key: cord-0959415-iuywogc1 authors: Cutolo, Carlo Alberto; Bonzano, Chiara; Scotto, Riccardo; Iester, Michele; Bagnis, Alessandro; Pizzorno, Chiara; Catti, Carlo; Traverso, Carlo Enrico title: Moving beyond the Slit-Lamp Gonioscopy: Challenges and Future Opportunities date: 2021-12-06 journal: Diagnostics (Basel) DOI: 10.3390/diagnostics11122279 sha: e190b05b60cb23fe342c17fb7d9d6d872efa19d1 doc_id: 959415 cord_uid: iuywogc1 After almost a century from its introduction in clinical practice, slit-lamp gonioscopy is still considered the reference standard for evaluating the anterior chamber angle (ACA). Gonioscopy is essential for diagnosing angle closure disease, and ACA features are included in glaucoma’s diagnostics and treatments algorithms. However, shortcomings of slit-lamp gonioscopy include a steep learning curve, lack of agreement between examiners and poor documentation. Thanks to advances in miniaturization and computing, new instruments for digital gonioscopy have been developed and marketed. This narrative review focuses on the Gonioscope GS-1, which permits semi-automated circumferential documentation of the ACA in real-colour photographs. Advantages and disadvantages of GS-1 compared with slit-lamp gonioscopy and other ACA imaging technologies such as optical coherence tomography are discussed. Finally, potential opportunities offered by this device for telemedicine, virtual clinics, and automatic classification with deep learning are presented. Slit-lamp gonioscopy is an essential part of adult eye examination to detect angleclosure or define secondary causes of intraocular pressure (IOP) elevation [1] . Moreover, many treatments targeted at the anterior segment of the eye are delivered through gonioscopy lenses and not performing gonioscopy in patients suspected of having glaucoma may cause misdiagnosis and suboptimal treatment with a negative outcome on the disease [2, 3] . For instance, laser trabeculoplasty requires an open angle, and trabecular meshwork pigment status is informative for setting adequate power [4] . Moreover, minimally invasive glaucoma surgery (MIGS), performed standalone or in combination with cataract surgery to reduce the number of IOP-lowering drugs, requires detailed identification of angle structures for the correct placement of the device [5] . High-risk diabetic patients and eyes suffering from vaso-occlusive retinal disease may regularly need the anterior chamber angle (ACA) assessed for signs of neovascular glaucoma [6] . Additionally, eyes treated with procedures for reopening the angle such as iridotomy or iridoplasty should have the angle assessed over time because angles may continue to close [7] . For all these reasons, gonioscopy is an essential skill for diagnosis and treatment. Since the introduction of gonioscopy in clinical practice, technology has evolved, offering more advanced methods to inspect the ACA that partially overcome shortcomings of slit-lamp gonioscopy [8] . In particular, slit-lamp gonioscopy is often limited by poor documentation of findings, steep learning curve, eye contact, partial agreement between examiners, impractical photography of the observed finding and limited grading or measurements [9, 10] . This review critically discusses the clinical value of two of the most promising technologies for document the ACA, namely optical coherence tomography (OCT) and digital gonioscopy. We performed a search of electronic databases such as PubMed and Web of Science to identify studies on digital gonioscopy using the keywords: "goniophotography", "GS-1", "automatic gonioscopy", "gonioscopy", "anterior segment OCT", "optical coherence tomography", "anterior chamber angle", "iridocorneal angle". Additional sources were obtained from the reference lists of the included literature. We imposed no language or publication year restrictions. Medical Subject Headings (MeSH) searches were also performed. This article is based on previously conducted studies and does not involve any new studies of human or animal subjects performed by any authors. The first limbus indentation able to identify the ACA was performed in 1907 by Trantas. However, except for rare cases of keratoconus, keratoglobus, or severe myopia, a direct view of the ACA was not possible due to the principle of total internal reflection [11] . Contact lenses and prisms have been introduced to overcome this problem. In 1914, Salzmann proposed the first gonioscopy lens, and in 1968, Goldmann enabled the visualization of angle structures through the gonioprism still used in our clinical practice. Based on the lens used to perform slit-lamp gonioscopy, a direct or indirect approach to the ACA is possible in daily practice. Koeppe's lens allows a direct visualization of the angle. Light passes from the anterior chamber through the cornea and this contact lens, allowing a direct examination of the ACA. The one or three-mirrored Goldmann lens or the four-mirrored Zeiss lens permit an indirect view of the angle structures. In this technique, the light from the AC is reflected in a mirror, providing an inverted view of the ACA. Compared to the three-mirrored lenses, in which one of the three mirrors points toward the ACA and the others provide views of the anterior retina and Ora Serrata, in the four-mirrored, such as Zeiss, Volk G, Posner, and Susmann, all mirrored lenses point toward the ACA. Moreover, their relatively small and flat contact area allows a dynamic gonioscopy, which is still considered the reference standard. Indentation gonioscopy through a gentle pressure placed on the cornea pushed the aqueous humor into the ACA, allowing to discriminate between an appositional (reversible) versus synechial (irreversible) angle closure. Indeed, in case of an iridotrabecular contact or apposition, the angle will open, and the angle structures become visible, whereas, in the presence of peripheral anterior synechiae (PAS), the angle will remain the same in the affected areas. Indentation gonioscopy is also extremely useful to figure out a plateau iris. The anatomical landmarks of the ACA identified through the gonioscopy examination from posterior to anterior are: the iris root insertion, the ciliary body (CB), the scleral spur (SS), the trabecular meshwork (TM) and the Schwalbe's line (SL). To provide a standardized description of the ACA, many classification systems have been established over time. For example, based on the visibility of the anatomical structures of the ACA, the Scheie system, based on angularity, the Shaffer system, based on the insertion of the iris, angularity, configuration, and pigmentation of the posterior TM, the Spaeth system [12] . Gonioscopy plays an essential role in recognizing glaucoma subtype and establishing proper medical or surgical treatment interventions [4, 5, 13] . In addition to its use in the classification of glaucoma, gonioscopy aids in diagnosing secondary causes of outflow obstruction such as pigment, new vessels, iris cysts, tumors, pseudoexfoliative material, angle-recession or foreign bodies in the ACA [1, 3, 6] . Both the experience of the examiner and the illumination condition could affect the gonioscopy interpretation. For example, too much inadvertent pressure on the cornea or excessive light exposure during the examination could overestimate the anterior chamber angle width. Comparison with other methods to inspect the ACA are summarized by Table 1 . OCT was introduced as a non-invasive methodology to inspect the ACA in 1994. At that time, the system had an axial resolution of 10 µm and a lateral resolution of 22 µm; the image acquisition rate was slow, and because 0.8-mm wavelength light cannot penetrate the sclera ACA structures were poorly visualized [14] . Time-domain OCT using 1.3-mm wavelength were then developed allowing increased penetration through scattering angle structures so that ACA morphology could be visualized with more detail. Using the scleral spur (SS) as a landmark, it was then possible to quantify the ACA width using parameters previously introduced for the UBM including the angle opening distance (AOD) at 500 µm from the scleral spur, the angle recess area (ARA) and the trabeculo-iris space [15] . With OCT technology, precise measurements made it possible to measure dynamic changes in the ACA occurring with dark-light variations [16] , quantify ACA morphology modifications after laser peripheral iridotomy [17, 18] or phacoemusification [19, 20] . While OCT offers non-contact and is less dependent on observer's training than slit-lamp gonioscopy, it also presents several drawbacks. Firstly, correct angle classification and measurement necessitate precise landmark identification. In a study comparing gonioscopy and OCT for ACA assessment in glaucoma patients, SS was identified only in 56% and 50% of quadrants with a time-domain OCT and a spectral-domain OCT, respectively. Over time spectral-domain (SD) systems have an increased image acquisition speed and improved image resolution compared with time-domain systems, the shorter wavelength of 840 nm limits tissue penetration and imaging of the ACA [21] . To overcome these limitations, the use of the Schwalbe lines as a landmark has been proposed and validated by several studies. Moreover, swept-source OCT permit fast image acquisition rate with higher tissue penetration compared to spectral-domain systems and resulted in more reliable ACA imaging; Secondly, OCT imaging does not permit acquiring features of the ACA that are essential for glaucoma diagnosis and classifications. Intensity and patterns of ACA pigmentation may offer clues on secondary causes of glaucoma as pseudoexfoliation or pigment dispersion. In case of neovascular glaucoma, fine abnormal neovessels of the ACA may precede the synechial angle closure and slit-lamp examination and slitlamp gonioscopy should not be replaced by OCT for follow-up high risk patients; Finally, moderate agreement exists in the detection of angle closure between slit-lamp gonioscopy and OCT. The OCT systems tend to overestimate angle closure compared to slit-lamp gonioscopy as the reference standard [8, 22] . Beside these limitations, OCT may foster ACA assessment by permitting imaging of the trabecular meshwork and on the Sclemm's canal and the band of extracanalicular limbal lamina with relevance to diagnosis and treatments [23] . Recently, features of angle dysgenesis evaluated with OCT in juvenile onset open-angle glaucoma were associated with poor selective laser trabeculoplasty success making this technology complementary to gonioscopy [24] . Moreover, microscopeintegrated OCT are commercially available and provide valuable information before, during and at the end of anterior segment or posterior pole surgeries [25] . A further advancement in ACA evaluation is represented by automatic gonioscopy that permits acquiring digital photographic documentation of the ACA. Figure 1 shows the NIDEK GS-1 Automated Gonioscope (NIDEK Co., Ltd., Gamagori, Japan) able to acquire 360 • true color images of the ACA [26, 27] . The instrument utilizes a multi-mirror prismatic lens with 16 reflecting facets, designed and developed specifically for this system and unique in its kind; in association with an illumination system provided with a white LED and a high-resolution color camera. The NIDEK multi-mirror prismatic lens was developed by optimizing each surface for automatic angle detection. Each of the 16 prism mirrored facets projects a white light onto a 22.5 • portion of the irido-corneal angle. The camera takes 17 pictures at varying focal planes, from each of the 16 gonioprism facets in order to obtain a good level of focus. The best focused image acquired in every facet is automatically selected by the software and can be rendered in a unique, linear or circular, image of the entire irido-corneal angle called Stitching. The examination takes at least about 1 min, thus requiring good fixation and reasonable patient cooperation; it is advisable to perform it after instillation of topical anaesthetic. This system uses immersion gels to ensure patient comfort, the multi-mirror prism The examination takes at least about 1 min, thus requiring good fixation and reasonable patient cooperation; it is advisable to perform it after instillation of topical anaesthetic. This system uses immersion gels to ensure patient comfort, the multi-mirror prism lens is not designed to contact directly the cornea. The acquisition procedure is divided into two phases, the first involving the correct alignment of the camera with the ocular angular structures. The images from the upper, lower, nasal and temporal areas of the eye will appear on the instrument screen; the device indicates the state of alignment suggesting the movements to be made by the operator in order to get the best image quality. A first click on the acquisition joystick allows obtaining an automatic alignment of the angular structures, which the software is able to recognize autonomously regardless of the eyes color. The second phase is the actual acquisition; once the correct viewing of the angle in at least three of four sectors has been achieved, the instrument automatically acquires the sixteen gonioscopic images at various focal planes by the rotating system. It is possible to make a manual acquisition by simply clicking the joystick again if the ideal alignment established by the software is not reached. Each of the sixteen acquired sectors can be evaluated and saved by the operator with the intent of selecting the image captured with the best focus helped by the software that automatically highlights the most in-focus image for each sector. With all sixteen sections it is possible to create a single panoramic view called Stitching, linear or circular, which represents the entire structure of the iridocorneal angle. The 360-degree view of the angle is available in 3 available formats (16-section display, a circular display, or a linear display). Images can be saved in the built-in storage system or exported and transmitted for remote assessment. Figure 2 shows examples of pathological and post-surgical findings imaged with GS-1. A pilot study on GS-1 performance reported the clinical utility of the instrument in assessing various clinical conditions including angle-closure, narrow angle and surgical devices and postsurgical conditions including CyPass, iStent, XEN, Baerveldt tube shunt, angle recession with iridodialysis, angle neovascularization, pigment dispersion, post-trabeculectomy sclerostomy and surgical iridectomy, and post-laser peripheral iridotomy [27] . Post-surgical findings were also evaluated by Barão et al. who analyzed the location of the XEN stent imaged by GS-1. They found that that different locations of XEN did not seem to significantly impact late success and that combined implantation of XEN with phacoemulsification does not seem to influence stent location [28] . More recently, Matsuo et al. used the GS-1 device to investigate the pattern of peripheral anterior synechia (PAS) formation after ab-interno trabeculotomy and they found that PAS formation rate was significantly higher within the incision compared to other angle sectors. GS-1 was also used to assess the pigmentation in the ACA using automatic and manual method with good agreement between the two techniques [29] . Agreement with slit-lamp gonioscopy has not yet been well characterized. In a pilot study, Texeira et al. reported that manual and automated gonioscopy showed only slight agreement for the assessment of iridocorneal angle opening status. In particular, angle closure was detected in 23.4% with slit-lamp gonioscopy in comparison with 4.3% using GS-1 [30] . Additionally, agreement between different graders of the ACA images acquired with the GS-1, although limited, could be improved by sufficient training and a solid consensus [29] . Peroni et al. recently better characterized the agreement on identification of the ACA structures between graders. Using a software annotation tool that permitted tracing the contours of anatomical layers, Peroni et al. found that scleral spur had the minimum agreement whereas it was best on iris root, trabecular meshwork and cornea [31] . Based on these studies, deep learning semantic segmentation algorithms for processing images acquired by GS-1 are currently in development [32] [33] [34] . A pilot study on GS-1 performance reported the clinical utility of the instrument in assessing various clinical conditions including angle-closure, narrow angle and surgical devices and postsurgical conditions including CyPass, iStent, XEN, Baerveldt tube shunt, angle recession with iridodialysis, angle neovascularization, pigment dispersion, post-trabeculectomy sclerostomy and surgical iridectomy, and post-laser peripheral iridotomy [27] . Post-surgical findings were also evaluated by Barão et al. who analyzed the location of the XEN stent imaged by GS-1. They found that that different locations of XEN did not seem to significantly impact late success and that combined implantation of XEN with phacoemulsification does not seem to influence stent location [28] . More recently, Matsuo et al. used the GS-1 device to investigate the pattern of peripheral anterior synechia (PAS) Poor accessibility to appropriate ophthalmological care is associated with inaccurate diagnosis and treatment of glaucoma. In particular, glaucoma requires prompt diagnosis and classification for an effective treatment as well as periodic follow-up visits to detect progression that requires additional therapies. Moreover, the number of patients affected by glaucoma is increasing worldwide and with the current COVID-19 pandemic healthcare systems are under pressure like never before [35, 36] . Virtual glaucoma clinics are designed to acquire patient's data in a reliable and efficient way by technicians to be reviewed by an ophthalmologist, thus maximising appointment capacity and reducing waiting times. There is evidence that patients report higher levels of satisfaction of this format and non-inferior knowledge of their condition than those seen in standard clinics [37] . Among the essential exams for glaucoma management there are slit-lamp examination, standard automated perimetry, optic nerve imaging or photography, tonometry, and ACA assessment. Among methods for angle assessments, both goniophotography and OCT have been tested for remote grading showing levels of agreement ranging from fair to excellent with the ground truth gonioscopic examination [38, 39] . When considering ACA assessment without standard slit-lamp gonioscopy, OCT could have advantages over goniophotography for screening for angle closure disease because it does not require contact and is easier to perform. However, goniophotography is more informative for detecting signs of secondary glaucoma or inspecting angle sectors subject to previous angle surgeries or device placement. EyeCam permits acquiring clinically gradable images but it is limited by an acquisition time of approximately 5 to 10 min per eye and by its requirement for supine patient positioning [39] . Viceversa, GS-1 is performed with the patient in sitting position and the entire angle is photographed in less than a minute, probably making this technology more appealing for a virtual clinics setting. Technologies for the assessment of the ACA have evolved in recent years thanks to advancements in the miniaturization of mechanical, optical, and electronic devices. Also, software development and image processing have permitted the realization of precise measurements of the morphology of the ACA [40] . On the other hand, slit-lamp gonioscopy is still considered the reference standard and an essential skill for the ophthalmologist [1] . New technologies have made it possible to acquire digital images of the ACA, ameliorating documentation, comparisons over time and permitting remote diagnosis and image analysis. Both OCT and digital gonioscopy are fast and easy techniques to perform and could be suitable for virtual clinics, whereas slit-lamp gonioscopy requires considerable training. Actual therapeutical algorithms and angle classification are based on slit-lamp gonioscopy findings and there is uncertainty over whether the use of new methods to assess the ACA are associated with better clinical outcomes. Digital gonioscopy and OCT role in patient's management is not yet completely understood. Ultrasound biomicroscopy, even if not discussed in detail in this review, has a role when complex angle-closure mechanisms are suspected because of its ability to image behind the iris [41] . All the complementary techniques for imaging the ACA should be better considered as complementary exams to slit-lamp gonioscopy and not as alternatives [41] . OCT has shown a potential role in predicting the outcome of laser treatments [42] [43] [44] [45] . Digital gonioscopy has been shown useful in documenting post-operative and abnormal findings of the ACA and has a great potential for teaching gonioscopy. Deep learning algorithms for automatic classification of the angle are on the horizon and their validity in the clinical setting should be accurately evaluated [31] [32] [33] . Informed Consent Statement: Written informed consent has been obtained from the patients to publish images in this paper. The authors declare no conflict of interest. Terminology and Guidelines for Glaucoma, 4th Edition-Part 1-Supported by the EGS Foundation The Philadelphia Glaucoma Detection and Treatment Project: Detection Rates and Initial Management Anterior Chamber Angle Assessment Techniques: A Review Selective Laser Trabeculoplasty versus Eye Drops for First-Line Treatment of Ocular Hypertension and Glaucoma (LiGHT): A Multicentre Randomised Controlled Trial Minimally-Invasive Glaucoma Surgeries (MIGS) for Open Angle Glaucoma: A Systematic Review and Meta-Analysis Evidence-Based Recommendations for the Diagnosis and Treatment of Neovascular Glaucoma1 1The Authors Do Not Have Any Proprietary or Financial Interest in Any Products or Devices Discussed in This Study Longitudinal Changes of Angle Configuration in Primary Angle-Closure Suspects: The Zhongshan Angle-Closure Prevention Trial Recent Advances in Anterior Chamber Angle Imaging Perceptions of Training in Gonioscopy Historical Notes on Gonioscopy Uses Old and New. The Inheritance of Occludable Angles Anterior chamber angle assessment techniques Micrometer-Scale Resolution Imaging of the Anterior Eye In Vivo With Optical Coherence Tomography Optical Coherence Tomography Imaging of the Anterior Chamber Angle Comparisons of Anterior Segment Biometry between Chinese and Caucasians Using Anterior Segment Optical Coherence Tomography Changes in Anterior Segment Morphology in Response to Illumination and after Laser Iridotomy in Asian Eyes: An Anterior Segment OCT Study Anterior Segment Optical Coherence Tomography for Imaging the Anterior Chamber after Laser Peripheral Iridotomy Comparing Laser Peripheral Iridotomy to Cataract Extraction in Narrow Angle Eyes Using Anterior Segment Optical Coherence Tomography Artificial Intelligence Applications and Cataract Management: A Systematic Review Evaluation of the Anterior Segment Angle-to-Angle Scan of Cirrus High-Definition Optical Coherence Tomography and Comparison With Gonioscopy and With the Visante OCT Assessment of the Scleral Spur in Anterior Segment Optical Coherence Tomography Images Characterizing Anterior Segment OCT Angle Landmarks of the Trabecular Meshwork Complex In Vivo Imaging of the Schlemm's Canal and the Response to Selective Laser Trabeculoplasty What Is the Impact of Intraoperative Microscope-Integrated OCT in Ophthalmic Surgery? Relevant Applications and Outcomes. A Systematic Review Anterior Chamber Angle Imaging with Automatic Gonio-Photography Novel and Semiautomated 360-Degree Gonioscopic Anterior Chamber Angle Imaging in Under 60 Seconds Automated Gonioscopy Assessment of XEN45 Gel Stent Angle Location After Isolated XEN or Combined Phaco-XEN Procedures: Clinical Implications Automated Anterior Chamber Angle Pigmentation Analyses Using 360 • Gonioscopy Automated Gonioscopy Photography for Iridocorneal Angle Grading On Clinical Agreement on the Visibility and Extent of Anatomical Layers in Digital Gonio Photographs A Deep Learning Approach for Semantic Segmentation of Gonioscopic Images to Support Glaucoma Categorization Semantic Segmentation of Gonio-Photographs via Adaptive ROI Localization and Uncertainty Estimation Semantic Segmentation of Gonioscopic Images Exploiting Adaptive ROI Localization and Uncertainty Estimation Effect of COVID-19-Related Lockdown on Ophthalmic Practice in Italy: A Report from 39 Institutional Centers Early impact of COVID-19 outbreak on eye care: Insights from EUROCOVCAT group Virtual Glaucoma Clinics: Patient Acceptance and Quality of Patient Education Compared to Standard Clinics Remote Grading of the Anterior Chamber Angle Using Goniophotographs and Optical Coherence Tomography: Implications for Telemedicine or Virtual Clinics A Population-Based Assessment of the Agreement Between Grading of Goniophotographic Images and Gonioscopy in the Chinese-American Eye Study (CHES) Differences in Anterior Chamber Angle Assessments Between Gonioscopy, EyeCam, and Anterior Segment OCT: The Chinese American Eye Study The Effects of Phacoemulsification and Intraocular Lens Implantation on Anatomical and Functional Parameters in Patients with Primary Angle Closure: A Prospective Study. (An American Ophthalmological Society Thesis) Anatomic Changes and Predictors of Angle Widening after Laser Peripheral Iridotomy: The Zhongshan Angle Closure Prevention Trial Changes in Intraocular Pressure and Angle Status After Phacoemulsification in Primary Angle Closure Hypertension Novel Gonioscopy Score and Predictive Factors for Intraocular Pressure Lowering After Phacoemulsification