key: cord-1009156-fu0aw024 authors: Prisk, G. Kim; West, John B. title: Non-invasive Measurement of Pulmonary Gas Exchange Efficiency: The Oxygen Deficit date: 2021-10-21 journal: Front Physiol DOI: 10.3389/fphys.2021.757857 sha: d4f88bf961799550213714ca00b75e6c9eb04356 doc_id: 1009156 cord_uid: fu0aw024 The efficiency of pulmonary gas exchange has long been assessed using the alveolar-arterial difference in PO(2), the A-aDO(2), a construct developed by Richard Riley ~70years ago. However, this measurement is invasive (requiring an arterial blood sample), time consuming, expensive, uncomfortable for the patients, and as such not ideal for serial measurements. Recent advances in the technology now provide for portable and rapidly responding measurement of the PO(2) and PCO(2) in expired gas, which combined with the well-established measurement of arterial oxygen saturation via pulse oximetry (SpO(2)) make practical a non-invasive surrogate measurement of the A-aDO(2), the oxygen deficit. The oxygen deficit is the difference between the end-tidal PO(2) and the calculated arterial PO(2) derived from the SpO(2) and taking into account the PCO(2), also measured from end-tidal gas. The oxygen deficit shares the underlying basis of the measurement of gas exchange efficiency that the A-aDO(2) uses, and thus the two measurements are well-correlated (r(2)~0.72). Studies have shown that the new approach is sensitive and can detect the age-related decline in gas exchange efficiency associated with healthy aging. In patients with lung disease the oxygen deficit is greatly elevated compared to normal subjects. The portable and non-invasive nature of the approach suggests potential uses in first responders, in military applications, and in underserved areas. Further, the completely non-invasive and rapid nature of the measurement makes it ideally suited to serial measurements of acutely ill patients including those with COVID-19, allowing patients to be closely monitored if required. For the lung to exchange gas (O 2 from the inspired air into the blood, and CO 2 from the blood to the expired gas), alveolar gas and pulmonary capillary blood must be brought into close apposition across the thin alveolar-capillary membrane. Any degree of spatial mismatch between ventilation and perfusion [ventilation-perfusion (VA/Q) inequality] will lower the efficiency of the exchange of any gas, resulting in a difference between the partial pressure of a gas in the arterial blood leaving the lung, and gas in the exhaled breath (Rahn and Fenn, 1955; West, 1969 For O 2 and CO 2 , the dissociation curves that describe the content of the gas in blood as a function of partial pressure, are markedly different. The sigmoidal shaped O 2 dissociation curve rapidly flattens at higher values of PO 2 . Thus, the presence of any regions of the lung with a low VA/Q ratio will result in the addition of poorly oxygenated blood to the arterial circulation, but a compensatory increase in overall ventilation (from chemoreceptive responses) cannot add more oxygen to blood exiting regions of high VA/Q. In contrast, the quasilinear CO 2 dissociation curve means that low and high VA/Q regions can compensate for each other. Thus, it is common to see patients with pulmonary disease with arterial hypoxemia, while having a normal arterial PCO 2 (West and Luks, 2016) . A small increase in VA/Q inequality occurs with healthy aging (Cardus et al., 1997) , and increased VA/Q inequality is a hallmark of virtually all pulmonary diseases (Hopkins and Wagner, 2017) . Therefore, the measurement of the alveolar-arterial PO 2 difference (A-aDO 2 ) has long been a mainstay in assessing the disruption to pulmonary gas exchange caused by disease (Filley et al., 1954) . While conceptually simple, measuring the A-aDO 2 is technically challenging. Richard Riley first showed that the PO 2 in arterial blood could be measured by equilibrating a small bubble of air with the blood and measuring the PO 2 in the gas (Riley et al., 1957) . However, at the time, the technical difficulties of reliably sampling alveolar gas were overwhelming. To bypass this problem, Riley developed the construct of the "ideal alveolar PO 2 . " This was the alveolar PO 2 in the lung that would have been present if there was no ventilation-perfusion inequality, the PCO 2 in the alveolar gas was the same as that in arterial blood, and the respiratory exchange ratio was the same as that in the actual lung. The ideal alveolar PO 2 can be obtained from an arterial blood sample using the alveolar gas equation (Rahn and Fenn, 1955) : where A refers to alveolar, a to arterial, I to inspired, and R is the respiratory exchange ratio, the CO 2 production divided by the O 2 consumption (generally assumed to be 0.8 at rest). The final term in this equation is often ignored as it typically has a magnitude of only a few mmHg. A more detailed description of Riley's innovative approach can be found in West et al. (2020) . This approach provides a number representing the alveolar PO 2 , and it does so without actually measuring alveolar gas. Compact, rapidly responding gas analysis devices are now readily available, allowing direct measurement of expired PO 2 and PCO 2 . The Riley construct utilizes the ideal alveolar PO 2 to obviate the need to make a "technically difficult" measurement to calculate the A-aDO 2 . The oxygen deficit (OD) comes from a direct measurement of expired gas partial pressures and uses a non-invasive means to determine what would otherwise be an invasive measurement, the arterial PO 2 . The approach is to continuously measure expired O 2 and CO 2 while the patient breathes quietly on a mouthpiece while wearing a noseclip. The final concentrations measured just before the abrupt transition to inspired gas are taken as the end-tidal values of PO 2 and PCO 2 . An example of the expired gas record is shown in Figure 1 . The end-tidal values for partial pressure are a good reflection of the values within the alveolus (discussed in detail in reference West et al., 2020) and are highly reproducible. Previous work has shown that the breath-to-breath within-subject standard deviation of normal subjects breathing air is ~1.4 mmHg for PO 2 and ~0.7 mmHg for PCO 2 , with somewhat lower numbers when breathing a hypoxic gas. A trend plot of the last 30 values of these (covering 1-2 min) provides a direct indication of whether the patient is in steady-state, an important consideration since highly variable breathing would result in considerable variation in end-tidal partial pressures for both O 2 and CO 2 . Having directly measured the alveolar partial pressure for O 2 (and CO 2 ), the A-aDO 2 could be measured directly by taking an arterial blood sample to measure arterial PO 2 . However, to make the process both rapid and non-invasive, arterial PO 2 is estimated from the arterial oxygen saturation (SaO 2 ) as measured by pulse oximetry (SpO 2 ) as determined simultaneously using a fingertip pulse oximeter. This is then used to determine the corresponding arterial PO 2 from the Hill equation: where P 50 is the PO 2 at a saturation of 50%, and the SaO 2 is expressed as the fractional saturation; range [0, 1]. PO 2 is determined by taking the logarithm of the equation and solving algebraically. The term n (commonly referred to as the "Hill-n") is that required to match the sigmoidal shape of the O 2 -Hb dissociation curve, and a value of 2.7 is commonly used (Severinghaus, 2002; Prisk and West, 2019) . While a Hill-n of 2.7 provides an excellent fit to the experimentally determined values of saturation and PO 2 over the entire range of saturation (Severinghaus, 1966; Prisk and West, 2019) , in practice, only blood oxygen saturations in the range of 75-100% are likely to be encountered in patients. Over this limited range, an improved fit to the experimental data is achieved with a Hill-n = 2.88 . The P 50 of blood is normally ~27 mmHg, however this varies with alterations in PCO 2 , body temperature, base excess, and levels of 2,3-diphosphoglycerate (2,3-dpg; West and Luks, 2016) . Because the end-tidal gas partial pressure reflects the alveolar PCO 2 , the leftward or rightward shift of the O 2 -Hb dissociation curve from a PCO 2 different to the normal value of 40 mmHg (the Bohr effect) can be accounted for. Using the Kelman routines (Kelman, 1966 (Kelman, , 1968 . Note that in this patient there is a steep alveolar plateau for O 2 and CO 2 , indicative of marked heterogeneity in the lung. Below these are plots of end-tidal PO 2 (red) and PCO 2 (blue) over the preceding 150 s. The lower traces allow an assessment of whether the patient is in a steady state. This is also indicated by the Steady State indicator in the lower right of the screen. At the top of the screen are numerical values for the end-tidal partial presses of O 2 and CO 2 , the respiratory quotient (RQ), respiratory rate (RR), barometric pressure (PBar), inspired PO 2 (PIO 2 ), pulse rate (PR) and arterial oxygen saturation via pulse oximetry (SpO 2 ). In the center at the top is the calculated arterial PO 2 (termed gPaO 2 , red text), and the O 2 Deficit (the difference between end-tidal PO 2 and calculated arterial PO 2 ). (B) A screen summarizing the data from (A) without the graphical displays. The cartoon of the lung on the right shows the measured end-tidal PO 2 (115 mmHg in this example), and the calculated arterial PO 2 (gPaO 2 , 66 mmHg in this example), which together result in the oxygen deficit in traffic-light color-coded text (49 mmHg in this example). The operator may toggle between this screen and that in (A) as desired. assuming otherwise normal conditions for temperature, base excess and 2,3-dpg is determined (Prisk and West, 2019) . This is used to correct for changes in alveolar PCO 2 , assuming this is equal to arterial PCO 2 , an equivalence that has been long established (Comroe and Dripps, 1944) . Alterations in base excess, 2,3-dpg, or body temperature are not accounted for, because a blood sample is not obtained. The difference between the calculated arterial PO 2 and the measured end-tidal (alveolar) PO 2 is termed the OD. This can be thought of as a surrogate measurement of the A-aDO 2 . In the latter case the arterial value is measured, and the alveolar value estimated as described by Riley (above) , while in the case of the OD, the alveolar value is measured, and the arterial value estimated. This should not be confused with the "oxygen deficit" that provides a measure of the anaerobic contribution during exercise (Krogh and Lindhard, 1920; Medbo et al., 1988) . In the normal lung there is variation in the regional alveolar PO 2 , and this is often exaggerated in lung disease. The expired gas is a mixture of gas from all regions of the lung, just as the arterial blood is a mixture of blood from all regions of the lung. Further as gas exchange continues throughout expiration, PO 2 continues to fall. However, provided the end-tidal values are measured at functional residual capacity (FRC), the naturally occurring volume for end expiration at rest, the effect of ongoing gas exchange is minimal. Thus, the end tidal PO 2 is a direct and useful measurement of the alveolar PO 2 . The obvious challenge of determining the OD is the estimation of the arterial PO 2 from the SpO 2 given the shape of the O 2 -Hb dissociation curve which is very flat at higher values of PO 2 . At high values, even small errors in SpO 2 translate into large differences in the calculated PaO 2 . However, this problem becomes smaller at lower values of SpO 2 as the O 2 -Hb dissociation curve becomes steeper. A study addressing the likely errors in calculated PaO 2 showed that for SpO 2 values of 94% and below, the error in the calculated PaO 2 was less than 5 mmHg (Prisk and West, 2019) . Above a SpO 2 of 94% the calculation of PaO 2 was, as expected, unreliable. However, if SpO 2 is greater than 94% while breathing air at sea level, then no major gas exchange impairment exists, and so there is no need to measure OD. Because the approach considers the alveolar PCO 2 as well as the PO 2 , the left or right shift in the O 2 -Hb dissociation curve from changes in alveolar PCO 2 (the Bohr effect) can be directly accounted for. This effect is the principal cause of shifts in the O 2 -Hb dissociation curve, and so the ability to correct for this is critical. Failure to do so would result in errors in the OD of >5 mmHg for SpO 2 values of 94% (see figure 4 of Prisk and West, 2019 for details). Shifts in the O 2 -Hb dissociation curve from other causes (base excess, temperature, 2-3 dpg) are not taken into account with the non-invasive approach. These however, are much smaller, and produce only minor errors in the calculated OD (Prisk and West, 2019) . A Monte Carlo simulation of the typical simultaneously present errors in the measurements of SpO 2 and alveolar PCO 2 showed that the calculated OD had a slight negative bias (<5 mmHg) and typical variability of ~5 mmHg at an SpO 2 of 94%, with both of these values decreasing as SpO 2 fell, showing the viability of the approach (Prisk and West, 2019) . Initial studies in normal subjects were performed with the subjects breathing a hypoxic gas mixture (FIO 2 = 0.125) to ensure that the SpO 2 fell into the range in which OD could be reliably measured (SpO 2 < 95%). A study of 20 young subjects (19-31 years) and 11 older subjects (47-88 years) showed a very small OD in the young cohort (~2 mmHg), which was increased in the older (~8 mmHg; West et al., 2018b) . The increase in OD with increasing age is consistent with the wellknown increase in the A-aDO 2 with healthy aging (Raine and Bishop, 1963) . A more extensive subsequent study explored the effects of varying the inspired oxygen between the previously used hypoxic gas (FIO 2 = 0.125) up to and including breathing air (FIO 2 values of 0.15, 0.175, and 0.21; Liu et al., 2020) . This study again showed a higher OD in the older cohort compared to the young, with the difference persisting at all values of inspired oxygen, including air. Importantly, there was no statistical difference in the measured values of OD between and FIO 2 of 0.125 and 0.175, although OD rose in both cohorts when the subjects were breathing air. The result is consistent with an expected reduction in the A-aDO 2 as inspired PO 2 is lowered due to minimization of the effect of VA/Q inequality as the saturation falls and gas exchange occurs on the steeper and more linear portion of the O 2 -Hb dissociation curve. The intrasubject variability in the measured OD was large at high values of SpO 2 , and fell rapidly as SpO 2 was reduced below 94%, consistent with the simulation studies performed (Prisk and West, 2019) . The study showed that the OD was sensitive to the mild gas exchange impairment associated with healthy aging, even while breathing air, but that individual errors at high values of SpO 2 meant that the measurement was not likely to be useful in individual subjects at SpO 2 values above 94%. A recent study has also demonstrated the validity of the non-invasive approach to measure a gas exchange deficit. A direct comparison between OD and arterial blood gas (ABG) was performed in 25 normal subjects during hypoxic exercise, showing a correlation between OD and A-aDO 2 with an r 2 = 0.71, and with a small bias between the two, with OD being on average 5.2 ± 5.0 mmHg higher than A-aDO 2 (Howe et al., 2020) . The non-invasive nature of the measurement serves to enable measurements in conditions in where it would be challenging to perform ABGs, and in particular, serial ABGs. Recent examples are measurements performed in trained breath-hold divers before and after dives at an open-water dive site (Patrician et al., 2021a,b The studies showed a substantial but transient decrement in gas exchange efficiency as measured by increased OD, in some cases to nearly 70 mmHg. This was likely due to the development of pulmonary edema from the hydrostatically induced lung compression (lung-squeeze). A small initial study in a cohort of ambulant patients from a general pulmonary outpatient clinic with a variety of pulmonary diseases showed that the OD was greatly elevated in this group compared to normal, with an average OD of ~49 mmHg (West et al., 2018a) . When the OD was directly compared to the A-aDO 2 measured by the collection of an ABG in 23 patients with an SpO 2 of 95% or less, there was a high correlation (r 2 = 0.72; West et al., 2018c) . There were similar strength correlations between the calculated PaO 2 and that measured from the ABG, and between end-tidal PCO 2 and that from the ABG. The calculated PaO 2 was on average ~4 mmHg higher than that measured from the ABG. This study showed that the non-invasive approach provided a convenient, low cost, and accurate alternative to the use of an ABG to measure the magnitude of the gas exchange disruption in patients with pulmonary disease. A recent case report highlighted the use of the non-invasive approach in determining the underlying cause of a gas exchange defects in a patient in whom ABGs could not readily be obtained (Amaza et al., 2021) . This report served to highlight the potential of the non-invasive approach, and further showed that the approach was useful as a teaching tool. In the context of the ongoing SARS-CoV-2 pandemic, a small preliminary study investigated the usefulness of the non-invasive approach to measuring the impairment of pulmonary gas exchange in patients with suspected COVID-19 considered to be at risk of deterioration before obvious respiratory failure had ensued (McGuire et al., 2021) . Patients were either breathing air, or on supplemental low-dose oxygen, which was temporarily discontinued for ~10 min before measurements were taken. Of 13 patients studied, five were discharged home and the other eight were admitted based on physician decision using standard procedures. The OD was significantly greater in the patients who were admitted than in those who were discharged (OD 55 ± 20 vs. 32 ± 14 respectively, p = 0.041), suggesting that the measurement was a potentially useful means of assessing severity. Similarly, the oxygen deficit was significantly higher in the patients requiring supplemental O 2 than in those who did not (65 ± 9 vs. 30 ± 1 respectively, p < 0.001) again suggesting that the non-invasive measurement of gas exchange impairment provided useful clinical insight in a rapid non-invasive manner. The studies performed to date show that a non-invasive approach can be used to quantitatively assess the gas exchange deficit in patients with pulmonary disease. The approach taken is in many respects comparable to the traditional measurement of the A-aDO 2 first developed by Riley, and there is considerable physiological overlap in the measurements. This is shown in Figure 2 which shows the classic O 2 -CO 2 diagram and highlights the aspects of the VA/Q inequality that the two methods encompass. The traditional method of Riley focuses on the consequences of the presence of regions of low VA/Q that serve to add end-capillary blood with a low PO 2 to the arterial blood FIGURE 2 | Classic O 2 -CO 2 diagram with the ventilation-perfusion line joining the points for mixed venous blood and inspired gas. The traditional Riley analysis is based on the composition of arterial blood and ideal alveolar gas, and it is strongly biased by lung units with low ventilation-perfusion ratios that lie to the left of the ideal point (hatched area). By contrast, the new test also includes contributions from lung units with high ventilation-perfusion ratios that are located to the right of the ideal point. Modified from West et al. (2018a) . (so-called venous admixture), the result being arterial hypoxemia and an increased A-aDO 2 . The measurement of oxygen deficit encompasses the effects of low VA/Q and high VA/Q, although the effect of high VA/Q regions on arterial PO 2 is small because of the shape of the O 2 -Hb dissociation curve. The large overlap in the areas of influence of the two methods (Figure 2 ) means that the two measurements would be expected to be highly correlated, albeit not the same, and this was demonstrated with a strong correlation between measures of A-aDO 2 and OD (West et al., 2018c; Howe et al., 2020) . It is reasonable to question what additional information is gained by measuring the OD as opposed to simply measuring SpO 2 . While both VA/Q mismatch and shunt will serve to decrease arterial PO 2 (and thus SpO 2 ) and increase OD, so too will hypoventilation. Because the alveolar PCO 2 is also measured, hypoventilation can readily be detected which may provide an important clinical distinction of the cause of hypoxemia in some patients. Further, the OD takes into account the effect of changes in PCO 2 on the O 2 -Hb dissociation curve. Thus, the oxygen deficit directly addresses the efficiency of gas exchange, in the same way that the A-aDO 2 does. The important clinical measurement of the A-aDO 2 has been performed using an invasive approach for ~70 years. However, its use has become less common in recent years, likely due to the cost, time required for the measurement, and the uncomfortable and invasive nature of the procedure. In contrast, the OD is a rapid, non-invasive measurement that can be readily performed on patients ranging from ambulatory to those on mechanical ventilation. The measurement takes only a few minutes, requiring only that the patient breathe quietly on a mouthpiece while wearing a noseclip for ~2 min, while wearing a fingertip pulse oximeter. The equipment is portable, making it suitable for use not only in the hospital, but in the field, and in underserved areas. GP wrote, edited, and approved the article. JW edited and approved the article. All authors contributed to the article and approved the submitted version. FUNDING Support for some was provided by MediPines Corporation which provided AGM100 TM units for use in some of the studies referenced. Support was provided by MediPines Corporation to cover publication costs. Case studies in physiology: untangling the cause of hypoxemia in a patient with obesity with acute leukemia Increase in pulmonary ventilation-perfusion inequality with age in healthy individuals The oxygen tension of arterial blood and alveolar air in normal human subjects Alveolar and arterial oxygen tensions and the significance of the alveolar-arterial oxygen tension difference in normal men The Multiple Inert Gas Elimination Technique Validation of a noninvasive assessment of pulmonary gas exchange during exercise in hypoxia Calculation of certain indices of cardio-pulmonary function, using a digital computer Computer program for the production of O 2 -CO 2 diagrams The changes in respiration at the transition from work to rest Oxygen deficit is a sensitive measure of mild gas exchange impairment at inspired O 2 between 12.5% and 21% The use of the alveolar gas meter for assessment of gas exchange abnormalities in COVID-19 Anaerobic capacity determined by maximal accumulated O 2 deficit Case studies in physiology: breath-hold diving beyond 100 meters-cardiopulmonary responses in world-champion divers Temporal changes in pulmonary gas exchange efficiency when breath-hold diving below residual volume Deriving the arterial PO 2 and oxygen deficit from expired gas and pulse oximetry A Graphical Analysis of the Respiratory Gas Exchange. The O2-CO2 Diagram A-a difference in O 2 tension and physiological dead space in normal man A bubble method for estimation of PCO 2 and PO 2 in whole blood Blood gas calculator The invention and development of blood gas analysis apparatus Ventilation-perfusion inequality and overall gas exchange in computer models of the lung A new, noninvasive method of measuring impaired pulmonary gas exchange in lung disease: an outpatient study Measuring the efficiency of pulmonary gas exchange using expired gas instead of arterial Frontiers in Physiology | www Respiratory Physiology: The Essentials Measurements of pulmonary gas exchange efficiency using expired gas and oximetry: results in normal subjects Noninvasive measurement of pulmonary gas exchange: comparison with data from arterial blood gases The University of California, San Diego, has exclusively licensed technology to MediPines Corporation, Orange County, CA, to develop a device (the AGM100 TM ) used in some studies referenced in this work. JW declares a financial interest with MediPines Corporation.The remaining author declares that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.Publisher's Note: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.Copyright © 2021 Prisk and West. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.