key: cord-0959114-ugtthiiw authors: Sultan, Laith R.; Chen, Yale Tung; Cary, Theodore W.; Ashi, Khalid; Sehgal, Chandra M. title: Quantitative pleural line characterization outperforms traditional lung texture ultrasound features in detection of COVID‐19 date: 2021-04-02 journal: J Am Coll Emerg Physicians Open DOI: 10.1002/emp2.12418 sha: 6e09e9b11c43daba73759ae5d8b7c9f6ca125e0d doc_id: 959114 cord_uid: ugtthiiw BACKGROUND AND OBJECTIVE: Lung ultrasound is an inherently user‐dependent modality that could benefit from quantitative image analysis. In this pilot study we evaluate the use of computer‐based pleural line (p‐line) ultrasound features in comparison to traditional lung texture (TLT) features to test the hypothesis that p‐line thickening and irregularity are highly suggestive of coronavirus disease 2019 (COVID‐19) and can be used to improve the disease diagnosis on lung ultrasound. METHODS: Twenty lung ultrasound images, including normal and COVID‐19 cases, were used for quantitative analysis. P‐lines were detected by a semiautomated segmentation method. Seven quantitative features describing thickness, margin morphology, and echo intensity were extracted. TLT lines were outlined, and texture features based on run‐length and gray‐level co‐occurrence matrix were extracted. The diagnostic performance of the 2 feature sets was measured and compared using receiver operating characteristics curve analysis. Observer agreements were evaluated by measuring interclass correlation coefficients (ICC) for each feature. RESULTS: Six of 7 p‐line features showed a significant difference between normal and COVID‐19 cases. Thickness of p‐lines was larger in COVID‐19 cases (6.27 ± 1.45 mm) compared to normal (1.00 ± 0.19 mm), P < 0.001. Among features describing p‐line margin morphology, projected intensity deviation showed the largest difference between COVID‐19 cases (4.08 ± 0.32) and normal (0.43 ± 0.06), P < 0.001. From the TLT line features, only 2 features, gray‐level non‐uniformity and run‐length non‐uniformity, showed a significant difference between normal cases (0.32 ± 0.06, 0.59 ± 0.06) and COVID‐19 (0.22 ± 0.02, 0.39 ± 0.05), P = 0.04, respectively. All features together for p‐line showed perfect sensitivity and specificity of 100; whereas, TLT features had a sensitivity of 90 and specificity of 70. Observer agreement for p‐lines (ICC = 0.65–0.85) was higher than for TLT features (ICC = 0.42–0.72). CONCLUSION: P‐line features characterize COVID‐19 changes with high accuracy and outperform TLT features. Quantitative p‐line features are promising diagnostic tools in the interpretation of lung ultrasound images in the context of COVID‐19. Coronavirus disease 2019 (COVID- 19) was declared a pandemic caused by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) after spreading to >180 countries by March 2020, 1 with >100 million cases confirmed and 2 million deaths. 2 Medical imaging plays a major role in COVID-19 diagnosis and management. 3 The most common modality for imaging COVID-19 patients is chest x-ray, which detects presence of a local or bilateral patchy shadowing infiltrate, but it is known for low sensitivity, with findings absent in >40% of the cases. 4 Computed tomography (CT) scan has higher sensitivity and shows ground-glass opacities 5 and, therefore, has been used in different therapeutic and triage strategies since the outbreak started. 6 The use of chest CT remains limited because of radiation exposure concerns and lack of availability in overextended healthcare facilities. 7 In the critically ill, the transport of unstable patients and exposure of infected patients also may outweigh the clinical benefit. There continues to be need for alternative imaging methods that enable quick, low-cost, easyto-use evaluation of COVID-19 patients. Lung ultrasound is a promising tool for lung monitoring in the intensive care unit (ICU), 8, 9 particularly for assessing lung aeration. [10] [11] [12] A decade of clinical and physical studies demonstrates that lung ultrasound is able to detect interstitial lung disease, subpleural consolidations, and acute respiratory distress syndrome with different etiologies. With new COVID-19, evolving evidence from clinical practice and several studies shows the usefulness of lung ultrasound for the management of pneumonia, from diagnosis to monitoring and follow-up. [13] [14] [15] [16] [17] Characteristic ultrasound findings have been reported that can help physicians detect and stage the disease, tracking its progression. 13 The common patterns observed include irregular, thickened pleural lines, multiple B-lines (vertical artifacts) ranging from focal to diffuse, which reflect the stage of inflammatory lung disease. These vertical artifacts of different shapes and lengths occur when the lung loses normal aeration but is not completely consolidated. These findings are usually bilateral with posterior basal predominance in the lungs. Although there is a general consensus on lung ultrasound's usefulness for COVID-19 pneumonia, 18 it is an inherently user-dependent modality and, without proper training, could result in errors. 19 Limited experience with COVID-19 further adds to the challenge of using lung ultrasound effectively. To overcome these limitations, a number of studies have evaluated quantitative methods for assessing lung ultrasound. [20] [21] [22] These studies assess the use of automated detection of the B-line to provide critical visual information to clinicians in real time for diagnosis. In this study, we propose a comprehensive approach for detecting This multiple-center, retrospective pilot study was conducted on 20 B-mode ultrasound images that were used to evaluate the proposed quantitative analysis. Ten images were acquired from COVID- This study evaluates a novel quantitative approach for defining characteristics of the lung pleural lines, which performed better in the detection of COVID-19 when compared with traditional qualitative analysis of lung ultrasound findings. from the lung images ROIs as grayscale first-order statistics, and determined by run-length and gray-level co-occurrence matrices (GLCM). 25 The 7 image features measured are described in Appendix 1. In brief, the TLT features computation indirectly represents the real texture of the tissue by the non-deterministic properties that govern the distributions and relationships between the grey levels of the ultrasound image. The quantitative features measured demonstrate the changes in lung tissue texture, and should capture and quantify the prominence of A-and B-lines used by human to discriminate lung ultrasound patterns found in COVID-19 patients from the appearance of normal lung. Standard descriptive statistics were computed for the features extracted from the images: arithmetic mean and standard error. The 2tailed Student's t test of unequal variance was used to determine the statistical significance of the difference between 2 groups. P < 0.05 was considered significant. To evaluate the reproducibility of the analysis, observer agreement for the same individual (intraobserver) and between 2 individuals Qualitative image findings were provided by an emergency physician experienced with lung ultrasound for both normal and COVID-19 images. Normal images were described as having a thin, well-defined From the 7 p-line features extracted from lung ultrasound images, 6 showed statistically significant differences (P < 0.05) between normal and COVID-19 cases. Figure 2 shows a comparison of the mean and The diagnostic performance of TLT features TLT model, whose features were closely distributed and even overlapping ( Figure 6 ). Figure 7 shows the cases that were incorrectly Similarly, in Figure 7B , a normal case was labeled incorrectly as COVID-19, but p-line features extracted by the computer correctly diagnosed the case. The cases that were misdiagnosed showed a TLT pattern that was overlapping between the two groups: COVID-19 and normal cases. The individual observer analyzed the same set of images 2 weeks following the first analysis. Interclass correlation coeffecients (ICCs) between the 2 analyses on average for p-line features showed excellent agreement of ICC of 0.85 ± 0.09 ( High agreement levels were recorded between 2 observers who analyzed the same set of images ( Table 2 ). The average ICC for p-line fea-tures was 0.83 ± 0.12, excellent agreement. The feature that showed the highest agreement was p-line thickness with ICC of 0.98, whereas p-line brightness showed the lowest agreement (ICC = 0.63). For TLT, overall agreement was good with ICC = 0.67 ± 0.10. Highest agreement in a feature was in GL non-uniformity (ICC = 0.87); whereas, the lowest was seen in TLT heterogeneity with ICC = 0.56 (Table 2 ). Intra-and interobserver agreement were calculated for analyses using another set of images ( Table 2 ). For both intra-and interobserver, ICCs The study is not without limitations. One limitation is the small sample Another limitation is that the current analysis is used to differentiate 1 category of inflammatory diseases related to COVID-19. In future, the methodology could be applied to a variety of lung inflammatory diseases. Ultrasound imaging is known to be user dependent where imaging presets, nature of the imaging device, and the training of the observer can influence diagnosis. Although quantitative analysis minimizes these effects, a broader study involving images acquired from multiple clinical devices with varying imaging presets used by different observers will generalize the use of quantitative p-line and TLT for machine-agnostic diagnosis. Ultrasound is an ideal imaging tool for COVID-19 diagnosis because of its high sensitivity, safety, portability, and affordability. 28 However, a significant disadvantage is that it is highly user dependent, and not all clinicians have training in performing lung ultrasound and reading the images. Operator experience may also affect specificity, becausee an expert will correlate different lung ultrasound patterns with different disease processes. 29 To overcome these problems, a number of studies have evaluated quantitative assessment methods [20] [21] [22] to help physicians in image interpretation. However, the focus of these studies is limited to developing semi-quantitative scoring systems With progression of disease, the lungs are filled with inflammatory cells as well as fluids that cause the ultrasound beam to be trapped between the inflammatory cells, producing the vertical artifacts called B-lines. 12, 13 These lines are brighter and more uniform compared to A-lines seen in normal lungs. We were able to quantify these findings but did not observe a significant difference between the 2 groups, and it is unknown if the difference could become significant with more where minute differences between images were found to significantly influence the scoring process and ultimately the final diagnosis. These findings underscore the importance of image selection in clinical settings. In conclusion, we introduced a computer-based system that captures p-line pattern changes associated with COVID-19 by quantitative analysis of lung ultrasound. Quantitative p-line features showed high accuracy in detecting COVID-19 cases compared to TLT features, which can be more uncertain. These results suggest that a comprehensive quantitative system that characterizes p-lines would improve the diagnostic accuracy of COVID-19 on lung ultrasound. However, future studies on a larger scale are needed before translating these techniques to clinical practice. In the current study, the automated methods were performed off-line but these methods can be easily integrated into the operation of scanner for real-time bedside assessment. Future larger studies will incorporate advanced machine learning methods to optimize the p-line detection algorithm for robustness and automation. Given more cases, fully automated machine learning segmentation could be performed. This technology when implemented successfully in clinical practice will increase confidence in diagnosis, especially in low-resource communities around the globe that lack experience in lung ultrasound. The authors thank Ms. Susan M. Schultz for her kind contribution to the study. The authors have declared no conflict of interest. For region R = {x i ,y i },ranging over N x horizontal values, the mean thickness is the mean of the y ranges taken through the region at each horizontal coordinate.d The perimeter of the lesion divided by the circumference of its best-fit ellipse, also called elliptically normalized circumference. where P is the region's perimeter and C is the circumference of a best-fit ellipse to the region. Projected deviation measures the horizontal (transverse) irregularity in a region's gray level by computing the standard deviation of the depth-projected mean intensity. Given image image I x,y of dimensions N x, N y , it is given by the formula to the right. The probability that the points in a region R = {x i ,y i } will lie on their linear regression fit a + bx i , where the regression minimizes χ 2 . This probability is the same as the probability of a t test with the 2 degrees of freedom (a,b) on the regression fit. , where χ 2 is the chi-square of a linear fit, given by: Brightness and heterogeneity are first-order histogram features: the mean brightness level of the segmented pleural lines and the measured variability in mean brightness, respectively. Coronavirus disease (COVID-2019) situation reports. World Heal Organ An interactive web-based dashboard to track COVID-19 in real time Correlation of chest CT and RT-PCR testing for coronavirus disease 2019 (COVID-19) in China: a report of 1014 cases Clinical characteristics of coronavirus disease 2019 in China Radiological findings from 81 patients with COVID-19 pneumonia in Wuhan, China: a descriptive study CT features of SARS-CoV-2 pneumonia according to clinical presentation: a retrospective analysis of 120 consecutive patients from Wuhan city Correlation between chest computed tomography and lung ultrasonography in patients with coronavirus disease 2019 (COVID-19) Ultrasound for "lung monitoring" of ventilated patients Imaging in acute respiratory distress syndrome Bedside ultrasound assessment of positive end-expiratory pressureinduced lung recruitment Ultrasound assessment of antibiotic-induced pulmonary reaeration in ventilator-associated pneumonia Ultrasound assessment of lung aeration loss during a successful weaning trial predicts postextubation distress* A review of early experience in lung ultrasound in the diagnosis and management of COVID-19 A preliminary study on the ultrasonic manifestations of peripulmonary lesions of non-critical novel coronavirus pneumonia (COVID-19) The continuing 2019-nCoV epidemic threat of novel coronaviruses to global health-The latest 2019 novel coronavirus outbreak in Wuhan The diagnosis of pneumonia in a pregnant woman with COVID-19 using maternal lung ultrasound Lung ultrasound and computed tomographic findings in pregnant woman with COVID-19 Lung ultrasound in the monitoring of COVID-19 infection Sources of error in emergency ultrasonography Automatic detection of B-Lines in in vivo lung ultrasound Localizing B-lines in lung ultrasonography by weakly supervised deep learning, in-vivo results Automatic classification of pediatric pneumonia based on lung ultrasound pattern recognition Lung ultrasound for patients with coronavirus disease 2019 pulmonary disease A region growing and merging algorithm to color segmentation Texture feature analysis for different resolution level of kidney ultrasound images Intraclass Correlation Coefficient Guidelines, criteria, and rules of thumb for evaluating normed and standardized assessment instruments in psychology The utility of lung ultrasound in COVID-19: a systematic scoping review Point of care lung ultrasound is useful when screening for CoVid-19 in emergency department patients Modified lung ultrasound score for assessing and monitoring pulmonary aeration Modifizierter Lungen-US-Score zur Bewertung und Überwachung der Belüftung der Lunge. Modif Lung Ultrasound Ultraschall Med Chest Computed tomography findings in COVID-19 and influenza: a narrative review First order-histogram features that describe the brightness of the tissue (mean) and the variability in brightness (standard deviation).where I x,y is the intensity (grayscale) of the echo line at coordinate [x,y]. This is a run-length-matrix (RLM) feature that measures disorderliness of homogeneous runs of gray along defined directions. The RLM p gives the length of homogeneous runs for each gray level, computed over 8 directions (vertical up and down, horizontal left and right, diagonals), so that element (i,j) is the number of homogeneous runs of j pixels with intensity i.where H is the total count of homogeneous runs in p, the run-length matrix defined at left. This is a run-length matrix (RLM) feature that measures the disorderliness of the lengths of homogeneous runs, computed similarly to gray-level non-uniformity above, with the inner and outer sums switched.where H is the total count of homogeneous runs, and p is the RLM defined above. The mean of the GLCM (p), the co-occurrence matrix which records the counts of pixel intensity combinations occurring between neighboring pixels, computed over 8 directions (up and down, left and right, diagonals). Element (i,j) is the count of times a pixel of intensity i had a pixel of intensity j next to it in any direction at some scale. This is a measure of orderliness, and records whether small patterns repeat themselves.where p is the GLCM as defined at left. Also called the inverse difference moment, homogeneity increases if the region has less contrast.where p is the GLCM as defined above. Larger entropy means the texture is more disordered.where p is the GLCM as defined above.