key: cord-0935331-ug8gl4uh authors: Kim, Noori; Han, Kyungsup; Su, Pei-Chen; Kim, Insup; Yoon, Yong-Jin title: A rotationally focused flow (RFF) microfluidic biosensor by density difference for early-stage detectable diagnosis date: 2021-04-29 journal: Sci Rep DOI: 10.1038/s41598-021-88647-0 sha: 9fc5d7e08c08df78a4a4685321ed13ee464b5969 doc_id: 935331 cord_uid: ug8gl4uh Label-free optical biosensors have received tremendous attention in point-of-care testing, especially in the emerging pandemic, COVID-19, since they advance toward early-detection, rapid, real-time, ease-of-use, and low-cost paradigms. Protein biomarkers testings require less sample modification process compared to nucleic-acid biomarkers’. However, challenges always are in detecting low-concentration for early-stage diagnosis. Here we present a Rotationally Focused Flow (RFF) method to enhance sensitivity(wavelength shift) of label-free optical sensors by increasing the detection probability of protein-based molecules. The RFF is structured by adding a less-dense fluid to focus the target-fluid in a T-shaped microchannel. It is integrated with label-free silicon microring resonators interacting with biotin-streptavidin. The suggested mechanism has demonstrated 0.19 fM concentration detection along with a significant magnitudes sensitivity enhancement compared to single flow methods. Verified by both CFD simulations and fluorescent flow-experiments, this study provides a promising proof-of-concept platform for next-generation lab-on-a-chip bioanalytics such as ultrafast and early-detection of COVID-19. www.nature.com/scientificreports/ thickness [23] [24] [25] . Typically, convective transportation is dominant in a microfluidic device due to its high Péclet number (Pe) over 10 3 based on an extremely narrow width of the microchannel and low diffusivity of analyte solutions 26, 27 . In this case, binding kinetics can be effectively augmented by increasing initial reactive flux, which is directly enhanced by the reduction of boundary layer thickness [28] [29] [30] . Reducing the height, however, may induce an excessively high-pressure drop (i.e., Darcy-Weisbach equation for the rectangular-shaped channel, P ∝ h −2 ) 26,27 that could cause breakage of the bonding between microfluidic channel above a particular bonding strength 31 . Therefore, it is desirable to develop a straightforward microfluidic approach that can be ubiquitous with any sensor platforms and induce enhanced performances for target applications. One of the SARS-CoV2′s genetic and structural features is a protein on its surface, which would give a genetic clue of contagiousness to the human body 32 . Recent proteomics reported 27 potential biomarkers, including proteins, to indicate severity levels of COVID-19 symptom. Although polymerase chain reaction (PCR) based diagnostics have benefits in test scalability, they provide limited information on the severity of the disease, nor the course of illness 33 . Also, the viral RNA preparation time in real-time PCR (aka RT-PCR) process may affect the diagnostic accuracy as well as slow down the testing time. In a general laboratory setting, biomarker isolation and simple preparation process with protein-based biomarkers are more uncomplicated and comfortable compared to a nucleic acid (DNA/RNA) or cell-based biomarkers 34 . The protein-based biomarkers can help in early-stage diagnostics requiring minimum modification of samples, such as no amplification. To address the limitations with the RT-PCR method, Seo et al. developed a biosensor that detects the SARS-CoV-2 antigen protein using a field-effect 35 . However, their approach relies on the sensor system that they designed, a graphenebased biosensing device, and also requires longer incubation time (i.e., 4 h) 34 . Also, various methods integrated with a microfluidic channel for streptavidin-biotin complex detection have been investigated [36] [37] [38] . Chen, Ting-Yang, et al., have developed an electrical detection with thin film transistor-based biosensor integrated with a microfluidic channel by monitoring the current response of biotin and streptavidin. Yang, Haoyue, and Toshiya Sakata also have employed a microfluidic system able to monitoring the electrical signals induced by the change in pH resulting from the biotin-streptavidin interaction by utilising a molecular charge contact method. However, these previously reported systems have a low detection limit (0.4 mM of biotin and 16.7 μM of streptavidin and 2.3 µM of streptavidin, respectively) and require complicated fabrication. Although Castro, David, et al. have presented a two-phase microfluidic system which provides a user-friendly interface and simple fabrication solution, it has the relatively low detection limit, 100 ng/mL. In this work, we introduce a T-shaped microfluidic device with a rotationally focused flow (RFF) method, enhancing the detection probable limit of a label-free optical biosensor without additional instruments even in a low concentration environment. Two fluids with different density are used to induce reduced boundary layer thickness without increasing pressure drop. For demonstrating the RFF method, three validations have been performed; simulations using Computational Fluid Dynamics (CFD) ACE + , a flow experiment with fluorescent beads, and a quantitative measurement of the enhanced detection limit with label-free silicon microring resonator sensors. Working principle of rotaional focused flow (RFF) Figure 1 represents the schematic diagram of the RFF method. The fluidic channel is simple T-shaped, where two fluids with different density are introduced from two short ends and are flowed through the main channel to another end. When a target fluid is injected from one of the inlets while a less-dense fluid is pumped from the other inlet, the self-rotational flow is formed. It results in a reduction of the distance between the target and immobilised probe molecules on the sensor surface. Due to microfluidics characteristics, laminar flow with strong surface tension is expected in the rotating stream. It also enables target molecules to be focused vertically to the bottom with a low diffusion rate. Hydrodynamically, the target biomolecules are focused near at sensors' surface. It is to achieve a higher binding probability with immobilised probe molecules. Therefore, a faster and less-dense fluid (than the target www.nature.com/scientificreports/ fluids) is injected to increase the volumetric fraction ratio of the less-dense fluid. This phenomenon results in enhancement of the binding probability, as shown to the enlarged view of a sensing part in Fig. 1 . The height of target fluids in RFF can be obtained by modifying an equation from Nguyen et al. 39 where H R , Q T , μ T , Q L , μ L , and H C are the height of target fluids in RFF, the volumetric flow rate of target fluids, the dynamic viscosity of target fluids, the volumetric flow rate of less-dense fluids, the dynamic viscosity of less-dense fluids, and the height of a microchannel respectively. The fluids rotation is completed at the sensor position, and laminar flow exists. When the rotation is done, the heights can be deduced with fluids and injection properties based on Eq. (1). Also the entrance length (L e ) can be considered where in Laminar flow L e is 0.06*R e *D h . For our system Reynolds number (Re) much less than 1 and D h is 200 μm. The sensor is located around at 4.8 mm, and L e for our system is much smaller than the sensor position. From this fact, we can see that the rotation is completed away before reaching the sensor location. For our experiment and simulation setting, we have fixed the ratio 3:1 between Q L and Q T and 5:4 between μ L and μ T for ethanol and target fluids diluted by water, respectively. Therefore, H R is reduced to 0.3 of H c by adding a less-dense liquid with three times faster flow rate than target fluids based on Eq. (1). Computational fluid dynamics (CFD) simulation. Pattern forecast of biomolecules is essential as controlling biomolecules affects the binding probability directly. Notably, the biomolecules in microfluidic devices are more easily controlled with the laminar flow due to the high surface tension 27, 40 . Fluidic simulations have been performed using CFD ACE + software (a commercial computational fluid dynamics solver, CFD Research Corporation, AL, USA). The software solution provides geometry, grid generation, solving, data visualisation. In our simulation, we utilised two representative modules in the software; "Two-fluid" and "Spray. " The "Twofluid" was chosen to apply two independent fluids with different densities, and the "Spray" was used to assign particles as target biomolecules. For the CFD simulation, a microchannel design with 300 µm of width, 200 µm of height, and 4800 µm of length was chosen. The 200 µm height was fixed as the maximum height of microchannel fabricated by UV-lithography 41 . The 300 µm width and 4800 µm length were determined considering suitable aspect ratio (0.5-1) for the stable generation of hydrodynamic focusing 42 . Immunopure streptavidin and EZ-Link NHS-PEG 4 -biotin were purchased from Thermo Scientific Pierce (Singapore). Biotin anti-human TNF-α was purchased from BioLegend (San Diego, CA, USA). Other chemicals were analytical reagent grade and were used as received. All samples and buffers were prepared using DI water and PBS. A simple T-shaped microchannel which has the same dimension with the CFD simulation was printed onto a film photomask. PDMS microfluidic channels were built by PDMS replica molding from SU-8/silicon masters that were fabricated by the conventional UV-photolithographic process using the film mask. For 200 µm of target thickness, SU-8 2100 of a photoresist was poured onto an 8-inch silicon substrate and spin-coated with the spin speed of 1400 rpm for 60 s. The spin-coated photoresist was then baked at 65 °C for 7 min and 95 °C for 40 min. The baked layer covered with the designed photomask was then exposed to a 315 mJ/cm 2 dose of the UV light. In a post-exposure baking process, the transferred photoresist was then baked on a hotplate at 65 °C for 5 min and 95 °C for 14 min. Finally, through a development process for 17 min using a SU-8 developer and rinse process by IPA and DI water, a SU-8 master for the replication of microfluidic devices was realised. PDMS is the most common polymer for the fabrication of microfluidic devices because of those advantages such as low-cost and straightforward replication process 17 . The fabricated SU-8/silicon master was coated for the silanisation using 97% trichlorosilane for 30 min. This process is to prevent the adhesion between the master and PDMS replicas 43 . The mixture of the PDMS prepolymer (sylgard 184, part A) and the curing agent (sylgard 184, part B) with a 10:1 weight ratio was evenly stirred for 10 min using a glass rod. The PDMS mixture was then poured onto a SU-8/silicon master. The PDMS mixture on the master was then degassed inside a vacuum desiccator and cured for one hour at 80 °C in a convection oven. There are two experiments; dummy and real tests. Hence bonding processes of molded PMDS replicas have been performed twice, one with sliding glass and the other with a silicon-based microring resonator sensor chip (Fig. 2) . The silicon-based microring resonator has been fabricated with 220 nm thickness of SOI wafers with 2 μm of buried oxide (BOX) layer in our previous work 20, 43, 44 . In brief, one linear input waveguide is connected www.nature.com/scientificreports/ with four microrings, and each microring has a corresponding output waveguide. One of the microrings's output is used as a reference while the 3 remaining microrings' are collected through a vertical grating coupler (Fig. 2a) . The dimension of racetrack-shaped microrings is 5 μm of the radius with 2.042 of coupling length. Besides, the width of waveguides is 500 nm, and the gap between a linear waveguide and a microring is 220 nm. A PDMS replica was pierced using a steel punch to connect tubes at inlet and outlet positions before a bonding process. Prepared PDMS replica and slide glass (or silicon-based microring resonator) surfaces were washed with running IPA and DI water, followed by drying under a nitrogen stream and by heating in a convection oven. They were then treated with the oxygen plasma to activate the hydroxyl group for the adhesion using the NT-2 plasma cleaning system (Anatech Ltd., CA) under 70 W of a treating power in 60 sccm of O 2 flow rate for 50 s. The treated PDMS replica was bonded with the sliding glass for the verification of the RFF. Also, the microchannel in the treated PDMS replica was aligned and bonded to the microring sensor chip guided by an inverted microscope with CCD camera (Fig. 2a) . Finally, two types of microfluidic devices (Fig. 2b,c) have built after a hardening process in a convection oven at 150 °C for 30 min. Results and discussion CFD simulation. Figure (1) 42 . For an ideal choice of less-dense fluid, the density difference between two fluids must be large enough to ensure rapid fluid rotation to minimise diffusion between them. In this project, therefore, ethanol was selected (Δρ = 251 kg/m 3 , an ethanol density of 789 kg/m 3 ) and the viscosity of the target fluid (PSB + Polystyrene bead) was 1040 kg/m 3 where the density of Polystyrene bead is 1050 kg/m3. One of the major concerns for using ethanol could be the probability of biomolecules denaturation 48 when mixing two fluids. However, in a typical microfluidic configuration, mixing is unlikely to happen due to negligible Reynolds number except for a little diffusion 27, 49 . The Reynolds number (Re) can be considered as Re = D * V * ρ µ , where D, V, ρ, and μ are dimension, velocity, density, and viscosity, respectively. In specific to our system, the ρ and μ values are the same as our simulation values given for the simulation, and micro (10 -6 ) can represent our experimental dimension. It leads us to have Re = 10 −6 * V * 1040 0.000855 ≈ 1.77 * V . Based on our experiment and simulation velocity of fluids (3 μL/min, 9 μL/min), we can confirm that our Reynolds number is much less than 1, satisfying the small Reynolds condition microfluidic environment. Our system method employed both Laminar flow and self-rotating. Small Reynolds number ensures Laminar flow domination which does not to cause diffusion. Fluids with different densities enable self-rotating in microchannel 42 , and it helps to focus the particles on the sensor surface (i.e., Fig. 3(b) safe region). Therefore, the denaturation of input biomolecules can be avoided by controlling the diffusion layer thickness consistency. The CFD simulation was carried out to confirm this phenomenon. Figure 3b represents the backside view (position 7) of a microchannel in a CFD simulation to investigate the diffusion layer thickness and a safe region from the denaturation. In this figure, red block, blue block, and ball typed dots indicate ethanol, PBS, and polystyrene beads, respectively. The simulation results show two distinct regions within the focused region of RFF; a diffusion region and a safety region where biomolecules do not mix with ethanol. The diffusion region forms beyond 42 μm above the bottom surface, which provides enough distance to prevent the diffusion of denatured molecules to the sensor surface. Therefore, the RFF method using ethanol is suitable to be used with our microfluidic device, and the method can enhance their binding probability by reducing a distance between target molecules and immobilised probe molecules. (Fig. 4a) was overlaid with the CFD simulation result (Fig. 4b) . Remarkably, both images are perfectly aligned, which confirms the generation of the RFF, as shown in Fig. 4c . Binding performance. The binding performance was investigated using a silicon-based microring sensor. The microring sensor is a label-free refractive index sensor by taking advantage of the evanescent filed present near the ring resonator surface 50 . Its resonant wavelength, λ, is directly related to the effective refractive index of the surrounding area and the antibody-antigen interaction can be monitored by measuring the shift in the resonant wavelength, Δλ 44,46,51 . The binding performance of the RFF method was compared with that of the single flow (SF) method using Streptavidin-biotin binding as an antigen-antibody model case. The SF is a typical flow method that target fluids are injected solo into a microchannel. The streptavidin solution was injected into a microchannel without using a less-dense fluid. The binding curve of 190 nM of Streptavidin solution injected by the SF method is shown in Fig. 5a . The microchannel was first filled with PBS to obtain the baseline then the Streptavidin solution was injected at 3 μL/min of a volumetric flow rate for 5 min followed by PBS washing for 10 min to remove any non-specific binding. The binding curves of various Streptavidin solutions ranging from 9.5 nM to 1900 nM concentrations were obtained. And the resonance shift values at 15 min (10 min after PBS washing) are plotted against the concentration of Streptavidin solution in Fig. 5b . The resonant wavelength shifts at 15 min were 25, 178, 385, and 590 pm corresponding to 9.5, 57, 190 , and 1900 nM of Streptavidin solutions, respectively. The graph shows a logarithmic growth, with a correlation coefficient of the least square (R 2 = 0.9838). In the RFF method experiment, ethanol was added as a less-dense fluid to a microchannel. Figure 5c represents a binding curve of 1.9 fM of Streptavidin solution injected by the RFF method. The microchannel was first filled with PBS to obtain the baseline; Streptavidin analytes were then injected with 3 μL/min of a volumetric flow rate, and ethanol was injected with 9 μL/min of a volumetric flow rate. After 5 min, the entire microchannel was washed with PBS for 10 min for removing non-specific binding. More Streptavidin molecules near the sensor surface were observed from the RFF method than the SF method. The molecules could accumulate near the sensor surface, and this could generate a large amount of nonspecific bindings and result in a significant drop-off in the wavelength shifts during PBS washing. The measured wavelength shifts at 15 min (10 min after PBS washing) by the biotin-Streptavidin interaction using the RFF method in various concentrations of Streptavidin solutions ranging from 0.19 fM to 190 fM were plotted as www.nature.com/scientificreports/ shown in Fig. 5d . The resonant wavelength shifts were 260, 460, 646, and 862 pm corresponding to 0.19, 1.9, 19, and 190 fM of Streptavidin solutions, respectively. Similar to the SF method, the conjugations tendency grows logarithmically with a correlation coefficient in the least square method (R 2 = 0.9992). One may read from the stable wavelength shift values from (a) and (c), they are similar (around 400 pm). Notice that results were measured from the SF method with 1.9*108fM while RFF with 1.9fM concentration. Remarkably, the sensitivity of the microring sensor with the RFF method was improved by 8-order of magnitudes compared to the SF method. Note that the purpose of Fig. 5 is not comparing each sensor's performance to the saturation value. It is rather to examine the sensing possible limit varying concentration. The essential focus was on "detection possibility, " as long as the binding quantity is reasonable to cause wavelength shift. The results can benefit early diagnosis where samples' concentration is relatively low. When concentration is very low, probability to molecule binding is low compared to the single flow. In usual case, with low concentration, the molecule does not reach to the sensor surface. But with the RFF system, even with a diluted sample (low concentration), the rotational force helped to position the molecule near to the sensor surface, hence increase binding probability. Binding curves for both methods were well fitted with the Langmuir model with correlation coefficients from plotted graphs. For the Langmuir model, the apparent dissociation constant (K d ) is defined as 44 where Δλ is the measured resonant wavelength shift, λ max is the maximum wavelength shift for bound streptavidin to immobilised biotin onto a microring resonator. And C s and K d are the concentration of Streptavidin solution and the apparent dissociation constant, respectively. In these experiments, over 190 fM of Streptavidin concentrations, wavelength shifts were varied from 800 to 920 pm. According to the data, we can estimate λ max as 860 pm. K d between Streptavidin molecules and Biological activity test. In order to confirm that enhanced apparent sensitivity of RFF method 44 is based on real increase in the sensitivity not due to other factors including non-specific binding of denatured protein clusters, we investigated the biological activity of bound Streptavidin molecules using biotinylated anti-human TNF-α (biotin-TNFα) as a secondary probe. Figure 6 shows binding procedures and the binding curves of the biotin-Streptavidin interaction using various concentrations of Streptavidin solutions (red line) with different flow methods followed by PBS washing (blue line) and subsequent interaction between bound Streptavidin and biotin-TNFα (green line). The schematic of the detailed binding procedures is shown in Fig. 6a . The binding curves for 190 nM of Streptavidin solution using the SF method, 0.19 fM and 1.9 fM of Streptavidin solution using RFF method are shown in Fig. 6b . Numerical values of wavelength shifts at 5 min (after Streptavidin binding), 15 min (after washing) and 20 min (after biotin-TNFα binding) are listed in Table 1 Potential migration of biomolecules by Dean vortex. As we discussed earlier, drastically enhanced sensitivity by the RFF method is beyond expectation. Besides a simple channel height reduction effect, the secondary flow effect has been considered to explain the target analyte flux increment further. Among various secondary flows, the transverse motion of particles by the Dean vortex is well-known to have a significant effect on the enhancement of the sensitivity 52-55 . Generally, particle migrations in laminar flow are changed when fluids flow in a curved, bent or twisted structures. Inside the structures, particles are forced from the centre to the outer side by centrifugal force or torsional force with a double symmetric vortex upward and downward [52] [53] [54] 56 . Figure 7 illustrates the generated vortex and how migrations of particles would happen in the RFF method. In the clockwise rotating region, the target fluid with biomolecules is forced according to the direction of rotation. In this case, the transverse vortex motion of biomolecules is expected as in a twisted-channel without passing through the interface of two fluids based on high surface tension and diffusion-limited transportation of microfluidic characteristics as shown in Fig. 7 . It could result in the active migration of biomolecules closer to the sensor surface. Therefore, Dean vortex is one explanation for the drastically enhanced sensitivity of the RFF method. However, further investigation should be carried out in the future to seek theoretical and experimental proof. We presented the RFF method for enhancing the sensitivity of label-free optical biosensors by adding a less-dense fluid horizontally. The hydrodynamically focused rate was controlled by changing the volumetric fraction ratio between the target fluid and the less-dense fluid. These phenomena were resulting in the enhancement of the binding. For the investigation, microfluidic devices were designed with the RFF method, and CFD simulations, flow experiments, and quantitative measurements were performed. The formation of RFF and behaviours of biomolecules inside a simple T-shaped microchannel were investigated in CFD simulations and fluorescent bead experiments. The CFD simulation confirms the hydrodynamic focusing of target fluids (3/10 of the channel height) when the volumetric flow rate ratio of less-dense fluid, ethanol, to the target fluid is fixed to 3:1 which agrees with theoretical expectation. For the quantitative characterisation of binding performances, optical measurements of silicon-based labelfree optical biosensors were carried out using biotin-streptavidin as a model antigen-antibody system with different flow methods. The results showed drastic improvement (8-order of magnitudes) in the sensitivity of the biotin-modified microring sensor in the RFF method compared to the SF method. For both methods, the www.nature.com/scientificreports/ conjugation tendency (between streptavidin and immobilised biotin) was logarithmically proportional to Streptavidin solutions' concentrations. The apparent dissociation constants obtained from the Langmuir models were to be 4.12 (± 0.53) × 10 -7 M for the SF method and 2.82 (± 3.12) × 10 -15 M for the RFF method. We conducted a series of experiments including bioactivity testing and comparison with the VFF method, which has the height reduction effect purely and verified the real effect of the RFF method for significant improvement in the sensitivity of the optical sensor. Also, a low concentration rate with 0.19 fM Streptavidin solution was tested with promising detection results, which could have been lowered further to check the lowest possible concentration limit. Along with the channel height reduction effect, the transverse motion of particles by the Dean vortex can explain one of the dominant impacts on the sensitivity enhancement. The RFF method offers a simple and effective way to enhance the sensitivity of the label-free optical sensor. It can be widely applied to any biosensor without requiring additional instruments. Among current diagnostic tests for COVID-19, nucleic acid testing (e.g., RT-PCR) is the most predominantly used methods 57, 58 . However, its time consuming and complicated process, including the preparation of viral RNA, the reverse transcription and amplification steps signify the need for the development of rapid and sensitive POC diagnostics for SARS-CoV-2, the viral aetiology of COVID-19 35 . Moreover, a recent study highlights the early-stage COVID-19 detection with protein-biomarker based testing 34 . The early-stage diagnosis can help reduce and slow-down the risk of viral transmission. Therefore, our RFF method, integrated with widely used microring resonators based on the antigen-antibody interaction, provides an approachable concept to develop essential tools for early detection of COVID-19 with shortening process time. It is a microfluidic-based sensor system which requires a minimal amount of sample. Although we used Streptavidin-biotin binding as a model case for antigen-antibody, further study for detection of COVID-19 using the RFF method with SARS-CoV-2 antigen should be investigated so that the suggested method can be quickly implemented in the current COVID-19 pandemic. Researchers and industries worldwide have been competing and collaborating to achieve a significant breakthrough to combat the COVID-19. Cutting-edge methods have advanced biosensing technologies to realise ultrasensitive, ultrafast, and early detection in pandemic diseases. Although most of the attempts are not practically viable, numerous research ideas have been published regarding COVID-19 keeping limitations and challenges; the research attempt will never be enough until we conquer the COVID-19 22 . Received: 25 September 2020; Accepted: 12 March 2021 gclid= CjwKC Ajwmf_ 4BRAB EiwAG hDfSc A37AN tr1Kp tMZwHf_ iYylH IzDg4 SPM-yinfx II5-MSBgg 8Lo7C iBoCj McQAvD_ BwE Epidemiological and clinical characteristics of 99 cases of 2019 novel coronavirus pneumonia in Wuhan, China: A descriptive study How the Coronavirus Escapes an Evolutionary Trade-Off That Helps Keep Other Pathogens in Check Data gaps and the Policy Response to the Novel Coronavirus Protecting health-care workers from subclinical coronavirus infection The Coronavirus Pandemic and Aerosols: Does COVID-19 Transmit Via Expiratory Particles? Theory, fabrication and applications of microfluidic and nanofluidic biosensors Optical biosensors for label-free detection of small molecules Highly-sensitive optical biosensor based on equal FSR cascaded microring resonator with intensity interrogation for detection of progesterone molecules Optofluidic microsystems for chemical and biological analysis Sensitivity comparison of surface plasmon resonance and plasmon-waveguide resonance biosensors 2012 Annual International Conference of the IEEE 523-526 Magnetic biosensors for genetic screening of cystic fibrosis Lectin-modified piezoelectric biosensors for bacteria recognition and quantification Sensitive optical biosensors for unlabeled targets: A review Enhanced sensitivity of subwavelength multibox waveguide microring resonator label-free biosensors Label-free biosensing using a microring resonator integrated with poly-(dimethylsiloxane) microfluidic channels Silicon-based integrated label-free optofluidic biosensors: latest advances and roadmap Measurement of biomarker proteins for point-of-care early detection and monitoring of cancer An isothermal, label-free, and rapid one-step RNA amplification/detection assay for diagnosis of respiratory viral infections Fast Accurate Point of Care Diagnosis of COVID-19 Pandemic Enabled Through Advanced Lab-on-Chip Optical Biosensors: Opportunities and Challenges Opportunities and challenges for biosensors and nanoscale analytical tools for pandemics: COVID-19 Dielectrophoresis for bioparticle manipulation Label-free attomolar detection of proteins using integrated nanoelectronic and electrokinetic devices Manipulation of nanoparticles and biomolecules by electric field and surface tension Diffusion coefficients of ethanol in aqueous solutions Microfluidics for Biotechnology Enhancement of affinity-based biosensors: effect of sensing chamber geometry on sensitivity Hydrodynamic focusing of conducting fluids for conductivity-based biosensors Numerical characterization and optimisation of the microfluidics for nanowire biosensors Novel poly(dimethylsiloxane) bonding strategy via room temperature "chemical gluing Why does the coronavirus spread so easily between people? Ultra-high-throughput clinical proteomics reveals classifiers of COVID-19 infection Protein based biomarkers for Non-invasive Covid-19 detection Rapid detection of COVID-19 causative virus (SARS-CoV-2) in human nasopharyngeal swab specimens using fieldeffect transistor-based biosensor Transient analysis of streptavidin-biotin complex detection using an IGZO thin film transistor-based biosensor integrated with a microfluidic channel Molecular-charge-contact-based ion-sensitive field-effect transistor sensor in microfluidic system for protein sensing High-throughput incubation and quantification of agglutination assays in a microfluidic system Microfluidic rheometer based on hydrodynamic focusing Fundamentals of Fluid Mechanics Effect of flow rates on generation of monodisperse clay-poly (N-isopropylacrylamide) embolic microspheres using hydrodynamic focusing microfluidic device An efficient 3-dimensional hydrodynamic focusing microfluidic device by means of locally increased aspect ratio Fabrication of microfluidic systems in poly (dimethylsiloxane) Label-free aptamer sensor based on silicon microring resonators Molecular recognition between genetically engineered streptavidin and surface-bound biotin Label-free biosensor based on an electrical tracing-assisted silicon microring resonator with a low-cost broadband source Label-free DNA sensor for detection of bladder cancer biomarkers in urine Entrapment of biomolecules in sol-gel matrix for applications in biosensors: Problems and future prospects An acoustic micromixer using low-powered voice coil actuation Optical biosensors based on silicon-on-insulator ring resonators: A review Label-free DNA sensor for detection of bladder cancer biomarkers in urine Dean vortices-induced enhancement of mass transfer through an interface separating two immiscible liquids Transport Phenomena in Micro Process Engineering Dean-coupled inertial migration and transient focusing of particles in a curved microscale pipe flow Optimization of microfluidic biosensor efficiency by means of fluid flow engineering Vortex-assisted DNA delivery Diagnosing COVID-19: The disease and tools for detection Laboratory diagnosis of COVID-19: Current issues and challenges /scientificreports/ conceived and supervised the project. N.K and K.H. are co-first authors. All authors contributed to writing the manuscript The authors declare no competing interests. Correspondence and requests for materials should be addressed to Y.-J.Y.Reprints and permissions information is available at www.nature.com/reprints.Publisher's note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http:// creat iveco mmons. org/ licen ses/ by/4. 0/.