key: cord-1009144-zikyb797 authors: Qian, Chungen; Zhou, Mi; Cheng, Fangming; Lin, Xiaotao; Gong, Yijun; Xie, Xiaobing; Li, Ping; Li, Zhiyong; Zhang, Pingan; Liu, Zejin; Hu, Fang; Wang, Yun; Li, Quan; Zhu, Yan; Duan, Guikai; Xing, Yinting; Song, Huanyu; Xu, Wenfang; Liu, Bi-Feng; Xia, Fuzhen title: Development and Multicenter Performance Evaluation of The First Fully Automated SARS-CoV-2 IgM and IgG Immunoassays date: 2020-04-21 journal: nan DOI: 10.1101/2020.04.16.20067231 sha: 169e190c1349941d1e62b5343fd90d998873a83c doc_id: 1009144 cord_uid: zikyb797 BACKGROUND: The outbreak of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) has rapidly spread globally. The laboratory diagnosis of SARS-CoV-2 infection has relied on nucleic acid tests. However, there are many limitations of nucleic acid tests, including low throughput and high rates of false negatives. More sensitive and accurate tests to effectively identify infected patients are needed. METHODS: This study has developed fully automated chemiluminescent immunoassays (CLIA) to determine IgM and IgG antibodies to SARS-CoV-2 in human serum. The assay performance has been evaluated at 10 hospitals. Clinical specificity was evaluated by measuring 972 hospitalized patients with diseases other than COVID-19, and 586 donors of a normal population. Clinical sensitivity was assessed on 503 confirmed cases of SARS-CoV-2 by RT-PCR and 52 suspected cases. RESULTS: The assays demonstrated satisfied assay precision with coefficient of variation (CV) of less than 4.45%. Inactivation of specimen does not affect assay measurement. SARS-CoV-2 IgM shows clinical specificity of 97.33% and 99.49% for hospitalized patients and normal population respectively. SARS-CoV-2 IgG shows clinical specificity of 97.43% and 99.15% for the hospitalized patients and the normal population respectively. SARS-CoV-2 IgM and IgG show clinical sensitivity of 85.88% and 96.62% respectively for confirmed SARS-Cov-2 infection with RT-PCR, of 73.08% and 86.54% respectively for suspected cases. CONCLUSIONS: we have developed fully automated immunoassays for detecting SARS-CoV-2 IgM and IgG antibodies in human serum. The assays demonstrated high clinical specificity and sensitivity, and add great value to nucleic acid testing in fighting against the global pandemic of the SARS-CoV-2 infection. Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) was initially identified in Wuhan, Hubei province, China, in December 2019, causing an ongoing outbreak of the novel coronavirus disease (COVID-19) (1, 2) . Although the epidemic in China has come under control through strict containment precautions over the past two months, SARS-CoV-2 has rapidly spread to more than 200 countries and regions (3). SARS-CoV-2 belongs to the same family of coronaviruses as severe acute respiratory syndrome (SARS-CoV) and Middle East respiratory syndrome (MERS-CoV), however, its transmission efficiency is much higher than SARS-CoV and MERS-CoV, and there is no indication that transmission will cease as the weather gets warmer. The World Health Organization (WHO) has announced COVID-19 is now a global pandemic, and that there is a long battle ahead to fight the virus (4). At present, there is no effective medication to treat the disease and it might take a year for a vaccine to be developed. The only way to control the outbreak of the virus is to identify and quarantine the infected individuals. COVID-19 is characterized by similar symptoms as the common cold such as fever, non-productive cough, fatigue, and featured chest CT patterns that may also cause fatal complications, as with severe acute respiratory syndrome (5, 6). The etiological diagnosis of SARS-CoV-2 infection relies on a nucleic acid test with Real Time-PCR (RT-PCR) using specimens collected via nasopharyngeal swab (7). However, high numbers of false negatives have been All rights reserved. No reuse allowed without permission. author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint (which was not peer-reviewed) is the . https://doi.org/10.1101/2020.04.16.20067231 doi: medRxiv preprint found in some laboratories, leading to a positive detection rate of RT-PCR around 50% of suspected clinical and epidemiological COVID-19 cases. There are many factors that can cause false negatives, and the specimen collection via nasopharyngeal swab could be the major challenge as the chance for the virus to move to the upper respiratory tract is smaller via non-productive cough. Degradation of the virus mRNA due to specimen inactivation at 56℃ is another cause for false negative result (8) . The limited detection efficacy of RT-PCR for identifying SARS-CoV-2 infection requires the need for a more sensitive test to effectively identify COVID-19 patients for a better control of the virus transmission and proper treatment of the disease. Detection of specific serum antibodies for SARS-CoV-2 could provide an alternative solution and compensate for the limitations of the nucleic acid test. It has been well established that IgM antibodies are generated and released into the blood soon after the viral infection, followed by IgG antibodies, therefore the detection of the specific antibodies in blood is a sensitive measurement for viral infection. It has been reported that the IgM antibodies can be detected in the blood from 7 days, and begins to decrease 10 days after the disease onset, while IgG antibodies can be detected [8] [9] days after the disease onset and continues to increase rapidly thereafter (9) . Commercial and non-commercial serological tests are currently under development. Based on the sequence of SARS-CoV-2 nucleic acid, we have constructed and expressed the viral nucleocapsid protein and spike protein. Using the purified All rights reserved. No reuse allowed without permission. author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint (which was not peer-reviewed) is the . https://doi.org/10.1101/2020.04.16.20067231 doi: medRxiv preprint recombinant antigen, we developed chemiluminescent immunoassays (CLIA) to determine the IgM and IgG antibodies to SARS-CoV-2 in human serum or plasma, which is performed automatically by an immunoassay analyzer. The performance of the assays has been evaluated at 10 hospitals, demonstrating high specificity and high sensitivity for SARS-CoV-2 detection. Being an effective way to compensate for the false negative issue of RT-PCR, the fully automated SARS-CoV-2 IgM and IgG immunoassays provided a powerful means of laboratory diagnosis for fighting against the global spread of the virus. Using pFastBac1 vector, the plasmids were constructed by inserting gene fragments for expressing the SARS-CoV-2 nucleocapsid protein and spike protein based on the published SARS-CoV-2 nucleic acid sequence on Genbank (MN908947.3). They were then transfected into the insect cell Sf9 to express SARS-CoV-2 fusion proteins, which were then purified by a combination of affinity chromatography and ion-exchange chromatography. Screening and verification for the specificity and sensitivity of the recombined SARS-CoV-2 fusion protein were performed using SARS-CoV-2 IgM and IgG immunoassay system. All rights reserved. No reuse allowed without permission. author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint (which was not peer-reviewed) is the . https://doi.org/10.1101/2020.04.16.20067231 doi: medRxiv preprint The SARS-CoV-2 IgM and IgG assays are two-step immunoassays for the qualitative detection of SARS-CoV-2 IgM and IgG antibodies in human serum and plasma, using direct chemiluminometric microparticle technology (10, 11) . The assays are performed on a fully automated immunoassay analyzer (manufactured by YHLO Biotech). In the first step, sample, recombinant SARS-Cov-2 antigens-coated paramagnetic microparticles, and a sample treating agent are combined. SARS-CoV-2 IgM or IgG antibodies present in the sample binds to the SARS-CoV-2 All rights reserved. No reuse allowed without permission. author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint (which was not peer-reviewed) is the All rights reserved. No reuse allowed without permission. author/funder, who has granted medRxiv a license to display the preprint in perpetuity. Four serum samples (one negative and three positives for SARS-CoV-2 IgM) and three serum samples (one negative and two positives for SARS-CoV-2 IgG) were measured twice-a-day in duplicates over the span of 20 days, and 80 data points were obtained on each serum. Results are shown in Table 1 and Table 2 . The repeatability of the SARS-CoV-2 IgM is from 2.80% to 4.32%, and the within-laboratory precision is from 3.02% to 4.45%. The repeatability of the SARS-CoV-2 IgG is from 3.11% to 4.30%, and the within-laboratory precision is from 3.12% to 5.13%. All rights reserved. No reuse allowed without permission. author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint (which was not peer-reviewed) is the . https://doi.org/10.1101/2020.04.16.20067231 doi: medRxiv preprint 0 author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint (which was not peer-reviewed) is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint (which was not peer-reviewed) is the . https://doi.org/10.1101/2020.04.16. All rights reserved. No reuse allowed without permission. author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint (which was not peer-reviewed) is the . https://doi.org/10.1101/2020.04.16.20067231 doi: medRxiv preprint author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint (which was not peer-reviewed) is the . https://doi.org/10.1101/2020.04.16.20067231 doi: medRxiv preprint author/funder, who has granted medRxiv a license to display the preprint in perpetuity. author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint (which was not peer-reviewed) is the . https://doi.org/10.1101/2020.04.16.20067231 doi: medRxiv preprint 7 author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint (which was not peer-reviewed) is the . https://doi.org/10.1101/2020.04.16.20067231 doi: medRxiv preprint The y-axis labelled "RLU" denotes the chemiluminescent signal measured as relative light unit. All rights reserved. No reuse allowed without permission. author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint (which was not peer-reviewed) is the . https://doi.org/10.1101/2020.04.16.20067231 doi: medRxiv preprint Severe acute respiratory syndrome-related coronavirus: The species and its viruses -a statement of the Coronavirus Detection of 2019 Novel Coronavirus (2019-nCoV) by Real-Time RT-PCR Effect of virus inactivation on weak positive results of nucleic acid test for 2019 Novel Coronavirus A pneumonia outbreak associated with a new coronavirus of probable bat origin Enhanced immunoassay sensitivity using chemiluminescent acridinium esters with increased light output Automated chemiluminescence immunoassay measurements National Health Commission of the People's Republic of China. The guideline of All rights reserved. No reuse allowed without permission. author/funder, who has granted medRxiv a license to display the preprint in perpetuity.