key: cord-0971413-gevguehk authors: Bonde, J.; Ejegod, D. M.; Pedersen, H.; Smith, B.; Cortes, D.; Leding, C.; Thomsen, T.; Benfield, T.; Schnieder, U. V.; Tingleff, J.; Arbyn, M.; Lisby, G.; Andersen, O. title: Clinical validation of point-of-care SARS-COV-2 BD Veritor antigen test by a single throat swab for rapid COVID-19 status on hospital patients predominantly without overt COVID symptoms date: 2021-04-17 journal: nan DOI: 10.1101/2021.04.12.21255299 sha: 4811e7713c0dda177d15c2e647ea9a54931974f3 doc_id: 971413 cord_uid: gevguehk BACKGROUND Fast identification of severe acute respiratory syndrome Coronavirus-2 (SARS-CoV-2) infected individuals is a strategically vital task to ensure correct management and quarantine. Rapid antigen test could be a supplement to the standard-of-care Nucleic Acid Amplification Test (NAAT). The aim of this study was to determine the accuracy of the BD Veritor SARS-CoV-2 antigen test as a screening instrument in a hospital setting. METHODS A cohort of prospective samples were collected from hospital staff and patients at the Emergency, Infectious Diseases and Pediatrics and Adolescent Medicine departments at Hvidovre Hospital. All samples were collected using oropharyngeal swabs, and BD Veritor Antigen test results were paired with routine NAAT test results. Sensitivity, specificity, positive and negative predictive values of the antigen test were calculated using NAAT as reference. RESULTS Overall, 809 samples from 674 individuals were included (average age 45 years, range 0-98 years). Among all samples, 8% were SARS-CoV-2 positive by NAAT testing and 5.3% by BD Veritor. The sensitivity of the antigen test was 63.1% and specificity 99.7%. The positive predictive value was 95.3%. False-positive rate was 4%. The cycle threshold value was significantly higher among individuals with false negative antigen tests compared to true positives. CONCLUSION The sensitivity, specificity and positive predictive values show that the BD Veritor antigen test from oropharyngeal collected specimens performs well. Antigen testing may be a supplement, but not substitute, to NAAT testing as the primary diagnostic modality in hospital settings where fast turnaround test results may assist in decisions regarding isolation and quarantine. Fast and easy-to-perform diagnostic methods is of high priority to curb the Coronavirus Disease 2019 (COVID- 19) pandemic. As the pandemic enters its 2 nd year, the political focus in many countries including Denmark is on easy test accessibility, fast time-to-result, accurate tests and cost-efficiency. Even when vaccination is fully rolled-out, it is anticipated that severe acute respiratory syndrome Coronavirus-2 (SARS-CoV-2) testing will remain a pivotal public health measure the coming years. The reference standard for SARS-CoV-2 testing is Nucleic Acid Amplification Test (NAAT) targeting the RNA genome of SARS-CoV-2 in upper-respiratory specimens and performed in centralized laboratory settings 1, 2 . NAAT systems deliver test-to-result in 2 to 6 hours depending on circumstances and this time-to-result has obtained widespread acceptance as a premise of protecting public health by identifying infected individuals and referring them into quarantine. However, many SARS-COV-2 test situations benefit from point-of-care testing (POCT) with fast time-to-result. In a hospital setting, rapid identification of SARS-CoV-2 infected patients is a strategically vital task to ensure correct clinical management, precaution management, isolation or quarantine of SARS-CoV-2 positive patients. SARS-CoV-2 carriers pose a potential risk to healthcare professionals who are in close contact with the patient during examinations or treatments for non-COVID-19 as well as COVID-19-related conditions, and constitute a risk to other patients in e.g. common waiting areas and shared hospital rooms in high through-put Emergency Departments were a flow-culture is prominent 3 . Antigen tests are immunoassays that detect a specific viral antigen. Rapid SARS-CoV-2 antigen detection (RAD) tests can be performed as a POCT, are easy to administer, and the results are typically available within 15-20 minutes. Combined with the solid political focus, especially on time-to-result and scalability, it has resulted in the implementation of SARS-CoV-2 antigen POCT in public and private test settings. The RAD test market is booming as the SARS-CoV-2 pandemic unfolds 4, 5 , but qualitative data on their performance is limited and biased by test cohort characteristics 6 . SARS-CoV-2 antigen POCT is a promising instrument to optimize SARS-CoV-2 control in conjuncture with NAAT test regimens, but principal concerns regarding antigen test performance include low sensitivity, reproducibility, and low positive predictive value, all depending on specific circumstances 6, 7 . This study aimed to conduct a clinical performance validation of the BD Veritor antigen Rapid test (Veritor) using oropharyngeal swabs from hospital staff and patients (inpatients and outpatients). The majority of tested were without COVID-19 symptoms tested as a routine screening. . CC-BY 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) The copyright holder for this preprint this version posted April 17, 2021 The "BD Veritor System for Rapid Detection of SARS-CoV-2" is a chromatographic lateral flow immunoassay for the qualitative detection of nucleocapsid antigens in a respiratory specimen. The dedicated Veritor throat swab was resuspended for 15´sec in the provided test tube containing a virus transport medium that neutralizes SARS-CoV-2 virus to a non-infectious level 8 . The test cassette incubates for 15 minutes in the BD Veritor analyzer (VA) and the result is reported by the VA. Patient ID was scanned into the VA ensuring safe patient ID identification. The manufacturer reports a test specificity of 100% and a sensitivity of 84% compared to qRT-PCR as a reference standard during the first 5 days after disease onset for use on superficial nasal specimens (Manufacturers insert). SARS-CoV-2 NAAT testing was part of the routine diagnostic for SARS-CoV-2 testing performed at either the The cohort size was determined on >80% power to detect a 5% difference in sensitivity between PCR and the antigen POCT. A SARS-CoV-2 positivity rate of <2% with an increasing trend was assumed. Based on the above, at least 773 paired samples were required. In total, 971 Veritor samples were collected from 779 persons ( Figure 1 ). Samples with a Veritor test (index) and a reference NAAT test result within ±5 days were included. Twelve samples were included with a time difference between index and reference test >5 ≤10 days based on the premise that the paired reference was the only available and that results were concordant. Ninety-one samples were excluded due to no matching reference test. Seven samples were excluded due to incorrect personal identifiers. Forty-nine samples were excluded as multiple samples collected from the same subject within a 24 hours window. Eleven samples were excluded due to a history of prolonged SARS-CoV-2 positivity rendering the patient irrelevant for antigen test. Four samples were excluded based on insufficient sampling material. Seventytwo individuals were tested twice or more during the study period (11%, Figure 1 ). The Department of Pediatrics and Adolescent Medicine provided 49 samples from in-and outpatients (average age; 3.8 years, . CC-BY 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. Overall, the sensitivity and specificity of Veritor using NAAT as reference standard were 63.1% and 99.7%, respectively (Table 1) . A sub-analysis of patients (N=29, average age 52 years, range 5-82 years) with known, recent NAAT confirmed SARS-CoV-2 showed a sensitivity and specificity of 73.9% and 83.3%, respectively ( Table 2) . For individuals with unknown SARS-CoV-2 status (N=780, average age 45 years, range 0-98), sensitivity and specificity were 57.1% and 99.9%, respectively (Table 2) . The overall sensitivity of the Veritor test was 63.1%. Though seemingly unimpressive at first glance, the Veritor test performs at par with similar antigen test evaluated on the general Danish population 11 or similar studies with cohorts not dominated by symptomatic persons 6, [12] [13] [14] . For the group with known, recent NAAT confirmed SARS-CoV-2, the Veritor test provided a sensitivity of 74% (N=29). The sensitivity amongst individuals with unknown SARS-CoV-2 status was 57% (N=780). In comparison, the recent Danish study by Jakobsen et al. using the BioSensor antigen test found a sensitivity of 78% for persons with selfreported symptoms, versus 49% amongst people without self-reported symptoms 11 . The specificity of the Veritor test was found to be uniformly high with 99.7% for the overall cohort, and 99.9% in individuals with unknown SARS-CoV-2 status at time of testing. Amongst those with unknown SARS-CoV-2 status, the Veritor test stands out for the very high PPV of 96% and low Veritor test false-positive rate of 4.0%, which provides confidence that a positive Veritor finding is indeed SARS-CoV-2 positive. This is in contrast to the asymptomatic cohort of the Jakobsen study 11 where the PPV of the BioSensor antigen test was 72.5%. Finally, the NPV of 97.6% is at par with the majority of . CC-BY 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted April 17, 2021. ; https://doi.org/10.1101/2021.04.12.21255299 doi: medRxiv preprint antigen test evaluated 6 , with the caveat that high NPV values are driven by the relatively low SARS-CoV-2 prevalence in ours and other´s studies. Infectious Diseases departments and staff undergoing routine testing. The age range is from 5 days old to 98 years. The broad age spectrum contrast with similar studies, most of which only included persons 18 years or older (Table 3) , demonstrating the easy operationalization of the Veritor antigen test using a throat collected specimen. The SARS-CoV-2 prevalence by NAAT, overall and amongst persons with unknown SARS-CoV-2 status, was 8.1% and 5.5%, respectively. The prevalence reflects the hospital setting, the enrichment with recently diagnosed, NAAT confirmed SARS-CoV-2 positive patients, and a peak prevalence of SARS-CoV-2 in Denmark at the time. Veritor test negative/PCR positive cases had significantly higher Ct scores indicating a lower viral load of the discordant samples compared to the Veritor-to-PCR concordant samples ( Figure 2 ). This is in line with earlier studies of various antigen test system performance [11] [12] [13] [15] [16] [17] . In the respiratory tract, SARS-CoV-2 peak viral load is observed at the time of symptom onset and subsequently declines 18, 19 . In a clinical update by Cevik et al. 20 they conclude that the highest infectiousness potential of SARS-CoV-2 is up to 5 days of symptom onset, but that NAAT tests continue to be positive for a mean of 17 days. Furthermore, Pekosz and colleagues using cell culture demonstrated that the Veritor test had similar performance in detecting replicating SARS-CoV-2 infections as PCR 21 . Nevertheless, as PCR/NAAT is more sensitive than antigen tests or viral cell culture and as human-to-human transmission may occur at even lower levels than observed in cell culture, NAAT remain the reference standard for detection of SARS-CoV-2. From a hospital setting perspective where a fast time-to-result turnaround is required, a positive antigen test should lead to isolation without awaiting confirmatory NAAT testing, whereas a negative antigen test result combined with a SARS-CoV-2 negative clinical evaluation should await NAAT confirmation as the Veritor test cannot exclude SARS-CoV-2 infection. Comparing antigen test performance across studies is hampered by dissimilarities in study design, cohort characteristics, age, sample collection sites, and reference NAAT test. Table 3 presents antigen test systems in use in Denmark, and it is noteworthy that manufacturer-provided performance data stem from cohorts dominated by very high COVID-19 symptomatic rates resulting in high clinical assay sensitivity estimates not confirmable by real-life, manufacturer-independent studies [11] [12] [13] . From a practical perspective, confirming SARS-CoV-2 infection in patients with overt COVID-19 symptoms remains an essential yet minor diagnostic application in absolute numbers. The largest application of antigen testing is screening of predominantly asymptomatic persons as part of efforts to curb dissemination of SARS-CoV-2 infection in society in general and characterized by low prevalence of SARS-CoV-2 infection. Hence the current level of manufacturer-provided documentation and studies focusing on symptomatic cohorts are at best . CC-BY 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted April 17, 2021. ; https://doi.org/10.1101/2021.04.12.21255299 doi: medRxiv preprint insufficient to guide on antigen test performance in a predominantly non-infected populations undergoing screening by antigen testing. Our study used an oropharyngeal (throat) collected specimen, which is off-label according to the manufacturers' specifications and other published studies [11] [12] [13] 17 . In Denmark, the National Board of Health recommends throat swabs as the preferential sampling site for the detection of SARS-CoV-2. However, the emergence of antigen testing has forced the implementation of nose swabs in Denmark given the lack of data validating antigen tests for throat collected specimens. Combining our study with the Van der Moeren study 17 , the Veritor antigen test could allow the choice of either a throat or nose swab, whatever is preferred by the individual undergoing testing. Moreover, from an operationalization point-of-view, if a patient must undergo an oropharyngeal collected sample for a SARS-CoV-2 NAAT test, then it is possible to do a Veritor antigen test in the same procedure, so that the patient gets two procedures in one test situation. In a Hospital POCT setting, we found the BD Veritor Antigen test system simple and easily operationalized with positive user experience feed-back. A strength of this study was that the testing was real-life, POCT at multiple departments and the enrollment of samples were representative of the patients seen at the participating departments. All samples were completed immediately on-site, and no frozen or preserved specimens were included. Our POCT setting distinguishes this study from studies using centralized sample testing 6 or stored samples. The VA unit provides a layer of safe patient identification as the health insurance card is scanned to link patient ID and sample allowing for systematic data reporting. The Veritor test capacity is easily scalable and performs similar whether it is employed using the VA unit or using visual reading of the test cassette 17 . The test buffer supplied neutralizes SARS-CoV-2 virus 22 which is a safety feature for the staff operating the test system. All samples were oropharyngeal specimens. This eliminate the potential bias of comparing clinical assay performance between nasopharyngeal, oropharyngeal, nasal, oral or saliva samples. Moreover, access to centralized test databases allowed a detailed record of previous SARS-CoV-2 tests at the individual subject level. A limitation of the study was the comprehensive NAAT technologies for concurrent NAAT testing of reference samples reflecting the real-life hospital situation. This nevertheless limited the power of the Ctanalysis by reducing the number of samples where a Ct score could be retrieved. Regarding Ct score, a general limitation of ours and other´s studies is the interpretation of Ct scores with respect to calling positive cases given the lack of international standardization of PCR Ct values 6 . WHO recently issued a notice of concern regarding interpretation of specimens at or near the limit for PCR detection 23 . A strength of our study was that all positives are called after manual inspection of curves. Yet until joint international standards are agreed, performance evaluations of antigen tests with NAAT as reference standard will benefit NAAT testing in terms of sensitivity. . CC-BY 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted April 17, 2021. ; https://doi.org/10.1101/2021.04.12.21255299 doi: medRxiv preprint In summary, the BD Veritor antigen test for SARS-CoV-2 detection performs well in a hospital POCT setting and could encourage re-thinking whether antigen tests can be used for rapid screening to manage inpatients and outpatients while SARS-CoV-2 status is confirmed by NAAT testing. The very high PPV and NPV in the group of persons with unknown SARS-CoV-2 status jointly demonstrate that the Veritor antigen test represents an analytical improvement for mass-screening compared to the antigen tests recently employed in Denmark. The Veritor antigen tests may serve as a supplement where fast turnaround test results can assist and aide in making clinical decisions regarding SARS-CoV-2 isolation and quarantine but antigen test in general cannot substitute NAAT testing as the primary diagnostic modality for the diagnosis and surveillance of SARS-CoV-2 infection. We would like to thank the staff of the participating AHH-departments who helped with the collection of samples and data retrieval. Sample collection, testing and project management was embedded into the routine of the participating departments. BD Diagnostics, Sparks, MD, donated BD Veritor SARS-CoV-2 antigen tests and Veritor analyzers free-of-charge. The donor had no role in data handling, data presentation, or resulting publication. Marc Arbyn was supported by the Validation of Coronavirus Assays (VALCOR) project (Sciensano, Brussels, Begium). . CC-BY 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted April 17, 2021. Organization WH: WHO Laboratory testing for coronavirus disease ( COVID-19) in suspected human cases: interim guidance Prevention CfDCa: Interim Guidance for Antigen Testing for SARS-CoV Implementing evidence-based practices in an emergency department: contradictions exposed when prioritising a flow culture The explosion of new coronavirus tests that could help to end the pandemic Fast coronavirus tests: what they can and can't do Van den Bruel A, Cochrane C-DTAG: Rapid, point-of-care antigen and molecular-based tests for diagnosis of SARS-CoV-2 infection Scaling up COVID-19 rapid antigen tests: promises and challenges Methods of Inactivation of SARS-CoV-2 for Downstream Biological Assays Detection of a novel human coronavirus by real-time reverse-transcription polymerase chain reaction An in-well direct lysis method for rapid detection of SARS-CoV-2 by real time RT-PCR in eSwab specimens Detection of SARS-CoV-2 infection by rapid antigen test in comparison with RT-PCR in a public setting Urgent need of rapid tests for SARS CoV-2 antigen detection: Evaluation of the SD-Biosensor antigen test for SARS-CoV-2 Field evaluation of a rapid antigen test (Panbio COVID-19 Ag Rapid Test Device) for COVID-19 diagnosis in primary healthcare centres Clinical Evaluation of BD Veritor SARS-CoV-2 Point-of-Care Test Performance Compared to PCR-Based Testing and versus the Sofia 2 SARS Antigen Point-of-Care Test Low performance of rapid antigen detection test as frontline testing for COVID-19 diagnosis Comparison of automated SARS-CoV-2 antigen test for COVID-19 infection with quantitative RT-PCR using 313 nasopharyngeal swabs, including from seven serially followed patients Performance evaluation of a SARS-CoV-2 rapid antigentest: test performance In the community in the Netherlands COVID-19 pandemic-a focused review for clinicians SARS-CoV-2 transmission dynamics should inform policy Virology, transmission, and pathogenesis of SARS-CoV-2 Antigen-Based Testing but Not Real-Time Polymerase Chain Reaction Correlates With Severe Acute Respiratory Syndrome Coronavirus 2 Viral Culture Methods of inactivation of SARS-CoV-2 for downstream biological assays Organization WH: WHO information notice for IVD users: nucleic acid testing (NAT) technologies that use realtime polymerase chain reaction (RT-PCR) for detection of SARSCoV-2