key: cord-1007224-0tj6ztlo authors: Marvaki, Apostolia; Papachristidis, Alexandros; Nakou, Eleni; Toth, Eva; O'Gallagher, Kevin; Fisher, Richard; Shah, Ajay M.; Monaghan, Mark J. title: Innovative transthoracic echocardiographic imaging on prone ventilated patients with COVID-19 using a transesophageal probe date: 2020-06-24 journal: JACC Cardiovasc Imaging DOI: 10.1016/j.jcmg.2020.06.005 sha: dd35e7a51c3ab7872de124789936b4bf6c513e44 doc_id: 1007224 cord_uid: 0tj6ztlo nan Coronavirus disease 19 (COVID-19) is causing severe pneumonia leading to acute respiratory distress syndrome, a condition where prone ventilation for > 12 hours/day has proven beneficial [1] . Transthoracic echocardiography (TTE) to assess COVID-19-related cardiac complications [2] is very challenging on patients in the prone position. The aim of this study was to examine the feasibility of an innovative use of a transesophageal echocardiography (TEE) probe to perform TTE's on patients with COVID-19 who were invasively ventilated in the prone position. Twenty-one patients were examined. All scans were performed with a Philips Affiniti or CVx systems and the X7-2t or X8-2t TEE probes (Philips Healthcare, Andover, MA). In a small number of patients several frequencies and harmonics were tested. The low fundamental frequency ("Pen" selection -2.4MHz) was found to provide best image quality as visually assessed by two operators. The average duration of a scan with the TEE probe was 16±3 minutes. In 18/21 patients (85.7%), conventional TTE's were performed using the same ultrasound machines and the S5-1 TTE probe (Philips, Andover, MA) with the patients in supine position. For those scans, the operators selected the frequency and harmonics function based upon their discretion for optimal visualization (HGen-1.6/3.2MHz or HPen-1.3/2.6MHz). The studies were performed as part of routine clinical assessment or under ethical approval granted by the NHS Health Research Authority (REC20/EE/0131). The mean time window between the two examinations was 3.6 days (range 0-12 days) and the order of the scans on each patient was random. In order to acquire transthoracic images with the TEE probe, the operator was standing on the left side of the patient at the level of their head, facing the patient's right side and caudally ( Figure 1A ). The probe was slid between the bed mattress and patient's chest wall. Ultrasound gel was applied to the transducer surface. The direction of the probe was oblique from the patient's shoulder towards the mid-sternum. The transducer was facing upwards and was supported underneath by operator's left hand. The probe shaft was held with the right hand to allow rotation and the machine was operated by another operator. The tip of the probe was placed in the conventional left parasternal position, in the third or fourth left intercostal space. Images of the long axis of the heart were obtained at zero angle as in the conventional TTE parasternal long-axis view ( Figure 1B) . By increasing the angle to 60 -120 degrees a short axis view of the heart could be obtained, usually at the level of the mitral valve. Rotating the probe clockwise, the mid-papillary level and apical short axis views were obtained ( Figure 1B) , whereas with counterclockwise rotation, the great arteries were imaged. The application of colour Doppler and pulsed and/or continuous wave Doppler allowed assessment of flows ( Figure 1C ). The majority of patients were male (81%) and 48% were of black ethnicity. The mean age of the cohort was 55.9 ± 7.7 years and the mean body mass index was 28.0 ± 4.6 kg/m 2 . Three patients (14.3%) had a history of coronary artery disease, 5 (23.8%) were diabetic, 3 (14.3%) were smokers and 4 (19.1%) had a history of lung disease. The TEE and TTE probe echocardiograms were reviewed and interpreted by two independent and blinded readers and the image quality was graded as poor, fair and good, using criteria previously described [3] . From the TTE probe scans 4 were graded as non-diagnostic (19.0%), 5 poor (23.8%), 1 fair (4.8%) and 11 good (52.4 %). From the 4 echocardiograms with the TEE probe which were non diagnostic, 2 correlated to poor conventional TTE image quality (both on female patients) and in the other 2 the conventional TTE was also nondiagnostic (both on male patients). The poor quality TEE probe scans corresponded to 3 poor and 1 good quality TTE probe scans (on 1 patient a TTE probe scan was not available). The fair quality TEE probe scan corresponded to fair quality TTE scan and the 11 good quality TEE probe scans correlated to 1 fair and 8 good quality TTE probe scans (on 2 patients TTE probe scans were not available). In 4 patients a D-shaped left ventricle and dilated right ventricle, compatible with right ventricular pressure overload, was found. In one patient the left ventricular systolic function was found to be mildly impaired and a small pericardial and pleural effusion were diagnosed in 2 patients. No valvular abnormalities were identified. The agreement between modalities in echocardiographic findings was 100%. The technique we describe has not been reported previously and in 81% of patients provided adequate image quality for basic left and right ventricular assessment, valve function and pericardial effusion. In 12 patients it was possible to acquire apical views and in 4 a subcostal view was feasible ( Figure 1C ). The unconventional and innovative use of a TEE probe to perform transthoracic echocardiography in prone positioned and invasively ventilated ICU patients, is feasible and can be of diagnostic quality in most cases. The described technique can be an alternative or even superior to conventional TTE on prone patients. COVID-19 ARDS:clinical features and differences to 'usual' pre-COVID ARDS Echocardiography during Prone-Position Mechanical Ventilation in Patients with COVID-19: A Proposal for a New Approach Standardized delineation of endocardial boundaries in three-Dimensional left ventricular echocardiograms