key: cord-0989696-dp23iq8i authors: Melo, Saulo Maia d’Avila; Macedo, Marília Ferraz de Oliveira; Pereira, Juliana Silva Santana title: Agreement among four portable wireless pulse oximeters and in-office evaluation of peripheral oxygen saturation date: 2021 journal: J Bras Pneumol DOI: 10.36416/1806-3756/e20200251 sha: 208ad81381e62e0033775dc8dd34a83d9dbcabf9 doc_id: 989696 cord_uid: dp23iq8i nan The use of portable pulse oximetry is well-established in the hospital environment. The use of portable wireless pulse oximeters in office visits is growing rapidly, and some patients present oxygen desaturation even when they feel relatively well. (1, 2, 5, 6) Many models of portable wireless pulse oximeters, from different manufacturers and at various prices, are sold in Brazil. Questions have arisen regarding the use of SpO 2 measurement in outpatient settings: What is the minimum time required to obtain a proper SpO 2 reading?; How long before the SpO 2 stabilizes?; and Is there agreement among the readings of the various oximeters used in Brazil? The objective answers to these questions have not been clearly established. Therefore, the objective of the present study was to evaluate the minimum time necessary for the appropriate measurement of SpO 2 , to determine the time necessary for the SpO 2 reading to stabilize, and to evaluate the agreement among the results of four different portable wireless pulse oximeters used in Brazil. This was a cross-sectional study, performed in a physician office and approved by the research ethics committee of the institution (CAAE No. 52677816.3.0000.5371). All the participants signed a free and informed consent form and there were no conflicts of interest. The sample size required in order to identify good agreement (above 0.81) among the four oximeters and to measure SpO 2 was estimated with an error below 0.20, a level of significance of 5%, a confidence interval amplitude of 0.1, and the addition of 10% to compensate for losses and refusals. (7, 8) Thus, we found that a minimum sample of 45 patients would be required in order to evaluate the agreement among the oximeters. We evaluated a convenience sample comprising all volunteers ≥ 18 years of age. Patients with hypotension, hypothermia, digital clubbing, Raynaud's phenomenon, significant anemia, or fever were excluded, as were those with fake or painted nails, those in whom there were hand movement artifacts, and those who had any cognitive deficit that would prevent them from filling out the questionnaire (in the absence of information provided by family members). The SpO 2 data were collected after the volunteers had rested for at least 5 min, during which time they were comfortably seated in a chair, with one hand resting on a table. The four portable wireless pulse oximeters used were as follows: GO 2 Achieve (Nonin Medical, Inc., Plymouth, MN, USA); ChoiceMMed (ChoiceMMed America Corp., Bristol, PA, USA); Rossmax SB100 (Rossmax, Taipei, Taiwan); and Finger Type & Oximeter (Beijing Choice Electronic Technology Co., Beijing, China). The oximeters were placed simultaneously and distributed randomly on the fingertips of the same hand. The SpO 2 was measured at three different time points (30 s, 60 s, and 120 s), determined with a stopwatch and verified photographically. To evaluate the agreement among the oximeters, we chose the highest SpO 2 reading obtained from each device at each of the three time points and calculated the intraclass correlation coefficients (ICCs). (7, 8) The results were interpreted using the criteria established by Landis and Koch, (9) We evaluated 133 patients, of whom 60 (45.1%) were male and 73 (54.9%) were female. The mean age in the sample as a whole was 55.34 ± 18.90 years (95% CI: 52.09-58.88; range, 18-95 years). The SpO 2 measurements did not differ significantly among the time points evaluated (30 s, 60 s, and 120 s), for any one device or among the four oximeters evaluated, and the SpO 2 remained stable for 120 s ( Table 1) . For each oximeter, the highest reading was found at the 120-s time point, and that value was used to evaluate the agreement among the devices. (9) The agreement among the four oximeters was considered excellent (ICC = 0.902; 95% CI: 0.857-0.933). Previous studies have not clearly defined the ideal time to begin reading a portable wireless pulse oximeter or the window of time necessary for the oximeter reading to stabilize. (5, 6, 10) In the present study, we have demonstrated that, in hemodynamically stable patients, the SpO 2 can be read at 30 s and remains stable up to 120 s, with no significant differences in readings among the three time points evaluated (Table 1 ). With the current technical qualification of the new portable pulse oximeters, future studies will be able to evaluate SpO 2 levels with confidence in less than 30 s. In the absence of previous studies on the objectives under discussion, our results can help health professionals (physicians, nurses, and physiotherapists) evaluate the SpO 2 objectively and safely in as little as 30 s, optimizing their outpatient consultation time and allowing a flexible choice in the acquisition of these different devices to be used in the daily practice, with a good cost-benefit ratio. Our study has some limitations. Because of the cross-sectional study design, the durability of the wireless pulse oximetry devices was not evaluated. To evaluate the accuracy of the oximeters used, it would be necessary to compare them with the goldstandard system (arterial blood gas analysis with the determination of the SaO 2 ). We concluded that, in a physician office, SpO 2 can be measured properly in as little as 30 s with any of the oximeters evaluated here. We also found that the reading remains stable for 120 s and that the agreement among the four oximeters was excellent. SMDM: idealization, conception, and planning of the study; interpretation of the data; drafting and revision of the manuscript; and approval of the final version. MFOM: conception and planning of the study; interpretation of the data; and drafting and revision of the manuscript. JSSP: interpretation of the data; and revision of the manuscript. Clinical use of pulse oximetry: official guidelines from the Thoracic Society of Australia and New Zealand British Thoracic Society Emergency Oxygen Guideline Group; BTS Emergency Oxygen Guideline Development Group. BTS guideline for oxygen use in adults in healthcare and emergency settings Protocolo de Manejo Clínico para o Novo Coronavírus (2019-nCoV) World Health Organization. Institutional Repository for Information Sharing World Health Organization; c2020 Clinical management of severe acute respiratory infection (SARI) when COVID-19 disease is suspected: interim guidance Pulse oximetry. Parsons PE, Finlay G Pulse Oximetry Sample size formulas for estimating intraclass correlation coefficients with precision and assurance Agreement analysis in clinical and experimental trials The measurement of observer agreement for categorical data Diretrizes para Testes de Função Pulmonar