key: cord-0956541-6xf7rjuf authors: Iredahl, Fredrik; Muller, David A.; Togö, Totte; Jonasson, Hanna; Baker, Ben; Anderson, Chris D.; Henricson, Joakim title: Local Response and Barrier Recovery in Elderly Skin Following the Application of High-Density Microarray Patches date: 2022-04-10 journal: Vaccines (Basel) DOI: 10.3390/vaccines10040583 sha: 6104fbaf3f106b4dcc97016baec287ef24f95156 doc_id: 956541 cord_uid: 6xf7rjuf The high-density microneedle array patch (HD-MAP) is a promising alternative vaccine delivery system device with broad application in disease, including SARS-CoV-2. Skin reactivity to HD-MAP applications has been extensively studied in young individuals, but not in the >65 years population, a risk group often requiring higher dose vaccines to produce protective immune responses. The primary aims of the present study were to characterise local inflammatory responses and barrier recovery to HD-MAPs in elderly skin. In twelve volunteers aged 69–84 years, HD-MAPs were applied to the forearm and deltoid regions. Measurements of transepidermal water loss (TEWL), dielectric permittivity and erythema were performed before and after HD-MAP application at t = 10 min, 30 min, 48 h, and 7 days. At all sites, TEWL (barrier damage), dielectric permittivity (superficial water);, and erythema measurements rapidly increased after HD-MAP application. After 7 days, the mean measures had recovered toward pre-application values. The fact that the degree and chronology of skin reactivity and recovery after HD-MAP was similar in elderly skin to that previously reported in younger adults suggests that the reactivity basis for physical immune enhancement observed in younger adults will also be achievable in the older population. As the world enters the third year of the COVID-19-pandemic, the emergence of new SARS-CoV-2 variants such as Omicron are expected to cause the continued need for additional vaccine boosters to protect risk groups [1] . At the beginning of the COVID-19 pandemic, age was quickly identified as a prominent risk factor [2] , making vaccination of the elderly a priority. Typically, most vaccines are administered via subcutaneous or intramuscular routes [3] . However, these vaccine delivery methods have disadvantages linked to the requirement of administration by health care professionals, the generation Vaccines 2022, 10, 583 2 of 11 of contaminated sharp waste, and the risk of local side effects related to injection [4] . An alternative route of vaccination is by microneedles, which aims at the intradermal delivery of the vaccine. We have experience with a high-density microarray patch (HD-MAP) vaccine delivery platform. The HD-MAP applied to the skin provides practical advantages, including the absence of sharps, a reduced dependence or total removal from the cold chain, and the possibility for self-application through the use of a disposable applicator [5] [6] [7] . The HD-MAP delivers the vaccine antigen to the epidermal and dermal layers of the skin containing a high density of dendritic cells. As a function of this targeted vaccine delivery, our group and other researchers have demonstrated that enhanced immune responses with lower doses of vaccine are achievable as compared to traditional needle-based vaccination methods [8, 9] . Previous studies have demonstrated the broad utility of the HD-MAP delivery platform, which has shown promising results against the dengue virus [10] [11] [12] , influenza [13, 14] , West Nile virus, Chikungunya virus [15] , Japanese encephalitis [16] , and poliovirus [17, 18] . Recently, pre-clinical studies using the HD-MAP have shown promising results producing broadly high titre neutralizing antibody responses against the SARS-CoV-2 ancestral virus and variants of concerns [19] . To date, previously reported clinical studies evaluating microneedle vaccinations have been performed in young adults in whom the skin is relatively reactive to external stimulus/provocation resulting in adaptive and innate immune reactivity [20] [21] [22] . Previous experience of vaccination in the elderly, for instance against influenza, has shown that it is necessary to apply an increased dose of the vaccine to maintain efficacy [23, 24] . The kinetics of skin barrier recovery following penetration by HD-MAP is relevant for subject acceptability. Quantification of transepidermal water loss (TEWL) by evaporimetry is a classic method for assessing the barrier function in the skin. While there is always a background level of water vapour escaping the surface, once the HD-MAP punctures the skin, the integrity of the stratum corneum is lost and the rate of water loss from the viable epidermis increases markedly. Thus, measurements of TEWL can serve as an indicator of successful HD-MAP engagement as well as being per se an illustration of the induction of an innate reactivity and of an individual's wound healing kinetics after the minor trauma caused by the HD-MAP application. In younger individuals, we have previously observed skin barrier recovery within 48 h observed by TEWL [20] , and the resistance to topical histamine application after the same time period [25] . Non-invasive methodology allows documentation of skin barrier penetration/damage, induction of reactive inflammation, and the possibility to follow changes over time, which are the advantageous attributes of microneedle vaccine administration, without the discomfort and possible minor side events of the skin biopsy technique. The aim of this study was therefore to study the reactive capability of elderly skin to the application of uncoated (not containing vaccine) HD-MAPs to the skin of the volar forearm and deltoid region, quantifying reactivity by measuring transepidermal waterloss as an indicator of both barrier penetration and reactivity as well as superficial skin hydration and erythema as indicators of inflammatory response. Twelve volunteers, with a mean age ± standard deviation (range min to max) of 75 ± 5 (69 to 84) years and a body mass index of 27 ± 5 (21 to 37) kg/m 2 , were recruited and gave their written informed consent. The volunteers were asked to refrain from drinking coffee, tea, or alcohol and from exercise on the day of each measurement. The volunteers were seated comfortably in a semi-supine position during the measurements, which started after 20 min of acclimatization. Provocations and measurements were undertaken at the forearm and deltoid regions. Lighting conditions in the room were kept constant by closing window blinds and turning off ambient lights during measurements with the polarisation spectroscopy device. The room temperature was maintained at 21 ± 1 • C. The non-invasive nature of the assessment methods facilitated the study of individual Vaccines 2022, 10, 583 3 of 11 phenotype, variability due to the presence or severity of disease as well as the possible influence of interactivity between therapeutic agents and vaccine effectiveness. The study was approved by the Regional Ethics Committee at Linköping University Hospital (Dnr. MB 2017-409-31, 18 October 2017). Participants received one HD-MAP applied to the volar forearm and one applied to the deltoid region on the non-dominant arm (Figure 1 ), as previously described by Muller et al. (2020) [20] . The application sites were free from hair, tattoos, scars, and visible veins. All areas were cleaned with an alcohol swab prior to application. Before application, the sites were compressed with a pre-calibrated skin conditioning ring (Vaxxas Pty Ltd., Brisbane, Australia) at a force of 30 N, to equalise the resistance of the skin between volunteers. A preloaded applicator device (Vaxxas Pty Ltd., Brisbane, Australia) was then docked into a skin conditioning ring. The applicator was triggered to apply the HD-MAP at a speed of 20 m/s. Once applied, HD-MAPs were left in place for 2 min. Immediately after removal, pain and local signs, erythema, oedema, and the presence of minimal breakthrough or minor "wet" bleeding were assessed clinically by trained staff, and skin physiological measurements were performed. Wet bleeding was defined as the presence, immediately after patch removal, of multiple pinpoint blood spotting on an applied tissue paper. Perceived pain on a scale from 0 (no pain) to 10 (worst imaginable pain) was reported by the volunteers at 10-and 30-min post-HD-MAP application. Measurements were recorded in the order of transepidermal water loss (evaporimetry), dielectric permittivity, erythema (polarisation spectroscopy), and by regular, high-resolution photography (dermoscopy). Dermatoscopic and dielectric permittivity images were subsequently assessed at the study's time points for the grading of the occurrence and degree of petechia, "black dots" (thought to be oxidation products in puncture points), scaling (dry flaking skin), and change in the hydration of dielectric permittivity image brightness at the HD-MAP site [20, 26] . Skin physiological measurements were performed in the same order at 10 min and 30 min after application and at return visits at 48 h and 7 days. invasive nature of the assessment methods facilitated the study of individual phenotype, variability due to the presence or severity of disease as well as the possible influence of interactivity between therapeutic agents and vaccine effectiveness. The study was approved by the Regional Ethics Committee at Linköping University Hospital (Dnr. MB 2017-409-31, 18 October 2017). Participants received one HD-MAP applied to the volar forearm and one applied to the deltoid region on the non-dominant arm (Figure 1 ), as previously described by Muller et al. (2020) [20] . The application sites were free from hair, tattoos, scars, and visible veins. All areas were cleaned with an alcohol swab prior to application. Before application, the sites were compressed with a pre-calibrated skin conditioning ring (Vaxxas Pty Ltd., Brisbane, Australia) at a force of 30 N, to equalise the resistance of the skin between volunteers. A preloaded applicator device (Vaxxas Pty Ltd., Brisbane, Australia) was then docked into a skin conditioning ring. The applicator was triggered to apply the HD-MAP at a speed of 20 m/s. Once applied, HD-MAPs were left in place for 2 min. Immediately after removal, pain and local signs, erythema, oedema, and the presence of minimal breakthrough or minor "wet" bleeding were assessed clinically by trained staff, and skin physiological measurements were performed. Wet bleeding was defined as the presence, immediately after patch removal, of multiple pinpoint blood spotting on an applied tissue paper. Perceived pain on a scale from 0 (no pain) to 10 (worst imaginable pain) was reported by the volunteers at 10-and 30-min post-HD-MAP application. Measurements were recorded in the order of transepidermal water loss (evaporimetry), dielectric permittivity, erythema (polarisation spectroscopy), and by regular, high-resolution photography (dermoscopy). Dermatoscopic and dielectric permittivity images were subsequently assessed at the study's time points for the grading of the occurrence and degree of petechia, "black dots" (thought to be oxidation products in puncture points), scaling (dry flaking skin), and change in the hydration of dielectric permittivity image brightness at the HD-MAP site [20, 26] . Skin physiological measurements were performed in the same order at 10 min and 30 min after application and at return visits at 48 h and 7 days. The HD-MAP is a 1 × 1 cm patch containing 3136 conically shaped, from 120 µm at the bottom to 25 µm at the top, solid microprojections with a length of 250 µm. It is constituted by a non-dissolvable liquid crystal polymer and has a max penetration depth of 150 µm but an average penetration of around 100 µm. The HD-MAPs used were non-coated and applied as described in the procedure section. Transepidermal water loss was used to indicate penetration of the epidermal barrier. The measurement was performed by a Tewameter-300 probe (Courage + Khazaka electronic GmbH, Köln, Germany) at each application site. Evaporimetry measures transepidermal water loss as an indirect measure of barrier integrity. An increase in TEWL is expected when the barrier is compromised. At each measurement, the Tewameter probe was placed over the HD-MAP application site, and measurements were collected every second for 30 s and stabilised values selected for analysis. Collected data were analysed by CK-MPA-Multi-Probe-AdapterFB, v2.4.2.1/207-11-10 (Courage + Khazaka electronic GmbH, Köln, Germany). An Epsilon™ device (Epsilon Modell E100, Biox, London, UK) was used to estimate the presence of water in the superficial layer, stratum corneum, of the skin. The measurement principle has previously been well described [27] . Briefly, the system utilises the measurement of the dielectric constant of the upper 5 µm of the skin to create a two-dimensional hydration image of the skin surface. Due to 76,800 sensors within the sensing area of 12.8 × 15 mm, skin surface hydration can be mapped. We used the standardised "burst mode" option with a 5 s delay after the first skin contact, a frame interval of 1 s, and a total measurement period of 30 s. Tissue Viability Imaging (TiVi, WheelsBridge AB, Linköping, Sweden) was used to estimate erythema (rubor). The measurement principle has previously been described in detail [28] . Briefly, based on the wavelength-dependent absorption properties of red blood cells (RBCs) and imaging-processing algorithms, a TiVi-value (in arbitrary units, AU) that is linearly proportional to the local RBC concentration in the skin is calculated for each pixel. Image analysis and the calculation of TiVi values were made using WheelsBridge AB Software, v1.2.20, November 2018, Linköping, Sweden. The TiVi system was set to single photo mode and a medium resolution and positioned approximately 25 cm above the observed site. Photographic documentation of the application sites was performed using a dermatoscope (iC1, Heine Optotechnik GmbH & Co., Gilching, Germany) attached to a mobile telephone (iPhone 6, Apple, Cupertino, CA, USA). Dermoscopy refers to the examination of the skin using skin surface microscopy. Dermoscopy is mainly used to evaluate pigmented skin lesions but is also used for detailed analysis and documentation of inflammatory events. Data are presented in terms of mean ± SD. One way ANOVA with Šidák's multiple comparison test was used to compare changes between pre-application values and values at 10 min, 30 min, 48 h, and 7 days. Differences between the forearm and deltoid regions were tested with the Student's t-test. The alpha level for statistical significance was set at 0.05. All statistical analyses were made with the aid of GraphPad Prism v9.1.2 for Windows (GraphPad Software, San Diego, CA, USA, www.graphpad.com (accessed on 1 March 2022)). HD-MAP applications caused minimal discomfort. Perceived pain on a scale from 0 (no pain) to 10 (worst imaginable pain) was reported by the volunteers at 10-and 30-min post-HD-MAP application. At 10 min, no volunteer reported perceived pain values higher than 3. At 30 min, all volunteers, except one that reported 2, reported 0 perceived pain (Supplementary Materials, Table S1 ). Minor wet bleeding was seen in 12 of the 24 HD-MAP applications immediately upon removal of the HD-MAP (Table 1 ). Petechiae were seen in most images at 10 and 30 min to be virtually absent at later time points. Black dots at skin puncture points had developed in about half the cases by 48 h and 7 days. The dermatoscopic images showed uneven dry and flaking skin in the pre-application images in 9 of 24 cases and almost twice as many at later time points (Table 1) . A representative series of dermatoscopic images can be seen in Figure 2 . Data are presented in terms of mean ± SD. One way ANOVA with Šidák's multiple comparison test was used to compare changes between pre-application values and values at 10 min, 30 min, 48 h, and 7 days. Differences between the forearm and deltoid regions were tested with the Student's t-test. The alpha level for statistical significance was set at 0.05. All statistical analyses were made with the aid of GraphPad Prism v9.1.2 for Windows (GraphPad Software, San Diego, CA, USA, www.graphpad.com). HD-MAP applications caused minimal discomfort. Perceived pain on a scale from 0 (no pain) to 10 (worst imaginable pain) was reported by the volunteers at 10-and 30-min post-HD-MAP application. At 10 min, no volunteer reported perceived pain values higher than 3. At 30 min, all volunteers, except one that reported 2, reported 0 perceived pain (Supplementary Materials, Table S1 ). Minor wet bleeding was seen in 12 of the 24 HD-MAP applications immediately upon removal of the HD-MAP (Table 1 ). Petechiae were seen in most images at 10 and 30 min to be virtually absent at later time points. Black dots at skin puncture points had developed in about half the cases by 48 h and 7 days. The dermatoscopic images showed uneven dry and flaking skin in the pre-application images in 9 of 24 cases and almost twice as many at later time points (Table 1) . A representative series of dermatoscopic images can be seen in Figure 2 . Skin barrier integrity was indirectly assessed as TEWL (g/hm 2 ) as measured by evaporimetry. Mean pre-application TEWL values did not differ significantly between regions Skin barrier integrity was indirectly assessed as TEWL (g/hm 2 ) as measured by evaporimetry. Mean pre-application TEWL values did not differ significantly between regions (forearm; 6.65 ± 1.08 g/hm 2 , deltoid; 6.63 ± 2.19 g/hm 2 , p = 0.42). At the 10and 30-min observations, TEWL was significantly increased compared to respective preapplication values (p < 0.0001) at both the forearm and deltoid ( Figure 3A,B) . The highest mean TEWL values were observed at 10 min, 67.50 ± 13.49 g/hm 2 at the forearm and 66.01 ± 10.95 g/hm 2 at the deltoid. At 7 days there was no significant difference compared to pre-application values at either region (forearm; p = 0.99, deltoid; p > 0.99). (forearm; 6.65 ± 1.08 g/hm 2 , deltoid; 6.63 ± 2.19 g/hm 2 , p = 0.42). At the 10-and 30 observations, TEWL was significantly increased compared to respective pre-applic values (p < 0.0001) at both the forearm and deltoid ( Figure 3A-B) . The highest mean T values were observed at 10 min, 67.50 ± 13.49 g/hm 2 at the forearm and 66.01 ± 10.95 g at the deltoid. At 7 days there was no significant difference compared to pre-applic values at either region (forearm; p = 0.99, deltoid; p > 0.99). Superficial skin hydration, i.e., the presence of water in the stratum corneum, was assessed as dielectric permittivity. Mean pre-application values for dielectric permittivity did not differ significantly between the application sites (forearm; 5.86 ± 2.05 F/m, deltoid; 5.31 ± 2.42 F/m; p = 0.11). The dielectric permittivity increased significantly (increased stratum corneum hydration) during the first 30 min at both sites (24 of 24 HD-MAPs) compared to respective pre-application values (forearm; pre-application vs. 10 min, p < 0.0001, pre-application vs. 30 min, p < 0.0001; deltoid; pre-application vs. 10 min p < 0.0001, pre-application vs. 30 min p < 0.0001) ( Figure 3C,D) . The highest values for Vaccines 2022, 10, 583 7 of 11 dielectric permittivity were observed at 10 min, 34.00 ± 12.23 F/m at the forearm, and 39.08 ± 18.53 F/m at the deltoid. At 48 h there were no significant differences between the pre-application values (forearm; pre-application vs. 48 h p > 0.99; deltoid; pre-application vs. 48 h, p = 0.99). There was no significant difference compared to the respective preapplication values at 7 days for either site (forearm; p = 0.76, deltoid; p = 0.88). However, decreased signals (decreased hydration) compared to pre-application values at both sites were observed in slightly less than half of the images at both 48 h and at 7 days (forearm; pre-application mean dielectric permittivity value 5.87 ± 2.054 F/m, 48 h; 6.15 ± 0.56 F/m; 7 days, 2.16 ± 0.56 F/m; deltoid; pre-application 5.31 ± 2.42 F/m; 48 h, 5.70 ± 2.3 F/m, 7 days 1.86 ± 0.45 F/m) (Figure 4 ). did not differ significantly between the application sites (forearm; 5.86 ± 2.05 F/m, deltoid 5.31 ± 2.42 F/m; p = 0.11). The dielectric permittivity increased significantly (increased stra tum corneum hydration) during the first 30 min at both sites (24 of 24 HD-MAPs) compare to respective pre-application values (forearm; pre-application vs. 10 min, p < 0.0001, pre application vs. 30 min, p < 0.0001; deltoid; pre-application vs. 10 min p < 0.0001, pre-appl cation vs. 30 min p < 0.0001) (Figures 3C-D) . The highest values for dielectric permittivit were observed at 10 min, 34.00 ± 12.23 F/m at the forearm, and 39.08 ± 18.53 F/m at th deltoid. At 48 h there were no significant differences between the pre-application value (forearm; pre-application vs. 48 h p > 0.99; deltoid; pre-application vs. 48 h, p = 0.99). Ther was no significant difference compared to the respective pre-application values at 7 day for either site (forearm; p = 0.76, deltoid; p = 0.88). However, decreased signals (decrease hydration) compared to pre-application values at both sites were observed in slightly les than half of the images at both 48 h and at 7 days (forearm; pre-application mean dielectri permittivity value 5.87 ± 2.054 F/m, 48 h; 6.15 ± 0.56 F/m; 7 days, 2.16 ± 0.56 F/m; deltoid pre-application 5.31 ± 2.42 F/m; 48 h, 5.70 ± 2.3 F/m, 7 days 1.86 ± 0.45 F/m) (Figure 4 ). Not that an initial increased response is observed within 30 min after application, followed by a de creased response compared to pre-application after 48 h and 7 days. The kinetics of the erythema reaction to HD-MAP application was quantified by TiV The mean TiVi value was 128.75 ± 43.51 AU at the forearm site and 126.08 ± 35.32 AU a the deltoid site and did not vary significantly between (p = 0.84). TiVi values at 10 min an 48 h were significantly increased (p ≤ 0.0001)️ at the forearm and deltoid sites compared t the respective pre-application values (Figures 3E-F) . TiVi values returned to pre-applica tion levels for both test sites at 7 days (forearm; p = 0.45, deltoid p = 0.99). TiVi value peaked at 30 min (275.1 ± 66.45 AU at the forearm site and 240.2 ± 63.19 AU at the deltoid site) with slightly lower values at 10 min (forearm; 260.6 ± 57.96 AU, deltoid; 233.3 ± 56.5 AU). Here, we reported the innate reactivity and recovery of the skin in older adults fo lowing the application of HD-MAPs. The kinetics of the reaction were similar to thos previously reported in younger skin [20] . Using a range of complementary skin physio logical methods of investigation, we found that HD-MAP application was associated wit an instant loss of barrier integrity as measured by transepidermal water loss, which im plies capacity for vaccine delivery. HD-MAP application also triggered a rapid increas in erythema and surface hydration, which indicates the inflammatory provocation of th skin as a result of direct cell damage or the axon reflex mechanism [20, 29] . The applicatio of the HD-MAP resulted in erythema at 48 h, indicating tissue reactivity at a later tim point than the rapid, direct cellular damage or axon-reflex mediated reaction. Evaporim etry and surface hydration values began by 48 h to approach pre-application values wit near complete recovery within a week, except for surface hydration which fell to lowe levels than at the outset. Decreased hydration fits with the dermoscopy images showin Note that an initial increased response is observed within 30 min after application, followed by a decreased response compared to pre-application after 48 h and 7 days. The kinetics of the erythema reaction to HD-MAP application was quantified by TiVi. The mean TiVi value was 128.75 ± 43.51 AU at the forearm site and 126.08 ± 35.32 AU at the deltoid site and did not vary significantly between (p = 0.84). TiVi values at 10 min and 48 h were significantly increased (p ≤ 0.0001) at the forearm and deltoid sites compared to the respective pre-application values ( Figure 3E,F Here, we reported the innate reactivity and recovery of the skin in older adults following the application of HD-MAPs. The kinetics of the reaction were similar to those previously reported in younger skin [20] . Using a range of complementary skin physiological methods of investigation, we found that HD-MAP application was associated with an instant loss of barrier integrity as measured by transepidermal water loss, which implies capacity for vaccine delivery. HD-MAP application also triggered a rapid increase in erythema and surface hydration, which indicates the inflammatory provocation of the skin as a result of direct cell damage or the axon reflex mechanism [20, 29] . The application of the HD-MAP resulted in erythema at 48 h, indicating tissue reactivity at a later time point than the rapid, direct cellular damage or axon-reflex mediated reaction. Evaporimetry and surface hydration values began by 48 h to approach pre-application values with near complete recovery within a week, except for surface hydration which fell to lower levels than at the outset. Decreased hydration fits with the dermoscopy images showing flaking skin, and we interpreted this as post inflammatory desquamation, a phenomenon not observed in younger skin. With the exception of the flaking skin, the results were similar chronologically and in absolute values to previously observed data in the younger individuals. The present findings support the capacity for HD-MAP application to cause inflammatory reactivity in the skin, thought to be one of the primary drivers of the enhanced immune response observed following microneedle vaccination, even in elderly individuals. These findings suggest that the immune-enhancing phenomena associated with HD-MAP vaccine delivery will likely be seen also in the older age group. The elderly skin did show some differences to the results seen in younger skin. The minor and transitory pinpoint wet bleedings observed after HD-MAP application in 50% of the present study's volunteers was more than the 28% observed in a previously published younger population [20] after HD-MAP application. Likewise, petechiae were more common at early time points in this older group. The appearance at later time points of flaking, dry skin caused, we believe, some subtle changes in the development of the black dots, which indicate points of microneedle penetration. This is likely due to structural changes in older skin [30, 31] which may also result in mechanical factors of relevance to the HD-MAP application. The measurement of the inflammatory component erythema (rubor) by reflectance spectroscopy showed a similar reactivity to that previously observed in younger skin. In fact, the reactive erythema observed by reflectance spectroscopy in our aged group distributed itself above the mean values for the younger group previously studied. The fact that the occurrence of petechiae was more frequent in the older group may have contributed to the higher TiVi signal since the method quantitates total blood concentration in the tissue. In addition to the structural changes of elderly skin, a low-grade increase in the basal inflammatory state in aging skin, "inflammaging" is a hypothesised characteristic of elderly skin, which could be a functional etiological explanation [32] [33] [34] . Other groups have studied different aspects of the effects of microneedle application, often with devices of differing attributes to our HD-MAP. Points of focus have been micropuncture attributes and chronology of closure. Other delivery platforms such as microneedle patches [22] or dissolving microneedle patches [21] have likewise shown promising results regarding safety and tolerability. In our study, evaporimetry, dielectric permittivity, and polarisation spectroscopy were applied to observe different aspects of HD-MAP application. We chose evaporimetry because of its ability to observe epidermal penetration resulting in water loss. An alternative would have been OCT [35, 36] . In our future work, we plan to perform direct comparative studies of young and old skin to better understand this matter further. This study was limited by the fact that indirect physiological signs of inflammation were measured instead of direct molecular markers. Non-invasive methodologies facilitate, however, the performance of studies on both healthy individuals and individuals with disease or on specific medication, without the need for invasive methods for the direct measurement of biomarkers. The possibility exists for the combination of skin physiological techniques with minimally invasive techniques such as biopsy or other sampling methods with a view to establishing a robust relationship to more convenient and broadly applicable surrogate skin physiological methods for use in method development and studies of efficacy. Another limitation is that only 4 of 12 subjects returned on the 7-day revisit. The fact that our previous study had shown that tissue recovery was obvious by 7 days makes it less likely that the smaller size of the number of observations we had data on at 7 days would have hidden a lack of recovery. Since the reactivity to HD-MAP application in this group of elderly volunteers (mean age 75 years) was similar to that seen in a previously published younger group (mean age 27 years), we conclude that skin reactivity in the older age group was maintained at a level which is likely sufficient to induce the physical adjuvance hypothesised to be one of the positive attributes of HD-MAP vaccination. A further detailed study of how skin reactivity is influenced by device features and application method, as well as the variability associated with subject phenotype, disease, or medication, is warranted. After the pandemic: Perspectives on the future trajectory of COVID-19 Factors associated with COVID-19-related death using OpenSAFELY Vaccine delivery systems. Hum. Vaccines Immunother Shoulder Injury Related to Vaccine Administration (SIRVA): Petitioner claims to the National Vaccine Injury Compensation Program Microneedle patches for vaccination in developing countries Improving the reach of vaccines to low-resource regions, with a needle-free vaccine delivery device and long-term thermostabilization Usability, acceptability, and feasibility of a High-Density Microarray Patch (HD-MAP) applicator as a delivery method for vaccination in clinical settings Potent response of QS-21 as a vaccine adjuvant in the skin when delivered with the Nanopatch, resulted in adjuvant dose sparing Microprojection arrays applied to skin generate mechanical stress, induce an inflammatory transcriptome and cell death, and improve vaccine-induced immune responses A chimeric dengue virus vaccine candidate delivered by high density microarray patches protects against infection in mice Efficient Delivery of Dengue Virus Subunit Vaccines to the Skin by Microprojection Arrays Developing a Stabilizing Formulation of a Live Chimeric Dengue Virus Vaccine Dry Coated on a High-Density Microarray Patch Safety, tolerability, acceptability and immunogenicity of an influenza vaccine delivered to human skin by a novel high-density microprojection array patch (Nanopatch™). Vaccines Safety, tolerability, and immunogenicity of influenza vaccination with a high-density microarray patch: Results from a randomized, controlled phase I clinical trial Nanopatch-targeted skin vaccination against West Nile Virus and Chikungunya virus in mice Safety and dose-sparing effect of Japanese encephalitis vaccine administered by microneedle patch in uninfected, healthy adults (MNA-J): A randomised, partly blinded, active-controlled, phase 1 trial Inactivated poliovirus type 2 vaccine delivered to rat skin via high density microprojection array elicits potent neutralising antibody responses High-density microprojection array delivery to rat skin of low doses of trivalent inactivated poliovirus vaccine elicits potent neutralising antibody responses Complete protection by a single-dose skin patch-delivered SARS-CoV-2 spike vaccine Innate local response and tissue recovery following application of high density microarray patches to human skin Clinical study and stability assessment of a novel transcutaneous influenza vaccination using a dissolving microneedle patch The safety, immunogenicity, and acceptability of inactivated influenza vaccine delivered by microneedle patch (TIV-MNP 2015): A randomised, partly blinded, placebo-controlled, phase 1 trial Efficacy of high-dose versus standard-dose influenza vaccine in older adults Comparative effectiveness of high-dose versus standard-dose influenza vaccination on numbers of US nursing home residents admitted to hospital: A cluster-randomised trial Micropuncture closure following high density microarray patch application in healthy subjects Safety, acceptability and tolerability of uncoated and excipient-coated high density silicon micro-projection array patches in human subjects Anatomical site variation of water content in human skin measured by the Epsilon: A pilot study Sub-epidermal imaging using polarized light spectroscopy for assessment of skin microcirculation Cellular responses at the application site of a high-density microarray patch delivering an influenza vaccine in a randomized, controlled phase I clinical trial A review of protection against exposome factors impacting facial skin barrier function with 89% mineralizing thermal water The dynamic anatomy and patterning of skin Shelter from the cytokine storm: Pitfalls and prospects in the development of SARS-CoV-2 vaccines for an elderly population Inflamm-aging. An evolutionary perspective on immunosenescence Optical coherence tomography is a valuable tool in the study of the effects of microneedle geometry on skin penetration characteristics and in-skin dissolution In-vivo dynamic characterization of microneedle skin penetration using optical coherence tomography