key: cord-0999494-wjdxeqio authors: Satomi, Natsuko; Enta, Kenji; Otsuka, Masato; Ishii, Yasuhiro; Asano, Ryota; Sawa, Shigeharu title: Left ventricular free wall perforation by a right ventricular pacemaker lead: a case report date: 2021-03-31 journal: Eur Heart J Case Rep DOI: 10.1093/ehjcr/ytab125 sha: 2e91ace4901ff201fac2e7afbff742d472c74647 doc_id: 999494 cord_uid: wjdxeqio BACKGROUND: Lead perforation is one of the major complications of pacemaker implantation, but cases of right ventricular (RV) lead perforation through the septum and left ventricle are rarely reported. We described a rare case of left ventricular (LV) free wall perforation by an RV lead and the management of this complication. CASE SUMMARY: An 84-year-old man was admitted with a dual-chamber pacemaker due to pacing failure caused by an RV lead fracture. New lead implantation was performed on the next day, but pacing failure occurred again on the second post-operative day (POD). We found the lead perforation on the fluoroscopy during temporary pacemaker insertion. Computed tomography scan and transthoracic echocardiogram showed that the added lead perforated through both the septum and LV free wall. A new lead was inserted on the fourth POD, and an off-pump open chest surgery for extraction of the penetrating lead was performed uneventfully on the 20th POD. DISCUSSION: We considered that some features of the lead (SelectSecure 3830-69, Medtronic) may be related to this complication, as the lead was very thin, had a non-retractable bare screw and was inserted with a dedicated delivery catheter. We have to be careful when performing implantation of this kind of lead to avoid such a rare complication. • Pacemaker lead perforation not only through the septum but also through the left ventricular free wall is a rare but serious complication following pacemaker implantation. • Despite initially unremarkable pacemaker performance measurements, lead perforation has to be considered when pacing failure occurs and further diagnostic examinations should be performed. • One should take extreme care not to screw a lead deeply and not to give too much slack. This is particularly the case for non-retractable bare screw leads that are implanted using a dedicated delivery catheter. Cardiac perforation by a pacemaker lead is a serious and rare complication, occurring in only 0.1-0.8% of pacemaker implantation procedures. 1 Although most perforations are recognized during or shortly after implantation, subacute and late perforations have also been reported. 2 The most frequent site of lead perforation is the right ventricular (RV) apex; however, only one case of lead penetration through both the ventricular septum and the left ventricular (LV) free wall has been reported. 3 Here, we described a case of both septum and LV free wall perforation by an RV pacemaker lead within a few days after the new lead insertion. An 84-year-old Japanese man with a dual-chamber pacemaker implanted via the left cephalic vein for complete atrioventricular block 7 years prior was admitted to our hospital with syncope due to pacing failure caused by an RV lead fracture (impedance: 1671 X, threshold: 2.25 V at 0.4 ms) ( Figure 1A) . He was on 10 mg amlodipine besylate and 100 mg aspirin once a day for hypertension and atherosclerotic stroke, and not on any medicine precipitating syncope. There were no signs of neurological deficits. The previous device check-up conducted 6 months prior to pacing failure showed normal impedance, threshold and pacing status (562 X, 1.25 V, 70% of atrial and 100% of ventricular pacing). He had no finding of heart failure, and the transthoracic echocardiogram (TTE) on admission showed normal LV contraction (ejection fraction was 55%) and mild tricuspid valve regurgitation. There was no visible dislocation or lead fracture on the chest X-ray. On the following day, a new RV lead, a thin lumenless lead with a diameter of 1.37 mm (SelectSecure 3830-69, Medtronic, Minneapolis, MN, USA) was inserted. It was carefully screwed into the RV septum with a non-deflectable delivery catheter (C315S10, Medtronic). This lead was selected because of its thinness as there was a narrowing of the left cephalic vein caused by the two previous leads. Moreover, we considered that it might be easy to screw it to a specific position by a non-deflectable delivery catheter despite of the obstruction by the previous RV lead. There were no abnormalities and no complications during and immediately after the surgery (intraoperative RV lead impedance and threshold were 814 X and 0.9 V, and those immediately after operation were 646 X and 1.0 V, respectively). On the first post-operative day (POD), transient ventricular pacing failure appeared with more than 10-s pause causing syncope. His electrocardiogram showed the right bundle branch block pattern ( Figure 1B) , but there was neither detectable pericardial effusion on TTE nor pneumothorax or dislodgement on X-ray ( Figure 1C) . However, such pacing failure became more frequent. Finally, on the second POD, he had a pacemaker exit block with a slow ventricular escape rhythm (20 b.p.m.) ( Figure 1D ). Right ventricular lead data did not show any remarkable change (RV lead impedance and threshold were 513 X and 1.75 V, respectively). A temporary transvenous pacemaker was inserted emergently, and we first found the additional lead penetrating through the cardiac wall on fluoroscopy ( Figure 2A) . Computed tomography (CT) and TTE showed the tip of the additional lead exiting in the left thoracic cavity passing through the septum and the LV free wall ( Figure 2B-D) . On the fourth POD, an additional new lead (CapSureFix NOVUS 5076-58, Medtronic) was inserted in the RV septum, and the temporary transvenous pacemaker was removed. Finally, an off-pump open chest surgery for the extraction of the penetrating lead was performed on the 20th POD, with prior discontinuation of aspirin for 2 weeks. The tip of the lead was found piercing through the LV posterior free wall into the thoracic cavity ( Figure 3 ). The parietal pleura was scratched by the tip of the lead without lung damage. The lead was extracted from the generator pocket smoothly, and the perforation wounds of the LV and pericardium were sutured using the felt sandwich technique. The clinical course after the open chest surgery was uneventful. On subsequent pathological examination, the tissue sample from the extracted lead showed the normal structure of the myocardium. The patient was discharged on the 34th POD, and he had been uneventful for a year. We experienced a rare case of RV pacemaker lead perforation through both the septum and the LV free wall. We speculate the patient's clinical course after the new lead insertion as follows: first, when the pacing failure occurred on the first POD, the lead perforated only through the septum; second, the failure occurred and resolved because of the intermittent LV wall pacing after the penetration of the septum; finally, the pacing failed completely when the lead penetrated through the LV free wall. We consider that certain features of the lead (SelectSecure 3830-69, Medtronic) are related to this complication. It is thin with a non-retractable bare screw and is inserted with a dedicated delivery catheter. 4 The tip load of the thin lead is quite heavy and may have created some stress on the cardiac wall. Moreover, the operator's torque was easily and directly conducted to the tip of the lead while positioning the lead into the Dual-chamber pacemaker was implanted due to complete atrioventricular block. Here, we considered that HBP and LBBAP would not be optimal for our elderly patient with preserved LV function; however, the features of the lead being thin and easily controllable were useful as described in the previous section. All the reported lead perforations related to HBP and LBBAP were caused by this lead and the delivery system. 5 There were some reports of septal perforation by this lead but not of LV free wall perforation before. To avoid such a lead complication with additional lead implantation, a leadless pacemaker in combination with the functional AAI system could be an alternative therapy. On initial examination, X-ray and TTE were not suggestive of lead perforation. According to a previous report on patients with lead perforation, pericardial effusion and tamponade occurred in only 38% and 19% of patients respectively, although 72% of all patients showed symptoms. 6 Hence, CT should provide an accurate diagnosis in cases with ambiguous findings. 7 There is only one previous case reporting asymptomatic RV lead perforation through both the septum and LV free wall. 3 According to that report, diagnosing lead perforation with no obvious findings was quite difficult; however, the lead perforation was found unexpectedly by X-ray at a routine 2-week follow-up With respect to lead extraction, a previous study reported transvenous extraction with cardiac surgery in standby as a safe method to treat the lead perforations; however, there has been no clear consensus on the appropriate management of LV perforation in stable patients. 6 Here, we chose an open chest surgery for the extraction of the penetrating lead because of the high risk of LV free wall penetration. As a result, there was no complication with the lead extraction. In conclusion, we experienced a rare case of RV lead perforation through the septum and the LV free wall. We recommended that while inserting thin leads with a non-retractable bare screw by a dedicated delivery catheter, extreme care should be taken to avoid screwing too deeply and creating excessive redundancy and to wait for a moment until the torque can be released. Supplementary material is available at European Heart Journal-Case Reports online. Slide sets: A fully edited slide set detailing this case and suitable for local presentation is available online as Supplementary data. The authors confirm that written consent for submission and publication of this case report including image(s) and associated text has been obtained from the patient in line with COPE guidance. Lead perforation: incidence in registries Delayed lead perforation: a disturbing trend Right ventricular lead perforation through the septum, left ventricle, and pleura, managed by an open surgical approach Multi-center clinical experience with a lumenless, catheter-delivered, bipolar, permanent pacemaker lead: implant safety and electrical performance Left bundle branch pacing is the best approach to physiological pacing Clinical, anatomical, and technical risk factors for postoperative pacemaker or defibrillator lead perforation with particular focus on myocardial thickness Subacute right ventricle perforation by pacemaker lead presenting with left hemothorax and shock Funding: None declared. Left ventricular free wall perforation