key: cord-1012158-l4h1uffp authors: Villalta, Danilo; Moratto, Anna; Salgarolo, Valeria; Re, Mirella Da; Giacomello, Roberto; Malipiero, Giacomo title: New-Generation Quantitative Immunoassays for SARS-CoV-2 Antibody Detection: Need for Harmonization date: 2022-01-01 journal: Ann Lab Med DOI: 10.3343/alm.2022.42.1.113 sha: 708df9c571f4c14ef0669e56cc0f7e040e7f037b doc_id: 1012158 cord_uid: l4h1uffp nan Dear Editor, Since March 2020, many qualitative or semi-quantitative serologic assays to detect antibodies against severe acute respiratory syndrome-coronavirus 2 (SARS-CoV-2) have become commercially available. Harristshøj, et al. [1] compared the results of 16 serological assays on a large number of patients and controls, showing differences in accuracy, despite the majority of automated assays for total (IgG+IgM) and IgG antibody detection reaching the predefined criteria for acceptable performance. Additionally, the antibody response appeared to be the strongest among patients with more severe disease [1] . Since quantitative determination of SARS-CoV-2 antibodies may be important for longitudinal monitoring of the antibody response in patients and in response to a vaccine, a new generation of quantitative immunoassays for SARS-CoV-2 antibody detection was recently launched [2] [3] [4] [5] [6] [7] . However, to date, few independent studies evaluating these assays have been published, and no studies have compared the quantitative data obtained with these novel immunoassays. We compared the results of four new-generation fully automated immunoassays for quantitative SARS-CoV-2 IgG antibody detection using a total of 123 serum samples, including 93 samples from patients diagnosed as having coronavirus disease 2019 (COVID-19; 44 with mild, 35 with severe, and 14 with critical disease) and 30 samples from patients with other acute viral diseases (Epstein-Barr virus, cytomegalovirus, influenza, adenovirus) as controls. Sera from patients with COVID-19 were collected between 19 and 30 days from symptom onset. All serum samples were collected at Pordenone Hospital (Italy) from March to June 2020. The study was approved by the Institutional Review Board of the Pordenone Hospital, and informed consent was obtained from the patients. Two of the assays use an antigen of the recombinant receptor-binding domain (RBD) of the spike (S) protein (sCOVG, Siemens, Tarrytow, NY, USA, and Maglumi SARS-CoV-2 S-RBD IgG, Snibe, Shenzhen, China), one uses a recombinant S1 protein (SARS-CoV-2-Sp1 IgG EliA, Thermo Fisher, Uppsala, Sweden), and one uses a recombinant trimeric protein (S1+S2+RBD) (Liaison SARS-CoV2 Trimeric S IgG, Diasorin, Saluggia, Italy). All samples were processed according to the manufacturers' instructions using an automated chemiluminescence immunoassay platform ( quantification were re-assayed after automated (manual for Diasorin) pre-dilution with the diluent provided by the manufacturer. Statistical analyses were performed using MedCalc software version 15.2.1 (Ostend, Belgium). Sensitivity was defined as the proportion of correctly identified patients with COVID-19 infection and specificity as 100 minus the proportion of control population found positive for SARS-Cov-2 antibodies. The Mann-Whitney U test was used for the comparison of continuous variables. P < 0.05 was considered statistically significant. Correlations between the different assays were calculated using correlation coefficient (r) and Passing-Bablok tests. Seven of the 93 samples from COVID-19 patients and all control samples showed a negative result with all four assays (sensitivity: 92.5%, 95% confidence interval [CI]: 85.1%-96.2%; specificity: 100%, 95% CI: 88.4%-100%; overall agreement: 100%). The seven COVID-19 patients who had a negative result from the antibody assays all showed mild symptoms. A significantly higher antibody level was detected in patients with severe or critical symptoms than in patients with mild symptoms (P ≤ 0.0002) (Fig. 1) . Quantitative comparison of SARS-CoV-2 antibody levels showed a very high correlation between the Siemens and Snibe assays (r = 0.920), between the Thermo Fisher and Siemens assays (r = 0.930), and between the Thermo Fisher and Snibe assays (r = 0.930). High correlations were observed between the Diasorin and Siemens assays (r = 0.871), between the Diasorin and Snibe assays (r = 0.797), and between the Diasorin and Thermo Fisher assays (r = 0.810). However, the Passing-Bablok regression test (Fig. 2) showed different absolute values among the assays and, with the exception of the Snibe vs. Siemens and Snibe vs. Thermo Fisher comparisons, also showed significant deviation from linearity. The relatively weak correlations between Diasorin and the other assays could be explained by the trimeric nature of the antigen used in this assay, likely detecting an antibody subset not detected by the other assays. In conclusion, with the limitation of the small sample size, perfect qualitative agreement of results was obtained among four new-generation immunoassays, whereas some differences in the quantitative results were found, indicating that commutabil- Comparison of 16 serological SARS-CoV-2 immunoassays in 16 clinical laboratories Serological assays for SARS-CoV-2 infectious disease: benefits, limitations and perspectives Quantitative measurement of anti-SARS-CoV-2 antibodies: analytical and clinical evaluation A novel highly quantitative and reproducible assay for the detection of anti-SARS-CoV-2 IgG and IgM antibodies Evaluation of the automated LIAISON SARS-CoV-2 TrimericS IgG assay for detection of circulating antibodies Clinical evaluation of the Abbott Alinity SARS-CoV-2 spike-spewww cific quantitative IgG and IgM assays among infected, recovered, and vaccinated groups Clinical validation of the Siemens quantitative SARS-CoV2 spike IgG assay (sCOVG) reveals improved sensitivity and a good correlation with virus neutralization titers