approaches. Here we use the Oxford Nanopore Rapid barcoding kit which enables relatively easy library preparation of these amplicons to achieve rapid, simple, and cost-effective sequencing. Five de-identified samples that had been assessed as positive by other methods were obtained from the New Zealand Auckland District Health Board. 80 µl of Viral Transport Media that had previously stored a nasopharyngeal swab from a patient infected with SARS-CoV-2 were used for RNA isolation using the QIAamp Viral RNA Mini spin kit (Qiagen, Cat No./ID: 52904) according to manufacturer specifications, with the following modifications: sample volume was brought up from 80 µl to 140 µl using 1x PBS, and RNA was eluted using two elutions with 40 µl Buffer AVE for a final volume of approximately 80 µl. To determine C q values, we performed quantitative RT PCR. We used the 2x SensiFAST Probe No-ROX One-Step Mix (Bioline, Cat No./ID: BIO-76005), according to manufacturer specifications using 2.5 µl RNA template. Primers and probes were from the CDC 2019-nCoV CDC Assay (IDT, Cat No./ID: 10006606) targeting the N1 and N2 regions of the nucleocapsid phosphoprotein of SARS-CoV-2. NP targets the RNase P gene for detection of human nucleic acids and serves as an internal control for sample integrity. Reactions were run at a total volume of 20 µl on the Liberty16 mobile real-time PCR device (Ubiquitome). The results are shown in Table 2 . The detailed methods for Reverse Transcription, PCR, and library preparation have been posted to protocols.io: https://www.protocols.io/view/ncov-2019-sequencing-protocol-rapid-barcoding-1200-bgggjttw Below we briefly summarize the method and provide additional notes on the genome assembly. Two separate PCR reactions are done for each SARS-CoV2 positive sample. Two pools of primers are made: 'Pool 1' contains thirty primers that generate the odd numbered tiled amplicons, while 'Pool 2' contains twenty eight primers that generate the even numbered tiled amplicons for the 1200 bp set (Fig. 1) . This tiled approach is necessary to minimise overlap between amplicons, which would decrease PCR efficiency as overlapping amplicons would anneal to each other. After the PCR is completed, we combine the two pools. Basecalling, assembly, and variant calling We used Guppy Version 3.6.0 for basecalling and demultiplexing all runs. We used the ARTIC Network bioinformatics protocol for all genome assembly and variant calling steps with the 1200 bp amplicon sets (https://artic.network/ncov-2019/ncov2019-bioinformatics-sop.html), with the .bed and .tsv files adjusted accordingly to accommodate the different primer sequences and binding locations. We used 250 bp as the minimum read length cut-off and 1500 bp as the maximum length cut-off, which resulted in between 20% and 30% of the reads being filtered. The primer sequences and .bed and .tsv files necessary for the ARTIC variant calling pipeline are available on Zenodo https://zenodo.org/record/3897530#.Xuk7oGpLjep. To quantify the number of ambiguous bases in the genome assemblies, we excluded the first and last 180 bp of the genome, which are outside of any amplicon set or which contain primer binding sites, and are thus usually masked during the ARTIC bioinformatics pipeline. https://mc.manuscriptcentral.com/bmp To calculate how sequencing depth affected genome coverage for each amplicon set, we first calculated the total number of Mbp collected for each amplicon set. We then scaled that total to the required sequencing depth. For example, if a total of 40 Mbp of data was collected for one amplicon set, we calculated genome coverage for that amplicon set, and then multiplied the coverage values by the relevant factor (e.g. to calculate coverage for 20 Mbp of data, we multiplied by 0.5). This removes the complicating issue that different amplicon sets have different distributions of read lengths. For read subsampling we used seqtk [9] https://github.com/lh3/seqtk [9] . To simulate read contamination, we used seqtk to subsample using a random seed, and concatenated the subsampled read sets from two different samples, performing this for all possible pairwise combinations. We then used the ARTIC Network bioinformatics pipeline after this subsampling to call variants. The steps in the pipeline include filtering to exclude short reads and likely chimeric reads, which we performed as noted above. Our first goal was to select an amplicon set resulting in even coverage across the SARS-CoV-2 genome, to ensure that we could obtain high-quality, complete genomes with minimal sequencing depth. We tested four different tiled, multiplexed amplicon sets: the current ARTIC network 400 base pair V3 primer pool, available from Integrated DNA technology (IDT catalog number: 10006788), and three sets designed using Primal Scheme [10] . These sets generate amplicons that are approximately 1200 bp, 1500 bp, and 2000 bp in size (File S1). In all cases, primers were multiplexed into two pools ('Pool 1' and 'Pool 2'), creating a tiled amplification of the entire SARS-CoV2 genome (Fig. 1) . Our second goal was to reduce the time and cost of library preparation. To this end, we used the Oxford Nanopore Rapid Barcoding kit. This kit uses a transposase to attach barcodes and motor proteins to DNA molecules, with library preparation taking less than 25 minutes of hands on time and requiring minimal reagents (Table S1 ). This contrasts with the Oxford Nanopore Ligation Sequencing kit (LSK-109) that is most commonly used [11] , which requires over two hours of hands on time for library preparation, as well as several third-party reagents. In addition to using the Rapid Barcoding kit approach, we omit a bead-based cleanup step after PCR, further reducing the time and cost. We tested all four primer sets on two RT-qPCR positive patient samples, one with a cycle quantification value (C q ) of 20.3, and the second with a C q of 31.2 ( Table 1) , which is similar to the mean C q found in sputum, feces, and pharyngeal swabs, and considerably higher than that found in nasal swabs [12] . We found that all four primers sets produced adequate results. However, the 1200 bp set exhibited the least variability, especially in the high C q sample ( Fig. 1) . We quantified this variation by testing how many Mbp of read data was required to achieve 30X coverage at every position in the genome (here we use Mbp of sequencing data and not read number to mitigate the effects of differences in average read length between the samples). We found that for the 1200 bp and 2000 bp amplicon sets, 30X read coverage across the 99.9% of the genome could be achieved with only 3Mbp of data (Fig. 2) . In contrast, for the 400 bp amplicon set, genome coverage was far more variable, and for the high C q sample, even with 20Mbp of data, 30X coverage was achieved for only 99% of the genome. We thus selected the 1200bp primer set for additional experiments, as this amplicon size yielded the most consistent results. We next tested the reliability of the 1200 bp amplicon set with the Rapid Barcode kit library prep using three additional SARS-CoV-2 positive patient samples. Across these five samples, C q values ranged from 20 to 31 (Table 1) . We multiplexed all five samples on a single flow cell for four hours. Again we found relatively even genome coverage for all five samples, although for higher C q samples, there again appeared to be greater variability across the genome (Fig. 3) . 7 We used the ARTIC Network bioinformatics pipeline (see Methods) for genome assembly and variant calling for these samples, and in all cases obtained assemblies with fewer than 10 ambiguous base pairs ( Table 2) , excluding the 5' and 3' ends of the genome, which are either not part of an amplicon or which are primer sequences. We again quantified the sequencing depth (number of reads) required to achieve sufficient genome coverage for assembly and variant calling. We downsampled the read sets for each sample, varying the number of reads from 2,500 to 30,000 (approximately 1.25 Mbp to 15 Mbp, assuming a pre-filtered average read length of 500 bp ( Table 2 )), and quantified the fraction of the genome covered at varying depths (Fig. 4) . We found that with low C q samples, very few reads were required to achieve high coverage depth across the genome. For the lowest C q samples, only 10,000 reads (approximately 5 Mbp) were required to achieve 50X coverage depth over more than 99.9% of the genome. Finally, we quantified how sequencing depth affected genome assembly. Again, for low C q samples, only 10,000 reads were required to reach fewer than 20 ambiguous bases, and with 15,000 reads (approximately 7.5 Mbp), fewer than 10 ambiguous bases remained in the genomes for all samples (Fig. 5) . The method we use here relies on the Rapid Barcode library prep (RBK-004), which is transposon based, and usually results in only a single barcode being present on a read. This contrasts with the ligation-based library prep (LSK-109), for which most reads contain two barcodes (one at each end of the molecule). For the ligation-based library prep, stringent demultiplexing can be imposed that requires the same barcode to be present at both ends. However, when demultiplexing samples prepared using the Rapid Barcoding method, such stringent demultiplexing is not possible. Thus, there may be some crossover between samples with different barcodes due either to (1) residual transposase activity after samples are pooled; (2) low rates of ligation or chimaera formation between molecules (we observe between 0.05% and 0.1% of all reads as being longer than the longest amplicon, although many of these are human transcripts); or (3) mis-classification during the demultiplexing step performed by the basecaller and demultiplexer, Guppy. We note that the latter phenomenon is unlikely to result in high levels of barcode crossover: we find generally that fewer than 0.1% of all reads are assigned to barcodes that have not been used in any specific experiment. For this reason, we tested how read contamination affected variant calling. We first subsampled 20,000 reads from all samples and called variants. In all cases, we found that the variant calls matched those when using all reads (between 45,000 and 115,000 for all samples ( Table 2) ). In order to analyze the effects of read contamination, we first removed sample AKL-MU-2 (C q 20.6) from consideration, as the variants in this sample were identical to those in Thus, the level of read contamination varied from 0.2% to 50%. For all combinations of read sets and contaminant levels, we called variants using the ARTIC Network bioinformatic pipeline, and tested whether these matched the variant calls when using the uncontaminated read sets. In no cases did we find any false positive SNPs (variants that were called despite not being present in the uncontaminated original sample). Unsurprisingly, we found that as we increased contamination levels, variants were no longer reliably called. However, in all cases this only occurred at contamination levels greater than 5% (Fig. 6) . In many cases all expected variants were called even when read contamination levels were in excess of 25%. Here we have shown that with a 1200 bp multiplexed amplicon set and the Oxford Nanopore Rapid Barcode kit we can achieve rapid, simple, and inexpensive genome sequencing of SARS-CoV-2. The sequence data results in even genome coverage, allowing high-quality viral genomes to be assembled at 50X coverage with fewer than 20,000 reads, or approximately 10 Mbp of sequencing data, and often far less. Furthermore, we have shown that even when there is read contamination, we can successfully call variants at levels of cross-sample contamination exceeding 5%. This suggests that the increased level of barcode crossover that might result when using the Rapid Barcode library prep method should rarely result in incorrect variant calls. It will be critical to test this method on high C q samples, as it is likely that achieving even genome coverage will be more difficult in these cases. Here we have only tested samples up to C q 31. for the high C q sample. In the case of the 400 bp amplicon set, 99% genome coverage at 30X required 19 Mbp of sequence data, and 99.9% was only achieved with 33 Mbp of sequence data. is achieved with fewer than 20K reads. Insets show genome coverage levels at the top end of the y-axis (range from 0.995 to 1). Each line indicates the coverage for one sample. Insets show higher resolution at the upper limit of the y-axis. The colours of each sample on these plots are the same as those in Fig. 3 . Note that the scale of the y-axis in the top left plot differs from the others. Mbp of sequence data Fraction of bp with > 30X coverage q q q q q q q q q q q q q q q q q q q 400 bp C q 20 q q q q q q q q q q q q q q q q q q 400 bp C q 31 q q q q q q q q q q q q q q q q q q q 1200 bp C q 20 q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q 1500 bp C q 31 q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q 0K 10K 20K 30K q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q q qq q q q q qq q q q qqq q q q q q q q q q q q q q qqq q q qqq q q q qq q q q q q q q q q q q q q q qqq q q qq q q q q q qq q q q q q q q q q q q q q q qq q q q qq q q q qqq q q q q q q q q q q q q q qqq q q qqq q q q qq q q q q q q q q q q q q q q qqq q q qqq q q q q q q q q q q q q q q q q q q q qqq q q q qq q q q q qq q q q q q q q q q q q q q qqq q q qqq q q q q q q q q q q q q q q q q q q q qqq q q qqq q q q q qq q q q q q q q q q q q q q qqq q q q qq q q q q q q q q q q q q q q q q q q q q qq q q qqq q q q q qq Rapid metagenomic characterization of a case of imported COVID-19 in Cambodia Cost-Effective Tailed Amplicon Method for Sequencing SARS-CoV-2 Amplicon based MinION sequencing of SARS-CoV-2 and metagenomic characterisation of nasopharyngeal swabs from patients with COVID-19 A proposal of alternative primers for the ARTIC Network's multiplex PCR to improve coverage of SARS-CoV-2 genome sequencing SARS-CoV-2 genomes recovered by long amplicon tiling multiplex approach using nanopore sequencing and applicable to other sequencing platforms An emergent clade of SARS-CoV-2 linked to returned travellers from Iran 2017) Data, disease and diplomacy: GISAID's innovative contribution to global health Multiplex PCR method for MinION and Illumina sequencing of Zika and other virus genomes directly from clinical samples nCoV-2019 sequencing protocol v2 v1 (protocols.io.bdp7i5rn) Detection of SARS-CoV-2 in Different Types of Clinical Specimens