key: cord-0970246-o7kpbnri authors: Jhou, You-Ru; Wang, Chih-Hung; Tsai, Huey-Pin; Shan, Yan-Shen; Lee, Gwo-Bin title: An Integrated Microfluidic Platform Featuring Real-Time Reverse Transcription Loop-Mediated Isothermal Amplification for Detection of COVID-19 date: 2022-01-26 journal: Sens Actuators B Chem DOI: 10.1016/j.snb.2022.131447 sha: 523279e362f5513acc0d3c1450863200c58b4c13 doc_id: 970246 cord_uid: o7kpbnri An integrated microfluidic platform (IMP) utilizing real-time reverse-transcription loop-mediated isothermal amplification (RT-LAMP) was developed here for detection and quantification of three genes of the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2; i.e., coronavirus diseases 2019 (COVID-19)): RNA-dependent RNA polymerase, the envelope gene, and the nucleocapsid gene for molecular diagnosis. The IMP comprised a microfluidic chip, a temperature control module, a fluidic control module that collectively carried out viral lysis, RNA extraction, RT-LAMP, and the real-time detection within 90 min in an automatic format. A limit of detection of 5×10(3) copies/reaction for each gene was determined with three samples including synthesized RNAs, inactive viruses, and RNAs extracted from clinical samples; this compact platform could be a useful tool for COVID-19 diagnostics. The coronavirus diseases 2019 (COVID- 19) pandemic, which began in China in late 2019, was caused by the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2; an RNA virus). The outbreak of SARS-CoV-2 soon became a global pandemic recognized by World Health Organization (WHO) [1] . Total deaths reached four million people worldwide by September 2021, in part due to high rates of transmission between asymptomatic individuals [2] . Developing fast and accurate diagnostic methods could effectively prevent the spread of infections, and SARS-CoV-2 detection devices have indeed been critical. This is normally undertaken via reverse-transcription polymerase chain reaction (RT-PCR), which is highly sensitive and specific [3] although it may require a lengthy thermal-cycling process on a costly apparatus. Coronavirus genomes consist of 8-10 open reading frames (ORFs) [4] ; For instance, ORF1 is translated into 16 non-structural proteins that include the RNA-dependent RNA polymerase enzyme (RdRp). Envelope (E), nucleocapsid (N), S spikes, and membrane proteins are also essential for completion of the viral replication cycle [5] , and RdRp, E, and N, genes have been widely used as SARS-CoV-2 biomarkers [6] . In addition to PCR [7] , virus antigen and antibody testing have been used [8] [9] ; such procedures are relatively fast and simple although they lack the specificity and sensitivity of nucleic acids-based approaches and may generate false-positive or false-negative results. In addition, antibodies are only observed in blood in the middle and later stages of the illness and thus are not suitable for early detection [9] [10] . RT-PCR is relatively time-consuming (~2 to 4 hr) and labor-intensive; furthermore, viral lysis and RNA extraction must be undertaken by well-trained personnel. In contrast, loop-mediated isothermal amplification (LAMP) can amplify nucleic acids without J o u r n a l P r e -p r o o f thermocycling [11] ; this is associated with reductions in detection time compared to RT-PCR [12] . Efforts have been made to miniaturize and automate LAMP [13] [14] [15] , and LAMP technology has been integrated with microfluidic technologies [16] [17] for rapid and sensitive detection of pathogens [18] . Compared with traditional LAMP, the main advantages of these microfluidics-based systems are 1) lower sample and reagent volumes, 2) compact size for on-site testing, and 3) capacity for automation via integration of microfluidic devices (e.g., micropumps, micromixer, & microreactors). For instance, we demonstrated that sample purification, RNA extraction, and LAMP could be performed on an integrated microfluidic platform without human intervention for influenza diagnosis [19] . However, this device was used for end-point testing only (i.e., only qualitative, not quantitative diagnosis), which may not be satisfactory for SARS-CoV-2 diagnostics since quantitative diagnosis of viral load is required. Several devices/systems for detecting SARS-CoV-2 have been reported [20] [21] [22] . For example, a portable platform for rapid virus detection was characterized by a limit of detection (LOD) of 5 x 10 4 copies/mL of the N gene in 30 min though RNA extraction was carried out off-chip [20] . Similarly, an LOD of only 1 copy/mL was reported for a device featuring an integrated microfluidic platform (IMP) controlled by a smartphone [21] . Another smartphone-controlled device, a centrifugal microfluidic platform, was characterized by an LOD of 10 4 copies/mL of the Orf1ab gene, and the detection time was only 60 min [22] . However, none of these approaches are fully automated (i.e., requiring human intervention during detection). They also feature only 1-2 genes while a 3-gene approach has been shown to reduce chances of false-negative or false-positive diagnoses [23] . Given these limitations, we devised an IMP for rapid detection and quantification of three viral genes (E, N, and RdRp) of SARS-CoV-2. The entire process including rival lysis, RNA extraction, RT-LAMP and quantification could be automated within 90 min with a limit of detection as low as 5x10 3 J o u r n a l P r e -p r o o f 5 copies/reaction (i.e., 2x10 2 copies/μL). Therefore, this new platform could provide useful and accurate detection for SARS-CoV-2. Experimental process for viral lysis, RNA extraction, RT-LAMP, and real-time quantification Released RNAs were transported to chambers (described below) to be mixed with 5 μL of 1.05 μm-diameter Dynabeads MyOne™ (carboxylic acid, Invitrogen, USA; 10 6 beads/μL) that had been conjugated with three amine-modified RNA-specific probes (Table S1) J o u r n a l P r e -p r o o f was placed under the chambers to collect the magnetic bead-RNA complexes for 1 min. Then, unbound RNAs and interfering materials were washed out, and the complexes were concentrated by the magnet for 5 min. Next, unbound RNAs were again washed away with 35 μL of double-distilled water (ddH 2 O) for 5 min. Next, RT-LAMP reagents (24 μL/reaction) were mixed with magnetic bead-RNA complexes and then transported to the corresponding chambers for RT-LAMP using calcein for fluorescent detection as in [19] . In addition to the The microfluidic chip ( Figures S2 & 2a) , which consisted of an air channel layer, a liquid control layer, 9 micropumps/mixers, 18 microvalves, 14 microchambers, and 18 air injection holes, was made from polydimethylsiloxane (PDMS, Sylgard 184A/B, Sil-More Industrial, USA) that had been poured into polymethymethacrylate (PMMA, Dateh Industrial, Taiwan) molds [24] . The PDMS was mechanically demolded from PMMA, cured at 80°C for 5 hr, Figure 2a ). Both samples and reagents were transported by pneumatically driven membrane-type micropumps ( Figure S1 ), which also functioned as micromixers. For liquid transport, a positive gauge pressure was provided by closing the left microvalves and activating the micropumps ( Figure S1a ); the liquid was drawn into the micropump with a negative gauge pressure provided ( Figure S1b ). The left microvalve was closed via a positive gauge pressure, and the right microvalve was opened via a negative gauge pressure for liquid transport ( Figure S1c ); in other words, liquid transport was achieved by providing a positive gauge pressure for activating the micropump and then providing a positive gauge pressure for closing the right microvalve ( Figure S1d ). For mixing, liquids were drawn into the micropump while providing a negative gauge pressure for opening the left microvalve and micropump. The micropump then transported the liquid from chamber to chamber in turn by providing positive and negative gauge pressures. The microfluidic platform included a fluidic control module, a temperature control module, and an optical detection module, all controlled by a microcontroller (Arduino UNO, Italy; Figure S6a ), which was designed to be compact for on-site applications. The second was for the optical detection module (45 cm x 30 cm x 30 cm; Figure S6 (b), as described below). Both units were fabricated by a 3D printer (i3 Mega, Anycubic, China). The optical detection module consisted of a PMT (C6271+R928, Hamamatsu, Japan), an optical shutter, a laser, a set of optical filters and mirrors, and a microcontroller ( Figure S4 ). The light source used for the optical detection module was a diode-pumped, solid-state laser (MBL-473, CNI, China; 473±1 nm, [for calcein]). A shutter controlled by a stepper motor (MG995, TowerPro, Taiwan) was used to block the laser when not needed ( Figure S5 ). The optical detection module was mounted with band-pass filters (500-550 nm) for fluorescence detection, and signals were captured by the PMT under microcontroller regulation. The fluorescence intensities (i.e. optical detection signals) of serially diluted test samples were evaluated with the negative control among triple experiments in this study. Thus, the two-tailed student t test analysis was used to calculate whether there was a significant difference (p<0.05) for LOD determination. During the 60-min RT-LAMP, signals were acquired every minute, and a threshold time could be defined as the time required for the fluorescence signal to cross over a threshold, which was set as in [18, 25, 26] ; since calcein was used, the threshold time was inversely proportional to the amount of initial target in the sample. With this approach, real-time fluorescence signal could be used to quantify the initial concentration of viruses. The data were fitted to a sigmoidal curve, and the threshold time was calculated as the sum of the mean fluorescence signal during the initial 5 min and 5 times of the standard deviation values during the same time as in the previous study [26] . Therefore, the threshold time, which was inversely proportional to the initial amount of the target, could be determined by the point of intersection of the threshold value and the fitted curve. Temperature variations within the two chambers were both <0.5°C ( Figure S7) when mixing released RNA with magnetic beads coated with RNA probes, the applied pressure could affect bead capture [27] . Therefore, the gauge pressures for the liquid mixing/washing steps were decreased to positive/negative gauge pressures of 10/-15 kPa, respectively. According to previous studies [13, 27] , a gentle mixing should be applied to achieve an optimal capture. If the mixing was too strong such that the resulting shear force was much higher than the binding force between released RNA and probes on the beads, then the capture rate could be deteriorated. The mixing performance of the micromixer was further explored. The mixing index of Agarose gel electrophoresis ( Figure 5) showed that only the genes of the SARS-CoV-2 virus could be successfully amplified, indicating that the designed primers exhibited satisfactory specificity. Next, E, N, and RdRp gene constructs were serially diluted from 5x10 4 to 5x10 1 copies/reaction for exploring the sensitivity of RNA extraction+RT-LAMP on-chip. After RNA extraction and RT-LAMP, each gene was successfully amplified ( Figure 6 ). The LODs were found to be 5x10 3 copies/reaction for E (Figure 6a ), N (Figure 6b ), and RdRp (Figure 6c) genes when using synthesized RNA. To evaluate the fluorescence intensities in real-time, the optical detection module was used to detect fluorescence every minute, and, after a 60-min reaction (Figure 6d ), fluorescent signals were evident and varied in intensity across differing starting concentrations. The two-tailed student t test showed that there was a significant difference (p<0.05) for LOD determination for 3 genes with a concentration higher than 5x10 3 copies/reaction, which indicated that the optical detection was reliable. The assay was then used with serially diluted inactive viruses (5x10 4 to 5x10 1 copies/reaction), and each gene was successfully amplified after on-chip viral lysis+RNAextraction+RT-LAMP ( Figure 7) ; LODs were 5x10 3 copies/reaction (i.e., 2x10 2 copies/μL) for all genes (Figures 7a-c) , lower than the 5x10 4 copies/μL value of a previous work [20] with the S and orf 8 genes. Note that the LOD of the N gene was only 50 copies/μL in that work, though the associated device could not detect all three genes simultaneously. The corresponding fluorescence detection results (Figure 7d ) also highlight the high performance of our novel device. Similarly, the two-tailed student t test revealed a significant difference J o u r n a l P r e -p r o o f (p<0.05) for LOD determination for each gene with a concentration higher than 5x10 3 copies/reaction. Hence, the optical detection could be conducted accurately. For clinical samples of lineage B.1.1.7. (i.e., the original COVID-19 virus) provided by NCKUH, samples were lysed, and viral RNAs were 1) extracted using protocols previously described [19] and 2) diluted from 5x10 4 to 5x10 1 copies/reaction. All three genes were successfully amplified (Figure 8) , and on-chip LOD was 5x10 3 copies/reaction (i.e., 2x10 5 copies/mL) for each gene (the same as for synthesized RNA & inactive viruses). The LOD of the developed device was measured to be 5x10 3 copies/reaction (i.e., 2x10 5 copies/mL or Ct of 32.6 [28] ) for each gene. According to the previous work [28] , threshold (Ct) values >30 or copy numbers <10 6 copies/mL are defined as "non-infectious," meaning that our device can detect clinically meaningful levels of viruses; this is further evidenced in Figure 8d , in which fluorescence differences can be observed across dilutions. Moreover, as for each gene, there was a significant difference (p<0.05) for LOD determination in the two-tailed student t test. Therefore, the device could detect each gene. cylinders [25] . Therefore, the optical detection module could provide more reliable optical detection than in previous works [19, 25] , and three COVID-19 genes could be detected in only 90 min. This platform could consequently revolutionize COVID-19 diagnostics. In this study, an automatic real-time RT-LAMP microfluidic system has been demonstrated. The system was consisted of 3 modules, including fluidic control, temperature control, and optical detection modules, which have been integrated in a compact packaging case. The entire process for viral lysis, RNA extraction, RT-LAMP, and optical detection/quantification could be automated. It could simultaneously amplify E, N, and RdRp genes in SARS-CoV-2 and the LOD was found to be 5x10 3 copies/reaction (i.e. 2x10 5 J o u r n a l P r e -p r o o f copies/mL) for all three genes. With this approach, the rapid detection of three genes could be completed in 90 min. To the best of our knowledge, it is the first time that three genes of the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) could be detected and quantified on an integrated microfluidic platform by automating the entire process (including rival lysis, RNA extraction, RT-LAMP and optical quantification). The device could also conduct real-time detection. The initial copy number of viruses could be determined accordingly. Therefore, the developed platform should meet the need of rapid detection of COVID-19. There are no conflicts to declare. The origin, transmission and clinical therapies on coronavirus disease 2019 (COVID-19) outbreak -an update on the status Asymptomatic cases in a family cluster with SARS-CoV-2 infection Detection of SARS-CoV-2 in different types of clinical specimens The genetic sequence, origin, and diagnosis of SARS-CoV-2 Coronaviruses: an oerview of their replication and pathogenesis Detection of 2019 Novel Coronavirus (2019-nCoV) by Real-time RT-PCR An overview of nucleic acid testing for the novel coronavirus SARS-CoV-2 SARS-CoV-2 seroconversion in humans: a detailed protocol for a serological assay, antigen production, and test setup Antibody Responses to SARS-CoV-2 in patients with COVID-19 Validation of serological and molecular methods for diagnosis of Zika virus infections A novel reverse transcription loop-mediated isothermal amplification method for rapid detection of SARS-CoV-2 Loop-mediated isothermal amplification (LAMP): a rapid, accurate, and xost-effective diagnostic method for infectious diseases Rapid isolation and detection of aquaculture pathogens in an integrated microfluidic system using loop-mediated isothermal amplification LAMP-on-a-chip: revising microfluidic platforms for loop-mediated DNA amplification Loop-mediated isothermal amplification integrated on microfluidic chips for point-of-care quantitative detection of pathogens Simpler, faster, and sensitive Zika virus assay using smartphone detection of loop-mediated isothermal amplification on paper microfluidic chips Real-time fluorescence loop mediated isothermal amplification for the diagnosis of malaria A Sample-to-answer, portable platform for rapid detection of pathogens with a smartphone interface Rapid isothermal amplification and portable detection system for SARS-CoV-2 Rapid molecular diagnostics of COVID-19 by RT-LAMP in a centrifugal polystyrene-toner based microdevice with end-point visual detection Sample-to-answer COVID-19 nucleic acid testing using a low-cost centrifugal microfluidic platform with bead-based signal enhancement and smartphone read-out Diagnosing COVID-19: The disease and tools for detection An automated and portable antimicrobial susceptibility testing system for urinary tract infections Vancomycin-resistant gene identification from live bacteria on an integrated microfluidic system by using low temperature lysis and loop-mediated isothermal amplification Development of a real-time fluorescence loop-mediated isothermal amplification assay for rapid and quantitative detection of Ustilago maydis Research on magnetic bead motion characteristics based on magnetic beads preset technology SARS-CoV-2, CT-values, and infectivity-conclusions to be drawn from side observations Formal analysis, Investigation, Methodology, Validation, Writing -Original Draft; Chih-Hung Wang: Conceptualization, Methodology, Writing -review & editing Writing -review & editing Shan Conceptualization, Resources, Writing -review & editing Conceptualization, Funding acquisition The authors thank the Ministry of Science and Technology ☒ The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.☐The authors declare the following financial interests/personal relationships which may be considered as potential competing interests: J o u r n a l P r e -p r o o f