key: cord-0940847-xy4lgrnv authors: Bae, Kyungsoo; Jeon, Kyung-Nyeo title: Diagnosis of Pulmonary Embolism in Unenhanced Dual Energy CT Using an Electron Density Image date: 2021-10-05 journal: Diagnostics (Basel) DOI: 10.3390/diagnostics11101841 sha: 8ca4530476ad06c238dd0155580222f963be8c63 doc_id: 940847 cord_uid: xy4lgrnv Dual-energy computed tomography (CT) is a promising tool, providing both anatomical information and material properties. Using spectral information such as iodine mapping and virtual monoenergetic reconstruction, dual-energy CT showed added value over pulmonary CT angiography in the diagnosis of pulmonary embolism. However, the role of non-contrast-enhanced dual energy CT in pulmonary embolism has never been reported. Here, we report a case of acute pulmonary embolism detected on an electron density image from an unenhanced dual-energy CT using a dual-layer detector system. Acute pulmonary embolism (PE) is a potentially fatal medical condition if its diagnosis is delayed. Computed tomography (CT) pulmonary angiography (CTPA) is the gold standard in the diagnosis of PE. Dual-energy CT (DECT) is a promising tool providing both anatomical information and material properties. An iodine map from DECT shows added value over CTPA in the diagnosis of PE [1] . In CT with suboptimal enhancement, monoenergetic image reconstruction can improve the diagnostic accuracy of PE by increasing vascular attenuation and the contrast-to-noise ratio [2] . However, in some patients, without clinical suspicion of PE, CT scans can be performed without intravenous contrast administration. The role of an unenhanced DECT in detecting PE has not been reported yet. Here, we report a case of acute PE detected on an electron density (ED) image from an unenhanced DECT obtained using a dual-layer detector system. This study was approved by the Institutional Review Board of our hospital. A 90-year-old woman was transferred to the emergency department with a cough, sputum, and dyspnea on exertion for a week. She was on medication for hypertension. Her white blood cell count was 12.33 × 10 3 µL and her erythrocyte sedimentation rate was 37 mm/h. Reverse transcription polymerase chain reaction (RT-PCR) test for SARS-CoV-2 was negative. Her serum creatinine level was 1.17 mg/dL. Her estimated glomerular filtration rate (eGFR) was 48 mL/min/1.73 m 2 . A chest radiograph showed subtle increased opacities in the right upper lobe and both lower lobes ( Figure 1 ). Since pneumonia was suspected, a chest CT scan was performed without contrast enhancement. CT was obtained using a dual-layer detector spectral CT scanner (IQon, Philips Healthcare, Best, The Netherlands). The scanning parameters were as follows: 120 kVp, 140-250 mA (reference mAs, 73), pitch of 0.609, rotation time of 0.4 s, 64 × 0.625 mm collimation, 1 mm slice thickness, 1 mm slice increment, and smooth filter (filter A). Chest CT images showed patchy areas of consolidation and ground-glass opacities in both lower lobes, suggestive of pneumonia ( Figure 2 ). There was a focal fibrotic scar in the right upper lobe. The patient was treated with antibiotics. Chest radiographs showed gradual improvement. However, the patient complained of persistence of dyspnea. Blood oxygen saturation level was 85% at room air and 99% after supplement of oxygen via nasal prongs at 3 L/min. A review of the unenhanced chest CT images revealed unremarkable findings except pneumonia. ED images were retrospectively reconstructed from a spectral based image of DECT. ED images showed multifocal areas of high ED in both interlobar, the left upper lobe, and both lower lobe pulmonary arteries ( Figure 3 ). The ED value measured at a high-density area was 108 %EDW, while the ED value of pulmonary artery without a high density was 104 %EDW (Figure 4 ). Serum D-dimer level was 10.46 µg/dL. With a suspicion of PE, CTPA was performed on the third day of admission. CTPA showed multiple thrombi in pulmonary arteries. The locations of thrombi corresponded to areas of high ED on ED images ( Figure 5) . Although there are several original research studies or case series on the detection of PE using signs of a hyperdense lumen in non-contrast-enhanced CT, density differences between intravascular blood and thrombus are usually not sufficiently big enough to be noticed without a contrast enhancement [3] [4] [5] [6] [7] . Protein content in red blood cells rather than fibrin in thrombus has been shown to be related to hyperdensity in CT [8] . Therefore, the visualization of PE on an unenhanced CT depends on the thrombus age and the patient's hematocrit level at the time of imaging [5] . According to a case control study by Sun et al. [9] , non-contrast CT showed a high negative predictive value (>90%) for central PE identification. However, it was neither sensitive nor specific enough to accurately detect central PE. Conventional CT images depict the amounts of total beam attenuation in each voxel of body tissue when an X-ray passes through the body. Total X-ray attenuation is a result of two physical interactions between X-ray photons and the body, including the photoelectric absorption and Compton scattering known to depend on X-ray energy, effective atomic number, and ED of the tissue. By measuring X-ray attenuation at two different energies, DECT allows the estimation of the effective atomic number and ED of each voxel. DECT can produce more accurate ED values than a conventional CT [10] [11] [12] . The ED value derived from DECT is expressed as a percentage relative to the ED of water (%EDW). Commercially available DECT systems are classified into two main categories: source-based (dual rotation, dual source, rapid kilovoltage switching) and detector-based (dual-layer) systems [13] . In a dual-layer detector CT system, as used in the preset study, spectral information is obtained from every scan performed using routine protocols. Therefore, tissue characterization could be retrospectively performed on demand by measuring iodine density, effective atomic number, and ED. ED generally corresponds to the physical density of the tissue. In clinical practice, ED has been used for optimizing the dose distribution in radiotherapy rather than for diagnostic purpose. Recently, Daoud et al. [14] have shown the potential benefit of ED in detecting lung lesions. ED images can improve the visualization of lung parenchymal lesions compared to conventional CT in patients with COVID-19 pneumonia [14] . In the present study, ED images derived from an unenhanced DECT clearly depicted PE, which could not be detected on conventional CT images. Due to the current pandemic situation, requests for non-contrast CT evaluations are increasing in patients with suspected COVID-19 [15] . In such patients, the pulmonary artery should also be carefully evaluated to detect potential complications, PE, particularly if the scan is performed using a dual-energy technique. In summary, we report a case of PE detected on an unenhanced DECT using ED images. This shows the potential of DECT for patients who cannot use an iodinated contrast agent in CT scans due to poor renal function, allergy to iodinated contrast, or pregnancy. Dual-Energy CT for Pulmonary Embolism: Current and Evolving Clinical Applications Improved Opacification of a Suboptimally Enhanced Pulmonary Artery in Chest CT: Experience Using a Dual-Layer Detector Spectral CT Hyperdense pulmonary artery sign-Detection of pulmonary embolism in patients with suspected COVID-19 using non-contrast chest CT. Radiol. Case Rep Never say never: Identification of acute pulmonary embolism on non-contrast computed tomography imaging Six Cases of Acute Central Pulmonary Embolism Revealed on Unenhanced Multidetector CT of the Chest The validity of hyperdense lumen sign in non-contrast chest CT scans in the detection of pulmonary thromboembolism Clinical Usefulness of Computed Tomography Study Without Contrast Injection in the Evaluation of Acute Pulmonary Embolism Differentiation of Clot Composition Using Conventional and Dual-Energy Computed Tomography Detection of central pulmonary embolism on non-contrast computed tomography: A case control study Measurement of electron density in dual-energy X-ray CT with monochromatic X rays and evaluation of its accuracy Dual-layer spectral computed tomography: Measuring relative electron density Accuracy of electron density, effective atomic number, and iodine concentration determination with a dual-layer dual-energy computed tomography system Spectral Computed Tomography: Fundamental Principles and Recent Developments Could Spectral CT Have a Potential Benefit in Coronavirus Disease (COVID-19)? Chest CT practice and protocols for COVID-19 from radiation dose management perspective