key: cord-1001020-40qzpbuz authors: Davies, John D.; Huang, Yuh Chin; MacIntyre, Neil R. title: Evaluation of a novel endotracheal tube suctioning system incorporating an inflatable sweeper date: 2021-10-12 journal: Can J Respir Ther DOI: 10.29390/cjrt-2021-026 sha: 0289713cd7eabcb9db4ac4db187048c1c25b07af doc_id: 1001020 cord_uid: 40qzpbuz INTRODUCTION: Accumulation of secretions in an endotracheal tube can increase the resistance to flow resulting in an increased patient work of breathing when the patient is interacting with the ventilator. Retained secretions can also serve as an infection risk. Standard suction catheters are limited in their ability to keep the lumen of the endotracheal tube clear. A novel closed-suction catheter has been introduced that incorporates a balloon at its distal end that, when inflated, physically scrapes secretions out of the endotracheal tube (CleanSweep catheter (CSC), Teleflex, Morrisville NC). We hypothesized that the CSC would be more efficient at removing secretions from inside the endotracheal tube than a standard suction catheter (SSC). METHODS: We performed a bench study examining resistive pressures across different sizes of endotracheal tubes when cleaned by the CSC as compared with an SSC. This study was followed by a prospective crossover study again comparing the CSC with an SSC in intubated intensive care unit patients receiving mechanical ventilation and requiring frequent suctioning. RESULTS: For the bench study the CSC was significantly better in reducing airway resistive pressures (P < 0.001). In the prospective crossover study the CSC over 2 h also removed significantly more secretions than the SSC (P < 0.05). CONCLUSION: Both our bench and crossover clinical study demonstrated improved clearance of secretions with the CSC vs an SSC. Further research is needed to ascertain the clinical outcome benefits of enhanced secretion removal. The endotracheal tube (ETT) is a life-saving device that can help protect the natural airway and that allows delivery of mechanical ventilator support. Unfortunately, these tubes compromise normal glottic function, impede coughing, and impair normal airway secretion clearance. Consequent secretion retention in the lungs and in the tube can serve as a bacterial breeding ground and can produce significant airway obstruction with consequent elevations in a patient's work of breathing when the patient is interacting with the ventilator [1, 2] . These all have the potential to prolong the need for mechanical ventilator support [3, 4] . ETT suctioning devices have been used for decades to facilitate secretion removal during mechanical ventilation [2, [5] [6] [7] [8] . While these devices can clearly remove secretions, the impact on clinical outcomes has not been carefully studied [8] . Indeed, these catheters have significant downsides: secretion removal is restricted to only regions near the catheter tip, tracheal stimulation can elicit unfavorable hemodynamic conditions and central nervous system reflexes (e.g., fluctuations in intracranial pressure), the catheter can traumatize the airways (and perhaps even stimulate secretion production), the applied negative pressure can create distal airway collapse and hypoxemia, and the procedure can cause considerable discomfort [2, 9] . Opening the ventilator circuit to insert the suction device can also cause a loss of positive end-expiratory pressure (PEEP) and an increase in pneumonia risk [6] . This latter issue has prompted the development of closed-suction catheters that are integrated into the ventilator circuit [6] . Strategies to make the suctioning process more effective can potentially improve lung function and also reduce the need for recurrent suctioning. Examples of such strategies include novel catheter tip designs and multiple small catheters imbedded in the ETT wall [10] [11] [12] [13] . Perhaps the most direct approach is to use mechanical devices to physically extract airway secretions out of the tube. First described using standard Fogarty catheters [14] , more modern versions use a variety of catheter designs to physically "scrape" the inside of the ETT [15] [16] [17] [18] [19] . These devices generally enhance the efficiency of secretion clearance but, unfortunately, they are usually standalone catheters requiring circuit disruption for use and the need for a separate suction system. A novel closed-system suctioning device with an integrated inflatable balloon at its tip was recently introduced (CleanSweep Teleflex, Morrisville NC). The balloon is not designed to be inflated with the catheter beyond the distal end of the ETT, and it is only inflated during catheter withdrawal thus "sweeping" the secretions off the interior wall of the ETT (Figure 1 ). Theoretically this design has appeal in regards to potentially enhancing secretion removal. The question we want to answer is whether the CleanSweep suction catheter is more efficient for secretion removal than the standard suction catheter (SSC; Halyard, Alpharetta, Georgia) that is currently being used at our institution. This study would be performed firstly in a bench model followed by a crossover clinical design in mechanically ventilated patients requiring frequent suctioning. In this bench study, the CleanSweep catheter (CSC) was compared with an SSC. The systems were connected through a standard mechanical ventilator circuit (ISO-Gard, Teleflex, Morrisville, North Carolina) to a mechanical ventilator (Getinge Servo-i) with the following settings: volume assist control mode, tidal volume (VT) 500 mL, respiratory rate (RR) 12/min, PEEP 5 cm H 2 O, inspiratory time (Ti) 0.8 s, Ti rise 0.40 (resulted in a constant flow of 50 L/min). The distal end of the ETT was connected to an Ingmar single chamber test lung with a resistance setting of 5 cm H 2 O/L/s. An upper airway model (Biovo Tecnologies, Rosh HaAyin, 4809173, Israel) was used and the head was positioned at approximately 30 degrees ( Figure 2 ). ETTs (Mallinckrodt Covidien, Boulder, Colorado) of three different diameters (7.0, 7.5, and 8.0 mm) were used. Artificial mucus was created based on a study published by Rozycki et al. [20] using Polyox water-soluble resin N-750 solution (Dow Chemicals, Woodbury, New Jersey) at a 2.5% concentration. The artificial mucus was made by heating 200 mL of water to 95-97°C and adding 2.5 g of Polyox per 100 mL of water. The solution was placed on a vibrating mixer for 2 h until the solution was homogeneous. The artificial mucus was dyed blue to enhance visualization [18] . Three millilitres of the artificial mucus was inserted into the ETT, and once the peak pressure stabilized (5-10 breaths) airway pressures (peak pressure (Ppeak) and plateau pressure (Pplat)) were recorded. Following mucus insertion, suctioning was performed using both catheters in random order (initial and follow-up suction). The suction catheter (14 French) was inserted past the distal end of the ETT and suction was applied at -120 mm Hg for 5 s while withdrawing the catheter. With the CSC, the balloon was deflated during catheter insertion and inflated once the catheter was in place. After each suctioning airway pressure measurements were repeated. Five runs on each size ETT using each system as both an initial and follow-up device were carried out. Five additional runs on a 7.0 mm ETT were made alternating between the suction systems with size 12 French catheters. Each individual catheter was cleaned and reused one time. Mucus volume and airway pressure measurements under each test condition were averaged and differences between the catheter systems were analyzed using two-sided paired t-test for each testing condition. Significance was taken at P < 0.05. The prospective crossover study was approved by the Duke Institutional Review Board (Pro00087066). Eight intubated mechanically ventilated patients requiring frequent ( 7 days) were less common than we anticipated. Second, physicians (and sometimes families) were quite reluctant to allow circuit disruptions for catheter placement. This latter point underscores the importance of using closed-system suctioning systems whenever possible. Further exacerbating this recruitment issue was the onset of the COVID-19 pandemic. When this occurred, all nonessential clinical studies involving human intervention were suspended at our institution. At the end of the day, we are left with a decision to use enhanced mechanical suctioning systems, such as the one we described, based on clinical judgment. Patients with copious secretions, prolonged ventilator requirements, and excessive work of breathing would certainly seem the most likely patients to consider as candidates as we await further studies. Support provided by an unrestricted grant from Teleflex, Morrisville NC. Teleflex was not involved with any of the study procedures, data acquisition or data analysis. Mr. Davies discloses that he is a clinical consultant for Teleflex Medical. Dr. Huang discloses a financial relationship with Windtree Therapeutics. Dr. MacIntyre discloses financial relationships with Inspirx, Ventec Life Systems, and Hillrom. All authors have completed the ICMJE uniform disclosure form. Endotracheal tube intraluminal volume loss among mechanically ventilated patients Principles and practice of mechanical ventilation Predicting extubation failure after successful completion of a spontaneous breathing trial Predictors of extubation outcome in patients who have successfully completed a spontaneous breathing trial Endotracheal tube intraluminal diameter narrowing after mechanical ventilation: use of acoustic reflectometry Clinical practice guidelines. Endotracheal suctioning of mechanically ventilated patients with artificial airways 2010 An in vitro evaluation of the effectiveness of endotracheal suction catheters Endotracheal suctioning of the adult intubated patient -What is the evidence? The effect of suction catheter insertion and tracheal stimulation in adults with severe brain injury The new NL tracheal suction catheter The influence of suction catheter tip design on tracheobronchial trauma and fluid aspiration efficiency Design and function of tracheal suction catheters A 72-hour study to test the efficacy and safety of the "Mucas Slurper" in mechanically ventilated sleep Extirpation of endotracheal tube secretions with a Fogarty arterial embolectomy catheter Evaluation of endotracheal tube scraping on airway resistance Endotracheal tubes cleaned with a novel mechanism for secretion removal: a randomized controlled clinical study Removal of endotracheal tube obstruction with a secretion clearance device Effect of the endOclear device on biofilm in endotracheal tubes A clinical assessment of the Mucus Shaver: a device to keep the endotracheal tube free from secretions Endotracheal tube seal and suction performance in a novel biorealistic tracheal model Endotracheal tube narrowing after closedsystem suctioning: prevalence and risk factors Endotracheal tube intraluminal volume loss among mechanically ventilated patients The prospective, crossover study was approved by the Duke Institutional Review Board (Pro00087066). The bench study did not involve human research so Duke Institutional Review Board approval was not needed. All authors were responsible for (i) substantial contributions to the conception and design of the work and the acquisition, analysis, and interpretation of data for the work; (ii) drafting the work and revising it critically for important intellectual content; (iii) final approval of the version to be published; and (iv) agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. All authors read and approved the final version of the manuscript.