key: cord-0944639-l10sf3qf authors: Harskamp, R.; Bekker, L.; Himmelreich, J.; De Clercq, L.; Karregat, E.; Sleeswijk, M.; Lucassen, W. title: Performance of popular pulse oximeters compared with simultaneous arterial oxygen saturation or clinical-grade pulse oximetry: a cross-sectional validation study in intensive care patients date: 2021-03-12 journal: nan DOI: 10.1101/2021.03.09.21253181 sha: c4f5e12a0bb5d4e84707c07e74a104efd9f360ab doc_id: 944639 cord_uid: l10sf3qf Background: The demand for pulse oximeters is high in the current COVID-19 pandemic. Despite their popularity, clinical studies to evaluate the reliability in obtaining information on a patient's oxygenation status, are lacking. Aim: To evaluate the performance of pulse oximeters under clinical conditions, with arterial blood gas measurement (SaO2) as reference standard. Methods: We studied the accuracy of ten top-selling pulse oximeters in Europe and the USA in an intensive care population at the Flevoziekenhuis in Almere, the Netherlands. The studied pulse oximeters were: AFAC FS10D, AGPTEK FS10C, ANAPULSE ANP 100, Cocobear, Contec CMS50D1, HYLOGY MD-H37, Mommed YM101, PRCMISEMED F4PRO, PULOX PO-200, and Zacurate Pro Series 500DL. Adult patients in whom an SaO2 measurement was obtained as part of routine care were asked for inclusion. Directly after obtaining the SaO2 blood sample we obtained the pulse oximeter readings (SpO2) of the investigational devices, in random order. Outcomes were bias (SpO2-SaO2) mean, root mean square difference (ARMS), mean absolute error (MAE), and accuracy in identifying hypoxemia (SaO2<=90%). As a clinical index test, we included a hospital-grade SpO2-monitor. Results: In 35 consecutive patients we obtained 2,258 SpO2-readings and 234 SaO2-samples. Mean bias ranged from -0.6 to -4.8. None of the pulse oximeters had an ARMS<=3%, the requirement set by ISO-standards and required for FDA 501(k) clearance. The MAE ranged from 2.3 to 5.1, and 5 out of 10 pulse oximeters met the requirements of [≤]3%. For hypoxemia, negative predictive values were 98-99%. Positive predictive values ranged from 11-30%. Highest accuracy (95%CI) were found for Contec CMS50D1; 91% (86-94) and Zacurate Pro Series 500DL; 90% (85-94). The hospital-grade SpO2-monitor had an ARMS of 3.0 and MAE of 1.9, and an accuracy of 95% (91-97%). Conclusion: Direct-to-consumer pulse oximeters do not meet ISO-standards required for FDA-clearance, but can accurately rule out hypoxemia. Performance of popular pulse oximeters compared with simultaneous arterial oxygen saturation or clinical-grade pulse oximetry: a cross-sectional validation study in intensive care patients T h e d e m a n d f o r p u l s e o x i m e t e r s i s h i g h i n t h e c u r r e n t C O V I D -1 9 p a n d e m i c . D e s p i t e t h e i r p o p u l a r i t y , c l i n i c a l s t u d i e s t o e v a l u a t e t h e r e l i a b i l i t y i n o b t a i n i n g i n f o r m a t i o n o n a p a t i e n t ' s o x y g e n a t i o n s t a t D i r e c t -t o -c o n s u m e r p u l s e o x i m e t e r s d o n o t m e e t I S O -s t a n d a r d s r e q u i r e d f o r F D A -c l e a r a n c e , b u t c a n a c c u r a t e l y r u l e o u t h y p o x e m i a . B A C K G R O U N D P u l s e o x i m e t r y h a s b e c o m e a n i n d i s p e n s a b l e , l o w -c o s t , n o n -i n v a s i v e , d i a g n o s t i c t o o l t o a s s e s s a p a t i e n t s ' o x y g e n s a t u r a t i o n . T y p i c a l l y , t h e s e d i a g n o s t i c i n s t r u m e n t s h a v e a c l i p t h a t c a n b e p u t o n a p a t i e n t ' s f i n g e r t o o b t a i n i n f o r m a t i o n o n t h e p e r i p h e r a l a r t e r i a l o x y g e n s a t u r a t i o n ( S p O 2 ) , w h i c h s e r v e s a s a p r o x y f o r t i s s u e o x y g e n a t i o n . [ 1 ] O v e r t h e p a s t d e c a d e , p u l s e o x i m e t e r s h a v e e v o l v e d t o p l a y a p i v o t a l r o l e i n r o u t i n e m e d i c a l c a r e , a n d a r e a n e s s e n t i a l b e d s i d e t o o l i n m a k i n g t r e a t m e n t a n d / o r r e f e r r a l d e c i s i o n s i n c o m m u n i t y -b a s e d h e a l t h c a r e s e t t i n g s . I n t h e p a s t y e i o n c r i t e r i a w e r e p a t i e n t s w i t h o u t a c l i n i c a l i n d i c a t i o n f o r a r t e r i a l a c c e s s , t h o s e w i t h i n h e r i t e d f o r m s o f a b n o r m a l h e m o g l o b i n , a n d t h o s e w h o r a p i d l y d e t e r i o r a t e d d u e t o a c u t e h e m o d y n a m i c c o m p r o m i s e , i n w h i c h t h e m e a s u r e m e n t s r e q u i r e d f o r t h i s s t u d y c o u l d h i n d e r m e d i c a l i n t e r v e n t i o n s a n d t h e r e b y n e g a t i v e l y a f f e c t p a t i e n t s a f e t y . S t u d y p r o c e d u r e s I n t e n s i v e C a r e p e r s o n n e l n o t i f i e d t h e s i t e i n v e s t i g a t o r ( L . B . ) o f a p o t e n t i a l l y e l i g i b l e p a t i e n t f o r s t u d y e n r o l m e n t . T h e s i t e i n v e s t i g a t o r e n r o l l e d e a c h p a t i e n t a f t e r v e r b a l a n d w r i t t e n c o n T . T a b l e 4 d i s p l a y s f a c t o r s t h a t w e r e a s s o c i a t e d w i t h b i a s ( S p O 2 -S a O 2 ) o f e a c h p u l s e o x i m e t e r . O v e r a l l , d a r k e r s k i n c o m p l e x i o n a n d a n i n a c c u r a t e p u l s e r a t e m e a s u r e m e n t ( d i f f e r e n c e b e t w e e n p u l s e o x i m e t e r p u l s e r a t e a n d " t r u e " p u l s e r a t e a s c a p t u r e d b y I C U m o n i t o r ) , w e r e a s s o c i a t e d w i t h p o o r e r S p O 2 p e r f o r m a n c e i n f i v e o u t o f t e n p u l s e o x i m e t e r s . O t h e r f a c t o r s t h a t n e g a t i v e l y a f f e c t e d r e s u l t s w e r e i n c r e a s e d s y s t o l i c b l o o d p r e s s u r e a n d c o l d h a n d s t o t o u c h , i n f o u r a n d t h r e e p u l s e o x i m e t e o f ≤ 3 % , h e r e b y m e e t i n g t h e I S O c r i t e r i a f o r a c c u r a c y . O f t h e t e s t e d o x i m e t e r s , t h e C o n t e c C M S 5 0 D L m a y b e c o m p a r a b l e w i t h t h e C o n t e c C M S 5 0 D 1 m o d e l t h a t w e t e s t e d . I n o u r s t u d y , w h i c h w a s p e r f o r m e d u n d e r c l i n i c a l c o n d i t i o n s , t h e C o n t e c d e v i c e d i d n o t f a r e a s w e l l , a l t h o u g h i t w a s s t i l l o n e o f t h e b e t t e r p u l s e o x i m e t e r s t h a t w e t e s t e d . P r i o r s t u d i e s f o u n d t h a t o x i m e t e r s p e r f o r m e d w o r s e i n h y p o x i c c o n d i t i o n s , w i t h m e a n b i a s i n c r e a s i n g a t l o w e r o x y g e . CC-BY-NC-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted March 12, 2021. D a t a a v a i l a b i l i t y : A l l d a t a r e l e v a n t t o t h e s t u d y a r e i n c l u d e d i n t h e a r t i c l e o r u p l o a d e d a s a n o n l i n e s u p p l e m e n t . T h e d a t a s e t s u s e d d u r i n g t h e c u r r e n t s t u d y a r e a v a i l a b l e f r o m t h e c o r r e s p o n d i n g a u t h o r o n r e a s o n a b l e r e q u e s . CC-BY-NC-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted March 12, 2021. ; https://doi.org/10.1101 /2021 . CC-BY-NC-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. P e r f o r m a n c e o f p u l s e o x i m e t e r s c o m p a r e d w i t h a r t e r i a l b l o o d g a s ( u s i n g t h e m e t r i c s : m e a n b i a s , r o o t m e a n s q u a r e d i f f e r e n c e a n d m e a n a b s o l u t e e r r o r ) A l l m e a s u r e m e n t s i n c l u d i n g o u t l i e r s E x c l u d i n g o u t l i e r s * M e a n b i a s ( ± S D A c c u r a c y o f p u l s e o x i m e t e r s i n d e t e c t i n g h y p o x e m i a ( p r e v a l e n c e o f S a O 2 ≤ 9 0 % i s 5 . 1 % ) P u l s e o x i m e t e r T P / F N F P / T N S e n s i t i v i t y S p e c i f i c i t y P P V N P V A c c u r a c y A F A C F S 1 0 D 9 / 3 4 6 / 1 7 4 7 5 ( 4 3 -9 5 ) 7 9 ( 7 3 -8 4 ) 1 6 ( 1 1 -2 3 ) 9 8 ( 9 6 -9 9 ) 7 9 ( 7 3 -8 4 ) A G P T E K F S 1 0 C 1 0 / 2 3 9 / 1 8 2 8 3 ( 5 2 -9 8 ) . CC-BY-NC-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) F a c t o r s a s s o c i a t e d w i t h p o o r p e r f o r m a n c e o f e a c h p u l s e o x i m e t e r P u l s e o x i m e t e r A s s o c i a t e d f a c t o r ( s ) B e t a S t d E r r o r P -v a l u e A F A C F S 1 0 D H e a r t r a t e b i a s S y s t o l i c b l o o d p r e s s u r e 0 . 0 4 0 . . CC-BY-NC-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) . CC-BY-NC-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) We squared the difference (either positive, negative or 0) between a pulse oximeter measurement and the reference standard measurement. Then we calculated the mean of the sum of squares, and determined the root of that mean. As the difference between the pulse oximeter and the reference standard is squared, the A RMS is always positive. MAE: mean absolute error; n: number of samples; : Pulse oximeter measurement; : Reference standard measurement We calculated the absolute difference (either positive or 0) between a pulse oximeter measurement and the reference standard measurement. We then determined the mean of these absolute differences. Similarly to the A RMS , the MAE is always positive, as it concerns an absolute difference. . CC-BY-NC-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) The copyright holder for this preprint this version posted March 12, 2021. ; https://doi.org/10.1101/2021.03.09.21253181 doi: medRxiv preprint Table S1 . Accuracy of pulse oximeters in detecting abnormal oxygenation status, using either SaO 2 ≤92% or SaO 2 ≤94% as threshold. The values for PPV, NPV and accuracy are dependent on disease prevalence TP= true positive, FP= false positive, FN= false negative, TN= true negative, PPV= positive predictive value, NPV= negative predictive value. . CC-BY-NC-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) The copyright holder for this preprint this version posted March 12, 2021. ; https://doi.org/10.1101/2021.03.09.21253181 doi: medRxiv preprint Figure S1 . Bland-Altman plots of performance of pulse oximeters (bias compared with SaO 2 . CC-BY-NC-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) The copyright holder for this preprint this version posted March 12, 2021. ; https://doi.org/10.1101 https://doi.org/10. /2021 PRCMISEMED F4 PRO