key: cord-1006580-v6hsy20x authors: Porte, Lorena; Legarraga, Paulette; Iruretagoyena, Mirentxu; Vollrath, Valeska; Pizarro, Gabriel; Munita, Jose; Araos, Rafael; Weitzel, Thomas title: Evaluation of two fluorescence immunoassays for the rapid detection of SARS-CoV-2 antigen—new tool to detect infective COVID-19 patients date: 2021-01-21 journal: PeerJ DOI: 10.7717/peerj.10801 sha: a7fe9bc22214750ddaa9888bbc84fbf53027f3c4 doc_id: 1006580 cord_uid: v6hsy20x BACKGROUND: Real-Time Reverse-Transcription Polymerase Chain Reaction (RT-PCR) is currently the only recommended diagnostic method for SARS-CoV-2. However, rapid immunoassays for SARS-CoV-2 antigen could significantly reduce the COVID-19 burden currently weighing on laboratories around the world. METHODS: We evaluated the performance of two rapid fluorescence immunoassays (FIAs), SOFIA SARS Antigen FIA (Quidel Corporation, San Diego, CA, USA) and STANDARD F COVID-19 Ag FIA (SD Biosensor Inc., Gyeonggi-do, Republic of Korea), which use an automated reader. The study used 64 RT-PCR characterized clinical samples (32 positive; 32 negative), which consisted of nasopharyngeal swabs in universal transport medium. RESULTS: Of the 32 positive specimens, all from patients within 5 days of symptom onset, the Quidel and SD Biosensor assays detected 30 (93.8%) and 29 (90.6%) samples, respectively. Among the 27 samples with high viral loads (Ct ≤ 25), the two tests had a sensitivity of 100%. Specificity was 96.9% for both kits. CONCLUSION: The high performance of the evaluated FIAs indicates a potential use as rapid and PCR-independent tools for COVID-19 diagnosis in early stages of infection. The excellent sensitivity to detect cases with viral loads above ~10(6) copies/mL (Ct values ≤ 25), the estimated threshold of contagiousness, suggests that the assays might serve to rapidly identify infective individuals. Since the declaration of the COVID-19 pandemic in March 2020, infection with the new SARS-CoV-2 has resulted in over 30 million confirmed cases and almost 1 million deaths worldwide, as of September 2020 (https://COVID19.who.int). Early detection of cases by highly sensitive and specific real-time reverse-transcription polymerase chain reaction (RT-PCR) is the currently recommended diagnostic strategy (World Health Organization, 2020a) . However, the high cost of RT-PCR, shortage of reagents, and need for trained personnel have limited the testing capacities of laboratories to provide results in a timely manner (World Health Organization, 2020b) . In addition, new aspects of SARS-CoV-2 testing include the estimation of infectivity to tailor control measures of known or suspected COVID-19 cases (Rhee et al., 2020) . Low viral counts detected by RT-PCR might not correspond to infectious virus, especially in asymptomatic individuals or in patients who continue shedding viral RNA after recovering from COVID-19, resulting in false positives and segregation of people who are no longer infectious (Jefferson et al., 2020; Wáng, 2020; Kim et al., 2020) . Therefore, alternative diagnostic tools allowing the rapid testing of large numbers of samples and the ability to evaluate infectivity are of high priority (European Centre for Disease Prevention and Control (ECDC), 2020). Recent developments include whole-genome sequencing, CRISPR-based technologies, droplet-based digital PCR, RT-LAMP, and immunoassays for the detection of antibodies or antigen (Giri et al., 2020) . Rapid antigen detection tests (Ag-RDT) using immunochromatographic (ICT) or fluorescence immunoassays (FIAs) have recently become available; many of which are CE-IVD licensed and some have received FDA emergency use authorization (EUA) (www.finddx.org/COVID-19/pipeline). These tests are fast, easy to use and do not require highly trained personnel or sophisticated equipment. As previously suggested FIAs are highly specific and can reach remarkably high sensitivities, if applied in samples from early phases of infection or with high viral loads Weitzel et al., 2020 , Liotti et al., 2020 . For this reason, they could complement RT-PCR testing by becoming a first line tool for rapid detection of infective individuals. Here we present the performance of two novel FIA automated antigen detection systems in samples from COVID-19 patients presenting within 5 days of symptom onset. Samples derived from patients attending Clínica Alemana in Santiago, Chile, for COVID-19 testing. Specimens consisted of naso-oropharyngeal flocked swabs obtained by trained personnel and placed in universal transport media (UTM-RT Ò System; Copan Diagnostics, Murrieta, CA, USA). Samples were examined for SARS-CoV-2 RNA by RT-PCR assay (COVID-19 Genesig, Primerdesign Ltd., Chander's Ford, UK). Samples exhibiting exponential amplification curves and cycle thresholds (Ct) values ≤40 were considered positive . RT-PCR characterized UTM samples were aliquoted and kept at −80 C until analysis by the two FIA kits, "SOFIA SARS Antigen FIA" (Quidel Corporation, San Diego, CA, USA) and "STANDARD F COVID-19 Ag FIA" (SD Biosensor Inc., Gyeonggi-do, Republic of Korea). Both tests detect SARS-CoV-2 nucleocapsid protein by lateral flow immunofluorescence, which is interpreted by automated analysers (Table 1 ). The kits are CE-IVD labelled; Quidel recently received EUA by the FDA. Manufacturers state that both tests should be performed using nasopharyngeal swabs collected from symptomatic individuals within 5 days of symptom onset. The use of samples stored in certain brands of universal transport media (including Copan UTM) is permitted for the SD Biosensor assay; the Quidel test initially also allowed using UTM, but a recent instruction update discourages the use of such prediluted samples. For the evaluation, 32 RT-PCR positive UTM samples, all collected within the first 5 days after symptom onset, and 32 negative specimens were selected. All positive samples were from symptomatic patients, 12 negative samples were from asymptomatic patients screened before surgery. Some of the positive (n = 27) and negative samples (n = 19) had been used in a previous evaluation . Assays were performed using the same sample aliquot, following manufacturers´instructions, by the same laboratory personnel, who were blinded to RT-PCR results. In brief, specimens were mixed with an extraction reagent, dispensed into the cassette's sample well, and read after incubation by an instrument. Testing procedures were performed under a BSL2 cabinet. RT-PCR served as reference method; in case of discordant results, tests were repeated. Demographic and clinical information was retrieved from the national surveillance system and anonymously analysed. Sensitivity, specificity, and accuracy were calculated as recommended by current Clinical and Laboratory Standards Institute guidelines (CLSI, 2008) ; for the Wilson score Confidence Interval (CI) at 95%, we used OpenEpi (version 3.01). Test performance was evaluated for all samples and for those with high viral loads (Ct ≤ 25), as previously described (World Health Organization, 2020c). Kits and analysers were provided by manufacturers at reduced costs for evaluation purposes. The study was approved by the Comité Etico Científico, Facultad de Medicina Clínica Alemana, Universidad del Desarrollo in Santiago, Chile, and a waiver of informed consent was granted (Project ID number 931). The study included a total of 64 samples, 32 were RT-PCR positive and 32 RT-PCR negative. The median age was 39 years (IQR 36.7-57) and 52% were male. Median days from symptom onset to RT-PCR testing of positive and negative cases were 2 (IQR 1-3) and 1 (IQR 0.75-4), respectively. Ct values had a median of 17.95 (IQR, ); 29/32 samples (90.6%) had a Ct ≤ 25 (Table 2) . Both assays demonstrated an overall sensitivity >90%, reaching 100% for samples with high viral loads (Table 3) . False negative results were observed with the Quidel and SD Biosensor assays in two and three samples, respectively, which had Cts of 30.89-32.57 and were taken on the 4th or 5th day after symptom onset. Specificity was 96.7% for both tests, that is, both kits displayed a single false positive result, from two distinct symptomatic RT-PCR negative cases (Table 3 ). The two assays were user friendly, included ready-touse reagents and required little hands-on time. Moreover, analysers were easy-to-use, stored the results, and included options for QR coding, printing, and connection to laboratory information systems. Antigen-based assays have risen as one of the most practical alternatives, but independent evaluations of their diagnostic performance are scarce and their role within the routine diagnostic workup is yet not defined (World Health Organization, 2020c; Dinnes et al., 2020) . One of the main concerns about antigen-based tests is that these assays per se have a higher detection threshold than RT-PCR, thus, they might miss cases with low viral replication during very early or late infections (World Health Organization, 2020c; Liotti et al., 2020) . However, the results of this and former studies indicate that antigen detection by immunofluorescence, especially when used with an automated reader, has an excellent sensitivity to detect SARS-CoV-2 in samples with estimated viral loads abovẽ 10 6 copies/mL (Ct values ≤ 25) (World Health Organization, 2020c), which are found in pre-symptomatic (1-3 days before symptom onset) and early symptomatic COVID-19 cases (up to 5-7 days after symptom onset) Lee et al., 2020; Zou et al., 2020) . According to recent modelling studies, elevated viral titers are associated to infectivity (Larremore et al., 2020) . This is in accordance with in vitro experiments, which showed no viral growth from samples with Cts > 24 or taken >8 days after symptom onset (Bullard et al., 2020 , Wölfel et al., 2020 . A viral load of 10 6 copies/mL has therefore been suggested as the limit of infectivity for clinical practice (Drosten, 2020) . However, until the exact threshold of contagiousness is known, other authors have considered a more conservative approach (1,000 copies/mL) (Jacot et al., 2020) . For samples with high viral loads, our comparative study showed that both tests were 100% sensitive. This is in accordance with manufacturers' information and a recent study from Italy of the STANDARD F assay, which showed a high detection rate of 95.2% (20/21) in samples with Cts < 25 (Liotti et al., 2020) . In our panel of positive samples, false negatives only occurred with Cts > 30, which translates to viral loads <10 4 for the used RT-PCR protocol (Scohy et al., 2020) , although this finding has to be confirmed with a larger number of specimens. The high-performance value coincides with recent studies of a similar FIA with automated reading (BioEasy, Shenzhen, China), which demonstrated sensitivities of 100% for samples with Cts ≤ 25 and of 98% for samples with Cts ≤ 30 (Diao et al., 2020) . In contrast, immunochromatographic SARS-CoV-2 antigen tests demonstrated lower sensitivity values of 74-85% for samples with Cts ≤ 25 Mertens et al., 2020; Lambert-Niclot et al., 2020) . The evaluation of test specificity was limited by the low number of included RT-PCR-negative samples. In our experience, the rate of false positives using the Biosensor assay in pre-surgical samples of low-risk asymptomatic patients was 1.6% (6/374) (L. Porte and T. Weitzel, 2020, unpublished observations) . Although additional studies with larger numbers of samples are needed, the excellent performance data of FIA Ag-RDTs suggest their potential use in the following scenarios, when RT-PCR is unavailable or impractical: (1) closed or semi-closed remote communities such as cruise ships or military camps (World Health Organization, 2020c), (2) High-risk congregate facilities including schools, care-homes, dormitories, etc., when testing daily or every other day could reduce secondary infections by 100% or 90%, respectively (Paltiel, Zheng & Walensky, 2020) , and (3) screening of asymptomatic attendees at potential superspreader events, like conferences, weddings, and sports or cultural events. In the future, due to their high sensitivity to detect infective patients, FIA Ag-RDTs might also play an important role within "test-out" strategies, that is, the early release of suspected cases from self-isolation or shortening quarantine for proven cases. The authors received no funding for this work. The authors declare that they have no competing interests. Lorena Porte conceived and designed the experiments, performed the experiments, analyzed the data, prepared figures and/or tables, authored or reviewed drafts of the paper, and approved the final draft. Paulette Legarraga conceived and designed the experiments, performed the experiments, analyzed the data, authored or reviewed drafts of the paper, and approved the final draft. Mirentxu Iruretagoyena conceived and designed the experiments, analyzed the data, authored or reviewed drafts of the paper, and approved the final draft. Valeska Vollrath conceived and designed the experiments, analyzed the data, authored or reviewed drafts of the paper, and approved the final draft. Gabriel Pizarro conceived and designed the experiments, performed the experiments, prepared figures and/or tables, and approved the final draft. Jose Munita conceived and designed the experiments, analyzed the data, authored or reviewed drafts of the paper, and approved the final draft. Rafael Araos conceived and designed the experiments, analyzed the data, authored or reviewed drafts of the paper, and approved the final draft. Thomas Weitzel conceived and designed the experiments, analyzed the data, prepared figures and/or tables, authored or reviewed drafts of the paper, and approved the final draft. The following information was supplied relating to ethical approvals (i.e., approving body and any reference numbers): The study was approved by the institutional review board (Comité Etico Científico, Facultad de Medicina Clínica Alemana, Universidad del Desarrollo, Santiago, Chile) and a waiver of informed consent was granted (Project ID number 931). The following information was supplied regarding data availability: Raw data are available as a Supplemental File. Predicting infectious SARS-CoV-2 from diagnostic samples User protocol for evaluation of qualitative test performance; approved guideline-second edition: CLSI document EP12-A2 Diagnosis of acute respiratory syndrome coronavirus 2 infection by detection of nucleocapsid protein Diagnostic Test Accuracy Group. 2020. Rapid, point-of-care antigen and molecular-based tests for diagnosis of SARS-CoV-2 infection European Centre for Disease Prevention and Control (ECDC). 2020. COVID-19 testing strategies and objectives Review of analytical performance of COVID-19 detection methods. Epub ahead of print 18 Temporal dynamics in viral shedding and transmissibility of COVID-19 Viral load of SARS-CoV-2 across patients and compared to other respiratory viruses Viral cultures for COVID-19 infectious potential assessment-a systematic review Detection and isolation of SARS-CoV-2 in serum, urine, and stool specimens of COVID-19 patients from the Republic of Korea Evaluation of a rapid diagnostic assay for Porte detection of SARS-CoV-2 antigen in nasopharyngeal swabs Test sensitivity is secondary to frequency and turnaround time for COVID-19 screening. Epub ahead of print Serological approaches for COVID-19: epidemiologic perspective on surveillance and control Performance of a novel diagnostic assay for rapid SARS-CoV-2 antigen detection in nasopharynx samples. Epub ahead of print 23 LHUB-ULB SARS-CoV-2 Working Diagnostic Group. 2020. Development and potential usefulness of the COVID-19 Ag respi-strip diagnostic assay in a pandemic context Assessment of SARS-CoV-2 screening strategies to permit the safe reopening of college campuses in the United States Evaluation of a novel antigen-based rapid detection test for the diagnosis of SARS-CoV-2 in respiratory samples Duration of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) infectivity: when is it safe to discontinue isolation? Epub ahead of print 25 Low performance of rapid antigen detection test as frontline testing for COVID-19 diagnosis CT suggests discharged COVID-19 patients who were retested RT-PCR positive again for SARS-CoV-2 more likely had false negative RT-PCR tests before discharging Comparative evaluation of four rapid SARS-CoV-2 antigen detection tests using universal transport medium Diagnostic testing for SARS-CoV-2: interim guidance Advice on the use of point-of-care immunodiagnostic tests for COVID-19: scientific brief Antigen-detection in the diagnosis of SARS-CoV-2 infection using rapid immunoassays: interim guidance Virological assessment of hospitalized patients with COVID-2019 SARS-CoV-2 viral load in upper respiratory specimens of infected patients Supplemental information for this article can be found online at http://dx.doi.org/10.7717/ peerj.10801#supplemental-information.