key: cord-0973524-ev9rdgp7 authors: Ozog, David M.; Sexton, Jonathan Z.; Narla, Shanthi; Pretto-Kernahan, Carla D.; Mirabelli, Carmen; Lim, Henry W.; Hamzavi, Iltefat H.; Tibbetts, Robert J.; Mi, Qing-Sheng title: The Effect of Ultraviolet C Radiation Against Different N95 Respirators Inoculated with SARS-CoV-2 date: 2020-09-03 journal: Int J Infect Dis DOI: 10.1016/j.ijid.2020.08.077 sha: 77e22ac2d2c7c54df9ba2cd9b43ab9418e9d88ea doc_id: 973524 cord_uid: ev9rdgp7 Objectives There are currently no studies that have examined whether one dosage can be uniformly applied to different respirator types to effectively decontaminate SARS-CoV-2 on N95 filtering facepiece respirators (FFRs). Health care workers have been using this disinfection method during the pandemic. Our objective was to determine the effect of UVC on SARS-CoV-2 inoculated N95 respirators and whether this was respirator material/model type dependent. Methods Four different locations (facepiece and strap) on 5 different N95 FFR models (3 M 1860, 8210, 8511, 9211; Moldex 1511) were inoculated with a 10 μL drop of SARS-CoV-2 viral stock (8 × 107 TCID50/mL). The outside-facing and wearer-facing surfaces of the respirators were each irradiated with a dose of 1.5 J/cm2 UVC (254 nm). Viable SARS-CoV-2 was quantified by a median tissue culture infectious dose assay (TCID50). Results UVC delivered using a dose of 1.5 J/cm2, to each side, was an effective method of decontamination for the facepieces of 3 M 1860 and Moldex 1511, and for the straps of 3 M 8210 and the Moldex 1511. Conclusion This dose is an appropriate decontamination method to facilitate reuse of respirators for healthcare personnel when applied to certain models/materials. In addition, some straps may require additional disinfection to maximize the safety to frontline workers. Implementation of widespread UVC decontamination methods requires a careful consideration of model, material type, design, and fit-testing following irradiation. The shortage of personal protective equipment (PPE) is affecting healthcare workers worldwide during the coronavirus disease 2019 (COVID-19) pandemic. The ability to decontaminate and reuse N95 filtering facepiece respirators (FFRs) is a partial solution to the current shortage. 1 We previously proposed decontaminating respirators with repurposed dermatology office phototherapy devices, which serve as a platform for ultraviolet C (UVC) (254nm) germicidal disinfection. 2 On March 31 st , during the height of the pandemic, Henry Ford Health System (HFHS) began decontaminating 3M 1860 respirators with UVC and returning J o u r n a l P r e -p r o o f them to their original users. UVC dosing was extrapolated from earlier virology work. Since then, several thousand respirators have been decontaminated; one HFHS hospital alone decontaminated 5,797 respirators from April 7, 2020-July 2, 2020 with a single device. Recently, Fischer et al. demonstrated that UVC (260-285nm) was effective at decontaminating N95 respirators inoculated with SARS-CoV-2. 3 However, previous studies have demonstrated that UVC (254 nm) decontamination is dependent on the material/model type of N95 respirators. 4, 5 Furthermore, there are currently no studies that have examined whether one dosage can be uniformly applied to different respirator types to effectively decontaminate SARS-CoV-2 on N95 filtering facepiece respirators (FFRs).Therefore, the objective of this study was to determine the effect of UVC on decontamination of SARS-CoV-2-innoculated N95 respirators using a variety of FFRs that are available to healthcare employees at Henry Ford Health System (HFHS) in Detroit, MI. The study was performed as a collaboration between the HFHS and the University of Michigan. All study procedures were approved and conducted according to the University of Michigan Institutional Biosafety Committee BSL3 (Biosafety Level 3). The appropriate training and medical surveillance for experimental procedures and manipulations performed in the BSL3 facility were satisfied by all individuals directly involved in laboratory testing at University of Michigan. The SARS-CoV-2 strain used was USA-WA1/2020 NR-52281. Viral stocks of SARS-COV-2 were obtained from the Biodefense and Emerging Infections Research Resources Repository and were propagated in Vero-E6 cells grown in Dulbecco's Modified Eagle Medium (DMEM) without phenol red, with 2% Fetal Bovine Serum (FBS), L-glutamine, penicillin/streptomycin, non-essential amino acids, and hydroxyethyl piperazineethanesulfonic acid (HEPES). The virus stock was purposely produced in a phenol red-free medium to avoid photodegradation or photooxidation that may affect the results. For stock virus titration, aliquots of viral stock were applied on confluent Vero-E6 cells in 96-well plates for a 50% tissue culture infectious dose (TCID50) assay. Viral stocks were determined to be 8 x 10 7 TCID50/mL. Respirators were tested 100% intact and included the following models: 3M 1860 (St. Intact FFRs in a donned position were inoculated on the outside-facing surface with a single 10 μL drop of viral stock (8 x 10 7 TCID50/mL) on 4 areas to account for differing received doses on complex surfaces: nosepiece, apex, chin-piece, and strap ( Figure 1A) . Inoculated respirators were dried in a biosafety cabinet at room temperature for 40 minutes. For each N95 respirator model, FFRs were UVC-irradiated or left untreated as positive controls for viral load recovery. The respirators were then placed under the UVGI device, in the center, and were individually treated with a dose of 1.5 J/cm 2 . Then, they were rotated and the wearer-facing side of the N95 was again irradiated with 1.5 J/cm 2 . The irradiation time for each side was approximately 60-70 seconds. The device does not generate any heat; as such, all FFRs were exposed to UVC at room temperature. Immediately after the completion of the irradiation, 4 mm circles containing the inoculated surface were obtained with a leather belt eyelet hole punch tool and were placed in 300 μL (microliters) of PBS for 1 hour at room temperature. Recovered viral loads were determined by TCID50 assay of the absorbed samples. Briefly, 25 μL aliquots of serially 10-fold diluted samples were inoculated into 96-well plates with a Vero-E6 cell monolayer in sextuplicate and cultured in DMEM with 2% FBS, Lglutamine, penicillin/streptomycin, non-essential amino acids, and HEPES. The plates were observed for cytopathic effects for 4 days. Viral titer was calculated with the Reed and Müench endpoint method. 7 Viral yields were expressed as total TCID50 recovered in 300 μL or TCID50/ 4mm punch. TCID50 negative controls were cells with media only and were included on each J o u r n a l P r e -p r o o f plate assayed. All negative controls had no cytopathic effect (CPE). The limit of detection (LOD) for the TCID50 assay was determined to be 10 1.3 TCID50 / 4 mm punch. If the amount of viral particles was below the LOD, then a theoretical, yet low content of viruses may be present. However, an absence of CPE in the Vero E6 cells at 4 days post inoculation indicates a loss of infectivity and is evidence of inactivation of the SARS-CoV-2 samples ( Figure 1B) . We considered effective decontamination to be results below the LOD with no CPE, the elimination of all infectious SARS-CoV-2. Following preliminary testing (Appendix Figure 1 , Appendix Table 1 ), virus inoculation was performed on all 5 types of respirators. For each type, three were irradiated with UVC, and one was not irradiated to serve as positive control. Similar results were seen for the 3M 1860 respirators as in the preliminary study. All facepiece locations were below the LOD with absent CPE. Two straps were above the LOD, and one strap was below the LOD with absent CPE. Sufficient virus (≤ 1 log reduction) was recovered from the untreated positive controls on all facepiece locations; however, a lower yield was recovered from the untreated control strap (Figure 2, Appendix Table 2 ). On the 3M 8210 respirators, location 1 had two respirators above the LOD, and one respirator below the LOD with absent CPE. Location 2 had one FFR at the LOD, and two FFRs below the LOD with absent CPE. Location 3 and all the straps were below the LOD with absent CPE. Lower virus yields were recovered from the untreated positive control on all facepiece locations. In contrast, the strap did not absorb the droplet and a sufficient yield was obtained. Table 2 ). Five N95 respirator models were inoculated with SARS-CoV-2 and tested. UVC showed a significant (≥ 3 log) reduction of H1N1 influenza viability. 8 J o u r n a l P r e -p r o o f Some respirator models have materials, such as the straps of the 3M 1860, that demonstrate hydrophilic characteristics when inoculated. Moreover, these seemingly hydrophilic surfaces showed consistently lower mean log reduction < 3 log10 TCID50. 5 In contrast, seemingly hydrophobic materials, such as the 3M 1860 facepiece, were found to demonstrate a > 3 log10 TCID50 reduction. 5 In our study using SARS-CoV-2, we observed similar results with the 3M 1860 facepiece and strap. Further, the facepieces of the 3M 8210 have hydrophilic properties which were reflected in the reduced decontamination results, while the straps did not readily absorb the droplets, and hence were adequately disinfected. The Moldex 1511 facepiece and straps also appeared to be hydrophobic and did not absorb the droplets. Some straps are prone to twisting. Consequently, when the respirator is flipped during the irradiation process, care must be taken to make sure the appropriate surface of the strap is Table 3 ). Additionally, manufacturers may consider using, for example, the same material as the straps of the 3M 8210 for all the other models of FFRs to improve UVC decontamination. Our dosage for this study was partially based on previous work with Influenza A (H1N1), A/Shanghai/1/2013, MERS-CoV, and SARS-CoV, 4, 5, 8, 9 where it was determined that all areas of a respirator should receive at least 1 J/cm 2 . Data from the Photomedicine and Photobiology Unit at HFHS demonstrated through theoretical and measured models that the curvature and the distance of the 3M 1860 N95 respirator from the light source affected the dosage delivered in a predictable way. Moreover, extrapolating from this model, after irradiating one side of the respirator with 1.5 J/cm 2 , some of the lateral aspects may only receive 900 mJ/cm 2 while the apex of the respirator may receive almost 3 J/cm 2 . Further, it was also observed that a certain percentage of the dosage received in an area (~10%) permeates to the other side. 6 Therefore, 1.5 J/cm 2 was chosen as the lowest radiant exposure (i.e. fluence) to ensure that all areas received at least 1 J/cm 2 . Increasing the dosage delivered may improve decontamination, but UVC radiation can degrade certain polymers in a dose dependent manner, and the effects may vary greatly between different models. 10 Lindsley et al exposed circular coupons from different N95 respirators to UVC doses ranging from 120-950 J/cm 2 . The exposure led to a small increase in particle penetration (up to 1.25%). At dosing > 950 J/cm 2 , a more significant effect on the strength of the layers of the respirator materials (in some cases, > 90% decrease) was observed. Less effect was seen on the respirator straps tested, requiring a dose of 2,360 J/cm 2 to reduce the breaking strength by 20-51%. 11 The limitation of this study was that fit-testing was not performed, which Additional testing was performed to determine if the droplet was drying larger than the 4 mm area tested. The results showed that there was limited to no virus in the periphery of the 4 mm area tested, and that no virus could be detected on the wearer-facing surface. Moreover, the lower yield reflects a diminished ability to resuspend the virus after drying. Limitations of the study include that no soiling agents were used. However, at Henry Ford Health System, as in other healthcare facilities, personnel are instructed not to reuse respirators that are visibly soiled. Further, it is still unclear what the infectious dose is for SARS-CoV-2; therefore, it is unknown if a significant reduction in viral load eliminates contagious risk. Based upon clinical observation at the HFHS sites, since beginning UVC decontamination of N95 respirators, it does not appear that there was a significant increase in healthcare personnel related COVID-19 infection that could be potentially attributed to wearing N95 respirators that were not decontaminated of infectious J o u r n a l P r e -p r o o f SARS-CoV-2. However, we were unable to trace whether COVID-19 positive employees had used UVC treated respirators, which is a major limitation to this assumption. In conclusion, UVC at a dose of 1.5 J/cm 2 applied to both sides is effective at decontaminating SARS-CoV-2 on some N95 respirators. This dose may only be an appropriate decontamination method to facilitate reuse of PPE for healthcare personnel when applied to certain models/materials. Further, this UVC dosage may not provide adequate decontamination against other more robust hospital-acquired respiratory infections; however, treatable respiratory infections have not been a significant cause of morbidity during the COVID-19 pandemic. In addition, some straps may require additional disinfection to maximize the safety to the frontline workers. Implementation of widespread UVC decontamination methods requires a careful consideration of model, material type, design, and fit-testing following irradiation. It should also be emphasized that similar cautions should be practiced for all other methods of respirator decontamination. The authors declare the following financial interests/personal relationships which may be considered as potential competing interests: SN is a sub-investigator for Biofrontera. DO is an investigator for Biofrontera. HWL is an investigator for LITE study which is funded by PCORI, and home phototherapy machines are provided by Daavlin, and has participated as a speaker in general educational session for Ra Medical System. IHH is an investigator for LITE study which This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors. All study procedures were approved and conducted according to the University of Michigan Prevention CfDCa. Decontamination and Reuse of Filtering Facepiece Respirators Ultraviolet germicidal irradiation: possible method for respirator disinfection to facilitate reuse during COVID-19 pandemic Effectiveness of N95 Respirator Decontamination and Reuse against SARS-CoV-2 Virus Ultraviolet germicidal irradiation of influenza-contaminated N95 filtering facepiece respirators Research to Mitigate a Shortage of Respiratory Protection Devices During Public Health Emergencies UVC Germicidal Units: Determination of Dose Received and Parameters to be Considered for N95 Respirator Decontamination and Reuse. Photochem Photobiol A SIMPLE METHOD OF ESTIMATING FIFTY PER CENT ENDPOINTS12 The importance of the minimum dosage necessary for UVC decontamination of N95 respirators during the COVID-19 pandemic for UV-C-Based N95 Reuse Risk Management Ultraviolet-C and other methods of decontamination of filtering facepiece N-95 respirators during the COVID-19 pandemic. Photochemical & photobiological sciences : Official journal of the European Photochemistry Association and the Effects of Ultraviolet Germicidal Irradiation (UVGI) on N95 Respirator Filtration Performance and Structural Integrity The Importance of Fit-Testing in Decontamination of N95 Respirators: A Cautionary Note