key: cord-0982560-c279s4go authors: Russell, Cameron title: Development of a Device to Reduce Oropharyngeal Aerosol Transmission date: 2020-06-07 journal: J Endod DOI: 10.1016/j.joen.2020.05.012 sha: 42ec6655f6442596dbf67835d0f32b740faf504f doc_id: 982560 cord_uid: c279s4go The world-wide COVID-19 pandemic has greatly impacted dental practice. Issues confronting practicing dentists include possible transmission of disease by droplets/ aerosol or contact with contaminated surfaces. Dentists are at increased risk due to their proximity to the oropharynx. In an effort to reduce potential exposure to aerosols generated during treatment, a device has been developed in which a polycarbonate shield is mounted to the dental operating microscope with an attached high velocity vacuum hose. Anemometer measurements demonstrate an exhaust outflow of 3.9 ft/min at a position approximating the patient’s oropharynx. More research may be warranted using this or similar approaches to mitigate aerosol transmission. The global COVID-19 pandemic, due to an exponential increase in SARS-CoV-2 infections [1] , has had a major impact on health care delivery [2, 3] . As of April 2020, the Center for Disease Control and Prevention (CDC) [4] and the American Dental Association [5] issued guidance to postpone all but emergency and urgent dental conditions. The rationale is to prevent disease transmission and to conserve personal protective equipment (PPE). The transmission of the SARS-CoV-2 virus is believed to primarily be via droplets/aerosol originating from the oropharynx or by contact with surfaces contaminated with the virus [2, 6, 7] . The CDC recommends that aerosol generating procedures performed on known or suspected COVID-19 patients should be in negative pressure/ airborne infection isolation rooms (AIIRs) [8] . However, an important additional clinical concern is how to deliver aerosol generating procedures to patients who are infected, but asymptomatic. The onset for COVID-19 infection is up to 14 days and about 80% of patients may be either asymptomatic or have symptoms reminiscent of mild flu or seasonal allergies [9] . Moreover, the growing recognition of nosocomial transmission via aerosol generation has prompted a re-examination of aerosol production in the dental operatory [10] . Given the lack of available negative pressure/AIIRs for dental treatment and the need to deliver endodontic treatment for patients who may or not be infected, there is a need to develop an engineering solution to mitigate aerosol exposure in the dental operatory. Here, is the description of a polycarbonate shield mounted to the dental operating microscope (DOM) with an attached high velocity vacuum hose. A description of its construction and measurement of evacuated airflow is provided. The materials used in the construction of this device are listed in Table 1 . Fig 1 and 2 illustrates the DOM-shield device and its use during endodontic treatment. Shield: Both polycarbonate (Lexan™) and acrylic (Plexiglass™) have advantages and disadvantages in this application. The benefit of Plexiglass™ is that it is more transparent, less inclined to chemical discoloration and tends to be more scratchresistant, but it does not have the same strength as polycarbonate. Polycarbonate has a higher impact resistance, allowing drilling without cracking. An 18X24 rectangular 0.093" thick sheet of clear polycarbonate (Part 1PC1824A; Home Depot) was cut into an 18" diameter disk using a band saw. Next, a template was created by first removing the objective housing from the Global 6A microscope body Shield: After removing the locking ring from the objective lens, the threads of the objective lens were placed into the 2 ½ inch hole in the polycarbonate shield (Fig 2A) . The locking ring was then reattached to the objective lens. The shield and objective lens were then threaded onto the body of the microscope. Note: the DOM positioning handles need to be placed in an upward direction (Fig 1A, B The other end of the flexible ducting was attached to the intake end of the duct fan ( Fig 1E, F) . The ducting was attached to the microscope-mounting pole with two 24-inch cable ties (Fig 1E, F) . The excess flexible ducting was placed on the exhaust end of the duct fan and placed through the drop ceiling, gypsum board, attic and out the roof. An AC Infinity carbon filter (Part AC-DCF4, AC Infinity Inc, City of Industry, CA) was used at the terminus. A suitable HEPA filter was not available due to current government COVID-19 restrictions. The acrylic shield and hosing can be disinfected with 0.5% NaOCl. Prior studies have demonstrated that a 1 min exposure of either 0.1% or 0.5% NaOCl is effective in reducing coronavirus levels ~1,000-fold on a stainless steel surface with serum as an additional contaminant [11] . The velocity of the exhaust system was determined by placing a handheld hot-wire anemometer (TES-1340, TES Electrical Electronic Corp., Taipei, Taiwan) directly under the vacuum inlet and 10" beneath the DOM objective lens (to approximate the location of the patient's oropharynx). The TES-1340 has a dynamic airflow range with a stated accuracy of + 3%. Airflow was measured over a 1 min period with recording of mean airflow. To date, patients have accepted the use of this exhaust device as contributing to both their safety as well as the dentist and staff. While noticeable, the sound of the exhaust is not disturbing. The results of the anemometer indicated that the mean exhaust airflow is 498 ft/min at the vacuum hose inlet. At the DOM objective lens, the mean airflow was 163 ft/min at 2.5" below the lens, 59 ft/min at 5" below the lens and 3.9 ft/min at 10" below the lens (10" is approximately at the patient's oropharynx). The CDC has issued guidelines that confirmed or suspected COVID-19 patients be treated in AIIR facilities. Yet a much broader range of patients who require endodontic treatment may be either uninfected or infected and asysmptomatic. To address this larger population, a device was developed with the potential to reduce oropharyngeal aerosol transmission. Measurements of airflow indicate that this device generated an exhaust airflow of about 3.9 ft/min at the approximate location of the patient's airway. The use of this device may mitigate aerosol transmission, serves as a sneeze guard, is accepted by patients, does not interfere with the delivery of endodontic treatment and is relatively inexpensive. Moreover, mitigation in aerosol transmission may have many benefits in reducing exposure to other airborne pathogens to the dentist and staff. However, there are limitations to this device. First, it does not replace the AIIR recommended by the CDC for treating COVID-19 confirmed or suspected patients. Second, the demonstration of substantial exhaust airflow from the region of the oropharynx is consistent, but not sufficient to establish a significant reduction in aerosol transmission. This latter point should be evaluated in future research in a proper research environment [12] . However, this device, coupled with appropriate PPE may offer benefits in delivery of endodontic treatment. Report of the WHO-China Joint Missionon Coronavirus Disease Rolling Update on Coronavirus Diease (CVVID-19) CDC Guidance for providing dental care during COVID-19 ADA urges dentists to heed April 30 interim postponement recommendation Aerosol and Surface Stability of SARS-CoV-2 as Compared with SARS-CoV-1 How COVID-19 Spreads Interim Infection Prevention and Control Recommendations for Patients with Suspected or Confirmed Coronavirus Disease 2019 (COVID-19) in Healthcare Settings Characteristics of and Important Lessons From the Coronavirus Disease 2019 (COVID-19) Outbreak in China: Summary of a Report of 72314 Cases From the Chinese Center for Disease Control and Prevention Insights into the May issue of the Journal of Endodontics Persistence of coronaviruses on inanimate surfaces and their inactivation with biocidal agents Recognition of aerosol transmission of infectious agents: a commentary