key: cord-0931054-zfek4vhd authors: Mutalib, Sharah Abdul; Mace, Michael; Seager, Chloe; Burdet, Etienne; Mathiowetz, Virgil; Goldsmith, Nicola title: Modernising grip dynamometry: Inter-instrument reliability between GripAble and Jamar date: 2022-01-24 journal: BMC Musculoskelet Disord DOI: 10.1186/s12891-022-05026-0 sha: fc166f6ccf9551667ae42061d7b8cbfb6a87a441 doc_id: 931054 cord_uid: zfek4vhd INTRODUCTION: Maximum grip strength (MGS) is a reliable biomarker of overall health and physiological well-being. Therefore, an accurate and reliable measurement device is vital for ensuring the validity of the MGS assessment. This paper presents GripAble, a mobile hand grip device for the assessment of MGS. GripAble’s performance was evaluated using an inter-instrument reliability test against the widely used Jamar PLUS+ dynamometer. METHODS: MGS data from sixty-three participants (N = 63, median (IQR) age = 29.0 (29.5) years, 33 M/30 F) from both hands using GripAble and Jamar PLUS+ were collected and compared. Intraclass correlation (ICC), regression, and Bland and Altman analysis were performed to evaluate the inter-instrument reliability and relationship in MGS measurements between GripAble and Jamar PLUS+ . RESULTS: GripAble demonstrates good-to-excellent inter-instrument reliability to the Jamar PLUS+ with ICC(3,1) = 0.906 (95% CI [0.87—0.94]). GripAble’s MGS measurement is equivalent to 69% (95% CI [0.67—0.71]%) of Jamar PLUS+’s measurement. There is a proportional difference in mean MGS between the two devices, with the difference in MGS between GripAble and Jamar PLUS+ increasing with MGS. CONCLUSION: The GripAble is a reliable tool for measuring grip strength. However, the MGS readings from GripAble and Jamar PLUS+ should not be interchanged for serial measurements of the same patient, nor be translated directly from one device to the other. A new normative MGS data using GripAble will be collected and accessed through the software for immediate comparison to age and gender-matched subpopulations. Maximum grip strength (MGS) is a ubiquitous objective outcome measure for delineating hand functions, including the severity of upper limb impairment, improvement after hand surgery and functional progress after rehabilitation (or lack thereof ). A weak grip is associated with poor health-related quality of life (HRQoL) [1] , and individuals with impaired grip strength pose an increased risk of having a heart attack, stroke, and cognitive loss [2, 3] . An accurate and reliable instrument for measuring MGS is crucial for ensuring the accuracy of grip assessment and the validity of the resulting clinical interpretation. MGS is typically measured using a handgrip device equipped with a force or pressure sensor -a dynamometer. The Jamar hydraulic hand dynamometer [4] is the Open Access *Correspondence: sharahamo@imperial.ac.uk 1 GripAble Limited, London, UK Full list of author information is available at the end of the article gold standard device recommended by the American Society of Hand Therapists (ASHT) for measuring MGS [5] . Jamar is used in many studies for quantifying MGS, most notably in the well-known normative grip strength studies on adults and children by Mathiowetz et. al. [6, 7] , which remain widely used by clinicians worldwide to compare patients' grip strength. Although Jamar hydraulic is accurate, reliable and has good-to-excellent test-retest reliability and excellent interrater reliability, it is also deemed outdated by many clinicians, as it is heavy at 680 g, has questionable robustness, insensitive for measuring low forces and requires regular recalibration. Therefore, many updated devices emerged that demonstrate good-to-excellent inter-instrument reliability with Jamar. These devices include Baseline [8] , Rolyan [9] , Grippit [10] , MyoGrip [11] and Bodygrip [12] . Nonetheless, Jamar remains the gold standard, mainly due to heavy reliance and continued use of the aforementioned 1980s normative grip strength datasets. However, a recent study found that this normative data may no longer be valid, as grip strength has declined in recent decades. For example, the grip strength of Americans ages 20-34 has weakened dramatically relative to their 1980s counterparts [13] . This finding suggests that a new normative dataset is needed to consider the anthropometric and lifestyle changes in the past 40 years that may have altered the general population's grip strength. Whilst many previous studies used Jamar hydraulic to benchmark against new grip strength assessment devices, there is potential for significant reader error when using the Jamar hydraulic due to the dial only showing 2 kg increments [14] . Hogrel [11] also observed that Jamar hydraulic tends to overestimate grip force due to the inertial movement of the needle, which jumps slightly higher than the actual reading. The Jamar PLUS+ Digital is a digital analogue of the Jamar hydraulic which uses an electronic measurement principle based on a load cell force sensor instead of a hydraulic sensor [15] . The only other difference between the two Jamar devices is the weight (490 g vs 680 g) and small differences in external materials and finish. Despite being analogous, the interinstrument reliability between the two Jamar models has been shown to be poor-to-moderate, with Jamar hydraulic measuring approximately 10% higher than Jamar PLUS+ [16] . This suggests that many device-specific factors contribute to differences in grip strength measurement including technological, as well as, size, weight, and material characteristics. With an aim to modernise the current dynamometer, this paper presents the GripAble hand grip device -a new digital, accurate, sensitive and robust device for measuring grip strength. GripAble incorporates an electronic force measurement system with dual load cell sensors and connects wirelessly to a mobile device e.g. tablet, enabling users to perform objective MGS assessments. The addition of motion sensors enables the measurement of range-of-motion and device orientation allowing the user's compliance to a standardised protocol, such as hand posture during MGS to be monitored. According to Fess, the two most crucial criteria of any assessment device are (1) excellent measurement reliability across multiple sessions, examiners and devices and (2) high validity when compared against existing validated instruments. Fess further recommends (3) administrative instructions, (4) equipment criteria, (5) normative data, (6) instruction for interpretation and (7) a bibliography [17] . This paper aims to address the validity of GripAble by establishing the inter-instrument reliability compared to the Jamar PLUS+ . The Jamar PLUS+ was chosen due to the similarities in the underlying technology (i.e. digital force sensing) and the aforementioned limitations associated with the Jamar hydraulic device. We hypothesised that there would be an excellent interdevice reliability between Jamar PLUS+ and GripAble. However, larger diameter devices have been shown to yield lower grip force, as smaller intrinsic muscle groups will be recruited [18] . Therefore, considering the size difference between GripAble and Jamar PLUS+, we also hypothesised that there would be proportional differences in measurement output between the two devices, where the difference will be higher for individuals with higher MGS, with GripAble producing a small force output due to a relative increase in it's circumference. GripAble is a device that comprises a dual load cell forcesensing mechanism that enables the handgrip to deform elastically when squeezed. GripAble has a sensitivity of up to 60 g and is accurate up to 2 kg of force [19] . It is also equipped with an IMU (accelerometer, gyroscope and magnetometer) for measuring wrist movement and device orientation [20] . The grip plate of GripAble can be set to either elastic or isometric modes [21] using a simple 'locking' button. It connects wirelessly via Bluetooth to an Android device with a custom app, which will read and record data associated with the user's hand grip and movement whilst using GripAble. Two GripAble hand grip devices and two Jamar PLUS+ Digital Hand Dynamometers (referred to as Jamar+ hereinafter) were used for measuring MGS. Table 1 presents the feature comparison between Gri-pAble and Jamar+ . All GripAbles and Jamar+ s were calibrated a maximum of two months prior to the start date of the data collection process and were otherwise unused to ensure a level playing field. The Jamar+ s were set on position 2 (second smallest handle position) following ASHT standardised position for measuring grip strength [5] . The GripAbles were used in isometric mode to match Jamar+ . Sixty-three healthy participants (N = 63) aged between 16 -80 were recruited for the study. Each participant was assigned an alphanumeric ID and randomly split into four different experiment groups: GripAble right first (GR), GripAble left first (GL), Jamar+ right first (JR), and Jamar+ left first (JL). Participants were included if they had no diagnosed upper limb pathology, pain in the hand, wrist or forearm, or history of neurological disorder affecting the upper quadrant. All participants had a good comprehension of English to understand verbal instructions and experimentation documentation. Participants outside these criteria were excluded. All participants gave informed consent before the study. The experiment was performed following the ethical standards laid down in the 1964 Declaration of Helsinki and was approved by the Imperial College Research Ethics Committee (ICREC) and Science, Engineering & Technology Research Ethics Committee (SETREC). The study took place during the COVID-19 pandemic, i.e. August-December 2020. Therefore, all experiments were conducted outdoors, with a 2-metre distance between the investigator and the participant strictly followed as per NHS England COVID-19 guidelines. At the start of each session, the participant and investigator washed their hands thoroughly, and both devices were sanitised with isopropyl alcohol (IPA) wipes between participants. Each participant was positioned upright in a seated position on a chair with no armrests during the experiment. The investigator ensured that each participant maintained similar positioning at all times, which was standardised according to the ASHT guidelines [5] , including legs uncrossed, the bottom of the spine positioned against the back of the chair with hips and knees positioned at approximately 90°, arm adducted, elbow flexed at 90°, forearm in neutral and wrist in comfortable 15° to 30° extension and 0° to 15° of ulnar deviation (Fig. 1a) . Before starting, the investigator first demonstrated a single maximum grip test protocol. Verbal instruction was given, i.e. "Gradually put on the force. Now squeeze as hard as you can. 3, 2, 1, and relax" by the investigator during each measurement, which was also displayed on the GripAble app screen and only visible to the investigator. Three MGS measurements were recorded from each hand, using each device (i.e. twelve measurements in total). The hand was alternated between trials (e.g. Gri-pAble R-L-R-L-R-L, followed by Jamar+ R-L-R-L-R-L). A minimum 15-s rest was enforced between each measurement of the same hand and a minimum of three-minute rest between devices, as per ASHT guidelines. The measurement from the opposite hand was taken during the 15-s rest. Two investigators collected all measurements using the same sets of Jamar+ and GripAble devices throughout the study (i.e. one set for each investigator). All measurements taken using GripAble were calculated automatically by the GripAble app, and results were displayed to the investigator at the end of the test. The screen showing the GripAble app faced the investigators the entire time. Fig. 1b and 1c show screenshots of the GripAble app's test pages displaying the test page and result page, respectively. The mean MGS of the three trials from each hand and device was used for data analysis, as per ASHT guidelines. The mean MGS of all participants from both hands (N = 126) were then tested for normality using a Shapiro-Wilk test. Statistical significance was calculated at a 95% confidence level (p < 0.05). A zero-intercept linear fit using total least squares (TLS) was used to analyse the relationship between the measurements from GripAble and Jamar+ . Meanwhile, the inter-instrument reliability between Jamar+ and Gri-pAble was tested using intraclass correlation coefficient (ICC) assuming average fixed raters, i.e. ICC 3,1 , which measures the inter-instrument reliability in terms of 'consistency' [22] . ICC values of < 0.5, 0.5 -0.75, 0.75 -0.9, and > 0.90 indicate poor, moderate, good and excellent reliability, respectively. The ICC was calculated across all participants, as well as separately for each gender, starting device order and starting hand order, for each device and hand. A Bland and Altman plot was used to evaluate the MGS values measured using Jamar+ and GripAble. The plot visualises the differences in MGS ( D GJ ) between the two devices against their means ( µ GJ ). An ordinary least square (OLS) regression analysis was performed on this relationship to the proportionality of relationship between these two terms. The regression of D GJ on µ GJ can be described in the form of a y-intercept and a slope of the regression. A slope that is significantly different from zero (i.e. p < 0.05) suggests a proportional difference in MGS between Jamar+ and GripAble, where the difference increases (or decreases) as the mean increases. In this case, the limits of agreement are calculated using a regression approach for nonuniform differences [23, 24] . Statistical differences of participants' age between genders (male vs female), between starting device orders (GripAble first vs Jamar+ first) and between starting hand orders (right first vs left first) were calculated using a Mann-Whitney U test. Statistical differences in MGS between genders (male vs female) was calculated using a Mann-Whitney U test, whilst between hands (right vs left) and devices (GripAble vs Jamar+) were calculated using Wilcoxon signed-rank tests. Finally, the influence of starting device and starting hand orders on MGS was also analysed using a Mann-Whitney U test, i.e. GripAble first vs Jamar+ first, and right hand first vs left hand first. A Shapiro-Wilk test revealed a non-normal distribution of MGS measured using both GripAble and Jamar+ (p < 0.001 for both). Therefore, results are presented as median (interquartile range; IQR) and non-parametric tests were used for statistical analysis. Table 2 describes the demographic information of the participants (N = 63). Participants were predominantly self-reported right-handed (N R = 57; N L = 6). Mann-Whitney U tests revealed no significant difference in participants' age between genders (male vs female), between starting device orders (GripAble first vs Jamar+ first) and between starting hand orders (right first vs left first) ( p > 0.05 for all three comparisons performed). Figure 2a shows the mean MGS for each participant comparing Jamar+ (x-axis) to GripAble (y-axis). The right (square markers) and left (triangular markers) hands are shown as connected points. Age and gender are indicated by the marker size and colour, respectively. A zero-intercept linear fit using TLS revealed that GripAble's measurement output is equivalent to 69% (95% CI [68-71]%) of Jamar+ 's measurement output. Inter-instrument reliability was tested using intra-class correlation (ICC) assuming average fixed raters, i.e. ICC 3, 1 The overall ICC value was computed across all data collectively, and an overall ICC of 0.909 (95% CI [0.87-0.94]) indicates a good-to-excellent inter-instrument consistency between the two devices. The ICC was also calculated separately for each gender, starting hand order and starting device order. The full ICC results are shown in Table 3 . Table 4 summarises the descriptive data on the mean MGS out of three trials from all participants using both GripAble and Jamar+. Figure 3 visualises the mean MGS from each device and hand from the three trials. Wilcoxon signed-rank tests revealed significant differences between the overall MGS, as well as right hands' and left hands' MGS between Gri-pAble and Jamar+ (p < 0.001 for all three comparisons). Figure 4 visualises the influence of gender and hand on MGS. A Mann-Whitney U test revealed significant differences between males and females (p < 0.001 for both GripAble and Jamar+), whilst a Wilcoxon signed-rank test revealed significant differences between right and left hands (p < 0.001 for both GripAble and Jamar+). A Mann-Whitney U test revealed no significant differences in MGS between participants that started with GripAble first or Jamar+ first (p = 0.45 for GripAble measurements, p = 0.97 for Jamar+), or between participants that started with the right hand first or left hand first (p = 0.28 for GripAble measurements, p = 0.29 for Jamar+), suggesting that the starting device and starting hand orders did not influence the MGS measurements. The MGS measurements from sixty-three participants using GripAble and Jamar+ show that GripAble has an overall good-to-excellent inter-instrument reliability with Jamar+ , with an overall ICC of 0.909 (95% CI 0.87-0.94). GripAble's force output is equivalent to 69% (95% CI [0.67-0.71]%) of Jamar+ 's force output. Jamar+ tended to read higher MGS than GripAble. The difference was not fixed in absolute value but proportionally increased with higher MGS. The difference in readings between the two devices is not unique to this study. In a previous study by Hogrel, Jamar was 14% higher than MyoGrip, and the difference proportionally increased with higher MGS [11] . Although they attributed the difference to the inertial movement of Jamar's needle, we hypothesised that physical differences between GripAble and Jamar+ , including circumference and weight, play a more significant role. Specifically, a larger diameter device such as GripAble encourages flexion at metacarpophalangeal (MCP) joints and therefore recruits smaller intrinsic muscles. In contrast, the smaller Jamar+ encourages intrinsic minus grip pattern with the proximal interphalangeal joints predominantly active, increasing the force output for the same individual. Moreover, GripAble is also nearly half the weight of Jamar+ , which may have further contributed to the difference. Psychological factors could also play a part. Jamar+ is made of metal and plastic, whereas GripAble is made entirely of plastic. The material difference and the lower weight of GripAble may have given the impression of it being more fragile, causing the participants to be more reserved from exerting maximal grip force on the device. Therefore, verbal reinforcement may be needed for future grip assessment studies to encourage the user to squeeze maximally when using GripAble. Although Jamar is considered the gold-standard device for measuring MGS, it is outdated and needs updating to reflect evolving digital mobile health (mHealth) technology trends. With the recent global COVID-19 pandemic, the rehabilitation paradigm has shifted from inpatient to outpatient venues, including remote management using virtual platforms. Presently, clinicians who are managing their patients remotely need to rely on self-reported grip strength capabilities. Patients may benchmark against a change in their ability to carry out ADLs, which is not sensitive or objective. GripAble's custom mobile software platform provides an objective measurement of MGS completely remotely. Such a feature allows patients to regularly test their grip strength whilst allowing therapists to monitor and assess patients both accurately and reliably over time without their physical presence. Moreover, the integrated sensors within GripAble and software can provide researchers with additional information on the exact positioning of the device and hand postures during MGS assessment. For example, the ASHT guideline for measuring MGS recommended that the device be held vertically and stably. Using the data provided by the motion sensors within GripAble, it would be possible to explore user compliance to this guideline. Hand grip is vital for activities of daily living, and an extensive body of research has shown that it is a reliable biomarker of overall physical health and wellbeing, including predicting mortality and morbidity risks [25] . Nevertheless, the true impact of impaired grip strength on one's functional performance of ADLs is poorly understood. Most commonly, grip strength is only assessed as a single maximum effort, which only measures capability during one short instance. In functional activities, grip needs to be employed in many different ways, such as sustained over a more extended period or repeatedly applied. Gri-pAble provides a range of grip tests, including but not limited to grip endurance, sustained gripping, grip in various forearm rotational positions [26] , rapid exchange [27] and sine wave grip accuracy tests [28] , to allow for a more holistic view of hand function. Ultimately, this opens the opportunity for a deeper understanding of the association between grip and functional performance, where the multiple aspects of grip may uncover such associations. This study provides evidence of GripAble's good-toexcellent inter-instrument reliability and consistency to the Jamar+ . There was a proportional bias between the two devices, where the differences in MGS readings between the two devices increased as their means increased. Therefore, whilst measurements from both devices can be considered valid, the MGS readings from these two devices should not be interchanged for serial measurements of the same patient, such as when comparing pre and post-intervention, nor be translated directly from one device to the other. GripAble can be used clinically as a dynamometer whilst providing additional software benefits, including the display of postural and procedural instructions, standardised narrative and automated result computation and analysis. Furthermore, integrating a dynamometer into an accessible multifunctional training device gives the potential for objective assessment both in the clinic and remotely for home users. Further studies investigating the other facets of grips, as well as test-retest and inter-rater reliability are needed. Updated devicespecific normative values for age and gender subsets are also needed and, once available, can be integrated into the software for immediate comparison. Is grip strength associated with health-related quality of life? Findings from the Hertfordshire Cohort Study Grip strength and the risk of cognitive decline and dementia: a systematic review and meta-analysis of longitudinal cohort studies Prognostic value of grip strength: Findings from the Prospective Urban Rural Epidemiology (PURE) study. The Lancet Grip test; the use of a dynamometer with adjustable handle spacings American Society of Hand Therapists; Clinical assessment recommendations 3rd edition: Impairment-based conditions Grip and pinch strength: normative data for adults Grip and Pinch Strength: Norms for 6-to 19-Year-Olds Comparison of Baseline instruments to the Jamar dynamometer and the B and L Engineering pinch gauge Comparison of Rolyan and Jamar dynamometers for measuring grip strength A comparative study of the Jamar ® and the Grippit ® for measuring handgrip strength in clinical practice Grip strength measured by high precision dynamometry in healthy subjects from 5 to 80 years Comparison of Jamar and Bodygrip Dynamometers for Handgrip Strength Measurement Comparative study of millennials' (age 20-34 years) grip and lateral pinch with the norms A review of the measurement of grip strength in clinical and epidemiological studies: towards a standardised approach Interinstrument reliability of the jamar electronic dynamometer and pinch gauge compared with the jamar hydraulic dynamometer and B&L engineering mechanical pinch gauge The need for reliability and validity in hand assessment instruments Grip Force Vectors for Varying Handle Diameters and Hand Sizes GripAble: an accurate, sensitive and robust digital device for measuring grip strength Democratizing Neurorehabilitation: How Accessible are Low-Cost Mobile-Gaming Technologies for Self-Rehabilitation of Arm Disability in Stroke? Elasticity improves handgrip performance and user experience during visuomotor control A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research Measuring agreement in method comparison studies Confidence in Altman-Bland plots: A critical review of the method of differences Grip strength, body composition, and mortality • fast, convenient online submission • thorough peer review by experienced researchers in your field • rapid publication on acceptance • support for research data, including large and complex data types • gold Open Access which fosters wider collaboration and increased citations maximum visibility for your research: over 100M website views per year your research ? Choose BMC The rapid exchange grip strength test and the detection of submaximal grip effort Motor dexterity and strength depend upon integrity of the attention-control system Publisher's Note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations The authors would like to thank all participants for taking part in the study Authors' contributions SAM wrote the manuscript. SAM and NG designed the study protocol. CS and NG collected the data. SAM, MM and NG contributed equally to analysis plan and result interpretation, with SAM and MM leading data (including MGS: Maximum grip strength; ICC: Intraclass correlation coefficient; IQR: Interquartile range; ADL: Activities of daily living; ASHT: American Society of Hand Therapists; IMU: Inertial measurement units.