key: cord-0936823-jtk7magy authors: Szabó, Gábor Tamás; Üveges, Áron; Tar, Balázs; Ágoston, András; Dorj, Azzaya; Jenei, Csaba; Kolozsvári, Rudolf; Csippa, Benjamin; Czuriga, Dániel; Kőszegi, Zsolt title: The Holistic Coronary Physiology Display: Calculation of the Flow Separation Index in Vessel-Specific Individual Flow Range during Fractional Flow Reserve Measurement Using 3D Coronary Reconstruction date: 2021-04-28 journal: J Clin Med DOI: 10.3390/jcm10091910 sha: 7de860a891989e5fb8c914c65ef8cd4f4f605f17 doc_id: 936823 cord_uid: jtk7magy In order to make optimal decisions on the treatment of atherosclerotic coronary heart disease (CHD), appropriate evaluation is necessary, including both the anatomical and physiological assessment of the coronary arteries. According to current guidelines, a fractional flow reserve (FFR)–based clinical decision is recommended, but coronary flow reserve (CFR) measurements and microvascular evaluation should also be considered in special cases for a detailed exploration of the coronary disease state. We aimed to generate an extended physiological evaluation during routine FFR measurement and define a new pathological flow–related prognostic factor. Fluid dynamic equations were applied to calculate CFR on the basis of the three-dimensional (3D) reconstruction of the invasively acquired coronary angiogram and the measured intracoronary pressure data. A new, potentially robust prognostic parameter of a coronary lesion called the “flow separation index” (FS(i)), which is thought to detect the pathological flow amount through a stenosis was introduced in a vessel-specific flow range. Correlations between FS(i) and the clinically established physiological indices (CFR and FFR) were determined. The FS(i) was calculated in 19 vessels of 16 patients, including data from the pre- and post-stent revascularization treatment of 3 patients. There was no significant correlation between the FS(i) and the CFR (r = −0.23, p = 0.34); however, there was significant negative correlation between the FS(i) and the FFR (r = −0.66, p = 0.002). An even stronger correlation was found between the FS(i) and the ratio of the resting pressure ratio and the FFR (r = 0.92, p < 0.0001). The diagnostic power of the FS(i) for predicting the FFR value of <0.80, as a gold standard prognostic factor, was tested by receiver operating characteristic analysis. FS(i) > 0.022 proved to be the cutoff value of the prediction of a pathologically low FFR with a 0.856 area under the curve (95% confidence interval: 0.620 to 0.972). The present flow–pressure–velocity display provides a comprehensive summary of patient-specific pathophysiology in CHD. The consequences of epicardial stenoses can be evaluated together with their complex relations to microvascular conditions. Based on these values, clinical decision-making concerning both pharmacological therapy and percutaneous or surgical revascularization may be more precisely guided. For optimal decision-making in the treatment of atherosclerotic coronary heart disease (CHD) including pharmacotherapy, as well as percutaneous coronary interventions (PCI) or surgical treatment, the adequate evaluation of coronary angiographic results with quantitative anatomical and physiological assessments is recommended [1] . According to current guidelines, in the case of intermediate stenosis without objective evidence of ischemia, the measurement of the fractional flow reserve (FFR) is preferred during invasive coronary angiography [2, 3] . FFR expresses the ratio between the mean distal coronary artery pressure measured by a pressure wire and the mean aortic pressure detected at the tip of the guiding catheter during pharmacologically induced maximal vasodilation [4] . Based on current clinical guidelines, an FFR value equal to or less than 0.80 is an indication for PCI [2, 3] . However, in the Fractional Flow Reserve versus Angiography for Multivessel Evaluation 2 (FAME 2) study, patients with an FFR value above the cutoff ratio (0.80) also demonstrated a significant risk of adverse cardiovascular events, and 12% of these study participants required PCI. On the other hand, 52.6% of the patients randomized to medical therapy under the cutoff FFR value did not suffer an event during the 2-year follow-up period [5] . These data indicate the need for a more specific prognostic parameter than FFR. A coronary lesion can also be evaluated by coronary flow reserve (CFR), which is defined as the ratio of the coronary flow (velocity) during maximal vasodilation to that in a resting state [6] . Hemodynamically significant coronary lesions are known to reduce CFR to as low as <2. In the case of diffuse coronary artery atherosclerosis, CFR can also be decreased to a pathological level; furthermore, an impaired CFR may indicate microvascular dysfunction. In patients with suspected CHD, CFR proved to be an independent factor in terms of prediction of mortality [7] . The quantitative relation between the values of CFR and prognosis has also been established, providing meaningful incremental risk stratification [8] . In contrast with FFR, which represents the hemodynamic consequences of the epicardial resistance, CFR reflects the flow capacity of the entire coronary artery system, including the microcirculation [9] . Moreover, as FFR and CFR utilize entirely different approaches to describe coronary pathophysiology, pronounced discrepancies may occur between these two parameters [10] . This fact highlights the need for the incorporation of both pressure and flow parameters when aiming to describe lesion physiology [11] . Several attempts have been made to perform the simultaneous measurement of pressure-derived FFR and CFR using pressure measurements only [12] [13] [14] [15] [16] . However, given the systematic underestimation of CFR values [14] , it has been concluded that CFR cannot be measured merely by pressure alone, as friction losses are not negligible across a native coronary artery stenosis, and the ratio of the flow separation resistance and the resistance from the laminar flow cannot be predicted from pressure data [13, 14] . Recent publications proposed the "pressure-bounded" CFR (CFR pb ) assessment for delineating the possible range of the CFR according to the measured resting and hyperemic pressures [15, 16] . However, with the classification of the lesions into three distinct CFR pb groups in retrospective studies, the approach has failed to specify whether CFR is in a normal or abnormal range in approximately half of the lesions [15, 16] . A three-dimensional (3D) coronary tree reconstruction can be performed out of standard orthogonal two-dimensional (2D) images. Lesion characteristics show limitations with 2D quantitative coronary angiography [17] , while the comparison of 2D vs. 3D coronary evaluations has demonstrated the superiority of 3D analysis in the detection of the functional severity of a lesion [18] . Another advantage of 3D reconstruction is related to the possibility of optimal flow modeling in the coronary arteries of interest. Here, we aimed to introduce a hemodynamic calculation method by setting up a model, where 3D angiographic reconstruction was also incorporated. Previous evaluations have demonstrated the superiority of 3D vs. 2D analysis in the detection of the functional severity of a lesion. Moreover, 3D reconstruction has the advantage of optimal flow modeling in the coronary arteries of interest. We made efforts to augment the results of simultaneous intracoronary pressure measurement by exploiting the 3D analysis-related flow assessment during both vasodilation and the resting state, expressing the classic physiological parameter of CFR. We established the flow separation index (FS i ) as a new evaluation parameter that can also take into account the pathological turbulent flow in connection with low and oscillating shear stress, which play an important role in the progression of an atherosclerotic plaque [19] [20] [21] . Therefore, FS i may potentially become a prognostic factor for the progression of CHD. Our main goal here was to explore the correlations between FS i and the already established prognostic factors, FFR and CFR. In our model, one of the stipulations was to avoid the use of Doppler or thermodilution methods, which can be technically challenging [10, [22] [23] [24] . Therefore, our new calculation method uses data only from the routinely performed intracoronary pressure measurement and 3D parameters to assess pressure-flow relations [11] . Our model describes the flowpressure relation. The flow chart of generating the holistic coronary physiology display is depicted in Figure 1 . Diagnostic coronary angiography was performed and recorded with an AXIOM Artis-X-Ray device (Siemens, Munich, Germany), using standard fluoroscopic views. FFR measurement was performed with the RadiAnalyzer TM device (St. Jude Medical, Saint Paul, MN, USA) and a pressure sensor guidewire PressureWire TM Certus TM (St. Jude Medical). The wire was calibrated before use, and pressures were equalized at the tip of the guiding catheter. The resting pressure curve was recorded after intracoronary (ic.) administration of glyceryl trinitrate and a flush of saline. Maximal vasodilation was reached with the ic. administration of 200 µg adenosine. For a distal wire position, the pressure sensor was introduced at least 20 mm below the coronary stenosis. For the calculation of CFR p-3D , 3D angiographic reconstruction was performed following invasive coronary angiography with a dedicated, commercially available software (QAngio XA Research Edition 1.0, Medis Specials bv, Leiden). For the reconstruction, two angiographic recordings of sufficient visual quality are necessary, with at least a 25 • difference between the projections. We modeled the interrogated vessel segment in 3D, from the coronary orifice to the level of the pressure wire sensor. In our display model, 3D imaging for acquiring highly accurate anatomical data was required. CFR p-3D and FS i values were calculated from the corresponding data of the precise anatomical reconstruction and intracoronary pressure measurement. The calculation model of CFR p-3D is depicted in Figure 2 . Classic fluid dynamic equations were used for the calculation of the flow-pressure function, as published previously [25] . According to this model, the total pressure difference (∆p), the linear coefficient of the viscous friction (f), and the quadratic term representing separation losses (s) as well as the volumetric flow value were determined. By calculating the flow rate in hyperemic and resting states, the CFR p-3D value was computed as the ratio of the two flow rates. The calculated CFR p-3D as a classic index of functional coronary impairment was projected to display the pressure-flow relation of coronary circulation, providing information on the patient-specific pathophysiology. The scheme is a conceptually modified version of what Papafaklis et al. proposed as a virtual functional assessment display [26] . Their approach displayed the computed ratio of distal to aortic pressure (P d /P a ) − Q relationship in a function of a hypothesized fixed flow range (0-4 mL/s). As opposed to this, in our scheme, the X axis represents the actual flow ratio to the resting flow, while the Y axis shows the distal to proximal (aortic) pressure ratio. Thus, the flow ratio can be read on the X axis, while the resting pressure ratio is shown on the Y axis and, at maximal hyperemia, the flow ratio shows the CFR, and the pressure ratio represents the FFR ( Figure 3 ). The calculated CFR p-3D always defines the exact CFR value within the pressurebounded CFR interval, as depicted in Figure 4 [15, 16] . [25] . By using the measured pressure data, the equations were solved to calculate the flow. Pa: pressure in the proximal reference area; MLA: minimum lumen area; P d : pressure distal to the stenosis; A p : proximal reference area; A s : average stenotic area; A d : distal reference area; L p : length of the proximal reference area; L s : length of the stenotic area; L d : length of the distal reference area; ∆p: total pressure drop along the target vessel; Q: volumetric flow; f : linear coefficient in the viscous friction pressure loss; s: quadratic coefficient in the separation-related pressure loss term; η: blood viscosity; ρ: blood density. Using only intracoronary pressure data, the lower and higher bounds of the CFR Pb interval can be calculated by assuming the minimal CFR, as this would be the case with only a quadratic pressure drop, and the higher bound is defined as the maximal CFR, as this would be the case with only a linear pressure drop. The exact combination of the two types of flow resistances can be determined for the individual vessel by including the 3D geometry of the coronary artery in the CFR Pb determination. CFR Pb equations are valid in these forms only when the aortic pressures are the same during the resting and hyperemic states. This requirement is approximately met in the case of regular FFR measurement, where the aortic pressure usually does not change significantly after ic. adenosis provocation. The original concept defines the interval by pressure gradients [15, 16] : where ∆p is the total pressure drop along the target vessel and CFR is the coronary flow resistance. Flow separation is regarded as a disturbed flow, which is considered to be an important factor in the progression of coronary atherosclerotic plaques [27] [28] [29] . The hemodynamic calculation presented in Figure 4 describes the calculation and value of flow separation in the vessel-specific flow range. After calculating the CFR p-3D , the pressure-flow relation curve could be plotted in a diagram, which comprehensively demonstrated the flow separation pressure loss (FS i ) (Figures 3 and 4) . The area of the pressure loss due to flow separation, as presented in the diagram, can be readily calculated by integrating the pressure ratio values into the patient-specific flow interval (presented as a red area in Figures 3 and 4 and as a red line in Figure 5 ). Before stent implantation (A) the intracoronary physiological assessment revealed a significant decrease in FFR. After stent implantation (B), the FFR, the CFR, and the maximal tissue perfusion increased (from 0.79 to 0.94, from 2.09 to 2.67, and from 1.37 to 1.68 mL/min/g heart muscle, respectively). The FS i decreased from 0.062 to 0.009. The maximal tissue perfusion was determined from the calculated hyperemic flow divided by the supplied myocardial mass assessed on the basis of echocardiographic measurement [30] and the supplied left ventricular segments of the culprit vessel [1, 31] . FS i : flow separation index; Max: maximum; mL: milliliter; min: minute; g: gram; Pa: pressure in the proximal reference area; Pd: pressure distal to the stenosis. FS i is a dimensionless parameter of the examined artery, and its value is not dependent on the size of the vessel or the absolute value of the volumetric flow. The FS i value is defined proportionally to the quadratic pressure drop along a coronary lesion in an artery-specific flow range. FS i is calculated with the equation: where FS i is the flow separation index, CFR is the coronary flow reserve, s is the quadratic coefficient in the separation-related pressure loss term, Q is the volumetric flow; Pa is the aortic pressure, Qact is the actual volumetric flow, and Qrest is the volumetric flow registered under resting state. Sixteen consecutive patients were enrolled into the study between 1 December 2018 and 1 April 2019. All patients had an indication for invasive coronary investigation. The main inclusion criteria of patients into our current study were single coronary lesion of left anterior descending coronary artery (LAD), left circumflex coronary artery (LCx), or right coronary artery (RCA) with intermediate stenosis. All patients were excluded who had multivessel disease, more than one lesion on the interrogated coronary artery, left main coronary artery disease, ostial stenosis, CABG in the anamnesis, or an acute indication of coronary angiography. The study was conducted according to the guidelines of the Declaration of Helsinki and approved by the Hungarian National Institute of Pharmacy and Nutrition (project identification code: OGYEI/61148/2018). Informed consent was obtained from all subjects involved in the study. Statistical evaluations were performed by using the MedCalc Statistical Software, Version 14.8.1 (MedCalc Software bvba, Ostend, Belgium). Sensitivity (SN), specificity (SP), positive predictive value (PPV), negative predictive value (NPV), and accuracy (AC) with confidence intervals (CI 95%) were calculated in a standard way. The Shapiro-Wilk test was performed to identify normal distribution. Parameter changes after stent implantation were compared by paired t-test. Spearman's correlation analysis was carried out to examine the correlations of FS i and CFR p-3D , as well as FFR, and a significant relationship was found between two variables if the associated p value was less than 0.05. The area under the curve (AUC) calculated by receiver operating characteristic (ROC) analysis was applied to determine the diagnostic power of FS i to predict an FFR < 0.80. To evaluate the correlation of FS i to FFR and CFR, the value was calculated for 19 coronary arteries in 16 patients. Baseline characteristics of patients are presented in Table 1 . Detailed hemodynamic parameters are summarized in Table 2 . In the case of the three arteries, pre-and post-PCI calculations were also performed, their detailed hemodynamic parameters and changes are featured in Table 2 , and the results are indicated with A (pre-stent measurement) or B (post-stent measurement). Hemodynamic changes of patient 14 after stent implantation are depicted in Figure 5 . The changes in the average values of FFR, CFR p-3D , and FS i in patients 14, 15, and 16, where PCI with stent implantation was performed, are depicted in Figure 6 . The changes between pre-and post-stent average values of FFR and FS i differed significantly (0.71 ± 0.04 to 0.85 ± 0.04, p = 0.001 and 0.058 ± 0.015 to 0.011 ± 0.005, p = 0.05, respectively), while there was no significant difference in the CFR p-3D (1.79 ± 0.26 to 2.03 ± 0.33, p = 0.35) in this small-sized group. In our study, no significant correlation was found between FS i and CFR p-3D (r = −0.23, p = 0.34) ( Figure 7A) ; however, there was a significant negative correlation between FS i and FFR (r = −0.66, p = 0.002) ( Figure 7B ). An even stronger correlation was found between FS i and the ratio of the resting pressures quotient (P d /P a ) and FFR (r = 0.92, p < 0.0001) ( Figure 7C ). The diagnostic power of FS i for predicting an FFR value of <0.80, as a gold standard prognostic factor, was tested by ROC analysis in this small-sized pilot study. An FS i greater than 0.022 proved to be the cutoff value of the prediction of pathologically low FFR with a 0.856 AUC with a 95% confidence interval between 0.620 to 0.972 ( Figure 7D ). Positive predictive value, negative predictive value, accuracy, sensitivity, and specificity were 0.900, 0.889, 0.895, 90.00%, and 88.89%, respectively. In our novel interpretation, the pressure-flow relation derived from routine FFR measurement and 3D parameters provides a new index for distinction between resistance of laminar (or friction) flow and resistance of flow separation. According to the latter component, the flow separation index (FS i ) was generated in the vessel-specific flow range (Figure 3 ). It seems reasonable to differentiate between benign laminar, viscous and pathologically disturbed, turbulent flows. Recent experimental results also support the idea that flow disturbance plays a causal role (lowered and multidirectional shear stress) in the progression of atherosclerotic plaques [32] . Our method for the combined determination of FFR p-3D and FS i is applicable for every clinically indicated invasive FFR measurement during coronary angiography after the target vessel's 3D reconstruction. The 3D reconstruction software, which can be either built into the X-ray equipment or operated separately on a personal computer, can provide the necessary data for the 3D reconstruction for the calculation of the flow values by using intracoronary pressure measurement results. The flow calculation itself performed by this module does not require extra calculation time and provides an online, comprehensive flow-pressure relation display. Thereby, a more comprehensive dataset can be gathered compared to the FFR measurement alone: the consequences of epicardial stenosis can be evaluated together with the complex relations of the microvascular state, while the endothelial function can also be assessed based on the correlation of CFR and FFR, where CFR/FFR will be an indicator of microvascular function. According to these values, repeated examinations during the follow-up of the patients may clearly show the effects of the treatment both on the epicardial and microvascular levels. The microvascular level has been in the focus of clinical research lately, as accumulating data underline the importance of microvascular injury in the pathogenesis of coronavirus disease 2019 (COVID-19)-related cardiovascular complications [33] . Endothelial injury and dysfunction affecting mainly the microvascular circulation provoked by immunemediated injury seems to be an important determinant of these complications [34, 35] . In this respect, the precise examination and reification of the microvascular circulation is warranted for the assessment of sustained symptoms of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) infection in relation to possible heart involvement. Especially in the case of SARS-CoV-2 infection-associated cardiac injury, the identification of longterm consequences is necessary by a validated method for the measurement and follow-up of microvascular function. Only an easily available diagnostic tool can open a broad field of assessment of the effect of potential treatments at the microvascular level. In our pilot experiment, significant correlation was identified between FS i and FFR. This finding is in line with previous observations that the pressure gradient during the hyperemic state of the coronary artery is linked to the increase in pathological flow separation [36] . On the other hand, FFR reflects only the hyperemic pressure gradient, and therefore, it cannot characterize the resting conditions. However, the resting pathological flow also plays an important role in the progression of the epicardial plaque for obvious reasons, exposing the low and oscillating flow shear stress in the area of the flow separation, even for longer time periods than the hyperemic one [37] . Thus, by the inclusion of basal flow characterization, FS i may be a potentially better prognostic marker than only hyperemic flow-related FFR, for cardiovascular endpoints related to the progression of a coronary plaque. In this concept, the higher the value of FS i , the higher the risk of the progression of CHD. As stated previously, the incorporation of flow parameters into descriptive coronary physiological examination is an important guide in clinical decision-making [38, 39] . The present flow-pressure-velocity display provides a comprehensive summary of patientspecific pathology in CHD. The consequences of epicardial stenosis can be evaluated together with the complex relations of microvascular conditions without the requirement of Doppler wire or thermodilution procedures. As disturbed flow is considered to be an important factor in the progression of epicardial coronary atherosclerotic plaques [20] , we have differentiated benign laminar and pathologically turbulent flow resistances during routine FFR measurements. Using 3D coronary reconstruction data, the calculated pressure-flow relation provided a new, potentially prognostic information related to the FS i in a vessel-specific flow range. Based on these results, a large-scale clinical follow-up trial is warranted to test the prognostic value of FS i . Our method may serve as a theoretical basis for prospective clinical trials investigating the natural history and effects of different CHD treatments by event-driven analysis. FS i is a promising and comparable parameter to identify each coronary vessel with an atherosclerotic lesion prone to progress. The patent of the method detailed in this paper was issued by the Hungarian Intellectual Property Office (appl. no.: PCT/HU2019/050008, applicant: University of Debrecen). Z.K. is the inventor of this patent. No other conflict of interest is declared. Informed Consent Statement: Informed consent was obtained from all subjects involved in the study. The data presented in this study are available on request from the corresponding author. The data are not publicly available due to privacy. The authors declare no conflict of interest. Holistic polar map for integrated evaluation of cardiac imaging results ESC Guidelines for the diagnosis and management of chronic coronary syndromes ESC/EACTS Guidelines on myocardial revascularization Experimental basis of determining maximum coronary, myocardial, and collateral blood flow by pressure measurements for assessing functional stenosis severity before and after percutaneous transluminal coronary angioplasty Fractional flow reserve-guided PCI for stable coronary artery disease A comparison of two methods to measure coronary flow reserve in the setting of coronary angioplasty: Intracoronary blood flow velocity measurements with a Doppler catheter, and digital subtraction cineangiography Coronary Flow Reserve During Dipyridamole Stress Echocardiography Predicts Mortality Improved Cardiac Risk Assessment With Noninvasive Measures of Coronary Flow Reserve Coronary Flow Reserve from Mouse to Man-From Mechanistic Understanding to Future Interventions Is Discordance of Coronary Flow Reserve and Fractional Flow Reserve Due to Methodology or Clinically Relevant Coronary Pathophysiology? Anatomic versus physiologic assessment of coronary artery disease. Role of coronary flow reserve, fractional flow reserve, and positron emission tomography imaging in revascularization decision-making Pressure-based simultaneous CFR and FFR measurements: Understanding the physiology of a stenosed vessel Assessment of coronary flowreserve by coronary pressure measurement: Comparison with flow-or Velocity-Derived coronary flow reserve Pressure-derived measurement of coronary flow reserve Fractional flow reserve and pressure-bounded coronary flow reserve to predict outcomes in coronary artery disease What can intracoronary pressure measurements tell us about flow reserve? Pressure-Bounded coronary flow reserve and example application to the randomized DEFER trial Comparison of Assessment of Native Coronary Arteries by Standard Versus Three-Dimensional Coronary Angiography Three-dimensional and two-dimensional quantitative coronary angiography, and their prediction of reduced fractional flow reserve The role of blood flow in determining the sites of atherosclerotic plaques Role of endothelial shear stress in the natural history of coronary atherosclerosis and vascular remodeling: Molecular, cellular, and vascular behavior Blood Flow Modeling in Coronary Arteries: A Review Problems of Coronary Flow Reserve Van Royen, N. 4172Doppler flow velocity and thermodilution to assess coronary flow reserve: Head-to head comparison with [ 15 O]H 2 O PET imaging Doppler Versus Thermodilution-Derived Coronary Microvascular Resistance to Predict Coronary Microvascular Dysfunction in Patients With Acute Myocardial Infarction or Stable Angina Pectoris Less invasive fractional flow reserve measurement from 3-dimensional quantitative coronary angiography and classic fluid dynamic equations Fast virtual functional assessment of intermediate coronary lesions using routine angiographic data and blood flow simulation in humans: Comparison with pressure wire -fractional flow reserve Effects of Disturbed Flow on Vascular Endothelium: Pathophysiological Basis and Clinical Perspectives Prediction of Progression of Coronary Artery Disease and Clinical Outcomes Using Vascular Profiling of Endothelial Shear Stress and Arterial Plaque Characteristics: The PREDICTION Study Expert recommendations on the assessment of wall shear stress in human coronary arteries: Existing methodologies, technical considerations, and clinical applications Echocardiographic determination of left ventricular mass in man Segmental comparison between coronary angiography and positron emission tomography reveals low predictive value of epicardial flow for viability Inducing Persistent Flow Disturbances Accelerates Atherogenesis and Promotes Thin Cap Fibroatheroma Development in D374Y-PCSK9 Hypercholesterolemic Minipigs ST-Segment Elevation in Patients with Covid-19-A Case Series Endothelial dysfunction in COVID-19: Current findings and therapeutic implications The compelling arguments for the need of microvascular investigation in COVID-19 critical patients Quantification of disturbed coronary flow by disturbed vorticity index and relation with fractional flow reserve High Coronary Shear Stress in Patients With Coronary Artery Disease Predicts Myocardial Infarction Diagnostic and Prognostic Implications of Coronary Flow Capacity: A Comprehensive Cross-Modality Physiological Concept in Ischemic Heart Disease Fundamentals in clinical coronary physiology: Why coronary flow is more important than coronary pressure