key: cord-0970595-s2yivfen authors: Haimovich, Adrian D.; Taylor, R. Andrew; Krumholz, Harlan M.; Venkatesh, Arjun K. title: Performance of Temporal Artery Temperature Measurement in Ruling Out Fever: Implications for COVID-19 Screening date: 2020-09-15 journal: J Gen Intern Med DOI: 10.1007/s11606-020-06205-2 sha: 9ce52f37c3ae63f432c972ce5ef4fe304bed10e9 doc_id: 970595 cord_uid: s2yivfen nan Current CDC recommendations for mitigation of community COVID-19 transmission include temperature screening. 1 Due to low cost and ease-of-use, temporal artery thermometers (TATs) applied to the forehead are widely employed to screen for fever, but prior literature has suggested poor sensitivity and high variability. 2, 3 Published data are largely limited by small per-study sample sizes and a focus on pediatric, surgical, and intensive care settings that are not generalizable to real-world screening populations. We sought to determine the real-world test performance of TATs for fever rule-out by utilizing a large electronic dataset of emergency department encounters for whom universal temperature screening was conducted. For reference standards, we included rectal temperature, a widely recognized core temperature, as well as oral temperature, which benefits from wide clinical acceptance and robust specificity. 2 Our primary objective was to determine TAT sensitivities and specificities across a range of temperatures in comparison with rectal and oral cutoffs of 100.4°F/38°C. Our secondary outcome was limit-ofagreement (LOA) by Bland-Altman analysis. We extracted temperature measurements and method of temperature assessment from electronic health record (EHR) data (Epic, Verona, WI) collected between March 2013 and June 2019 within a large hospital system comprising ten acute care sites. We identified paired perpatient data where a TAT measurement was documented within 15 minutes of a rectal temperature measurement or oral temperature. When multiple measurements were taken with a single modality within the defined interval, the mean value was used. In the Bland-Altman analysis, we calculate the mean of the differences between the paired measurements and the limits-of-agreement (LOA) as defined by 95% confidence interval bounds. We identified 1.84 million adult (age > 18 years) emergency department visits by 602,089 patients with over 4.6 million temperature readings; there were 1293 paired readings from 1276 encounters that met our inclusion for TAT versus rectal measurement and 16,132 readings from 16,031 encounters for TAT versus oral measurement. The admission rate across the paired measurement patient population was 50.0%. The prevalence of fever in the rectal and oral temperature populations was 34.4% and 4.3%, respectively. Using a threshold of 100.4°F, TAT measurement identified fever compared with the rectal reference with sensitivity 0.27 (95% CI 0.27-0.31), specificity 0.98 (0.96-0.99), PPV 0.85 (0.79-0.91), and NPV 0.72 (0.69-0.74). TAT measurement identified fever compared with the oral reference with a sensitivity 0.23 (95% CI 0.20-0.26), specificity 0.99 (0.99-0.99), PPV 0.53 (0.48-0.59), and NPV 0.97 (0.96-0.97). We did not observe significant differences in fever sensitivity when limiting our paired analysis to 5-or 10-minute windows. Decreasing the threshold for fever using TAT improved sensitivity with respect to the rectal temperature gold standard of 100.4°F (Fig. 1) . Using a TAT cutoff of 99°F, for example, increased sensitivity to 0.63 (0.58-0.67) and NPV to 0.82 (0.79-0.84), while specificity decreased to 0.86 (0.83-0.88) and PPV to 0.7 (0.65-0.74). The mean differences and LOA between TAT and rectal temperatures and TAT and oral temperatures were − 1.13 (LOA − 6.69 to 4.42)°F, and − 0.45 (− 3.89-2.99)°F, respectively (Fig. 2) . Consistent with prior work, we found TATs had poor test performance, identifying less than one in three positive cases. [2] [3] [4] We observed that decreasing TAT temperature thresholds yielded significantly improved test sensitivity with modest losses in specificity and note that this approach is being recommended by some state departments of public health. 5 Generalizability of our analysis may be limited by reliance on emergency department visits with paired measures, as well as possible non-uniformity of testing devices over time. 4 Based on current CDC recommendations for fever measurement (cutoff of 100.4°F) and TAT screening, the majority of people who would meet the criteria for fever would be wrongly classified as afebrile. A lower TAT fever cutoff would necessarily increase false positive rates, likely prompting testing for COVID-19 and potentially unnecessary self-isolation and workplace withdrawal for uninfected individuals. Public and private entities seeking to mitigate the transmission of COVID-19 in the workplace or high-risk settings should consider a combination of lower thresholds (e.g., 99°F), alternative technologies, or serial temperature assessment approaches as well as using temperature alongside symptom screening to ensure that public health goals are achieved. 6 Implementation of Mitigation Strategies for Communities with Local COVID-19 Transmission Accuracy of Peripheral Thermometers for Estimating Temperature: A Sy st e ma t i c Re v i ew a n d Me t a -a na l y s i s . A n n I n t e rn M e d Temperature measurements with a temporal scanner: systematic review and meta-analysis Temperature measurement in the adult emergency department: oral, tympanic membrane and temporal artery temperatures versus rectal temperature Essential Services Screening Recommendations for COVID-19 Pandemic Symptom Screening at Illness Onset of Health Care Personnel With SARS-CoV-2 Infection Publisher's Note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations Acknowledgments: IRB: This work was deemed exempt by the Yale Institutional Review Board.