key: cord-0965736-b1fzvv3n authors: Coll De Peña, Adriana; Hill, Nicole; Lapizco-Encinas, Blanca H. title: Determination of the Empirical Electrokinetic Equilibrium Condition of Microorganisms in Microfluidic Devices date: 2020-10-19 journal: Biosensors (Basel) DOI: 10.3390/bios10100148 sha: a83cf2c057c8ec092863c2464025ba898c42428f doc_id: 965736 cord_uid: b1fzvv3n The increased concern regarding emerging pathogens and antibiotic resistance has drawn interest in the development of rapid and robust microfluidic techniques to analyze microorganisms. The novel parameter known as the electrokinetic equilibrium condition ([Formula: see text]) was presented in recent studies, providing an approach to analyze microparticles in microchannels employing unique electrokinetic (EK) signatures. While the [Formula: see text] shows great promise, current estimation approaches can be time-consuming or heavily user-dependent for accurate values. The present contribution aims to analyze existing approaches for estimating this parameter and modify the process into an accurate yet simple technique for estimating the EK behavior of microorganisms in insulator-based microfluidic devices. The technique presented here yields the parameter called the empirical electrokinetic equilibrium condition ([Formula: see text]) which works well as a value for initial approximations of trapping conditions in insulator-based EK (iEK) microfluidic systems. A total of six types of microorganisms were analyzed in this study (three bacteria and three bacteriophages). The proposed approach estimated [Formula: see text] values employing images of trapped microorganisms, yielding high reproducibility (SD 5.0–8.8%). Furthermore, stable trapping voltages (sTVs) were estimated from [Formula: see text] values for distinct channel designs to test that this parameter is system-independent and good agreement was obtained when comparing estimated sTVs vs. experimental values (SD 0.3–19.6%). The encouraging results from this work were used to generate an EK library of data, available on our laboratory website. The data in this library can be used to design tailored iEK microfluidic devices for the analysis of microorganisms. Insulator-based electrokinetic (iEK) microfluidic techniques, especially dielectrophoresis (DEP), have been used for a large variety of bioanalytical applications [1] . Two primary modes of EK microfluidic techniques, electrode-based and insulator-based, have been used in numerous applications with DEP [2] . Electrode-based EK techniques are popular [3] [4] [5] [6] and have demonstrated promising capabilities including the ability to analyze different strains of bacteria in a co-culture and track their behavior [7] and even categorize erythrocytes by their ABO-Rh blood types [8] . However, iEK microfluidic techniques are generally simpler to fabricate, inexpensive, disposable, and less affected to fouling than traditional electrode-based EK devices. Insulator-based EK microfluidic techniques have been used for the rapid identification, detection, and purification of both viruses [9] [10] [11] and cells [12] [13] [14] [15] , as these techniques can distinguish quickly between very similar organisms. For images of microorganisms electrokinetically trapped between insulating posts, forming a stable band. We used the term stable trapping voltage (sTV) to identify the applied potential, which is higher than the traditional sufficient trapping voltage (TV), defined as the required voltage to obtain an observable band of trapped particles in iEK devices [9] . The use of channels with insulating post to determine eE EEC was presented in a recent preliminary study using bacterial and yeast cells [13] . The present contribution aims to review previous approaches, draw distinctions between E EEC and eE EEC , and provide an accurate and simple method for estimating eE EEC and estimate electrokinetic trapping conditions. The results obtained demonstrated that once eE EEC has been determined for particular microorganism, this parameter can be used to estimate a starting range for the stable voltage (sTV) at which the microorganism will be trapped in any iEK microfluidic channel. Good agreement was found between estimated and experimentally determined sTV values. These encouraging results were used to generate an EK library of data, available on our laboratory website [43] . The eE EEC values presented here were obtained with a methodology that is simpler, faster, and less labor-intensive than the process for determining E EEC values [40] . The parameter eE EEC introduced here shows great potential for the rapid differentiation and identification of microorganisms in iEK microfluidic devices by employing eE EEC as a unique electrokinetic signature. Biosensors 2020, 10, x FOR PEER REVIEW 3 of 14 trapped between insulating posts, forming a stable band. We used the term stable trapping voltage (sTV) to identify the applied potential, which is higher than the traditional sufficient trapping voltage (TV), defined as the required voltage to obtain an observable band of trapped particles in iEK devices [9] . The use of channels with insulating post to determine was presented in a recent preliminary study using bacterial and yeast cells [13] . The present contribution aims to review previous approaches, draw distinctions between and , and provide an accurate and simple method for estimating and estimate electrokinetic trapping conditions. The results obtained demonstrated that once has been determined for particular microorganism, this parameter can be used to estimate a starting range for the stable voltage (sTV) at which the microorganism will be trapped in any iEK microfluidic channel. Good agreement was found between estimated and experimentally determined sTV values. These encouraging results were used to generate an EK library of data, available on our laboratory website [43] . The values presented here were obtained with a methodology that is simpler, faster, and less labor-intensive than the process for determining values [40] . The parameter introduced here shows great potential for the rapid differentiation and identification of microorganisms in iEK microfluidic devices by employing as a unique electrokinetic signature. Electrokinetic phenomena are classified as linear and non-linear, determined by their dependence on the electric field [39] . Linear EK phenomena include , the flow of liquid relative to a stationary surface due to the electrical double layer (EDL) of the channel walls, and (1) , the motion of electrically charged particles relative to a stationary fluid. The velocity expressions for and (1) are: where and (1) represent the and (1) mobilities, respectively, and are the zeta potentials of the wall and particle, respectively, and and are the media permittivity and viscosity, respectively. Non-linear EK phenomena include , particle motion due to polarization Electrokinetic phenomena are classified as linear and non-linear, determined by their dependence on the electric field [39] . Linear EK phenomena include EO, the flow of liquid relative to a stationary surface due to the electrical double layer (EDL) of the channel walls, and EP (1) , the motion of electrically charged particles relative to a stationary fluid. The velocity expressions for EO and EP (1) where µ EO and µ (1) EP represent the EO and EP (1) mobilities, respectively, ζ W and ζ P are the zeta potentials of the wall and particle, respectively, and ε m and η are the media permittivity and viscosity, respectively. Non-linear EK phenomena include DEP, particle motion due to polarization effects caused by a non-uniform electric field, and EP of the second kind EP (3) , an electrophoretic response that appears at higher E [36] [37] [38] [39] . The expressions for DEP, EP (3) , and the total particle velocity (v P ) considering all four EK effects are: where µ DEP and µ EP are the DEP and EP (3) mobilities, respectively. A recent study [13] performed in devices with insulating posts (Figures 1 and 2a-c) suggested that the DEP contribution to total particle velocity, in comparison to the EP (3) contribution, is only 0.89-5.95%. Therefore, it was decided for simplification purposes to neglect DEP from the expression for total particle velocity, thus: Biosensors 2020, 10, x FOR PEER REVIEW 4 of 14 effects caused by a non-uniform electric field, and EP of the second kind ( (3) ), an electrophoretic response that appears at higher [36] [37] [38] [39] . The expressions for , (3) , and the total particle velocity ( ) considering all four EK effects are: where and (3) are the and (3) mobilities, respectively. A recent study [13] performed in devices with insulating posts (Figures 1 and 2a -c) suggested that the contribution to total particle velocity, in comparison to the (3) contribution, is only 0.89-5.95%. Therefore, it was decided for simplification purposes to neglect from the expression for total particle velocity, thus: Distribution of E within the constriction region at 1000 V, where red represents the highest E magnitude. The black contour lines are isometric lines across which E is uniform, the red cross represents where the new method would estimate the empirical electrokinetic equilibrium condition ( ), and the white cross represents where the preliminary study [13] estimated the . (d) Schematic representation of post-less channel employed in the estimation of using particle velocimetry measurements [40] . (e) Forces acting on particles in post-less channels depending on their surface charges. Equation (6) is also an accurate expression of particle velocity in a post-less microchannel ( Figure 2d -e), as is not present [40] . From this expression the parameter , which is the E at which a particle stops migrating and is electrokinetically trapped ( = 0), can be derived [13, 40] : Distribution of E within the constriction region at 1000 V, where red represents the highest E magnitude. The black contour lines are isometric lines across which E is uniform, the red cross represents where the new method would estimate the empirical electrokinetic equilibrium condition (eE EEC ), and the white cross represents where the preliminary study [13] estimated the eE EEC . (d) Schematic representation of post-less channel employed in the estimation of E EEC using particle velocimetry measurements [40] . (e) Forces acting on particles in post-less channels depending on their surface charges. Equation (6) is also an accurate expression of particle velocity in a post-less microchannel (Figure 2d -e), as DEP is not present [40] . From this expression the parameter E EEC , which is the E at which a particle stops migrating and is electrokinetically trapped (v P = 0), can be derived [13, 40] : Biosensors 2020, 10, 148 5 of 14 Equation (8), obtained by isolating E from the particle velocity equation, can be used to determine accurate EP (3) mobility data (µ EP ) once E EEC has been identified from particle velocimetry measurements in post-less microchannels [40] . However, since significant experimentation and data analysis are required by numerous particle velocimetry measurements, the present study proposes the use of the empirical-E EEC (eE EEC ) which can be determined in much more straight forward manner. The eE EEC parameter is determined employing microchannels with insulating posts, where DEP forces are present and particle-particle interactions are more significant (due to higher particle concentrations) and influence particle trapping [44] . As DEP is still present in these channels, there is not a straightforward equation for eE EEC as there is for E EEC . The eE EEC parameter offers a good estimation of the trapping voltages when testing microorganisms in an iEK system. Both E EEC and eE EEC are considered system-independent, but they are not media-independent, as the mobilities in Equation (8) depend on the suspending medium characteristics. For this study, standard soft lithography techniques were used to cast polydimethylsiloxane (PDMS) microfluidic channels ( Figure 1a ) [13] . Upon curation with heat, the cast was removed from the mold, holes were punched to create channel liquid reservoirs and the cast was sealed with a PDMS-coated glass wafer using a plasma corona wand (Electro Technic Products, Chicago, IL, USA). This was done to ensure all the internal surfaces of the channel were made of the same material ensuring uniform EO flow. The devices were 10.16 mm long, 0.88 mm wide, and 40 µm tall. Microchannels with three distinct post shapes were made: circles, ovals and diamonds (Figure 1b-d). For this study, three distinct bacteria species and three distinct bacteriophage species were used (Table 1) . Size information for all microorganisms was measured in our lab or obtained from the literature [43] . These organisms were fluorescently labeled with one of the DNA intercalating Syto dyes, 11 or 45 (Invitrogen, Carlsbad, CA, USA). All organisms had a final concentration in the order of 10 6 particles/mL or higher after labeling. The suspending media consisted of DI water purified using a RiOs™ 200 Water Purification System 120 V (MilliporeSigma, Burlington, MA, USA), and had a conductivity of 15.1 ± 6.1 µS/cm and a pH of 6.7 ± 0.5. A Leica DMi8 inverted microscope (Wetzlar, Germany) paired with a Leica DFC7000 T camera and the software LASX provided by the manufacturer were used to record the behavior of the microorganisms. A high voltage supply (Model HVS6000D, LabSmith, Livermore, CA, USA) was used to apply direct current (DC) electric potentials. COMSOL Multiphysics ® 4.4 was used to simulate our system and estimate the eE EEC of each microorganism. A sample of 5-25 µL of the selected microorganism was introduced into the microfluidic channel with the desired post shape (Figure 1b-d) . Platinum wires were placed at the inlet and outlet reservoirs and a DC electric potential was applied from left to right creating zones of higher field intensity within the insulating posts (Figure 2a ). Voltages were gradually increased above TV to a voltage where stable bands were established, referred to as stable trapping voltage (sTV). Images of trapped microorganisms at sTV were captured and the sTV values were recorded. The location of the center of the "front" side of the band of trapped microorganisms was determined from the images, as depicted by the red cross in Figure 2b . This location at the "front" side of the band was selected because at this exact location is where particle velocity is zero, that is, all forces exerted on the microorganisms are in equilibrium. After the first microorganisms are trapped at this equilibrium location, additional microorganisms accumulate behind them, resulting in a band of trapped microorganisms. Therefore, the location of the "front" or inner side of the band is where the forces are in equilibrium, that is, the eE EEC value. These experiments were conducted in triplicate for each organism and each post shape. The sTV values were then input into COMSOL to simulate the distribution of E within the channel (Figure 2a ) and the eE EEC was then estimated as the E at the location determined from the COMSOL image (red cross in Figure 2c ), where v P = 0. Cardenas-Benitez et al. [40] was the first to estimate the novel parameter E EEC for four types of polystyrene particles using a post-less channel (Figure 2d) and particle tracking velocimetry (PTV) measurements. The forces acting on the particles in this system are illustrated in Figure 2e. Each particle type was run individually, and particle velocities were tracked at both low and high E regimes until particle motion switched directions, i.e., particles started to migrate backwards. When the particle switched directions, the E at which v P = 0 was recorded and used to estimate the E EEC for each particle. Therefore, numerous videos of each particle type had to be analyzed at different applied voltages to determine the E EEC . In a study by Coll De Peña et al. [13] that employed channels with insulating posts, the DEP contribution to the particle velocity was neglected (Equation (6)), as it was estimated to be between 0.89-5.95% in comparison to the EP (3) contribution. This study estimated eE EEC , since microfluidic devices with posts were used, by combining experimental information with COMSOL estimations of the electric field distribution between posts (Figure 2a) . Experiments in that preliminary study were focused on estimating the trapping voltage (TV), which is the applied voltage when the particle trapping becomes observable. Experiments with five strains of bacteria and one of yeast cells were performed in microfluidic devices containing circle and oval shaped posts (Figure 1c,d) . Each organism was run separately in microchannels with each post shape, and the TV was visually determined, as the example shown in Figure 2b , by gradually increasing the applied voltage. Upon obtaining the TV experimentally, a COMSOL model was used to estimate E, which is equivalent to the eE EEC at v P = 0; E was estimated at the center of the constriction (where E is max), as represented with the white cross in Figure 2c . The new approach presented here has two goals: (i) clearly distinguish between E EEC [40] and eE EEC [13] and (ii) accurately estimate the new parameter eE EEC by employing a new distinct location (red cross in Figure 2b ,c) within the constriction region. The E EEC values presented in the seminal paper by Cardenas-Benitez et al. [40] are determined through significant experimental and data analysis work with polystyrene particles in devices without posts, ensuring that only three forces are present: EP (1) , EO, and, at higher voltages, EP (3) . Numerous experiments were needed to be performed at several different voltages, the resulting particle velocity data required extensive analysis, leading to a labor intensive and time-consuming process to properly determine the E EEC values. The seminal study by Cardenas-Benitez et al. [40] presented fundamentals of E EEC and EP (3) in iEK devices, as shown in Equation (8) . The present contribution proposes a different approach, as it focuses instead on the application of experiments in iEK devices with insulating posts and higher particle concentrations. Higher particle concentrations lead to particle-particle interactions [44] , and the presence of insulating posts leads to DEP effects, which can influence the eE EEC value. Bacterial cells and bacteriophages were used in the eE EEC estimations. While E EEC is the more accurate parameter, the eE EEC value is easier to obtain, requires much less experimental work, and can still provide valuable estimations that can be used as an EK signature for the rapid identification of particle and microorganisms. The eE EEC value is determined with a simpler, faster, and far less labor-intensive method that works well to provide an initial estimation of stable trapping voltages for different species. Another important advantage of employing eE EEC is that lower voltages are required, since the presence of insulating posts produces higher local electric fields; allowing reaching electrokinetic equilibrium at lower applied voltages than those required in PIV channels. A characterized species might be expected to have similar E EEC and eE EEC values, though the contribution of DEP and particle-particle interactions could potentially contribute to deviations in reported values. The goal of eE EEC values and addressing eE EEC and the technique to determine it is to give a good estimate for trapping conditions within an insulating electrokinetic microfluidic device. Addressing the second goal, between the preliminary study to determine eE EEC [13] and this study, the use of observable trapping voltage (TV) vs. stable trapping voltage (sTV) is the primary distinction. This is crucial as the TV can be affected by sample concentration given that a band may not be observable until highly concentrated, introducing subjectivity into the system. Thus, rather than estimating the eE EEC at the center of the constriction [13] (white cross, Figure 2b ,c), the new location for estimating eE EEC was determined by experimental images at sTV values, i.e., when the band of trapped microorganism is clear and stable (red cross, Figure 2b ,c). To emphasize this point, the eE EEC values from the prior preliminary study [13] for the bacteria were compared to the values obtained with the new approach presented here. All of the eE EEC values from the prior study were higher and had a higher variance (34-76%) than those found in this present study. It is clear from the findings of the present study that the new method proposed for estimating eE EEC is robust and reliable. For every microorganism, experiments were conducted in triplicate in each post shape, and for each trial an image of the EK trapping of the microorganism at the sTV was used for estimating the eE EEC at the location illustrated in Figure 2b ,c (red cross). The estimations of eE EEC are reported in Figure 3a ,b for cells and bacteriophages, respectively. The average combined eE EEC for each distinct microorganism species across the distinct post shapes are shown in Figure 3c , where, as expected due to larger size, bacteria have lower eE EEC values than the bacteriophages [41] . The standard deviation across post shapes for all species ranged from 0.8 to 9.9%; in particular, it ranged from 0.8 to 8.3% for the bacteria (Figure 3a ) and from 2.5 to 9.9% for the bacteriophages (Figure 3b ). well within our tolerance. The EK library reflecting these results is publicly available on our laboratory website [43]. To demonstrate the applicability of the new parameter for estimating sTV values in iEK microfluidic devices with different post shapes, the sTV values for one bacteria species (S. enterica) and one bacteriophage species (SPN3US) were estimated using COMSOL. The values at the locations at which the sTV values were estimated for Figure 4a -e can be seen in Figure 3a ,b. To vary these estimations, three distinct locations were determined for each post shape by employing the previously estimated location, plus and minus one standard deviation. In all the simulations, the third column of posts within the channel was used to perform a voltage sweep to determine the values. As these values were found at single positions, the values were linearly dependent on the voltages. Linear interpolation was used to estimate the corresponding voltages to the value for each distinct post shape. Then, the estimated voltages were compared with the experimentally obtained voltages. This illustrates the versatility of , since it can be applied to channels with other post shapes to estimate the applied voltage needed to achieve stable EK trapping of the microorganism at specific positions. The sTV values found at each position were averaged together; these values are reported in Figure 4a ,b,f,g for bacteria S. enterica and Figure 4c The standard deviation of the estimation for the circle, oval, and diamond-shaped posts were 4.6%, 3.6%, and 5.8%, respectively. These low values suggest that this technique is not biased towards any particular post shape or feature. Thus, given that eE EEC is post shape-independent, the eE EEC values obtained for each microorganism in each post shape were then averaged together (Figure 3c and Table 1 ), yielding low standard deviations (5.2-8.8% for bacterial and 5.0-8.4% for bacteriophages). This low variability in eE EEC values illustrates the high precision and accuracy of this method. Considering that a portion of this variability could be introduced by neglecting the DEP contribution, and that biological samples have an innate population diversity, these variations are well within our tolerance. The EK library reflecting these results is publicly available on our laboratory website [43]. To demonstrate the applicability of the new eE EEC parameter for estimating sTV values in iEK microfluidic devices with different post shapes, the sTV values for one bacteria species (S. enterica) and one bacteriophage species (SPN3US) were estimated using COMSOL. The E values at the locations at which the sTV values were estimated for Figure 4a -e can be seen in Figure 3a ,b. To vary these estimations, three distinct locations were determined for each post shape by employing the previously estimated location, plus and minus one standard deviation. In all the simulations, the third column of posts within the channel was used to perform a voltage sweep to determine the E values. As these E values were found at single positions, the E values were linearly dependent on the voltages. Linear interpolation was used to estimate the corresponding voltages to the eE EEC value for each distinct post shape. Then, the estimated voltages were compared with the experimentally obtained voltages. This illustrates the versatility of eE EEC , since it can be applied to channels with other post shapes to estimate the applied voltage needed to achieve stable EK trapping of the microorganism at specific positions. The sTV values found at each position were averaged together; these values are reported in Figure 4a ,b,f,g for bacteria S. enterica and Figure 4c The standard deviation in the estimation of sTV ranged between 0.3% and 19.6% with an average of only 5.2% for S. enterica and 5.7% for SPN3US bacteriophage. Components of these deviations can be the accuracy for determining the band positions (represented by the red crosses in Figure 4f -h), variations in suspending media characteristics, microorganism population distribution, and the choice to neglect effects. The applicability and good accuracy of this approach are supported by the little overlap that exists between the small error bars of the experimental and estimated sTV values in Figure 4b -e, suggesting that there is no significant statistical difference between experimental and estimated values. It is important to note that Figure 4a does show that the estimated sTV values based on the value for the oval posts is outside of the error bar range for the experimental voltages. This is not the case for the results in Figure 4b -e, where estimated sTV values are within the error bars of the experimental sTV values. As expected from previous reports [41, 45] , the sTV values in the circle posts are the highest, followed very closely by the diamonds, when present, and then by the oval posts. This is consistent throughout several different experiments. Variations in the system might be the cause behind this discrepancy, but the lower values for oval The standard deviation in the estimation of sTV ranged between 0.3% and 19.6% with an average of only 5.2% for S. enterica and 5.7% for SPN3US bacteriophage. Components of these deviations can be the accuracy for determining the band positions (represented by the red crosses in Figure 4f -h), variations in suspending media characteristics, microorganism population distribution, and the choice to neglect DEP effects. The applicability and good accuracy of this approach are supported by the little overlap that exists between the small error bars of the experimental and estimated sTV values in Figure 4b -e, suggesting that there is no significant statistical difference between experimental and estimated values. It is important to note that Figure 4a does show that the estimated sTV values based on the eE EEC value for the oval posts is outside of the error bar range for the experimental voltages. This is not the case for the results in Figure 4b -e, where estimated sTV values are within the error bars of the experimental sTV values. As expected from previous reports [41, 45] , the sTV values in the circle posts are the highest, followed very closely by the diamonds, when present, and then by the oval posts. This is consistent throughout several different experiments. Variations in the system might be the cause behind this discrepancy, but the lower values for oval posts versus the higher values for circle posts should also be noted. In general, this approach does demonstrate reasonable estimations for sTV value range. In addition, it can also be observed that the sTV values for SPN3US are significantly higher than those for S. enterica (Figure 4a-e) , which is also expected based on their significant size difference. These anticipated trends further support the validity of this approach. Taking this technique a step forward, its potential applicability is demonstrated by predicting potential sTV values in three extra post shapes from one of our recent studies [46] . Three distinct asymmetrical post shapes (Figure 5a ,c,e) were used for predicting sTV values for S. enterica and SPN3US by employing their respective EK signature (eE EEC ). The specific positions considered for these estimations are marked with a black cross in Figure 5b ,d,f, which were 25% of the length of the left-half of the constriction. This specific location was used because it is the average distance from the center in all images analyzed. For symmetric post shapes, the highest E value exists at the exact center of the constriction, where the vertical spacing is smallest. For asymmetric posts, the highest E value is very close to the center, where the vertical spacing is smallest and the posts divide. For consistency with our previous results, in all of these simulations, the third column of posts within the channel was used. Biosensors 2020, 10, x FOR PEER REVIEW 10 of 14 posts versus the higher values for circle posts should also be noted. In general, this approach does demonstrate reasonable estimations for sTV value range. In addition, it can also be observed that the sTV values for SPN3US are significantly higher than those for S. enterica (Figure 4a-e) , which is also expected based on their significant size difference. These anticipated trends further support the validity of this approach. Taking this technique a step forward, its potential applicability is demonstrated by predicting potential sTV values in three extra post shapes from one of our recent studies [46] . Three distinct asymmetrical post shapes (Figure 5a ,c,e) were used for predicting sTV values for S. enterica and SPN3US by employing their respective EK signature ( ). The specific positions considered for these estimations are marked with a black cross in Figure 5b ,d,f, which were 25% of the length of the left-half of the constriction. This specific location was used because it is the average distance from the center in all images analyzed. For symmetric post shapes, the highest value exists at the exact center of the constriction, where the vertical spacing is smallest. For asymmetric posts, the highest value is very close to the center, where the vertical spacing is smallest and the posts divide. For consistency with our previous results, in all of these simulations, the third column of posts within the channel was used. From Figure 5 it can be observed that the estimated sTV values were highest in the oval-wideoval posts (Figure 5c,d) , followed by the oval-wide-diamond posts (Figure 5e ,f) and then by the ovalnarrow-oval posts (Figure 5a,b) for both microorganisms as reported in Table 2 . This trend agrees with theory as shown by a previous study involving these asymmetric posts [46] . As expected, the significantly larger microorganism, S. enterica bacterial cells, had lower predicted sTV values than the smaller SPN3US bacteriophages. While these values are modeling predictions only, they were estimated based on results from experimental data in other post shapes, and the trends observed in these predictions are encouraging. These results illustrate the ability to use to predict sTV in alternate post shapes, which can be crucial in the design of more robust experiments where high resolution separations are required. From Figure 5 it can be observed that the estimated sTV values were highest in the oval-wide-oval posts (Figure 5c,d) , followed by the oval-wide-diamond posts (Figure 5e ,f) and then by the oval-narrow-oval posts (Figure 5a,b) for both microorganisms as reported in Table 2 . This trend agrees with theory as shown by a previous study involving these asymmetric posts [46] . As expected, the significantly larger microorganism, S. enterica bacterial cells, had lower predicted sTV values than the smaller SPN3US bacteriophages. While these values are modeling predictions only, they were estimated based on results from experimental data in other post shapes, and the trends observed in these predictions are encouraging. These results illustrate the ability to use eE EEC to predict sTV in alternate post shapes, which can be crucial in the design of more robust experiments where high resolution separations are required. Table 2 . Predicted sTV values for S. enterica and SPN3US in each insulator-based electrokinetic (iEK) microfluidic channel. Oval-narrow-oval (Figure 5a ,b) 519 Oval-wide-oval (Figure 5c The present work illustrates a new methodology for estimating the empirical electrokinetic equilibrium condition (eE EEC ) in insulator-based devices containing posts. The EK behavior of three distinct species of bacteria and three species of bacteriophage viruses was analyzed. While there are prior approaches reporting estimations of E EEC [40] and E EEC values [13] , both approaches have limitations. The particle velocimetry approach in post-less microchannels [40] is the most accurate and can be used to also estimate mobility data, but it is highly labor intensive. In an attempt to simplify this process, Coll De Peña et al. [13] presented the second approach in which experiments are conducted in microchannels with insulating posts and relies on determination of the trapping voltage (TV). This technique requires significantly less experimentation but has to contend with the impact of DEP and particle-particle interactions on the system, hence classifying it here as eE EEC , rather than E EEC . The goal of the present study is to improve the preliminary method for the determination of the eE EEC by utilizing images of bands of microorganisms electrokinetically trapped at the stable trapping voltage (sTV) for estimating the new eE EEC parameter. The present study properly introduces the new eE EEC parameter, which has a determination process that is simpler, faster, and less labor-intensive than that of the E EEC parameter determined in devices without insulating posts [40] . The approach presented here is more robust, as it has lower user-dependence than the previous report illustrating determination in devices with posts [13] . The results from this work illustrate good precision in the estimations of eE EEC , with low standard deviations (<10%). The applicability eE EEC was further validated by performing estimations of sTV values for two microorganisms, bacteria S. enterica and bacteriophage SPN3US, which produced great agreement between the estimations and experimentally obtained sTV values (average species deviations < 6%). Moreover, the eE EEC values identified here were also used to predict the sTV values for three additional post shape designs to illustrate the versatility of eE EEC as a post shape-independent parameter. A summary of all the results obtained here, which are the unique EK signatures (eE EEC ) of these microorganisms can be found on our EK library available on our laboratory website [43] . The data in this library can be used to design iEK microfluidic devices for any applications that require the rapid identification of microorganisms. Review: Microbial analysis in dielectrophoretic microfluidic systems Review-Where Is Dielectrophoresis (DEP) Going? Alternating Current-Dielectrophoresis Collection and Chaining of Phytoplankton on Chip: Comparison of Individual Species and Artificial Communities Automated "pick and transfer" of targeted cells using dielectrophoresis Combined detection of AC-electrokinetic effects: Experiments with three-axial chicken red blood cells High-throughput dielectrophoretic filtration of sub-micron and micro particles in macroscopic porous materials Tracking Inhibitory Alterations during Interstrain Clostridium difficile Interactions by Monitoring Cell Envelope Capacitance Explorations of ABO-Rh antigen expressions on erythrocyte dielectrophoresis: Changes in cross-over frequency Analysis of bacteriophages with insulator-based dielectrophoresis Concentration of Sindbis virus with optimized gradient insulator-based dielectrophoresis Virus enrichment for single virus infection by using 3D insulator based dielectrophoresis Isolation and identification of Listeria monocytogenes utilizing DC insulator-based dielectrophoresis Creation of an electrokinetic characterization library for the detection and identification of biological cells Development of the resolution theory for gradient insulator-based dielectrophoresis Differentiation of Escherichia coli serotypes using DC gradient insulator dielectrophoresis Selective trapping of live and dead mammalian cells using insulator-based dielectrophoresis within open-top microstructures Selective isolation of live/dead cells using contactless dielectrophoresis (cDEP) Dielectrophoretic Concentration and Separation of Live and Dead Bacteria in an Array of Insulators Biophysical separation of Staphylococcus epidermidis strains based on antibiotic resistance Quantitative dielectrophoretic tracking for characterization and separation of persistent subpopulations of Cryptosporidium parvum Identification of Essential Genes in the Salmonella Phage SPN3US Reveals Novel Insights into Giant Phage Head Structure and Assembly Transcatheter Aortic Valve Implantation Represents an Anti-Inflammatory Therapy via Reduction of Shear Stress-Induced, Piezo-1-Mediated Monocyte Activation Continuous Separation of DNA Molecules by Size Using Insulator-Based Dielectrophoresis Six-Helix Bundle and Triangle DNA Origami Insulator-Based Dielectrophoresis DNA manipulation by means of insulator-based dielectrophoresis employing direct current electric fields Immunoglobulin G and bovine serum albumin streaming dielectrophoresis in a microfluidic device Protein manipulation with insulator-based dielectrophoresis and direct current electric fields Microscale electrokinetic assessments of proteins employing insulating structures Insulator-based dielectrophoresis of microorganisms: Theoretical and experimental results Passive Dielectrophoretic Focusing of Particles and Cells in Ratchet Microchannels. Micromachines 2020 Isolation and enrichment of low abundant particles with insulator-based dielectrophoresis Effects of electrothermal vortices on insulator-based dielectrophoresis for circulating tumor cell separation Dielectric Characterization and Separation Optimization of Infiltrating Ductal Adenocarcinoma via Insulator-Dielectrophoresis Biofluid pretreatment using gradient insulator-based dielectrophoresis: Separating cells from biomarkers Selective trapping of single mammalian breast cancer cells by insulator-based dielectrophoresis Nonlinear electrophoresis at arbitrary field strengths: Small-Dukhin-number analysis Weakly nonlinear electrophoresis of a highly charged colloidal particle Field-induced disturbance of the double layer electro-neutrality and non-linear electrophoresis Electrokinetic phenomena of the second kind and their applications Direct Current Electrokinetic Particle Trapping in Insulator-Based Microfluidics: Theory and Experiments Experimental and theoretical study of dielectrophoretic particle trapping in arrays of insulating structures: Effect of particle size and shape On the use of correction factors for the mathematical modeling of insulator based dielectrophoretic devices Exploiting Particle Mutual Interactions To Enable Challenging Dielectrophoretic Processes Effect of insulating posts geometry on particle manipulation in insulator based dielectrophoretic devices Continuous flow separation of particles with insulator-based dielectrophoresis chromatography The authors would like to thank Julie A. Thomas from Rochester Institute of Technology for providing the laboratory with S. enterica, SPN3US, ΦKZ and 201Φ2-1, and Abbi Miller and Victor Perez-Gonzalez for valuable insights on the manuscript. The authors declare no conflict of interest.