key: cord-0998813-ec3pmvih authors: Begley, J. L.; Lavery, K. E.; Nickson, C. P.; Brewster, D. J. title: The aerosol box for intubation in coronavirus disease 2019 patients: an in‐situ simulation crossover study date: 2020-06-01 journal: Anaesthesia DOI: 10.1111/anae.15115 sha: b25406fedda8c98e076135cef12df228ca6fd4f1 doc_id: 998813 cord_uid: ec3pmvih The coronavirus disease 2019 pandemic has led to the manufacturing of novel devices to protect clinicians from the risk of transmission, including the aerosol box for use during tracheal intubation. We evaluated the impact of two aerosol boxes (an early‐generation box and a latest‐generation box) on intubations in patients with severe coronavirus disease 2019 with an in‐situ simulation crossover study. The simulated process complied with the Safe Airway Society coronavirus disease 2019 airway management guidelines. The primary outcome was intubation time; secondary outcomes included first‐pass success and breaches to personal protective equipment. All intubations were performed by specialist (consultant) anaesthetists and video recorded. Twelve anaesthetists performed 36 intubations. Intubation time with no aerosol box was significantly shorter than with the early‐generation box (median (IQR [range]) 42.9 (32.9–46.9 [30.9–57.6])s vs. 82.1 (45.1–98.3 [30.8–180.0])s p = 0.002) and the latest‐generation box (52.4 (43.1–70.3 [35.7–169.2])s, p = 0.008). No intubations without a box took more than 1 min, whereas 14 (58%) intubations with a box took over 1 min and 4 (17%) took over 2 min (including one failure). Without an aerosol box, all anaesthetists obtained first‐pass success. With the early‐generation and latest‐generation boxes, 9 (75%) and 10 (83%) participants obtained first‐pass success, respectively. One breach of personal protective equipment occurred using the early‐generation box and seven breaches occurred using the latest‐generation box. Aerosol boxes may increase intubation times and therefore expose patients to the risk of hypoxia. They may cause damage to conventional personal protective equipment and therefore place clinicians at risk of infection. Further research is required before these devices can be considered safe for clinical use. The severe acute respiratory syndrome-coronavirus-2 (SARS-CoV-2) pandemic has put the lives of healthcare workers at great risk, as evidenced by the growing numbers of healthcare worker deaths in many nations [1, 2] . The threat to healthcare workers from the virus, and the associated fear, has been further exacerbated by a worldwide shortage of personal protective equipment (PPE) [3, 4] . SARS-CoV-2, which causes coronavirus disease 2019 , is predominantly spread by droplets but has the potential for aerosol transmission [5] . This has seen a significant focus on the utility and application of PPE during aerosol-generating procedures including intubation [6] . Data from the severe acute respiratory syndrome (SARS) epidemic of [2002] [2003] suggest that healthcare workers involved in intubation were at six times the risk of acquiring that coronavirus [7] . The correct application of PPE and infection control techniques were shown to significantly reduce this risk [8, 9] . Recently, international guidelines have been developed to assist healthcare workers involved in the airway management (including intubation) of COVID-19 patients [10, 11] . Ongoing clinician concerns and anxiety about safety, especially in light of PPE shortages, has driven solutions including increased manufacturing of PPE, suggestions regarding the reuse of PPE and in some cases the creation of novel devices to protect healthcare workers [12, 13] . Such innovation has virtue, especially in times of supply shortages, but the inherent risks of this approach are also described. For example, while regulated equipment is usually subjected to testing in a range of circumstances by various users, improvised equipment is not and is therefore likely to have a higher failure rate in clinical use [14] . One such novel device is the aerosol box. First described by Dr Lai Hsien-yung [15] , these devices typically consist of a transparent plastic cube covering a patient's head and shoulders, with access holes for the intubating proceduralist's arms and sometimes additional holes for an assistant. These devices have been discussed in both the traditional medical literature [16] [17] [18] but more commonly on social media and medical education websites, often being praised for their ingenuity [19] [20] [21] [22] . Despite there being no published research on the safety or efficacy of the aerosol boxes, these devices are being used in clinical practice and manufacturers have already distributed hundreds of units to hospitals in the USA, the UK and Australia [12, 17, [23] [24] [25] [26] [27] . There is no doubt that genuine fear is motivating the distribution and use of these devices, especially in systems with critical PPE shortages. Additional drivers of this rapid implementation may also include 'Gizmo Idolatry' (the implicit conviction that a more technological approach is intrinsically better than one that is less technological [28] ), and 'MacGyver bias' (the inherent attraction of our own personal improvised devices [14] ). We contend that these devices require further evaluation before implementation. We sought to investigate the use of airway boxes and the time to successful intubation of a COVID-19 patient in a simulated environment. We aimed to also explore their effects on pre-oxygenation, laryngoscopy and conventional PPE. We evaluated the impact of two aerosol boxes on tracheal intubation of a simulated patient with severe COVID-19 in the intensive care unit (ICU) in an in-situ simulation crossover study. The simulated process of intubation was designed to comply with the Safe Airway Society COVID-19 airway management guidelines [10] , hereafter referred to as the 'SAS guidelines'. The SAS guidelines are endorsed by 13 specialist colleges, societies and associations in Australia and New Zealand and thus inform the practice of most frontline workers involved in COVID-19 airway management in Australia. They are largely consistent with other international guidelines including the Difficult Airway Society guidelines [11] . The study received ethics approval from the Cabrini Institute Ethics Committee. Written consent was obtained from research participants and additional consent was obtained for the publication of images. Local designers provided the aerosol boxes. Designs for a variety of aerosol boxes have been published online and are being mass produced. We studied two devices; one which closely resembled the original box described by Canelli et al. [16] (the 'early-generation' box) and another which was the most modified and advanced version available to us (the 'latest-generation' box) (Fig. 1 ). The early-generation box resembles that originally designed by Dr Lai Hsien-yung of Taiwan [15] , but with modifications similar to those suggested by an American manufacturer in order to make it suitable for larger body habitus patients [23] . The latestgeneration box represents a local designer's most recent adaption and includes holes for the assistant's hands, a hole on top for insertion of a bougie to aid intubation, and ports for applying suction to generate a negative pressure (not used in this study). On this model, we followed the designer's recommendation that occlusive dressings (Tegaderm Film, 3M, St Paul, MN, USA) cover the holes to create a tighter seal, and a plastic drape be placed over the patient's chest in place of the plastic wall which had been removed in this model. Table 1 . (Table 1) . There were eight breaches of PPE (see also supporting information, Figure S2 ). One was a tear in a gown sleeve and the remainder were gowns pulled back from the glove exposing skin. One breach occurred using the earlygeneration box and the remainder (including the gown tear) occurred using the latest-generation box. All PPE breaches appeared to be due to the gown becoming stuck or held-up at the arm hole, in particular becoming caught on the occlusive dressings used on the latest-generation box. Free-form commentary was sought from participants immediately after each simulated intubation ( Table 2 ). The most frequently mentioned factors relating to the aerosol boxes were discomfort of the arms, back or knees (6 out of 12 participants; 50%) and an increased cognitive load (4 out of 12 participants; 33%). Before a novel medical device is considered for clinical use, it must be assessed for efficacy and safety. In the case of aerosol boxes, efficacy is the ability of the device to provide protection to healthcare workers from infection. Safety of both the patient and staff must also be considered, which includes allowing efficient intubation (to avoid patient harm from hypoxia) as well as the device not causing injury to either the patient or healthcare worker during its use. To our knowledge this is the first formal study of these devices that has been conducted. This study found that use of either of the two different aerosol boxes significantly slowed intubation times when used by experienced airway specialists. First-pass success rates were lower with both boxes, although the study size was small. Contributing factors to procedural difficulty include reduced ability to manipulate devices within the box, reduced arm movement and increased cognitive load. Cognitive overload is recognised to impair decision making and lead to potential patient harm during airway management [29, 30] . The observed delays in intubation are important; the desaturation of patients with COVID-19 on induction of anaesthesia may be rapid and profound The frequency and severity of PPE damage observed in this study was concerning. Seven events occurred with the latest-generation box which may be contributed to by the occlusive dressings used on the arm holes (intended to create a tight seal around the arms), but one event also occurred on the early-generation box without these Figure 3 Comparison of paired intubation times with the latest generation box (abscissa) and the early generation box (ordinate). Unsuccessful intubation was censored and is shown at 180 s. There was no association between the intubation times of the two aerosol boxes. This study examined pre-oxygenation and simple intubation; however, airway management is a process that involves much more than these two procedures. Oropharyngeal suctioning; hyper-angulated videolaryngoscopy; supraglottic airway insertion; patient repositioning; front-of-neck access (cricothyrotomy); and fibreoptic intubation were not examined. These procedures remain untested in these novel devices and should be studied before aerosol boxes can be used safely. Similarly, the implications for management of the morbidly obese patient, the agitated patient or the patient requiring erect pre-oxygenation also need to be studied for safety. This study did not examine the efficacy of aerosol boxes in reducing the viral exposure risk to clinicians. There are no published studies examining whether aerosol boxes reduce healthcare worker viral exposure. This cannot be assumed and research is required. It has been suggested that the box should be removed if difficulty is encountered [16, 17] . We did not allow this in our study but it is likely that a number of boxes would have been removed. We did not examine the process of the emergency removal of the boxes; further investigation would be needed to examine if this was achievable in a timely and safe manner. Concerns with an emergency removal of the box during airway management include patient or healthcare worker injury, dispersal of aerosols and droplets from within the box, and the contamination of healthcare workers from the box surfaces. Finally, we examined the use of these devices by the most experienced airway specialists, being specialist (consultant) anaesthetists. The volume of practice for intubations among intensive care specialists is significantly less [33] . It is possible that a non-anaesthetist group of specialists would experience more difficulty with these techniques. The SARS-CoV-2 pandemic has caused healthcare workers to fear for their personal safety at work and the genuine motivation to protect oneself from harm is easily understood. Nevertheless, this study demonstrates significant patient safety concerns in the use of aerosol boxes for intubation of COVID-19 patients. Furthermore, aerosol boxes may paradoxically increase the risk to clinicians involved in COVID-19 airway management by causing breaches of PPE. Understanding and addressing sources of anxiety among health care professionals during the COVID-19 pandemic Memoriam: healthcare workers who have e46. Epub 17 April Waiting for something positive Aerosol and surface stability of SARS-CoV-2 as compared with SARS-CoV-1 Staff safety during emergency airway management for COVID-19 in Hong Kong Aerosol generating procedures and risk of transmission of acute respiratory infections to healthcare workers: a systematic review Risk factors for SARS transmission from patients requiring intubation: a multicentre investigation in Toronto Consensus statement: Safe Airway Society principles of airway management and tracheal intubation specific to the COVID-19 adult patient group Consensus guidelines for managing the airway in patients with COVID-19: Guidelines from the Difficult Airway Society, the Association of Anaesthetists the Intensive Care Society, the Faculty of Intensive Care Medicine and the Royal College of Anaesthetists Australian doctors design and make life-saving equipment needed for coronavirus pandemic Staying ahead of the wave The MacGyver bias and attraction of homemade devices in healthcare Taiwanese doctor invents device to protect US doctors against coronavirus Barrier enclosure during endotracheal Intubation Aerosol box, an operating room security measure in COVID-19 pandemic Intubation boxes for managing the airway in patients with COVID-19 Aerosol Boxes (Intubation Boxes) Should we use an "aerosol box" for intubation? Learning From Excellence. LFE in COVID19 Anesthesiologist trials taiwanese intubating box Intubation Box is an inexpensive and reusable personal protective device that functions as a barrier between Covid-19 patients and healthcare providers Here's how Shore Plastics spearheaded local production of doc-protecting Aerosol Boxes Waltham company makes aerosol boxes to help prevent spread of COVID-19 among health care workers Limavady firm's aerosol box protects medics from Covid-19 in NHS 'hour of need Minolta looks to help Aus healthcare workers with 3D printing Gizmo idolatry Difficult Airway Society 2015 guidelines for management of unanticipated difficult intubation in adults Cognitive errors detected in anaesthesiology: a literature review and pilot study The Italian coronavirus disease 2019 outbreak: recommendations from clinical practice. Anaesthesia 2020. Epub 27 March Simulating hypoxia and modelling the airway Should ongoing airway education be a mandatory component of continuing professional development for College of Intensive Care Medicine Fellows? No external funding or competing interests declared. Additional supporting information may be found online via the journal website.