key: cord-0992155-89wrq4cd authors: Vaquer, Andreu; Alba-Patiño, Alejandra; Adrover-Jaume, Cristina; Russell, Steven M.; Aranda, María; Borges, Marcio; Mena, Joana; Castillo, Alberto del; Socias, Antonia; Martín, Luisa; Arellano, María Magdalena; Agudo, Miguel; Gonzalez-Freire, Marta; Besalduch, Manuela; Clemente, Antonio; Barón, Enrique; de la Rica, Roberto title: Nanoparticle Transfer Biosensors for the Non-Invasive Detection of SARS-CoV-2 Antigens Trapped in Surgical Face Masks date: 2021-06-24 journal: Sens Actuators B Chem DOI: 10.1016/j.snb.2021.130347 sha: 246651fabc527997f3393f2aa7bf83b58228de53 doc_id: 992155 cord_uid: 89wrq4cd Detecting SARS-CoV-2 antigens in respiratory tract samples has become a widespread method for screening new SARS-CoV-2 infections. This requires a nasopharyngeal swab performed by a trained healthcare worker, which puts strain on saturated healthcare services. In this manuscript we describe a new approach for non-invasive COVID-19 diagnosis. It consists of using mobile biosensors for detecting viral antigens trapped in surgical face masks worn by patients. The biosensors are made of filter paper containing a nanoparticle reservoir. The nanoparticles transfer from the biosensor to the mask on contact, where they generate colorimetric signals that are quantified with a smartphone app. Sample collection requires wearing a surgical mask for 30 minutes, and the total assay time is shorter than 10 min. When tested in a cohort of 27 patients with mild or no symptoms, an area under the receiving operating curve (AUROC) of 0.99 was obtained (96.2% sensitivity and 100% specificity). Serial measurements revealed a high sensitivity and specificity when masks were worn up to 6 days after diagnosis. Surgical face masks are inexpensive and widely available, which makes this approach easy to implement anywhere. The excellent sensitivity, even when tested with asymptomatic patient samples, along with the mobile detection scheme and non-invasive sampling procedure, makes this biosensor design ideal for mass screening. The continued spread of COVID-19 has had devastating social and economic repercussions. Mass screening programs have been implemented by many governments in order to detect presymptomatic and non-symptomatic patients so that contagions can be reduced through contacttracing and self-isolation. Currently, the detection of viral nucleic acids with RT-PCR or the immunodetection of viral antigens are the predominant methods for screening early SARS-CoV-2 infections [1] [2] [3] . RT-PCR is advantageous because it is sufficiently sensitive to detect the virus in pre-symptomatic and non-symptomatic patients. This method, however, requires expensive reagents and can only be performed by trained personnel in laboratories with immobile instruments, which severely limits the number of tests that can be performed on a daily basis [2] . The other predominant screening method, antigen tests, consist of colorimetric lateral flow immunoassays that detect viral proteins. Although their analytical sensitivity is lower than RT-PCR, their low cost and point-of-need operation have made them popular alternatives for mass J o u r n a l P r e -p r o o f screening [2] . In both cases, nasopharyngeal swabs are typically used for sample collection. The drawback of nasopharyngeal swabs is the invasive and uncomfortable sampling procedure that must be performed by a trained healthcare worker [4] . Inadequate sampling may lead to contamination with samples from the upper respiratory tract, which increases the number of false positive results [4, 5] . This puts additional pressure on healthcare providers, who need to divert healthcare workers from patient care to perform COVID-19 tests [6] . Saliva specimens are emerging as a non-invasive alternative to sample gathering for nucleic acid testing [7] [8] [9] [10] [11] . Unfortunately, the antigen tests currently available yield a high number of false negatives when tested on saliva samples [6] . To overcome these limitations, it would be desirable to find a new, non-invasive antigen test with high sensitivity that can be used at the point of need [12, 13] . Such a method could enable mass screening for COVID-19 without overburdening healthcare systems since both sample collection and analysis could be performed by laymen [14, 15] . In this manuscript we propose, for the first in the literature, a non-invasive COVID-19 antigen test based on detecting viral proteins trapped in surgical face masks. Commercial surgical masks contain a hydrophobic polypropylene (PP) layer that provides liquid barrier protection (Fig. 1A ). When wearing a surgical mask, the aerosols and droplets that are produced as the wearer breaths and speaks are captured by this layer [16, 17] . We hypothesized that SARS-CoV-2 antigens would therefore accumulate in this layer as well, and that they could subsequently be detected with nanoparticle transfer biosensors. These biosensors consist of a piece of filter paper modified with a polymer in order to store antibody-decorated gold nanoparticles (Fig. 1B, Inset and Fig. S1 ) [18] . The polymer avoids irreversible interactions between the paper and the nanoparticles in dry conditions, which ensures a complete release of nanoparticles when liquid is added [19] . Paper areas outside the reservoir are modified with wax, which improves the release of nanoparticles J o u r n a l P r e -p r o o f from the reservoir compared to our original design [18] . In traditional lateral flow tests, nanoparticles are transferred horizontally from a reservoir to a detection area [20] [21] [22] [23] [24] . In contrast, our biosensors transfer nanoparticles vertically upon contact with a receiving substrate [18, 19] . While this method was proposed for fabricating paper-only biosensors (e.g. origami biosensors), here we introduce a new application for detecting analytes in exhaled breath by combining the nanoparticle transfer properties of paper immunosensors with the antigen-binding properties of polypropylene face masks (Fig. 1B) . Upon contact with the PP layer of the mask, the nanoparticles specifically bind to viral antigens. After 5 min the paper biosensor is removed and, after a quick washing step, the presence of antigens attached to the mask is evaluated by quantifying the colorimetric signal with a smartphone app (Figs. 1C-1D). The whole process takes less than 10 min to be completed and can be easily performed at the point of need. The app does not require any smartphone modifications or light boxes to yield consistent readings, which is also advantageous for in-field applications [25] . When tested in a cohort of 27 low-and nonsymptomatic patients who wore a surgical mask for 30 min, an area under the receiving operating curve (AUROC) of 0.99 was obtained. Serial measurements revealed an inverse correlation between the days elapsed since diagnosis and the biosensor signal, which indicates that the sensor signal depends on the viral load. Unmodified surgical face masks are used for sample collection, which are inexpensive and widely available. This, along with the rapid assay time and point-ofneed detection scheme make our biosensors promising for non-invasive COVID-19 mass testing. After washing, nanoparticles remain attached to the paper mainly through antibody-antigen interactions; (D) The colorimetric signal S (pixel intensity) is calculated in a few seconds with a smartphone app that fixes the imaging conditions with an interactive augmented reality design that compensates for variations in light conditions; (E) Signal generation mechanism based on the specific recognition of masked-trapped antigens by antibody-decorated nanoparticles. J o u r n a l P r e -p r o o f Whatman filter paper #1 was purchased from GE Healthcare Life Sciences. Wax paper squares (6") A previously described citrate-reducing method was used to obtain gold nanoparticles with a diameter of ca. 40 nm [25] . The citrate around the nanoparticles was substituted for thiol-PEGacid. The carboxylate moieties were transformed into sulfo-NHS esters by adding EDC (1 mg rpm, 6 min). Whatman paper #1 was cut into 1.5 cm x 1.5 cm squares, which were then printed with wax with the following procedure. First, wax paper was cut into 2 cm x 2 cm squares. A 5 mm circle was removed from the centre with a hole puncher. The remaining wax was aligned with the Whatman paper and the paraffin was transferred by pressing a hot iron against it for 5 s. The procedure was repeated on the other side of the paper to completely seal it. The nanoparticle reservoir was fabricated by adding 3 µL of 30% PSS in the unwaxed paper area and letting it dry at room temperature (RT) followed by adding 1 µL of antibody-decorated gold nanoparticles and letting it dry at RT. SARS-CoV-2 Nucleocapsid protein solutions at different final concentrations were prepared in PBS. Then each dilution was transferred to a spray bottle. Surgical masks cut into 1.5 cm x 2 cm squares were sprayed twice with each solution at a fixed 5 cm distance. After drying at RT, the three layers were separated, and the hydrophobic polypropylene layer was hydrated with 1 mL of PBS-BSA. Immediately afterwards, the nanoparticle reservoir was placed onto the surgical facemask filter and held with the aid of a clamp for 5 min. The reservoir was then peeled off and the filter was washed with 1 mL of PBST. The resulting colorimetric signal was measured immediately afterwards with the mobile app as described in previous publications [18, 25] . In brief, the biosensor is placed on a gray card that is used to calibrate the camera and compensate J o u r n a l P r e -p r o o f varying light conditions. A virtual square on the smartphone screen is aligned with the test in order to control the distance and angle. The app automatically segments the paper biosensors from the background and finds the region of interest. Then it starts calculating the pixel intensity and subtracts it from the background calculated at four points around the region of interest. Inconsistent light conditions (e.g. shadows) are controlled for automatically, i.e. the app stops taking measurements if the lighting becomes uneven. This informs the user that the smartphone needs to be realigned, at which point the measurements resume. The data set is evaluated after 50 valid measurements are captured. The resulting colorimetric signal S ranges from 255 (black) to 0 (white). Measurements are obtained in less than 10 s. Healthcare professionals that were self-isolated due to a recent positive COVID-19 test were contacted through the risk prevention service from the Balearic Islands healthcare system. All participants gave informed consent (Ethics committee protocol IB 4251/20 PI). Surgical masks were collected and kept at RT for no longer than 48 h before storing them at -80 o C in the Biobank IdISBa and CIBERES Pulmonary Biobank Consortium, a network currently formed by twelve tertiary Spanish hospitals integrated in the Spanish National Biobanks Network. They were processed following standard procedures with the appropriate approval of the Ethics and Scientific Committees and with the collaboration of the healthcare services of the Hospital Universitario Son Espases and Hospital Son Llàtzer. Two 1.5 cm x 2 cm squares were cut in the centre of each mask. The hydrophobic polypropylene layer was then isolated, and antigens were detected with the method described above. J o u r n a l P r e -p r o o f Statistical analysis was performed using the GraphPad Prism software (San Diego, CA). Correlation between variables was measured using Spearman's correlation coefficient. Diagnostic test accuracy was evaluated by receiver operating characteristic (ROC) curve analysis. A p value <0.05 was considered statistically significant. The detection mechanism in the proposed biosensors is based on dispensing nanoparticles to a receiving substrate through contact transfer. While this concept has been previously demonstrated using paper to absorb the nanoparticles [19] , it has never been shown to work using the PP layer of a face mask as the receiving substrate. Furthermore, the original design used avidin-modified nanoparticles covered with biotinylated antibodies as nanoprobes. The nanosensors described here used secondary antibodies instead of avidin and therefore have different surface properties. These differences in nanoprobe design and substrate material could have a great impact in nanoparticle transference and immunodetection, which are crucial to detect antigens trapped on face masks. After demonstrating that nanoparticles can be transferred to PP substrates, we sought to study the detection of SARS-CoV-2 antigens in liquid droplets. This required dispensing droplets of solutions containing a known concentration of N-antigen onto face masks, which were subsequently analyzed as schematically shown in Fig. 1 . The setup in Fig. 3A was devised in order to generate the droplets. It consisted of placing a piece of face mask (containing all three layers) at a fixed distance of 5 cm from a spray bottle (Fig. S2 ). After spraying face masks twice, the PP J o u r n a l P r e -p r o o f layer was isolated and tested with the procedure schematized in Fig. 1. In Fig. 3B , there is linear relationship between the logarithm of the antigen concentration and the colorimetric signal in the concentration range between 0.3 and 10 2 ng·mL -1 . The limit of detection, expressed as the first sample that yields a signal above two times the standard deviation of the blank, is 3 ng·mL -1 . Control experiments performed by spraying BSA instead of N-antigen yielded lower signals, which demonstrated that color is mainly originated by specific antibody-antigen interactions. These experiments demonstrate that nanoparticle transfer biosensors can detect SARS-CoV-2 antigens present in droplets that have been captured on a surgical face mask. Other proteins were detected with the same procedure in a dose-dependent manner, which suggests that nanoparticle transfer biosensors could be used to detect other relevant biomarkers in droplets trapped by face masks (Fig. S3 ). It should be noted that the limit of detection and sensitivity of this calibration method are only indicative, as it could change depending on the number of times that antigen solutions are sprayed on the face masks, among other factors. Likewise, in real applications, the sensitivity may depend on the time that the patient has worn the mask. Below we explore these points further by analyzing face masks worn by volunteers for different amounts of time. J o u r n a l P r e -p r o o f To study the detection of SARS-CoV-2 antigens trapped in face masks, healthcare workers and their family members who had been recently diagnosed as COVID-19 positive with either RT-PCR or a commercial antigen test were contacted to participate in our study. All participants were self-isolated in their homes with mild (21) or no (6) Fig. 4C face masks worn by the same volunteers on the same day for 120 min yielded three samples below the limit of detection, as well as one COVID-19 negative sample above the limit of detection. The average colorimetric signal for positive patients changed from 35  8 for 30 min to 31  13 for 120 min, which means that measurements were less precise (relative standard deviation RSD 22% and 41%, respectively). Control experiments did not change substantially (11  5 vs 12  6) . In Fig. 4D , the AUC also decreased to 0.92 with an associated sensitivity of 88.5% and specificity of 97.2% for S = 20.3 (positive and negative likelihood ratios are 31.9 and 0.12, respectively). These experiments demonstrate that SARS-CoV-2 antigens can be detected in Next, we studied serial measurements over time with nanoparticle transfer biosensors. In Fig. 5A the colorimetric signal S is plotted as a function of the time elapsed since the official diagnostic J o u r n a l P r e -p r o o f with RT-PCR or a commercial antigen test when masks were worn for 30 min. In this Fig., there is a statistically significant negative correlation between both parameters, as expected from the fact that the viral load decreases as time passes for patients with mild or asymptomatic COVID-19 [26] . The same analysis showed a poor correlation that was not statistically significant when the analysis was made with masks worn by the same patients for 120 min (Fig. S4 ). Fig. 5B shows the AUC analysis for face masks worn by patients within the 3 first days after the first diagnosis, 4-6 days afterwards, and 7 or more days afterwards. In these plots the AUC stays high within the diagnosis. This is summarized in Fig. 5D , where the variation of signal S with time was calculated as the slope of a linear regression plot crossing the three points shown in Fig. 5C . The slope was then plotted as a function of the first day that the mask was worn since diagnosis. In this Fig., there is a statistically significant negative correlation between both parameters. S decreases faster for patients that are diagnosed at a later stage, in agreement with the observation that viral replication is the highest during the first stages of infection [27] . These experiments confirm that there is a relationship between time elapsed since diagnosis and the sensor signal, which is presumably linked to variations in the viral load. J o u r n a l P r e -p r o o f Several aspects of the experiments described here deserve special consideration. The results shown in Fig. 3 demonstrate that the signal is specific and dose-dependent when the only parameter that changes is the concentration of the antigen. The amount of antigen trapped by the mask depends on the distance between the spray and the mask, as well as the number of times that the solution is sprayed onto the mask. As such, the Figs. of merit reported in Fig. 3 are indicative only. The setup does not pretend to fully mimic the way that aerosols and droplets are generated in the human body. A recent paper describes the airborne transmission of SARS-CoV-2 as aerosols and/or droplets [16] . Droplets are larger and emitted when the patient sneezes or coughs, whereas aerosols consist of smaller particles generated by exhaled breath. It should be noted that none of the participants in the study reported cough or sneezing as a symptom, although 26% did report rhinorrhea or nasal congestion. The patient that yielded a signal below the limit of detection in Fig. 4A reported fever and headache symptoms and had been diagnosed with a commercial antigen test one day prior. Masks worn by the same patients in the following days yielded a positive result. The antibodies used in this study for detecting N-antigen differentiate between the same protein produced by SARS-CoV-1 and SARS-CoV-2 ( Fig. S5 ), which means that interference from an infection by another virus from the Coronaviridae family is highly unlikely. Since the patient showed COVID-19 symptoms and the patient tested positive afterwards, the false negative was probably caused by a biosensor malfunction rather than by a low viral load or a false positive result in the original test. Better specificity and sensitivity were always found when masks were worn for only 30 min. It has been proposed that proteins adsorb onto polypropylene fibers following a two-stage mechanism [28] . During the first stage a protein monolayer is formed onto the pristine material, whereas during the second stage more materials adsorb onto the first monolayer. The first layer is strongly bound to PP via hydrophobic interactions, whereas during the second stage additional J o u r n a l P r e -p r o o f proteins are bound via weaker non-specific protein-protein interactions. We hypothesized that when masks are worn for long periods of time, antigens attach to the PP layer following the second mechanism. Since antigens are weakly bound, antibody-decorated nanoparticles recognizing them are easily removed during the washing step. This decreases the signal and increases the variability of the assay as observed in Fig. 4 and Fig. S4 . Additional experiments will be undertaken to determine whether the mask wearing time can be reduced below 30 min. In this study, the time elapsed between the first diagnosis and sample gathering was used as a proxy for studying the evolution of the viral load. Participants in the study were self-confined and were not allowed to travel for RT-PCR analysis. Healthcare resources could not be diverted for nurses to visit them and take a sample. Nevertheless, this cohort of patients allowed us to access chronological data from mild or asymptomatic COVID-19 patients, which are the main target population for mass screening. Our results show that the biosensors signal decreases as time evolves, and that when the diagnosis is delayed the chances of obtaining a negative result increase. However, since we do not have data on actual viral loads it is unknown whether the patients are still contagious when the biosensor signal drops. In conclusion, we have reported a new non-invasive method for COVID-19 diagnosis. It is based on using nanoparticle transfer biosensors for detecting SARS-CoV-2 antigens trapped in surgical face masks. The biosensors show an excellent sensitivity and specificity, even when tested with asymptomatic patients, when masks were worn for only 30 min. The biosensor signal decreases chronologically with serial testing. The total turnaround time is fast (under 10 min) and signals can be read with a smartphone, which is ideal for decentralized, point-of-need mass screenings. Circumventing the need for nasopharyngeal swabs in sample collection ensures an uncomplicated J o u r n a l P r e -p r o o f screening process where participants wear a face mask for 30 min and receive a test result in 10 min. The antigen-based detection mechanism makes them complementary to recent approaches for nucleic acid detection in saliva with mobile biosensors. In the future, the proposed biosensors could be used for detecting antigens from other pathogens as well as biomarkers of inflammation such as interleukins, which have been shown to be present in exhaled breath [29] . ASSOCIATED CONTENT Supplementary material. The manuscript was written through contributions of all authors. All authors have given approval to the final version of the manuscript. ACKNOWLEDGMENT Causative Virus (SARS-CoV-2) in Human Nasopharyngeal Swab Specimens Using Field-Effect Transistor-Based Biosensor Rapid Scaling Up of Covid-19 Diagnostic Testing in the United States -The NIH RADx Initiative Development of a SERS-based lateral flow immunoassay for rapid and ultrasensitive detection of anti-SARS-CoV-2 IgM/IgG in clinical samples SARS-CoV-2 Nasopharyngeal Swab Testing-False-Negative Results From a Pervasive Anatomical Misconception Evaluation of the rapid antigen test Panbio COVID-19 in saliva and nasal swabs in a population-based point-of-care study A smartphone-read ultrasensitive and quantitative saliva test for COVID-19 Saliva as an Alternative Specimen for Molecular COVID-19 Testing in Community Settings and Population-Based Screening Pan, Rapid, ultrasensitive, and quantitative detection of sars-cov-2 using antisense oligonucleotides directed electrochemical biosensor chip Saliva as a Candidate for COVID-19 Diagnostic Testing: J o u r n a l P r e -p r o o f A Meta-Analysis Dual lateral flow optical/chemiluminescence immunosensors for the rapid detection of salivary and serum IgA in patients with COVID-19 disease Rapid diagnostics of coronavirus disease 2019 in early stages using nanobiosensors: Challenges and opportunities Ultraviolet-induced in situ gold nanoparticles for point-of-care testing of infectious diseases in loop-mediated isothermal amplification Recent advances in nanomaterials based biosensors for point of care (PoC) diagnosis of Covid-19 -A minireview Portable magnetic particle spectrometer (MPS) for future rapid and wash-free bioassays Airborne Transmission of COVID-19: Aerosol Dispersion, Lung Deposition, and Virus-Receptor Interactions Multiplexed Nanomaterial-Based Sensor Array for Detection of COVID-19 in Exhaled Breath Paper biosensors for detecting elevated IL-6 levels in blood and respiratory samples from COVID-19 patients Nanoparticle Reservoirs for Paper-Only Immunosensors SARS-CoV-2 Coronavirus Nucleocapsid Antigen-Detecting Half-Strip Lateral Flow Assay Toward the Development of Point of Care Tests Using Commercially Available Reagents Development of a Lateral Flow Strip Membrane Assay for Rapid and Sensitive Detection of the SARS-CoV-2 Rapid and Sensitive Detection of anti-SARS-CoV-2 IgG, Using Lanthanide-Doped Nanoparticles-Based Lateral Flow Immunoassay Development of a Sensitive Immunochromatographic Method Using Lanthanide Fluorescent Microsphere for Rapid Serodiagnosis of COVID-19 A multi-target lateral flow immunoassay enabling the specific and sensitive detection of total antibodies to SARS COV-2 Nanoparticle-Based Mobile Biosensors for the Rapid Detection of Sepsis Biomarkers in Whole Blood Chronological Changes of Viral Shedding in Adult Inpatients With COVID-19 in SARS-CoV-2, SARS-CoV, and MERS-CoV viral load dynamics, duration of viral shedding, and infectiousness: a systematic review and meta-analysis Evaluation of loading characteristics and IgG binding performance of Staphylococcal protein A on polypropylene capillary-channeled polymer fibers Automated in Vivo Nanosensing of Breath-Borne Protein Biomarkers He is currently on the first year of the MSc Science and Chemical technology at the University of the Balearic Islands. He is a member of the Multidisciplinary Sepsis Group Alejandra Alba-Patiño obtained her MSc degree in Chemical Science and Technology from Balearic Islands University where she is currently doing her PhD. Thanks to a scholarship funded by Roche pharma, she is performing her research at the Balearic Islands Health Research Institute (IdISBa) Cristina Adrover obtained her degree in Chemistry in 2019 and her MSc degree in Science and Chemical Technology in 2020, both from the Balearic Islands University (UIB). Currently, she is in the first year of her PhD degree in the Balearic Islands Health Research Institute (IdISBa) His work applies real-time image processing to biosensing. He is also interested in the 3D modelling of biochemical interactions and in developing paper-based microfluidic devices She specialized in Critical Care Medicine at HU Son Llàtzer hospital in 2015 and from 2015 she is working at HU Son Llàtzer as intensivist and is member of Sepsis Unit from its creation Associated Professor in Infectious Diseases by Balearic Islands University (UIB) She is currently working in Hospital Universitari Son Llàtzer of the Balearic Islands Alberto del Castillo obtained his degree in Medicine from University of Salamanca in 2002 and completed the Intensive Care specialty at the Hospital Dr.Peset (Valencia) in 2008. He is currently working in the Intensive Care Unit of the University Hospital Son Llàtzer. He is a member of the Multidisciplinary Sepsis Group She specialized in Critical Care Medicine at Vall d'Hebron hospital and from 2006 she is working at HU Son Llàtzer as intensivist and is member of Sepsis Unit from its creation She specialized in Internal Medicine and Infectious Diseases at Son Espases Hospital and currently, she is working at the Emergency Department of Son Llàtzer Hospital She specialized in Family and Community Medicine at Son Dureta Hospital in Palma de Mallorca After that she started a postdoctoral fellowship at the National Institutes of Health (NIH), USA, for almost 6 years. She has experience in epidemiology studies and biology of aging and inflammation. She is currently a Miguel Servet Investigator, a tenure track position, at the Balearic Islands Health Research Institute She has developed her work life in the Health Service from Balearic Islands and currently, she performs as head of the Prevention Department. of nanomaterials, biosensors and bio-enabled nanofabrication RT-PCR, real-time polymerase chain reaction; IgG, immunoglobulin G; PP, polypropylene; AUC-ROC, area under the ROC curve, where ROC is the receiver operating characteristic curve; COPD chronic obstructive pulmonary disease. Dr. Roberto de la Rica completed his PhD at the National Center for Microelectronics in Barcelona, Spain. He has worked at the City University of New York, at the University of Twente and at Imperial College London. He has experience in nanolithography, synthesis