key: cord-0979976-r7l1mdz8 authors: Sander, Ingrid; Kespohl, Sabine; Zahradnik, Eva; Göcke, Philipp; Hosbach, Ingolf; Herrmann, Burkhard L; Brüning, Thomas; Raulf, Monika title: Quantitative measurement of IgG to SARS‐CoV‐2 antigens using monoclonal antibody‐based enzyme‐linked immunosorbent assays date: 2022-01-30 journal: Clin Transl Immunology DOI: 10.1002/cti2.1369 sha: 317f6138da283bb8b77b1814952c150569601ace doc_id: 979976 cord_uid: r7l1mdz8 OBJECTIVE: Standardised quantitative analysis of the humoral immune response to SARS‐CoV‐2 antigens may be useful for estimating the extent and duration of immunity. The aim was to develop enzyme‐linked immunosorbent assays (ELISAs) for the quantification of human IgG antibodies against SARS‐CoV‐2 antigens. METHODS: Enzyme‐linked immunosorbent assays were developed based on monoclonal antibodies against human IgG and recombinant SARS‐CoV‐2 antigens (Spike‐S1 and Nucleocapsid). The WHO 67/086 immunoglobulin and WHO 20/136 SARS‐CoV‐2 references were used for standardisation. Sera of a study group of COVID‐19‐positive subjects (n = 144), pre‐pandemic controls (n = 135) and individuals vaccinated with BioNTech–Pfizer BNT162b2 vaccine (n = 48) were analysed. The study group sera were also tested using EuroImmun SARS‐CoV‐2‐ELISAs and a quantitative S1‐specific fluorescence enzyme immunoassay (FEIA) from Thermo Fisher. RESULTS: The ELISA results were repeatable and traceable to international units because of their parallelism to both WHO references. In the study group, median anti‐S1‐IgG concentrations were 102 BAU mL(−1), compared to 100 and 1457 BAU mL(−1) in the vaccination group after first and second vaccination, respectively. The ELISAs achieved an area under the curve (AUC) of 0.965 (S1) and 0.955 (Nucleocapsid) in receiver operating characteristic (ROC) analysis, and a specificity of 1 (S1) and 0.963 (Nucleocapsid) and sensitivity of 0.903 (S1) and 0.833 (Nucleocapsid) at the maximum Youden index. In comparison, the commercial assays (S1‐FEIA, S1 and Nucleocapsid ELISA EuroImmun) achieved sensitivities of 0.764, 0.875 and 0.882 in the study group, respectively. CONCLUSIONS: The quantitative ELISAs to measure IgG binding to SARS‐CoV‐2 antigens have good analytical and clinical performance characteristics and units traceable to international standards. Specific detection of the immune response to SARS-CoV-2 infection plays an important role in tracking the spread of COVID-19 in the population. [1] [2] [3] [4] [5] [6] [7] IgG antibodies in particular remain detectable after the end of the disease or even after infections without symptoms. 8, 9 Numerous immunoassays have been developed to detect IgG to the SARS-CoV-2 antigens, 10, 11 some of which are characterised by high specificity and sensitivity. [12] [13] [14] [15] [16] Both the spike protein that binds to the receptors on the host cell for virus entry and the nucleocapsid protein required for virus replication are recognised by human antibodies and serve as antigens for the serological detection of infection. 17, 18 Normally, the nucleocapsid protein is present in cells at higher copy numbers than the spike protein 19 and is responsible for the high sensitivity of immunoassays based on this antigen due to its high immunogenicity. The spike protein is located on the surface of the enveloped RNA virus and consists of trimers of two glycosylated subunits (S1 and S2). The S1 subunit contains the receptor-binding region (RBD) and shows less homology to other coronaviruses than the S2 protein. 20 Because antibodies with proven neutralising activity bind to the spike protein and the RBD in particular, [21] [22] [23] the spike protein is also the major antigen target for vaccines. After the first vaccination, antibody responses are elicited and then boosted with the second vaccine dose to protein-binding IgG levels above those of human convalescent sera. 24, 25 Specific IgG antibody titres decrease over time but remain elevated up to 6 months after a complete vaccination. 26, 27 The long-term persistence of SARS-CoV-2-specific antibodies is unclear as is whether the presence of antibodies confers protective immunity against SARS-CoV-2. 28 As the vaccination campaign progresses and breakthrough infections occur even after full vaccination, 29 immunoassays for the quantitative detection of anti-SARS-CoV-2 antibodies are becoming increasingly import. On the one hand, a significant increase in antibody concentration indicates successful vaccination; on the other hand, the antibody level may indicate the probability and duration of immune protection and sterilising immunity. Neutralising antibody titres were found to be strongly associated with the magnitude of the IgG response [30] [31] [32] [33] and were highly predictive of the immune protection from symptomatic SARS-CoV-2 infection. 34 Quantitative immunoassays are part of the standard repertoire in serology diagnostics and research. For example, the quantification of antigen-specific human IgG antibodies is an important component in the diagnosis of hypersensitivity pneumonitis. 35 Antigen-specific IgG concentrations are determined using a calibration curve traceable to international human immunoglobulin standards, 36 thus enabling values to be compared across laboratories if the same antigens and assay components are used. In the case of anti-SARS-CoV-2 immunoglobulin, an international standard now exists 37 that has been evaluated in a WHO collaborative study. 38 The aim of this study was to develop quantitative enzyme-linked immunosorbent assays (ELISA) in order to detect the binding of human IgG to the Spike-S1 and nucleocapsid virus proteins. The measurement of antibodies against both immunodominant SARS-CoV-2 antigens usually allows for the differentiation of the immune response after vaccination and infection, as almost all licenced vaccines target the spike protein. In order to validate the developed ELISAs, a study group of individuals with positive SARS-CoV-2 PCR test results (true positives) and a control group with sera taken before the COVID-19 pandemic (true negatives) were used for ROC analysis. The results of the study group using our newly developed ELISAs were compared to those of commercially available anti-SARS-CoV-2 IgG immunoassays from EuroImmun, L€ ubeck, Germany and Thermo Fisher Scientific, Darmstadt, Germany. For standardisation and to calculate conversion factors, international WHO references for immunoglobulins 36 and SARS-CoV-2-specific antibodies 37 were used. Finally, sera of vaccinated individuals were tested with the new ELISAs before and after vaccination. Sera classified as positive by PCR test (study group) and pre-pandemic sera (control group) were used for immunoassay development and validation ( Table 1 ). The participants in the study group were older than those in the control group (mean difference was 8 years). For the quantification of human IgG, wellcharacterised monoclonal antibodies were used, which bind to all IgG subclasses with high specificity and affinity. 39 A biotinylated version of mAb HP6017, which binds to the Fc part of human IgG, was used as the detection antibody. For the reference curve, the capture antibody mAb HP6045 was used, which binds to the Fd part of IgG. A steep sigmoidal curve with the inflection point at approximately 0.12 µg A mL À1 was obtained using a pool serum with known IgG concentration. The dilution series of the WHO 67/ 086 Immunoglobulin (Ig) reference preparation and the WHO 20/136 SARS-CoV-2 reference binding to the Spike-S1 HEK or Nucleocapsid E. coli antigens resulted in parallel curves ( Figure 1 ). The WHO 20/136 reference pool showed higher binding to the Nucleocapsid compared to the Spike-S1 antigen. Parallelism to the reference curve was controlled by the intra-assay coefficient of variation (CV). The CV% of repeated experiments and of the study groups' dilution series are presented in Table 2 . The intra-assay CV for the WHO references dilution series was 5% or lower, and the inter-assay CV was below 8%. Because of the parallelism of the curves, conversion factors could be calculated ( Table 3) . The WHO 67/87 Ig reference was measured to contain 8482 µg A mL À1 (8197-8768 µg A mL À1 95% confidence interval CI), and the WHO 20/136 SARS-CoV-2 reference had a binding value of 40 µg A mL À1 (37-43 µg A mL À1 95% CI) to S1 HEK and 150 µg A mL À1 (142-158 µg A mL À1 95% CI) to the Nucleocapsid E. coli . One mg A IgG of the laboratory reference pool is equivalent to 11.13 IU of the WHO Ig reference and to the binding activity units (BAU), which are shown in Table 3 . Furthermore, the results of the study groups' dilution series and the repeatability of the results of rerun samples were analysed ( Table 2 ). The intraassay CV did not differ significantly between the different antigens and was at mean 6.80% (6.29-7.31% 95% CI) for the measurements to all three antigens in the study group. The variation among the low values (< 5 mg A L À1 ) was significantly higher than that of the high or medium values. With respect to repeatability, the medium values had the lowest CV, but the differences between high, median and low values were not significant. According to the European Medicines Agency, a CV of 15% or lower is acceptable for repeatability. 40 The ELISAs clearly met the The IgG results of the study group and the controls including ROC analyses are shown in Figure 2 . The AUC for anti-S1 HEK IgG, anti-Nucleocapsid HEK IgG and anti-Nucleocapsid E. coli IgG was 0.965, 0.890 and 0.955, respectively. The cut-off values for optimal discrimination between controls and the study group were estimated by the maximal YI (YI max ), which were 0.903, 0.667 and 0.797 for anti-S1 HEK IgG, anti-Nucleocapsid HEK IgG and anti-Nucleocapsid E. coli IgG, respectively. IgG values to nucleocapsids were higher in the study group as well as in the control group compared to the values in response to the Spike-S1 antigen. The maxima reached 1522, 890 and 300 mg A L À1 , and the 90% quantiles were at 68, 55 and 35 mg A L À1 for the Nucleocapsid E. coli , the Nucleocapsid HEK and the Spike-S1 Hek , respectively. There were no significant differences in IgG concentrations between the subjects with symptoms and those where information on symptoms was lacking (data not shown). For comparison, all sera from the study group participants were also tested using the semiquantitative IgG anti-SARS-CoV-2 ELISAs from EuroImmun and the quantitative Spike-S1 EliA FEIA from Thermo Fisher. The sensitivities of the immunoassays are shown in Table 4 . The Spike-S1 HEK ELISA had a similar sensitivity as the EuroImmun S1 ELISA, but a higher sensitivity than the Spike-S1 EliA. The sensitivity of the Nucleocapsid E. coli was lower than that of the EuroImmun Nucleocapsid ELISA. When the anti-S1 HEK results were combined with the anti-Nucleocapsid E. coli IgG results (positive if at least one test positive), sensitivity was 94.4% and specificity was 96.3%. Spearman correlations between all assay results in the study group were calculated, and scatter plots of the IgG values are shown in Figure 3 . All assay results correlated highly significantly with one another. The correlation coefficients between anti-S1 IgG results were in the range of 0.89-0.93 and between anti-Nucleocapsid IgG results in the range of 0.73-0.87 ( Figure 3a) . Conversely, the correlation coefficients between S1 and Nucleocapsid IgG results were between 0.58 and (Figure 3b and c) ; however, the semi-quantitative EuroImmun ELISA became saturated at ratios around 10. In most cases above the cut-off, the nucleocapsid expressed in Escherichia coli resulted in higher values than the nucleocapsid expressed in HEK cells (Figure 3c ). The anti-Nucleocapsid E. coli and anti-Spike-S1 HEK IgG ELISAs were used for quantification of antibodies after vaccination of 10 previously PCRpositive individuals and a group of 48 volunteers without SARS-CoV-2 infection. The basic data of these groups are presented in Table 5 . Only four people did not develop anti-S1 antibodies after the 1 st vaccination and only one still remained negative after the 2 nd vaccination with BNT162b2. The IgG results were converted to BAU mL À1 by the factors shown in Table 3 . IgG to S1 increased significantly after 1 st and 2 nd vaccination in the vaccinated group (Figure 4a ) in contrast to IgG to the nucleocapsid which remained low (Figure 4b ). On average, the anti-Spike-S1 IgG concentration was 23 times higher after the 2 nd vaccination than after the 1 st vaccination. Antibody concentrations tended to be lower with increasing age of the vaccinated. This effect was significant in the linear regression models of log-transformed values as a function of age and more pronounced for anti-Spike-S1 IgG after the 1 st vaccination (P-value = 0.0074) than after the 2 nd vaccination (P-value = 0.0172). The 10 individuals vaccinated after SARS-CoV-2 infection had even higher mean and median anti-Spike-S1 IgG antibody concentrations than the vaccinated group after the second vaccination. The median IgG level of the 10 individuals after their positive PCR test had been 151 BAU mL À1 (range 13-2144 BAU mL À1 ) to Spike-S1 and 129 BAU mL À1 (range 11-833 BAU mL À1 ) to the Nucleocapsid (data not shown). Quantification of the immune response to SARS-CoV-2 may be used as an indicator of immunity. Analysis of the T-cell response or of neutralising antibodies requires specialised laboratories, sometimes also with increased safety measures, whereas the measurement of antibodies by ELISA is a standard method that can be performed in most immunological laboratories. In SARS-CoV-2 infections, different classes of immunoglobulins are produced almost simultaneously within the first weeks after infection. 17, [41] [42] [43] Since the titres for IgG were the highest, longest lasting and proportional to the neutralising activity, [30] [31] [32] 41 we restricted our ELISA development to this immunoglobulin class. Currently, there exist wellcharacterised mAbs for human IgG that are optimal for standardised detection in immunoassays. 39, 44 We combined two pansubclass-IgG-specific mAbs, with the Fd-specific HP6045 as the capture antibody in the reference curve, and the Fc-specific HP6017 for detection. Thus, the epitope for mAb binding on the human IgG antibodies should be distant from the paratopes for binding to SARS-CoV-2 epitopes in order to avoid steric hindrance. Using an internal serum pool with known total IgG concentrations for the reference curve, we obtained results in mg A L À1 . The 'A' in the index is intended to indicate that the IgG concentrations are antigenspecific and depend on the avidity of IgG and affinity of mAbs. Strictly speaking, it is not the antibody concentration that is measured, but rather the antibody-binding capacity. Because of the parallelism to the binding curve of the WHO At the start of our project, we used the sequence of the original Wuhan virus, as well as the spike and nucleocapsid antigens, which were available at the time, as the basis to develop our ELISA method. However, SARS-CoV-2 antigen selection is . Correlations between immunoassay results and scatter plots. Heat map (a) of Spearman correlations (r S ) between the study group's immunoassay results and scatter plots of Spike-S1 immunoassay results (b) and nucleocapsid results (c). The cut-off values (see Table 4 ) for positivity and borderline results in EuroImmun ELISAs and in the Spike-S1 EliA are indicated by dashed horizontal lines. The cut-off values for the IPA ELISAs are shown with solid lines, the LODs are indicated using dotted lines, and the line of identity is shown for the Nucleocapsid ELISAs from IPA. one of the crucial aspects of assay development. 28 Recombinant proteins are produced either by prokaryotic or by eukaryotic expression systems. To include post-translational modifications, such as glycosylation or phosphorylation, we acquired recombinant S1 and nucleocapsid SARS-CoV-2 antigens that were expressed in a human cell line, which contained a His-tag and were affinitypurified. Mean intra-assay CVs of 7% and interassay CVs of 6-14% revealed parallelism to the IgG reference curve and high repeatability, respectively. However, the clinical performance by ROC analyses was better for the S1 ELISA (AUC = 0.965) compared to the Nucleocapsid HEK ELISA (AUC = 0.890). Therefore, we used an additional antigena commercially available nucleocapsid that was expressed in E. coli. We observed improved clinical performance than in the Nucleocapsid HEK ELISA with an AUC of 0.955, and a specificity of 96.3% and sensitivity of 83.3%. Importantly combining the anti-S1 HEK and anti-Nucleocapsid E. coli , IgG results improved sensitivity to 94.4%. The criterion for inclusion into our study group was a positive PCR test. In contrast, COVID-19 patients including those who were hospitalised were selected for the positive collective in many other validation studies of antibody tests. 15, 16, 43, 46 It was observed that severely ill patients had higher antibody titers, 7, 43, 47, 48 while not all PCRpositive patients developed an immune response. 49 Consequently, a direct comparison of test sensitivities reported in the literature is of limited value. Therefore, we examined our study group with commercial IgG tests that were feasible with our laboratory equipment. These included semiquantitative ELISAs from the company EuroImmun, which were validated and used in several previous studies, 12, 15, 16, 32, 43, 48, 50 and the quantitative FEIA from Thermo Fisher, which can be processed using the automated Phadia 250 system. All test results correlated highly significantly with one another. In particular, the newly developed Spike-S1 ELISA was equivalent or even superior to the commercial immunoassays in terms of clinical performance. In contrast, the Nucleocapsid E. coli ELISA did not quite reach the clinical sensitivity of the EuroImmun ELISA. A comparatively high non-specific binding of the Nucleocapsid E. coli ELISA led to a significantly higher cut-off value compared to the S1 ELISA, and the clinical specificity was also suboptimal. It is possible that minimal cross reactions to endemic human beta-coronaviruses are responsible for the reactivity in the controls; alternately, the degree of purity of the recombinant protein was too low. However, it is known that a modified protein is used in the Euroimmun Nucleocapsid ELISA, and as a result, a high specificity could be achieved. 48 While the results of the different immunoassays were highly correlative, the lack of agreement among the values was expected because of their which are also traceable to international references, allow for comparability. This was addressed in recently published work in which IgG was measured using ImmunoCAP. 46 The authors reported a median level of 34.7 µg mL À1 to the SARS-CoV-2 spike-RBD and 24.5 µg mL À1 to a SARS-CoV-2 nucleocapsid (both proteins were expressed in HEK cells, biotinylated and bound to Streptavidin ImmunoCAPs) in 36 patients who were evaluated at a follow-up subsequent to their hospital stay. These median values are higher than those obtained with the subjects in our study, which is plausible because of the generally more severe disease experienced by hospitalised patients. Other factors that can also significantly influence the results include the different antigens used and how they are bound to the solid surfaces. A comparison of five different commercial quantitative immunoassays to quantify IgG levels against the SARS-CoV-2 spike protein showed good overall agreement, but they were not interchangeable, even when converted to BAU mL À1 using the WHO international standard for SARS-CoV-2 immunoglobulin. 51 According to the authors and in agreement with our results, this highlights the need for further standardisation of SARS-CoV-2 serology. anti-SARS-CoV2-Spike S1 IgG [BAU mL -1 ] anti-SARS-CoV2-Spike S1 IgG [mg A L -1 ] The newly developed ELISAs were used to study antibody responses of 48 healthy volunteers before and after vaccination with BioNTech-Pfizer BNT162b2 vaccine. After 1 st and 2 nd vaccination, 92% and 98% of the participants had anti-spike S1 IgG (medians 100 BAU mL À1 and 1457 BAU mL À1 , respectively). Ten individuals previously infected with COVID-19 developed even higher antibody concentrations after vaccination (median 2438 BAU mL À1 ). Similar strong immune responses after vaccination with BNT162b2 have also been published previously. 52,53 A limitation of our results after vaccination is the small number of samples, especially in the previously infected group. A robust correlation was seen between binding antibody titres and efficacy across different vaccines. 54 Initial correlates of antibody concentrations sufficient for protection against COVID-19 were reported 55 : The anti-spike IgG level associated with 80% vaccine efficacy against primary symptomatic SARS-CoV-2 infection during a 4-to 6month period was estimated to be 264 BAU mL À1 . However, the analyses by Feng et al. 55 had been conducted on samples taken 28 days after two doses of Oxford-AstraZeneca ChAdOx1 nCoV-19 and might not apply to protection afforded by the BNT162b2 vaccine used in our study. The ELISA method for the quantification of specific human IgG showed good intra-and inter-assay reproducibility and parallelism to international reference sera. Furthermore, the clinical performance with the S1 spike protein expressed in HEK cells was not inferior to commercial test systems. However, improvements are required for the Nucleocapsid ELISA. One advantage of the ELISA method presented here is that adaptation to new antigens, for example mutated variants of the spike protein, can be easily implemented. In contrast to other methods for antibody measurements against SARS-CoV-2, 56,57 no SARS-CoV-2-specific components are used apart from the SARS-CoV-2 antigen. In addition, the instruments used here are all part of the standard equipment available in most immunological laboratories, which makes the method suitable for many research laboratories. In order to directly compare values from different laboratories, however, standardisation of the SARS-CoV-2 antigens is necessary, as is the use of international reference sera. The immunoassays could then be used to determine vaccinedependent estimates of protective antibody concentrations, which will then represent a simpler approximation compared to neutralising antibodies. The study was carried out in accordance with the Code of Fifty sera samples from individuals (age 19-64, mean 42 years), collected between July 2006 and September 2007, were pooled, frozen in aliquots at À20°C and used as an internal laboratory reference. The total IgG concentration of this serum pool was previously measured by an external accredited medical laboratory and reported as 11.6 mg mL À1 (6.17 mg mL À1 IgG1, 4.38 mg mL À1 IgG2, 0.67 mg mL À1 IgG3 and 0.38 mg mL À1 IgG4). Recombinant SARS-CoV-2 S1 protein (Val16-Arg685 from NCBI sequence: YP_009724390.1) with a poly-histidine tag, expressed in human embryonic kidney 293 (HEK-293) cells, was purchased from Sino Biological, Eschborn, Germany (40591-V08H); recombinant SARS-CoV-2 nucleocapsid (N) protein (Met1-Ala419 from GenBank sequence QHD43423) with a C-terminal poly-histidine tag expressed in HEK-293 cells was purchased from RayBiotech (Peachtree Corners, GA, USA, 230-30164); and recombinant SARS-CoV-2 nucleocapsid (N) protein (Met1-Ala419 from GenBank sequence QHD43423) with a C-terminal serine and polyhistidine tag expressed in E. coli cells was purchased from Trenzyme, Konstanz, Germany (P2020). The following commercial anti-SARS-CoV-2 IgG immunoassays were used according to the manufacturers' instruction: anti-S1 IgG ELISA (Anti-SARS-CoV-2, EI 2606-9601 G; Euroimmun), anti-Nucleocapsid IgG ELISA (Anti-SARS-CoV-2-NCP, EI 2606-9620-2 G, Euroimmun) and EliA SARS-CoV-2-Sp1 IgG Fluorescence Enzyme Immunoassay (FEIA) on Phadia 250 instrument (Thermo Fisher Scientific, Freiburg, Germany). If the sample concentrations exceeded the range of the EliA assay, then the samples were manually pre-diluted at 1:10. For the IPA-established ELISAs, three strips (F8) of Nunc 96-well Flat MaxiSorp Immunoplates (Thermo Fisher Scientific) were coated with 100 µL per well (400 ng per well) of the pan-anti-human-FD-IgG monoclonal antibody HP6045 (05-4500, Thermo Fisher Scientific) for the calibration curve, and a further nine strips were coated with 100 µL of the SARS-CoV-2 antigen (250-400 ng per well) in 0.1 M carbonate-bicarbonate buffer pH 9.6 overnight at 4°C. On the following day, plates were blocked with 1% Tween-20 in phosphate-buffered saline (PBS) for 1 h at 22°C. The calibration curve was obtained by adding 100 µL of 10 serial 3/5 dilutions of the internal laboratory reference pool serum with IgG concentrations ranging from 5.85 to 580 ng mL À1 to the wells coated with anti-human IgG antibody. Each serum sample was tested using three serial 3/5 dilutions, which were added to the wells with the SARS-CoV-2 antigens. The start sample was diluted 1:100 using PBS with 0.05% Tween-20 (PBST) for both the references and samples. Only in the case of the SARS-CoV-2 nucleocapsid expressed in E. coli, the sample and reference dilution buffer additionally contained 20 µg mL À1 E. coli protein to reduce unspecific binding. After incubation for 1 h at 22°C, the bound human IgG was detected with 100 µL biotinylated pan-anti-human-Fc-IgG antibody HP6017 (05-4240, Thermo Fisher Scientific) diluted 1:1000 in PBST, followed by 100 µL/well of streptavidin-peroxidase conjugate (S5512, Sigma, Steinheim, Germany; diluted 1/20 000 with PBST) for 1 h at 22°C and finally 100 µL ABTS substrate [2,20-azinobis(3ethylbenzothiazoline-6-sulfonic acid) diammonium salt; Sigma] in 50 mM phosphate-citrate buffer, pH 4.2, with 0.015% hydrogen peroxide. The reaction was stopped with 0.32% sodium fluoride, and the absorbance was read at 414 nm in a SpectraMax M2 (Molecular Devices, Sunnyvale, CA, USA). Sample concentrations were calculated by interpolation of optical density (OD) values on a fourparameter fitted reference curve using Softmax Pro 7.0.3 (Molecular Devices) and reported as ng A mL À1 . The lower limit of detection (LOD) was the concentration corresponding to OD 414 = 0.05 above the minimal value of the four-parameter curve fit function (parameter A) which corresponds to a mean of 5 ng A mL À1 . The upper limit depended on the linear range of the sigmoidal curve corresponding at mean to 340 ng A mL À1 . In case, two or more of the sample dilutions were above the upper detection limit, and then, the sample was rerun using higher dilutions. The measurements were also repeated if the coefficient of variation of the calculated concentrations of the sample dilutions in range was greater than 25%. To investigate parallelism between the IgG calibration curve and the sample dilutions, the intra-assay CV value of the sample dilution series was calculated for the newly developed ELISAs. Samples with results > 15 mg A L À1 were considered 'high', results > 5 and ≤ 15 mg A L À1 were considered 'medium', and results ≤ 5 mg A L À1 were considered 'low'. To investigate the repeatability of the Spike-S1 HEK IPA and Nucleocapsid E. coli IPA ELISA, 21 samples with previously high, medium and low results were repeatedly tested on different plates and days and the mean inter-assay CV was calculated. For standardisation, the WHO international standard for human immunoglobulin G, A, M 36 and the first WHO international standard anti-SARS-CoV-2 for human immunoglobulin 37 [67/086 and 20/136, respectively, National Institute for Biological Standards and Control (NIBSC), Hertfordshire, UK] were used. The solution for the former (WHO 67/086) was prepared by dissolving the contents of one ampoule in 1 mL of distilled water with a resulting concentration of 94.4 IU mL À1 , 36 and the WHO 20/ 136 solution was prepared by dissolving the contents of one ampoule in 250 µL distilled water with a reported resulting concentration of 1000 BAU mL À1 with regard to binding to SARS-CoV-2 antigens. 37 Parallelism and repeatability were tested as described above, and converting factors between mg A L À1 to IU L À1 and BAU L À1 were calculated. Statistical analyses using Mann-Whitney and Kruskal-Wallis tests, two-way ANOVA with Bonferroni multiple comparison, calculation of confidence intervals (CI) and Spearman correlations, linear regression of log-transformed values and ROC analyses were performed with GraphPad Prism 9 (GraphPad Software, San Diego, CA, USA). For ROC, correlation analyses and graphs, IgG values less than the detection limit were set to 2/3 of this limit. The Youden index (YI = Sensitivity + Specificity À 1) was calculated using Excel (Office 2019, Microsoft, M€ unchen, Germany). Humoral immune response to SARS-CoV-2 in Iceland Prevalence of SARS-CoV-2 specific neutralising antibodies in blood donors from the Lodi red zone in Lombardy, Italy, as at 06 Prevalence of SARS-CoV-2 in Spain (ENE-COVID): a nationwide, population-based seroepidemiological study Serology-and PCR-based cumulative incidence of SARS-CoV-2 infection in adults in a successfully contained early hotspot (CoMoLo study) Infection fatality rate of SARS-CoV2 in a super-spreading event in Germany Seroprevalence of anti-SARS-CoV-2 IgG antibodies in Kenyan blood donors Seroprevalence of SARS-CoV-2 antibodies in over 6000 healthcare workers in Spain Persistence and decay of human antibody responses to the receptor binding domain of SARS-CoV-2 spike protein in COVID-19 patients Clinical and immunological assessment of asymptomatic SARS-CoV-2 infections Antibody tests for identification of current and past infection with SARS-CoV-2 Testing for SARS-CoV-2 (COVID-19): a systematic review and clinical guide to molecular and serological in-vitro diagnostic assays Evaluation of the EUROIMMUN anti-SARS-CoV-2 ELISA assay for detection of IgA and IgG antibodies Performance characteristics of the Abbott architect SARS-CoV-2 IgG assay and seroprevalence in Sensitivity in detection of antibodies to nucleocapsid and spike proteins of severe acute respiratory syndrome coronavirus 2 in patients with coronavirus disease 2019 Antibody response against SARS-CoV-2 spike protein and nucleoprotein evaluated by four automated immunoassays and three ELISAs Validation of a commercially available SARS-CoV-2 serological immunoassay SARS-CoV-2 S1 and N-based serological assays reveal rapid seroconversion and induction of specific antibody response in COVID-19 patients A comparison of four serological assays for detecting anti-SARS-CoV-2 antibodies in human serum samples from different populations SARS-CoV-2 (COVID-19) by the numbers Preexisting and de novo humoral immunity to SARS-CoV-2 in humans Potent neutralizing antibodies against SARS-CoV-2 identified by high-throughput single-cell sequencing of convalescent patients' B cells Isolation of potent SARS-CoV-2 neutralizing antibodies and protection from disease in a small animal model Potent neutralizing antibodies against multiple epitopes on SARS-CoV-2 spike Review of COVID-19 mRNA vaccines: BNT162b2 and mRNA-1273 Immune responses against SARS-CoV-2 variants after heterologous and homologous ChAdOx1 nCoV-19/ BNT162b2 vaccination Antibody persistence through 6 months after the second dose of mRNA-1273 vaccine for Covid-19 COVID-19 vaccines: modes of immune activation and future challenges SARS-CoV-2 antibody testing-questions to be asked Effectiveness of SARS-CoV-2 mRNA vaccines for preventing covid-19 hospitalizations in the United States Seroprevalence and correlates of SARS-CoV-2 neutralizing antibodies from a population-based study in Bonn Robust neutralizing antibodies to SARS-CoV-2 infection persist for months Correlation between SARS-COV-2 antibody screening by immunoassay and neutralizing antibody testing Neutralization of SARS-CoV-2 variants by convalescent and BNT162b2 vaccinated serum Neutralizing antibody levels are highly predictive of immune protection from symptomatic SARS-CoV-2 infection Update of reference values for IgG antibodies against typical antigens of hypersensitivity pneumonitis data of a German multicentre study A research standard for human serum immunoglobulins IgG WHO international standard for anti-SARS-CoV-2 immunoglobulin WHO/ BS.2020.2403 establishment of the WHO international standard and reference panel for anti-SARS-CoV-2 antibody Monoclonal antibody-based immunoenzymetric assays for quantification of human IgG and its four subclasses Guideline on bioanalytical method validation: European medicines agency Seroprevalence of anti-SARS-CoV-2 antibodies in COVID-19 patients and healthy volunteers up to 6 months post disease onset Comparison of SARS-CoV-2 serological tests with different antigen targets Evaluating the serological status of COVID-19 patients using an indirect immunofluorescent assay, France Evaluation of monoclonal antibodies having specificity for human IgG sub-classes: results of an IUIS/WHO collaborative study An international reference preparation for human serum immunoglobulins G, A and M: content of immunoglobulins by weight Quantitative measurement of IgG to SARS-CoV-2 antigens using monoclonal antibody-based enzymelinked immunosorbent assays Distinct antibody responses to SARS-CoV-2 in children and adults across the COVID-19 clinical spectrum Diagnostic and analytical performance evaluation of ten commercial assays for detecting SARS-CoV-2 humoral immune response Antibody response using six different serological assays in a completely PCR-tested community after a coronavirus disease outbreak-the CoNAN study Analysis of COVID-19 convalescent plasma for SARS-CoV-2 IgG using two commercial immunoassays Antispike protein assays to determine SARS-CoV-2 antibody levels: a head-to-head comparison of five quantitative assays Quantitative SARS-CoV-2 anti-spike responses to Pfizer-BioNTech and Oxford-AstraZeneca vaccines by previous infection status Differences in the concentration of anti-SARS-CoV-2 IgG antibodies post-COVID-19 recovery or post-vaccination Evidence for antibody as a protective correlate for COVID-19 vaccines Correlates of protection against symptomatic and asymptomatic SARS-CoV-2 infection A novel highly quantitative and reproducible assay for the detection of anti-SARS-CoV-2 IgG and IgM antibodies SARS-CoV-2-specific ELISA development This study was funded by the DGUV (German Social Accident Insurance; IPA project 145). We thank Christina Czibor, Silke Maryska, Sabine Bernard and Ursula Meurer for performing the immunoassays and Rosemarie Marchan for improving the language. Open Access funding enabled and organized by Projekt DEAL. The authors declare that they have no conflict of interest to disclose in the preparation of this manuscript.