key: cord-0998731-eq9gq6ni authors: Wigginton, Krista R; Arts, Peter J; Clack, Herek L; Fitzsimmons, William J; Gamba, Mirko; Harrison, Katherine R; LeBar, William; Lauring, Adam S; Li, Lucinda; Roberts, William W; Rockey, Nicole C; Torreblanca, Jania; Young, Carol; Anderegg, Loïc G; Cohn, Amy M; Doyle, John M; Meisenhelder, Cole M; Raskin, Lutgarde; Love, Nancy G; Kaye, Keith S title: Validation of N95 Filtering Facepiece Respirator Decontamination Methods Available at a Large University Hospital date: 2020-12-15 journal: Open Forum Infect Dis DOI: 10.1093/ofid/ofaa610 sha: c0337e4fab2b620f4164b78c581ed99ab2f4a20a doc_id: 998731 cord_uid: eq9gq6ni BACKGROUND: Due to unprecedented shortages in N95 filtering facepiece respirators, healthcare systems have explored N95 reprocessing. No single, full-scale reprocessing publication has reported an evaluation including multiple viruses, bacteria, and fungi along with respirator filtration and fit. METHODS: We explored reprocessing methods using new 3M 1860 N95 respirators, including moist (50%–75% relative humidity [RH]) heat (80–82°C for 30 minutes), ethylene oxide (EtO), pulsed xenon UV-C (UV-PX), hydrogen peroxide gas plasma (HPGP), and hydrogen peroxide vapor (HPV). Respirator samples were analyzed using 4 viruses (MS2, phi6, influenza A virus [IAV], murine hepatitis virus [MHV)]), 3 bacteria (Escherichia coli, Staphylococcus aureus, Geobacillus stearothermophilus spores, and vegetative bacteria), and Aspergillus niger. Different application media were tested. Decontaminated respirators were evaluated for filtration integrity and fit. RESULTS: Heat with moderate RH most effectively inactivated virus, resulting in reductions of >6.6-log(10) MS2, >6.7-log(10) Phi6, >2.7-log(10) MHV, and >3.9-log(10) IAV and prokaryotes, except for G stearothermohphilus. Hydrogen peroxide vapor was moderately effective at inactivating tested viruses, resulting in 1.5- to >4-log(10) observable inactivation. Staphylococcus aureus inactivation by HPV was limited. Filtration efficiency and proper fit were maintained after 5 cycles of heat with moderate RH and HPV. Although it was effective at decontamination, HPGP resulted in decreased filtration efficiency, and EtO treatment raised toxicity concerns. Observed virus inactivation varied depending upon the application media used. CONCLUSIONS: Both moist heat and HPV are scalable N95 reprocessing options because they achieve high levels of biological indicator inactivation while maintaining respirator fit and integrity. The N95 respirator is the most commonly used filtering facepiece respirator (FFR) and removes at least 95% of airborne particles [1] . When caring for patients infected with severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), the Centers for Disease Control and Prevention (CDC) recommends the use of an N95 or better respirator [2] . Due to unprecedented N95 respirator shortages, many healthcare systems have implemented reprocessing of N95 respirators [3] . Until recently, few studies addressed effectiveness and feasibility of large-scale N95 decontamination and reuse [4] [5] [6] [7] [8] [9] . In March 2020, the CDC issued guidance stating that hydrogen peroxide vapor (HPV), ultraviolet germicidal irradiation, and moist heat were "the most promising FFR decontamination methods" [10] . The US Food and Drug Administration (FDA) issued Emergency Use Authorizations (EUA) for N95 respirators, but important data are lacking with regards to virucidal and bactericidal efficacy of different modalities. The FDA recommendations for decontamination [11] offer a hierarchy of decontamination EUAs, ranging from ≥6-log 10 inactivation of resistant spores or Mycobacterium (Tier 1) to ≥3-log 10 inactivation of a nonenveloped virus or 2 vegetative bacteria (Tier 3). The recommendations do not specify an experimental medium for depositing microorganisms. Most FFR decontamination studies have used single biological indicators [4-6, 8, 12, 13] . Maintaining respirator fit [14] and N95 filtration performance are also critical safety components to consider. The National Institute for Occupational Safety and Health (NIOSH) standard for N95 respirators requires that 95% of particles sized between ~10 and 300 nm are prevented from passing through during respirator challenge tests [1] . We evaluated 5 different FFR decontamination strategies that were deemed practical for the University of Michigan Health System (UMHS) with regards to inactivation of several biological indicators, respirator fit, and filtration efficiency. We also addressed issues of practicality, scalability, and experimental approach. We analyzed several FFR treatment options, including moist heat, EtO, UV-PX, HPGP, and HPV for biological inactivation, filtration integrity, and fit. Biological inactivation was tested only on 3M 1860 respirators, whereas integrity and fit tests were conducted for other respirator types as well (Supplement Table S1 ). Particle filtration efficiency tests were conducted using a custom-built apparatus (Supplement Figure S1 ; Supplement Table S1 ). Assessments were based on differences in the number of NaCl aerosol particles that penetrated FFRs compared with the number used in the challenge test for each particle size and resulted in a calculated filtration efficiency (Supplement Figure S2 ). Filtration testing was conducted at higher flow rates than what is specified by NIOSH standard TEB-APR-STP-0059. This resulted in lower calculated filtration (Supplement Figure S3 ) and more conservative evaluation of FFR integrity. Breathability was assessed using differential pressure measured across FFRs at the test aerosol flow rate. Fit testing was performed by the Occupational Health Services [14] on the same individual preand postdecontamination. Two approaches were tested to achieve moderate RH. First, we used Steris Vision Reliance Washers with controlled local humidity by placing FFR in Ziploc containers [15] . In brief, ~300 µL deionized water was added to a 7.5 × 7.5-cm 2 paper towel in sealed Ziploc medium square polypropylene containers. When run in the dry cycle of the instrument washer at 82°C for 30 minutes (after a 15-to 30-minute warm up), the inside of the container achieved RH between 65% and 75% and 80°C (Supplement Figure S4 ). The second method involved humidity controlled ovens (Supplement Figure S5) . A laboratoryscale oven was initially used (123H; TestEquity, Moorpark, CA). Later, a hospital-scale humidity controlled oven was used for decontamination (Memmert model HCP240). Ovens were equilibrated at the desired RH (40%-90%) and heat (180-190°F) for >30 minutes and then run for a cycle time of 30 minutes. We used a Bioquell Q10 whole room decontamination system. N95 respirators or respirator coupons were placed on racks with clips strung on wire. Treatment consisted of 3 phases: a Gassing phase in which hydrogen peroxide was vaporized and emitted into the room; a Dwell phase in which hydrogen peroxide levels were maintained; and an Aeration phase in which hydrogen peroxide was filtered from air. This process was tested under 2 conditions, with the second test mimicking the FDA EUA certified method (Supplement) [16] . A LightStrike Pulsed Xenon UV lamp (model PXUV4D; Xenex) was used to deliver polychromatic (200-315 nm) wavelengths of UV light across the UV-C and UV-B range (Supplement Table S2 ). Coupons were placed with the outer facing layers toward the light source, 1.8 m from the lamp. For EtO decontamination, we used a 3M Steri-Vac 5XL Sterilizer/Aerator. A low temperature Sterrad 100NX system was used for HPGP treatment (Supplement). We used 4 model viruses (Supplement Table S3 ): 3 bacteria and 1 fungus. The viruses included +ssRNA bacteriophage MS2 and dsRNA bacteriophage Phi6 because they (1) are common human virus indicators [17, 18] , (2) can be produced at high concentrations, (3) have rapid turnaround times for culture-based enumeration, and (4) do not require biosafety level (BSL)2 or BSL3 facilities. We also used a recombinant influenza A H3N2 strain (IAV) that produces firefly luciferase in infected cells because, like SARS-CoV-2, it is an enveloped ssRNA virus that is transmitted via droplets and perhaps aerosols [19] . We used the mouse coronavirus, murine hepatitis virus (MHV), which is in the same genus as SARS-CoV-2. For bacterial inactivation experiments, we used Escherichia coli obtained from the American Type Culture Collection (ATCC 25922), Staphylococcus aureus (ATCC 29213), Geobacillus stereothermophilus (ATCC 12980; both vegetative bacteria and spore tablets [Pharmaceutical Biological Indicator For 6-Log Reduction: HPV-BI; Bioquell Inc., United Kingdom] were used), and Aspergillus niger (a patient isolate) (Supplement). Circular coupons with 1-inch diameter were prepared from 3M 1860 N95 FFRs. Each coupon was weighed and placed in a petri dish. The MHV stock in its culture Dulbecco's modified Eagle's medium ([DMEM]; MHV medium) (Supplement Table S4 ) was deposited directly on coupons. The IAV was deposited in its culture DMEM (IAV medium) (Supplement Table S4 ). The MS2/Phi6 deposition solutions consisted of the combined stock diluted to ~10 10 plaque-forming units (pfu)/mL in either 1× phosphate-buffered saline (PBS) (Supplement Table S5 ) or IAV medium (Supplement Table S4 ). Fifty microliters of virus solutions were deposited on each coupon in 2-µL droplets resulting in a total of ~10 6 TCID 50 influenza, 5 × 10 8 pfu/mL MS2 and phi6, and ~10 4 pfu/mL MHV deposited on each coupon. For the EtO experiment, MS2 was sprayed onto the coupon. Coupons were dried in biosafety cabinets for 1.25 hours. For bacteria and fungus, overnight grown cultures were diluted in saline to a final concentration of 10 6 -10 7 colony-forming units (CFU)/mL. For bacteria and fungus, overnight grown cultures were diluted in saline to a final concentration of 109 CFU/mL. For each prokaryote, 50 μL of a 109 suspension was applied as multiple drops to 1.25 × 0.25-inch rectangular coupons from 3M 1860 N95 FFRs. Coupons were dried before treatment. Each coupon with a specific indicator microorganism had a corresponding no-treatment control. After treatments, biological indicators were extracted and enumerated with infectivity assays. For viruses, coupons were cut into 5-6 small pieces and then deposited in 1.3 mL of extraction solutions, consisting of (1) PBS with 1% bovine serum albumin (BSA) for MS2 and Phi6, (2) IAV medium supplemented with 1% BSA, HEPES, and antibiotics for IAV, and (3) MHV medium for MHV. The tubes were vortexed for 1 minute at half-speed. Viruses in the extract solutions were titered with their respective assays [20] . For bacteria and fungus, coupons were placed into 8 mL trypticase soy broth and vortexed for 10 minutes. Liquid aliquots (1, 10, or ~60 μL) were plated in duplicate in trypticase soy sheep blood agar for all strains except A niger, which was plated on Sabouraud Dextrose Agar. Staphylococcus aureus and E coli were incubated at 35°C, G stereothermophilus was incubated at 56°C, and A niger was incubated at 30°C. Plates were enumerated after 72 hours of incubation. The recoveries of the deposited viruses after drying and extraction ranged from 12% to 113%. For all culture-based measurements, plates with at least 1 infectious unit (CFU) were included in calculations. For plates with no detectable colonies, we assume <1 infection unit for the least diluted sample when calculating concentrations. No-treatment controls were used for log reduction value (LRV) calculations. Filtering facepiece respirators met the NIOSH-approved N95 respirator standard through several cycles of decontamination by moderate RH heat, HPV, EtO, and UV-PX, but not HPGP treatment (Table 1 and Supplement Table S6 ). Although particle sizedependent filtration efficiency varied across cycles, it consistently remained above 95% (Supplement Figure S6 ). Breathability resistance was maintained between 16.5 and 17.5 mm H 2 O, which are the same as new FFRs and higher than other reports [9, 21] . Fit testing was conducted only for decontamination methods that were considered promising, and it passed in all cases across multiple cycles (Table 1 and Supplement Table S6 ). Pulsed Xenon UV-C Treatment resulted in 0.5-to 1.3-log 10 MS2, 0.0-to 2.0-log 10 phi6, 0.8-to 1.7-log 10 IAV, and >1.3-to 1.7-log 10 MHV inactivation ( Figure 1 ). The deposition solution impacted the extent of inactivation, with viruses deposited in IAV medium exhibiting less inactivation than viruses deposited in PBS. This effect was more pronounced for Phi6 than MS2. Prokaryotic results showed that UV-PX yielded zero to 0.3log 10 inactivation for G stereothermophilus, and the Geobacillus spore tablet tested positive after treatment (Table 2) . Aspergillus niger experienced 2.3-log 10 inactivation. Staphylococcus aureus inactivation was 0.5 log 10 for one test and >0.3 log 10 for another, although the maximum detectable LRV was <1 log 10 . Escherichia coli experienced 0.85-to >1.0-log 10 inactivation. Our industrial instrument washers had low RHs, even when wet towels were added ( Figure 2 ). Moderate RH heat was initially achieved using Ziploc containers inside the washers as recently reported [15] . In these containers, the temperature stabilized at 80°C, and RH stabilized between 65% and 75% after ~13 minutes, and these conditions were maintained for ~15 minutes (Supplement Figure S4 ). The conditions resulted in high levels of virus inactivation for all tested viruses (Figure 2A) . MS2, Phi6, IAV, and MHV inactivation was >6.9 log 10 , >7.2 log 10 , >3.4 log 10 , and >.4 log 10 , respectively. Subsequently, UMHS purchased an industrialscale, humidity-controlled oven. A pilot experiment with the humidity-controlled oven at 82°C and 45%-55% RH for 30 minutes resulted in inactivation levels exceeding the dynamic range of each virus ( Figure 2B ); >6.6-log 10 MS2, >6.7log 10 Phi6, >3.9-log 10 IAV, and >2.7-log 10 MHV inactivation. Moderate RH heat decontamination was not effective at inactivating the bacterial thermophile G stereothermophilus ( Table 2) . Geobacillus spore tablets were positive after treatment. Inactivation of A niger was >2.8 log 10 . Staphylococcus aureus inactivation was >2.4 log 10 in the humidity-controlled containers and >3.2 log 10 in the humidity-controlled hospital oven. Escherichia coli colonies were not detected on treated coupons for either method. The Sterrad HPGP system resulted in high levels of virus inactivation (Supplement Figure S7 ). Phi6 and IAV inactivation exceeded the assay dynamic range after treatment, corresponding to >8.2-and >4.0-log 10 inactivation, respectively. MS2 was inactivated by an average of 5.6 log 10 . Respirator integrity declined after 1 cycle and dropped below 95% after 3 cycles ( Table 1) . The Bioquell HPV decontamination system was tested under 2 conditions. Condition 2 was consistent with the FDA EUAapproved protocol [16] , and Condition 1 generated ~40% less vaporous hydrogen peroxide. Virus inactivation was inconsistent after both treatments, ranging from 1.5 log 10 for Phi6 to >4.5 log 10 for MS2 (Figure 3 ). More inactivation was observed when MS2 and Phi6 were deposited in IAV medium than in PBS. Geobacillus spore tablets were negative after treatment with both Conditions 1 and 2 ( Table 2 ). The G stereothermophilus spore culture was inactivated to >1.4 log 10 , >3.0 log 10 , and 2.8 log 10 under conditions 1, 2, and 2, respectively. Escherichia coli exhibited >3.8 log 10 inactivation. Staphylococcus aureus was inactivated only 1.0 log 10 under Condition 1 and >1.6 and >2.3 log 10 under Condition 2. Staphylococcus aureus colonies were detected on all treated coupons. We conducted a single experiment with EtO and achieved > 4.8log reduction of MS2 after treatment (Supplement Figure S8 ). Although effective for inactivating the MS2 virus, we did not expand testing due to concerns about residual EtO and toxicity to FFR wearers. This work provides comprehensive information about multiple N95 FFR decontamination methods evaluated by assessing the impact of treatment method on respirator fit and integrity, the ability to inactivate multiple biological indicators, and the ability to scale each treatment approach to reprocess hundreds of FFRs daily. Three treatment approaches (UV, heat, and HPV) were selected for more intensive assessment under a variety of media matrix conditions. Integrity testing did not reveal significant changes in filtration efficiency, breathability resistance, or fit for most decontamination methods after at least 5 cycles. The exception was HPGP, which reduced filtration integrity beyond standard criteria after 3 cycles. Our results are consistent with prior studies, which showed that moist heat [15] , UVC [22] , and HPV [23] are not harmful to the integrity of 3M 1860 FFRs after at least 5 cycles. Although there are no data on filtration integrity after UV-PX treatment, 3M 1860 FFRs maintained filtration efficiency after 4 cycles when treated with both UV-PX and moderate RH heat (Supplement Table S6 ). All respirators passed fit testing for moderate RH heat and HPV after 5 cycles (Supplement Table S6 ). Using the FDA's EUA recommendation of 3-log 10 virus inactivation as a baseline for decontamination of viruses, UV-PX Table S4 and Supplement Table S5 ). MHV, murine hepatitis virus. alone was insufficient. Others have reported higher MS2 and IAV log 10 reductions on N95 respirators after UVC treatment [4, 5, 8, 24] , although the lamps used in these studies generated primarily UV 254 and provided a larger fluence rate (>1000 mJ/cm 2 ) than the Xenex UV-PX unit, which provided 200-280 nm at a distance of 1.8 m (~24 mJ/cm 2 ). No studies have directly compared inactivation of microorganisms with UVC 254 to UV-PX on N95 respirators. Overall, UV-PX did not effectively inactivate prokaryotic indicators. At moderate RH (50%-70%), heat treatment was effective, inactivating all viruses beyond the dynamic assay ranges. Previous work has demonstrated that IAV and MS2 are susceptible to moist heat treatment [4, 5, 25] . In contrast, the conventional sterilization bacterial spore indicator Geobacillus showed poor inactivation; notably, our heat temperature was well below the condition in autoclaves where this indicator is typically used. Poor inactivation of spore formers with moist heat is a notable limitation, and it has important infection control implications with regards to Clostridium difficile. When the Bioquell HPV system was operated according to the FDA EUA-approved conditions, virus inactivation was less robust then with moist heat. Not all virus indicators exhibited greater than 3-log 10 inactivation. This result is different from that of Kenney et al [26] who report ≥6-log 10 inactivation of Phi6; however, they treated at 19.4 g/m 3 (our study used 12.8 g/m 3 ) with a much shorter dwell time but with more than twice the aeration time. The current study suggests that HPV results may be sensitive to methods used, and it is necessary to provide detailed protocols for comparisons to be made between studies. Although Geobacillus, E coli, and A niger inactivation levels were ~3 log 10 , we could not affirm >3-log 10 inactivation for S aureus, and S aureus colonies were always detected on coupons. There is no standard for S aureus inactivation with HPV treatment, and it is known to be resistant to HPV treatment [27, 28] . The infection control implications of this are important, because S aureus is a nosocomial threat. In addition, S aureus commonly colonizes the nares of healthcare workers [29] , and there is risk that the inner surfaces of respirators can become colonized with S aureus. Experimental protocols vary, and we find that they can significantly influence results. The FDA EUA recommends that treatments achieve >3-log 10 removal of viruses, but it does not give guidance on the application medium used to deposit viruses. The outcomes of this study suggest that achieving >3 log 10 for a specific virus depends not only on the virus tested but also on the deposition solution used. With HPV, MS2 exhibited >3-log 10 removal when it had been dried in IAV medium but <2 log 10 when it had been dried in PBS. Thus, utilization of a cell culture medium for depositing SARS-CoV-2 may lead to overestimating the inactivation that takes place in respiratory droplets. Our complimentary study [20] further explores the role of deposition matrix on virus inactivation by heat. Future studies on decontamination should clearly specify the microorganism deposition media that is used and, if possible, use the appropriate human material (eg, saliva). The methodology in this study, where a diverse set of viruses were analyzed, can be used for evaluation of a wide range of viruses and for potential future pandemic scenarios. Murine hepatitis virus most closely resembles SARS-CoV-2; however, demonstrating 3-log 10 reduction with this virus was technically challenging. The IAV experiments had larger dynamic range, up to 4 log 10 . The bacteriophage surrogates added additional experimental value because they had much higher dynamic ranges, usually >6-log 10 inactivation, and experiments could be carried out quickly, without BSL2 facilities. The results from this study present important issues that should be addressed by future research studies. Based on our results and the results of others, there seems to be a complex relationship between drying conditions, matrix, heat, and humidity, but a mechanistic understanding of how the organisms are inactivated does not yet exist. Mechanistic descriptions of virus inactivation in solution have been published, but they are lacking for viruses on fomites. For example, we observed an enhancement of inactivation when viruses were deposited in their culture medium. Future work should seek to describe the drivers for this enhanced inactivation. Mechanistic understandings of inactivation would help guide more accurate design and evaluation of N95 reprocessing. When evaluating FFR decontamination modalities, hospitals should consider practical and logistical issues, including costs and aspects of implementation and maintenance ( Table 3 ). The UV-PX had the shortest treatment cycle duration and HPV had the longest. The smallest dedicated space was required for moist heat, and the most space was required for HPV. Capital costs will vary by product and purchaser, and personnel effort required for each approach will also vary by site; consequently, 9 A B Inactivation, log 10 Table S4 and Supplement Table S5 ). MHV, murine hepatitis virus. Table S4 and Supplement Table S5 ). MHV, murine hepatitis virus. these costs were not considered in our analysis. With regards to operation and maintenance costs, heat with moderate RH and UV-PX were least expensive, with minimal ongoing costs after purchase. Hydrogen peroxide vapor was the costliest. Ease of implementation was similar for the 3 processes. This research will help hospitals understand the capabilities and limitations of several N95 FFR decontamination approaches. Heat with moderate RH performed the best with regards to pathogen inactivation and was the reprocessing method implemented by our hospital; however, no approach was without limitation (Table 3) . Ultimately, the needs of hospitals are likely to vary with regards to reprocessing volume, available capacity, and infrastructure needed to address N95 decontamination demands. Consequently, we conclude that no single FFR decontamination solution will work across all hospitals. Our study also highlights the importance of methodologic issues (eg, type of deposition media used) in evaluating FFR decontamination modalities. Future studies to validate FFR decontamination should include several viruses and deposit viruses in a range of solutions. Supplementary materials are available at Open Forum Infectious Diseases online. Consisting of data provided by the authors to benefit the reader, the posted materials are not copyedited and are the sole responsibility of the authors, so questions or comments should be addressed to the corresponding author. Centers for Disease Control and Prevention. NIOSH-Approved N95 Particulate Filtering Facepiece Respirators -3M Suppliers List Infection Control: Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) Institution of a novel process for N95 respirator disinfection with vaporized hydrogen peroxide in the setting of the COVID-19 pandemic at a large academic medical center Effectiveness of three decontamination treatments against influenza virus applied to filtering facepiece respirators A pandemic influenza preparedness study: use of energetic methods to decontaminate filtering facepiece respirators contaminated with H1N1 aerosols and droplets Ultraviolet germicidal irradiation of influenza-contaminated N95 filtering facepiece respirators Evaluation of five decontamination methods for filtering facepiece respirators A method to determine the available UV-C dose for the decontamination of filtering facepiece respirators Evaluation of multiple (3-cycle) decontamination processing for filtering facepiece respirators COVID-19 Decontamination and Reuse of Filtering Facepiece Respirators Enforcement Policy for Face Masks and Respirators During the Coronavirus Disease (COVID-19) Public Health Emergency (Revised) Guidance for Industry and Food and Drug Administration Staff Preface Public Comment Cleaning of filtering facepiece respirators contaminated with mucin and Staphylococcus aureus Assessment of N95 respirator decontamination and re-use for SARS-CoV-2 Department of Licensing and Regulatory Affairs. Occupational Health Standards A scalable method of applying heat and humidity for decontamination of N95 respirators during the COVID-19 crisis Battelle CCDS Critical Care Decontamination system Evaluation of the survivability of MS2 viral aerosols deposited on filtering face piece respirator samples incorporating antimicrobial technologies Survival of the enveloped virus Phi6 in droplets as a function of relative humidity, absolute humidity, and temperature Infectious virus in exhaled breath of symptomatic seasonal influenza cases from a college community Humidity and deposition solution play a critical role in virus inactivation by heat treatment on N95 respirators. mSphere (in review Contribution of breathing frequency and inhalation flow rate on performance of n95 filtering facepiece respirators Mitigate a Shortage of Respiratory Protection Devices Title: Research to Mitigate a Shortage of Respiratory Protection Devices During Public Health Emergencies Final Report for the Bioquell Hydrogen Peroxide Vapor (HPV) Decontamination for Reuse of N95 Respirators Development of a test system to evaluate procedures for decontamination of respirators containing viral droplets It's not the heat, it's the humidity: Effectiveness of a rice cooker-steamer for decontamination of cloth and surgical face masks and N95 respirators Hydrogen peroxide vapor sterilization of N95 respirators for reuse Effectiveness of ultraviolet devices and hydrogen peroxide systems for terminal room decontamination: Focus on clinical trials Meticillin-resistant Staphylococcus aureus is more resistant to vaporized hydrogen peroxide than commercial Geobacillus stearothermophilus biological indicators Reduction of nasal Staphylococcus aureus carriage in health care professionals by treatment with a nonantibiotic, alcohol-based nasal antiseptic