key: cord-1002105-8ipf31rz authors: Johnson, J Edward; Selvakumar, A title: Oxygen consumption of non-invasive ventilation modes in ICU ventilators date: 2021-12-22 journal: Indian J Anaesth DOI: 10.4103/ija.ija_761_21 sha: 21789da29ecda46cbcbcbd251383b3a6fcec1516 doc_id: 1002105 cord_uid: 8ipf31rz nan Armstrong For all oxygen delivery devices, we know the exact rate of supply of oxygen per minute. So, we can adjust the flow of oxygen and conserve oxygen utilisation. But in ICU ventilators for NIV and invasive ventilation, we do not know the exact requirement of oxygen per minute. Stuart and Klein [1] mentioned a method of estimating gas consumption in an ICU ventilator [ Table 1 ]. Since this calculation is for intubated patients on ventilator, a high flow rate may be needed for NIV because of the leak and the oxygen utilisation is likely to go up. The mechanical ventilator may generate flow by using the pressure gradient between the supply piped gas and the atmosphere. Alternatively, the ventilator may generate flow by means of a gas compressor turbine. These methods have the advantage of being able to generate the sort of high flows which are required for NIV (to compensate for leak around the edges of the mask). For most ICU ventilators, the maximum flow rate is approximately 200-250 L/min. According to a study by Anderson et al., [2] the peak inspiratory flow in a calmly breathing person seems to be 5-10 L/min, but with exercise (or respiratory distress) it can go up to 30-40 L/min. At peak exercise, the authors recorded peak inspiratory flow rates (PIFRs) of around 250 L/min from their healthy volunteers. There is a direct relationship between increase in PIFR and handgrip exercise. [3] Amato et al. [4] observed eight ventilated patients (four patients had acute respiratory distress syndrome) and found that some ended up with PIFR of up to 155 L/min. Then, if a 25% leak is added as the acceptable maximum ceiling (a leak in the circuit, a deflated cuff, an ill-fitted mask), an adult ventilator would probably need to produce at least around 200 L/min of flow. This is indeed the case for most of the existing models. Obviously, the ability to produce these high flow rates is dependent on the reliability of the hospitals' gas supply. If the central supply pressure is inadequate, the ventilator will not be able to produce such high flows. The technical specifications for these devices usually quote a minimum supply gas flow rate as 120 L/min, or a minimum supply pressure of 200 kPa. We can refer the service manual of the ventilators for the maximum flow that the ventilator can generate. To meet the flow rate in NIV, how much oxygen will be utilised is a million-dollar question. To know the oxygen utilisation per minute for different ICU ventilators, we have conducted a study by connecting D-type cylinders to the ICU ventilators and calculated the oxygen utilisation per minute. The minute flow rate (adjusting respiratory rate and tidal volume) and flow trigger were kept constant in all patients. For the commonly used ICU ventilators, the minimum usage of oxygen per minute when on 100% fractional inspired concentration of oxygen and a tight fitting mask can vary according to the make and model of the ventilator [ Table 2 ]. The average usage of oxygen per minute for NIV in an ICU ventilator is around 40-50 L. That will be around 57,600-72,000 L/day of gaseous oxygen equivalent to 67-84 L of liquid oxygen per day for a patient. So, it is easy to calculate the one-day usage of oxygen in each ICU. Since constant up-gradation of central oxygen supply and Medical Gas Pipeline System is to be done based upon anticipated future requirements as per Indian Society of Anaesthesiologists (ISA National) Advisory and Position Statement, this calculation of usage of oxygen for NIV will be much useful. [5] But future studies are needed to calculate the flow rate for ICU ventilators and their oxygen requirement depending upon the lung compliance. This table is based on flow rates under experimental conditions. Please note that flow rates may vary considerably depending on patient factors, oxygen supply factors and device/ventilator modes, most notably bias flow settings. Please consult manufacturer specifications when making any estimations as newer data may be available Enhanced recovery after surgery (ERAS)…. still a distant speck on the horizon ! Enhanced recovery after surgery (ERAS): Are anaesthesiologists prepared for the paradigm shift in perioperative care? A prospective cross-sectional survey in India Adopting newer strategies of perioperative quality improvement: The bandwagon moves on… Portable ventilators Peak inspiratory flows of adults exercising at light, moderate and heavy workloads Association between peak inspiratory flow rate and hand grip muscle strength in hospitalized patients with acute exacerbation of chronic obstructive pulmonary disease Volume-Assured Pressure Support Ventilation (VAPSV). A new approach for reducing muscle workload during acute respiratory failure Second wave of COVID-19 pandemic and the surge of mucormycosis: Lessons learnt and future preparedness: Indian Society of Anaesthesiologists (ISA National) advisory and position statement Oxygen consumption of non-invasive ventilation modes in ICU ventilators This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms. Quick response code Website: www.ijaweb.org DOI: 10.4103/ija.ija_761_21