key: cord-0954052-g9n83zqm authors: Zhou, Jin-an; Zeng, Hao-long; Deng, Ling-yan; Li, Hui-jun title: Clinical Performance of SARS-CoV-2 IgG and IgM Tests Using an Automated Chemiluminescent Assay date: 2021-04-20 journal: Curr Med Sci DOI: 10.1007/s11596-021-2349-7 sha: 91bfbf4e56c4b8654e629adb885a637d72d3631e doc_id: 954052 cord_uid: g9n83zqm Serology tests for viral antibodies provide an important tool to support nucleic acid testing for diagnosis of the novel coronavirus disease 2019 (COVID-19) and is useful for documenting previous exposures to SARS-CoV-2, the etiological agent of COVID-19. The sensitivities of the chemiluminescent SARS-CoV-2 IgG/IgM immunoassay were assessed by using serum samples collected from 728 patients testing positive for SARS-CoV-2 RNA. The specificity was evaluated on a panel of 60 serum samples from non-COVID-19 patients with high levels of rheumatoid factor, antinuclear antibody, or antibodies against Epstein-Barr virus (EBV), cytomegalovirus (CMV), mycoplasma pneumonia, human respiratory syncytial virus (RSV), adenovirus, influenza A or influenza B. The imprecision and interference were assessed by adopting the Clinical and Laboratory Standards Institute (CLSI) EP15-A2 and EP7-A2, respectively. Sensitivities between 1 and 65 days after onset of symptoms were 94.4% and 78.7%, for IgG and IgM test, respectively. The sensitivity increased with the time after symptom onset, and rose to the top on the 22nd to 28th days. The total imprecision (CVs) was less than 6.0% for IgG and less than 6.5% for IgM. Limited cross-reactions with antibodies against EBV, CMV, mycoplasma pneumonia, human RSV, adenovirus, influenza A or influenza B were found. These data suggested the chemiluminescent SARS-CoV-2 IgG and IgM, assay with reliable utility and sensitivity, could be used for rapid screening and retrospective surveillance of COVID-19. It has been reported that, of the total number of patients confirmed with coronavirus disease 2019 (COVID-19), more than 80% were mild cases, with older adults and people who have other conditions being most at risk [1] . This made the retrospective surveillance an important means to screen for severe acute respiratory syndrome (SARS)-associated coronavirus (CoV)-2 infection and interrupt undetected chains of disease transmission, as such surveillance may be key to tracking those mild cases [2] . The current laboratory method for identifying COVID-19 is mainly nucleic acid RT-PCR test, which is believed to be the most sensitive method for early identification of infection [3] . However, the RT-PCR test may not be an ideal tool for using in retrospective mass screening, as viral load rapidly declines 9 or 10 days after disease onset [4, 5] . More importantly, quite a number of false negatives occurred due to the variations during throat swab collection. The long turnaround times for RT-PCR of SARS-CoV-2 should also be considered [6] . Therefore, development and evaluation of new tests for specific antibodies of SARS-CoV-2 in patient blood are in urgent need. It is widely accepted that IgM provides the first line of defense during viral infections, prior to generation of adaptive, high affinity IgG responses that are important for long-term immunity and immunological memory [7, 8] . It was reported that after SARS-CoV infection, IgM antibody could be detected in patient blood after 3-6 days and IgG could be detected after 8 days [9] . While for Middle East respiratory syndrome coronavirus (MERS) infection, IgG and IgM tests showed optimal sensitivity after 21 days post onset of illness [10] . In the present study, we aimed to investigate the performance of SARS-CoV-2 IgG and IgM test using a new automated chemiluminescence assay on well-characterized samples from laboratory-confirmed COVID-19 patients and to challenge the test against a panel of samples known to cause possible crossreactions. A total of 728 serum samples from 728 COVID-19 patients were collected to estimate the diagnostic accuracy of the SARS-CoV-2 antibody test. These patients were all residents of Wuhan, China, and admitted to the Tongji Hospital of Tongji Medical College, Huazhong University of Science and Technology in Wuhan of China, from January 27 to March 2, 2020, and confirmed to have COVID-19 according to the Guidelines of the Diagnosis and Treatment of New Coronavirus Pneumonia published by the National Health Commission of China [11] . Of these patients, 320 (44.0%) were men and the median age was 63 years (range: . A second panel consisted of 60 serum samples were used to assess the degree of possible crossreactions. These samples were from patients with high levels of rheumatoid factor (RF) (range: >1500 IU/mL, n=2; ≤1500 IU/mL, n=5), antinuclear antibody (ANA) (1:320-1:3200, n=14), or IgM and IgG antibodies against Epstein-Barr virus (EBV) (n=10), cytomegalovirus (CMV) (n=3), mycoplasma pneumonia (n=5), human respiratory syncytial virus (RSV) (n=2), adenovirus (n=2), influenza A (n=11) or influenza B virus (n=6). The study was conducted in accordance with guidelines approved by the Ethics Committees of Tongji Hospital (China) (ID:TJ-IRB20200220). Oral informed consents were obtained from all the patients or patients' family members. SARS-CoV-2 antibodies (IgG, IgM) were determined by using a fully-automated, two-step chemilumi-nescent immunoassay kit (iFlash SARS-CoV-2 IgG/IgM, YHLO Biotech. Shenzhen, China), with the use of specific recombinant proteins (nucleocapsid protein and spike protein of the SARS-CoV-2). Paramagnetic carboxylated-microparticles were coated with the recombinant proteins through cross-linking by N-ethyl-N′-(3-dimethylaminopropyl) carbodiimide. Mouse monoclonal antihuman IgM or IgG was conjugated with NSP-DMAE-NHS. In the first step, sample, specimen treating agent (for IgM tests, sample dilution is needed), and recombinant SARS-Cov-2 antigen-coated paramagnetic microparticles were combined. The reaction mixture was washed and anti-human acridinium-labeled conjugate was added to form a reaction complex in the second step. Following another wash cycle, pre-trigger and trigger solutions were added to the reaction mixture. The final chemiluminescent reaction was expressed as relative light units (RLUs). The concentration of SARS-CoV-2 IgM or IgG in the sample was determined by comparing the RLU of a sample to the RLU determined from two calibrators. The manufacturer's suggested cutoff for SARS-CoV-2 IgG and IgM was >10 AU/mL. The repeatability and reproducibility was ≤10% and ≤15%, respectively. Imprecision was verified over five consecutive days using two levels of pooled human serum samples, tested in triplicate in accordance with Clinical and Laboratory Standards Institute (CLSI) document EP15-A2 [12] . Two fresh serum pools indicated two different concentration levels, were prepared, split to aliquots, stored at −70°C, and then thawed immediately in the 37°C water until measurement. During the imprecision verification, reagent of the same lot number was used by one operator. Interference testing of hemoglobin, bilirubin and lipaemia was performed based on CLSI document EP7-A2 [13] . Two base serum samples with two different concentrations of SARS-CoV-2 IgG/IgM were prepared. The Sysmex Interference Check.A Plus kit (Sysmex, Japan) was used to test the potential interferences of hemoglobin, bilirubin and lipaemia as previously reported [14] . The SPSS version 22.0 (SPSS, USA) was used as statistical package to estimate sensitivity and specificity with 95% confidence intervals. The crossreactivity study was treated as a descriptive analysis due to the insufficient sample size and therefore no statistical analyses were conducted. Of fig. 1 Interferences of hemoglobin (up to 5 g/L), unconjugated bilirubin (up to 20 mg/dL) and lipaemia (up to 1410 FTU) were estimated and could be accepted with changes less than or around 10%. As listed in table 3, we found at the low level (Level 1) of the SARS-CoV-2 antibodies, result of the IgG test was interfered by hemoglobin (up to 500 mg/dL) with the bias of 10.86%, while the IgM test was interfered by bilirubin (conjugated) (up to 20 mg/dL) with the bias of -12.18%. Determination of SARS-CoV-2 IgG and IgM antibodies was very precise, and within the range of repeatability and reproducibility reported by the manufacturer (table 4) . In the present study, we assessed the diagnostic performance for SARS-CoV-2 IgG and IgM test using a new automated chemiluminescent assay. Serum [6] , which was nearly equivalent to our estimation data. Seroconversion for IgG and IgM occurred nearly simultaneously in patients infected with SARS-CoV [15] , but the IgM against SARS-CoV-2 in this study showed a lower positive rate than IgG. This difference of sensitivities between IgG and IgM detection was also reported by a previous study of SARS-CoV [16] . We speculated that the difference could be due to the concentration of IgM was lower than that of IgG in blood after SARS-CoV-2 infection, also possibly due to the different capture sensitivity between the IgG and IgM regents in our assay. Moreover, previous studies of SARS-CoV indicated that the IgG and IgM increased to detectable levels at the second or third week after onset of symptoms [16, 17] . In our study, both the IgG and IgM against SARS-CoV-2 could be detected as early as on the 5th to 7th days post onset of illness, with the positive rate of 46.2% and 30.8%, respectively. The positive rate of the two antibodies rose in the following days and reached the peak levels on the 22nd to 28th days post onset of symptoms. These data suggested the detection sensitivity seemed to be correlated with the concentrations of the antibodies in blood. As we know, IgM appears mainly in the early stage of infection. The SARS-CoV IgM antibody was reported to be decreased on the 28th day and dropped to below a detectable level on about the 90th day post onset of illnes [8] . Our data showed the IgM positive rate of SARS-CoV-2 was also decreased 28 days after onset of symptoms, which seems to be consistent with the variations of antibody titers. Many serological tests have been developed for the diagnosis of virus infection, but false-positive reactions in negative serum may occur because of the immunological cross-reactivity with other anti-virus antibodies [2, 9] . We challenged the test with samples containing potential cross-reacting antibodies. No obvious cross-reactions were found with the other common pathogens including EBV, CMV, mycoplasma pneumonia, human RSV, adenovirus, and influenza A or influenza B virus, and with autoimmune diseases like ANA. However, cross-reaction was found in samples with high concentration of RF (>1500 IU/mL) for both SARS-CoV-2 IgG and IgM tests. Hence, interference of RF should be taken into account for patients with high concentration of RF. Additionally, since lipemia, hemolysis, and icterus commonly interfere with laboratory tests that use optical methods [18] , and considering the automated chemiluminescent assay used in the present study to determine the SARS-CoV-2 antibodies, we further estimated the interferences of these factors based on Clinical & Laboratory Standards Institute (CLSI) Interference Testing in Clinical Chemistry; Approved Guideline-Second Edition (EP7-A2), and found hemolysis and icterus may interfere with the test results when the antibody titer was low. The imprecision estimated according to that CLSI EP15-A2 is all less than 10% for withinrun and between run. All these results suggested the automated chemiluminescent assay in this study could be a reliable test for determination of SARS-CoV-2 IgG and IgM. This study has some limitations. The patient cohort used mainly consisted of COVID-19 patients living in Wuhan, and the number of patients infected with other pathogens or with autoimmune diseases was small. Further assessments of the test are needed for samples from COVID-19 patients of other endemic regions. In addition, we did not have the opportunity to include samples from patients of more than 65 days post symptom onset. Therefore, further studies are needed to estimate the sensitivity in samples beyond this date. In conclusion, The IgG and IgM test for determination of SARS-CoV-2 antibodies, using an automated chemiluminescent assay, showed high sensitivities even in early samples in combination with high precision, limited cross-reaction and interference. The data makes the IgG and IgM test a reliable test in surveillance for SARS-CoV-2 infection and may be key to interrupt undetected chains of disease transmission. The epidemiological characteristics of an outbreak of 2019 novel coronavirus diseases (COVID-19) in China SARS antibody test for serosurveillance A rapid advice guideline for the diagnosis and treatment of 2019 novel coronavirus (2019-nCoV) infected pneumonia (standard version) Detection of SARS coronavirus in plasma by real-time RT-PCR Clinical progression and viral load in a community outbreak of coronavirusassociated SARS pneumonia: a prospective study Clinical features of patients infected with 2019 novel coronavirus in Wuhan IgM in microbial infections: taken for granted? Immunol lett Profile of specific antibodies to the SARS-associated coronavirus Production of specific antibodies against SARS-coronavirus nucleocapsid protein without cross reactivity with human coronaviruses 229E and OC43 Suggested new breakpoints of anti-MERS-CoV antibody ELISA titers: performance analysis of serologic tests Diagnosis and Treatment Scheme of New Coronavirus Infected Pneumonia User Verification of Performance for Precision and Trueness; Approved Guideline, second edition Interference Testing in Clinical Chemistry Analytical validation and investigation on reference intervals of aldosterone and renin in Chinese Han population by using fully automated chemiluminescence immunoassays Chronological evolution of IgM, IgA, IgG and neutralisation antibodies after infection with SARS-associated coronavirus Differential sensitivities of severe acute respiratory syndrome (SARS) coronavirus spike polypeptide enzyme-linked immunosorbent assay (ELISA) and SARS coronavirus nucleocapsid protein ELISA for serodiagnosis of SARS coronavirus pneumonia Longitudinal profile of immunoglobulin G (IgG), IgM, and IgA antibodies against the severe acute respiratory syndrome (SARS) coronavirus nucleocapsid protein in patients with pneumonia due to the SARS coronavirus Evaluating interference caused by lipemia The authors declare that they have no conflict of interest.