key: cord-0979312-sepawqdp authors: Bedell, Kurt; Buchaklian, Adam; Perlman, Stanley title: Efficacy of an automated multi-emitter whole room UV-C disinfection system against Coronaviruses MHV and MERS-CoV date: 2016-01-28 journal: Infection Control & Hospital Epidemiology DOI: 10.1017/ice.2015.348 sha: 7ace5c93e0c8f732ea8c1e966eedf4f7608637ad doc_id: 979312 cord_uid: sepawqdp The virus responsible for Middle Eastern respiratory syndrome, MERS-CoV is a lineage C betacoronavirus similar to the mouse hepatitis virus type A59 (MHV-A59). The first reported case of MERS occurred in Saudi Arabia in 2012 and resulted in 76 deaths(1). Outbreaks of MERS have since occurred not only in the Middle East but South Korea as well(2). Rapid, efficient, and automated methods of disinfecting surfaces contaminated with the MERS-CoV virus may prevent the spread of the virus in the healthcare setting. Here we report on the use of an automated triple-emitter whole room disinfection system to inactivate the MHV-A59 and the MERS-CoV viruses on surfaces with a greater than 5 log(10) reduction on MERS in 5 minutes of UV-C exposure. The Surfacide multiple-emitter automated UV-C disinfection system consists of 3 ultraviolet C (UV-C) emitters with a diameter of 59 cm and a height of 195 cm. All systems include both a touch screen control tablet and a motion sensor that double as jointers to connect the emitters together for transport. The system utilizes a laser to identify the size of the disinfection space and to identify near-field objects. Based on these measurements the system then calculates the disinfection cycle time accordingly while rotating 360°. Triplicate 10-µL aliquots of MHV-A59 were loaded onto sterile glass coverslips at a concentration of 1.6e8 pfu/mL and allowed to dry. Coverslips containing dried virus were then placed in a UV transparent Petri dish and either left untreated or exposed to UV-C radiation from the Surfacide system placed 4 feet (1.22 m) from the samples for predetermined times between 0 and 60 minutes. The coverslips containing dried virus were then placed in Dulbecco's Modified Eagle's Medium (DMEM) media on ice to resuspend the virus. Serial 1:10 dilutions were inoculated onto Hela cells expressing the MHV receptor 7 and allowed to incubate for 45 minutes at 37°C. Medium containing virus was removed and the cells were allowed to incubate overnight in D10 media. Plaque counts were determined the following day by combining 1% neutral red with 2 × media plus agarose and incubating the cells for approximately 3 hours. All studies were conducted in triplicate with replicate experiments. Testing MERS-CoV sensitivity to UV-C was conducted under strict containment BSL3 laboratory protocols due to the infective nature of the virus, and all steps were carried out above bleachsoaked pads to prevent virus spread. The moist environment created by the bleach soaked pads in the biosafety cabinet prevented drying of the virus onto glass coverslips. MERS-CoV was therefore loaded onto glass coverslips as small droplets at a concentration of 8.9e5 pfu/mL in UV-C permeable dishes and exposed to UV-C energy from the Surfacide emitter placed 4 feet (1.22 m) from the dishes. Samples were removed at several time points between 0 and 30 minutes. Virus resuspension and dilution were carried out as described above. Virus dilutions were placed onto 90% confluent African green monkey kidney cell-line (Vero81) cells at 37°C for 40 minutes. Following a 40-minute incubation, medium containing virus was removed and replaced with 2 × medium plus 1.2% agarose and allowed to solidify. DMEM was then layered on top to prevent drying, and the plates were incubated at 37°C for 3 days to allow plaque formation. On day 3, formalin was added to fix the cells for 20 minutes. Formalin, media, and agarose were then removed and replaced with 0.1% crystal violet for a 5-minute incubation. This procedure was followed by a phosphate-buffered saline (PBS) wash prior to counting plaques. Again, all samples were prepared in triplicate. the UV-C energy emitted from the Surfacide disinfection system was able to reduce the viral titers of MHV-A59 by an average of 2.71 log 10 in 5 minutes and 6.11 log 10 in 10 minutes of exposure, resulting in undetectable levels of MHV-A59 virus (Figure 1 ). For MERS-CoV, a UV-C exposure time of only 5 minutes resulted in undetectable virus levels that remained undetectable following 30 minutes of total exposure for a 5.9 log 10 reduction (Figure 2 ). The multiple-emitter continuous UV-C disinfection system was >99.999% effective against MHV-A59, a mouse analog of MERS-CoV, and SARS-CoV in 10 minutes. Applying those same studies to droplets of MERS-CoV resulted in undetectable levels of MERS-CoV virus after only 5 minutes of exposure to the UV-C emitter or a percent reduction of >99.999%. Notably, UV-C disinfection is most effective when used as an adjunct to standard terminal cleaning of hard surfaces and should not be used to replace good cleaning practices. 8 This study is the first to document the effectiveness of a multiple-emitter, automated, whole-room UV-C system or any whole-room disinfection system against RNA viruses like MHV-A59, MERS-CoV, and SARS-CoV. The use of the Surfacide whole-room UV-C disinfection system during MERS outbreaks may prevent the nosocomial spread of the virus and protect staff in the process. acknowledgments Adam Buchaklian, PhD, is the Director of Research for Surfacide, LLC, and provided the UV-C disinfection system as well as recommendations on study design. All studies were conducted independently in the laboratory of Dr. Stanley Perlman at the University of Iowa. All other authors report no conflict of interest. Advances in clinical diagnosis and treatment of severe acute respiratory syndrome Middle East respiratory syndrome coronavirus (MERS-CoV); announcement of the Coronavirus Study Group Middle East respiratory syndrome Isolation of a novel coronavirus from a man with pneumonia in Saudi Arabia Lessons to learn from MERS-CoV outbreak in South Korea Middle East respiratory syndrome coronavirus (MERS-CoV): a cluster analysis with implications for global management of suspected cases Receptor for mouse hepatitis virus is a member of the carcinoembryonic antigen family of glycoproteins Decontamination of targeted pathogens from patient rooms using an automated ultraviolet-C-emitting device Virus was reduced by 2.71 log 10 in 5 minutes and by 6.11 log 10 in 10 minutes to undetectable levels. Results are the means of triplicate samples. figure 2. Log reduction of MERS-CoV during exposure to Surfacide UV-C at 4 feet (1.22 m). Log reduction of MERS-CoV on glass coverslips following Surfacide UV-C exposure at a 4-foot distance