key: cord-0949194-74l7esvf authors: nan title: Quantitative Analysis and Automated Lung Ultrasound Scoring for Evaluating COVID-19 Pneumonia With Neural Networks date: 2021-04-02 journal: IEEE Trans Ultrason Ferroelectr Freq Control DOI: 10.1109/tuffc.2021.3070696 sha: c62467fae21d59bc08bc29145812f52029dd965f doc_id: 949194 cord_uid: 74l7esvf As being radiation-free, portable, and capable of repetitive use, ultrasonography is playing an important role in diagnosing and evaluating the COVID-19 Pneumonia (PN) in this epidemic. By virtue of lung ultrasound scores (LUSS), lung ultrasound (LUS) was used to estimate the excessive lung fluid that is an important clinical manifestation of COVID-19 PN, with high sensitivity and specificity. However, as a qualitative method, LUSS suffered from large interobserver variations and requirement for experienced clinicians. Considering this limitation, we developed a quantitative and automatic lung ultrasound scoring system for evaluating the COVID-19 PN. A total of 1527 ultrasound images prospectively collected from 31 COVID-19 PN patients with different clinical conditions were evaluated and scored with LUSS by experienced clinicians. All images were processed via a series of computer-aided analysis, including curve-to-linear conversion, pleural line detection, region-of-interest (ROI) selection, and feature extraction. A collection of 28 features extracted from the ROI was specifically defined for mimicking the LUSS. Multilayer fully connected neural networks, support vector machines, and decision trees were developed for scoring LUS images using the fivefold cross validation. The model with [Formula: see text] two fully connected layers gave the best accuracy of 87%. It is concluded that the proposed method could assess the ultrasound images by assigning LUSS automatically with high accuracy, potentially applicable to the clinics. U LTRASONOGRAPHY is a radiation-free, easy to use, and portable imaging modality that competes the magnetic resonance imaging (MRI) and computed tomography (CT) in emergency and intensive care [1] . In this epidemic of COVID-19 Pneumonia (PN), ultrasonography was the only imaging modality with the access to the intensive care unit (ICU) and used as bedside supports in infected areas. Lung ultrasound (LUS) is appealing for PN diagnosis with the sensitivity and specificity superior to bedside chest X-ray, even close to CT [1] , [2] . LUS has been intensively explored in evaluation of lung properties and increasingly applied in diagnosis of various lung diseases, including COVID-19 PN [3] - [10] . There existed several protocols for LUS evaluation of the COVID-19 PN, the impacts of which were analyzed by Mento et al. [11] . Lung ultrasound score (LUSS) is utilized in clinics for semiquantitative assessment of pulmonary edema with good correlation with excessive lung fluid [12] , [13] , which is an important clinical manifestation of COVID-19 PN [14] . In clinical practice, there existed different methods to calculate the LUSS. As the most exhaustive method, the 28-sector approach accumulated all the numbers of B-lines in the 28 intercostal spaces to achieve a score [11] . Alternatively, a method calculated the LUSS by dividing the chest wall into eight zones, including two anterior and two lateral zones of each hemithorax, with the B-lines evaluated in one space of each zone [15] . In other practices, the clinicians assigned one of the four numbers (0, 1, 2, and 3) to an ultrasound image as the LUSS, depending on the observation of B-lines (appearance, number, and confluence) [16] , [17] . Other methods for calculating the LUSS could be referred to [18] and [19] . Although the methods for assigning LUSS provided semiobjective quantitative assessment for pulmonary edema, they suffered from subjectivity and large interobserver variations [16] , [20] , [21] . In addition, accurate assignment of the LUSS depended on the experience of clinicians, which limited the efficacy of LUSS in clinics, especially for caring patients with COVID-19 PN where the medical staffs are lacking [21] . Recently, some computer-aided methods have been proposed for quantitative analysis on LUS images for objective assessment of the lung conditions [9] , [20] - [23] . Brattain et al. [23] developed an image processing algorithm to detect the number of B-lines along angular slices to determine LUSS. Corradi et al. [21] proposed to quantitatively analyze the LUS images containing consolidation based on frequency distribution of gray scales. Brusasco et al. [20] developed an image segmentation method for automated detection of B-lines. Zong et al. [16] developed an animal model for quantitative evaluation of the pulmonary water content using LUSS. Recent endeavors were focused on the application of different classifiers for assigning the LUSS, including machine learning and deep learning [24] , [25] . Although existing techniques provided objective lung evaluations with success to different extents, there is a lack of methods for automated B-line scoring that may be most potentially applicable to the clinics. In view of this, we developed a quantitative and automatic LUS scoring system using multilayer fully connected neural networks for evaluating the LUS of COVID-19 PN. From February 23, 2020 to April 2, 2020, 31 patients (age: 55 ± 21 years old, male: 19, and female: 12) admitted to the ICU ward of Huoshenshan Hospital, a newly built hospital specialized for caring patients with COVID-19 PN in Wuhan, were included. All the patients had the symptoms of fever and dyspnea. Chest CT (United Imaging, uCT760) showed bilateral ground-glass opacities with peripheral, posterior, and basal predominance, which was in line with the international agreement [26] . The patients were confirmed as affected by the COVID-19 virus with a positive RT-PCR test after admission to the hospital. In addition, the patients were confirmed with COVID-19 PN of different conditions: critical (ten cases, 32.3%), severe (nine cases, 29.0%), common (seven cases, 22.6%), and mild (five cases, 16.1%) according to the diagnosis and treatment standard for COVID-19 PN in China [27] . The height and weight of the patients were recorded for the calculation of body mass index (BMI) (height: 168 ± 13 cm and weight: 70 ± 18 Kg). All patients were monitored in terms of oxygen index, positive end-expiratory pressure, static lung compliance, respiratory index, blood pressure, and body temperature. Conditions of all patients were evaluated daily by the clinicians. The study was approved by the Ethics Committee of Huoshenshan Hospital, Wuhan, China (Approval number: HSSLL030). All patients provided written informed consents by themselves or family members. All patients underwent LUS examinations for 12 standard fields on both hemithorax, including the upper and lower halves of the anterior, lateral, and posterior fields [28] . Repeated examinations were performed for part of the patients whose conditions changed during the treatment. A total of 45 patient times were finally performed with eight patients scanned twice and three patients scanned three times. The ultrasound equipment LOGIQ e (GE Healthcare, Wauwatosa, WI, USA) was utilized with a curved array low-frequency transducer (1) (2) (3) (4) (5) with acquisition details as image depth of 15 cm, focal depth of 7.5 cm, mechanical index (MI) of 1.2, thermal index (TI) of 0.7, and operation mode: penetration. Three ultrasound images that contained A-lines or B-lines were collected in each field and stored in the DICOM format. All scans were operated in the transverse plane to avoid the acoustic shadows of ribs. The time gain compensation was maintained the same for all patients in order to minimize variations in intensity among the collected images. A total of 1620 images were collected and scored blindly by two experienced clinicians (>6 years in using LUS) following the criteria as Score 0: normal defined as absence of B-lines and appearance of A-lines, Score 1: septal syndrome defined as B-lines at regular distances of about 7 mm, Score 2: interstitial-alveolar syndrome defined as the B-line distance less than 7 mm with some confluent, and Score 3: white lung designated for coalesced B-lines with the confluence more than 80% [17] . As an assistant, the distance between two adjacent B-lines whose distance was visually the maximum among others on an image was measured when determining different scores, in particular for Scores 1 and 2. The measurement was performed by placing two points on the middle of the B-linewidth along the middle line of the image height on the ultrasound machine. Images that were assigned with different scores by the two clinicians were discarded (see Fig. 1 ). Finally, there were 1527 images assigned with scores and included in the study with the composition of Score 0: 413 (27.0%), Score 1: 370 (24.2%), Score 2: 417 (27.3%), and Score 3: 327 (21.4%). The data will be provided on our lab website in the future (ht. tps://bio-hsi.ecnu.edu.cn/). Fig. 2 (a)-(d) representatively shows the ultrasound images with Scores 0, 1, 2, and 3, respectively. The collected LUS images were automatically processed via four steps: 1) curve-to-linear conversion; 2) pleural line detection; 3) region-of-interest (ROI) selection; and 4) feature extraction, to extract features for the subsequent classification task. 1) Curve-to-Linear Conversion: The ultrasound images (see Fig. 2 ) were converted to the linear format through the curve-to-linear conversion in an automated manner using a customized MATLAB code. Considering a matrix containing a curved ROI denoted as Cmatrix [see 3) To calculate the angle of the ROI by the slope of Lines A-C and B-D, denoted as AO'B. 4) To calculate the distance between pixels by where D p , L A−C , and N A−C is the distance between adjacent pixels, length of Line A-C, and number of pixels along Line A-C, respectively. gray values after the conversion involved bilinear interpolation [29] . Thereby, the image quality remained. Applied the curve-to-linear conversion, the images in the curve format in Fig. 2 were converted to their corresponding linear format, as shown in Fig. 4 . 2) Pleural Line Detection: Our previous work described an automated method for the pleural line detection in the linear images as simplified in the following steps [30] . value inside the rectangle was the maximum when shifting the rectangle from the left to right with the depth of the rectangle kept constant as determined in Step 1. The vertical position was chosen 5 pixels below the pleural line with two considerations: 1) it could avoid covering part of the pleural line which had certain thickness and 2) 5 pixels could be long enough to make the ROI away from the pleural line. 4) Ultrasound Image Feature Extraction: Two curves were created by averaging all the pixels' intensity of the ROI along each row and column in horizontal and vertical directions and normalized with respect to the maximum intensity of the ROI, denoted as C H and C V , respectively. The normalization was intended to avoid the influence of variations in intensity level among the images. Figs. 5 and 6 show the curves of C H and C V correspondingly obtained from the ROIs in Fig. 4 , respectively. The peaks of the curves were automatically detected and indicated by red circles in the figures. Recognizing that the bases for LUSS include the presence of A-lines (Score: 0) and B-lines (Scores: 1-3), as well as the confluent degree of B-lines are a kind of 1-D problem; 28 features from the C H , C V , and the ROI were creatively extracted in an automated manner and grouped into four categories: peak information in the horizontal direction, peak information in the vertical direction, area under the curve, and image feature, as listed in Table I . Noted that the peaks were detected via several steps: 1) filtering the projection curves using "Gaussian" filter, making the curve smooth; 2) removing the mean value or linear trend of the signal using MATLAB command: "detrend;" and 3) detecting the peaks by moving a window (three points in size) along the curve to detect the peaks. A peak was detected when the value of the middle point was higher than that of the left and right points with a predefined constraint, i.e., the peak value should be higher than 0.1 to avoid the influence of noises. The features were intentionally to mimic the clinician's interpretation when assigning the B-line scores, e.g., the number of peaks in C H represented the number of B-line; the average distance between peaks and average area between peaks in C H reflected the convergence of B-lines. With the features extracted from the ROI as input, we applied one-and two-layer fully connected neural networks to learn the data with a set of different number of hidden nodes, including one layer with 4, 25, 256, and 1024 hidden nodes, two layers with 4 × 8, 128 × 256, and 512 × 1024 hidden nodes. Adam optimizer was applied to fit the data with ReLU as the activation function, learning rate as 0.0001, batch size as 32, and epoch as 1000. In addition, machine learning models, including support vector machine (SVM) with linear and Gaussian cores and decision trees of 4 and 8 layers, were applied to the data. All the proposed models were evaluated via the fivefold cross validation, i.e., the data were randomly divided into five equal parts with four as the training data sets and the other one as the testing data set. Such a division between the training and testing data sets was repeated five times under each of the two configurations: 1) based on patient information, i.e., there were no data from the same patient falling in both the training and testing data sets (Config-1) and 2) based on all the data neglecting patient information (Config-2). The neural networks have been developed under both Config-1 and Config-2. In addition, machine learning models were trained under Config-1. The model was developed on the platform of Pytorch 1.4 with Python 3.7 and trained on a personal computer (CPU: amd4800u, RAM: 16 g 4266 MHz). As an example, the development of the 512 × 1024 neural network with 1000 epochs costs 198 s. As a summary, the proposed method included five steps: 1) curve-to-linear conversion; 2) pleural line detection; 3) ROI selection; 4) ultrasound image feature extraction; and 5) classification. Each step of the process was essential to perform the whole analysis. Step (1) transferred all the images from curve to linear, which made all B-lines parallel to each other, and A-lines straight. This would benefit for the subsequent feature extraction from the projection curves in the vertical and horizontal directions. 2) Step (2) performed the pleural line detection that supported for the succedent ROI selection. Step (3) selected the ROI from which the features were extracted for the classification task. 4) In Step (4), a total of 28 defined features that mimicked the human interpretation were extracted. Step (5), the deep learning model for automated lung ultrasound scoring was developed. Statistical analysis was performed using SPSS 22.0 for Windows system (SPSS Inc., Chicago, IL, USA). For a number of peaks, the Kruskal-Wallis test was used for the comparison among the four scoring subgroups, while one-way analysis of variance (ANOVA) was used for data comparison for all other quantitative parameters listed in Table I . The testing results of the fivefold cross validation of all the proposed models were compared using one-way repeated measures ANOVA. It was considered statistically significant if the P value was less than 0.05. The statistical analysis indicated that all features showed a statistically significant difference among the four scoring subgroups (P < 0.05). Representatively, Fig. 7(a)-(c) shows the average number of peaks, average distance between peaks, and average area between peaks in C H and C V of all the images in terms of LUSS:0 ∼ 3, respectively. The accuracies of different neural networks developed under Config-1 and Config-2 were plotted in terms of the computing epoch in Fig. 8(a) and (b) , respectively. It is noted that the reported performances of the proposed models were the averages of all the testing results of the cross validation. Statistical analysis showed that the testing results of the cross validation of different models presented a significant difference (p < 0.001). In addition, the accuracies of different neural networks trained under Config-1 and Config-2 were compared in Fig. 8(c) . For each neural network, the accuracy of the model trained under Config-2 was higher than that under Config-1. All the models trained under Config-1, including the SVM and decision tree, were compared in Fig. 9 , in which the neural network with 128 × 256 neurons gave the highest accuracy of 87%. In this study, we proposed several classification models for automated scoring of LUSS using the fivefold cross validation. The ultrasound images were collected from 31 COVID-19 PN patients with different conditions and labeled as Scores 0 ∼ 3 by experienced clinicians. A collection of 28 features were predefined and extracted from the ROI and ROI-generated curves for the classification task. Several classification models were trained and evaluated. The performance of the model (two layers with 128 × 256 hidden neurons) in the testing data demonstrated high accuracy for scoring LUS, indicating the promise for clinical applications. The statistical analysis indicated that each feature used in the study was sensitive to different subgroups and may be used to statistically discriminate images with Scores 0 ∼ 3. As representatives, three features (number of peaks, average distance between peaks, and average area between peaks) were plotted in Fig. 7 . The number of peaks in C H and C V in Fig. 7(a) potentially mirrored the number of B-lines and A-lines in the ROI, respectively. In Fig. 7(a) , significant distinction can be found between Score 0 and Scores 1-3 for the number of A-lines (i.e., peaks in C V ) and between Scores 1 and 2 for B-lines (i.e., peaks in C H ). A-lines demonstrates as parallel lines horizontally, resulting in multiple distinct peaks in C V , as shown in Figs. 4(a) and 6(a). While displayed as vertical lines/beams, B-lines rarely show peaks in C V , as shown in Figs. 4(b) and (c) and 6(b) and (c). The significant difference between A-lines and B-lines in the horizontal direction guaranteed the obvious distinction in peak number in C V for different LUSSs as learned from Fig. 7(a) . The same pattern also applied to the vertical direction for both A-lines and B-lines. The observations form Fig. 7 (a) may also be applied to Fig. 7(b) and (c) , which shows the average distances and areas between peaks in C H and C V , respectively. It is deserved to discuss that the average distances and areas between peaks in C H decreased as the LUSS increased from 1 to 3. Such a phenomenon was because B-lines were getting closer and more confluent when the LUSS increased (i.e., more edema), as shown in Fig. 5(b)-(d) . In addition, B-lines rarely had peaks in C V , resulting in almost zero value for average distance and area between peaks in C V , as shown in Fig. 7(b) and (c) . Note that the appearance of B-lines may be affected by the imaging frequency and bandwidth of the ultrasound probes/scanners [32] . Although this problem was mitigated in the study since the same probe was used, it is worth noting that counting the artifacts does not provide an absolute measure but only for a relative one. The change in B-line number may be more important than the number itself in the clinical practice. It is noticed from Fig. 8 (c) that the accuracy of the model trained under Config-2 was higher than that under Config-1 for each neural network. Such a phenomenon may be attributed to the fact that in Config-2, the training and testing data sets were divided without considering patient information, making those trained under Config-2 overfitting, since the data collected from the same patient may be of high similarity and grouped into the training and testing data sets. With such a consideration, the models trained under Config-1 were preferred. Among all the models trained under Config-1, the neural network with 128 × 256 neurons gave the highest prediction accuracy of 87% (see Fig. 9 ). One innovation of the study is the introduction of features extracted from 1-D curves obtained by projecting the pixels' intensities in the ROI in horizontal and vertical directions. Such an operation is based on the instinctive observation from the LUS images, e.g., B-lines are shown as parallel vertical lines/beams that can be imagined as a curve with several crests if looking from the bottom edge of the image toward the pleural line; the same applies to the A-line if looking from the left to the right side of the image, as shown in Figs. 4-6. In this study, 28 features that were well defined by mimicking the visual observations of the clinicians were extracted for training the classification models. In virtue of such an operation, the proposed model showed good performance based on the simple model structure and limited training data. It is noted that the curve-to-linear conversion is an important process to transfer all the curved images to linear, making the B-lines vertical and parallel to each other. The curve-to-linear conversion made the 2-D problem of the B-line evaluation in curved images to the 1-D problem of the projection curve analysis in linear images. Classification approaches have been applied to analyze ultrasound images and recently utilized to detect and evaluate the B-lines in an automated manner [20] , [23] , [33] - [36] . Brusasco et al. [20] applied the K-means classification to divide pixels in a filtered image into two subsets, i.e., B-lines and no B-lines. However, the method was validated on a small number of subjects, weakening its potential for real clinical applications. Brattain et al. [23] detected the B-lines via classification based on five features extracted from the angular slices. The limited number of features was not enough to depict the characteristics of the B-lines. In addition, the study did not mention any intelligent models for the classification. Soummer et al. [28] developed a sophisticated deep learning model for classification and localization of ultrasound COVID-19 markers by training on both ultrasound frames and videos. The model contained many hidden neurons which required data of a large number and powerful computing capability for training. In this study, we tried different classification models with artificial feature extraction as an anterior step to undertake the task of the classification with the advantages of: 1) no dependence on arbitrary and human-decided thresholds; 2) capability of getting robust classification models based on limited training data sets; and 3) avoiding overfitting by considering several parameters including the activation function, learning rate, batch size, and epoch. The accuracy of the neural network with 128 × 256 two fully connected layers trained under Config-1 was as high as 87% after the 700th epoch, indicating that the neural network could provide good performance for scoring LUS with high accuracy and low computation load. The model could then be potentially integrated into mobile ultrasound for clinical applications. There are several limitations of the study. First, the LUSS used in the study was only focused on the B-lines without considering the subpleural consolidations, which is not specially designed for the COVID-19 PN and may not fully reflect the status of the disease. Second, the study relied on visual determination of the B-score that is not a reliable gold standard. Third, the features to portray the LUS images were limited to 28 predefined parameters, which may be not enough for the current classification task although the proposed model showed good accuracy. In future studies, the scoring system developed specifically for COVID-19 PN will be applied as suggested by Perrone et al. [37] . In addition, the proposed technique needs to be improved with validation on reliable gold standard, e.g., using the postmortem gravimetry in animal studies or with the thermodilution in human studies as the reference in the future [20] , [38] . In this study, we proposed an automated scoring method to evaluate the ultrasound images of COVID-19 PN using different classification models. The results unveiled that the proposed method could assign LUSS for LUS images of COVID-19 PN with high accuracy, making it promise for automated LUS scoring. In addition, the proposed automated LUSS model has the potential to be integrated into portable and mobile ultrasound equipment for clinical use in hospitals of different levels as well as prehospital settings such as the ambulance. Thoracic ultrasonography: A narrative review Findings of lung ultrasonography of novel corona virus pneumonia during the 2019-2020 epidemic Proposal for international standardization of the use of lung ultrasound for patients with COVID-19: A simple, quantitative, reproducible method On the influence of imaging parameters on lung ultrasound B-line artifacts, in vitro study Is there a role for lung ultrasound during the COVID-19 pandemic? Lung ultrasound for COVID-19 patchy pneumonia: Extended or limited evaluations? Physical mechanisms providing clinical information from ultrasound lung images: Hypotheses and early confirmations Real-time multifrequency ultrasound imaging for quantitative lung ultrasound-first clinical results Assessment of interstitial lung disease using lung ultrasound surface wave elastography Characterization of the lung parenchyma using ultrasound multiple scattering On the impact of different lung ultrasound imaging protocols in the evaluation of patients affected by coronavirus disease 2019: How many acquisitions are needed? Ultrasound assessment for extravascular lung water in patients undergoing hemodialysis: Time course for resolution Assessment of extravascular lung water by quantitative ultrasound and CT in isolated bovine lung A rapid advice guideline for the diagnosis and treatment of 2019 novel coronavirus (2019-nCoV) infected pneumonia (standard version) Inter-rater reliability of quantifying pleural B-lines using multiple counting methods Using lung ultrasound to quantitatively evaluate pulmonary water content A simplified ultrasound comet tail grading scoring to assess pulmonary congestion in patients with heart failure Ultrasound lung comets in systemic sclerosis: A chest sonography hallmark of pulmonary interstitial fibrosis Relevance of lung ultrasound in the diagnosis of acute respiratory failure: The BLUE protocol Quantitative lung ultrasonography: A putative new algorithm for automatic detection and quantification of B-lines Computer-aided quantitative ultrasonography for detection of pulmonary edema in mechanically ventilated cardiac surgery patients Lung ultrasound and B-lines quantification inaccuracy: B sure to have the right solution Automated B-Line scoring on thoracic sonography Deep learning for classification and localization of COVID-19 markers in point-of-care lung ultrasound Automatic pleural line extraction and COVID-19 scoring from lung ultrasound data Chest CT in COVID-19 pneumonia: A review of current knowledge Released by National Health Commission & National Administration of Traditional Chinese Medicine on March 3 Ultrasound assessment of lung aeration loss during a successful weaning trial predicts postextubation distress Bilinear and bicubic interpolation methods for division of focal plane polarimeters Automated pleural line detection based on radon transform using ultrasound Ultrasound determination of chest wall thickness: Implications for needle thoracostomy Quantitative lung ultrasound spectroscopy applied to the diagnosis of pulmonary fibrosis: The first clinical study Deep learning in medical ultrasound analysis: A review Localizing B-Lines in lung ultrasonography by weakly supervised deep learning, in-vivo results Automated lung ultrasound B-Line assessment using a deep learning algorithm Automatic detection of B-lines in in vivo lung ultrasound A new lung ultrasound protocol able to predict worsening in patients affected by severe acute respiratory syndrome coronavirus 2 pneumonia Estimation of extravascular lung water using the transpulmonary ultrasound dilution (TPUD) method: A validation study in neonatal lambs The authors would like to thank the reviewers in advance for their comments and suggestions.